Comparison of different calibration strategies for the analysis of zinc and other pure metals by using the GD-MS instruments VG 9000 and element GD
- Synthetic pressed metal powder standards doped with standard solutions were used for the calibration of both commercially available high resolution GD-MS instruments Element GD and VG 9000 for zinc matrix. Different quantification procedures (IBR, Standard RSF, matrix matched RSF from the calibration with CRMs and use of doped synthetic standards) are compared using zinc matrix as an example, whereas the calibration with doped pellets turned out to be the best quantification technique for high-purity materials. The applicability of the Standard RSF concept is scrutinised. In this context, RSF values for several Matrices (Co, Cu, Fe, In and Zn) are reported additionally.
Autor*innen: | Tamara Gusarova, T. Hofmann, Heinrich Kipphardt, C. Venzago, Ralf MatschatORCiD, Ulrich PanneORCiD |
---|---|
Dokumenttyp: | Zeitschriftenartikel |
Veröffentlichungsform: | Verlagsliteratur |
Sprache: | Englisch |
Titel des übergeordneten Werkes (Englisch): | Journal of analytical atomic spectrometry |
Jahr der Erstveröffentlichung: | 2010 |
Verlag: | Royal Society of Chemistry |
Verlagsort: | London |
Jahrgang/Band: | 25 |
Erste Seite: | 314 |
Letzte Seite: | 321 |
DOI: | 10.1039/b921649a |
ISSN: | 0267-9477 |
ISSN: | 1364-5544 |
Verfügbarkeit des Dokuments: | Physisches Exemplar in der Bibliothek der BAM vorhanden ("Hardcopy Access") |
Bibliotheksstandort: | Sonderstandort: Publica-Schrank |
Datum der Freischaltung: | 19.02.2016 |
Referierte Publikation: | Ja |
Datum der Eintragung als referierte Publikation: | 18.11.2010 |