Purity statements of high purity materials serving as primary standards by GDMS.
Primary standards are materials known for their total purity and therefore appropriate to realise the link with the International System of Units (SI). The realisation and dissemination of primary standards is of fundamental importance for comparability of measurement results through traceability in all fields of chemical analysis.
To serve as a primary standard for element determination the total purity of a high purity material needs to be measured. In order to achieve a sufficiently small uncertainty (i.e. < 0.01%) this involves determining all possible impurity contributions and to subtract their sum from the ideal purity of 100%.
GD-MS has the potential to reduce the effort for purity determination, as a fast sensitive multi-element analysis without extensive sample preparation. Similar to other solid sampling techniques, glow discharge
requires calibration link the measured signal and the content of the impurities in the matrix.
The use of the concept Relative Sensitivity Factors (RSFs) provides good approximations especially for high purity materials. However, it only works with a wide uncertainty of the results.
The preparation of synthetic pressed powder samples by different modifications of liquid and powder doping used for the determination of both metallic and none-metallic impurities are described. Efficient determination of metallic impurities by GD-MS via differential and absolute measurements with a significantly reduced target uncertainty of 20 % for absolute measurements using liquid doped pressed powder calibration could be achieved. The determination of impurities has been evaluated by independent analytical methods such as HR-ICP-MS, ETAAS; CGHE and combustion analysis
Due to its advantages of being a direct comparison method, quantitative NMR spectroscopy (qNMR) becomes more and more popular in industry. While conventional high-field NMR systems are often associated with high investment and operational costs, the upcoming market of permanent-magnet based benchtop NMR systems show a considerable option for a lot of applications. The mobility of these systems allows to bring them more closely to the real production environment, e.g. for at-line quality control.
In this work we present an interlaboratory comparison study investigating the qNMR performance of state-of-the-art benchtop NMR spectrometers. Therefore, BAM prepared two samples of a mixture of NMR reference standards tetramethylbenzene (TMB) and tetrachloronitrobenzene (TCNB) at concentration levels of 200 mM and 10 mM. These “ready-to-use” samples were sent to participant laboratories, which performed analysis on their benchtop NMR equipment of different vendors and fields from 43 to 80 MHz. Raw data was reported back and further investigated by using different data analysis methods at BAM.
After this very first qNMR comparison study of benchtop NMR spectrometers show promising results, following studies are planned to cover more parts of the qNMR process, e.g. sample preparation and weighing, but also data analysis, as commonly done in similar studies for high-field NMR spectroscopy in industry and metrology.
Introduction
NMR spectroscopy is one of the most important analytical methods in organic chemistry. While most analyses are carried out qualitatively with the aim of substance identification and structure elucidation, quantitative NMR spectroscopy (qNMR) is increasingly gaining importance in research and industry. qNMR provides the most universally applicable form of direct purity determination without need for reference materials of impurities or the calculation of response factors but only exhibiting suitable NMR properties.
Methods
One of the most attractive features of quantitative NMR spectroscopy is that the NMR peak areas can be used directly for concentration quantification without further calibration. Another advantage of NMR spectroscopy is that the method has a high linearity between absolute signal area and sample concentration, which makes it an absolute analytical comparison method that is independent of the matrix. This enables automated robust data evaluation strategies that can be used for online applications of qNMR spectroscopy.
Jancke et al. proposed NMR spectroscopy as a relative primary analytical method because it can be fully described by mathematical equations from which a complete uncertainty budget can be derived, allowing it to be used at the highest metrological level. Weber et al. discussed in detail important aspects of the procedure that enable the realisation of low measurement uncertainties in qNMR measurements. Since certification of CRM requires expanded mea¬sure¬ment uncertainties of less than 0.5 % (relative), the work of Weber et al. demonstrated for the first time that qNMR can fulfil this criterion.
Results
To date, further comparative studies have been carried out in metrology and industry, demonstrating the performance of quantitative NMR spectroscopy and further reducing measurement uncertainties. The development of validation concepts and the commercial availability of suitable certified reference materials facilitate the application, especially in the usually highly regulated industrial environment. Users can thus accelerate the development of analytical methods. The talk will cover a wide range of topics from current metrological activities to new challenges for qNMR spectroscopy and also deals with aspects such as validation and accreditation.
Innovative aspects
• qNMR provides the most universally applicable form of direct purity determination
• Expanded measurement uncertainties lower than 0.15 % (relative) possible
• Benchtop NMR instruments increasingly used for qNMR spectroscopy