Comparison of peristaltic pumps used for sample introduction in Inductively Coupled Plasma-Atomic Emission Spectroscopy (ICP-AES)
- In this investigation, two conventional peristaltic pumps are compared with a new pump based on the “easy click” principle using a simultaneous ICP-AES instrument with standard operating conditions. It is found that the figures of merit achieved are quite comparable for all three pumps. Relative standard deviations (RSDs) range between 0.2% and 1.8%, and limits of detection as low as 0.1 μg/L have been achieved , demonstrating that the easy click principle of the new pump does not compromise the analytical figures of merit.
Autor*innen: | N. Jakubowski, S. Borrmann, Sebastian RecknagelORCiD, Janina Roik, F. Rickert |
---|---|
Dokumenttyp: | Zeitschriftenartikel |
Veröffentlichungsform: | Verlagsliteratur |
Sprache: | Englisch |
Titel des übergeordneten Werkes (Englisch): | Spectroscopy |
Jahr der Erstveröffentlichung: | 2020 |
Organisationseinheit der BAM: | 1 Analytische Chemie; Referenzmaterialien |
1 Analytische Chemie; Referenzmaterialien / 1.6 Anorganische Referenzmaterialien | |
Ausgabe/Heft: | 35 / S4 |
Erste Seite: | 6 |
Freie Schlagwörter: | ICP-AES; ICP-MS; Peristaltic pumps |
Themenfelder/Aktivitätsfelder der BAM: | Chemie und Prozesstechnik |
Chemie und Prozesstechnik / Chemische Charakterisierung und Spurenanalytik | |
Verfügbarkeit des Dokuments: | Datei im Netzwerk der BAM verfügbar ("Closed Access") |
Datum der Freischaltung: | 20.01.2021 |
Referierte Publikation: | Nein |