The measurement characteristics of every conventional accelerometer are temperature dependent - often to an extent which is highly relevant for the addressed application. For instance, at BAM Container drop tests are performed to investigate and evaluate the structural integrity of Containers for transport and storage of dangerous goods. Often, the test program includes drop tests at specific temperatures, also as part of approval procedures. The applied sensors are exposed to these test conditions and get influenced by them. Acceleration sensors show a considerable temperature influence on their fünction and characteristics. Particularly the damping mechanism of the seismic mass is a critical part. In regard to such applications, capable equipment and methods are required to consider and investigate this aspect in an adequate way. A fortiori as Manufacturer’s infonnation often is deficient. This study presents the setup and results of a method for testing and calibration of acceleration sensors under high dynamic irnpact. It combines a Hopkinson Bar with a temperature chamber.
Multichannel measuring systems are used to measure strains and accelerations during drop tests of containments for dangerous goods. Conventional systems require cabling of each sensor and co-falling of the cable harness, causing problems in the test preparation and execution. Promising results of a single cable measuring system, consisting of measuring modules with data bus connection and local data acquisition were obtained in laboratory investigations and full-scale drop tests.
One of the main unsolved issues of under-ground storages for, e.g., CO2, H2, and natural gas is the comprehensive surveillance of these areas with reasonable effort and costs. Conventional sensors (e.g., soil air probes or borehole probes), however, can only be used for punctual or locally limited measurements; further their application can cause structural influences (invasive application).
In this paper, we describe in detail the setup of a CO2 injection soil test field. This test field will be used to enhance and validate an innovative ap-proach for distributed subsurface monitoring of gas storage areas. To the author’s knowledge, this is the first time that, for this purpose, a test field is built in an application relevant scale.