• Treffer 1 von 1
Zurück zur Trefferliste
Zitieren Sie bitte immer diesen URN: urn:nbn:de:kobv:b43-531315

Observation of systematic deviations between Faraday cup aerosol electrometers for varying particle sizes and flow rates - results of the AEROMET FCAE workshop

  • Condensation particle counters (CPCs) are widely used for the measurement of aerosol particle number concentrations in the size range from approximately 3 nm to 3 μm. For an SI-traceable calibration of the size-dependent counting efficiency, which is advisable on a regular basis and required in several applications, Faraday cup aerosol electrometers (FCAEs) are considered to be a suitable SI-traceable reference.While the volumetric aerosol inlet flowrate and the electrical current measurement in FCAEs can be related to respective SI references, inter-comparison exercises for FCAEs are still performed on a regular basis to establish reliable uncertainty budgets and to further investigate the influences of designs and operational parameters on comparability. This is strongly demanded in the international community of metrological institutes and aerosol calibration facilities around the world, which provide CPC calibrations. In the present study, the performance of FCAEs wasCondensation particle counters (CPCs) are widely used for the measurement of aerosol particle number concentrations in the size range from approximately 3 nm to 3 μm. For an SI-traceable calibration of the size-dependent counting efficiency, which is advisable on a regular basis and required in several applications, Faraday cup aerosol electrometers (FCAEs) are considered to be a suitable SI-traceable reference.While the volumetric aerosol inlet flowrate and the electrical current measurement in FCAEs can be related to respective SI references, inter-comparison exercises for FCAEs are still performed on a regular basis to establish reliable uncertainty budgets and to further investigate the influences of designs and operational parameters on comparability. This is strongly demanded in the international community of metrological institutes and aerosol calibration facilities around the world, which provide CPC calibrations. In the present study, the performance of FCAEs was investigated,using Ag test aerosol particles with a 30 nm particle diameter by varying the inlet flowrates from 0.5 l min−1 to 4 l min−1. From our experimental results, significant deviations were observed in FCAE currents at sample flowrates smaller than 1.5 l min−1. It is recommended that these discrepancies should be quantified before an FCAE is used for CPC calibration at low sample flowrates and small particle sizes in the sub-30 nm size range.zeige mehrzeige weniger

Volltext Dateien herunterladen

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:A. Ahlawat, Stefan SeegerORCiD, Martin Gottschalk, T. Tuch, A. Wiedensohler
Dokumenttyp:Zeitschriftenartikel
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):Metrologia
Jahr der Erstveröffentlichung:2021
Organisationseinheit der BAM:4 Material und Umwelt
4 Material und Umwelt / 4.2 Material-Mikrobiom Wechselwirkungen
Veröffentlichende Institution:Bundesanstalt für Materialforschung und -prüfung (BAM)
Verlag:Institute of Physics (IOP)
Verlagsort:Bristol, UK
Jahrgang/Band:58
Ausgabe/Heft:5
Aufsatznummer:055006
Erste Seite:1
Letzte Seite:8
DDC-Klassifikation:Technik, Medizin, angewandte Wissenschaften / Ingenieurwissenschaften / Sanitär- und Kommunaltechnik; Umwelttechnik
Freie Schlagwörter:Aerosol measurement instruments; CPC calibration; Faraday cup aerosol electrometer; Inter-comparison
Themenfelder/Aktivitätsfelder der BAM:Umwelt
Umwelt / Umwelt-Material-Interaktionen
DOI:10.1088/1681-7575/ac0710
URN:urn:nbn:de:kobv:b43-531315
ISSN:1681-7575
ISSN:0026-1394
Verfügbarkeit des Dokuments:Datei für die Öffentlichkeit verfügbar ("Open Access")
Lizenz (Deutsch):License LogoCreative Commons - CC BY - Namensnennung 4.0 International
Datum der Freischaltung:24.08.2021
Referierte Publikation:Ja
Datum der Eintragung als referierte Publikation:20.09.2021
Schriftenreihen ohne Nummerierung:Wissenschaftliche Artikel der BAM
Einverstanden