TY - JOUR A1 - Bell, Jérémy A1 - Climent Terol, Estela A1 - Hecht, Mandy A1 - Buurman, Merwe A1 - Rurack, Knut T1 - Combining a droplet-based microfluidic tubing system with gated indicator releasing nanoparticles for mercury trace detection JF - ACS Sensors N2 - A droplet-based microfluidic sensor was developed for the detection of Hg2+ traces in water. The approach uses gated mesoporous nanoparticles loaded with a fluorescent BODIPY dye. The squaraine-based gating mechanism is highly selective for Hg2+ and the indicator release mechanism ensures sensitive detection. The microfluidic system is modular and was assembled from simple PTFE/PFA tubes, while detection was realized with standard optomechanic, optic, and electronic parts. The sensor shows a stable response without memory effects and allows the detection of Hg2+ in water down to 20 ppt. KW - Microfluidic sensor KW - Gated delivery system KW - Fluorescence KW - Mercury KW - Hybrid nanoparticles PY - 2016 DO - https://doi.org/10.1021/acssensors.5b00303 SN - 2379-3694 VL - 1 IS - 4 SP - 334 EP - 338 PB - American Chemical Society CY - Washington, DC AN - OPUS4-35831 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -