The combination of acoustically levitated droplets, mid-IR laser evaporation, and subsequent post-ionization by secondary electrospray ionization was applied for monitoring the enzymatic digestion of various proteins. Acoustically levitated droplets are an ideal, wall-free model reactor, readily allowing compartmentalized microfluidic trypsin digestions. Time-resolved interrogation of the droplets yielded real-time information on the progress of the reaction and thus provided insights into reaction kinetics. After 30 min of digestion in the acoustic levitator, the obtained protein sequence coverages were identical to the reference overnight digestions. Importantly, our results clearly demonstrate that the applied experimental setup can be used for the real-time investigation of chemical reactions. Furthermore, the described methodology only uses a fraction of the typically applied amounts of solvent, analyte, and trypsin. Thus, the results exemplify the use of acoustic levitation as a green analytical chemistry alternative to the currently used batch reactions.
The composition of acoustically levitated droplets was probed by a novel combination of mid-IR laser evaporation and subsequent postionization via secondary electrospray ionization. The combination of microliter samples and subnanoliter sampling provided time-resolved interrogation of droplets and allowed for a kinetic investigation of the laser-induced release of the analyte, which was found to strongly depend on the analytes. The observed substancespecific delayed release of the analytes permitted baseline-separated discrimination of the analytes, ideal for the study of complex samples.
The additionally applied postionization scheme was found to enable efficient detection of small volatile compounds as well as peptides. The detection of small molecules and peptides occurred under very different sampling geometries, pointing to two distinct underlying ionization mechanisms. Overall, our results suggest that the experimental setup presented in this study can serve as a widely applicable platform to study chemical reactions in acoustically levitated droplets as model reactors.
Acoustically levitated droplets have been suggested as compartmentalized, yet wall-less microreactors for high-throughput reaction optimization purposes. The absence of walls is envisioned to simplify up-scaling of the optimized reaction conditions found in the microliter volumes. A consequent pursuance of high-throughput chemistry calls for a fast, robust and sensitive analysis suited for online interrogation. For reaction optimization, targeted Analysis with relatively low sensitivity suffices, while a fast, robust and automated sampling is paramount. To follow this approach, in this contribution, a direct coupling of levitated droplets to a homebuilt ion mobility spectrometer (IMS) is presented. The sampling, Transfer to the gas phase, as well as the ionization are all performed by a single exposure of the sampling volume to the resonant output of a mid-IR laser. Once formed, the nascent spatially and temporally evolving analyte ion cloud needs to be guided out of the acoustically confined trap into the inlet of the ion mobility spectrometer. Since the IMS is operated at ambient pressure, no fluid dynamic along a pressure Gradient can be employed. Instead, the transfer is achieved by the electrostatic potential gradient inside a dual ring electrode ion optics, guiding the analyte ion cloud into the first stage of the IMS linear drift tube accelerator. The design of the appropriate atmospheric pressure ion optics is based on the original vacuum ion optics design of Wiley and McLaren. The obtained experimental results nicely coincide with ion trajectory calculations based on a collisional model.
Polynuclear cationic aluminum hydroxide phases are known intermediates in the formation of aluminum oxides or (oxide) ydroxides upon hydrolysis of aluminum salt solutions. In the presence of sulfate anions, these aluminum polyoxocations (Al13) can form crystalline Al13 sulfates with varying chemical composition.
The formation of these Al13 sulfates in aqueous solution has been poorly understood. Here, we investigate the early stage crystallization of Al13 clusters in a sulfate-containing solution, in situ and in real time. Dynamics associated with Al13 sulfate formation have been obtained for the first time, using Synchrotron X-ray diffraction (XRD) of solutions suspended by acoustic levitation. Time-resolved in situ data show that the cubic phase, Na [(AlO4)Al12(OH)24(H2O)12](SO4)4*10H2O, forms after only minutes. The Formation mechanism of Al13 sulfates was found to depend on the sulfate:aluminum (SO4:Al) ratio. Ex situ XRD of the product Al13 sulfates in solution shows that for SO4:Al ratio ≤ 1.5 two other crystalline phases form, and convert to the cubic phase upon washing and drying. In situ XRD for the same ratio shows transient formation of an intermediate during the crystallization process.
Mechanochemistry is increasingly used for synthesizing soft matter materials including metal organic compounds and cocrystals. The ever-increasing interest in this method is contrasted by a limited mechanistic understanding of the mechanochemical reactivity and selectivity. Time-resolved in situ investigations of milling reactions provide direct insights in the underlying mechanisms. We introduced a setup enabling in situ investigation of mechanochemical reactions using synchrotron XRD combined with Raman spectroscopy. The specific combination allows to study milling processes comprehensively on the level of the molecular and crystalline structure and thus obtaining reliable data for mechanistic studies. We discuss our recent results investigating the formation of (polymorphic) cocrystals. First investigations of a mechanochemical synthesis under controlled temperature which allow determining the activation barrier are presented.6 Furthermore, X-ray diffraction and in situ Raman spectroscopy coupled with thermography revealed a low temperature increase during milling reactions due to the mechanical impact and clear temperature increases as a result of the reaction heat. Our results indicate that in situ investigation of milling reactions offer a new approach to tune and optimize mechanochemically synthesized compounds.
The upscaling from small scale academic reactors to large industrial processes typically suffers from a large change in surface-to-volume ratio. A promising approach is the general avoidance of surfaces as in levitated droplet techniques. However, up to now, no mass spectrometric interface for online reaction monitoring in levitated droplets has been provided.
As model reaction the photoinitiated thiol-ene coupling between N-boc cysteine methyl ester and allyl alcohol was studied. A droplet of 5 µL reactand solution was provisioned into an acoustic trap aligned lateral to the MS inlet. Contactless sampling by laser ablation (LA) is followed by dielectric barrier discharge (DBD) postionization. The latter is needed to address non polar reaction partners.
Real time monitoring of chemical reactions has become a key step in industrial processes due to constantly increasing demands on product performance and environmental compatibility. In contrast to spectroscopic methods that usually require sample pretreatment, mass spectrometry (MS) has been proven as a robust method for multicomponent analysis. As an especially demanding reaction process, we here report successfully on the entirely contactless conduction and interrogation of a chemical reaction inside an acoustically levitated microliter droplet of reaction mixture. The approach represents a proof of concept study for fast reaction optimization approaches with minimal resource consumption. The sampling is done by repeated laser desorption of small fractions of the droplets volume spread over the timescale of the reaction process.
This paper presents a numerical and an experimental procedure to obtain the pressure field in single-axis acoustic levitators. Numerically, the pressure field is determined by a matrix method based on the Rayleigh integral that take into account the multiple wave reflections that occur between the transducer and the reflector. The numerical pressure field is compared with the acoustic pressure measured by an earplug microphone, that is connected to a hollow needle. The tip of the needle is moved by a µm translation stage through the field and the signal is recorded using a lock-in amplifier locked to the levitator frequency. The pressure field obtained numerically show good agreement with that obtained experimentally.
Carbamazepine (CBZ) is known for its variety of anhydrous and hydrous polymorphs. Herein, a thorough analysis of the crystallization behavior of CBZ is presented. The influence of the solvent and the concentration on the crystallization behavior under different environmental conditions is investigated by combined in situ XRD and Raman spectroscopy measurements. Crystallization studies conducted via conventional crystallization in glass tubes and in levitated droplets using an acoustic levitator indicate a dependence of the crystallization process from solvent and surface.