Planar coulometric sensors were tested in humidified synthetic air at various gas temperatures. Generated frost point temperature in the gas ranged from -30 °C to -60 °C and were measured by coulometric sensors and in addition by a calibrated dew point hygrometer. The gas temperatures, which were measured by a calibrated Pt100 sensor, were set to -20 °C, 0 °C, 23 °C, 40 °C, 50 °C, and 60 °C during the experiments. Empiric nonlinear functions were calculated between the generated humidity and the sensor signals. In comparison to the measured signals at 23 °C, the sensor signals were lower at the other gas temperatures. The measurements at 60 °C showed indistinct results due to a great signal noise. The response behavior of the sensors was similar at 23 °C, 40 °C and 50 °C. In contrast to that, the sensors reacted slowly at a gas temperature of -20 °C and 0 °C. In summary, with coulometric sensors it was possible to measure continuously trace humidity with an expanded uncertainty below 2 K.
This work presents a first attempt to use ionic liquids as a new coating for planar coulometric sensors. These sensors are used for the measurement of trace humidity in various gases. Usually, the coating of the sensors is tetraphosphorus decaoxide and its hydrolysis products. Instead, a hygroscopic ionic liquid was used as sensor coating in this work. Generated frost point temperatures tf in the gas ranged from -80 °C to -30 °C, which is equivalent to vapour mole fractions xv from 0.5 μmol·mol-1 to 376 μmol·mol-1. In addition to the coulometric sensors, the generated humidity is determined by a precision dew point hygrometer as reference. First results show that it is possible to measure humidity with ionic liquid based coulometric sensors.
Planar coulometric sensors were investigated in humidified synthetic air at various absolute gas pressures, i. e. 2 bar, 5, bar, 10 bar, and 15 bar. Humidified gas flow at adjusted gas pressure was split into two flows, one passed a coulometric sensor and the other one passed a reference hygrometer after decompression. Both signals were recorded and then compared after calculation of resulting frost point temperature. Calculation is based on a calibration function obtained at ambient pressure. Comparison showed that an increasing pressure resulted in a higher derivation between sensor signal (calculated frost point temperature) and reference frost point temperature. At an absolute pressure of 2 bar the differences were minor in consideration of the uncertainty, however at 15 bar the differences were 6.77 K. Nevertheless, it was possible to measure the gas humidity at higher pressure with coulometric trace humidity sensors.
Trace humidity was measured by using miniaturized planar coulometric sensors in technical gases such as hydrogen, nitrogen, helium, nitrous oxide, and synthetic air. Frost point temperatures tf in the gases ranged from −60 °C to −30 °C, which is equivalent to a vapour mole fraction xv from 10 μmol mol−1 to 376 μmol mol−1. In addition, the generated humidity was determined by using a precision dew point hygrometer as reference. Nonlinear calibration functions were calculated that correlated electric current (sensor signal) and reference humidity. Parameters of functions were tested with one-way analysis of variances (ANOVA) to prove if all used sensors had a similar behavior in the same gas during experiments. Results of ANOVA confirmed that averaged functions can be applied for trace humidity measurement in nitrogen, helium, nitrous oxide, and synthetic air. The calculated functions were negligibly different for nitrogen, helium, and synthetic air. In humidified nitrous oxide, a minor change of parameters was observed due to lower electrical currents. In total contrast to that, the measured sensor signals were significantly higher in humidified hydrogen and each sensor required its own calibration function. The reason was a recombination effect that favoured multiple measurements of water molecules. Nevertheless, it was possible to measure continuously trace humidity in all tested gases by using coulometric sensors with an expanded uncertainty below 2 K (k = 2).