Filtern
Dokumenttyp
- Vortrag (5)
- Posterpräsentation (3)
Sprache
- Englisch (8)
Referierte Publikation
- nein (8)
Schlagworte
- Isotopic shift (8) (entfernen)
Organisationseinheit der BAM
Eingeladener Vortrag
- nein (5)
Isotope analysis can be used to determine the age and provenance of geological samples. Modern techniques in optical spectrometry allow us a stand-off isotope analysis. This seminar will discuss how planetary science with the next Moon and Mars missions drive optical spectrometry into precise and accurate isotope analysis and how BAM will contribute.
Mass spectrometric methods (MC-ICP-MS and TIMS) are without doubt the working horse of stable isotope analysis. However, drawbacks of these methods include the high costs for instruments and their operation, experienced operators and elaborate chromatographic sample preparation which are time consuming.
We propose a lower-cost and faster optical alternative for the analysis of isotope ratios of selected elements: high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Stable isotope amount compositions of boron (B) and magnesium (Mg) were determined based on the absorption spectra of in-situ generated mono-hydrides and halide (MH and MX) using graphite furnace HR-CS-MAS. The use of a modular simultaneous echelle spectrograph (MOSES) helps to find the maximal isotope shift in the diatomic molecular spectra produced in a graphite furnace by using isotopic spike solutions. Isotopes of boron (10B and 11B) were studied via their hydrides for the electronic transition X 1Σ+ → A 1Π. The spectrum of a given sample is a linear combination of the 10BH molecule and its isotopologue 11BH. Therefore, the isotopic composition of samples can be calculated by a partial least square regression (PLS). For this, a spectral library was built by using samples and spikes with known isotope composition. Boron isotope ratios measured by HR-CS-MAS are identical with those measured by mass spectrometric methods at the 0.15 ‰ level. Similar results were obtained for a multiple isotope system like Mg (24Mg, 25Mg, and 26Mg), where isotope shifts of their isotopologues can be resolved in the MgF molecule for the electronic transition X 2Σ → A 2 Πi. The extension of this methodology to other elements like Li, Ca, Cu, and Sr is discussed.
Variations in the isotope amount composition of some elements like lithium, boron, magnesium, calcium and copper have been used as proof of provenance of a sample and to describe geological processes. Routinely, isotope compositions are determinate by mass spectrometry; the working horse of the isotope analysis. However, mass spectrometric methods are expensive, time consuming and they require a high qualified analysist.
Here an alternative faster and low cost optical method for isotope ratio determination is proposed: high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Stable isotope amount composition of X= Li, B, Mg, Ca, Cu and Sr have been determined by monitoring the absorption spectrum of their monohydride (XH) in graphite furnace HR-CS-MAS.
Analysis of stable isotopes has been used as proof of provenance of mineral and biological samples, to estimate a contamination source and to determine geological processes. This kind of analysis needs high accuracy and precision for reliable conclusions. Currently, stable isotope analysis is dominated by mass spectrometric techniques that are time consuming and expensive. Here we present a fast and low cost alternative for isotope analysis of boron and magnesium: high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GFMAS). Two stable isotope systems were evaluated separately: boron (10B:11B) and magnesium (24Mg:25Mg:26Mg). Their isotope amount ratios were estimated by monitoring their absorption spectra in-situ generated monohydrides. The molecular absorption spectrum of a XH molecule (X= B or Mg) with n isotopes would be a linear combination of n isotopologue spectra and the amount of each component (isotope) could be calculated by a multivariate regression (n= 2 and 3 for B and Mg respectively). For the analysis of boron certified reference materials (CRM), the band 1→1 for the electronic transition X1Σ+ → A1Π was measured around wavelength 437.1 nm. Since boron has a memory effect in graphite furnaces, a combination of 2 % (v/v) hydrogen gas in argon, 1 % trifluoromethane in argon, an acid solution of calcium chloride and mannitol as chemical modifiers were used during the BH vaporization at 2600 °C. Partial least square regression (PLS) for analysis of samples was applied. For this, a spectral library with different isotope ratios for PLS regression was created. Magnesium does not have memory effect. Therefore, only 2 % of hydrogen in argon as gas modifier during vaporization at 2500 °C was employed for analysis of magnesium CRM. Absorption spectra of MgH for the X2Σ→A2Π electronic transition (band 0→0) were recorded around wavelength 513.45 nm. A similar PLS procedure to the BH was applied. Results for B and Mg CRM are metrologically compatible with those reported by mass spectrometric methods. An accuracy of 0.08 ‰ for B and 0.1 ‰ Mg was obtained as the average deviation from the isotope CRM. Expanded uncertainties with a coverage factor of k = 2 range between 0.10 - 0.40 ‰.
Variations in the isotope amount composition of some elements like lithium, boron, magnesium, calcium, copper and strontium have been used as proof of provenance of a sample and to describe geological processes. Routinely, isotope compositions are determinate by mass spectrometry; the working horse of the isotope analysis. However, mass spectrometric methods are expensive, time consuming and they require a high qualified analysist.
Here an alternative faster and low cost optical method for isotope ratio determination is proposed: high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Stable isotope amount composition of X= Li, B, Mg, Ca, Cu and Sr have been determined by monitoring the absorption spectrum of their monohydride (XH) in graphite furnace HR-CS-MAS. For example, for the three Mg isotopes (24Mg, 25Mg and 26Mg) band (0→0) for the electronic transition X1Σ+ → A1Π was evaluated around wavelength 513.4 nm (Fig. 1). Partial least square regression (PLS) for analysis of samples and reference materials were applied. For this, a spectral library with different isotopes ratios for PLS regression were built. Results are metrological compatible with those reported by mass spectrometric methods.
Boron presents two stable isotopes, 10B and 11B and due to their relatively large mass difference (~ 10%) isotope fractionation leads to considerable isotope amount ratio variations n(10B)/n(11B) in natural occurrence. These have been used as a proof of provenance of mineral and biological samples, to estimate a contamination source and to the determination of geological processes by erosion or subduction. Additionally, boron is employed in the nuclear industry due to the capability of its isotope 10B to thermal-neutron capture and therefore 10B enriched boric acid solutions are used in the cooling system of thermonuclear facilities and in the alloying of steel and carbides for protective shielding. Usually, isotope ratio variations are determined by mass spectroscopic methods.
Here an alternative faster and low cost method for isotope ratio determination is proposed: high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Boron isotope amount ratios have been determined by monitoring the absorption spectrum of boron monohydride (BH) in graphite furnace HR-CS-MAS. Bands (0→0) and (1→1) for the electronic transition X1Σ+ → A1Π were evaluated around wavelengths 433.1 nm and 437.1 nm respectively. Partial least square regression (PLS) for analysis of samples and reference materials were applied. For this, a spectral library with different isotopes ratios for PLS regression was built. Results obtained around the 433.1 nm and 437.1 nm spectral regions are metrologically compatible with those reported by mass spectrometric methods. Moreover, a precision and accuracy of the method of ± 0.5 ‰ for the evaluated spectral region around 437.1 nm is reported here. This accuracy and precision is comparable with those obtained by thermal ionization mass spectrometry (TIMS) and multiple collector inductively coupled plasma mass spectrometry (MC-ICP-MS) for boron isotope ratio measurements.
Boron presents two stable isotopes, 10B and 11B and due to their relatively large mass difference (~ 10%) isotope fractionation leads to considerable isotope amount ratio variations n(10B)/n(11B) in natural occurrence. These have been used as a proof of provenance of mineral and biological samples, to estimate a contamination source and to the determination of geological processes by erosion or subduction. Additionally, boron is employed in the nuclear industry due to the capability of its isotope 10B to thermal-neutron capture and therefore 10B enriched boric acid solutions are used in the cooling system of thermonuclear facilities and in the alloying of steel and carbides for protective shielding. Usually, isotope ratio variations are determined by mass spectroscopic methods.
Here an alternative faster and low cost method for isotope ratio determination is proposed: high-resolution continuum source molecular absorption spectrometry (HR-CS-MAS). Boron isotope amount ratios have been determined by monitoring the absorption spectrum of boron monohydride (BH) in graphite furnace HR-CS-MAS. Bands (0→0) and (1→1) for the electronic transition X1Σ+ → A1Π were evaluated around wavelengths 433.1 nm and 437.1 nm respectively. Partial least square regression (PLS) for analysis of samples and reference materials were applied. For this, a spectral library with different isotopes ratios for PLS regression was built. Results obtained around the 433.1 nm and 437.1 nm spectral regions are metrologically compatible with those reported by mass spectrometric methods. Moreover, a precision and accuracy of the method of ± 0.5 ‰ for the evaluated spectral region around 437.1 nm is reported here. This accuracy and precision is comparable with those obtained by thermal ionization mass spectrometry (TIMS) and multiple collector inductively coupled plasma mass spectrometry (MC-ICP-MS) for boron isotope ratio measurements.