Determination of the total purity of a high-purity silver material to be used as a primary standard for element determination
- A candidate material for the use as primary standard for silver determination was characterized with respect to total purity. Except the radioactive elements and He, all possible impurities were considered. Based on glow discharge mass spectrometry, inductively coupled plasma-mass spectrometry and carrier gas hot extraction measurements, the demonstrated total purity and its standard uncertainty is w(Ag) = (99.999 52 ± 0.000 11) %. The purity value and its uncertainty is dominated by the contributions from the measurements of the nonmetallic impurities, namely S, N, C and O.
Autor*innen: | T. Zhou, Silke RichterORCiD, Ralf MatschatORCiD, Heinrich Kipphardt |
---|---|
Dokumenttyp: | Zeitschriftenartikel |
Veröffentlichungsform: | Verlagsliteratur |
Sprache: | Englisch |
Titel des übergeordneten Werkes (Englisch): | Accreditation and quality assurance |
Jahr der Erstveröffentlichung: | 2013 |
Verlag: | Springer |
Verlagsort: | Berlin |
Jahrgang/Band: | 18 |
Ausgabe/Heft: | 4 |
Erste Seite: | 341 |
Letzte Seite: | 349 |
Freie Schlagwörter: | Element determination; Primary standard; Purity; Silver; Total purity |
DOI: | 10.1007/s00769-013-0988-5 |
ISSN: | 0949-1775 |
ISSN: | 1432-0517 |
Verfügbarkeit des Dokuments: | Datei im Netzwerk der BAM verfügbar ("Closed Access") |
Datum der Freischaltung: | 20.02.2016 |
Referierte Publikation: | Ja |
Datum der Eintragung als referierte Publikation: | 26.08.2013 |