An interlaboratory comparison (ILC)was organised to characterise 87Sr/86Sr isotope ratios in geological and industrial reference materials by applying the so-called conventional method for determining 87Sr/86Sr isotope ratios. Four cements (VDZ 100a,VDZ 200a, VDZ 300a, IAG OPC-1), one limestone (IAG CGL ML-3) and one slate (IAG OU-6) reference materials were selected, covering a wide range of naturally occurring Sr isotopic signatures. Thirteen laboratories received aliquots of these six reference materials together with a detailed technical protocol. The consensus values for the six reference materials and their associated measurement uncertainties were obtained by applying a Gaussian, linear mixed effects model fitted to all the measurement results. By combining the consensus values and their uncertainties with an uncertainty contribution for potential heterogeneity, reference values ranging from 0.708134 mol mol-1 to 0.729778 mol mol-1 were obtained with relative expanded uncertainties of ≤ 0.007 %. This study represents an ILC on conventional 87Sr/86Sr isotope ratios, within which metrological principles were considered and the compatibility of measurement results obtained by MC-ICP-MS and by MC-TIMS is demonstrated. The materials characterised in this study can be used as reference materials for validation and quality control purposes and to estimate measurement uncertainties in conventional 87Sr/86Sr isotope ratio measurement.
Isotope ratio applications are on the increase and a major part of which are delta measurements, because they are easier to perform than the determination of absolute isotope ratios while offering lower measurement uncertainties. Delta measurements use artefact-based scales and therefore scale conversions are required due to the lack of the scale defining standards. Such scale conversions often form the basis for comparing data being generated in numerous projects andtherefore need to be as accurate as possible. In practice, users are tempted to apply linear approximations, which are not sufficiently exact, because delta values are defined by nonlinear relationships. The bias of such approximations often is beyond typical measurement uncertainties and its extent can hardly be predicted. Therefore, exact calculations are advised. Here, the exact equations and the bias of the approximations are presented, and calculations are illustrated by real-world examples. Measurement uncertainty is indispensable in this context and therefore, its calculation is described as well for determining delta values but also for scale conversions. Approaches for obtaining a single delta measurement and for repeated measurements are presented. For the latter case, a new approach for calculating the measurement uncertainty is presented, which considers covariances between the isotope ratios.
The ability of industrial X-ray computed tomography (CT) to scan an object with several internal and external features at once causes increasing adoption in dimensional metrology. In order to evaluate the quality of a measurement value, the task-specific measurement uncertainty has to be determined. Currently, VDI/VDE 2630 part 2.1 gives a guideline to determine the uncertainty of CT measurements experimentally by conducting repeated measurements. This is costly and time-consuming. Thus, the aim is to determine the task-specific measurement uncertainty numerically by simulations (e. g. according to the guide to expression of uncertainty in measurement (GUM) Supplement 1). To achieve that, a digital twin is necessary. This contribution presents a simple first approach how a digital twin can be built. In order to evaluate this approach, a study comparing measurements and simulations of different real CT systems was carried out by determining the differences between the measurement results of the digital twin and of the measurement results of the real-world CT systems. The results have shown a moderate agreement between real and simulated data. To improve on this aspect, a standardized method to characterize CT systems and methods to implement CT parameters into the simulation with sufficient accuracy will be developed.
Making optimal decisions about the reliability of existing structures requires that the information used in assessment adequately represents the properties and the condition of the structures. The knowledge gap regarding a structure to be assessed can be successively filled by individually purposeful observations on site. This paper gives an overview of an approach for utilizing nondestructively gathered measurement results in reliability assessment of existing structures. An essential part of measurement-based stochastic modeling of basic variables is the calculation of measurement uncertainties, which serves to establish confidence in measurement, to ensure the comparability of unambiguously expressed measurement results, and to quantify the quality of the measured information. Regarding the current discourse on how to treat information collected on-site in the context of assessment, the authors recommend that measurement uncertainty becomes an uncertainty component mandatorily to be represented in measurement-based stochastic models. The main steps of the proposed concept are presented, and the advantages of its application are emphasized by means of a prestressed concrete bridge as case study. The bridge is assessed regarding the serviceability limit state decompression using ultrasonic and radar data measured at the structure.
Laboratory test results are of vital importance for correctly classifying and labelling chemicals as “hazardous” as defined in the UN Globally Harmonized System (GHS) / EC CLP Regulation or as “dangerous goods” as defined in the UN Recommendations on the Transport of Dangerous Goods. Interlaboratory tests play a decisive role in assessing the reliability of laboratory test results. Interlaboratory tests performed over the last 10 years have examined different laboratory test methods. After analysing the results of these interlaboratory tests, the following conclusions can be drawn:
1. There is a need for improvement and validation for all laboratory test methods examined.
2. To avoid any discrepancy concerning the classification and labelling of chemicals, the use of validated laboratory test methods should be state of the art, with the results accompanied by the measurement uncertainty and (if applicable) the probability of incorrect classification.
This paper addresses the probability of correct/incorrect classification (for example, as dangerous goods) on the basis of the measurement deviation obtained from interlaboratory tests performed by the Centre for quality assurance for testing of dangerous goods and hazardous substances (CEQAT-DGHS) to validate laboratory test methods. This paper outlines typical results (e.g. so-called “Shark profiles” – the probability of incorrect classification as a function of the true value estimated from interlaboratory test data) as well as general conclusions and steps to be taken to guarantee that laboratory test results are fit for purpose and of high quality.
In most chemical reactions, stable isotopes are fractionated in a mass-dependent manner, yielding correlated isotope ratios in elements with three or more stable isotopes. The proportionality between isotope ratios is set by the triple isotope fractionation exponent θ that can be determined precisely for, e.g., sulfur and oxygen by IRMS, but not for metal(loid) elements due to the lower precision of MC-ICP-MS analysis and smaller isotopic variations. Here, using Mg as a test case, we compute a complete metrologically robust uncertainty budget for apparent θ values and, with reference to this, present a new measurement Approach that reduces uncertainty on θ values by 30%. This approach, namely, direct educt-product bracketing (sample−sample bracketing), allows apparent θ values of metal(loid) isotopes to be determined precisely enough to distinguish slopes in three-isotope space. For the example of Mg, we assess appropriate quality Control standards for interference-to-signal ratios and Report apparent θ values of carbonate−seawater pairs. We determined apparent θ values for marine biogenic carbonates, where the foraminifera Globorotalia menardii yields 0.514 ± 0.005 (2 SD), the coral Porites, 0.515 ± 0.006 (2 SD), and two specimens of the giant clam Tridacna gigas, 0.508 ± 0.007 (2 SD) and 0.509 ± 0.006 (2 SD), documenting differences in the uptake pathway of Mg among marine calcifiers. The capability to measure apparent θ values more precisely adds a new dimension to metal(loid) δ values, with the potential to allow us to resolve different modes of fractionation in industrial and natural processes.
Lead isotope amount ratios are commonly used in diverse fields such as archaeometry, geochemistry and forensic science. Currently, five reference materials with certified lead isotope amount ratios are available, namely NIST SRM 981, 982 and 983, GBW-04442 and NMIJ 3681-a. Only NIST SRM 981 and NMIJ 3681-a have approximately natural isotopic compositions, and NIST SRM 981 is predominantly used for correcting mass discrimination/mass fractionation in the applied mass spectrometric procedures. Consequently, there is no other certified reference material available to be used for validation and/or quality control of the analytical procedures applied to lead isotope amount ratio measurements. To fill this gap, two new reference materials have been produced and certified for their lead isotope amount ratios. For both certified reference materials, complete uncertainty budgets have been calculated and SI traceability has been established. This provides the users with independent means for validating and verifying their analytical procedures and for conducting quality control measures. ERM-EB400 is a bronze material with a nominal lead mass fraction of 45 mg kg-1 and certified lead isotope amount ratios of n(206Pb)/n(204Pb) = 18.072(17) mol mol-1, n(207Pb)/n(204Pb) = 15.578(18) mol mol-1 and n(208Pb)/n(204Pb) = 38.075(46) mol mol-1 with the associated expanded uncertainties (k = 2) given in brackets. ERM-AE142 is a high-purity solution of lead in 2% nitric acid with a nominal mass fraction of 100 mg kg-1 and certified Pb isotope amount ratios of n(206Pb)/n(204Pb) = 21.114(17) mol mol-1, n(207Pb)/n(204Pb) = 15.944(17) mol mol-1 and n(208Pb)/n(204Pb) = 39.850(44) mol mol-1 with the associated expanded uncertainties (k = 2) given in brackets. Both materials are specifically designed to fall within the natural lead isotopic variation and to assist users with the validation and verification of their analytical procedures. Note that while one of these reference materials requires the chemical separation of Pb from its matrix (ERM-EB400), the other does not (ERM-AE142). As additional information, δ208/206PbNIST SRM981 values are provided for both materials. For ERM-AE142, a delta value of δ208/206PbNIST SRM981 = -28.21(30) ‰ was obtained, and for ERM-EB400, a delta value of δ208/206PbNIST SRM981 = -129.47(38) ‰ was obtained, with the associated expanded uncertainties (k = 2) given in brackets.
Especially trace amounts of water vapour in gases can be reliably determined by coulometric trace humidity sensors. The principle of these sensors is based on water vapour absorption in a hygroscopic layer and its subsequent electrolytic decomposition. The calibration of sensors was performed in the humidity range, expressed as frost point temperature, from −30°C to −80°C . This range is equivalent to volume fractions smaller than 376 µL·L−1. Generated humidity was measured with coulometric sensors and a chilled dew point hygrometer that was used as reference. An empirical non-linear function was found between sensor signal and measured reference humidity. This function consists of two parameters with a measurement uncertainty. Both calibration parameters were checked by means of one-way analysis of variance. It showed that gas specific function can be used for humidity measurement in nitrogen, hydrogen, dinitrogen monoxide, compressed and synthetic air. It is possible to determine trace humidity in all tested gases with an expanded uncertainty less than 2.1 K (coverage factor k=2 ) regarding frost point temperature.
Trace humidity was measured by using miniaturized planar coulometric sensors in technical gases such as hydrogen, nitrogen, helium, nitrous oxide, and synthetic air. Frost point temperatures tf in the gases ranged from −60 °C to −30 °C, which is equivalent to a vapour mole fraction xv from 10 μmol mol−1 to 376 μmol mol−1. In addition, the generated humidity was determined by using a precision dew point hygrometer as reference. Nonlinear calibration functions were calculated that correlated electric current (sensor signal) and reference humidity. Parameters of functions were tested with one-way analysis of variances (ANOVA) to prove if all used sensors had a similar behavior in the same gas during experiments. Results of ANOVA confirmed that averaged functions can be applied for trace humidity measurement in nitrogen, helium, nitrous oxide, and synthetic air. The calculated functions were negligibly different for nitrogen, helium, and synthetic air. In humidified nitrous oxide, a minor change of parameters was observed due to lower electrical currents. In total contrast to that, the measured sensor signals were significantly higher in humidified hydrogen and each sensor required its own calibration function. The reason was a recombination effect that favoured multiple measurements of water molecules. Nevertheless, it was possible to measure continuously trace humidity in all tested gases by using coulometric sensors with an expanded uncertainty below 2 K (k = 2).
The primary crystallite size of titania powder relates to its properties in a number of applications. Transmission electron microscopy was used in this interlaboratory comparison (ILC) to measure primary crystallite size and shape distributions for a commercial aggregated titania powder. Data of four size descriptors and two shape descriptors were evaluated across nine laboratories. Data repeatability and reproducibility was evaluated by analysis of variance. One-third of the laboratory pairs had similar size descriptor data, but 83% of the pairs had similar aspect ratio data. Scale descriptor distributions were generally unimodal and were well-described by lognormal reference models. Shape descriptor distributions were multi-modal but data visualization plots demonstrated that the Weibull distribution was preferred to the normal distribution. For the equivalent circular diameter size descriptor, measurement uncertainties of the lognormal distribution scale and width parameters were 9.5% and 22%, respectively. For the aspect ratio shape descriptor, the measurement uncertainties of the Weibull distribution scale and width parameters were 7.0% and 26%, respectively. Both measurement uncertainty estimates and data visualizations should be used to analyze size and shape distributions of particles on the nanoscale.
For the accuracy of inertial navigation systems for indoor localization it is important to get high quality sensor data of the multi-sensor system. This can be realized using high quality sensors or the developed 3D-self-calibration-method for low cost sensors. Based on the calibration procedure of the accelerometer (ACC) and the magnetic field sensor (MAG), the additional integration of the gyroscope (GYRO) leads to a reduction of the indoor positioning error. This improves both the approximation for the accelerometer, the magnetic field sensor and the gyroscope so that the standard deviation of a single sensor is minimized. There are errors in the whole system. To determine these error sources it is important to define the measurement uncertainty. In this paper it is presented that the measurement uncertainty can be reduced by the application of the developed 3D-self-calibration method.
The article gives a condensed version of the keynote lecture held at the International Mass Spectrometry Conference 2012 in Kyoto. Starting with some examples for isotope research the key requirements for metrologically valid procedures enabling traceable and comparable isotope data are discussed. Of course multi-collector mass spectrometers are required which offer sufficiently high isotope ratio precision for the intended research work. Following this, corrections for mass fractionation/discrimination, validation of the analytical procedure including chemical sample preparation and complete uncertainty budgets are the most important issues for obtaining a metrologically valid procedure for isotope ratio determination. Only the application of such metrologically valid procedures enables the generation of traceable and comparable isotope data. To realize this suitable isotope and/or δ-reference materials are required, which currently are not sufficiently available for most isotope systems. Boron is given as an example, for which the situation regarding isotope and δ-reference materials is excellent. Boron may therefore serve as prototype for other isotope systems.
Isotope reference materials are needed to calibrate and validate analytical procedures used for the determination of isotope amount ratios, procedurally defined isotope ratios or so-called δ values. In contrast to the huge analytical progress in isotope ratio analytics, the production of isotope reference materials has not kept pace with the increasing needs of isotope analysts. Three representative isotope systems are used to explain the technical and non-technical difficulties and drawbacks, on one hand, and to demonstrate what can be achieved at its best, on the other hand. A clear statement is given that new isotope reference materials are needed to obtain traceable and thus comparable data, which is essential for all kinds of isotope research. The range of available isotope reference materials and δ reference materials should be increased and matrix reference materials certified for isotope compositions or δ values, which do not exist yet, should be provided.
The validation of a robust quantification procedure for EtG in hair using GC–NCI–MS is presented. Aqueous extraction is followed by complete lyophylization of the extract and derivatization with pentafluoropropionic anhydride (PFPA) under controlled temperature and duration. Clean-up of extracts was dispensable and standard single quadrupole MS displayed sufficient selectivity and sensitivity. The method displayed a wide linearity range and enabled LOD of 0.68 pg/mg, LOQ of 2.4 pg/mg, and precision below 8.12%. Since EtG was seen to display prolonged stability in the aqueous extracts and after derivatization with PFPA this straightforward procedure allows a routine throughput of large quantities of samples with little proneness to procedural scatter of results. The method was applied to demonstrate the homogeneity of two hair reference materials with mean EtG contents of 8.48 pg/mg and 22.0 pg/mg. Aside from the application in homogeneity studies of hair reference materials predominantly in the concentration range of 10–50 pg/mg the method was also designed for daily routine quantification of real-world sample with regard to drinking behavior assessment.
Dieser Artikel enthält aktuelle Forschungsergebnisse im Bereich des überwachungsgestützten Managements der Tragwerksintegrität. Das Management der Tragwerksintegrität hat zum Ziel, die Funktionalität eines Tragwerks während des gesamten Lebenszyklus durch Einhaltung der Zielzuverlässigkeiten sicherzustellen. Ausgehend von Forschungsergebnissen über die Charakteristik von Überwachungsverfahren und -daten im Rahmen von Zuverlässigkeitsanalysen, wird ein Ansatz vorgestellt, wie die erwarteten Kosten für das Management der Tragwerksintegrität durch Überwachungsverfahren optimiert werden können. Dazu wird eine Kosten-Nutzen-Analyse für den Lebenszyklus einer Gründungstruktur von Offshore-Windenergieanlagen durchgeführt. Im Rahmen der Bayes'schen Entscheidungstheorie werden durch eine Prä-posteriori-Entscheidungsanalyse die optimalen Entscheidungsparameter für den Entwurf eines Überwachungssystems bestimmt. Als Fallstudie wird ein Überwachungssystem für die Gründungsstruktur eines Multibrid-M5000-Prototyps konzipiert und die mögliche Reduktion der Risiken und erwarteten Betriebskosten quantifiziert. --------------------------------------------------------------------------------------------------------------------------------------------------------------------
This paper contains recent research results in the field of monitoring supported structural integrity management. The structural integrity management aims at ensuring the functioning of a structure throughout the life cycle by compliance with the target reliabilities. Building upon research results concerning the characteristics of monitoring techniques and data within the framework of reliability analyses, an approach is introduced for the optimisation of the expected structural integrity management costs by monitoring information. For this aim, a life cycle cost benefit analysis for offshore wind turbine support structures is formulated. The optimal monitoring decision parameters are then assessed utilizing the Bayesian pre-posterior decision theory. As a case study a monitoring system for a Multibrid M5000 offshore wind turbine prototype support structure is designed and possible reductions of the risks and expected costs are quantified.
Triple IDMS has been applied for the first time to the quantification of element concentrations. It has been compared with single and double IDMS obtained on the same sample set in order to evaluate the advantages and disadvantages of triple IDMS over single and double IDMS as an analytical reference procedure. The measurement results of single, double and triple IDMS are indistinguishable, considering rounding due to the individual measurement uncertainties. As expected, the relative expanded uncertainties (k = 2) achieved with double IDMS (0.08 %) are dramatically smaller than those obtained with single IDMS (1.4 %). Triple IDMS yields the smallest relative expanded uncertainties (k = 2, 0.077 %) unfortunately at the expense of a much higher workload. Nevertheless triple IDMS has the huge advantage that the isotope ratio of the spike does not need to be determined. Elements with high memory effects, highly enriched spikes or highest metrological requirements may be typical applications for triple IDMS.
Risikobasierte Überwachungsverfahren zur Lebensdauerverlängerung von bestehenden und neuen Bauwerken
(2011)
Diese Veröffentlichung enthält Konzepte zur Lebensdauerverlängerung von Bauwerken auf der Grundlage der Tragwerkszuverlässigkeitstheorie. Es werden aktuelle Forschungsergebnisse aus nationalen und europäischen Forschungsvorhaben vorgestellt. Den Schwerpunkt bildet die Verwendung von Bauwerksüberwachungsdaten unter Berücksichtigung von Messunsicherheiten zur Bestimmung der Tragwerkszuverlässigkeit. Dazu wird ein neuer Ansatz zur Bestimmung der A-posteriori Messunsicherheit vorgestellt. Es wird gezeigt, aufgrund welcher Eigenschaften der verwendeten probabilistischen Modelle die Zuverlässigkeit im Grenzzustand der Ermüdung in Bezug auf eine Lebensdauerverlängerung beeinflusst werden kann.
The measurement uncertainty is useful to estimate the confidence of analytical results. Nowadays, a result without the uncertainty statement cannot be considered reliable, but the scientific literature still lacks examples of the estimate of the measurement uncertainty. This paper presents a practical and reliable description of the measurement uncertainty estimation of the analytical determination of ethyl carbamate in cachaça by GC-IDMS. The isotope dilution technique (ID) associated with GC-MS was used to improve the accuracy. The uncertainty estimated corresponds to 10% of the mass fraction of ethyl carbamate (115 ± 11) ng/g, which is in agreement with ppb level.
Die Mikro- und Nanotechnologie gehört zu den Schlüsseltechnologien des 21. Jahrhunderts mit hohen Wachstumsprognosen, wie auch die im Auftrag des BMBF durchgeführte Studie “Nanotechnologie als wirtschaftlicher Wachstumsmarkt” von 2004 ausführlich darstellt. Aus diesem Trend resultiert ein steigender Bedarf an Messsystemen, die Nanostrukturen prozessnah bzw. im Fertigungsprozess charakterisieren können. Virtuelle Messgeräte liefern Erkenntnisse zur Entwicklung neuartiger Messsysteme, Analyse und Optimierung bestehender Verfahren sowie die Bestimmung der Messunsicherheit und modellbasierten Korrektur systematischer Fehler. Der virtuelle Messprozess umfasst neben dem Messmittel auch die Probe und die Wechselwirkungen zwischen beiden. In diesem Beitrag werden virtuelle Messgeräte vorgestellt sowie deren Anwendung diskutiert.