A Round Robin study has been carried out to estimate the impact of the human element in small-angle scattering data analysis. Four corrected datasets were provided to participants ready for analysis. All datasets were measured on samples containing spherical scatterers, with two datasets in dilute dispersions, and two from powders.
Most of the 46 participants correctly identified the number of populations in the dilute dispersions, with half of the population mean entries within 1.5 % and half of the population width entries within 40 %, respectively. Due to the added complexity of the structure factor, much fewer people submitted answers on the powder datasets.
For those that did, half of the entries for the means and widths were within 44 % and 86 % respectively. This Round Robin experiment highlights several causes for the discrepancies, for which solutions are proposed.
Continuing progress in the field of X-ray scattering methods empowers scientists with new possibilities to capture the most important piece of information about the structure of the sample - its 3D electron density.
Although the first methods appeared almost a century ago, recovering the density structure of a sample is still very problematic. Most avail-able imaging techniques transform a 3D electron density of a realspace structure into the 2D Fourier Transform of the intensity of scattered waves in the reciprocal space. This process causes a loss of information.
Firstly, instead of a 3D sample, a 2D image is created, and secondly, the phase information of the scattered waves is lost. The latter is known as the ”phase problem” and poses a serious obstacle on a way to recover a 3D electron density. In this work, we draw attention to the problem of forward and inverse Small Angle X-Ray Scattering. In the first, forward, part, we rethink the existing pipelines to computationally simulate such scattering experiments. Although there are efficient implementations of fast Fourier transformation, they often have some drawbacks. For instance, to calculate a 3D fast Fourier transform it is required to place its density in the RAM. For high-resolution structures of size > 1024 3 , this becomes very problematic, as the whole density structure requires more than 16 GB of memory.
CUDA solution allows for a very fast and parallelizable implementation of high-resolution data on hundreds of last-generation machines.
Such computations are very pricy and inaccessible for most scientists.
To bypass this limitation, we propose a solution for a split-up 3D fast Fourier transform, which is implemented as a sequence of 2D and 1D operations. We compare our implementation on the simulated 3D shapes and show the result of a proof-of-concept on 4096 3 Metallorganic framework density structure. In the second, inverse problem, we train an invertible neural network, that given scattering data can predict the shape and its parameters. The architecture is built such, that the inverse problem is learned together with the forward process - the Fourier Transformation. We achieved very good results with this architecture, nonetheless, further testing is required, as the current training set only encompasses three simple shapes: sphere, hard sphere and cylinder.
All code to reproduce and analyze the results is available at https: //github.com/sofyalaski/SAXS-simulations.
Nanomaterials and nanoparticles are finding applications across a wide range of technology sectors, from medicine and food to transportation and construction. In order to assess these new materials for potential risks to health and the environment, they need to be well-characterised. The measurement of constituent nanoparticle size, shape, and size distribution are important factors for the risk evaluation process.
EMPIR project Improved traceability chain of nanoparticle size measurements (17NRM04, nPSize) is working to assess a range of traceable nanoparticle measurement approaches, including Scanning Electron Microscopy (also in Transmission Mode), Atomic Force Microscopy and Small Angle X-ray Scattering, and deliver improved calibration methods to users. For the techniques under investigation, physical models of their response to a range of nanoparticle types are developed. Validated reference materials will also be used for an inter-comparison of measurement systems, with an evaluation of the associated measurement uncertainty. With project contributions to standards development work, manufacturers will be better placed to assess the human and environmental risks posed by nanomaterials across a whole range of products.