• Treffer 6 von 13
Zurück zur Trefferliste

Hopkinson bar method for temperature dependent testing and calibration of accelerometers

  • The measurement characteristics of every conventional accelerometer are temperature dependent - often to an extent which is highly relevant for the addressed application. For instance, at BAM Container drop tests are performed to investigate and evaluate the structural integrity of Containers for transport and storage of dangerous goods. Often, the test program includes drop tests at specific temperatures, also as part of approval procedures. The applied sensors are exposed to these test conditions and get influenced by them. Acceleration sensors show a considerable temperature influence on their fünction and characteristics. Particularly the damping mechanism of the seismic mass is a critical part. In regard to such applications, capable equipment and methods are required to consider and investigate this aspect in an adequate way. A fortiori as Manufacturer’s infonnation often is deficient. This study presents the setup and results of a method for testing and calibration ofThe measurement characteristics of every conventional accelerometer are temperature dependent - often to an extent which is highly relevant for the addressed application. For instance, at BAM Container drop tests are performed to investigate and evaluate the structural integrity of Containers for transport and storage of dangerous goods. Often, the test program includes drop tests at specific temperatures, also as part of approval procedures. The applied sensors are exposed to these test conditions and get influenced by them. Acceleration sensors show a considerable temperature influence on their fünction and characteristics. Particularly the damping mechanism of the seismic mass is a critical part. In regard to such applications, capable equipment and methods are required to consider and investigate this aspect in an adequate way. A fortiori as Manufacturer’s infonnation often is deficient. This study presents the setup and results of a method for testing and calibration of acceleration sensors under high dynamic irnpact. It combines a Hopkinson Bar with a temperature chamber.zeige mehrzeige weniger

Volltext Dateien herunterladen

  • Hopkinson bar method for temperature dependent testing and calibration of accelerometers.pdf
    eng

Metadaten exportieren

Weitere Dienste

Teilen auf Twitter Suche bei Google Scholar Anzahl der Zugriffe auf dieses Dokument
Metadaten
Autor*innen:Matthias BartholmaiORCiD, Klaus-Dieter Werner, Michael Kammermeier
Dokumenttyp:Beitrag zu einem Tagungsband
Veröffentlichungsform:Verlagsliteratur
Sprache:Englisch
Titel des übergeordneten Werkes (Englisch):31st Danubia-Adria Symposium on advances in experimental mechanics (Proceedings)
Jahr der Erstveröffentlichung:2014
Herausgeber (Institution):Verein Deutscher Ingenieure (VDI)
Erste Seite:244
Letzte Seite:245
Veranstaltung:31st Danubia-Adria Symposium on advances in experimental mechanics
Veranstaltungsort:Kempten, Germany
Beginndatum der Veranstaltung:24.09.2014
Enddatum der Veranstaltung:27.09.2014
ISBN:978-3-00-046740-0
Verfügbarkeit des Dokuments:Datei im Netzwerk der BAM verfügbar ("Closed Access")
Datum der Freischaltung:20.02.2016
Referierte Publikation:Ja
Datum der Eintragung als referierte Publikation:30.10.2014