TY - CONF A1 - Resch-Genger, Ute T1 - How to get it right with photoluminescence measurements in the UV/VIS/NIR N2 - Optical measurements of transparent solutions of organic dyes and semiconductor quantum dots and scattering materials such as luminescent nanocomposites and microparticles and phosphors dispersed in liquid and solid matrices play an important role in energy conversion, solid-state lighting, medical diagnostics, and security barcoding. A key performance parameter is the photoluminescence quantum yield QY, i.e., the number of emitted per number of absorbed photons. QY of transparent luminophore solutions can be obtained relative to a fluorescence QY standard of known QY. Such relative QY measurements require a calibrated spectrofluorometer. Here, we address typical sources of uncertainty for luminescence measurements and relative and absolute QY measurements from the instrument and material side and provide recommendations to circumvent or at least minimize them for transparent and scattering samples in the UV/vis/NIR/SWIR. T2 - Workshop "Photoluminescence Spectroscopy and its application in Materials Research" CY - Vienna, Austria DA - 19.02.2026 KW - Quality assurance KW - Reference material KW - Fluorescence KW - Dye KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - NIR KW - Instrument performance validation PY - 2026 AN - OPUS4-65575 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Reliable measurements of the photoluminescence quantum yield of transparent and scattering luminophores N2 - Optical measurements of transparent solutions of organic dyes and semiconductor quantum dots and scattering materials such as luminescent nanocomposites and microparticles and phosphors dispersed in liquid and solid matrices play an important role in energy conversion, solid-state lighting, medical diagnostics, and security barcoding. A key performance parameter is the photoluminescence quantum yield QY, i.e., the number of emitted per number of absorbed photons. QY of transparent luminophore solutions can be obtained relative to a fluorescence QY standard of known QY.[1] Meanwhile, a first set of certified fluorescence QY standards is available.[2] Such relative QY measurements require a calibrated spectrofluorometer.[1,3] For determining QY of scattering liquid and solid samples, absolute measurements of QY with a calibrated integrating sphere setup are mandatory.[1,4,5] However, scattering QY standards are not available and uncertainties of such measurements have not yet been assessed in interlaboratory comparisons (ILCs). To determine typical sources of uncertainty of absolute QY measurements, we assessed the influence of the measurement geometry and the optical properties of the blank for determining the number of incident photons absorbed by the sample in an ILC using commercial integrating sphere setups and a custom-designed integrating sphere setup. Samples examined included transparent and scattering dye solutions, solid phosphors such as YAG:Ce optoceramics used as LED converter material, and polymer films stained with different amounts of phosphor microparticles. Matching QY values could be obtained for transparent dye solutions and scattering dispersions with a blank with scattering properties closely matching those of the sample, while QY measurements of optoceramic samples with different blanks revealed substantial differences of more than 20 %. Based on our data, we recommend non-absorbing blank materials with a high reflectivity (>95 %) such as a 2 mm-thick PTFE target placed on the sample holder as blanks. T2 - eMRS 2025 Fall Meeting CY - Warsaw, Poland DA - 15.09.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Advanced material KW - Synthesis KW - Characterization KW - Fluorescence quantum yield KW - Phosphor KW - Absolute KW - Integrating sphere spectroscopy KW - Dye KW - Standardization KW - Reference material KW - Interlaboratory comparison KW - Uncertainty PY - 2025 AN - OPUS4-64184 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - From molecular and nanoscale chromophores with UV/VIS/NIR/SWIR luminescence to multi method characterization of surface coatings N2 - An overview of the research activities of division Biophotonics is presented covering examples for photophysical studies of different types of molecular and nanocrystalline luminophores, luminescent particles, and sensor systems in solution, in dispersion, and in the solid state and multi-method charactreization workflow for the characterization of surface-functionalized engineered nanomaterials. In addition, the importance of reliable optical measurements, particularly standardized workflows for the determination of the key performance parameter luminescence quantum yield of transparent and scattering luminescent samples with fluorescence and integrating sphere spectroscopy, and validated methods for quantifying surface functional groups and ligands on nanomaterials is highlighted. Thereby, also ongoing standardization activities are presented as well as certified reference materials and reference materials from division Biophotonics. T2 - Chinese-German Chemical Association - Annual Meeting CY - Berlin, Germany DA - 22.08.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Synthesis KW - Quantum yield KW - Characterization KW - Advanced material KW - Surface KW - Lifetime KW - Sensor KW - Oxygen KW - Ph KW - Standardization KW - Fluorescent probe KW - Reference material PY - 2025 AN - OPUS4-64181 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Molecular and nanoscale emitters photophysics , photoluminescence quantum yields, and surface chemistry N2 - Inorganic nanocrystals such as spectrally shifting lanthanide-based nanoparticles (LnNCs) like NaYF4: Yb, Er and semiconductor quantum dots, organic and inorganic particles stained with sensor molecules, and organic dyes showing aggregation-induced emission are meanwhile broadly applied in the life and material sciences. The identification of optimum particle architectures and molecular structures for photonic applications requires quantitative spectroscopic studies and methods to control and analyse particle surface chemistry. In the following, photoluminescence studies of different emitter classes are presented, thereby addressing the measurement of particle brightness and photoluminescence quantum yields in different spectral windows parameters required for an in-depth mechanistic understanding. In addition, examples for the quantification of surface functional groups on nanomaterials with optical spectroscopy are given. T2 - GdCH Kolloquium CY - Düsseldorf, Germany DA - 11.11.2025 KW - Dye KW - Fluorescence KW - Quantum yield KW - Lifetime KW - Sensor materials KW - Temperature KW - Cr(III) complex KW - Nano KW - Particle KW - Silica KW - Polymer KW - Metrology KW - Quality assurance KW - Reference material KW - Surface chemistry KW - Size KW - Shape KW - Particle number concentration KW - Method KW - Optical assay KW - qNMR KW - Validation KW - Potentiometry KW - XPS PY - 2025 AN - OPUS4-64728 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Quantifying functional groups and coatings on nanoobjects N2 - Engineered nanomaterials (NM) of different size, shape, chemical composition, and surface chemistry are increasingly used for many key technologies of the 21st century and consumer products. This includes polymer and silica or silica-coated nanoparticles (NP) with covalently bound surface groups, semiconductor quantum dots (QD), metal and metal oxide NP, and lanthanide NP with coordinatively or electrostatically bound ligands, as well as surface-coated nanostructures like micellar encapsulated NP. Decisive for most applications of NMs are their specific surface properties, which are largely determined by the chemical nature and amounts of ligands and functional groups (FGs) on the NM surface. The surface chemistry can strongly affect the physicochemical properties of NM, their charge, hydrophilicity/hydrophobicity, reactivity, function, stability, and processability and thereby their impact on human health and environment. Knowledge of NM surface chemistry plays an important role for NM functionality and performance in (bio)applications and the fate, exposure, dissolution, transformation, and accumulation of NM, and thus, the potential risks for human health and the environment. This highlights the importance of reliable, validated, and eventually standardized analytical methods for analyzing and quantifying NM surface chemistry for process and quality control of NM production, safe use of NMs, design of novel NM, and sustainable concepts for NM fabrication.[1-3] In this context, interlaboratory comparisons (ILCs) are needed to assess method reliability and reference materials with known surface chemistries for establishing surface analytical methods and their performance validation.[2,4] Also, to respond to the increasing number of samples to be analyzed, cost-efficient automation concepts for surface analysis are needed that can be realized with affordable and preferably commercial instrumentation.[5] Here, we provide an overview of analytical methods for FG analysis and quantification used by us for quantifying broadly utilized FGs and ligands on different types of NMs with electrochemical titration methods, optical assays, nuclear magnetic resonance (NMR) and vibrational (IR) spectroscopy, and X-ray based and thermal analysis methods.[1,2] Thereby, method- and material-related challenges are addressed, and the importance of multi-method characterization approaches easing method validation by method cross-validation. Special emphasis is dedicated to simple, versatile, and cost-efficient methods such as optical assays and electrochemical titration methods. T2 - eMRS Fall Meeting 2025 CY - Warsaw, Poland DA - 15.09.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Synthesis KW - Characterization KW - Advanced material KW - Surface KW - Standardization KW - Reference material KW - Functional group KW - Quantification KW - Coating KW - Interlaboratory comparison KW - Uncertainty PY - 2025 AN - OPUS4-64183 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Assessing different types of silica networks for the complete protection of nir luminescent molecular rubies from oxygen quenching in air N2 - The application of emerging luminophores such as near-infrared (NIR) emissive earth-abundant chromium(III) (CrIII) complexes and triplet-triplet annihilation upconversion (TTA-UC) systems in air as optical reporters for bioimaging or photonic materials for energy conversion requires simple and efficient strategies for their complete protection from luminescence quenching by oxygen. Therefore, we explored the influence of sol-gel synthesis routes on the oxygen protection efficiency of the resulting core and core/shell silica nanoparticles (SiO2 NPs), utilizing the molecular ruby-type luminophores CrPF6 ([Cr(ddpd)2](PF6)3; ddpd = N,N’-dimethyl- N,N’-dipyridin-2-ylpyridin-2,6-diamine) and CrBF4 ([Cr(ddpd)2](BF4)3) with their oxygen-dependent, but polarity-, proticity-, viscosity-, and concentration-independent luminescence as optical probes for oxygen permeability. The sol-gel chemistry routes we assessed include the classical Stöber method and the underexplored larginine approach, which relies on the controlled hydrolysis of tetraethoxysilane (TEOS) in a biphasic cyclohexane/water system with the catalyst l-arginine. As demonstrated by luminescence measurements of air- and argon-saturated dispersions of CrPF6- and CrBF4-stained SiO2 NPs of different size and particle architecture, utilizing the luminescence decay kinetics of argon-saturated solutions of CrPF6 and CrBF4 in acetonitrile (ACN) as benchmarks, only SiO2 NPs or shells synthesized by the l-arginine approach provided complete oxygen protection of the CrIII complexes under ambient conditions. We ascribe the different oxygen shielding efficiencies of the silica networks explored to differences in density and surface chemistry of the resulting nanomaterials and coatings, leading to different oxygen permeabilities. Our l-arginine based silica encapsulation strategy can open the door for the efficient usage of oxygen-sensitive luminophores and TTA-UC systems as optical reporters and spectral shifters in air in the future. T2 - eMRS Strasbourg CY - Strasbourg, France DA - 26.05.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Synthesis KW - Quantum yield KW - NIR KW - Mechanism KW - Characterization KW - Electron microscopy KW - Silica KW - Oxygen sensing KW - Surface KW - Doping KW - Lifetime KW - Cr(III) complex PY - 2025 AN - OPUS4-63300 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Quantifying the total and accessible amount of surface functionalities and ligands on nanomaterials N2 - Engineered nanomaterials (NMs) of various chemical composition and surface functionalization are routinely fabricated for industrial applications such as medical diagnostics, drug delivery, sensing, catalysis, energy conversion and storage, opto-electronics, and information storage. NM dispersibility, stability, processability, and function as well as the interaction with biological species and environmental fate are largely determined by NM surface functionalities, i.e., functional groups (FGs) and ligands. Therefore, reliable, reproducible, and eventually standardized surface characterization methods are vital for quality control of NMs, and mandatory to meet increasing concerns regarding their safety. Suitable methods for determining surface functionalities on ligand-stabilized core and core/shell NPs include advanced techniques such as traceable quantitative nuclear magnetic resonance (qNMR) as well as X-ray electron spectroscopy (XPS) and time of flight secondary ion mass spectrometry (ToF-SIMS), and simpler optical and electrochemical methods.[1] The latter less costly and fast methods, which can be automated, are often used by NM producers for process and quality control.[1,2] To validate methods, establish measurement uncertain-ties, test reference materials, and produce reference data, multi-method characterization studies are needed.[3,4] as well as interlaboratory comparisons (ILC) on determining NM surface chemistry and well characterized test and reference NMs providing benchmark values.[5,6] Here, we present examples for quantifying common surface FGs such as amino and carboxyl groups on functional NMs of different chemical composition such as silica, polymer, iron oxide, and lanthanide-based upconversion nanoparticles with optical assays, electrochemical titration methods, qNMR, and chromatographic separation techniques. In addition, ongoing interlaboratory comparisons will be presented. T2 - Yucomat 2025 CY - Herec Novi, Montenegro DA - 01.09.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Synthesis KW - Characterization KW - Advanced material KW - Surface KW - Standardization KW - Reference material KW - Functional group KW - Quantification KW - Coating KW - Interlaboratory comparison KW - Uncertainty PY - 2025 AN - OPUS4-64182 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Gawlitza, Kornelia T1 - Using Dual Fluorescent Molecularly Imprinted Particles Coupled with a Miniaturized Opto-Microfluidic Platform for On-Site Detection of Perfluoroalkyl Carboxylic Acids N2 - Per- and polyfluoroalkyl substances (PFAS) are a group of synthetic organofluorine chemicals widely used in the production of various materials, including firefighting foams, adhesives, and coatings that resist stains and oil. In recent years, PFAS have gained attention as emerging environmental contaminants, with particular emphasis on perfluoroalkyl carboxylic acids (PFCAs), the most common type of PFAS. PFCAs are defined by a fully fluorinated carbon chain and a charged carboxylic acid group. They have been classified as Substances of Very High Concern and included in the REACH Candidate List due to their persistence, resistance to biodegradation, and toxicological impacts. Traditional methods for analyzing PFCAs, like GC-MS, HRMS, and HPLC-based techniques, are time-consuming, non-portable, expensive, and require specialized expertise. On the other hand, fluorescence assays offer a user-friendly, portable, and cost-effective alternative with high sensitivity and quick results, particularly when the binding of the analyte causes a specific increase in the probe’s fluorescence. Combining these probes with a carrier platform and a miniaturized optofluidic device presents a promising approach for PFCA monitoring. In this study, a new guanidine BODIPY fluorescent indicator monomer was synthesized, characterized, and incorporated into a molecularly imprinted polymer (MIP) designed for the specific detection of perfluorooctanoic acid (PFOA). The MIP layer was formed on silica core nanoparticles doped with tris(bipyridine)ruthenium(II) chloride, serving as an optical internal reference for calibration-free assays. In combination with an extraction step prior to sample analysis, this system enables selective and reliable detection of PFCAs in surface water samples, minimizing interference from competing substances, matrix effects, and other factors. When integrated into an opto-microfluidic setup, the assay provided a compact, user-friendly detection system capable of detecting micromolar levels of PFOA in under 15 minutes from surface water samples. T2 - ANAKON2025 CY - Leipzig, Germany DA - 10.03.2025 KW - Perfluorooctanoic Acid (PFOA) KW - On-site detection KW - Fluorescence KW - Microfluidics KW - Molecularly Imprinted Polymers PY - 2025 AN - OPUS4-62712 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Wegner, Karl David T1 - Advancing Short-Wave Infrared (SWIR) Emission N2 - There is a growing interest in the exploitation of the short-wave infrared (SWIR), which refers to the wavelength band of light between 900 nm and 2500 nm. Luminophores that emit in the SWIR are used in various areas of telecommunications, photovoltaics, security systems (night vision), and in biomedicine. In particular for biomedical applications, the SWIR range is highly promising because light scattering, absorption, and autofluorescence of tissue and biological compounds are strongly reduced compared to the visible (400–700 nm) and NIR (~700–900 nm). The benefits of SWIR-emissive QDs have been demonstrated for a variety of applications, such as in thermal sensing, as photoelectrochemical biosensor, in in vivo vascular imaging, and for fluorescence-guided surgery.[1] Full exploitation of SWIR photoluminescence (PL) imaging and sensing is currently hampered by i.) a lack of suitable advanced nanomaterials with a high PL quantum yield (PL QY) and a high brightness, that can be used safely in vivo and ii.) a lack of quantitative and reliable data on the optical properties of many SWIR emitters. Promising nanomaterials for the SWIR are heavy metal-free Ag2S quantum dots (QDs). Aiming for the development of SWIR advanced nanomaterials with optimum performance, we have dived deeper into the photophysical processes occurring in these nanomaterials, thereby exploring in depth how the environment such as temperature, surface ligand composition, and the incorporation of transition metals influence the optical properties Ag2S QDs. We observed a strong enhancement of the SWIR emission of upon addition of metal ions such as Zn2+, yielding PL quantum yields of about 10% and thus making them highly suitable for non-invasive deep imaging of vascular networks and 3D fluid flow mapping. T2 - Anakon 2025 CY - Leipzig, Germany DA - 10.03.2025 KW - Quantum dots KW - Nano KW - Particle KW - SWIR KW - Fluorescence KW - Temperature KW - Ag2S KW - Quality assuarance KW - Ligand KW - Sensor PY - 2025 AN - OPUS4-62769 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Scholtz, Lena T1 - Luminescent, Semiconductor Nanoparticle-Loadedpolymer Microbeads–Comparingparticlearchitectures N2 - Luminescent polymer microparticles (PMPs) are applied in various (bio)analytical and diagnostic processes.[1] The staining of these beads is important for the realization of optically distinguishable barcodes that can be read out, e.g., by a flow cytometer or fluorescence microscope. Typically, luminescent semiconductor nanoparticles (NPs) absorb in a broad wavelength range and show narrow emission bands, which enables simultaneous excitation of differently colored luminophores and facilitates a spectral discrimination.[1] This makes them ideal candidates for this purpose and encouraged us to explore and develop a simple, effective approach to luminescent semiconductor NP encoding of polystyrene PMPs and identify suitable synthesis conditions.[2] Until now, mainly semiconductor quantum dots (QDs) have been used for the synthesis of luminescent PMPs, although NPs with different shapes could introduce beneficial new features. Aiming for the application of our developed procedure to non-spherical NPs, we systematically investigated the luminescence properties of the resulting NP-stained beads using fluorescence and integrating sphere spectroscopy as well as fluorescence and electron microscopy. These studies showed that the suitability of semiconductor NPs for the synthesis of luminescent PMPs depends not only on their shape, but also heavily on their surface chemistry.[3] The successful incorporation of nonspherical NPs opens the path to include even more NPs, and the results can help to deduce future applications for the beads which best suit their specific properties. T2 - E-MRS Fall Meeting 2025 CY - Warsaw, Poland DA - 15.09.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Synthesis KW - Quantum dot KW - Quantum rod KW - Platelet KW - Quantum yield KW - Polymer particle KW - Encoding KW - Surface chemistry KW - Mechanism KW - Characterization KW - Lifetime KW - Barcode KW - Polymerization KW - Method PY - 2025 AN - OPUS4-64242 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - The emp project smurfnano – Standardizing the quantification of surface functionalities, ligands, and coatings on nanomaterials N2 - For industrial applications such as medical diagnostics, drug delivery, sensing, catalysis, energy conversion and storage, opto-electronics, and information storage, meanwhile engineered nanoparticles (NPs) with various chemical compositions and surface functionalities are routinely fabricated. NP function, interaction with biological species, and environmental fate are largely determined by surface functionalities. Reliable, reproducible, and standardized surface characterization methods are therefore vital for quality control of NPs, and mandatory to meet increasing concerns regarding their safety. Also, industry, international standardization organizations, regulatory agencies, and policymakers need validated and standardized measurement methods and reference materials. These needs are addressed by the recently started European metrology project SMURFnano involving 12 partners from different National Metrology Institutes, designated institutes, and research institutes, two university groups as well as one large company and one SME producing NPs. This project as well as first results derived from the development of test and reference materials with a well characterized surface chemistry and ongoing interlaboratory comparisons will be presented. T2 - LNE Workshop CY - Paris, France DA - 04.11.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Synthesis KW - Characterization KW - Advanced material KW - Surface KW - Standardization KW - Reference material KW - Functional group KW - Quantification KW - Coating KW - Automation KW - Potentiometry KW - Method KW - Validation KW - Optical assay KW - Fluram KW - qNMR KW - Comparison KW - ILC PY - 2025 AN - OPUS4-64725 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - The SMURFnano project - standardized measurements of surface functionalities on nanoparticles N2 - Engineered nanoparticles (NPs) with various chemical compositions and surface functionalities are routinely fabricated for industrial applications such as medical diagnostics, drug delivery, sensing, catalysis, energy conversion and storage, optoelectronics, and information storage. NP function, interaction with biological species, and environmental fate are largely determined by surface functionalities. Reliable, reproducible, and standardized surface characterization methods are therefore vital for quality control of NPs, and mandatory to meet increasing concerns regarding their safety. Also, industry, international standardization organizations, regulatory agencies, and policymakers need validated and standardized measurement methods and reference materials. However, methodologies for determining NP surface properties, including the amount, chemical composition, and homogeneity of surface functionalities and coatings are largely non-standardized. Suitable methods for determining surface functionalities on ligand-stabilized core and core/shell NPs include advanced techniques such as traceable quantitative nuclear magnetic resonance (qNMR) as well as X-ray electron spectroscopy (XPS) and time of flight secondary ion mass spectrometry (ToF-SIMS), and simpler optical and electrochemical methods. The latter less costly methods are often used by SMEs, e.g., for quality control. To validate methods, establish measurement uncertainties, test reference materials, and produce reference data, international interlaboratory comparisons (ILC) on NP surface functionalization measurements are required and well characterized test and reference nanomaterials providing benchmark values.[1] These needs are addressed by the recently started European metrology project SMURFnano involving 12 partners from different National Metrology Institutes, designated institutes, and research institutes, two university groups as well as one large company and one SME producing NPs. This project as well as first results derived from the development of test and reference materials with a well characterized surface chemistry and ongoing interlaboratory comparisons will be presented. T2 - eMRS CY - Strasbourg, France DA - 27.05.2025 KW - Quality assurance KW - Reference analysis KW - Standardization KW - Metrology KW - Reference products KW - Reference materials KW - Mission KW - Surface chemistry KW - Nano KW - Particle KW - qNMR KW - XPS KW - Fluorescence KW - Optical assays KW - Interlaboratory comparison KW - Uncertainty PY - 2025 AN - OPUS4-63243 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tavernaro, Isabella A1 - Abram, Sarah-Luise A1 - Homann, Christian A1 - Scholtz, Lena A1 - Radnik, Jörg A1 - Resch-Genger, Ute T1 - Validated and standardized measurements and quantification of surface functionalities on nanoparticles N2 - Surface-functionalized organic and inorganic engineered nanomaterials (NMs) are widely applied in the life and materials sciences. NM performance depends on key factors such as particle size and shape, crystal phase, morphology, chemical composition, and surface chemistry, i.e., surface coatings, functional groups (FGs), and ligands.1 The latter controls their processability and interaction with the environment and largely their possible toxicity. Thus, methods for FG quantification are important tools for quality control of NM production processes and can foster the sustainable development of functional and safe(r) NMs. This underlines the importance of validated and standardized analytical methods for surface analysis and reference materials.2 This encouraged us to explore simple and versatile tools for quantifying common bioanalytically relevant FGs such as optical assays, electrochemical titration methods, quantitative nuclear magnetic resonance spectroscopy (qNMR), and X-Ray photoelectron spectroscopy (XPS) and to perform a first interlaboratory comparison (ILC) on surface FG quantification.3,4 In a follow-up ILC, BAM and NRC explored qNMR sample preparation, measurement, and data evaluation protocols for commercial and custom-made aminated SiO2 NPs with sizes of 20-100 nm, different amounts of surface amino FGs, and different porosity.5,6 First, the number of amino FGs accessible for a dye reporter was determined with a cost-efficient, automated optical fluorescamine assay. Then, qNMR workflows and protocols were stepwise fine-tuned. The qNMR ILC was complemented by joint XPS measurements. BAM also examined the applicability of fast and automatable potentiometric titrations to screen the total amount of (de)protonable FGs on aminated SiO2 NPs. Our results underline the need to evaluate protocols for FG quantification in ILCs and the advantages of multi-method characterization strategies for efficient method cross validation. T2 - Surface and Micro/Nano Analysis Working Group CY - Paris, France DA - 08.04.2025 KW - Quality assurance KW - Reference material KW - Fluorescence KW - Nano KW - Particle KW - Surface analysis KW - XPS KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - QNMR KW - Potentiometry PY - 2025 AN - OPUS4-62969 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Quantifying the total and accessible amount of surface functionalities and ligands on nano-materials: Overview and recommended methods N2 - Engineered nanoparticles (NPs) with various chemical compositions and surface functionalities are routinely fabricated for industrial applications such as medical diagnostics, drug delivery, sensing, catalysis, energy conversion and storage, opto-electronics, and information storage. NP function, interaction with biological species, and environmental fate are largely determined by surface functionalities. This calls for reliable, reproducible, and standardized surface characterization methods, which are vital for quality control of NPs, and mandatory to meet increasing concerns regarding their safety. Validated and standardized workflows for surface analysis are also increasingly requested by industry, international standardization organizations, regulatory agencies, and policymakers. To establish comparable measurements of surface functionalities across different labs and ease instrument performance validation, reference test materials and reference materials of known surface chemistry as well as reference data are needed. In the following, different methods for determining surface functionalities on ligand-stabilized core and core/shell NPs include advanced techniques are presented and discussed regarding method-inherent advantages and limitations. Special emphasis is dedicated to traceable quantitative nuclear magnetic resonance (qNMR), X-ray electron spectroscopy (XPS) and time of flight secondary ion mass spectrometry (ToF-SIMS), and simpler optical and electrochemical methods. T2 - LNE Workshop CY - Paris, France DA - 04.11.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Synthesis KW - Characterization KW - Advanced material KW - Surface KW - Standardization KW - Reference material KW - Functional group KW - Quantification KW - Coating KW - Automation KW - Potentiometry KW - Method KW - Validation KW - Optical assay KW - Fluram KW - Fluorescamine KW - qNMR KW - Comparison KW - ILC PY - 2025 AN - OPUS4-64726 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Buchholz, Michelle T1 - Dual Fluorescent Molecularly Imprinted Polymers (MIPs) for Detection of the Prevalent Anti-Inflammatory Drug Diclofenac N2 - Ensuring the purity of air and water is essential for the overall well-being of life on earth and the sustainability of the planet's diverse ecosystems. To achieve the goal of zero pollution, as outlined in the 2020 European Green Deal by the European Commission,[1] significant efforts are in progress. A key aspect of this commitment involves advancing more efficient and economically viable methods for treating wastewater. This includes the systematic monitoring of harmful pollutants such as heavy metals, microplastics, pesticides, and pharmaceuticals. One example is the presence of the anti-inflammatory drug diclofenac in water systems, primarily originating from its use as a gel or lotion for joint pain treatment. Diclofenac contamination in surface waters has been detected at approximately 10 μg L-1 (0.03 μM)[2] which is not solely due to widespread usage but also because of the drug's resistance to microbial degradation. Conventional wastewater treatment plants (WWTPs), which rely on biodegradation, sludge sorption, ozone oxidation, and powdered activated carbon treatment, struggle to efficiently remove diclofenac from wastewater.[3],[4] For instance, to enable WWTPs to efficiently monitor and optimize their processes, it would be advantageous to develop on-site detection and extraction methods for persistent pharmaceutical residues in aqueous samples. In this work, a sol-gel process was used to prepare Nile blue-doped silica nanoparticles (dSiO2-NPs) with a diameter of ca. 30 nm that were further functionalized to enable reversible-addition-fragmentation chain-transfer (RAFT) polymerization. To achieve fluorescence detection, a fluorescent monomer was used as a probe for diclofenac in ethyl acetate, generating stable complexes through hydrogen bond formation. The diclofenac/fluorescent monomer complexes were imprinted into thin molecularly imprinted polymer (MIP) shells on the surface of the dSiO2-NPs. Thus, the MIP binding behaviour could be easily evaluated by fluorescence titrations to monitor the spectral changes upon addition of the analyte. Doping the core substrate with Nile blue generates effective dual fluorescent signal transduction. This approach does not solely depend on a single fluorescence emission band in response to analyte recognition. Instead, it enables the fluorescent core to function as an internal reference, minimizing analyte-independent factors such as background fluorescence, instrumental fluctuation, and operational parameters.[5] Rebinding studies showed that the MIP particles have excellent selectivity towards the imprinted template and good discrimination against the competitor ibuprofen, with a discrimination factor of 2.5. Additionally, the limit of detection was determined to be 0.6 μM. Thus, with further optimization of the MIP, there is potential for the development of a MIP-based biphasic extract-&-detect fluorescence assay for simple, sensitive and specific sensing of diclofenac in aqueous samples down to the required concentrations of 0.03 μM. T2 - MIP2024: The 12th International Conference on Molecular Imprinting CY - Verona, Italy DA - 18.06.2024 KW - Sensor KW - Diclofenac KW - Molecularly Imprinted Polymers KW - Fluorescence KW - Pollutant PY - 2024 AN - OPUS4-60439 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tavernaro, Isabella T1 - Amorphous, fluorescent silica particles for bioimaging applications N2 - Nowadays amorphous silica nanoparticles (SiO2 NP) are one of the most abundant engineered nanomaterials, with an annual production of hundreds of thousands of tons, that are used in a broad field of industrial products and processes. Since SiO2 NP are highly stable and easily produced on a large scale at low cost, they are widely employed as fillers for rubbers and composites, absorbents, catalysts, advanced coating additives as well as plant growth agents in agriculture, anti-caking agents in food products, or as carrier material in cosmetic industry. Moreover, they are promising candidates for colloidal scaffolds in biomedical applications like bioimaging, sensing or controlled drug delivery. SiO2 NP modified with luminescent chromophores have several advantages as compared to conventional molecular probes like enhanced brightness, ease of designing ratiometric systems, and increased photostability. Here we present our work on multicolored SiO2 NP for imaging and sensing applications. T2 - FUNGLASS Workshop CY - Berlin, Germany DA - 06.03.2024 KW - Sensors KW - Nano KW - Particles KW - Silica KW - Luminescence KW - Fluorescence KW - Quality assurance KW - Method KW - Synthesis KW - Dye KW - pH KW - Surface analysis PY - 2024 AN - OPUS4-62167 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Schweizer, S. T1 - Interlaboratory comparison on absolute PL quantum yield measurements of scattering luminescent materials N2 - Optical measurements of scattering luminescent materials dispersed in liquid and solid matrices and luminescent powders play an important role in fundamental research and industry. Typical examples are luminescent nano- and microparticles and phosphors of different composition in different matrices or incorporated into ceramics with applications in energy conversion, solid-state lighting, medical diagnostics, and security barcoding. The key parameter for the performance of these materials is the photoluminescence quantum yield QY, i.e., the number of emitted photons per number of absorbed photons. QY of transparent luminophore solutions can be determined relatively to a fluorescence quantum yield standard of known QY. Such standards are meanwhile available as certified reference materials.[1] The determination of QY of scattering liquid and solid samples like dispersions of luminescent nanoparticles, solid phosphors, and optoceramics requires, however, absolute measurements with an integrating sphere setup. Although the importance of reliable absolute QY measurements has been recognized, no interlaboratory comparisons (ILCs) on measurement uncertainties and the identification of typical sources of uncertainty have been yet reported. Also, no scattering reference materials with known QY are available. We present here the results of a first ILC of 3 laboratories from academia and industry performed to identify and quantify sources of uncertainty of absolute QY measurements of scattering samples. Thereby, two types of commercial stand-alone integrating sphere setups with different illumination and detection geometries were utilized for measuring QY of transparent and scattering dye solutions and solid phosphors. As representative and industrially relevant solid and scattering samples, YAG:Ce optoceramics of varying surface roughness were chosen, applied, e.g., as converter materials for blue light emitting diodes. Special emphasis was dedicated to the influence of the measurement geometry, the optical properties of the blank, utilized to determine the number of photons of the incident excitation light absorbed by the sample, and the sample-specific surface roughness. While matching QY values could be obtained for transparent dye solutions and scattering dispersions, here using a blank with scattering properties closely matching those of the sample, QY measurements of optoceramic samples with different blanks revealed substantial differences, with the blank's optical properties accounting for measurement uncertainties of more than 20 %. Based upon the ILC results, we recommend non-absorbing blank materials with a high reflectivity (>95 %) such as a 2 mm-thick PTFE target placed on the sample holder which reveals a near-Lambertian light scattering behavior, yielding a homogeneous light distribution within the integrating sphere. T2 - 8th IWASOM CY - Gdánsk, Poland DA - 07.07.2024 KW - Nano KW - Particle KW - Fluorescence KW - Quantum yield KW - Method KW - Sensor KW - Film KW - Absolute quantum yield KW - Scattering KW - YAG:Ce KW - Quality assurance KW - Lanthanide KW - Upconverter KW - LED KW - Uncertainty KW - Standardization KW - Reference material PY - 2024 AN - OPUS4-62100 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Rurack, Knut T1 - Sensitive and multiplexed assays for point-of-need applications: innovations for robust, reliable, and user-friendly diagnostics N2 - The development of portable analytical assays, particularly during the SARS-CoV-2 pandemic, has revolutionized diagnostics and expanded their use to areas such as food safety, environmental monitoring and forensics. These assays offer the advantage of rapid on-site decision making without the need for laboratory facilities. The omnipresence of mobile devices with advanced cameras and processing power further increases their usability. However, most current assays are limited to detecting single parameters. The challenge now is to develop robust multiplexed assays that can simultaneously detect multiple parameters with high sensitivity. This lecture will present generic approaches developed at BAM with a focus on supramolecular chemistry, luminescence detection, nanomaterials and miniaturization of devices. Examples include mesoporous nanomaterials, gated indicator systems, imprinted polymers, microfluidic devices, test strips and smartphone-based analysis. T2 - Kolloquium Optische Technologien der FH Münster CY - Steinfurt, Germany DA - 18.12.2024 KW - Rapid testing KW - Fluorescence KW - Explosives KW - Environmental contaminants KW - Multiplexing PY - 2024 AN - OPUS4-62322 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Getting it right with photoluminescence quantum yields of molecular and nanoscale luminophores and luminescent particles N2 - Photophysical and mechanistic studies, the comparison of different emitter classes, and the rational design of the next generation of molecular and nanoscale reporters require quantitative photoluminescence measurements and the reliable determination of the key performance parameter photoluminescence quantum yield (QY), i.e., the number of emitted per absorbed photons. This is of special importance for all photoluminescence applications in the life and material sciences in the UV/vis/NIR/SWIR. To improve the reliability and comparability of photoluminescence and QY measurements across laboratories, pitfalls, achievable uncertainties, and material-specific effects related to certain emitter classes must be explored. Also, suitable protocols and reference materials are needed which have been validated in interlaboratory comparisons for different wavelength regions and transparent and scattering luminophores.[1] Based on absolute and relative photoluminescence measurements of functional dyes and nanomaterials like semiconductor quantum dots and rods, spectrally shifting lanthanide upconversion nanocrystals, perovskites, and YAG:Cer converter materials, reliable methods for determining QY of transparent and scattering luminophores, nonlinear emitters, and solid luminescent nanomaterials have been developed.[2,3] Thereby, material- and method-related uncertainties of relative and absolute QY measurements and achievable uncertainties could be quantified for linear and nonlinear UV/vis/NIR/SWIR emitters and lately for also luminescent and scattering materials and solid phoshors, here in an interlaboratory comparison of three labs utilizing integrating sphere spectroscopy.[4,5] In addition, to provide simple tools for a better comparability of QY measurements, recently, a first set of UV/vis/NIR quantum yield standards has been developed and certified with complete uncertainty budgets.[6] In the following, the outcome of these studies will be presented, thereby addressing common pitfalls and providing recommendations on the performance of reliable QY measurements of linear and non-linear emitters in transparent, scattering, and solid samples. T2 - 29th Lecture Conference on Photochemistry (LCP 2024) GDCh CY - Mainz, Germany DA - 16.09.2024 KW - Nanoparticle KW - Nano KW - Luminescence KW - Quality assurance KW - Synthesis KW - Standardization KW - Reference material KW - Quantum yield KW - Fluorescence KW - Reference data KW - Integrating sphere spectroscopy KW - ILC KW - Converter material KW - YAG:Ce KW - Optoceramic PY - 2024 AN - OPUS4-61075 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Scholtz, Lena T1 - Luminescent, Semiconductor Nanoparticle-loaded Polymer Microbeads - Comparing Particle Architectures N2 - The synthesis of luminescent polymer microparticles stained with different, including non-spherical, semiconductor nanoparticles (NPs) was developed and optimized while ensuring a good preservation of the particle photoluminescence. The aplicability of the different semiconductor NPs was determined to depend on their surface chemistry and shape. Thus, a first correlation of semiconductor NP architecture and applicability could be deduced, which has not been reported before. T2 - 821. WE-Heraeus-Seminar CY - Bad Honnef, Germany DA - 12.11.2024 KW - Nanoparticle KW - Surface chemistry KW - Synthesis KW - Fluorescence KW - Luminescence PY - 2024 AN - OPUS4-61732 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Photoluminescence quantum yields of molecular & nanoscale luminophores in the UV/VIS/NIR/SWIR in dispersion and in the solid state N2 - Photophysical and mechanistic studies, the comparison of different emitter classes, and the rational design of the next generation of molecular and nanoscale reporters require quantitative photoluminescence measurements and the reliable determination of the key performance parameter photoluminescence quantum yield (QY), i.e., the number of emitted per absorbed photons. This is of special importance for all photoluminescence applications in the life and material sciences in the UV/vis/NIR/SWIR. T2 - MAF 2024 Conference CY - Valencia, Spain DA - 08.09.2024 KW - Nanoparticle KW - Nano KW - Luminescence KW - Quality assurance KW - Synthesis KW - Standardization KW - Reference material KW - Quantum yield KW - Fluorescence KW - Reference data KW - Integrating sphere spectroscopy KW - ILC KW - Converter material PY - 2024 AN - OPUS4-61072 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Absolute Determination of Photoluminescence Quantum Yields of Scattering LED Converter Materials – How to Get it Right N2 - Optical measurements of scattering luminescent materials dispersed in liquid and solid matrices and luminescent powders play an important role in fundamental research and industry. Typical examples are luminescent nano- and microparticles and phosphors of different composition in different matrices or incorporated into ceramics with applications in energy conversion, solid-state lighting, medical diagnostics, and security barcoding. The key parameter for the performance of these materials is the photoluminescence quantum yield QY, i.e., the number of emitted photons per number of absorbed photons. QY of transparent luminophore solutions can be determined relatively to a fluorescence quantum yield standard of known QY. Such standards are meanwhile available as certified reference materials.[1] The determination of QY of scattering liquid and solid samples like dispersions of luminescent nanoparticles, solid phosphors, and optoceramics requires, however, absolute measurements with an integrating sphere setup. Although the importance of reliable absolute QY measurements has been recognized, no interlaboratory comparisons (ILCs) on measurement uncertainties and the identification of typical sources of uncertainty have been yet reported. Also, no scattering reference materials with known QY are available. We present here the results of a first ILC of 3 laboratories from academia and industry performed to identify and quantify sources of uncertainty of absolute QY measurements of scattering samples. Thereby, two types of commercial stand-alone integrating sphere setups with different illumination and detection geometries were utilized for measuring QY of transparent and scattering dye solutions and solid phosphors. As representative and industrially relevant solid and scattering samples, YAG:Ce optoceramics of varying surface roughness were chosen, applied, e.g., as converter materials for blue light emitting diodes. Special emphasis was dedicated to the influence of the measurement geometry, the optical properties of the blank, utilized to determine the number of photons of the incident excitation light absorbed by the sample, and the sample-specific surface roughness. While matching QY values could be obtained for transparent dye solutions and scattering dispersions, here using a blank with scattering properties closely matching those of the sample, QY measurements of optoceramic samples with different blanks revealed substantial differences, with the blank's optical properties accounting for measurement uncertainties of more than 20 %. Based upon the ILC results, we recommend non-absorbing blank materials with a high reflectivity (>95 %) such as a 2 mm-thick PTFE target placed on the sample holder which reveals a near-Lambertian light scattering behavior, yielding a homogeneous light distribution within the integrating sphere. T2 - e-MRS 2024 CY - Strasbourg, France DA - 27.05.2024 KW - Luminescence KW - Quantitative spectroscopy KW - Nano KW - Particle KW - Quantum yield KW - Lifetime KW - Quality assurance KW - Phosphor KW - Converter marterial KW - Fluorescence KW - Interlaboratory KW - Comparison KW - Method KW - Uncertainty KW - Reference material PY - 2024 AN - OPUS4-60490 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tavernaro, Isabella T1 - Rational Design of Analyte-responsive Fluorescent Particle Sensors for Life Sciences Applications N2 - Engineered and tailor-made nanomaterials (NM) are of increasing relevance for current and future developments in the life and material sciences for applications, e.g., as drug carriers, fluorescent sensors, and multimodal labels in bioanalytical assays, and reporters for imaging applications. For instance, NM-based reporters and sensors, that are labelled or stained with a multitude of conventional or sensor dyes, have several advantages as compared to molecular probes like enhanced brightness, i.e., amplified signals, ease of designing ratiometric systems by combining analyte-sensitive and inert reference dyes, and increased photostability. Moreover, this can enable the use of hydrophobic dyes in aqueous environments. For rational NM design, choosing and tailoring the intrinsic physicochemical properties, such as particle size, size distribution, morphology, and surface chemistry of the NM application-specific considerations like biocompatibility, ease and low cost of preparation, and colloidal stability and performance in the targeted environment must be considered. In this lecture, different design concepts of inorganic, organic, and hybrid NM and microparticles with hydrophilic surface chemistries and different functionalities are presented that can be used for the targeting of lysosomes; and to monitor functional parameters of endo-lysosomal compartments, like pH or enable oxygen sensing. T2 - Chemical Probes for Lysosomal Biology CY - York, United Kingdom DA - 09.09.2024 KW - Luminescence KW - Quantification KW - Nano KW - Particle KW - Quality assurance KW - Fluorescence KW - Method KW - Uncertainty KW - Reference material KW - Sensor KW - Synthesis KW - Dye KW - pH KW - Silica KW - Polystyrene KW - Surface analysis PY - 2024 AN - OPUS4-62173 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tavernaro, Isabella T1 - pH- and O2-Responsive Nanoparticles – The MiGraGen Project N2 - In recent years, the demand for reliable, versatile, fluorescent pH and oxygen sensors has increased rapidly in many biomedical applications since these analytes are important indicators of cell function or certain diseases. Therefore, sensor particles are needed that are small enough to penetrate cells, non-toxic, and allow for close-up optical monitoring. When developing such sensor systems, one must consider the pH and oxygen range detectable by the sensor dye and the matrix material of the used carrier particles. Here, we present the development of pH- and oxygen-responsive polymeric beads functionalized with fluorescent dyad molecules that consist of an analyte-responsive fluorophore and an analyte-inert dye. T2 - MiGraGen Project Meeting 09.08.2023 CY - Online Meeting DA - 09.08.2023 KW - Nano- and microsensors KW - Functionalized silica and polymeric particles KW - pH sensing KW - Oxygen sensing KW - Fluorescence PY - 2023 AN - OPUS4-58071 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Design and Quantitative Characterization of Functional Molecular Chromophores and Nanomaterials with UV/vis/NIR/IR Emission – An Overview of Research Activities in Division Biophotonics N2 - In the focus of division Biophotonics are the design, preparation, analytical and spectroscopic characterization, and application of molecular and nanoscale functional materials, particularly materials with a photoluminescence in the visible, near infrared (NIR) and short-wave infrared (SWIR). This includes optical reporters for bioimaging and sensing, security and authentication barcodes, and materials for solid state lighting, energy conversion, and photovoltaics. For the identification of optimum particle structures quantitative spectroscopic studies are performed under application-relevant conditions, focusing on the key performance parameter photoluminescence quantum yield. In addition, simple, cost-efficient, and standardizable strategies for quantifying functional groups on the surface of nano- and microparticles are developed, here with a focus on optical assays and electrochemical titration methods, cross-validated by more advanced methods such as quantitative NMR. In addition, reference materials and reference products are developed for optical methods, particularly luminescence techniques, and for analytical methods utilized for the characterization of nanomaterials. T2 - Projekttreffen Nile-Chrome 2.0 CY - Mainz, Germany DA - 11.12.2023 KW - Fluorescence KW - Quantum yield KW - Optical spectroscopy KW - Reference material KW - Reference data KW - Quality assurance KW - Dye KW - Reference product KW - NIR KW - SWIR KW - Nano KW - Particle KW - Silica KW - Polymer KW - Surface group analysis KW - Sensor molecules PY - 2023 AN - OPUS4-59123 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Wegner, Karl David T1 - Luminescent Quantum dots – the next-generation nano light bulbs N2 - Fluorescent semiconductor nanocrystals, also known as quantum dots (QDs), enabled many advancements in biotechnology, photovoltaics, photocatalysis, quantum computing and display devices. The high versatility of this nanomaterial is based on their unique size-tunable photoluminescence properties, which can be adjusted from the visible to the near-infrared range. In contrast to other nanomaterials, QDs made the transition from a laboratory curiosity to the utilization in commercial products, like the QLED television screen or in smartphone displays. The best investigated QDs are composed of heavy metals like cadmium or lead, which is not the best choice in terms of toxicity and environmental pollution. A more promising material is Indium Phosphide (InP), which is also currently used by Samsung, Sony and co. in the QLED displays. In this contribution, I would like to give you a sneak peek behind the curtains of nanomaterial synthesis and show how this material is produced, how to stabilize their structural properties, and assess their toxicity in environmentally relevant conditions. Furthermore, I would like to present a synthesis method to accomplish the last open challenge in display technology of a blue luminescent LED based on QDs by introducing a new element to the InP QDs. T2 - The Berlin Postdoc Day CY - Berlin, Germany DA - 03.11.2022 KW - InP KW - Quantum dots KW - Fluorescence KW - Aging KW - Doping KW - Nanomaterial KW - Cytotoxicity PY - 2022 AN - OPUS4-56194 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Pauli, Jutta A1 - Würth, Christian A1 - Güttler, Arne A1 - Resch-Genger, Ute T1 - Reliable Determination of the Signal-Relevant Spectroscopic Key Characteristics of Luminescent Reporters and Optical Probes for Imaging in the vis/NIR/SWIR N2 - Introduction. Comparing different emitter classes and rationally designing the next generation of molecular and nanoscale probes for bioimaging applications require accurate and quantitative methods for the measurement of the key parameter photoluminescence quantum yield f.1 f equals the number of emitted per number of absorbed photons. This is particularly relevant for increasingly used fluorescence imaging in the short wave-infrared region (SWIR) ≥ 900 nm providing deeper penetration depths, a better image resolution, and an improved signal-to-noise or tumor-to-background ratio.2, 3 However, spectroscopic measurements in the SWIR are more challenging and require specific calibrations and standards. T2 - EMIM 2021 CY - Göttingen, Germany DA - 24.08.2021 KW - Fluorescence KW - Optical probe KW - Dye KW - Photophysics KW - Quantum yield KW - Mechanism KW - NIR KW - SWIR KW - Imaging KW - Reference material KW - Reliability KW - Nano KW - Particle KW - Method KW - Quality assurance PY - 2021 AN - OPUS4-53233 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Measuring the Upconversion Luminescence of Ensemble and Single Particle Lanthanide-Based Upconversion Nanocrystals N2 - Lanthanide-based upconversion nanoparticles (UCNPs) like hexagonal 𝛽-NaYF4 UCNPs doped with Yb3+ and Er3+, which efficiently convert 976 nm light to ultraviolet, visible, and near infrared photons, offer new strategies for luminescence-based sensing, barcoding, and Imaging. Their upconversion (UC) luminescence (UCL) features like UCL intensity, quantum yield, relative spectral distribution / UCL luminescence color, and luminescence decay kinetics are, however, strongly influenced by particle size, dopant ion concentration, particle architecture, surface chemistry including presence and thickness of surface passivation and shielding shells, microenvironment/presence of quenchers with high energy vibrations, and excitation power density (P). We present here a comprehensive study of the influence of excitation power density on the UCL features of different types of UCNPs, focusing on Yb3+ and Er3+ co-doped NaYF4 core-only and core-shell nanostructures with different sizes and doping ion concentration, which underlines the importance of P-dependent optimum dopant concentrations for UCNP performance and the potential of P-tuning of UCL. T2 - Materials Challenges in Alternative & Renewable Energy 2021 (MCARE 2021) CY - Online meeting DA - 19.07.2021 KW - Fluorescence KW - Lifetime KW - Method KW - Stability KW - Coating KW - Surface chemistry KW - Lanthanide KW - Upconversion KW - Nano KW - Particle KW - Single particle spectroscopy KW - Quantum yield KW - Microscopy KW - Photophysics PY - 2021 AN - OPUS4-53111 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Influence of Surface Chemistry and Size on the Stability of β-NaYF4:Yb,Er Nanocrystals in Various Environments N2 - The use of inorganic lanthanide-doped upconversion nanoparticles (UCNP) in bioimaging and cellular studies requires biocompatible particles. One possible cause of UCNP toxicity is the release of potentially harmful fluoride and lanthanide ions as revealed by dilution studies in aqueous environments, particularly under high dilution conditions. To address this issue, suitable surface coatings preventing such effects in combination with fast screening methods suited for online monitoring and in situ analyses are desired. Here we present systematic studies of differently sized β-NaYF4:Yb,Er UCNP stabilized with different surface coatings and hydrophilic ligands varying in binding strength to the particle surface in various aqueous environments at different temperatures and UCNP concentrations. The concentration of the fluoride and lanthanide ions released upon particle dissolution was quantified electrochemically with a fluoride ion-sensitive electrode and inductively coupled plasma optical emission spectrometry (ICP-OES) and monitored fluorometrically, thereby exploiting the sensitivity of the upconversion luminescence to changes in size and surface chemistry. Moreover, changes in surface chemistry were determined with X-Ray photoelectron spectroscopy (XPS). Based upon our results, we could derive optimum screening parameters for UCNP stability studies and determine conditions and coating procedures and ligands for enhancing UCNP stability in aqueous environments. T2 - UPCON2021 CY - Online meeting DA - 06.04.2021 KW - Fluorescence KW - Lifetime KW - Method KW - Quantification KW - Stability KW - Coating KW - Surface chemistry KW - Lanthanide KW - Fluoride KW - Electrochemistry KW - ICP-OES KW - Upconversion KW - Nano KW - Particle KW - Aging KW - Quality assurance KW - Mass spectrometry KW - XPS PY - 2021 AN - OPUS4-52411 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Introduction to Fluorescence Spectroscopy N2 - A brief introduction to fluorescence spectroscopy will be provided, ranging from typically measured fluorescence quantities over instrument-specific contributions to measured fluorescence signals to selected applications. In this context, an overview of the photoluminescence properties of molecular and nanoscale luminescence reporters will be given including a brief insight into their photophysics and fluorescence standards designed by division Biophotonics for the calibration and instrument performance validation of fluorescence measuring devices will be presented. T2 - Analytical Academy CY - Berlin, Germany DA - 13.01.2020 KW - Linearity KW - Fluorescence KW - Dye KW - Quality assurance KW - Nnano particle KW - Method KW - Measurement uncertainty KW - Quantification PY - 2020 AN - OPUS4-51618 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Introduction to Fluorescence Spectroscopy N2 - A brief introduction to fluorescence spectroscopy will be provided, ranging from typically measured fluorescence quantities over instrument-specific contributions to measured fluorescence signals to selected applications. In this context, an overview of the photoluminescence properties of molecular and nanoscale luminescence reporters will be given including a brief insight into their photophysics and fluorescence standards designed by division Biophotonics for the calibration and instrument performance validation of fluorescence measuring devices will be presented. T2 - Analytical Academy BAM CY - Online meeting DA - 17.11.2020 KW - Fluorescence KW - Quality assurance KW - Quantification KW - Linearity KW - Measurement uncertainty KW - Method KW - Nano particle KW - Dye PY - 2020 AN - OPUS4-51619 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Design and Quantitative Characterization of Functional Molecular Chromophores and Nanomaterials with UV/vis/NIR/IR Emission and Traceable Optical Measurements N2 - Research of division Biophotonics at the Federal Institute for Materials Research and Testing (BAM) covers several topics including photophysics of molecular and nanocrystalline emitters, the development of signal enhancement, multiplexing, and barcoding strategies, surface group quantification, the rational design of different types of stimuli-responsive optical probes, and concepts and reference materials for the validation of optical-spectroscopic measurements. In the following representative examples for each of these topics are given. Also, current developments like single particle spectroscopy and flow cytometry with lifetime detection and newly certified fluorescence quantum yield standards are presented. T2 - Kolloquium BfR CY - Berlin, Germany DA - 12.10.2020 KW - Fluorescence KW - Quantitative spectroscopy KW - Single particle spectroscopy KW - Multiplexing KW - Reference materials KW - Optical probes KW - Sensor molecules KW - Assay KW - Dye KW - Quantum yield KW - Method development KW - surface group analysis KW - synthesis KW - fluorescence standards PY - 2020 AN - OPUS4-51449 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Photoluminescence Properties of Different Types of Nanocrystals at the Ensemble and Single Emitter Level N2 - Correlating the photoluminescence (PL) properties of nanomaterials like semiconductor nanocrystals (QDs) and upconversion nanocrystals (UCNPs) assessed in ensemble studies and at the single particle level is increasingly relevant for applications of these nanomaterials in the life sciences like bioimaging studies or their use as reporters in microfluidic assays. Here we present a comparison of the spectroscopic properties of ensembles and single emitters for QDs like II/VI QDs and cadmium-free AIS/ZnS QDs as well as different UCNPs. The overall goal of this study was to derive particle architectures well suited for spectroscopic and microscopic applications. T2 - BIOSSPIE CY - San Francisco, CA, USA DA - 02.02.2019 KW - Quantum yield KW - Nanomaterial KW - Photoluminescence KW - Absolute fluorometry KW - Integrating sphere spectroscopy, KW - NIR KW - IR KW - Fluorescence KW - Nanoparticle KW - Semiconductor KW - Quantum dot KW - Single particle spectroscopy KW - Surface chemistry PY - 2019 AN - OPUS4-47358 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Relative and Absolute Methods for Measuring Photoluminescence Quantum Yields of UV/vis/NIR Emitters N2 - One of the key spectroscopic performance parameters of molecular and particulate emitters is the photoluminescence quantum yield (PL QY) that provides a direct measure for the number of emitted per absorbed photons. This triggered the interest in methods suitable for measuring this property for emitters in various environments in the UV/vis/NIR and above 1000 nm as well as on the ensemble and single emitter level. Moreover, for nonlinear emitters like lanthanide-based upconversion nanocrystals methods including instrumentation for power density-dependent PL QY studies are required. An overview of the research activities in Division Biophotonics of BAM is given and suitable relative and absolute methods for the deter-mination of PL QY of organic dyes and different types of application-relevant nanomaterials in dispersion and in the solid state are presen-ted. This covers also the design and calibration of integrating sphere setups, achievable uncertainties, and candidates for PL QY reference materials. T2 - OSRAM Veranstaltung CY - Regensburg, Germany DA - 09.01.2019 KW - Quantum yield KW - Calibration KW - Reference material KW - Uncertainty KW - Absolute fluorometry KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Fluorescence KW - Nanoparticle KW - Dye PY - 2019 AN - OPUS4-47263 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Advanced characterization of nanomaterials N2 - The rational synthesis and use of nanomaterials require the characterization of many different properties, ranging from particle size and size distribution over surface chemistry to more applicationrelevant features like optical, electrochemical, and magnetic properties. In the following, several methods for the characterization of functional groups on nanomaterials, like polymer and silica nanoparticles, semiconductor quantum dots, and lanthanide-based upconversion nanocrystals are presented. Additionally, procedures for the measurement of the key spectroscopic performance parameters of nanomaterials with linear and nonlinear photoluminescence, such as the photoluminescence quantum yield, are presented for the UV/vis/NIR/SWIR. T2 - Summerschool CY - Bad Honnef, Germany DA - 22.07.2019 KW - Quantum yield KW - Nanoparticle KW - Fluorescence KW - Quantum dot KW - NIR KW - SWIR KW - Quality assurance KW - Calibration PY - 2019 AN - OPUS4-48630 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Determining Photoluminescence Quantum Yields of Molecular and Nanocrystal Emitters in the UV/vis/NIR/SWIR N2 - The comparison of different emitter classes and the rational design of the next generation of molecular and nanoscale reporters require accurate and quantitative photo-luminescence measurements. This is of special importance for all photoluminescence applications in the life and material sciences and nanobiophotonics. In the following, procedures for the determination of the spectroscopic key parameter photoluminescence quantum yield, i.e., the number of emitted per absorbed photons, in the UV/vis/NIR/SWIR are presented including pitfalls and achievable uncertainties and material-specific effects related to certain emitter classes are addressed. T2 - Kolloqium National Center for Nanoscience and Technology CY - Peking, People's Republic of China DA - 18.10.2019 KW - Fluorescence KW - Quantum yield KW - Integrating sphere spectroscopy KW - Dye KW - Nanocrystal KW - NIR KW - SWIR KW - Quantum dot KW - Fluorescence standard KW - Uncertainty KW - Calibration PY - 2019 AN - OPUS4-49362 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Hübner, Oskar T1 - Luminescence lifetime encoding for flow cytometry with quantum-dot-encoded beads N2 - Spectral encoding of cells or particles and the discrimination of multiple spectral codes are a critical process in flow cytometry (FCM). Typical issues in spectral encoding are, e.g., the spectral overlap of codes, or the increasing complexity of instruments . The exploitation of the photoluminescence lifetime (LT) as an encoding parameter could be used to circumvent both of these issues, as it adds another dimension to the parameter space, or, when used as a stand-alone parameter, requiring only one excitation light source and one detector. While LT encoding was considered already decades ago it is still not implemented as a routine technique in FCM yet, mainly due to the challenge of very few photons being available within the limited transition time of a cell or particle through the laser spot. Recently, we demonstrated LT-FCM based on luminophores with ns LTs in a compact and low-cost flow cytometer. Measurements on polymer microbeads containing luminophores with distinctly different excited state LTs enabled the complete discrimination of three LT codes and five codes in total could be identified. Now, we have extended our approach towards considerably longer LTs by custom-made polymer microbeads loaded with different ratios of InP/ZnS and AgInS2 quantum dots. The use of these materials significantly expands the usable time range for LT encoding to up to several hundred ns. Our studies demonstrate the possibility to further increase the number of viable LT codes for multiplexing in LT-FCM without the need for extensive hardware modifications. T2 - Visions in Cytometry - 29th Annual Conference of the German Society for Cytometry CY - Berlin, Germany DA - 25.07.2019 KW - Multiplexing KW - Lifetime KW - Bead KW - Flow cytometry KW - Fluorescence KW - Quantum dot KW - InP KW - AIS KW - Dye KW - Encoding KW - Barcoding KW - Assay KW - Method PY - 2019 AN - OPUS4-49390 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Luminescent nanocrystals – Photophysics and applications for lifetime multiplexing N2 - Bioanalytical, diagnostic, and security applications require the fast and sensitive determination of a steadily increasing number of analytes or events in parallel in a broad variety of detection formats.[1,2] Ideal candidates for spectral encoding and multiplexing schemes are luminescent nanocrystals like semiconductor quantum dots (QDs), particularly Cd-containing II/VI QDs with their narrow and symmetric emission bands. With the availability of relatively simple and inexpensive instrumentation for time-resolved fluorescence measurements, similar strategies utilizing the compound-specific parameter fluorescence lifetime or fluorescence decay kinetics become increasingly attractive.[3-5] The potential of different types of QDs like II/VI, III/V and Cd-free ternary QDs such as AgInS (AIS) QDs for lifetime-based encoding and multiplexing has been, however, barely utilized, although the lifetimes of these nanocrystals cover a time windows which is barely accessible with other fluorophores. Here we present a brief insight into the photophysics of AIS QDs and show the potential of dye- and QD-encoded beads for lifetime-based encoding and detection schemes in conjunction with flow cytometry and fluorescence lifetime imaging microscopy T2 - Nanax 2019 CY - Hamburg, Germany DA - 16.09.2019 KW - Nano KW - Microparticle KW - Bead KW - Encoding KW - Lifetime KW - Multiplexing KW - Flow cytometry KW - Bead-based assay KW - Fluorescence KW - Dye KW - LT-FCM KW - Time-resolved flow cytometry KW - Method PY - 2019 AN - OPUS4-49039 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Quantification of Parameters Affecting the Upconversion Luminescence of Lanthanide-Based Upconversion Nanocrystals N2 - Lanthanide-based upconversion nanoparticles (UCNPs) like hexagonal Beta-NaYF4 UCNPs doped with Yb3+ and Er3+, which efficiently convert 976 nm light to ultraviolet, visible, and near infrared photons, offer new strategies for luminescence-based sensing, barcoding, and imaging. The properties of their upconversion (UC) luminescence (UCL) are, however, strongly influenced by particle size, the concentration and spatial arrangement of the dopant ions, surface chemistry including presence and thickness of surface passivation and shielding shells, microenvironment/presence of quenchers with high energy vibrations, and excitation power density (P). We present here a comprehensive study of the influence of UCNP size and particle architecture for Yb3+ and Er3+ co-doped NaYF4 core-only and core-shell nanostructures in the size range of about 5 nm to 50 nm, which underlines the importance of particle synthesis, surface chemistry, and quantitative luminescence measurements for mechanistic insights and the determination of application-relevant matrix- and P-dependent optimum dopand concentrations. T2 - Materials Challenges in Alternative and Renewable Energy (MCARE) 2018 CY - Vancouver, BC, Canada DA - 20.08.2018 KW - Upconverting nanoparticles KW - Size KW - Energy transfer KW - Fluorescence KW - Absolute fluorescence quantum yield KW - Fluorescence decay kinetics KW - Power density dependence PY - 2018 AN - OPUS4-46393 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Spectroscopic characterization of semiconductor and lanthanide-based nanocrystals with vis and NIR emission N2 - The increasing interest in molecular and nanoscale emitters with photoluminescence > 800 nm and recently also > 1000 nm for bioanalysis, medical diagnostics, bioimaging, and safety Barcodes requires quantitative spectroscopic studies, which are, however still challenging in this long wavelength region. This is of special relevance for nanocrystalline emitters like semiconductor quantum dots and rods as well as lanthanide-based upconversion and downconversion nanocrystals, where surface states and the accessibility of emissive states by quenchers largely control accomplishable photoluminescence quantum yields and hence, signal sizes and detection sensitivities from the reporter side. Moreover, nonlinear emitters like lanthanide-based upconversion nanocrystals require also power density-dependent studies of their luminescence spectra, quantum yields, and decay kinetics. Here, we present suitable absolute methods and underline the impact of such measurements on a profound mechanistic understanding of the nonradiative deactivation pathways in semiconductor and upconversion nanocrystals of different chemical composition and particle architecture. T2 - PCNSPA 2018 - Photonic Colloidal Nanostructures: Synthesis, Properties, and Applications CY - St. Petersburg, Russia DA - 04.06.2018 KW - Semiconductor KW - Nanoparticle KW - Quantum dot KW - Absolute fluorometry KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Fluorescence KW - Surface chemistry KW - Deactivation pathways KW - Photophysics KW - Modeling PY - 2018 AN - OPUS4-45796 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Automated determination of genotoxicity of nanoparticles with DNA-based optical assays - The NANOGENOTOX project N2 - The overall interest in nanotoxicity, triggered by the increasing use of nanomaterials in the material and life sciences, and the synthesis of an ever increasing number of new functional nanoparticles calls for standardized test procedures1,2 and for efficient approaches to screen the potential genotoxicity of these materials. Aiming at the development of fast and easy to use, automated microscopic methods for the determination of the genotoxicity of different types of nanoparticles, we assess the potential of the fluorometric γH2AX assay for this purpose. This assay, which can be run on an automated microscopic detection system, relies on the detection of DNA double strand breaks as a sign for genotoxicity3. Here, we provide first results obtained with broadly used nanomaterials like CdSe/CdS and InP/ZnS quantum dots as well as iron oxide, gold, and polymer particles of different surface chemistry with previously tested colloidal stability and different cell lines like Hep-2 and 8E11 cells, which reveal a dependence of the genotoxicity on the chemical composition as well as the surface chemistry of these nanomaterials. These studies will be also used to establish nanomaterials as positive and negative genotoxicity controls or standards for assay performance validation for users of this fluorometric genotoxicity assay. In the future, after proper validation, this microscopic platform technology will be expanded to other typical toxicity assays. T2 - SPIE 2018 CY - San Francisco, USA DA - 27.01.2018 KW - Nanoparticle KW - Fluorescence KW - Surface chemistry KW - Size KW - Assay KW - Microscopy KW - Nanotoxicity KW - Toxicity KW - Automation KW - Calibration KW - Standard PY - 2018 AN - OPUS4-44186 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Geißler, Daniel T1 - NanoGenotox - Automatable Determination of the Genotoxicity of Nanoparticles with DNA-based Optical Assays N2 - Nanomaterials are used in many different applications in the material and life sciences. Examples are optical reporters, barcodes, and nanosensors, magnetic and optical contrast agents, and catalysts. Due to their small size and large surface area, there are also concerns about their interaction with and uptake by biological systems. This has initiated an ever increasing number of cyctoxicity studies of nanomaterials of different chemical composition and surface chemistry, but until now, the toxicological results presented by different research groups often do not address or differ regarding a potential genotoxicity of these nanomaterials. This underlines the need for a standardized test procedure to detect genotoxicity.1,2 Aiming at the development of fast, easy to use, and automatable microscopic methods for the determination of the genotoxicity of different types of nanoparticles, we assess the potential of the fluorometric γH2AX assay for this purpose. This assay, which can be run on an automated microscopic detection system, relies on the determination of DNA double strand breaks as a sign for genotoxicity.3 Here, we present first results obtained with broadly used nanomaterials like CdSe/CdS and InP/ZnS quantum dots as well as iron oxide, gold, and polymer particles of different surface chemistry with previously tested colloidal stability. These studies will be also used to establish nanomaterials as positive and negative genotoxicity controls or standards for assay performance validation for users of this fluorometric genotoxicity assay. In the future, after proper validation, this microscopic platform technology will be expanded to other typical toxicity assays. References. (1) Landsiedel, R.; Kapp, M. D.; Schulz, M.; Wiench, K.; Oesch, F., Reviews in Mutation Research 2009, 681, 241-258. (2) Henriksen-Lacey, M.; Carregal-Romero, S.; Liz-Marzán, L. M., Bioconjugate Chem. 2016, 28, 212-221. (3) Willitzki, A.; Lorenz, S.; Hiemann, R.; Guttek, K.; Goihl, A.; Hartig, R.; Conrad, K.; Feist, E.; Sack, U.; Schierack, P., Cytometry Part A 2013, 83, 1017-1026. T2 - 9th International Conference on Nanotoxicology - New tools in risk assessment of nanomaterials CY - Dusseldorf/Neuss, Germany DA - 18.09.2018 KW - Nano KW - Nanotoxicity KW - Fluorescence KW - Quantum dot KW - Surface KW - Passivation shell KW - Automated assay KW - Nanoparticle PY - 2018 AN - OPUS4-47540 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Absolute Fluorescence Measurements > 800 nm - Setup Design, Challenges, and Characterization of Semiconductor and Lanthanide-based Nanocrystals N2 - There is an increasing interest in optical reporters like semiconductor and lanthanide-based nanocrystals with emission > 800 nm and recently also > 1000 nm for bioanalysis, medical diagnostics, and safety barcodes. Mandatory for the comparison of different emitter classes and the rational design of the next generation of reporters for the short wavelength infrared (SWIR) region are reliable and quantitative photoluminescence measurements in this challenging wavelength region. This is of special relevance for nanocrystalline emitters like semiconductor quantum dots and rods as well as for upconversion and downconversion nanocrystals, where surface states and the accessibility of emissive states by quenchers largely control accomplishable quantum yields and hence, signal sizes and detection sensitivities from the reporter side. Such measurements are currently hampered by the lack of suitable methods and standards for instrument calibration and validation as well as by the lack of quantum yield standards with emission > 800 nm and especially > 1000 nm. In this respect, we present the design of integrating sphere setups for absolute and excitation power densitydependent measurements of emission spectra and quantum yields in the wavelength region of 650 to 1650 nm including calibration strategies and first candidates for potential fluorescence standards. Subsequently, the photoluminescence properties of different types of nanocrystals are presented and discussed including absolute photoluminescence measurements of upconversion and down conversion emission in different solvents. T2 - MRS 2017 CY - Boston, MA, USA DA - 26.11.2017 KW - Upconversion KW - Nanoparticle KW - Absolute fluorometry KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Fluorescence KW - Surface chemistry KW - Deactivation pathways KW - Semiconductor quantum dot KW - SWIR KW - Quantum yield KW - Energy transfer KW - Size PY - 2017 AN - OPUS4-43202 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Effect of Particle Size and Excitation Power Density on the Luminescence Efficiency of Upconversion Nanocrystals in Different Dispersion Media N2 - Upconversion nanoparticles (UCNPs) offer new strategies for luminescence-based sensing and imaging. One of the best studied materials are ..-NaYF4 UCNPs doped with 20 % Yb3+ and 2 % Er3+, which efficiently convert 976 nm light to photons emitted at 540 nm, 655 nm, and 845 nm, respectively, reveal long luminescence lifetimes (> 100 µs), are photostable and chemically inert. Their upconversion (UC) luminescence (UCL) properties are, however, strongly influenced by particle size, surface chemistry, and microenvironment. In addition, the multiphotonic absorption processes responsible for UCL render UCL excitation power density (..) dependent. This makes quantitative UCL measurements as well as the determination of UC quantum yields (.UC) very challenging. The rational design of brighter UCNPs particle architectures and the interest in identifying optimum particle architectures for FRET-based sensing and imaging schemes, which often rely on core-only UCNPs to minimize donor-acceptor distances, encouraged us to assess the influence of particle size and P on UCL. Here, we present the photophysical properties of a series of hexagonal NaYF4 UCNPs with sizes from 10 to 43 nm with different surface ligands dispersed in organic solvents and water studied by steady state and time-resolved fluorometry as well as quantitatively by integrating sphere spectrometry with P varied over about three orders of magnitude. Our results underline the need for really quantitative luminescence studies for mechanistic insights and the potential of high P to compensate for UCL quenching due to high energy phonons and surface effects. T2 - 15th Conference on Methods and Applications in Fluorescence CY - Bruges, Belgium DA - 10.09.2017 KW - Upconversion KW - Nanoparticle KW - Absolute fluorometry KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Fluorescence KW - Surface chemistry KW - Deactivation pathways KW - Photophysics KW - Quantum yield PY - 2017 AN - OPUS4-43181 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Spectroscopic properties of upconversion nanoparticles N2 - Lanthanide-doped up-converting nanoparticles (UCNPs) are promising reporters for medical diagnostics and bioimaging. Current limitations present their relative low absorption cross sections and low fluorescence efficiencies, with the latter being affected by particle size, surface chemistry, and microenvironment, particularly water. Here, we present results from systematic studies of the excitation power density dependent upconversion luminescence spectra, intensities/intensity ratios of the individual emission bands, slope factors, and quantum yields of UCNPs of varying size, dopant concentration, and surface chemistry in different microenvironments as well as the up- and downconversion luminescence decay kinetics of the different emission bands. Moreover, first studies of the energy transfer from UCNPs to surface-bound organic dyes acting as fluorescence acceptors are shown. Based upon these measurements, fluorescence deactivation channels are identified and spectroscopic parameters for the screening of material performance are derived. T2 - Projekttreffen COST CY - Straßbourg, France DA - 09.01.2017 KW - Upconverting nanoparticles KW - Size KW - FRET KW - Fluorescence KW - Absolute fluorescence quantum yield KW - Fluorescence decay kinetics KW - Power density dependence PY - 2017 AN - OPUS4-39076 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Recent insights in the spectroscopic properties of upconversion nanoparticles N2 - Lanthanide-doped up-converting nanoparticles (UCNPs) are promising reporters for medical diagnostics and bioimaging, which are excited in the near infrared (NIR) by multiphoton absorption processes, and show multiple narrow emission bands in the visible (vis) and NIR, long luminescence lifetimes in the μs range, and excellent photostability. Current limitations present their relative low absorption cross sections and low fluorescence efficiencies, with the latter being affected by particle size, surface chemistry, and microenvironment, particularly water. Here, we present results from systematic studies of the excitation power density dependent upconversion luminescence spectra, intensities/intensity ratios of the individual emission bands, slope factors, and quantum yields of UCNPs of varying size, dopant concentration, and surface chemistry in different microenvironments as well as the up- and downconversion luminescence decay kinetics of the different emission bands. Moreover, first studies of the energy transfer from UCNPs to surface-bound organic dyes acting as fluorescence acceptors are shown. Based upon these measurements, fluorescence deactivation channels are identified and spectroscopic parameters for the screening of material performance are derived. T2 - Spie Photonics west 2017 CY - San Francisco, USA DA - 28.01.2017 KW - Upconverting nanoparticles KW - Size KW - FRET KW - Fluorescence KW - Absolute fluorescence quantum yield KW - Fluorescence decay kinetics KW - Power density dependence PY - 2017 AN - OPUS4-39075 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Semiconductor nanocrystals with VIS and NIR/IR emission - spectroscopy properties and surface chemistry N2 - Semiconductor nanocrystals with a spherical (QDs) core and a spherical or a rod-shaped Shell, u.a., so-called Quantum dot-Quantum rods (QDQRs) are increasingly used as fluorescent Reporters or optically active components in the life and material science, e.g., in solid state lightening including Plasma Displays. (1,2) Morever, there is an increasing interest in materials with emission >800 nm for bioanalysis, medical diagnostics, and safety barcodes. Prerequisites for the mechanistic understanding of nonradiativ decay channels needed for the rational design of improved nanomaterials and the comparison of material Performance are reliable fluorescence measurements and validated methods for the assessment of their surface chemistry. (3,4) The latter is of particular importance for nanocrystalline Emitters, where surface states and the accessibility of emissive states by quenchers largely control photoluminescence properties. (5) Here, we present results from systematic spectroscopic studies including absolutely measured photolumunescence Quantum yields of different vissible and NIR emisisve QD and QDQRs Systems of varying particle architecture size and surface chemistries in Dispersion and embedded in salt crystals. (6,7) T2 - MCare 2017 CY - Jeju, South Korea DA - 20.02.2017 KW - Semiconductor quantum dot KW - Nanoparticle KW - Absolute fluorometry KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Fluorescence KW - Surface chemistry KW - Photophysics KW - Quantum yield KW - Single particle PY - 2017 AN - OPUS4-43133 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Quantitative Characterization of Functional Nanomaterials with vis/NIR Emission N2 - The rational design of functional nanomaterials for optical applications in the material and life sciences requires optical-spectroscopic methods for the quantitative characterization of their signal-relevant optical properties. Additionally, methods for the simple and quantitative analysis of the surface chemistry are desired as the chemical nature and number of the surface groups and ligands can affect the optical features and controls the interaction of these nanomaterials with their environment. Here, we present quantitative photoluminescenvce studies of different types of vis/NIR-emissive nanomaterials like semiconductor quantum dots and upconversion nanocrystals will be presented and their relevance for the mechanistic understanding of nonradiative decay channels and the rational design of new nanomaterials will be underpinned. In this respect, also validation concepts for such measurements and absolute fluorometry will be introduced. In addition, the potential of optical spectroscopy for surface group and ligand analysis surface chemistry will be demonstrated exemplarily for semiconductor quantum dots T2 - AK Prof. Voss CY - Uni Brunswick, Germany DA - 03.08.2017 KW - Semiconductor KW - Upconversion KW - Nanoparticle KW - Fluorescence KW - Surface group analysis KW - NIR KW - Absolute fluoreometry KW - Integrating sphare spectroscopy KW - Optical probe KW - Assay KW - Thiol ligand PY - 2017 N1 - Geburtsname von Nirmalananthan-Budau, Nithiya: Nirmalananthan, N. - Birth name of Nirmalananthan-Budau, Nithiya: Nirmalananthan, N. AN - OPUS4-41366 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Challenges and examples for quantitative fluorescence measurements > 800 nm with semiconductor and lanthanide-doped nanocrystals N2 - There is an increasing interest in molecular and nanoscale with emission > 800 nm and recently also > 1000 nm for bioanalysis, medical diagnostics, bioimaging, and safety barcodes. Mandatory for the comparison of different emitter classes and the rational design of the next generation of reporters for the short wavelength infrared (SWIR) Region are reliable and quantitative photoluminescence measurements in this challenging wavelength region. This is of special relevance for nanocrystalline emitters like semiconductor quantum dots and rods as well as lanthanide-based upconversion and downconversion nanocrystals, where surface states and the accessibility of emissive states by quenchers largely control accomplishable photoluminescence quantum yields and hence, signal sizes and detection sensitivities from the reporter side. Such measurements are currently hampered by the lack of suitable methods and standards for instrument calibration and validation and quantum yield standards with emission > 800 nm and especially > 1000 nm. In this respect, we present the design of integrating sphere setups for absolute and excitation power density-dependent measurements of emission spectra and photoluminescence quantum yields in the wavelength Region of 650 to 1650 nm including calibration strategies and first candidates for potential fluorescence standards. Subsequently, the photoluminescence properties of different types of nanocrystals are presented including the upconversion and downconversion emission of differently sized and surface functionalized lanthanide-doped nanoparticles and photoluminescence quenching effects are quantified. T2 - SHIFT 2017 CY - Teneriffa, Spain DA - 13.11.2017 KW - Upconversion KW - Nanoparticle KW - Absolute fluorometry KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Fluorescence KW - Surface chemistry KW - Deactivation pathways KW - Semiconductor quantum dot KW - SWIR KW - Quantum yield PY - 2017 AN - OPUS4-43203 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - The European Upconversion Network: From the Design of Photon-upconverting Nanomaterials to (Biomedical) Applications (CM1403) WG1 Materials Research & Photophysical Characterization N2 - Lanthanide-doped photon-upconversion nanoparticles (UCNPs) have been in the focus of many research activities in the material and life sciences over the last 15 years because of their potential to convert light between different spectral regions and their unique photophysical properties. In order to fully exploit the application potential of these fascinating nanomaterials, a number of challenges has to be overcome such as the low brightness particularly of small UCNPs and the reliable quantification of the excitation power density (P)-dependent upconversion luminescence (UCL). Here, the need and requirements on the characterization of the optical properties of UCNPs are discussed with special focus on the reliabiliy and comparability of relative and quantitative luminescence measurements and prerequisites for their standardization. T2 - COST Treffen CY - Aveiro, Portugal DA - 26.06.2017 KW - Upconversion KW - Nanoparticle KW - Fluorescence KW - NIR KW - Absolute fluorometry KW - Integrating sphere spectroscopy KW - Reference maerial KW - Standardization KW - Quantum yield PY - 2017 AN - OPUS4-41367 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Instrument Calibration and Standardization of Fluorescence Measurements in the UV/vis/NIR/IR N2 - Comparison of fluorescence measurements performed on different fluorescence instruments, analyte quantification from fluorescence intensities as well as the determination of fluorescence quantum yields require instrument calibration and consideration of the wavelength-dependent instrument-specific quantities spectral photon flux reaching the sample and spectral responsivity. Here, we present guidelines and recommendations for the qualification of fluorescence instruments and introduce suitable chromophore-based reference materials. Moreover, the design concepts of the different BAM fluorescence standards are discussed. T2 - COST 2017 CY - Turku, Finland DA - 03.04.2017 KW - Fluorescence KW - Dye KW - Glass KW - Calibration KW - Fluorescence standard KW - Integrating sphere spectroscopy KW - Fluorescence quantum yield KW - Instrument qualification KW - Quality assurance KW - NIR KW - IR PY - 2017 AN - OPUS4-43174 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Effect of surface chemistry on optical, chemical and electronic properties of blue luminescent graphene quantum dots N2 - Due to their unique physical properties, particularly their electronic and luminescent properties, graphene quantum dots (GQDs) are expected to be suitable for a wide range of applications in bioimaging, electro-optical and photonic materials or energy harvesting among others.1 Tuning the surface chemistry provides an efficient approach to modulate the fluorescence and distinct electronic properties of GQDs.2 Nevertheless, the role of surface chemistry on the electronic structure of GQDs remains poorly understood. In this presentation, we will compare systematically the electronic and chemical structures of GQDs functionalized with carboxylic and aminated groups to those of non-functionalized GQDs, combining theoretical and experimental approaches, here various photon-based spectroscopies. First, the electronic structure of GQDs was characterized by soft X-ray absorption (XA) and X-ray emission (XE) spectroscopies, probing unoccupied and occupied electronic states, respectively, at the carbon K edge for the first time. The interpretation of the XA/XE spectra was done based on theoretical calculations. Then, the chemical structure of the GQDs was characterized in situ by ATR-FTIR in water, thereby accounting for the importance of the interface between GQDs and water believed to play a central role in the chemical reactivity and the optical properties. We previously demonstrated that monitoring the OH vibrations of water molecules during exposure to humid air was a powerful method to probe H-bonding environment around carbon nanomaterials.3 For GQDs, clear surface-dependent water adsorption profiles are observed and discussed. Finally, UV/Vis absorption and photoluminescence measurements were done to characterize the optical properties of these GQDs. Our results suggest that the surface chemistry of the GQDs affects significantly their electronic structure and optical properties. These findings will contribute to an improved understanding of the structure–activity relationship of GQDs and other carbon nanomaterials with surface modifications. T2 - MRS Fall Meeting 2017 CY - Boston, USA DA - 26.11.2017 KW - Carbon KW - Nanoparticle KW - Fluorescence KW - NIR KW - IR KW - Surface chemistry KW - Deactivation pathways KW - Lifetime KW - Size KW - Giant carbon dot KW - Quantum yield PY - 2017 AN - OPUS4-43494 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Wagner, Sabine T1 - Combining fluorescent molecularly imprinted polymer sensor particles with microfluidic devices for the detection of herbicides N2 - 2,4-Dichlorophenoxyacetic acid (2,4-D) is one important and well-known herbicide that is widely used in agriculture because of its advantages to regulate plant growth. However, the use of large quantities of the treated plants as animal feed leads to residues in meat, milk and eggs. Furthermore, the herbicide can drain away and contaminate ground and drinking water. The ingestion of 2,4-D-contaminated food and water causes damage to the inner organs of humans and animals, e.g., the kidneys and the liver. Analytical assays based on molecularly imprinted polymers (MIPs) have emerged as a valuable tool in the field of environmental analysis due to the low production costs, stability, format adaptability and the possibility to imprint and thus their ability to recognize a wide variety of target analytes. With regard to optical sensing technologies, however, MIPs have only been used in considerably few applications, especially in fluorescence sensors. This limitation is basically due to the fact that the incorporation of a fluorescently responding moiety into a polymer matrix is challenging. One way to overcome this limitation is to use tailor-made fluorescent indicator monomers for direct transfer of the binding event into an optical signal and coat the MIP via reversible addition-fragmentation chain transfer (RAFT) polymerization as a thin layer onto the surface of silica nanoparticles as primary sensing element. Here, we present the response behavior of the fluorescent MIP sensor particles in terms of sensitivity for 2,4-D detection (the so-called imprinting factor), discrimination ability against structurally similar compounds and performance in a phase-transfer assay (PTA) on chip, i.e., the implementation of the assay into a microfluidic chip environment, offering a novel simple and rapid way for the detection of herbicides. T2 - 10. Interdisziplinäres Doktorandenseminar CY - Berlin, Germany DA - 28.02.2016 KW - Molecularly imprinted polymer KW - Fluorescence KW - Microfluidic devices PY - 2016 AN - OPUS4-38165 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Methods for the determination of the optical properties and the surface chemistry of fluorescent particles N2 - Fluorescent particles like nm- and m-sized polymeric beads doped or labeled with different types of fluorophores and nanocrystalline systems like quantum dots and upconversion phosphors emitting in the visible (vis), near-infrared (NIR), and IR (infrared) region are of increasing importance as fluorescent reporters for bioanalysis and medical diagnostics. The assessment and comparison of material performance and the development of rational design strategies for improved systems requires suitable spectroscopic tools for the determination of signal-relevant optical properties and analytical tools for the determination of the number of surface groups, ligands, biomolecules and /or fluorophores per bead. In this respect, suitable spectroscopic tools for the characterization of the optical properties of such materials like photoluminescence quantum yields and brightness values and the determination of their surface chemistry are introduced. This includes integrating sphere setups for absolute measurements of fluorescence quantum yields of liquid and solid, transparent and scattering materials in the wavelength region of 350 nm to 1600 nm at varying excitation power densities for the study of multi-photon processes and simple optical assays, validated by comparison with established analytical techniques relying on different detection principles. Here, different examples for the optical and analytical characterization of different types of nanoscale reporters are presented. T2 - MoLife Research Seminar CY - Bremen, Germany DA - 26.04.2016 KW - Optical assay KW - Fluorescence KW - Integrating sphere spectroscopy KW - Quantum yield KW - NIR KW - IR KW - Optical spectroscopy KW - Nanomaterials KW - Semiconducor nanocrystals KW - Upconversion nanocrystals KW - Surface analysis KW - Ligand analysis PY - 2016 AN - OPUS4-35954 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Absolute spectroscopic characterization of the optical properties of semiconductor and upconversion nanocrystals in the vis and IR N2 - Nanocrystalline fluorophores like semiconductor quantum dots and rods and recently also lanthanide-based upconversion phosphors with emission in the visible (vis), near-infrared (NIR), and IR (infrared) region are increasingly being used in bioimaging studies and fluorescence assays as well as in photovoltaics and solid state lighting. The assessment and comparison of material performance as well as the development of rational design strategies for improved systems require spectroscopic tools, which enable the determination of the signal-relevant optical properties like photoluminescence quantum yields and brightness values. In the case of nonlinear fluorescence as shown by upconversion materials, such measurements must be also performed as function of excitation power density. In this work, we report on methods for the absolute determination of the photoluminescence quantum yield and brightness of fluorescent particles in dispersion and as powders based on integrating sphere spectroscopy and underline the importance of such measurements for the understanding of the photophysics of such nanocrystals. T2 - International Conference on Fundamental Processes in Semiconductor Nanocrystals (FQDots16) CY - Berlin, Germany DA - 05.09.2016 KW - Fluorescence KW - Nanoparticle KW - Semiconductor quantum dot KW - Upconversion nanocrystal KW - NIR KW - Integrating sphere spectroscopy KW - Fluorescence quantum yield KW - Method PY - 2016 AN - OPUS4-38695 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Characterization of nanocrystalline particles with NIR emission - Spectroscopic properties and surface group analysis N2 - There is an increasing interest in optical reporters like semiconductor quantum dots and upconversion nanocrystals with emission > 800 nm for bioanalysis, medical diagnostics, and safety barcodes. Prerequisites for the comparison of material performance, the mechanistic understanding of nonradiative decay channels, and the rational design of new nanomaterials with improved properties are reliable fluorescence measurements and validated methods for the assessment of their surface chemistry. The latter is of special relevance for nanocrystalline emitters, where surface states and the accessibility of emissive states by quenchers largely control accomplishable photoluminescence quantum yields and hence, signal sizes and detection sensitivities from the reporter side. Here, we present the design of integrating sphere setups for the excitation power density-dependent absolute measurement of emission spectra and photoluminescence quantum yields in the wavelength region of 350 to 1600 nm and results from spectroscopic studies of semiconductor quantum dots and upconversion nanocrystals of different size and surface chemistries in various environments. Subsequently, examples for simple approaches to surface group and ligand analysis are presented. T2 - 1st International Biophotonics Conference CY - Singapure DA - 25.07.2016 KW - Fluorescence KW - Integrating sphere spectroscopy KW - Quantum yield KW - NIR KW - IR KW - Optical spectroscopy KW - Nanomaterials KW - Semiconducor nanocrystals KW - Surface analysis KW - Ligand analysis KW - Optical assay PY - 2016 AN - OPUS4-36998 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Suitable geometries for the measurement of photoluminescence quantum yields of luminescent and scattering samples – The DIN TNS project N2 - The characterization of the optical properties of photoluminescent systems, that scatter, like dispersions of nanoparticles with sizes exceeding about 25 nm or solid nanophosphors is of increasing importance for many applications in the life and material sciences. Examples present nanoscale optical reporters and dye-doped microparticles for bioimaging, fluorescence assays or DNA sequencing as well as nanocrystalline emitters like semiconductor quantum dots and rods or lanthanide-based nanophosphors embedded into solid matrices for solid state lighting, display technologies, or barcoding/security applications. The assessment and comparison of material performance as well as the development of rational design strategies for improved systems require spectroscopic tools, which enable the determination of the signal-relevant optical properties like photoluminescence quantum yields and brightness values. This encouraged us to built up an integrating sphere setup enabling absolute measurements of photoluminescence spectra and quantum yields of transparent and scattering photoluminescent dispersions and solid samples in different measurement geometries, i.e., direct and indirect illumination and the combination of both geometries and perform first measurements with selected emitters. Here, the design of this setup is presented and first recommendations concerning suitable measurement geometries are given. T2 - DKE-Sitzung CY - Frankfurt am Main, Germany DA - 31.08.2016 KW - Nanoparticle KW - Integrating sphere KW - Fluorescence KW - Quantum yield KW - Method KW - Standardization KW - Calibration KW - Reference material PY - 2016 AN - OPUS4-38643 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Simple methods, validation concepts, and reference materials for the characterization of functional nanomaterials and microparticles N2 - The surface chemistry / functionalization of nanomaterials and microparticles largely controls the stability of these materials as well as their solubility and subsequent biofunctionalization and their interactions with biological systems. Moreover, in the case of some nanomaterials like semiconductor quantum dots or lanthanide-based upconversion nanocrystals, the ligand shell strongly affects their optical properties, e.g., via passivation of surface states and traps that favor luminescence quenching or the protection of surface atoms from quenching water molecules. This renders analytical methods for the quantification of surface groups like functionalities very important. Targets of broad interest are here amino, carboxyl, alkine and maleimide groups used for common bioconjugation reactions and typical ligands like thiols and polyethylene glycol (PEG) molecules of varying length, used for the tuning of material hydrophilicity and biocompatibility, minimization of unspecific interactions, prevention of biofouling, and enhancement of blood circulation times as well as surface-bound biomolecules like streptavidin or other biomolecules relevant e.g., for diagnostic assays. Here, we focus on simple optical methods relying on standard laboratory instrumentation, validated by method comparison and/or mass balances and present examples for their use for the characterization of different types of nanomaterials and microparticles. T2 - Innovationsforum Senftenberg CY - Senftenberg, Germany DA - 01.06.2016 KW - Surface chemistry KW - Functional group analysis KW - Thiol assay KW - Fluorescence KW - Nanomaterial KW - Nanoparticle KW - PEG KW - Ligand KW - Semiconductor quantum dot KW - Quantum yield KW - Quantification KW - Method validation KW - Integrating sphere spectroscopy KW - Fluorescence standard PY - 2016 N1 - Geburtsname von Nirmalananthan-Budau, Nithiya: Nirmalananthan, N. - Birth name of Nirmalananthan-Budau, Nithiya: Nirmalananthan, N. AN - OPUS4-37111 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Standardization of fluorescence measurements: Design, development, validation, and fabrication of format-adaptable fluorescence standards for intensity, spectral, and temporal quantities N2 - Photoluminescence techniques are amongst the most widely used tools in the material and life sciences, with new and exciting applications continuously emerging, due to their many advantages like comparative ease of use, unique sensitivity, non-invasive character, and potential for multiplexing, remote sensing, and miniaturization. Drawbacks are , however, signals, that contain unwanted wavelength- and polarization contributions from instrument-dependent effects, which are time-dependent due to the aging of instrument components, and difficulties to measure absolute fluorescence intensities. Thus, there is a considerable need for standards for intensity, spectral, and temporal fluorescence quantities to meet the increasing need for instrument performance validation and global trends to harmonize physicochemical measurements. In this respect, instrument calibration strategies together with different types of fluorescence standards are presented as well as design concepts for robust, easy-to-use, and format-adaptable fluorescence standards useable for the determination of different fluorescence parameters and a broad variety of fluorescence techniques. T2 - SALSA-Kolloquien CY - Berlin, Germany DA - 07.06.2016 KW - Fluorescence KW - Standard KW - Quality assurance KW - Spectral correction KW - Quantum yield PY - 2016 AN - OPUS4-37070 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Applications and challenges of luminescence-based detection methods in the life and material sciences N2 - Luminescence-based detection methods, ranging from fluorescence spectroscopy for photophysical and mechanistic studies over sensing applications, chromatographic separation techniques and the microarray technology with fluorescence detection to fluorescence microscopy, flow cytometry, single molecule spectroscopy, and molecular imaging to integrating sphere spectroscopy, are among the most widely used methods in the life and material sciences. This is due to e.g., their unique sensitivity enabling the detection of single molecules, potential for multiplexing, ease of combination with spatial resolution, and suitability for remote sensing. Many of these advantages are closely linked to the choice of suitable molecular and nanoscale fluorescent reporters, typically required for signal generation. This includes organic dyes without and with sensor function, fluorophore-encoded polymeric and silica nanoparticles as well as nanocrystalline systems like semiconductor quantum dots and upconversion phosphors, emitting in the visible (vis), near-infrared (NIR), and IR (infrared). Current challenges present the environment sensitivity of most fluorophores, rendering fluorescence spectra, measured intensities/fluorescence quantum yields, and fluorescence decay kinetics matrix-dependent, and instrument-specific distortions of measured fluorescence signals that need to be considered for quantification and comparability of data, particularly fluorescence spectra. Here, current applications of luminescence-based methods and different types of reporters will be presented. In this context, suitable spectroscopic tools for the characteri-zation of the optical properties of fluorescent reporters and fluorophore-encoded microparticles, analytical tools for the determination of the surface chemistry of different types of particles, and different multiplexing strategies will be discussed. T2 - 9th Meeting of Engineering of Functional Interfaces CY - Wildau,Germany DA - 03.07.2016 KW - Fluorescence KW - Multiplexing KW - Lifetime KW - Nanomaterial KW - Nanoparticle KW - PEG KW - Ligand KW - Semiconductor quantum dot KW - Quantum yield KW - Quantification KW - Upconversion nanoparticle KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Fluorescence standard KW - Calibration PY - 2016 AN - OPUS4-37112 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -