The scope of this paper is to illustrate the need for an improved quality assurance in fluorometry. For this purpose, instrumental sources of error and their influences on the reliability and comparability of fluorescence data are highlighted for frequently used photoluminescence techniques ranging from conventional macro- and microfluorometry over fluorescence microscopy and flow cytometry to microarray technology as well as in vivo fluorescence imaging. Particularly, the need for and requirements on fluorescence standards for the characterization and performance validation of fluorescence instruments, to enhance the comparability of fluorescence data, and to enable quantitative fluorescence analysis are discussed. Special emphasis is dedicated to spectral fluorescence standards and fluorescence intensity standards.
This report focuses on the characterization of CD4 expression level in terms of equivalent number of reference fluorophores (ERF). Twelve different flow cytometer platforms across sixteen laboratories were utilized in this study. As a first step the participants were asked to calibrate the fluorescein isothiocyanate (FITC) channel of each flow cytometer using commercially available calibration standard consisting of five populations of microspheres. Each population had an assigned value of equivalent fluorescein fluorophores (EFF denotes a special case of the generic term ERF with FITC as the reference fluorophore). The EFF values were assigned at the National Institute of Standards and Technology (NIST). A surface-labelled lyophilized cell preparation was provided by the National Institute of Biological Standards and Control (NIBSC), using human peripheral blood mononuclear cells (PBMC) pre-labeled with a FITC conjugated anti-CD4 monoclonal antibody. Three PBMC sample vials, provided to each participant, were used for the CD4 expression analysis. The PBMC are purported to have a fixed number of surface CD4 receptors. On the basis of the microsphere calibration, the EFF value of the PBMC samples was measured to characterize the population average CD4 expression level of the PBMC preparations. Both the results of data analysis performed by each participant and the results of centralized analysis of all participants' raw data are reported. Centralized analysis gave a mean EFF value of 22,300 and an uncertainty of 750, corresponding to 3.3% (level of confidence 68%) of the mean EFF value. The next step will entail the measurement of the ERF values of the lyophilized PBMC stained with labels for other fluorescence channels. The ultimate goal is to show that lyophilized PBMC is a suitable biological reference cell material for multicolor flow cytometry and that it can be used to present multicolor flow cytometry measurements in terms of ABC (antibodies bound per cell) units.