• Treffer 1 von 1
Zurück zur Trefferliste

Fluorescence polarization immunoassays for the quantification of caffeine in beverages

  • Homogeneous fluorescence polarization immunoassays (FPIAs) were developed and compared for the determination of caffeine in beverages and cosmetics. FPIAs were performed in cuvettes in a spectrometer for kinetic FP measurements as well as in microtiter plates (MTPs) on a multimode reader. Both FPIAs showed measurement ranges in the µg/L range and were performed within 2 and 20 min, respectively. For the application on real samples, high coefficients of variations (CVs) were observed for the performance in MTPs; the CVs for FPIAs in cuvettes were below 4%. The correlations between this method and reference methods were satisfying. The sensitivity was sufficient for all tested samples including decaffeinated coffee without preconcentration steps. The FPIA in cuvettes allows a fast, precise, and automated quantitative analysis of caffeine in consumer products, whereas FPIAs in MTPs are suitable for semiquantitative high-throughput screenings. Moreover, specific quality criteria forHomogeneous fluorescence polarization immunoassays (FPIAs) were developed and compared for the determination of caffeine in beverages and cosmetics. FPIAs were performed in cuvettes in a spectrometer for kinetic FP measurements as well as in microtiter plates (MTPs) on a multimode reader. Both FPIAs showed measurement ranges in the µg/L range and were performed within 2 and 20 min, respectively. For the application on real samples, high coefficients of variations (CVs) were observed for the performance in MTPs; the CVs for FPIAs in cuvettes were below 4%. The correlations between this method and reference methods were satisfying. The sensitivity was sufficient for all tested samples including decaffeinated coffee without preconcentration steps. The FPIA in cuvettes allows a fast, precise, and automated quantitative analysis of caffeine in consumer products, whereas FPIAs in MTPs are suitable for semiquantitative high-throughput screenings. Moreover, specific quality criteria for heterogeneous assays were applied to homogeneous immunoassays.zeige mehrzeige weniger

Volltext Dateien herunterladen

  • Oberleitner_Fluorescence polarization immunoassays.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Lidia Oberleitner, Julia Grandke, F. Mallwitz, Ute Resch-GengerORCiD, L.-A. Garbe, Rudolf SchneiderORCiD
Dokumenttyp:Zeitschriftenartikel
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):Journal of agricultural and food chemistry
Jahr der Erstveröffentlichung:2014
Verlag:American Chemical Society
Verlagsort:Columbus, Ohio
Jahrgang/Band:62
Ausgabe/Heft:11
Erste Seite:2337
Letzte Seite:2343
Freie Schlagwörter:Decaffeinated coffee; Fluorescence polarization immunoassay; Homogeneous immunoassay; Quality assurance criteria; caffeine
DOI:10.1021/jf4053226
ISSN:0021-8561
ISSN:1520-5118
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")
Datum der Freischaltung:20.02.2016
Referierte Publikation:Ja
Datum der Eintragung als referierte Publikation:28.04.2014
Einverstanden
Diese Webseite verwendet technisch erforderliche Session-Cookies. Durch die weitere Nutzung der Webseite stimmen Sie diesem zu. Unsere Datenschutzerklärung finden Sie hier.