Trace-Level Ammonia–Water Interactions in Hydrogen: Challenges in Gas Purity Analysis Using Optical-Feedback Cavity-Enhanced Absorption Spectroscopy (OF-CEAS)

  • Ammonia is a critical impurity in hydrogen fuel due to its irreversible poisoning effect on proton exchange membrane fuel cells. Therefore, international standards (e.g., ISO 14687) set a stringent threshold of 100 nmol/mol. Furthermore, with the growing potential use of ammonia as a hydrogen carrier, its accurate quantification is becoming increasingly important. However, the presence of trace humidity poses analytical challenges, as ammonia may interact with water or interfaces, thereby affecting its detectability. Therefore, the goal of this work is to enable accurate trace ammonia quantification for hydrogen purity measurements through fundamental studies of the methodological challenges. Here, low-pressure sampling (ultra)long-path Optical-Feedback Cavity-Enhanced Absorption Spectroscopy (OF-CEAS) was applied with an effective optical path length of approximately 6.17 km. We studied three average amounts of ammonia: (38.2 ± 0.8) nmol/mol, (74.8 ± 0.7) nmol/mol, and (112.1 ± 1.2)Ammonia is a critical impurity in hydrogen fuel due to its irreversible poisoning effect on proton exchange membrane fuel cells. Therefore, international standards (e.g., ISO 14687) set a stringent threshold of 100 nmol/mol. Furthermore, with the growing potential use of ammonia as a hydrogen carrier, its accurate quantification is becoming increasingly important. However, the presence of trace humidity poses analytical challenges, as ammonia may interact with water or interfaces, thereby affecting its detectability. Therefore, the goal of this work is to enable accurate trace ammonia quantification for hydrogen purity measurements through fundamental studies of the methodological challenges. Here, low-pressure sampling (ultra)long-path Optical-Feedback Cavity-Enhanced Absorption Spectroscopy (OF-CEAS) was applied with an effective optical path length of approximately 6.17 km. We studied three average amounts of ammonia: (38.2 ± 0.8) nmol/mol, (74.8 ± 0.7) nmol/mol, and (112.1 ± 1.2) nmol/mol. Furthermore, these amounts were investigated at trace-humidity levels ranging from 0.8 to 8.5 ppmV. We observed a systematic, nonlinear, and humidity-dependent positive measurement bias of up to + (1.0 ± 0.2) nmol/mol at the maximum investigated trace-humidity volume fraction of 8.5 ppmV. This bias was not caused by spectral interference but rather by water-induced accumulation of ammonia within the optical cavity. Moreover, time-resolved measurements in the presence of trace ammonia showed that water desorption follows first-order kinetics, whereas water adsorption followed mixed-order kinetics with an apparent reaction order of 1.57 ± 0.03. Distinct hydration states of surface-bound ammonia were identified, whereas under dry conditions and with increasing amounts of ammonia, enhanced surface adhesion through intermolecular clustering was observed. In addition, the presence of ammonium species within the sorption layer was indirectly confirmed by our experiments. In conclusion, we provide a deeper insight into trace-level ammonia–water interactions and establish a framework for optimizing methodologies, particularly for (ultra)long-path optical gas measurement systems.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Mehmet Emin BayatORCiD, Heinrich KipphardtORCiD, Carlo TiebeORCiD, Dirk TumaORCiD, Carsten EngelhardORCiD
Dokumenttyp:Zeitschriftenartikel
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):ACS Measurement Science Au
Jahr der Erstveröffentlichung:2025
Organisationseinheit der BAM:1 Analytische Chemie; Referenzmaterialien
1 Analytische Chemie; Referenzmaterialien / 1.0 Abteilungsleitung und andere
1 Analytische Chemie; Referenzmaterialien / 1.6 Anorganische Referenzmaterialien und Gasanalytik
8 Zerstörungsfreie Prüfung
8 Zerstörungsfreie Prüfung / 8.1 Sensorik, mess- und prüftechnische Verfahren
Veröffentlichende Institution:Bundesanstalt für Materialforschung und -prüfung (BAM)
Verlag:American Chemical Society (ACS)
Erste Seite:1
Letzte Seite:15
DDC-Klassifikation:Technik, Medizin, angewandte Wissenschaften / Ingenieurwissenschaften / Angewandte Physik
Freie Schlagwörter:Adsorption; Ammonia; Humidity; Hydrogen; OF-CEAS; Surface Interactions
Themenfelder/Aktivitätsfelder der BAM:Energie
Energie / Wasserstoff
DOI:10.1021/acsmeasuresciau.5c00105
URN:urn:nbn:de:kobv:b43-643627
ISSN:2694-250X
Verfügbarkeit des Dokuments:Datei für die Öffentlichkeit verfügbar ("Open Access")
Lizenz (Deutsch):License LogoCreative Commons - CC BY - Namensnennung 4.0 International
Datum der Freischaltung:15.10.2025
Referierte Publikation:Ja
Datum der Eintragung als referierte Publikation:27.10.2025
Schriftenreihen ohne Nummerierung:Wissenschaftliche Artikel der BAM
Einverstanden
Diese Webseite verwendet technisch erforderliche Session-Cookies. Durch die weitere Nutzung der Webseite stimmen Sie diesem zu. Unsere Datenschutzerklärung finden Sie hier.