On-site calibration system for trace humidity sensors

  • A portable device for calibration of trace humidity sensors and an adopted calibration procedure have been developed. The calibration device is based on humidity generation by permeating water through polymeric membrane tubes. Water vapour transmission rates for various polymers were experimentally determined in order to select the most suitable polymeric material. The developed trace humidity generator consists of a gas-flow polymeric hose immersed in a water reservoir thermostated by a sensorcontrolled heater. Mole fractions of water vapour between 1 µmol/ molˉ¹ and 350 µmol/molˉ¹ (equivalent to frost-point temperatures from -76 °C to -31 °C) were generated by varying either the operating temperature or gas flow. The operating temperature can be varied from 20 °C to 60 °C and kept stable within 0.1 K. Uncertainty analysis indicated that the trace humidity generator produces gas flows of constant humidity amounts with a relative expanded uncertainty less than 3.4% (k = 2) of theA portable device for calibration of trace humidity sensors and an adopted calibration procedure have been developed. The calibration device is based on humidity generation by permeating water through polymeric membrane tubes. Water vapour transmission rates for various polymers were experimentally determined in order to select the most suitable polymeric material. The developed trace humidity generator consists of a gas-flow polymeric hose immersed in a water reservoir thermostated by a sensorcontrolled heater. Mole fractions of water vapour between 1 µmol/ molˉ¹ and 350 µmol/molˉ¹ (equivalent to frost-point temperatures from -76 °C to -31 °C) were generated by varying either the operating temperature or gas flow. The operating temperature can be varied from 20 °C to 60 °C and kept stable within 0.1 K. Uncertainty analysis indicated that the trace humidity generator produces gas flows of constant humidity amounts with a relative expanded uncertainty less than 3.4% (k = 2) of the generated value.zeige mehrzeige weniger

Volltext Dateien herunterladen

  • Measurement2016.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Thomas Hübert, Carlo Tiebe, Marc Detjens, J. Majewski
Dokumenttyp:Zeitschriftenartikel
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):Measurement
Jahr der Erstveröffentlichung:2016
Verlag:Elsevier Ltd.
Jahrgang/Band:91
Ausgabe/Heft:9
Erste Seite:251
Letzte Seite:257
DDC-Klassifikation:Naturwissenschaften und Mathematik / Chemie / Analytische Chemie
Freie Schlagwörter:Measurements; Polymeric tubes; Sensor calibration; Trace Humidity; Water vapour permeation
DOI:10.1016/j.measurement.2016.05.013
ISSN:0263-2241
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")
Datum der Freischaltung:06.07.2016
Referierte Publikation:Ja
Datum der Eintragung als referierte Publikation:05.08.2016
Einverstanden
Diese Webseite verwendet technisch erforderliche Session-Cookies. Durch die weitere Nutzung der Webseite stimmen Sie diesem zu. Unsere Datenschutzerklärung finden Sie hier.