Filtern
Erscheinungsjahr
- 2014 (3)
Dokumenttyp
- Zeitschriftenartikel (3) (entfernen)
Referierte Publikation
- ja (3)
Schlagworte
- 1H (3) (entfernen)
Elimination of Artifacts in NMR SpectroscopY (EASY) is a simple but very effective tool to remove simultaneously any real NMR probe background signal, any spectral distortions due to deadtime ringdown effects and -specifically- severe acoustic ringing artifacts in NMR spectra of low-gamma nuclei. EASY enables and maintains quantitative NMR (qNMR) as only a single pulse (preferably 90°) is used for data acquisition. After the acquisition of the first scan (it contains the wanted NMR signal and the background/deadtime/ringing artifacts) the same experiment is repeated immediately afterwards before the T1 waiting delay. This second scan contains only the background/deadtime/ringing parts. Hence, the simple difference of both yields clean NMR line shapes free of artefacts.
In this Part I various examples for complete 1H, 11B, 13C, 19F probe background removal due to construction parts of the NMR probes are presented. Furthermore, 25Mg EASY of Mg(OH)2 is presented and this example shows how extremely strong acoustic ringing can be suppressed (more than a factor of 200) such that phase and baseline correction for spectra acquired with a single pulse is no longer a problem. EASY is also a step towards deadtime-free data acquisition as these effects are also canceled completely. EASY can be combined with any other NMR experiment, including 2D NMR, if baseline distortions are a big problem.
Medium resolution nuclear magnetic resonance (MR-NMR) spectroscopy is currently a fast developing field, which has an enormous potential to become an important analytical tool for reaction monitoring, in hyphenated techniques, and for systematic investigations of complex mixtures. The recent developments of innovative MR-NMR spectrometers are therefore remarkable due to their possible applications in quality control, education, and process monitoring. MR-NMR spectroscopy can beneficially be applied for fast, non-invasive, and volume integrating analyses under rough environmental conditions.
Within this study, a simple 1/16'' fluorinated ethylene propylene (FEP) tube with an ID of 0.04'' (1.02 mm) was used as a flow cell in combination with a 5 mm glass Dewar tube inserted into a benchtop MR-NMR spectrometer with a 1H Larmor frequency of 43.32 MHz and 40.68 MHz for 19F. For the first time, quasi-simultaneous proton and fluorine NMR spectra were recorded with a series of alternating 19F and 1H single scan spectra along the reaction time coordinate of a homogeneously catalysed esterification model reaction containing fluorinated compounds. The results were compared to quantitative NMR spectra from a hyphenated 500 MHz online NMR instrument for validation. Automation of handling, pre-processing, and analysis of NMR data becomes increasingly important for process monitoring applications of online NMR spectroscopy and for its technical and practical acceptance. Thus, NMR spectra were automatically baseline corrected and phased using the minimum entropy method. Data analysis schemes were designed such that they are based on simple direct integration or first principle line fitting, with the aim that the analysis directly revealed molar concentrations from the spectra.
Finally, the performance of 1/16'' FEP tube set-up with an ID of 1.02 mm was characterised regarding the limit of detection (LOQ (1H) = 0.335 mol L-1 and LOQ (19F) = 0.130 mol L-1 for trifluoroethanol in D2O (single scan)) and maximum quantitative flow rates up to 0.3 mL min-1. Thus, a series of single scan 19F and 1H NMR spectra acquired with this simple set-up already presents a valuable basis for quantitative reaction monitoring.
Die Entwicklung innovativer Medium-Resolution-NMR-Spektrometer eröffnet eine Reihe bemerkenswerter Applikationen für die Qualitätskontrolle, die akademische Ausbildung und die Prozessanalytik. Die Verwendung von Permanentmagneten erlaubt den Einsatz der NMR-Spektroskopie in industriellen Prozessumgebungen.