Multi-element analysis in different matrices using nitrogen microwave inductively coupled atmospheric pressure plasma mass spectrometry (MICAP-MS)

  • Inductively coupled plasma mass spectrometry (ICP-MS) emerged as a powerful technique for trace analysis of soil due to its multi-element capability, high sensitivity and low sample consumption. However, despite its success and widespread use, ICP-MS has several persistent drawbacks, such as high argon gas consumption, argon-based polyatomic interferences and the need for complicated RF-power generators. Unlike argon-based ICP, the nitrogen microwave inductively coupled atmospheric pressure mass spectrometry (MICAP-MS) uses nitrogen as plasma gas, which eliminates high operating costs associated with argon-gas consumption as well as the argon-based interferences1. In this work, the applicability of MICAP-MS for elemental analysis in different matrices is investigated. For this purpose, reference soil samples and steel samples are digested with aqua regia and used for analysis. Concentrations of selected elements are determined using MICAP-MS and validated with ICP-MS und certifiedInductively coupled plasma mass spectrometry (ICP-MS) emerged as a powerful technique for trace analysis of soil due to its multi-element capability, high sensitivity and low sample consumption. However, despite its success and widespread use, ICP-MS has several persistent drawbacks, such as high argon gas consumption, argon-based polyatomic interferences and the need for complicated RF-power generators. Unlike argon-based ICP, the nitrogen microwave inductively coupled atmospheric pressure mass spectrometry (MICAP-MS) uses nitrogen as plasma gas, which eliminates high operating costs associated with argon-gas consumption as well as the argon-based interferences1. In this work, the applicability of MICAP-MS for elemental analysis in different matrices is investigated. For this purpose, reference soil samples and steel samples are digested with aqua regia and used for analysis. Concentrations of selected elements are determined using MICAP-MS and validated with ICP-MS und certified values. Sensitivities, limits of detection and gas consumption for both methods are compared and discussed in detail. Performance of MICAP-MS under different nitrogen plasma gas concentrations is investigated and compared. Moreover, the performance of MICAP-MS in alloy matrices is investigated and discussed.zeige mehrzeige weniger

Volltext Dateien herunterladen

  • EWCPS2023.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Zengchao You
Koautor*innen:Alexander Winckelmann, Janina Roik, Wolfram Weisheit, Sebastian Recknagel, Carlos Enrique Abad Andrade
Dokumenttyp:Vortrag
Veröffentlichungsform:Präsentation
Sprache:Englisch
Jahr der Erstveröffentlichung:2023
Organisationseinheit der BAM:1 Analytische Chemie; Referenzmaterialien
1 Analytische Chemie; Referenzmaterialien / 1.6 Anorganische Referenzmaterialien
DDC-Klassifikation:Naturwissenschaften und Mathematik / Chemie / Analytische Chemie
Freie Schlagwörter:Multi-element analysis; Nitrogen plasma; Soil; Steel
Microwave inductively coupled atmospheric pressure mass spectrometry (MICAP-MS)
Themenfelder/Aktivitätsfelder der BAM:Chemie und Prozesstechnik
Chemie und Prozesstechnik / Chemische Charakterisierung und Spurenanalytik
Veranstaltung:EWCPS 2023
Veranstaltungsort:Ljubljana, Slovenia
Beginndatum der Veranstaltung:29.01.2023
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")