Matrix segregation as the major cause for sample inhomogeneity in MALDI dried droplet spots
- The segregation in dried droplet MALDI sample spots was analyzed with regard to the matrix-to-sample ratio using optical microscopy, MALDI imaging mass spectrometry (MALDI MSI) and IR imaging spectroscopy. In this context, different polymer/matrix/solvent systems usually applied in the analysis of synthetic polymers were investigated. The use of typical matrix concentrations (10 mg mL-1) in almost every case resulted in ring patterns, whereas higher concentrated matrix solutions always led to homogeneous sample spot layers. The data revealed that segregation is predominantly caused by matrix transport in the drying droplet, whereas polymer segregation seems to be only secondary.
Autor*innen: | Stefan Gabriel, C. Schwarzinger, B. Schwarzinger, Ulrich PanneORCiD, Steffen WeidnerORCiD |
---|---|
Dokumenttyp: | Zeitschriftenartikel |
Veröffentlichungsform: | Verlagsliteratur |
Sprache: | Englisch |
Titel des übergeordneten Werkes (Englisch): | Journal of the American society for mass spectrometry |
Jahr der Erstveröffentlichung: | 2014 |
Verlag: | American Society for Mass Spectrometry |
Verlagsort: | New York, NY |
Jahrgang/Band: | 25 |
Ausgabe/Heft: | 8 |
Erste Seite: | 1356 |
Letzte Seite: | 1363 |
Freie Schlagwörter: | Dried droplet; Imaging MS; MALDI; Matrix segregation; Polymers; Sample preparation |
DOI: | 10.1007/s13361-014-0913-0 |
ISSN: | 1044-0305 |
Verfügbarkeit des Dokuments: | Datei im Netzwerk der BAM verfügbar ("Closed Access") |
Datum der Freischaltung: | 20.02.2016 |
Referierte Publikation: | Ja |
Datum der Eintragung als referierte Publikation: | 19.02.2015 |