TY - CONF A1 - Resch-Genger, Ute A1 - Fallisch, A. A1 - Petrov, E. P. A1 - Salhany, R. A1 - Forthman, C. A1 - Guttenberg, Z. A1 - Nitschke, R. T1 - Spectral fluorescence standards for the calibration and performance validation of fluorescence microscopes N2 - The standardization and calibration of fluorescence microscopy have become increasingly vital due to the wide-spread use of in life and materials sciences. As the demand for reliable and user-friendly methods to assess micro-scope performance grows, universal calibration tools accessible to both researchers and vendors are needed. To support the standardization of characterization methods in microscopy, it is crucial to provide calibration tools together with standardized operating procedures for their effective implementation. The public-funded project "FluMiKal"* develops calibration tools in the shape of typical microscopic slides to assess key parameters such as spatial resolution, point spread function, spectral sensitivity, linearity and sensitivity of the detection system. The focus is on creating calibration tools that are user-friendly, robust, and versatile in their application. This work addresses the critical parameter of wavelength-dependent spectral sensitivity, which affects the meas-ured signals from the instrument side, yielding instrument-specific data and instrument aging-induced changes over time. For this purpose, μ-slides from ibidi with six channels are used, allowing them to be filled with different solutions containing molecular or nanoscale fluorophores with well-characterized absorption and fluorescence properties. The certified spectral fluorescence standards BAM-F003, F004, F005, and F007 assessed provided as ethanolic solutions by the Federal Institute for Materials Research and Testing (BAM), cover a broad spectral range from the blue to the near-infrared [1], [2]. Dye-based slide prototypes have been used to determine the spectral sensitivity of confocal microscopes from different vendors with various detector types by acquiring the spectral data of the BAM dyes under standardized measurement conditions, demonstrating the applicability of this concept. Proof-of-concept experiments could demonstrate the proper sealing of the slides. Further experiments will explore long-term stability and their potential as standards for relative intensity calibrations. * FluMiKal is funded by the Federal Ministry for Economic Affairs and Climate Action, Germany (WIPANO FKZ 03TN0047B) [1] doi: 10.1007/4243_2008_028. [2] doi: 10.1007/s00216-024-05723-w. T2 - European Light Microscopy Initiative - ELMI 2025 CY - Heidelberg, Germany DA - 03.06.2025 KW - Fluorescence KW - Advanced material KW - Validation KW - Calibration KW - Method comparison KW - Reference material KW - Dye KW - Fluorescence standard KW - Microscopy KW - CLSM KW - Imaging PY - 2025 AN - OPUS4-64206 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tavernaro, Isabella A1 - Sander, P. A1 - Andresen, Elina A1 - Schedler, U. A1 - Resch-Genger, Ute T1 - Potentiometric and Optical Titration for Cost- Efficient Quantification of Surface Functional Groups on Silica Nanoparticles N2 - Surface chemistry of engineered nanomaterials (NMs) plays a critical role not only in determining their interactions with the environment but also in their stability, safety, and functionality across diverse applications ranging from catalysis to biomedicine. Accurate quantification of surface functional groups (FGs) is therefore essential for quality control, risk assessment, and performance optimization.[1] However, many existing analytical techniques are either cost-intensive, require specialized instrumentation, or lack scalability for routine use. In this study, we present a comparative evaluation of potentiometric and optical titration as two simple, cost-efficient, and automatable methods for quantifying surface functional groups on a variety of surface-modified silica nanoparticles (SiO₂ NPs). These NPs were chosen as they are among the most frequently utilized engineered NMs in the life and material sciences. Potentiometric titration, based on pH monitoring during acid-base neutralization, offers a direct and label-free approach to determine the total amount of FGs. Optical titration provides a complementary method with potential for high-throughput screening. To examine the accuracy and robustness of our stepwise-optimized workflows and the achievable relative standard deviations (RSDs), measurements were performed by multiple operators in two laboratories. Method validation was conducted through cross-comparison with traceable, chemo-selective quantitative nuclear magnetic resonance spectroscopy (qNMR) and thermogravimetric analysis (TGA). A comparison with optical assays highlights the importance of measuring both quantities for comprehensive characterization of surface-modified NMs.[2] A combined NM surface analysis using optical assays and pH titration will simplify quality control of NM production processes and stability studies, and can yield large datasets for NM grouping in sustainable and safe(r)-by-design studies. T2 - eMRS Fall Meeting 2025 CY - Warsaw, Poland DA - 15.09.2025 KW - Fluorescence KW - Advanced material KW - Synthesis KW - Characterization KW - Nano KW - Particle KW - Silica KW - Surface analysis KW - Validation KW - qNMR KW - Fluram assay KW - Functional group KW - Quantification KW - Potentiometry KW - Amino groups KW - Fluorescamine KW - Calibration KW - Method comparison PY - 2025 AN - OPUS4-64205 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Optical spectroscopy – Techniques, instrumentation, and typical molecular and nanoscale reporters N2 - Different types of optical spectroscopies are introduced with special emphasis on method-inherent limitations and reliable instrument calibration and performance validation. In addition, different classes of molecular and nanocrystalline emitters are presented and the underlying photophysical processes are briefly described. T2 - Fakultät der University of Ottawa CY - Ottawa, Canada DA - 11.12.2018 KW - quality assurcance KW - Optical spectroscopy KW - Method comparison KW - Photoluminescence KW - Calibration KW - Performance validation KW - Dye KW - Nanoparticle KW - Quantum dots KW - Quantum yields KW - Uncertainty KW - Method validation PY - 2018 AN - OPUS4-47631 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Luminescence measurements, calibration strategies and photoluminescence quantum yields N2 - Different types of optical spectroscopies are introduced with special emphasis on method-inherent limitations and reliable instrument calibration and performance validation. In addition, procedures for the determination of spectroscopic key parameters like the photoluminescence quantum yield are presented including required instrument calibrations and material-specific effects related to certain emitters. T2 - Fakultät der University of Ottawa CY - Ottawa, Canada DA - 11.12.2018 KW - Quality assurcance KW - Optical spectroscopy KW - Method comparison KW - Photoluminescence KW - Calibration KW - Performance validation KW - Dye KW - Nanoparticle KW - Quantum dots KW - Quantum yields KW - Uncertainty KW - Method validation PY - 2019 AN - OPUS4-47630 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -