TY - JOUR A1 - Phukphatthanachai, Pranee A1 - Vogl, Jochen A1 - Traub, Heike A1 - Jakubowski, Norbert A1 - Panne, Ulrich T1 - A new approach of using polyethylene frits for the quantification of sulphur in copper metals by isotope dilution LA-ICP-MS and comparison with conventional IDMS techniques N2 - Polyethylene (PE) frits were used to quantify sulphur in copper and its alloys by isotope dilution combined with LA-ICP-MS as an alternative approach to conventional sample preparation: the copper samples were spiked, the spiked samples were dissolved, the resulting solutions were absorbed in the PE frits and finally the PE frits were analysed by LA-ICP-MS. A prerequisite for such a support material is a low sulphur blank and thus PE was selected for this purpose. The absorption efficiency of the PE frits was studied for varying sulphur amounts ranging from 2 mg S to 80 mg S showing that more than 99.5% of the loaded sulphur was absorbed by the frit. The so prepared PE frits were measured by LA-ICP-MS and yielded a good linearity (R2 ¼ 0.999) for the sulphur ion intensities corresponding to sulphur amounts up to 40 mg S; the associated sensitivity is approximately 3.4 x 10⁴ cps μg⁻¹ for ³²S. For the validation of the developed procedure the reference materials BAM-M376a, BAM-228 and BAM-227 were applied such that 2 μg S, 5 μg S and 11 μg S were absorbed in the PE frits, respectively. These samples were pre-quantified for the adsorbed sulphur amount by external calibration LA-ICP-MS yielding sulphur amounts of 0.9 μg, 5.1 μg and 8.5 μg (quantified for ³²S only), respectively. Relative Standard deviations of the isotope ratios were below 5% in average (n ¼ 3 lines) in all cases (except for the pure spike solution). These samples were then analysed by LA-ICP-IDMS and the measurement results were validated by comparing them with the results obtained by conventional ICP-IDMS. The obtained relative expanded measurement uncertainties ranged between 10% and 26%. Pearson's coefficient was used to express the correlation between both techniques; the obtained value was 0.999 demonstrating a strong correlation. Contrary to most published LA-ICP-IDMS procedures, the developed procedure enables SI-traceability for the measurement results. The metrological traceability to the SI for the sulphur mass fractions in copper was established by an unbroken chain of comparisons, each accompanied by an uncertainty budget. Thus, the measurement results are considered reliable, acceptable and comparable within the stated measurement uncertainty. The metrological traceability chain from the kg down to mass fraction in the samples obtained by LA-ICP-IDMS is presented as well. KW - Laser ablation KW - IDMS KW - Traceability KW - Uncertainty KW - Purity PY - 2018 DO - https://doi.org/10.1039/c8ja00116b SN - 0267-9477 SN - 1364-5544 VL - 33 IS - 9 SP - 1506 EP - 1517 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-45899 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Xiao, J. A1 - Vogl, Jochen A1 - Rosner, M. A1 - Deng, L. A1 - Jin, Z. T1 - A validated analytical procedure for boron isotope analysis in plants by MC-ICP-MS N2 - Boron (B) is an essential micronutrient for plant growth. Lack of valid methods for pretreatment and measurement of δ11B in plant restrict applications of it in the biosphere. Dry ashing, one step cation exchange and micro-sublimation were combined to separate and purify boron (B) in plant tissues. The low procedure blank, high B recovery and the accurate δ11B values of the plant reference materials demonstrate that this method is suitable and valid for B pretreatment and δ11B measurement in plant samples by MC-ICP-MS. Based on this method, the δ11B in different plants (Brassica napus, Chenopodium album L, moss, lichen, and Nostoc commune) was analyzed. For Brassica napus, δ11B increased gradually from root to leaf, and then decreased to rapeseed. For the same parts, the δ11B increased from the lower parts to the higher parts. This variation may be due to the B(OH)3 transporter of NIP6;1 and the incorporation of B into the cell. The reason for lower δ11B values in shell and rapeseed compared to those in leaves presumably is to the preferred Transport of borate in the phloem. The largest δ11B fractionation between leaf and root in Brassica napus and Chenopodium album L was +24.2‰ and +26.6‰, respectively. The large variation and fractionation of δ11B within plants indicates that δ11B is a good tracer to study the B translocation mechanisms and metabolism within plants. The δ11B in Nostoc commune, lichen, and moss showed variations of -4.1‰ to +21.5‰, −9.4‰ to +7.3‰, and −18.3‰ to +11. 9‰, respectively. In the same site, δ11B in different plants ranked Nostoc commune>moss>lichen and δ11B in mosses growing in different environment ranked soil>tree>rock. Rain and soil available B are the main B sources for these plants. The δ11B in Nostoc commune, lichen, and moss may be a useful tracer to study the atmospheric B input. In the future, plants culture experiments under certain environments and studies from molecular level are necessary to decipher the variation of δ11B and fractionation mechanisms within plants. KW - Boron isotope KW - Isotopic fractionation KW - Micro-sublimation KW - Plant tissue KW - MC-ICP-MS PY - 2019 DO - https://doi.org/10.1016/j.talanta.2018.12.087 SN - 0039-9140 SN - 1873-3573 VL - 196 SP - 389 EP - 394 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-47160 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schannor, Mathias A1 - Oelze, Marcus A1 - Traub, Heike A1 - He, Yubei A1 - Schmidt, Robin A1 - Heidemann, Luisa A1 - Savic, Lynn Jeanette A1 - Vogl, Jochen A1 - Meermann, Björn T1 - Advancing Biomarker Research: In Situ Cu Isotope Analysis in Liver Tumors by LA-MC-ICP-MS N2 - Stable metal isotopes have received increasing attention as medical biomarkers due to their potential to detect changes in metal metabolism related to diseases. In particular, copper stable isotopes are a powerful tool to identify isotopic variation between tumors and healthy tissue, suggesting application in cancer diagnosis. However, potential mechanisms causing isotope fractionation, such as redox- or bond-forming reactions and interactions of metals during transmembrane import and export, are less well understood. Here, we established an in situ method using laser ablation-multicollector-inductively coupled plasma-mass spectrometry (LA-MC-ICP-MS) to advance our understanding of the underlying processes responsible for isotope fractionation between normal and diseased tissues. Gelatin-based bracketing standards and quality control reference materials, crucial for laser ablation analysis, were developed to allow correction for instrumentally induced isotope fractionation during LA-MC-ICP-MS analysis. Using such matrix-matched standards, the method achieved intermediate precisions for delta values of better than 0.15 ‰ (2 s) for inorganic reference materials and of better than 0.17 ‰ (2 s) for biological reference materials. The developed routine was tested on rabbit VX2 liver tumor samples, a model system resembling human hepatocellular carcinoma (HCC) used to study liver cancer. In situ Cu isotope compositions between healthy (𝛿65/63NIST976(Cu) = −1.5 ‰ to 0.2 ‰) and tumorous (𝛿65/63NIST976(Cu) = 0.0 ‰ to 1.3 ‰) liver tissue show distinct differences in their isotope ratios. The observed isotopic dichotomy is consistent with previous solution-based MC-ICP-MS work, showing enrichment of heavy 65Cu in cancer biopsies relative to healthy tissue. KW - Biomarker KW - Laser Ablation KW - Reference Material PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-626865 DO - https://doi.org/10.1021/acs.analchem.4c05626 SN - 1520-6882 VL - 97 IS - 8 SP - 4425 EP - 4432 PB - American Chemical Society (ACS) AN - OPUS4-62686 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Flierl, Lukas A1 - Rienitz, Olaf A1 - Vogl, Jochen A1 - Pramann, Axel T1 - An advancement of the gravimetric isotope mixture method rendering the knowledge of the spike purity superfluous N2 - The gravimetric isotope mixture method is the primary method to determine absolute isotope ratios. This method, however, depends on the existence of suitable spike materials and knowledge of their purities. Determining the purity of the spikes can be tedious and labour-intensive. In this publication, an advancement of the gravimetric isotope mixture method, rendering the determination of the purity of the spike materials unnecessary, is presented. The advancement combines mass spectrometry and ion chromatography leading to an approach being independent of the purity of the spike materials. In the manuscript the mathematical background and the basic idea of the novel approach are described using a two-isotope system like copper or lithium. KW - Isotope amount ratios KW - Metrology KW - Mass spectrometry KW - Ion chromatography PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-608726 DO - https://doi.org/10.1007/s00216-024-05465-9 SP - 5325 EP - 5333 PB - Springer Science and Business Media LLC AN - OPUS4-60872 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dariz, P. A1 - Wortmann, U. G. A1 - Vogl, Jochen A1 - Schmid, Thomas T1 - Beautiful Pietàs in South Tyrol (Northern Italy): local or imported works of art? N2 - The study, dedicated to Beautiful Pietàs conserved in South Tyrol (Northern Italy), aims to establish, for the first time, a connection between Austroalpine raw materials and the high-fired gypsum mortars constituting the Gothic figure groups in question. The origin and chronology of this stylistically and qualitatively differing ensemble have been subject of art historical debate for nearly a century. The discourse is dominated by three main hypotheses: itinerary of an Austrian artist versus itinerary of the work of art created in an artist’s workshop in Austria versus itinerary of the stylistic vocabulary via graphical or three-dimensional models. The comparison of the δ34S values and the 87Sr/86Sr ratios of the gypsum mortars and Austroalpine sulphate deposits (in a compilation of own reference samples and literature data) points to the exploitation of sediments in the Salzkammergut and possibly also in the evaporite district of the Eastern Calcareous Alps, thus evidencing the import of the sculptures and not the activities of local South Tyrolean or itinerant artists. Two geochronological units are distinguishable: The Pietà in the Church St. Martin in Göflan can be assigned to Upper Permian raw material, whereas the metrologically consistent sculptures in the Church of Our Lady of the Benedictine Abbey Marienberg and in the Chapel St. Ann in Mölten correlate with deposits of the Early Triassic (or the Lower-Middle Triassic transition). The medieval gypsum mortars also differ in their mineralogical characteristics, i.e. in their geologically related minor components, as in the first case, characterised by a significant proportion of primary anhydrite, natural carbonate impurities mainly consist of calcite (partly converted to lime-lump-like aggregates), whereas in the second group dolomite (or rather its hydration products after pyrometamorphic decomposition) predominates, accompanied by celestine, quartz and potassium feldspar. The Pietà in the Cathedral Maria Himmelfahrt in Bozen turned out to be made of Breitenbrunn calcareous sandstone (Leitha Mountains, Burgenland, Austria), which is why the sample is not considered in the geochemical analysis. KW - High-fired gypsum mortar KW - Sulphur isotope KW - Strontium isotope KW - Polarised light microscopy KW - Raman microspectroscopy PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-545610 DO - https://doi.org/10.1186/s40494-022-00678-6 SN - 2050-7445 VL - 10 IS - 1 SP - 1 EP - 17 PB - Springer Nature AN - OPUS4-54561 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Xiao, J. A1 - Vogl, Jochen A1 - Rosner, M. A1 - Jin, Z. T1 - Boron isotope fractionation in soil-plant systems and its influence on biogeochemical cycling N2 - Boron (B) is an essential mineral nutrient for higher plants. Although B plant nutrition is well studied, the B isotope fractionation at the soil-plant interface, within plant metabolism, and its influence on biogeochemical cycling is not fully understood. Boron concentrations and isotope variations (δ11B) of the dicotyledonous plants of Chenopodium album and Brassica napus and their growing soils along a climatic gradient were analyzed to decipher these unresolved issues of the B behavior. The boron concentrations and δ11B values show an increasing trend from roots to leaves for both plants, while a decreasing trend from flower to shell and to seed for Brassica napus. A large boron isotope fractionation occurs within the plants with median Δ11Bleaf-root ≈ +20‰, which is related to different boron transporters and transportation ways. Formation of borate dimerized rhamnogalacturonan II in cell and B(OH)3 transportation in xylem lead to heavier δ11B values from root to stem and leaf while B(OH)4􀀀 transportation in phloem lead to lighter δ11B values from flower to shell and seed. Although samples cover a distinct transect with systematically different climatic conditions, Δδ11B within the individual plant compartments and between the bulk plants and the soil available B do not show any systematic variation. This suggests that B uptake from the soil into Chenopodium album and Brassica napus occurs without a distinct isotope fractionation at the soil-plant interface (median Δ11Bbulkplant-soil = 􀀀 0.2‰) and plants are able to regulate boron uptake. Both the observed large B fractionation within plant and low or absent B isotope fractionation at the soil-plant interface may have profound implications for the biological and geological B cycle. If this observed boron behavior also exists in other plants, their litters would be an important source for exporting 11B-rich biological material from continental ecosystems via rivers to the global oceans. This may be helpful for the explanation of ocean B cycle and the increasing δ11B values over the Cenozoic. KW - Boron isotopic composition KW - Boron isotope fractionation KW - Soil available boron KW - Biological boron recycling KW - Chenopodium album KW - Brassica napus PY - 2022 DO - https://doi.org/10.1016/j.chemgeo.2022.120972 SN - 0009-2541 VL - 606 SP - 1 EP - 8 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-55031 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Geilert, Sonja A1 - Vogl, Jochen A1 - Rosner, M. A1 - Eichert, T. T1 - Boron isotope variability related to boron speciation (change during uptake and transport) in bell pepper plants and SI traceable n(11B)/n(10B) ratios for plant reference materials N2 - Rationale: Boron (B) is an essential micronutrient in plants and its isotope variations are used to gain insights into plant metabolism, which is important for crop plant cultivation. B isotope variations were used to trace intra‐plant fractionation mechanisms in response to the B concentration in the irrigation water spanning the range from B depletion to toxic levels. Methods: A fully validated analytical procedure based on multi‐collector inductively coupled plasma mass spectrometry (MC‐ICP‐MS), sample decomposition and B Matrix separation was applied to study B isotope fractionation. The Validation was accomplished by establishing a complete uncertainty budget and by applying reference materials, yielding expanded measurement uncertainties of 0.8‰ for pure boric acid solutions and ≤1.5‰ for processed samples. With this validated procedure SI traceable B isotope amount ratios were determined in plant reference materials for the first time. Results: The B isotope compositions of Irrigation water and bell pepper samples suggest passive diffusion of the heavy 11B isotope into the roots during low to high B concentrations while uptake of the light 10B isotope was promoted during B depletion, probably by active processes. A systematic enrichment of the heavy 11B isotope in higher located plant parts was observed (average Δ11Bleaf‐roots = 20.3 ± 2.8‰ (1 SD)), possibly by a facilitated transport of the heavy 11B isotope to growing Meristems by B transporters. Conclusions: The B isotopes can be used to identify plant metabolism in Response to the B concentration in the irrigation water and during intra‐plant B transfer. The large B isotope fractionation within the plants demonstrates the importance of biological B cycling for the global B cycle. KW - isotope fractionation KW - boron KW - delta value KW - metabolism KW - bell pepper KW - SI traceability KW - measurement uncertainty PY - 2019 DO - https://doi.org/10.1002/rcm.8455 SN - 1097-0231 SN - 0951-4198 VL - 33 IS - 13 SP - 1137 EP - 1147 PB - John Wiley & Sons Ltd. AN - OPUS4-48213 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rienitz, O. A1 - Jährling, R. A1 - Noordmann, J. A1 - Pape, C. A1 - Röhker, K. A1 - Vogl, Jochen A1 - Manzano, J. V. L. A1 - Kozlowski, W. A1 - Caciano de Siena, R. A1 - Marques Rodrigues, J. A1 - Galli, A. H. A1 - Yim, Y.-H. A1 - Lee, K.-S. A1 - Lee, J. H. A1 - Min, H.-S. A1 - Chingbo, C. A1 - Naijie, S. A1 - Qian, W. A1 - Ren, T. A1 - Jun, W. A1 - Tangpaisarnkul, N. A1 - Suzuki, T. A1 - Nonose, N. A1 - Mester, Z. A1 - Yang, L. A1 - Pagliano, E. A1 - Greenberg, P. A1 - Mariassy, M. A1 - Näykki, T. A1 - Cankur, O. A1 - Coskun, F. G. A1 - Ari, B. A1 - Can, S. Z. T1 - CCQM-K122 "Anionic impurities and lead in salt solutions" N2 - The determination of the mass fractions of bromide, sulfate, and lead as well as the isotopic composition of the lead (expressed as the molar mass and the amount fractions of all four stable lead isotopes) in an aqueous solution of sodium chloride with a mass fraction of 0.15 g/g was the subject of this comparison. Even though the mass fractions ranged from 3 μg/g (bromide) to 50 ng/g (lead), almost all results reported agreed with the according KCRVs. KW - Absolute isotope ratio KW - Lead isotope ratios KW - Metrology KW - Traceability PY - 2020 DO - https://doi.org/10.1088/0026-1394/57/1A/08012 VL - 57 IS - 1A SP - 8012 PB - IOP Science AN - OPUS4-51156 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Molloy, J. L. A1 - Winchester, M. R. A1 - Butler, T. A. A1 - Possolo, A. M. A1 - Rienitz, O. A1 - Roethke, A. A1 - Goerlitz, V. A1 - Caciano de Sena, R. A1 - Dominguez Almeida, M. A1 - Yang, L. A1 - Methven, B. A1 - Nadeau, K. A1 - Romero Arancibia, P. A1 - Bing, W. A1 - Tao, Z. A1 - Snell, J. A1 - Vogl, Jochen A1 - Koenig, Maren A1 - Kotnala, R. K. A1 - Swarupa Tripathy, S. A1 - Elishian, C. A1 - Ketrin, R. A1 - Suzuki, T. A1 - Oduor Okumu, T. A1 - Yim, Y.-H. A1 - Heo, S. W. A1 - Min, H. S. A1 - Sub Han, M. A1 - Lim, Y. A1 - Velina Lara Manzano, J. A1 - Segoviano Regalado, F. A1 - Arvizu Torres, M. A1 - Valle Moya, E. A1 - Buzoianu, M. A1 - Sobina, A. A1 - Zyskin, V. A1 - Sobina, E. A1 - Migal, P. A1 - Linsky, M. A1 - Can, S. Z. A1 - Ari, B. A1 - Goenaga Infante, H. T1 - CCQM-K143 Comparison of Copper Calibration Solutions Prepared by NMIs/DIs N2 - CCQM-K143 is a key comparison that assesses participants’ ability to prepare single element calibration solutions. Preparing calibration solutions properly is the cornerstone of establishing a traceability link to the International System of Units (SI), and therefore should be tested in order to confirm the validity of CCQM comparisons of more complex materials. CCQM-K143 consisted of participants each preparing a single copper calibration solution at 10 g/kg copper mass fraction and shipping 10 bottled aliquots of that solution to the coordinating laboratory, the National Institute of Standards and Technology (NIST). The masses and mass fraction for the prepared solutions were documented with the submitted samples. The solutions prepared by all participants were measured at NIST by high performance inductively coupled plasma optical emission spectroscopy (HP-ICP-OES). The intensity measurements for copper were not mapped onto values of mass fraction via calibration. Instead, ratios were computed between the measurements for copper and simultaneous measurements for manganese, the internal standard, and all subsequent data reductions, including the computation of the KCRV and the degrees of equivalence, were based on these ratios. Other than for two participants whose measurement results appeared to suffer from calculation or preparation errors, all unilateral degrees of equivalence showed that the measured values did not differ significantly from the KCRV. These results were confirmed by a second set of ICP-OES measurements performed by the Physikalisch-Technische Bundesanstalt (PTB). CCQM-K143 showed that participants are capable of preparing calibration solutions starting from high purity, assayed copper metal. Similar steps are involved when preparing solutions for other elements, so it seems safe to infer that similar capabilities should prevail when preparing many different, single-element solutions. KW - Metrology KW - Primary calibration solution KW - Traceability PY - 2020 DO - https://doi.org/10.1088/0026-1394/58/1A/08006 SN - 0026-1394 VL - 58 IS - 1A SP - 08006 PB - IOP Science CY - Cambridge AN - OPUS4-51983 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hill, Sarah A1 - Infante, Heidi Goenaga A1 - Entwisle, John A1 - Strekopytov, Stanislav A1 - Ward-Deitrich, Christian A1 - Cowen, Simon A1 - Rienitz, Olaf A1 - Roethke, Anita A1 - Goerlitz, Volker A1 - Schulz, Ursula A1 - Pape, Carola A1 - Vogl, Jochen A1 - Koenig, Maren A1 - Jacimovic, Radojko A1 - Fisicaro, Paola A1 - Ren, Tongxiang A1 - Wang, Song A1 - Song, Panshu A1 - Li, Haifeng A1 - Linsky, Maré A1 - Sobina, Egor A1 - Lozano, Hernán Ezequiel A1 - Puelles, Mabel A1 - Yamani, Randa A1 - Haraldsson, Conny T1 - CCQM-K160: platinum group elements in automotive catalyst N2 - The platinum group elements (PGEs) play an important role in reducing emissions from automotive vehicles through their use in catalytic convertors but also for catalysis in the pharmaceutical industry. The immense economic value of platinum (Pt), palladium (Pd) and rhodium (Rh) highlights the importance of highly accurate measurements. Therefore, there is a need for National Metrology Institutes (NMIs) and Designated Institutes (DIs) to demonstrate measurement capability in this space. A pilot comparison (CCQM-P63) for precious metals in automotive catalyst took place in 2006, but with a limited number of institutes participating. Furthermore, this study was performed over 17 years ago. Therefore, there was a need to maintain existing capability and demonstrate new capability in a key comparison, in order to claim calibration and measurement capability claims (CMCs). With the core capability matrix, this study falls into the "Difficult to dissolve metals/metal oxides" which will support CMC categories 8 (Metal and metal alloys), 9 (Advanced materials) and 14 (Other materials). Eleven NMIs and DIs participated in the Key Comparison CCQM-K160 Platinum Group Elements in Automotive Catalyst. Participants were requested to evaluate the mass fractions of Pt, Pd and Rh in mg/kg in an unused autocatalyst material (cordierite ceramic base). The Key Comparison Reference Values (KCRVs) and Degrees of Equivalence (DoEs) were calculated utilising the NIST Decision Tree for the measurands. The participants utilised a number of sample preparation and analytical methods including hot plate digestion, microwave digestion and sodium fusion, followed by either atomic absorption spectroscopy (AAS), inductively coupled plasma optical emission spectroscopy (ICP-OES) or inductively coupled plasma mass spectrometry (ICP-MS) detection. Several calibration techniques were used, namely external calibration, standard addition, isotope dilution mass spectrometry (IDMS) and an exact matching procedure. Additionally, one participant employed instrumental neutron activation analysis (INAA) with k0 standardisation which is a direct solid analysis method. The majority of participants claimed traceability to NIST primary calibrants or their own CRMs. Furthermore, several matrix CRMs were included or spiked samples for quality control. All institutes were required to determine the dry mass fraction using the stipulated protocol. The NIST decision tree was implemented for the calculation of the KCRVs and DoEs. The participant results overall showed good agreement with the KCRV, despite the variety of dissolution procedures and measurement techniques for this highly complex matrix and challenging measurands. Successful participation in CCQM-K160 demonstrated measurement capabilities for the determination of mass fraction of Pt, Pd and Rh in the mg/kg range and will support broad scope CMC claims for a wide range of challenging matrices. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). KW - Metrology in Chemistry KW - Traceability KW - Uncertainty PY - 2024 DO - https://doi.org/10.1088/0026-1394/61/1A/08011 VL - 61 IS - 1A SP - 1 EP - 39 PB - IOP Publishing AN - OPUS4-60524 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -