This is the final report on the proficiency test (PT) „Measurement of the particle size distribution of ceramic powders by laser diffraction”. The PT was organized by the Federal Institute for Materials Research and Testing (BAM), division 5.5 “Advanced technical ceramics”. The measurements took place from 12/2021 to 03/2022. The aim of the interlaboratory comparison was the proficiency assessment of the participating laboratories.
Management and realization of PT were performed in accordance with DIN EN ISO/IEC 17043:2010.
The basis of the harmonized measuring procedure and the instructions for the interlaboratory comparison was the standard ISO 13320:2020.
The statistical analysis was performed in accordance with DIN ISO 13528:2020 by use of the software PROLab Plus (QuoData GmbH, Dresden, Germany).
A group of 44 laboratories from 17 countries participated in the PT. Laser diffraction analyzers produced by 7 different manufacturers were used. The three test materials were commercial products. Overall, the measured characteristic values of the particle distributions (d10, d50 and d90) were in the size range between 0.5 and 25 μm. The final report contains all individual results in an anonymous way.
The Institute for Reference Materials and Measurements (IRMM) of the Joint Research Centre (JRC), a Directorate General of the European Commission, operates the International Measurement Evaluation Program (IMEP). IMEP organises inter-laboratory comparisons in support of European Union policies. This paper presents the results of two proficiency tests (PTs): IMEP-116 and IMEP-39, organised for the determination of total Cd, Pb, As, Hg and inorganic As (iAs) in mushrooms. Participation in IMEP-116 was restricted to National Reference Laboratories (NRLs) officially appointed by national authorities in European Union member states. IMEP-39 was open to all other laboratories wishing to participate. Thirty-seven participants from 25 countries reported results in IMEP-116, and 62 laboratories from 36 countries reported for the IMEP-39 study. Both PTs were organised in support to Regulation (EC) No. 1881/2006, which sets the maximum levels for certain contaminants in food. The test item used in both PTs was a blend of mushrooms of the variety shiitake (Lentinula edodes). Five laboratories, with demonstrated measurement capability in the field, provided results to establish the assigned values (Xref). The standard uncertainties associated to the assigned values (uref) were calculated by combining the uncertainty of the characterisation (uchar) with a contribution for homogeneity (ubb) and for stability (ust), whilst uchar was calculated following ISO 13528. Laboratory results were rated with z- and zeta (ζ)-scores in accordance with ISO 13528. The standard deviation for proficiency assessment, σp, ranged from 10% to 20% depending on the analyte. The percentage of satisfactory z-scores ranged from 81% (iAs) to 97% (total Cd) in IMEP-116 and from 64% (iAs) to 84% (total Hg) in IMEP-39.
An interlaboratory comparison for the determination of aniline, nine methylated anilines and chlorinated anilines in contaminated groundwater (each compound between 1 µg L-1 and 750 µg L-1) was conducted as proficiency test for the selection of contract laboratories for a groundwater monitoring campaign. For the 13 participants, two different test sample series were prepared from a stabilized real case groundwater. Series A was the groundwater as sampled in the field and series B was prepared from series A by spiking with selected anilines. Homogeneity and stability investigations revealed that contaminated groundwater provides a viable basis for the preparation of ring test samples for the determination of anilines. Analytical procedures were limited to the standardized liquid–liquid extraction or solid-phase extraction in combination with gas chromatography/mass spectrometry (DIN 38407-16:1999). The robust consensus values were evaluated according to the standardized protocol of DIN 38402-45:2013. Robust reproducibility standard deviations ranged largely between 20?% and 60?% depending on the analyte. The proficiency assessment of individual participants combined the qualitative aspect of correct peak identification with the quantitative determination of individual concentrations within set limits in a unified approach. It could be shown that the accreditation status of laboratory and the existence of a standardized analytical procedure do not substitute a problem-related proficiency assessment of potential contractors.