A novel approach is suggested to strongly suppress artifacts in radiography and computed tomography caused by the effect of diffuse background signals 'backlighting' of 2D X-ray detectors. Depending on the detector geometry the mechanism may be different. Either based on the optical scattering of the fluorescent screen materials into the optical detection devices or Compton or X-ray fluorescence scattering by the detector components. Consequently, these erroneous intensity portions result in locally different violations of Lambert Beer's law in single projections (radiographs) as a function of the detector area coverage and the magnitude of the attenuation. The absorption of multiple metal sheets is investigated by monochromatic synchrotron radiation, thus excluding beam hardening. The proposed correction procedure simply requires the individual subtraction of one and the same fraction of the primary and transmitted mean intensity, as a constant (non-local) scattering mechanism is assumed.
A novel approach to strongly suppress artifacts in radiography and computed tomography caused by the effect of
diffuse background signals ('backlighting') of 2D X-ray detectors is suggested. Depending on the detector
geometry the mechanism may be different, either based on the optical scattering of the fluorescent screen
materials into optical detection devices or Compton or X-ray fluorescence scattering by the detector components.
Consequently, these erroneous intensity portions result in locally different violations of Lambert Beers law in
single projections (radiographs) as function of the detector area coverage and the magnitude of the attenuation.
The absorption of multiple metal sheets is investigated by monochromatic synchrotron radiation, thus excluding
beam hardening. The proposed correction procedure simply requires the individual subtraction of one and the
same fraction of the primary and transmitted mean intensity, as a constant (non-local) scattering mechanism is
assumed.
A novel approach to strongly suppress artifacts in radiography and computed tomography caused by the effect of diffuse background signals ('backlighting') of 2D X-ray detectors is suggested. Depending on the detector geometry the mechanism may be different, either based on the optical scattering by the fluorescent screen materials into optical detection devices or Compton or X-ray fluorescence scattering by the detector components. Consequently, these erroneous intensity portions result in locally different violations of Lambert–Beer's law in single projections (radiographs). When used as input data for computed tomography these violations are directly observed via modulation of the projected mass as a function of the rotation phase and the samples aspect ratio (dynamics). The magnitude of the diffuse background signal depends on the detector area covered by the projected sample. They are more pronounced the smaller the shadowed area and the stronger the total attenuation. This implies that the reconstruction suffers from additional anisotropic artifacts caused by elongated sample structures. This issue is studied simply by absorption of flat plates in a conventional laboratory radiography set-up and at a synchrotron radiation facility. In the latter case beam hardening artifacts can be excluded due to the monochromatic radiation. The proposed correction procedure requires simple integral intensity offsets as a constant (non-local) light scattering mechanism is assumed.