• Treffer 1 von 1
Zurück zur Trefferliste

Comparison of ppb-level gas measurements with a metal-oxide semiconductor gas sensor in two independent laboratories

  • In this work, we use a gas sensor system consisting of a commercially available gas sensor in temperature cycled operation. It is trained with an extensive gas profile for detection and quantification of hazardous volatile organic compounds (VOC) in the ppb range independent of a varying background of other, less harmful VOCs and inorganic interfering gases like humidity or hydrogen. This training was then validated using a different gas mixture generation apparatus at an independent lab providing analytical methods as reference. While the varying background impedes selective detection of benzene and naphthalene at the low concentrations supplied, both formaldehyde and total VOC can well be quantified, after calibration transfer, by models trained with data from one system and evaluated with data from the other system. The lowest achievable root mean squared errors of prediction were 49 ppb for formaldehyde (in a concentration range of 20–200 ppb) and 150 μg/m³ (in a concentrationIn this work, we use a gas sensor system consisting of a commercially available gas sensor in temperature cycled operation. It is trained with an extensive gas profile for detection and quantification of hazardous volatile organic compounds (VOC) in the ppb range independent of a varying background of other, less harmful VOCs and inorganic interfering gases like humidity or hydrogen. This training was then validated using a different gas mixture generation apparatus at an independent lab providing analytical methods as reference. While the varying background impedes selective detection of benzene and naphthalene at the low concentrations supplied, both formaldehyde and total VOC can well be quantified, after calibration transfer, by models trained with data from one system and evaluated with data from the other system. The lowest achievable root mean squared errors of prediction were 49 ppb for formaldehyde (in a concentration range of 20–200 ppb) and 150 μg/m³ (in a concentration range of 25–450 μg/m³) for total VOC. The latter uncertainty improves to 13 μg/m³ with a more confined model range of 220–320 μg/m³. The data from the second lab indicate an interfering gas which cannot be detected analytically but strongly influences the sensor signal. This demonstrates the need to take into account all sensor relevant gases, like, e.g., hydrogen and carbon monoxide, in analytical reference measurements.zeige mehrzeige weniger

Volltext Dateien herunterladen

  • Bastuck et al-2018-Comparison of ppb-level gas measurements with a metal-oxide semiconductor gas sensor in two independent laboratories.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:M. Bastuck, T. Baur, Matthias RichterORCiD, B. Mull, A. Schütze, T. Sauerwald
Dokumenttyp:Zeitschriftenartikel
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):Sensors and Actuators B: Chemical
Jahr der Erstveröffentlichung:2018
Organisationseinheit der BAM:4 Material und Umwelt
4 Material und Umwelt / 4.2 Material-Mikrobiom Wechselwirkungen
Verlag:Elsevier B.V.
Jahrgang/Band:273
Erste Seite:1037
Letzte Seite:1046
DDC-Klassifikation:Naturwissenschaften und Mathematik / Chemie / Analytische Chemie
Freie Schlagwörter:Calibration transfer; Indoor air quality; Inter-lab comparison; Selective quantification; Volatile organic compounds
Themenfelder/Aktivitätsfelder der BAM:Umwelt
Umwelt / Sensorik
DOI:10.1016/j.snb.2018.06.097
ISSN:0925-4005