A big part of the infrastructure is made of concrete, which is a multiphase system made of cement, aggregates, water and pores. Concrete is often used in combination with steel as reinforced concrete. Environmental influences, especially the ingress of harmful ions in combination with the ingress of water, trigger different damage processes which reduce the designed lifetime of a structure. The ingress of chlorides from de-icing salt or sea water leads to corrosion of the reinforcement. In addition, the carbonation of the concrete may trigger the corrosion of the reinforcement. The ingress of alkalis from deicing salts may cause the expansion of the amorphous silica aggregates (alkali-silica reaction) through
formation of a swelling gel of calcium silicate hydrate if water is present. The ingress of sulphates may cause spalling of the concrete surface due to ettringite formation.
For the standard procedure in civil engineering cores are taken, cut in slices, grinded and the obtained homogenized powder is solved in acid and investigated by standard procedures.
BAM has developed the LIBS technique for the 2D evaluation of the chemical composition of concrete. The technique is established for automated laboratory use with high numbers of samples to investigate transport processes of harmful species (Cl-, CO2 , SO4 2-and alkalis) in concrete. Information about ingress depth and the quantitative values are important to estimate the remaining lifetime of the infrastructure. LIBS is a surface technique. To get information about the ingress depth, a drill core has to be taken and cut in the middle. The measurements are carried out at the cross section. The main advantages of LIBS are the direct measurement on the surface of the concrete, fast analysis (sample rate 100 Hz) with a spatial resolution of up to 100 µm, the consideration of the heterogeneity of the concrete and the possibility of automated measurements which save a lot of manpower and time. As an example the investigation of ingress profiles for standard diffusion and migration tests in civil engineering takes hours in comparison to just a few minutes using LIBS. At the same time a 2D-evaluation provides information about hot spots of elemental concentration which may not be found by standard methods.
The state of the art of LIBS technique for applications in civil engineering will be presented, including typical results of 2D investigation of concrete in laboratory. The performance is also demonstrated by examples for onsite applications using a mobile LIBS system. The road map to standardization is presented as well.
Concrete is often used in combination with steel as reinforced concrete. Environmental influences, especially the ingress of harmful ions in combination with the ingress of water, trigger different damage processes which reduce the designed lifetime of a structure. The ingress of chlorides from de-icing salt or sea water leads to corrosion of the reinforcement. Also the carbonation of the concrete may trigger the corrosion of the reinforcement. The ingress of alkalis from de-icing salts may cause the expansion of the amorphous silica aggregates (alkali-silica reaction) through formation of a swelling gel of calcium silicate hydrate if water is present. The ingress of sulfates may cause spalling of the concrete surface due to ettringite formation.
BAM has developed the LIBS technique for automated laboratory use with high numbers of samples to investigate transport processes of harmful species (Cl-, CO2, SO42- and alkalis) in concrete. Information about ingress depth and the quantitative values are important to estimate the remaining lifetime of the infrastructure. To get information about the ingress depth, a core has to be taken and cut in the middle. The measurements are carried out at the cross section. The main advantages of LIBS are the direct measurement on the surface of the concrete, fast analysis (sample rate 100 Hz) with a spatial resolution of up to 100 µm, the consideration of the heterogeneity of the concrete. The possibility of automated measurements saves a lot of manpower and time. At the same time a 2D-evaluation provides information about hot spots of elemental concentration which may not be found by standard methods.
Typical results of 2D investigation of concrete in laboratory will be presented. The performance is also demonstrated by examples for onsite applications using a mobile LIBS system. The road map to standardization is presented as well.
The majority of the built infrastructure is made of concrete, which is a multiphase system made of cement, aggregates, water and pores. Concrete is often used in combination with steel as reinforced concrete. Environmental influences, especially the ingress of harmful ions in combination with the ingress of water, trigger different damage processes which reduce the designed lifetime of a structure. The ingress of chlorides from de-icing salt or sea water leads to corrosion of the reinforcement. Also the carbonation of the concrete may trigger the corrosion of the reinforcement. The ingress of alkalis from de-icing salts may cause the expansion of the amorphous silica aggregates (alkali-silica reaction) through formation of a swelling gel of calcium silicate hydrate if water is present. The ingress of sulfates may cause spalling of the concrete surface due to ettringite formation.
For the standard procedure in civil engineering cores are taken, cut in slices, grinded and the obtained homogenized powder is solved in acid and investigated by standard procedures.
BAM has developed the LIBS technique for the 2D evaluation of the chemical composition of concrete [1-11]. The technique is established for automated laboratory use with high numbers of samples to investigate transport processes of harmful species (Cl-, CO2, SO42- and alkalis) in concrete. Information about ingress depth and the quantitative values are important to estimate the remaining lifetime of the infrastructure. LIBS is a surface technique. To get information about the ingress depth, a core has to be taken and cut in the middle. The measurements are carried out at the cross section. The main advantages of LIBS are the direct measurement on the surface of the concrete, fast analysis (sample rate 100 Hz) with a spatial resolution of up to 100 µm, the consideration of the heterogeneity of the concrete and the possibility of automated measurements which save a lot of manpower and time. As an example the investigation of ingress profiles for standard diffusion and migration tests in civil engineering takes hours in comparison to just a few minutes using LIBS. At the same time a 2D-evaluation provides information about hot spots of elemental concentration which may not be found by standard methods.
Ingress of chlorides due to a crack in a repair mortar. Left: Photo of the cross section of a concrete core and the surface investigated by LIBS (area 70 mm x 70 mm). Right: Color coded chlorine intensity on the cross section of a concrete core, dark red represents high chlorine content.
The state of the art of LIBS technique for applications in civil engineering will be presented, including typical results of 2D investigation of concrete in laboratory. The performance is also demonstrated by examples for onsite applications using a mobile LIBS system. The road map to standardization is presented as well.
The main application of LIBS in civil engineering is the detection of harmful ions in concrete, which can penetrate the component through the porous concrete structure. The advantages of LIBS over standard methods are the possibility of multi-element analysis, measurement speed, spatially resolved measurements, and minimal sample preparation. The spatially resolved measurements of LIBS allow the assessment of the heterogeneity of the concrete by measuring separately the chemical composition of the aggregates and the binder matrix. The latter is particularly relevant because the determined elemental distribution can be directly related to the binder matrix. This is not possible with standard methods, since the material is homogenized to powder during sample preparation stage and the determined concentration is thus related to the total mass. In addition to the use of LIBS for the specific analysis of individual harmful ions, LIBS can also be used to estimate the concrete composition and thus determine, for example, the type of cement used. Corresponding information are relevant for the estimation of the remaining service life and for the preparation of a maintenance concept. In recent years, LIBS has been increasingly used in civil engineering. Currently, however, it is primarily used in research institutions and only occasionally in building materials laboratories. Special commercial devices have also been developed, which greatly simplify the application due to the high degree of automation. Mobile LIBS systems allow on-site application. A central point, which limits the use of LIBS in the commercial sector, is the lack of norms and standards. Therefore, within the framework of a project funded by the German government, work has been carried out on the preparation of a leaflet on quantitative chlorine determination in concrete, which will be published this year. In interlaboratory comparisons the robustness and accuracy for the practical application was demonstrated. LIBS also has great potential in the recycling of construction waste in conjunction with hyperspectral sensors. This issue is currently being addressed in a national project. During the presentation, the state of the art of LIBS in civil engineering will be presented, next steps will be discussed, and future challenges will be outlined.
Responsible treatment of the environment and resources is a key element of sustainability. The building and construction industry is one of the largest consumers of natural resources. Consequently, there is a particular need for regulations and technologies that help to create closed material cycles. From the technological point of view, such efforts are complicated by the growing material diversity and the amount of composites contained in present and future construction and demolition waste (CDW). Nowadays, simple but proven techniques like manual sorting are mainly used. However, this practice not only poses health risks and dangers to the staff performing the work, but also relies on merely obvious, visually striking differences. Automated, sensor-based sorting of these building materials could complement or replace this practice to improve processing speed, recycling rates, sorting quality, and prevailing health conditions. The preliminary results for the identification of a wide variety of building materials with LIBS are presented.
Two calibration-free LIBS techniques are used for the quantitative analysis of synthetic cement samples: the CF-LIBS based on the Boltzmann plot method and the Monte Carlo (MC) LIBS based on the iterative spectrum fitting. In CF-LIBS, the inverse problem is solved, i.e. the elemental concentrations are determined by the reconstruction of plasma parameters from spectra. The MC-LIBS technique solves the direct problem by finding the highest correlation between the model-generated and experimental spectrum. The accuracy of both calibration-free LIBS methods suffers from factors such as inaccurately determined instrumental function, the deviation of experimental plasma from the mathematical model used, not taking into account the collection geometry and from the uncertainty of spectroscopic data. Therefore, the both calibration-free LIBS approaches are applied to synthetic spectra which perfectly suit the mathematical model of the method. This test yields the accuracy of both the approaches for the ideal case. In addition, the accuracy of both methods is investigated for non-isothermal plasma, because real laser-induced plasma often has high gradients in temperature. Both methods assume an isothermal plasma.
For the determination of the remaining life-time and the degree of damage of reinforced concrete structures such as marine construction, bridges or parking decks, a highly precise measurement of harmful species is also required for trace elements. One of the most interesting elements is chlorine, because above a certain threshold corrosion is triggered. To increase the intensity of the chlorine line, helium is usually used, which is costly. To overcome this problem, low electrical discharge reheating is used which operates in air atmosphere. A comparison between results obtained by measuring with helium and reheating by electrical discharge is presented. The performance is compared by the resulting calibration curves and the calculated limit of detection obtained by 15 reference samples based on cement with NaCl. Concentrations of reference samples range from 0.05 to 2.5 wt% chlorine.
The determination of chloride is still one of the main tasks for the evaluation of reinforced concrete structures.
The corrosion of the reinforcement induced by the penetrating chlorides is the dominant damage process affecting the lifetime of concrete structures. In the recent years different research groups demonstrated that LIBS can be a fast and reliable method to quantify chlorine in cement-bound materials. Because chlorine in concrete can only occur as solved ions in the pore solution or bound in salts or hydrated cement phases, the detected emission of chlorine can be correlated with the chloride concentration determined e.g. with potentiometric titration. This work inter alia describes the production of reference samples and possible side effects during the production process. Due to transport processes in the porous matrix of the cement a misinterpretation of the concentrations is possible. It is shown how to overcome these effects and higher precisions of the single measurements can be realised. Using the calibration method, blank sample method and noise method, three different ways of calculating the limit of detection (LOD) and limit of quantification (LOQ) are compared. Due to the preparation of the reference samples a precision of the whole calibration model of sx0 = 0.023 wt% is determined.
The validation of the model is based on different test sets, which are varying in their composition
(different Cl-salts, water-to-cement ratios and additives). The determined mean error of the validation is
0.595 ± 0.063 wt%, which is comparable to standardised methods like potentiometric titration, direct potentiometry
or photometry (0.40 ± 0.06 wt%) [1].
In civil engineering, the investigation of existing infrastructure is of major importance for maintaining and ensuring stability of the structures. To ensure durability, uniform regulations and standards apply e.g. the European standard EN 206-1. In some countries, the EN-standard is supplemented by additional standards, as in Germany with DIN 1045-2. Here, specific application rules are described, e.g. for the cement type, to ensure the resistant to different exposures. Therefore, the knowledge of the materials originally used is important in assessing the condition of existing concrete structure. Unfortunately, these are often unknown and must be determined retrospectively. Therefore, we present the application of the laser-induced breakdown spectroscopy to distinguish between different types of cement. Spectral information’s are used to build a classification model. First, the accuracy of the classification is analyzed on ten pure laboratory cement samples. To investigate possible sources of error, the model was then applied to cement samples with different moisture content. The study shows that LIBS is a promising tool for distinguishing between cement types. For further industrial application, however, factors influencing the LIBS signal must be included to ensure a robust model.