The role of matrix reference materials in the process of demonstrating the degree of equivalence of measurement results obtained from intercomparisons is outlined, reviewing exemplary selected experience gained at BAM regarding the determination of organic contaminants in environmental matrices.
The specific characteristics of reference materials employed in the process of demonstrating equivalence between laboratories in the course of proficiency testing as well as the development, comparison and validation of methods are elaborated. The demand is for series of appropriately characterised samples which are fit for the purpose and it is seen from representative examples that the utilisation of such tailor-made RM designed to tackle the specific need of an analytical problem dominates over certified matrix reference materials in this context. Concluding, the role of certified matrix reference materials in quality assurance is briefly looked at both from the users and providers points of view.
The interlaboratory variability in the quantification of adsorbable organically bound halogens (AOX) in industrially contaminated soil is presented. Three consecutive rounds of a proficiency testing scheme, in which between 88 and 119 routine laboratories participated, yielded relative reproducibility standard deviations between 7 and 20% at AOX contents between 10.9 and 268 mg kg-1. Nineteen laboratories with established proficiency were invited to participate in the certification of the AOX content in three soil reference materials meant for the internal quality control in analytical laboratories. The certified values are (1349 ± 59) mg kg-1, (80 ± 7) mg kg-1 and (102 ± 8) mg kg-1, respectively.