Demands for improvement of temperature stability of extensometers

  • The presentation describes why the issue of thermal drift of extensometers has become important. It shows the possible sources of thermal drift of extensometers. It shows with a simple theoretical example calculation that the permitted strain error in standards can be exceeded if the temperature fluctuation at the extensometer is too high, but this effect remains mostly unknown because of missing data of the extensometer temperature throughout the test. This will be discussed in fatigue standard committees with the aim to improve the quality of fatigue testing results.

Volltext Dateien herunterladen

  • demands for improvement of extensometer stability.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Hellmuth Klingelhöffer
Dokumenttyp:Vortrag
Veröffentlichungsform:Präsentation
Sprache:Englisch
Jahr der Erstveröffentlichung:2017
DDC-Klassifikation:Technik, Medizin, angewandte Wissenschaften
Freie Schlagwörter:fatigue standard committee; temperature fluctuation of extensometers
Veranstaltung:8th Int Conf on LCF (LCF8)
Veranstaltungsort:Dresden, Germany
Beginndatum der Veranstaltung:27.06.2017
Enddatum der Veranstaltung:29.06.2017
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")
Datum der Freischaltung:26.11.2018
Referierte Publikation:Nein
Eingeladener Vortrag:Nein
Einverstanden