On-line monitoring of polyhydroxyalkanoate extraction process using compact NMR spectroscopy
- Portable benchtop NMR spectrometers enable real-time process and reaction monitoring in contrast to conventional laboratory based off-line gas chromatography or high-field NMR measurements. In this study, benchtop NMR spectroscopy is demonstrated as a process analytical technology (PAT) tool for the application of the solvent extraction step in downstream processing of polyhydroxyalkanoate (PHA) biopolymers. These are one of the few thermoplastic polymers synthesized 100 % via biotechnological routes which fully biodegrade in common natural environments. This makes them excellent candidate materials for sustainable replacement of conventional plastic materials.
Online NMR experiments were conducted using a fully automated setup, employing commercially available PTFE tubing as a flow-cell assembly. Extraction was carried out in a thermostated stirred batch reactor in lab-scale. Single-scan NMR spectra allowed continuous monitoring of the extraction of the PHA copolymerPortable benchtop NMR spectrometers enable real-time process and reaction monitoring in contrast to conventional laboratory based off-line gas chromatography or high-field NMR measurements. In this study, benchtop NMR spectroscopy is demonstrated as a process analytical technology (PAT) tool for the application of the solvent extraction step in downstream processing of polyhydroxyalkanoate (PHA) biopolymers. These are one of the few thermoplastic polymers synthesized 100 % via biotechnological routes which fully biodegrade in common natural environments. This makes them excellent candidate materials for sustainable replacement of conventional plastic materials.
Online NMR experiments were conducted using a fully automated setup, employing commercially available PTFE tubing as a flow-cell assembly. Extraction was carried out in a thermostated stirred batch reactor in lab-scale. Single-scan NMR spectra allowed continuous monitoring of the extraction of the PHA copolymer poly-(hydroxybutyrate-co-hydroxyhexanoate) containing 13.5 mol-% hydroxyhexanoate from Ralstonia eutropha biomass. Extractions were performed in chloroform and acetone across ground lyophilized cell loadings ranging from 20 to 120 g/L. The reproducibility and reliability of compact NMR spectroscopy closely matched with high-field NMR measurements. A strong correlation was observed between online low-field NMR data and offline gas chromatography (GC) analysis. The study highlights the versatility of compact NMR for process monitoring, facilitating endpoint determination and enhancing extraction efficiency by optimizing process parameters. Steady-state conditions were achieved within 6 to 10 minutes for chloroform and acetone, respectively, underscoring the method’s value in supporting downstream process development and optimization for PHA recovery.…


| Autor*innen: | Klas MeyerORCiD |
|---|---|
| Koautor*innen: | Isabel Thiele, Björn Weiske, Sebastian L. Riedel |
| Dokumenttyp: | Posterpräsentation |
| Veröffentlichungsform: | Präsentation |
| Sprache: | Englisch |
| Jahr der Erstveröffentlichung: | 2025 |
| Organisationseinheit der BAM: | 1 Analytische Chemie; Referenzmaterialien |
| 1 Analytische Chemie; Referenzmaterialien / 1.3 Instrumentelle Analytik | |
| DDC-Klassifikation: | Naturwissenschaften und Mathematik / Chemie / Analytische Chemie |
| Freie Schlagwörter: | Benchtop-NMR; Downstream processing; Polyhydroxyalkanoate; Process Analytical Technology |
| Themenfelder/Aktivitätsfelder der BAM: | Chemie und Prozesstechnik |
| Chemie und Prozesstechnik / Chemische Charakterisierung und Spurenanalytik | |
| Veranstaltung: | 20. Herbstkolloquium Arbeitskreis Prozessanalytik |
| Veranstaltungsort: | Frankfurt am Main, Germany |
| Beginndatum der Veranstaltung: | 03.12.2025 |
| Enddatum der Veranstaltung: | 05.12.2025 |
| Verfügbarkeit des Dokuments: | Datei im Netzwerk der BAM verfügbar ("Closed Access") |
| Datum der Freischaltung: | 10.12.2025 |
| Referierte Publikation: | Nein |