In a color X-ray camera spatial resolution is achieved by means of a polycapillary optic conducting X-ray photons from small regions on a sample to distinct energy dispersive pixels on a CCD matrix. At present, the resolution limit of color X-ray camera systems can go down to several microns and is mainly restricted by Pixel dimensions. The recent development of an efficient subpixel resolution algorithm allows a release from pixel size, limiting the resolution only to the quality of theoptics. In this work polycapillary properties that influence the spatial resolution are systematized and assessed both theoretically and experimentally. It is demonstrated that with the current technological Level reaching one micron resolution is challenging, but possible.
Aufgrund der natürlichen und vom Menschen verursachten Heterogenität von Böden erfordert der Erhalt und die Verbesserung ihrer Fruchtbarkeit durch landwirtschaftliche Bewirtschaftungsmaßnahmen sorgfältig geplante Ent-scheidungen, die auf einer detaillierten Erfassung der Bodeneigenschaften und einem vertieften Verständnis der Bodenprozesse beruhen. Es ist bekannt, dass die konventionelle, flächen-einheitliche Bewirtschaftung Ertragsver-luste durch zu geringe Bewirtschaftungsintensität (z. B. Düngung) auf einen Teil der Fläche verursacht, während andere Teile des Feldes zu hohe Dosen erhalten und es dadurch zu Verschwendung von Ressourcen und zu Um-weltbelastungen kommt. Trotz der Verfügbarkeit von Technologien für ortsspezifische Düngung ist die Akzeptanz ortsspezifischer Bewirtschaftung (Precision Agriculture) in der Praxis noch gering. Ein wesentlicher Grund dafür ist das Fehlen von kostengünstigen Methoden zur Erfassung der bewirtschaftungsrelevanten Bodenmerkmale. So-wohl für die elementanalytische Untersuchung von Böden als auch für die Untersuchung von Ausgangsmaterialien (z.B. Klärschlammaschen) für ihre Verwendung als Dünger ist eine zuverlässige Analytik und Prozessüberwachung erforderlich.
Es wurde ein Online-Analysenverfahren zur Bestimmung von Elementgehalten in Klär-schlammaschen und deren Produkten aus einem thermochemischen Verfahren entwickelt. Durch das thermochemische Verfahren können Wertstoffe aus den Aschen wieder für die Kreislaufwirtschaft nutzbar gemacht und Schadstoffe sicher entfernt werden. Das Analysensystem ist eine Methodenkombination von Online-Laser induzierter Plasma Spektroskopie (LIBS) und Online-Röntgenfluoreszenzanalyse (RFA). Robustheit und modulare Bauweise des Verfahrens sind wichtige Anforderungen um auch für andere Anwendungen nutzbar zu sein.
The color X-ray camera SLcam® is a full-field, single photon detector providing scanning-free, energy and spatially resolved X-ray imaging. Spatial resolution is achieved with the use of polycapillary optics guiding X-ray photons from small regions on a sample to distinct energy dispersive pixels on a charged-coupled device detector. Applying sub-pixel resolution, signals from individual capillary channels can be distinguished. Therefore, the SLcam® spatial resolution, which is normally limited to the pixel size of the charge-coupled device, can be improved to the size of individual polycapillary channels. In this work a new approach to a sub-pixel resolution algorithm comprising photon events also from the pixel centers is proposed. The details of the employed numerical method and several sub-pixel resolution examples are presented and discussed.
Examples of XRF and PIXE imaging with few microns resolution using SLcam® a color X-ray camera
(2015)
We present results of recent development of the color X-ray camera, type SLcam®, allowing detection of X-ray images with few microns resolution. Such spectral resolution is achieved with the use of high-quality polycapillary optics combined with sub-pixel resolution. Imaging of Siemens star resolution test chart reveals that the resolution limit of SLcam® can go down to nearly 5µm. Several real sample examples of measurements carried out at the laboratory, synchrotron, and particle-induced X-ray emission beamlines are shown. This is the first time SLcam® is used as particle-induced X-ray emission detector.
Elementanalyse mit Röntgenfluoreszenzspektrometrie direkt am Materialstrom in Produktionsprozessen gewinnt zunehmend an Bedeutung. Die Röntgenfluoreszenzspektrometrie kann aufgrund der technischen Entwicklung mittlerweile ohne aufwändige Probenvorbereitung als aussagekräftige Informationsquelle für Stoffparameter im industriellen Prozess eingesetzt werden. Der Einsatz von online Röntgenfluoreszenzanalyse zahlt sich sowohl finanziell als auch zeitlich schnell aus, da sie sowohl für den ökonomischen Einsatz von Rohmaterialien, als auch z. B. im Materialrecycling zu gebrauchen ist. In diesem Beitrag werden vier unterschiedliche Einsatzmöglichkeiten beschrieben.
We present a new high resolution X-ray imager based on a pnCCD detector and a polycapillary optics. The properties of the pnCCD like high quantum efficiency, high energy resolution and radiation hardness are maintained, while color corrected polycapillary lenses are used to direct the fluorescence photons from every spot on a sample to a corresponding pixel on the detector. The camera is sensitive to photons from 3 to 40 keV with still 30% quantum efficiency at 20 keV. The pnCCD is operated in split frame mode allowing a high frame rate of 400 Hz with an energy resolution of 152 eV for Mn Kα (5.9 keV) at 450 kcps. In single-photon counting mode (SPC), the time, energy and position of every fluorescence photon is recorded for every frame. A dedicated software enables the visualization of the elements distribution in real time without the need of post-processing the data. A description of the key components including detector, X-ray optics and camera is given. First experiments show the capability of the camera to perform fast full-field X-Ray Fluorescence (FF-XRF) for element analysis. The imaging performance with a magnifying optics (3×) has also been successfully tested.
Compact pnCCD-based X-ray camera with high spatial and energy resolution: a color X-ray camera
(2011)
For many applications there is a requirement for nondestructive analytical investigation of the elemental distribution in a sample. With the improvement of X-ray optics and spectroscopic X-ray imagers, full field X-ray fluorescence (FF-XRF) methods are feasible. A new device for high-resolution X-ray imaging, an energy and spatial resolving X-ray camera, is presented. The basic idea behind this so-called 'color X-ray camera' (CXC) is to combine an energy dispersive array detector for X-rays, in this case a pnCCD, with polycapillary optics. Imaging is achieved using multiframe recording of the energy and the point of impact of single photons. The camera was tested using a laboratory 30 µm microfocus X-ray tube and synchrotron radiation from BESSY II at the BAMline facility. These experiments demonstrate the suitability of the camera for X-ray fluorescence analytics. The camera simultaneously records 69696 spectra with an energy resolution of 152 eV for manganese Kα with a spatial resolution of 50 µm over an imaging area of 12.7 × 12.7 mm². It is sensitive to photons in the energy region between 3 and 40 keV, limited by a 50 µm beryllium window, and the sensitive thickness of 450 µm of the chip. Online preview of the sample is possible as the software updates the sums of the counts for certain energy channel ranges during the measurement and displays 2-D false-color maps as well as spectra of selected regions. The complete data cube of 264 × 264 spectra is saved for further qualitative and quantitative processing.
The propagation of synchrotron radiation through nanocapillary structures with channel sizes of 200 nm and periods in the micrometer size has been studied experimentally. It was shown that the propagation through individual capillary channels has a mode formation character. Furthermore it was shown that during the propagation through capillary channels the coherence of synchrotron radiation is partially conserved. Interference of beams propagating through different capillary channels is observed which leads to a periodically modulated distribution of the radiation intensity in a plane far from the exit of the structure. These investigations are of high relevance for the understanding of X-ray transmission through nanocapillaries and the appearance of wave properties at this size scale.
X-ray Fluorescence (XRF) with a scanning electron microscope (SEM) is a valuable completion of the analytical capabilities of SEMs. Small and compact micro-focus x-ray sources are mounted to the microscope chamber, and the x-ray spectra are monitored with conventional EDS systems. Up to now the x-ray tubes used for the micro-focus x-ray sources are equipped with beryllium windows about 100 µm thick. The poly-capillary x-ray lenses have their transmission maximum at photon energies around 10 keV. It drops down in both low- and high-energy ranges. Hence, L-radiation from an Mo or Rh target will be strongly attenuated, and the excitation of fluorescence in the soft x-ray range becomes very ineffective. A new micro-focus x-ray source was developed. It is characterised by a lower self-absorption in the tube target, thin beryllium windows and an x-ray optics having a large distance between its foci and the maximum of transmission at about 5 keV. Thus K line fluorescence of light elements becomes effectively excited by the L-radiation from Mo or Rh tube targets. The detection limit for sodium oxide in glass was found to be below 1 mass%.