• Treffer 2 von 0
Zurück zur Trefferliste

A new experimental setup for time- and laterally-resolved X-ray absorption fine structure spectroscopy in a 'single shot'

  • In this work, a new setup for dispersive XAFS measurements is presented. This reproducible and scanningfree setup yields both time- and laterally-resolved XAFS experiments in a ‘single-shot’. It allows a straightforward adjustment for probing different elements covering many relevant applications in materials science. An incoming energetic broadband beam is diffracted by a Si (111) crystal after passing through the sample and collected by an area sensitive detector. Depending on the energy range of the incoming beam, XANES and/or EXAFS spectra can be recorded with a time resolution down to 1 s. The feasibility of this setup was demonstrated at the BAMline at BESSY II (Berlin, Germany) with reference Fe and Cu foils and the results are hereby presented and discussed. Additionally, an application where time resolution on the second scale is required is briefly evaluated. The presented example concerns studying early stages of zinc(II)2-methylimidazolate (ZIF-8) crystallization. This isIn this work, a new setup for dispersive XAFS measurements is presented. This reproducible and scanningfree setup yields both time- and laterally-resolved XAFS experiments in a ‘single-shot’. It allows a straightforward adjustment for probing different elements covering many relevant applications in materials science. An incoming energetic broadband beam is diffracted by a Si (111) crystal after passing through the sample and collected by an area sensitive detector. Depending on the energy range of the incoming beam, XANES and/or EXAFS spectra can be recorded with a time resolution down to 1 s. The feasibility of this setup was demonstrated at the BAMline at BESSY II (Berlin, Germany) with reference Fe and Cu foils and the results are hereby presented and discussed. Additionally, an application where time resolution on the second scale is required is briefly evaluated. The presented example concerns studying early stages of zinc(II)2-methylimidazolate (ZIF-8) crystallization. This is particularly important for biomedical applications.zeige mehrzeige weniger

Volltext Dateien herunterladen

  • S2XAFS.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Anicó KulowORCiD, S. Witte, S. Beyer, Ana de Oliveira Guilherme BuzanichORCiD, Martin RadtkeORCiD, Uwe Reinholz, H. Riesemeier, C. Streli
Dokumenttyp:Zeitschriftenartikel
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):Journal of Analytical Atomic Spectrometry
Jahr der Erstveröffentlichung:2019
Organisationseinheit der BAM:6 Materialchemie
6 Materialchemie / 6.3 Strukturanalytik
Verlag:Royal Society of Chemistry
Verlagsort:London
Jahrgang/Band:34
Ausgabe/Heft:1
Erste Seite:239
Letzte Seite:246
DDC-Klassifikation:Naturwissenschaften und Mathematik / Chemie / Analytische Chemie
Freie Schlagwörter:Experimental setup; Laterally-resolved; Time-resolved; X-ray absorption fine structure; X-ray spectroscopy
Themenfelder/Aktivitätsfelder der BAM:Chemie und Prozesstechnik
DOI:10.1039/c8ja00313k
ISSN:0267-9477
ISSN:1364-5544
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")
Datum der Freischaltung:22.01.2019
Referierte Publikation:Ja
Datum der Eintragung als referierte Publikation:22.01.2019
Einverstanden