Atomic absorption spectroscopy and machine learning for lithium isotopic study
- The isotopic analysis of lithium is also relevant to the study of geological phenomena.1 In this work we propose improvements to the method for the isotopic analysis of lithium using a high-resolution continuum source atomic absorption spectrometer (HR-CS-AAS) coupled to a double echelle modular spectrometer (DEMON). 2 This tool for isotopic ratio determination is based on monitoring the isotopic components of lithium by their spin-orbit coupling and its isotopic shift of about 15 pm for the 22P←2 2S electronic transition around 670.788 nm. The data analysis was carried out by using a decision-tree-based ensemble machine learning (ML) algorithm. For the training of the algorithm (XGBoost), a set of samples with 6Li isotope amount fractions ranging from 0.06 to 0.99 mol-1 was used. Subsequently, the procedure was validated of a set of stock chemicals (Li2CO3, LiNO3, LiCl and LiOH) and a BAM candidate reference material, the cathode material LiNi1/3Mn1/3Co1/3O2 (NMC111). Finally, the MLThe isotopic analysis of lithium is also relevant to the study of geological phenomena.1 In this work we propose improvements to the method for the isotopic analysis of lithium using a high-resolution continuum source atomic absorption spectrometer (HR-CS-AAS) coupled to a double echelle modular spectrometer (DEMON). 2 This tool for isotopic ratio determination is based on monitoring the isotopic components of lithium by their spin-orbit coupling and its isotopic shift of about 15 pm for the 22P←2 2S electronic transition around 670.788 nm. The data analysis was carried out by using a decision-tree-based ensemble machine learning (ML) algorithm. For the training of the algorithm (XGBoost), a set of samples with 6Li isotope amount fractions ranging from 0.06 to 0.99 mol-1 was used. Subsequently, the procedure was validated of a set of stock chemicals (Li2CO3, LiNO3, LiCl and LiOH) and a BAM candidate reference material, the cathode material LiNi1/3Mn1/3Co1/3O2 (NMC111). Finally, the ML model was applied to the set of geological samples, previously digested, for the determination of their isotope ratio. The optical resolution was improved from 140,000 to 790,000 to better deconvolution the lithium isotopic components in the atomic spectrum. And the method was compared with multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS). The results are metrologically comparable.…
Autor*innen: | Dalia Morcillo |
---|---|
Koautor*innen: | Carlos Enrique Abad AndradeORCiD, Alexander WinckelmannORCiD, D. Frick, L. Jacobsen, Silke Richter, Sebastian RecknagelORCiD, Jochen VoglORCiD, Ulrich PanneORCiD |
Dokumenttyp: | Posterpräsentation |
Veröffentlichungsform: | Präsentation |
Sprache: | Englisch |
Jahr der Erstveröffentlichung: | 2022 |
Organisationseinheit der BAM: | 1 Analytische Chemie; Referenzmaterialien |
1 Analytische Chemie; Referenzmaterialien / 1.6 Anorganische Referenzmaterialien | |
DDC-Klassifikation: | Naturwissenschaften und Mathematik / Chemie / Analytische Chemie |
Freie Schlagwörter: | Atomic absorption spectroscopy; Lithium isotope |
Themenfelder/Aktivitätsfelder der BAM: | Chemie und Prozesstechnik |
Chemie und Prozesstechnik / Chemische Charakterisierung und Spurenanalytik | |
Veranstaltung: | Colloquium Spectroscopicum Internationale XLII (CSI XLII) |
Veranstaltungsort: | Gijón, Spain |
Beginndatum der Veranstaltung: | 30.05.2022 |
Verfügbarkeit des Dokuments: | Datei im Netzwerk der BAM verfügbar ("Closed Access") |
Datum der Freischaltung: | 23.11.2022 |
Referierte Publikation: | Nein |