TY - CONF A1 - Melzer, Michael T1 - Flexible GMR sensors for novel flux leakage testing capabilities N2 - In order to detect small and hidden material defects or micro-cracks, weak and localized magnetic stray fields need to be detected and characterized on the magnetic surface. Current flux leakage sensor elements are rigid and rely on air gaps to detect stray fields several hundred micrometers above the surface. Furthermore, the increasing complexity of critical components, enabled by novel fabrication technologies, primarily additive manufacturing, poses a significant challenge to the available flux leakage testing technologies. The emergence of flexible magneto-electronics paves the way to a compliant form of magnetic sensors that, on one hand, are able to adapt to highly curved surfaces, allowing these measurements to be conducted on irregularly formed components, which are inaccessible for nowadays electromagnetic NDT equipment and, on the other hand, could enhance the detection limit for small magnetic stray fields, by allowing the magneto-sensitive elements to approximate the object’s surface until only a few µm, mediated by thin and flexible packaging. T2 - 38th Danubia-Adria Symposium on Advances in Experimental Mechanics CY - Poros, Greece DA - 20.09.2022 KW - Non-destructive testing KW - Flux leakage testing KW - Flexible sensorics KW - GMR Sensors KW - Flexible electronics PY - 2022 AN - OPUS4-56628 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Nowatzky, Yannek T1 - Mistle: Metaproteomic index and spectral library search engine N2 - Introduction: With the introduction of accurate deep learning predictors, spectral matching applications might experience a renaissance in tandem mass spectrometry (MS/MS) driven proteomics. Deep learning models, e.g., Prosit, predict complete MS/MS spectra from peptide sequences and give the unprecedented ability to accurately predict mass spectra that may arise from any given proteome. However, the amount of spectral data is enormous when querying large search spaces, e.g., metaproteomes composed of many different species. Current spectral library search software, such as SpectraST, is not equipped to meet run time and memory constraints imposed by such large MS/MS databases, covering several millions of peptide spectrum predictions. Methods: Inspired by the fragment index data structure that had been introduced with MSFragger, we implement an efficient peak matching algorithm for computing spectral similarity between query and library spectra. Mistle (Metaproteomic index and spectral library search engine) uses index partitioning and SIMD (Single instruction, multiple data) intrinsics, which greatly improves speed and memory efficiency for searching large spectral libraries. Mistle is written in C++20 and highly parallelized. Results: We demonstrate the efficiency of Mistle on two predicted spectral libraries for the lab-assembled microbial communities 9MM and SIHUMIx. Compared to the spectral library search engine SpectraST, Mistle shows a >10-fold runtime improvement and is also faster than msSLASH, which uses locality-sensitive hashing. Although Mistle is slower than MSFragger, Mistle‘s memory footprint is an order of magnitude smaller. Furthermore, we find evidence that the spectral matching approach to predicted libraries identifies peptides with higher precision. Mistle detects peptides not found by database search via MSFragger and in turn uncovers unnoticed false discoveries among their matches. Conclusion: In this study, we show that predicted spectral libraries can enhance peptide identification for metaproteomics. Mistle provides the means to efficiently search large-scale spectral libraries, highlighted for the microbiota 9MM and SIHUMIx. T2 - HUPO2022 CY - Cancun, Mexico DA - 04.12.2022 KW - Proteomics KW - Mass spectrometry KW - Algorithms KW - Metaproteomics PY - 2022 AN - OPUS4-56695 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ermilova, Elena T1 - Modell based ellipsometry standards for different application cases and their application for real non-ideal systems N2 - Ellipsometry enables the investigation of material properties over a broad spectral range. As a fast and non-destructive method, it is widely used in industry for quality assurance. Despite the wide application of ellipsometry and its high industrial relevance there are some material-specific standards and there have been no generally accepted standards dealing with model validation and measurement uncertainties. The first German standard DIN50989-1: 2018 Ellipsometry – Part1: Principles (currently international standard ISO 23131: 2021) marks the beginning of a 6-part standard series for ellipsometry, which was developed under consideration of GUM. T2 - International Conference on Spectroscopic Ellipsometry (ICSE-9) CY - Beijing, China DA - 22.05.2022 KW - Spectroscopic and imaging ellipsometry KW - Standards KW - Uncertainties PY - 2022 AN - OPUS4-56698 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ermilova, Elena T1 - Hybrid optical measurement technique for detection of defects in epitaxially grown 4H-SiC layers N2 - Recent developments in power electronics require the use of new wide bandgap compound semiconductors. Silicon carbide (SiC) is one of the most promising materials for power electronics due to its outstanding properties and commercial availability. Some types of defects in the SiC substrates or homoepitaxial SiC layers can affect the performance of electronic devices in a serious manner or make its operation even impossible. Optical methods such as imaging ellipsometry (IE) and white light interference microscopy (WLIM) were applied for fast and non-contact investigation of defects in the epitaxially grown 12 µm 4H-SiC layers on 4H- SiC substrates. T2 - EOS Annual Meeting (EOSAM 2022) CY - Porto, Portugal DA - 12.09.2022 KW - Imaging ellipsometry KW - White light interference microscopy KW - Epitaxially grown 4H-SiC layers KW - Wide bandgap compound semiconductors PY - 2022 AN - OPUS4-56705 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Kazlagić, Anera A1 - Rosner, M. A1 - Vogl, Jochen T1 - Investigating the differences between MC-ICP-MS and MC-TIMS using conventional 87Sr/86Sr isotope ratios in limestone and slate reference materials N2 - The Federal Institute for Materials Research and Testing (BAM) organised an interlaboratory comparison (ILC) for the characterisation of 87Sr/86Sr isotope ratios in limestone (IAG/CGL ML-3) and Penrhyn slate (IAG OU-6) reference materials by applying the conventional method for 87Sr/86Sr isotope ratios. Samples were sent to thirteen analytical laboratories . Since both samples are powdered, rock materials, dissolution of the sample and Sr isolation via ion exchange chromatography were mandatory. This was done using acid, microwave/acid, bomb/acid digestion or borate fusion and subsequent isolation of Sr by means of commercially available ion exchange resins. In this study, we present and discuss the potential effects that differences between laboratories, and between two instrumental measurement techniques (i.e., MC-ICP-MS and MC-TIMS), may have upon the dispersion of measurement results of the 87Sr/86Sr isotope ratio in the two aforementioned reference materials. We used a statistical mixed effects model to assess the potential effects of both the laboratory and the measurement technique. Consensus values for both materials and associated standard uncertainties {(IAG/CGL ML-3 (0.708245±0.000004) mol/mol; IAG OU-6 (0.729769±0.000008) mol/mol} were estimated by fitting a linear, Gaussian mixed effects model (Pinheiro and Bates 2000) using the R function “lmer” defined in package “lme4”. The statistical results showed that there is no significant effect attributable to differences between instrumental techniques when both materials are considered together, or separately. The p-value of the test of significance of the measurement technique effect is greater than 0.54. For both materials there were statistically significant effects attributable to differences between laboratories when the measurement results for both materials were considered together and separately. This effect is less than 0.00004 in absolute value. However, for neither material did consideration or disregard for such differences induce significant changes in the estimate of the consensus value for the 87Sr/86Sr isotope ratio. Therefore, the effects attributable to differences between instrumental techniques or between laboratories can safely be disregarded when computing the best estimate for the true value of 87Sr/86Sr isotope ratio in these materials, by the community of expert laboratories represented in this study. T2 - GeoAnalysis 2022 CY - Freiberg, Germany DA - 06.08.2022 KW - Isotope ratio KW - Conventional isotope ratio KW - ILC KW - Traceability KW - Uncertainty KW - Isotope reference materials PY - 2022 AN - OPUS4-56848 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Johann, Sergej T1 - Multi-sensor system for long-term monitoring with WiFi and LoRaWAN technology N2 - In many cases, science falls back on self-developed prototype systems, which are used and developed for the measurement and execution of the tasks. About 80 % of the development is based on the same hardware design, which is used in only one application scenario and then discarded. For the most part, there are also uncalibrated sensors, since it is costly to calibrate a complicated sensor measurement system or the entire measurement chain since access to the sensor systems is not always available. This paper describes a conceptual design to implement a versatile sensor system with the motivation to fuse the data recording and data reception, which can cover large areas with the help of LoRaWAN (Long Range Wide Area Network) technology [1, 2]. To overcome disadvantages of LoRaWAN, namely the slow data rates, the proposed sensor system can also cover smaller areas with the widespread WiFi technology. An enormous advantage over individual complete systems in the form of a prototype, is the rapid expansion, uncomplicated calibration of the individual sensors and the ecological relief. A modular design is used, where individual stacks with sensors and peripherals can be added separately. The stacks are standalone low-power systems and can be calibrated, maintained, and replaced separately and do not require the entire measurement chain. The measured and sent values are stored locally on the main stack and sent to the data collector (gateway) and evaluated by means of the automated selection between WiFi and LoRaWAN. T2 - 38th Danubia-Adria Symposium on Advances in Experimental Mechanics CY - Athens, Greece DA - 20.09.2022 KW - Multisensor system KW - LoRaWAN KW - WiFi KW - Long term monitoring PY - 2022 AN - OPUS4-56464 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Piechotta, Christian T1 - EDC-WFD: A project to deliver reliable measurements for better monitoring survey and risks assessments. N2 - Monitoring programs should generate high-quality data on the concentrations of substances and other pollutants in the aquatic environment to enable reliable risk assessment. Furthermore, the need for comparability over space and time is critical for analysis of trends and evaluation of restoration of natural environment. Additionally, research work and exercises at the European level have highlighted that reliable measurements of estrogenic substances at the PNEC level are still challenging to achieve. The project EDC-WFD Metrology for monitoring endocrine disrupting compounds under the EU Water Framework Directive aims to develop traceable analytical methods for determining endocrine disrupting compounds and their effects, with a specific focus on three estrogens of the first watch list (17-beta-estradiol (17βE2), 17-alpha-ethinylestradiol (EE2), and estrone (E1)). Estrogens 17-alpha-estradiol (17E2) and estriol (E3) will be included to demonstrate the reliability of the developed methods and to support the requirements of Directive 2013/39/EC, Directive 2009/90/EC and Commission Implementation Decision (EU) 2018/840, hence improving the comparability and compatibility of measurement results within Europe. During the EDC-WFD project four selected effect-based methods (EBM) will be deeply investigated in order to improve their rationale use and their support in water quality assessment. In particular, the EBM sensitivity, specificity and accuracy on reference materials with single or mixture solutions of the five substances at a concentration of EQS values will be explored. This contribution will present the objectives and methods applied within the EDC-WFD project. T2 - EuChemS CY - Lisbon, Portugal DA - 29.08.2022 KW - WFD KW - Estrogens KW - Whole water samples KW - EQS PY - 2022 AN - OPUS4-57096 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Piechotta, Christian T1 - A method for monitoring estrogens in whole surface waters by GC-MS/MS N2 - Natural and synthetic estrogens are key endocrine-disrupting chemicals. Despite occurring at ultra-trace levels (below ng L-1), it is believed that they are contributing to an increase in feminized fish and other endocrine disruptive effects, and hence, their inclusion in the Watch list was not unexpected. One of the main sources ofestrogens to surface waters is wastewater effluent. Once in surface waters, they can partition into different compartments, i.e., water and suspended particulate matter. For this reason, there is an urgent need for a methodology to monitor estrogen levels below the environmental quality standards (EQS) set by the Water Framework Directive requirements. In this study, a precise and accurate gas chromatography-mass spectrometry method (GC-MS/MS) for the analysis of estrone (E1), 17β-estradiol (17β-E2), 17α-estradiol (17α- E2), 17-alpha-ethinylestradiol (EE2), and estriol (E3) in whole water samples with ng L-1 limit of quantification (LOQ) was developed and validated in accordance with CEN/TS 16800:2020 guidelines. T2 - IMSC 2022 CY - Maastricht, Netherlands DA - 27.08.2022 KW - EDC KW - WFD KW - GC-MS PY - 2022 AN - OPUS4-57097 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Portesi, C. T1 - Creation of a European Metrology Network for Safe and Sustainable Food N2 - Safe food is a prerequisite for good human health, however currently almost one in ten people in the world fall ill after eating contaminated food. National food safety authorities need to be able to manage food safety risks along the entire food chain. The Official Controls Regulation (EU) 2017/625 was put in place to ensure that food and feed law stipulates the need for validated analytical methods. However, whilst method development for food control and proficiency testing are well established by European Union Reference Laboratories and National Reference Laboratories, validation of the measurements involved, certified reference materials and internationally recognised calibration and measurement capabilities are currently often (?) lacking. This may erode trust in the accuracy of the measurements. This poster presents Food-MetNet, a coordinated preparatory initiative to establish the European Metrology Network for Safe and Sustainable Food, recently approved by the European Association of National Metrology Institutes (EURAMET). Food-MetNet aims to establish an ongoing dialogue between the metrology community, reference laboratories and regulatory bodies, in order to (i) identify stakeholders’ needs, (ii) develop a sustainable knowledge-sharing programme and web-based platform for stakeholders and (iii) develop roadmaps and a strategic research agenda. Improved access to more reliable and accurate food safety measurements will enable reference laboratories to more confidently and effectively compare their measurement results and support accreditation. This is particularly important where national food safety authorities need to assess potential new contaminants, novel food ingredients and newly emerging food risks, and new measurements methods may be required as well as input into new documentary standards. The activities for sustainable food intend to boost the activity within the "EU green deal framework" to reduce food waste and losses, and the development of standards and methods that allow for the reuse of food waste. The support to new business opportunities and innovation will allow increasing the market competitiveness of EU food industries and the valorisation of waste by reducing environmental impacts. T2 - 10th International Symposium on RECENT ADVANCES IN FOOD ANALYSIS CY - Prague, Czech Republic DA - 06.09.2022 KW - Food KW - Analysis KW - EMN KW - Metrology KW - Network PY - 2022 AN - OPUS4-56133 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Simon, Fabian T1 - HR-CS-GFMAS as a new screening method for emerging pollutants – per- and polyfluoroalkyl substances (PFASs) in the environment N2 - Per- and polyfluorinated alkyl substances (PFASs) are a group of several thousand individual compounds. Many PFASs are extremely persistent, bioaccumulative and toxic. The analysis of PFASs is challenging because of their various chemical and physical properties as well as the high number of compounds. Target-based approaches (e.g., LC-MS/MS) are limited to the availability of analytical grade standards and are not suitable for the analysis of new/unknown PFASs and transformation products. Therefore, PFAS sum parameter methods become increasingly important to indicate realistic PFAS pollution levels. PFAS sum parameters display the proportion of organically bound fluorine that can either be extracted (EOF) or adsorbed to activated carbon (AOF). For the instrumental analysis of such sum parameters, a fluorine selective detector is needed. High resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) is a sensitive and highly selective tool for fluorine determination. The method is based on the in situ formation of diatomic gallium-mono fluoride (GaF) in a graphite furnace at a temperature of 1550°C. The molecular absorption of GaF can be detected at its most sensitive wavelength at 211.248 nm providing limits of quantification of c(F) 2.7 µg/L. In the present work a HR-CS-GFMAS method for EOF analysis was utilized for suspended particulate matter (SPM) analysis in German rivers in time lines from 2005–2020. Therefore, time and spatial resolved trends were investigated. In addition, SPM samples were analyzed for the Σ41PFAS using target analysis based on UHPLC-HRMS. Overall, target analysis drastically underestimated the total PFAS burden in SPM of German rivers compared to the EOF HR-CS-GFMAS analysis. Using a fluorine mass balance approach, only 0.2% to 38.6% of the EOF was explainable using the Σ41PFAS determined with target analysis. Our study highlights the need to integrate PFAS sum parameters (e.g., EOF with HR-CS-GFMAS) in PFAS risk assessment strategies. T2 - Adlershofer Forschungsforum 2022 CY - Berlin, Germany DA - 11.11.2022 KW - PFAS KW - HR-CS-GFMAS KW - Fluorine KW - SPM PY - 2022 AN - OPUS4-56392 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Riedel, Soraya T1 - Tracing toxic fumonisins - Fumonisin sensing by Aspergillus niger Fumonisin Amine Oxidase (AnFAO) and amperometric hydrogen peroxide detection N2 - Fumonisins are a class of toxic secondary metabolites produced by various Fusarium species. The two most important producers of fumonisins are F. verticillioides and F. proliferatum but also Aspergillus niger is known to produce fumonisins. Most frequently they occur on maize, but also other grains can be contaminated with this group of mycotoxins. Exposure to fumonisins by dietary intake can have serious health effects on farm animals such as equine leukoencephalomalacia and porcine pulmonary oedema and is associated with neural tube defects and esophageal cancer in humans. Thus, the European Commission sets legal limits for fumonisins in foodstuffs. The detection of fumonisins is frequently performed in laboratories by chromatographic methods, which are costly and require trained personnel. Simplifying the analysis is therefore a major goal using portable detection systems. Electrochemical enzymatic biosensors offer great promise to meet this demand. Here we report for the first time an enzymatic fumonisin sensing approach with amperometric detection. For this purpose, an Aspergillus niger fumonisin amine oxidase (AnFAO) catalyzing the oxidative deamination of fumonisins, producing hydrogen peroxide, was recombinantly produced in E. coli. For the first time, the specific enzyme activity of AnFAO was determined using a horseradish peroxidase-based fluorescence assay. It was found that the specific activity of AnFAO using 20 μM Fumonisin B1 as substrate is higher than for 20 μM Fumonisin B2 with 0.122 U mg-1 and 0.058 U mg-1, respectively. It was possible to show a dependence of enzyme activity with enzyme – and substrate-concentration. For fumonisin B1 detection, the enzyme was coupled covalently to magnetic particles and the enzymatically produced H2O2 was detected amperometrically in a flow injection system using Prussian blue carbon electrodes. The developed method allows to quantify fumonisin B1 concentrations down to 1.5 µM and demonstrates that the recombinantly produced AnFAO was able to deaminate different concentrations of fumonisin even in immobilized form. Thus, this enzyme is well suited to develop an enzyme based electrochemical biosensor for fumonisin contaminated food and feed. T2 - Posterschau Adlershofer Forschungsforum CY - Berlin, Germany DA - 11.11.2022 KW - Biosensor KW - Amperometry KW - Mycotoxins KW - Electrochemical sensor KW - Recombinant protein expression PY - 2022 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. AN - OPUS4-56248 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Döring, Sarah T1 - Cost-effective method for full-length sequencing of monoclonal antibodies N2 - With the development of hybridoma technology, monoclonal antibodies are increasingly important in therapeutic and analytical applications. For recombinant expression in mammalian cells, knowledge of the variable regions of both, heavy (VH) and light (VL) chain, is a necessary prerequisite for generating expression plasmids. Furthermore, cloning antibody sequences including constant regions CH1 to CH3 and CL reduces impact of hybridoma cell loss and associated full natural antibody sequence leakage. Here, we show a cost-effective workflow for amplification of IgG antibody variable regions in combination with advanced methods for full-length cloning of monoclonal antibodies. T2 - Adlershofer Forschungsforum CY - Berlin, Germany DA - 11.11.2022 KW - Monoclonal Antibodies KW - DNA-Sequencing KW - PCR PY - 2022 AN - OPUS4-56275 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Munsch, Sarah Mandy T1 - Ermittlung der Oberflächenrelaxivität in Sandsteinen - ein Methodenvergleich N2 - Mit dem zerstörungsfreien Verfahren der nuklear magnetischen Resonanz (NMR) mit Fokus auf 1H-Protonen lässt sich der Feuchtegehalt in porösen Materialien bestimmen und bei einem vollgesättigten Medium eine Porengrößenverteilung ableiten. Für die Ermittlung der Porengrößenverteilung muss die gemessene T2-Relaxationszeitenverteilung mit dem Parameter Oberflächenrelaxivität (OR) umgerechnet werden. Ist dieser Parameter für das zu untersuchende Medium unbekannt, kann dieser mithilfe von Vergleichsverfahren abgeschätzt werden. Dabei kommen z. B. Gassorption, Quecksilberporosimetrie (MIP), Dünnschliff-Mikroskopie oder µ-Computer-Tomographie (µ-CT) zum Einsatz. Da die verschiedenen Verfahren allerdings unterschiedliche Auflösungsgrenzen aufweisen und auf verschiedenen physikalischen Prinzipien beruhen, können die ermittelten ORs für ein Medium stark variieren. Um ein besseres Verständnis für die OR und die Einflüsse der Vergleichsverfahren zu bekommen, wurden 19 verschiedene Sandsteintypen untersucht. Die OR wurde zum einen basierend auf der spezifischen inneren Oberfläche aus der dynamischen Dampfsorption (DVS) und der Gassorption abgeleitet. Zum anderen wurden die Relaxationszeitenverteilungen durch manuelles Anpassen der OR derart in Porengrößenverteilungen umgerechnet, sodass es eine Übereinstimmung mit den Porengrößenverteilungen aus MIP und/oder µ-CT gibt. Die Umrechnung von Relaxationszeitverteilungen in eine Porengrößenverteilung berücksichtigt dabei sowohl fast- als auch slow-diffusion. Insgesamt ergab sich für die Sandsteine ein breites Spektrum an ORs, bei dem die über die innere spezifische Oberfläche ermittelten ORs andere Größenordnungen aufwiesen (< 10 µm/s), als bei dem Vergleich mit MIP oder µ-CT (bis 600 µm/s). Die mit MIP und µ-CT ermittelten Werten lagen dabei um einen Faktor von 1 bis 30 auseinander. Dieser umfangreiche Datensatz, der vier verschiedene Vergleichsverfahren einbezieht, soll neue Erkenntnisse in der Charakterisierung von porösen Materialien (z. B. bezüglich der Oberflächenrauigkeit) und in der Ursache für die starke Variation der OR ermöglichen. Durch das Berücksichtigen von fast- als auch slow-diffusion, können dabei sogar Daten in eine Porengrößenverteilung umgerechnet werden, bei denen das fast-diffusion-Kriterium nach Brownstein und Tarr nicht erfüllt ist. Außerdem ergibt sich aus der Berücksichtigung des slow-diffusion-Regimes ein neues Verständnis für die obere Auflösungsgrenze von NMR. T2 - 82. Jahrestagung der Deutschen Geophysikalischen Gesellschaft CY - Online meeting DA - 07.03.2022 KW - NMR KW - Sandsteine KW - Oberflächenrelaxivität KW - Porengrößenverteilung PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-544721 UR - https://e-docs.geo-leo.de/handle/11858/9672 DO - https://doi.org/10.23689/fidgeo-5330 AN - OPUS4-54472 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Strangfeld, Christoph T1 - Bore hole probe for ultrasound-based quality control of sealing constructions N2 - Several countries, which use or used nuclear power, started to destruct nuclear power plants. Thus, large amounts of radioactive waste have to be stored safely during the next decades. Large underground caverns will be used a nuclear waste disposal. Once the cavern is full, a sealing construction is required. The sealing should avoid the contact to any fluids or substances of the storage barrels and reduce hazardous radiation. T2 - International Symposium Non-Destructive Testing in Civil Engineering CY - Zurich, Switzerland DA - 16.08.2022 KW - Ultrasonic testing KW - Nuclear waste PY - 2022 AN - OPUS4-55530 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Strangfeld, Christoph T1 - Frequency modulated, air-coupled ultrasound generated by fluidic oscillators N2 - The majority of ultrasonic devices used for non-destructive testing in civil engineering require contact with the surface of the concrete (specimen), which significantly increases the time required for the measurement. This makes it impractical for extensive investigation of large-scale structures such as bridge decks, foundations, or tunnels. In a pioneering approach, fluidic oscillators are used as contact free ultrasonic sources to overcome the aforementioned limitations. These robust and cost-effective actuators require only pressurised air and are ideally suited for harsh environments. At a constant supply pressure, they generate a continuous monofrequent actuation signal. Further, varying the supply pressure via a fast pressure regulator was found to generate a frequency modulated signal which enabled time-of-flight measurement with an added advantage of increased signal to noise ratio. To demonstrate the feasibility of this novel idea of non-contact ultrasound, the results of the initial tests are presented. T2 - IEEE International Ultrasound Symposium CY - Venice, Italy DA - 11.10.2022 KW - Air coupled ultrasound KW - Frequency modulation KW - Non-destructive testing KW - Civil engineering KW - Building materials KW - Fluidic oscillators PY - 2022 AN - OPUS4-56069 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ecke, Alexander T1 - An Immunosensor for Pharmaceutical Contaminants in Water Supply N2 - Contamination of drinking water with pharmaceuticals is an emerging problem and a potential health threat. Immunochemical methods based on the binding of the analyte to specific antibodies enable fast & cost-effective on-site analyses. Magnetic bead-based immunoassays (MBBAs) allow for implementation into an immunosensor for online testing. Particles are prepared by either coupling the analyte molecule (diclofenac) or the respective antibody (amoxicillin) to the surface. For miniaturization, detection is performed electrochemically (chronoamperometry) on a microfluidic chip. The developed immunosensor will enable detection of pharmaceuticals directly in water supply pipes. T2 - Analytica Conference CY - München, Germany DA - 21.06.2022 KW - Immunosensor KW - Magnetic Particles KW - Microfluidics KW - Amperometry KW - Diclofenac KW - Amoxicillin PY - 2022 AN - OPUS4-55381 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von Rüden, H. T1 - Röntgen-Computertomographie (CT) für die Kulturguterhaltung N2 - Als bildgebendes und zerstörungsfreies Verfahren eignet sich die Röntgen-Computertomographie (CT) insbesondere bei der Untersuchung von Kunst- und Kulturgütern. Die für Werkstoffprüfung und Materialanalyse konzipierten CT-Anlagen bieten dank hoher Röntgenleistung die Möglichkeit zusätzlich zu Holz, Keramiken und Kunststoffen auch stark schwächende Materialien, wie bspw. Metalle, zu durchdringen. Für eine hohe räumliche Auflösung im unteren Mikrometerbereich (2-200µm) sorgen entsprechend ausgelegte CT-Anlagen. In der Bundesanstalt für Materialforschung und -prüfung (BAM) stehen mehrere solcher CT-Anlagen für unterschiedliche Fragestellungen sowie Probengrößen und - materialien zur Verfügung. Einige Beispiele aus den vergangenen Arbeiten der BAM veranschaulichen das große Potential dieser Untersuchungsmethode. Da bei einer CT-Messung in der Regel das gesamte Untersuchungsobjekt erfasst und in ein digitales Volumenmodell überführt wird, eröffnet sich für Archäologen und Restoratoren die Möglichkeit Untersuchungen hinsichtlich Materialzusammensetzung, Erhaltungszustand und Herstellungstechnik am virtuellen Objekt vorzunehmen, ohne die Originalsubstanz zu beeinträchtigen. Gegenüber der klassischen Radiografie, bei der nur eine zweidimensionale Abbildung erreicht wird, bietet die CT-Untersuchung die Möglichkeit innenliegende Strukturen dreidimensional zu erfassen. Die notwendige Bestimmung der Objektoberfläche erlaubt zudem die Erstellung eines Oberflächenmodells des untersuchten Gegenstandes. Damit lassen sich unter anderem mechanische Simulationen (Belastung, Durchbiegung, Durchströmung) durchführen. Der erzeugte Datensatz kann somit auch zur Herstellung eines Replikats im 3D-Druckverfahren genutzt werden. Anhand der Untersuchung der Mandoline von Smorsone wird gezeigt, wie eine CT-Messung durchgeführt wird und wie in dem anschließend rekonstruierten 3D-Datensatz mittels Koordinatenmesstechnik Maße (Abstände, Wandstärken, Winkel) exakt ermittelt werden können. T2 - Jahrestagung der Gesellschaft für Musikforschung CY - Berlin, Germany DA - 28.09.2022 KW - Mikro-CT KW - Bildverarbeitung KW - Holz KW - Bronze KW - ZfP PY - 2022 AN - OPUS4-56412 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Meyer, Klas T1 - Compact NMR Spectroscopy in the field: A Versatile Tool for Automated Continuous-Flow Production N2 - Chemical companies must find new paths to successfully survive in a changing environment. The potential of digital technologies belongs to these. Flexible and modular chemical plants can produce various high-quality products using multi-purpose equipment with short downtimes between campaigns and reduce time to market for new products. Intensified continuous production plants allow for difficult to produce compounds. Therefore, fully automated “chemical” process control along with real-time quality control are prerequisites to such concepts and thus should be based on “chemical” information. A commercially available benchtop NMR spectrometer was integrated to the full requirements of an automated chemical production environment such as, e.g., explosion safety, field communication, and robust evaluation of sensor data. It was thereof used for direct loop advanced process control and real-time optimization of the process. Field studies in modular and conventional production plant setups show promising results gaining process knowledge for further optimization. NMR appeared as preeminent online analytical method and allow using a modular data analysis tool, which even served as reliable reference method for further PAT applications (e.g. NIR spectroscopy). In the future, such fully integrated and intelligently interconnecting “smart” systems and processes can speed up the high-quality production of specialty chemicals and pharmaceuticals. T2 - DECHEMA Workshop "Sensorik für die Digitalisierung chemischer Produktionsanlagen" CY - Frankfurt am Main, Germany DA - 13.06.2022 KW - Compact NMR KW - Process Control KW - Modular Production KW - Process Analytical Technology PY - 2022 AN - OPUS4-55037 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Recknagel, Sebastian T1 - Certified reference material for the determination of hydrolytic resistance of borosilicate glass N2 - The preparation of certified reference material for the determination of hydrolytic resistance of borosilicate glass is described. Certified properties are the acid consumption according to the relevant standard procedures described in ISO 720, USP<660> and Ph.Eur. 3.2.1. T2 - 26. International Congress on Glass CY - Berlin, Germany DA - 03.07.2022 KW - Borosilicate glass KW - CRM KW - Hydrolytic resistance PY - 2022 AN - OPUS4-55521 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Meyer, Klas T1 - Compact NMR Spectroscopy in the field: A Versatile PAT Tool for Production of Specialty Chemicals N2 - Companies of chemical industry find themselves more often in a rapidly changing environment, e.g., due to variability of raw material quality or energy costs and efficiency. Process optimization and new process concepts become more and more important. For example, flexible and modular chemical plants can produce various high-quality products using multi-purpose equipment with short downtimes between campaigns and reduce time to market for new products. Intensified continuous production plants allow for difficult to produce compounds like exothermic reactions with high heat dissipation. Therefore, fully automated “chemical” process control along with real-time quality control are prerequisites to such concepts and thus should be based on “chemical” information. A commercially available benchtop NMR spectrometer was integrated to the full requirements of an automated chemical production environment such as explosion safety, field communication, and robust evaluation of sensor data. Field studies in modular and conventional production plant setups show promising results gaining process knowledge for further optimization. NMR appeared as preeminent online analytical method and allow using a modular data analysis approach, which can even serve as reliable reference method for further calibration-dependent PAT applications (e.g., NIR spectroscopy). Based on experiences from these field studies an improved analyzer enclosure setup was developed and built, including the option of a secondary method besides NMR spectroscopy. Integrated control systems allow for a flexible implementation based on the available automation infrastructure at the chemical plant or pilot plant setup. In the future, fully integrated and intelligently interconnecting “smart” PAT systems and processes have the potential speed up the setup of production equipment for chemicals and pharmaceuticals and therefore help to reduce the time-to-market. T2 - Practical Applications of NMR in Industry Conference (PANIC) 2022 CY - La Jolla, CA, USA DA - 16.10.2022 KW - Process Analytical Technology KW - Benchtop NMR KW - Modular production PY - 2022 AN - OPUS4-56090 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -