• Treffer 1 von 2
Zurück zur Trefferliste

Improved Data Processing for Accurate Plasma Diagnostics with Implications for Calibration-Free LIBS

  • Many LIBS papers report the determination of plasma temperature using the Boltzmann plot method or the determination of electron density using the Stark line broadening relation. This requires measuring the integrated intensities of the spectral lines and the linewidth under the assumption of optical thinness. It is taken for granted that this can be easily done either by working with the raw spectra or by fitting the appropriate function to the observed spectral lines. However, extracting the necessary information from raw spectral data is not as easy as it might seem. The quality of such extraction will depend to a decisive extent on the type of spectral instrument used. The spectrum emitted by the plasma is distorted by the device; an example is shown in Fig. 1. The elimination of this distortion belongs to the class of inverse problems, the successful solution of which fundamentally depends on the quality of the available information. When it comes to spectroscopy, the quality ofMany LIBS papers report the determination of plasma temperature using the Boltzmann plot method or the determination of electron density using the Stark line broadening relation. This requires measuring the integrated intensities of the spectral lines and the linewidth under the assumption of optical thinness. It is taken for granted that this can be easily done either by working with the raw spectra or by fitting the appropriate function to the observed spectral lines. However, extracting the necessary information from raw spectral data is not as easy as it might seem. The quality of such extraction will depend to a decisive extent on the type of spectral instrument used. The spectrum emitted by the plasma is distorted by the device; an example is shown in Fig. 1. The elimination of this distortion belongs to the class of inverse problems, the successful solution of which fundamentally depends on the quality of the available information. When it comes to spectroscopy, the quality of information primarily means high spectral resolution and low noise. Not all spectrometers used in LIBS can provide the quality needed to solve the inverse problem; this casts doubt on many published plasma measurements. This communication will be devoted to the general shortcomings of spectral data processing and the inaccuracies in determining the plasma parameters resulting from these shortcomings. The analysis is based on the use of synthetic spectra generated by plasma with known temperature, particle density and electron density. The estimation of errors caused by inadequate processing of spectral data is made by comparing the initial and measured plasma parameters from the spectra. Recommendations will be made for which analytic function best approximates the observed spectral lines, and how data processing errors affect the accuracy of calibration-free LIBS will be discussed. These issues were only partially covered in previously published works, for example [1, 2].zeige mehrzeige weniger

Volltext Dateien herunterladen

  • Gornushkin CSI Tokushima Jun 22 2023.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Igor Gornushkin
Dokumenttyp:Vortrag
Veröffentlichungsform:Präsentation
Sprache:Englisch
Jahr der Erstveröffentlichung:2023
Organisationseinheit der BAM:1 Analytische Chemie; Referenzmaterialien
1 Analytische Chemie; Referenzmaterialien / 1.3 Instrumentelle Analytik
DDC-Klassifikation:Naturwissenschaften und Mathematik / Chemie / Analytische Chemie
Freie Schlagwörter:Calibration-free LIBS; Data processing; Laser induced breakdown spectroscopy
Themenfelder/Aktivitätsfelder der BAM:Chemie und Prozesstechnik
Veranstaltung:Colloquium Spectroscopicum Internationale XLIII, The 5th Asian Symposium on Laser Induced Breakdown Spectroscopy June 26-30, 2023
Veranstaltungsort:Tokushima, Japan
Beginndatum der Veranstaltung:26.06.2023
Enddatum der Veranstaltung:30.06.2023
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")
Datum der Freischaltung:17.10.2023
Referierte Publikation:Nein
Eingeladener Vortrag:Nein