Determination of the effective EDS detector area using experimental and theoretical X-ray emission yields

  • An energy dispersive X-ray spectrometer operating with a semiconductor detector should be specified in compliance with the ISO standard 15632 [1]. Requirements for specification are: a) a general description of the spectrometer to evaluate its performance, b) the energy resolution with corresponding dead time, c) the P/B ratio in the Fe55 spectrum and, finally, d) the L/K intensity Ratio in a Ni or Cu spectrum to estimate spectrometer efficiency at low energies. Items b) to d) can be easily checked by the user. Related procedures are recommended in the annexes of the standard.

Volltext Dateien herunterladen

  • determination_of_the_effective_eds_detector_area_using_experimental_and_theoretical_xray_emission_yields.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Vasile-Dan HodoroabaORCiD, R. Terborg, M. Procop
Dokumenttyp:Zeitschriftenartikel
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):Microscopy and microanalysis
Jahr der Erstveröffentlichung:2015
Herausgeber (Institution):Microscopy Society of America
Verlag:Cambridge University Press
Verlagsort:New York, NY
Jahrgang/Band:21
Ausgabe/Heft:Suppl. 3
Erste Seite:Paper 0739, 1481
Letzte Seite:1482
Freie Schlagwörter:EDS; Effective solid angle; X-rays
DOI:10.1017/S1431927615008181
ISSN:1431-9276
ISSN:1435-8115
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")
Datum der Freischaltung:20.02.2016
Referierte Publikation:Ja
Datum der Eintragung als referierte Publikation:05.10.2015