Coulometric sensors were tested in humidified synthetic air at various gas temperatures. Generated frost point temperature in the gas ranged from -30 °C to -60 °C and were measured by coulometric sensors and in addition by a calibrated dew point hygrometer. The gas temperatures, which were measured by a calibrated Pt100 sensor, were set to -20 °C, 0 °C, 23 °C, 40 °C, 50 °C, and 60 °C during the experiments. Empiric nonlinear functions were calculated between the humidity and the sensor signal. In comparison to the measured signals at 23 °C, the sensor signals were lower at the other gas temperatures. Measurements at 60 °C showed indistinct results due to a great signal noise. The response behavior of the sensors was similar at 23 °C, 40 °C and 50 °C. In contrast to that, the sensors reacted slowly at a gas temperature of -20 °C and 0 °C. In summary, with coulometric sensors it was possible to measure continuously trace humidity with an expanded uncertainty below 2 K.
This work presents a first attempt to use ionic liquids as a new coating for planar coulometric sensors. These sensors are used for the measurement of trace humidity in various gases. Usually, the coating of the sensors is tetraphosphorus decaoxide and its hydrolysis products. Instead, a hygroscopic ionic liquid was used as sensor coating in this work. Generated frost point temperatures tf in the gas ranged from -80 °C to -30 °C, which is equivalent to vapour mole fractions xv from 0.5 μmol·mol-1 to 376 μmol·mol-1. In addition to the coulometric sensors, the generated humidity is determined by a precision dew point hygrometer as reference. First results show that it is possible to measure humidity withionic liquid based coulometric sensors.
Planar coulometric sensors were investigated in humidified synthetic air at various absolute gas pressures, i. e. 2 bar, 5, bar, 10 bar, and 15 bar. Humidified gas flow at adjusted gas pressure was split into two flows, one passed a coulometric sensor and the other one passed a reference hygrometer after decompression. Both signals were recorded and then compared after calculation of resulting frost point temperature. Calculation is based on a calibration function obtained at ambient pressure. Comparison showed that an increasing pressure resulted in a higher derivation between sensor signal (calculated frost point temperature) and reference frost point temperature. At an absolute pressure of 2 bar the differences were minor in consideration of the uncertainty, however at 15 bar the differences were 6.77 K. Nevertheless, it was possible to measure the gas humidity at higher pressure with coulometric trace humidity sensors.
Die Spurenfeuchte ist eine wichtige Messgröße bei der Qualitätsbeurteilung von technischen und medizinischen Gasen und muss daher zuverlässig gemessen werden. Der Spurenbereich ist definiert durch einen Wasserdampf-Stoffmengenanteil kleiner als 2600 µmol/mol bzw. einer Frostpunkttemperatur kleiner -10 °C. Eine geeignete Methode für die kontinuierliche Spurenfeuchtemessung stellt das Elektrolyseverfahren bzw. coulometrische Messprinzip dar. Bei diesem Prinzip strömt ein feuchtes Gas über eine hygroskopische Phosphorpentoxid-Schicht, welche kontinuierlich Wasserdampf absorbiert. Durch das Anlegen einer Zersetzungsspannung wird das absorbierte Wasser elektrolysiert und der resultierende Elektrolysestrom korreliert zum Feuchtegehalt im Gas. Neben dem Einfluss von unterschiedlichen Gasen auf das Messverfahren, wurde der Einfluss der Druckfeuchte im Druckbereich von 2 bis 15 bar untersucht. Des Weiteren wurde eine neuartige Sensorbeschichtung basierend auf einer ionischen Flüssigkeit getestet und es wurde ein Prototyp für die abgestufte Bestimmung der Wasseraktivität und der Materialfeuchte aufgebaut.
Die Messung der Gasfeuchte im Spurenbereich in unterschiedlichen technisch und
medizinisch relevanten Gasen erfolgte mit Sensoren, die auf dem coulometrischen
Messprinzip basieren. Die verwendeten planaren coulometrischen Sensoren wurden im
Frostpunkttemperaturbereich von -70 °C bis -30 °C in synthetischer Luft, Wasserstoff,
Stickstoff und Distickstoffmonoxid (Lachgas) untersucht. Als Referenz für die Messungen
wurde ein Präzisionstaupunktspiegel-Hygrometer verwendet. Die Gasfeuchte kann mit Hilfe
der Sensoren in allen untersuchten Gasen, aber mit geänderten Kalibrierfunktionen, mit einer
erweiterten Unsicherheit von kleiner als 2,1 K bestimmt werden. Für die Messungen in
unterschiedlichen Gasen muss also der Einfluss der Gasmatrix auf das Sensorsignal
berücksichtigt werden. So ist zum Beispiel bei Messungen in befeuchtetem Wasserstoff,
aufgrund von Rekombinationseffekten, ein signifikant geändertes Sensorsignal zu
beobachten.
Coulometric sensors are applied for trace humidity measurements in various technical gases. The use of this sensor type is demanded in some standards by the European Pharmacopoeia for medical gases. Coulometric sensors allow the measurement of water vapour in gases such as e.g. air, Cl2, H2, N2, N2O, CH4 between the concentration ranging from 0.1 to 2,500 µmol∙mol-1 which corresponds to frost point temperature -90 °C to -10 °C, respectively.
The sensing principle is based on Faraday’s law of electrolysis whereby water is decomposed to hydrogen and oxygen. The sensor signal is the measured electrical current which is proportional to the mass of water that is absorbed on the hygroscopic phosphorous pentoxide layer. The signal is dependent on the gas flow at a given voltage, gas pressure and temperature.
The sensors need to be calibrated to measure the accurate signal for humidity in air. However, the signal is dependent on the type of gas matrix. This dependency has not been quantified so far. Therefore, the impact of reactive gases such as hydrogen and nitrous oxide on the calibration curve was investigated. Furthermore, a possible rationale in relation to the interaction of water with the gas matrix and its impact on the electrode reactions is suggested.
The experimental setup consists of a gas supply, dryer, humidifier, test chamber and reference hygrometer. The test gas is generated by mixing the dry and the wet gas flow. First the carrier gas is dried by an activated carbon filter and then split into two flows. One flow is dried again with a molecular sieve. The other flow is humidified by passing it through a bubbler filled with pure water. After this, the rate of both flows is controlled by mass flow controllers and then mixed with the test gas. A calibrated precision chilled mirror dew-point hygrometer is used as a reference instrument.
The coulometric sensors were exposed to different humidified gases and the sensor signal was recorded till a constant value was obtained. Calibration curves were calculated for the frost point temperature in the range of about -70 °C to -10 °C according to the equation, tf = A + B∙ln(I), (I is the electrolysis current and A, B are constants) followed with a linear regression fit.
Comparison of the results of air and nitrogen showed no significant differences. In contrast, there are remarkable differences for humidified hydrogen and nitrous oxide, respectively.
The difference might be due to increased recombination of hydrogen with the produced oxygen to form new water molecules in humidified hydrogen.
In conclusion, coulometric trace humidity sensor is a robust hygrometer for various technical applications. However it needs to be calibrated for the specific gas matrix.
Feuchte, sei es Material- oder Gasfeuchte, ist eine wichtige Messgröße bei der Qualitätsbeurteilung von Kunststoffen, landwirtschaftlichen Erzeugnissen, Energieträgern, Arzneimitteln, industriell und medizinisch verwendeten Gasen. Deswegen gibt es ein Interesse Feuchtemesserfahren hinsichtlich Präzision, Wiederholbarkeit, Rückführbarkeit und Stabilität kontinuierlich zu verbessern. Ein bewährtes Messprinzip für diese Aufgabe wurde bereits 1959 von Keidel entwickelt und basiert auf der Absorption und Elektrolyse von Wasserdampf. Der einfache Aufbau dieses Prinzips besteht aus einem Sensorelement, einer Gleichspannungsquelle, einem Digitalmultimeter und einen geregelten Gasstrom über den Sensor. Nach dem Faraday’schen Gesetz der Elektrolyse korreliert bei dem Messprinzip die Ladungsmenge mit der elektrolysierten Wassermasse. Jedoch bedarf es in der heutigen Zeit einer Validierung der Sensoren, weil durch gezielte Miniaturisierung weniger aktive Fläche vorhanden ist und somit das Faraday’sche Gesetz nicht vollständig anwendbar ist. In dieser Arbeit wurden coulometrische Sensoren mit einer planaren Elektrodenstruktur hinsichtlich der Einflüsse von unterschiedlichen Gasen, der Gastemperatur und dem -druck untersucht. Zusätzlich wurde eine neuartige Sensorbeschichtung basierend auf einer ionischen Flüssigkeit getestet. Des Weiteren wurde ein Messgerät für die abgestufte Bestimmung der Materialfeuchte und Wasseraktivität entwickelt und dessen messtechnischer Einsatz untersucht.
Ionische Flüssigkeiten als neuartige Sensorbeschichtung für coulometrische Spurenfeuchtesensoren
(2018)
Die Arbeit präsentiert erste Messungen mit coulometrischen Sensoren, die mit einer ionischen Flüssigkeit beschichtet sind. Diese Sensoren werden in der Industrie verwendet, um die Spurenfeuchte in unterschiedlichen Gasen zu messen. Normalerweise erfolgt eine Beschichtung der Sensoren mit Phosphorpentoxid und dessen Hydrolyseprodukte. In dieser Arbeit wurde stattdessen eine hygroskopische ionische Flüssigkeit benutzt. Die generierte Gasfeuchte für die Untersuchungen betrug bezogen auf die Frostpunkttemperatur -80 °C bis -30 °C, was gleichbedeutend mit einem Wasserdampf-Stoffmengenanteil von 0,5 μmol·mol-1 bis 376 μmol·mol-1 ist. Zusätzlich zu der Messung mit coulometrischen Sensoren erfolgte eine Referenzmessung der Gasfeuchte mittels eines Taupunkthygrometers. Erste Ergebnisse haben gezeigt, dass es möglich ist mit dieser Art der Beschichtung unterschiedliche Gasfeuchten zu messen. Aber es ergaben sich trotz gleicher Feuchte unterschiedliche Signalintensitäten bei den benutzen Sensoren und es zeigte sich eine Drift des Sensorsignales.
Die Messung der Gasfeuchte im Spurenbereich in unterschiedlichen technisch und medizinisch relevanten Gasen erfolgte mit Sensoren, die auf dem coulometrischen Messprinzip basieren. Die verwendeten planaren coulometrischen Sensoren wurden im Frostpunkttemperaturbereich von -70 °C bis -30 °C in synthetischer Luft, Wasserstoff, Stickstoff und Distickstoffmonoxid (Lachgas) untersucht. Als Referenz für die Messungen wurde ein Präzisionstaupunktspiegel-Hygrometer verwendet. Die Gasfeuchte kann mit Hilfe der Sensoren in allen untersuchten Gasen, aber mit geänderten Kalibrierfunktionen, mit einer erweiterten Unsicherheit von kleiner als 2,1 K bestimmt werden. Für die Messungen in unterschiedlichen Gasen muss also der Einfluss der Gasmatrix auf das Sensorsignal berücksichtigt werden. So ist zum Beispiel bei Messungen in befeuchtetem Wasserstoff, aufgrund von Rekombinationseffekten, ein signifikant geändertes Sensorsignal zu beobachten.
Planar coulometric sensors were tested in humidified synthetic air at various gas temperatures. Generated frost point temperature in the gas ranged from -30 °C to -60 °C and were measured by coulometric sensors and in addition by a calibrated dew point hygrometer. The gas temperatures, which were measured by a calibrated Pt100 sensor, were set to -20 °C, 0 °C, 23 °C, 40 °C, 50 °C, and 60 °C during the experiments. Empiric nonlinear functions were calculated between the generated humidity and the sensor signals. In comparison to the measured signals at 23 °C, the sensor signals were lower at the other gas temperatures. The measurements at 60 °C showed indistinct results due to a great signal noise. The response behavior of the sensors was similar at 23 °C, 40 °C and 50 °C. In contrast to that, the sensors reacted slowly at a gas temperature of -20 °C and 0 °C. In summary, with coulometric sensors it was possible to measure continuously trace humidity with an expanded uncertainty below 2 K.