TY - JOUR A1 - Hendriks, L. A1 - Brunjes, R. A1 - Taskula, S. A1 - Kocic, J. A1 - Hattendorf, B. A1 - Bland, G. A1 - Lowry, G. A1 - Bolea-Fernandez, E. A1 - Vanhaecke, F. A1 - Wang, J. A1 - Baalousha, M. A1 - von der Au, Marcus A1 - Meermann, Björn A1 - Holbrook, T. A1 - Wagner, S. A1 - Harycki, S. A1 - Gundlach-Graham, A. A1 - von der Kammer, F. T1 - Results of an interlaboratory comparison for characterization of Pt nanoparticles using single-particle ICP-TOFMS N2 - This study describes an interlaboratory comparison (ILC) among nine (9) laboratories to evaluate and validate the standard operation procedure (SOP) for single-particle (sp) ICP-TOFMS developed within the context of the Horizon 2020 project ACEnano. The ILC was based on the characterization of two different Pt nanoparticle (NP) suspensions in terms of particle mass, particle number concentration, and isotopic composition. The two Pt NP suspensions were measured using icpTOF instruments (TOFWERK AG, Switzerland). Two Pt NP samples were characterized and mass equivalent spherical sizes (MESSs) of 40.4 ± 7 nm and 58.8 ± 8 nm were obtained, respectively. MESSs showed <16% relative standard deviation (RSD) among all participating labs and <4% RSD after exclusion of the two outliers. A good agreement was achieved between the different participating laboratories regarding particle mass, but the particle number concentration results were more scattered, with <53% RSD among all laboratories, which is consistent with results from previous ILC studies conducted using ICP-MS instrumentation equipped with a sequential mass spectrometer. Additionally, the capabilities of sp-ICP-TOFMS to determine masses on a particle basis are discussed with respect to the potential for particle density determination. Finally, because quasi-simultaneous multi-isotope and multielement determinations are a strength of ICP-TOFMS instrumentation, the precision and trueness of isotope ratio determinations were assessed. The average of 1000 measured particles yielded a precision of below ±1% for intensity ratios of the most abundant Pt isotopes, i.e. 194Pt and 195Pt, while the accuracy of isotope ratios with the lower abundant isotopes was limited by counting statistics. KW - ILC KW - spICP-MS KW - PtNP KW - Nanopartikel PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-580353 DO - https://doi.org/10.1039/d3nr00435j SN - 2040-3364 VL - 15 IS - 26 SP - 11268 EP - 11279 PB - Royal Society of Chemistry (RSC) AN - OPUS4-58035 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Chronakis, Michail Ioannis A1 - Meermann, Björn A1 - von der Au, Marcus T1 - The evolution of data treatment tools in single-particle and single-cell ICP-MS analytics N2 - Single-particle inductively coupled plasma-mass spectrometry (sp-ICP-MS) is one of the most powerful tools in the thriving field of nanomaterial analysis. Along the same lines, single-cell ICP-MS (sc-ICP-MS) has become an invaluable tool in the study of the variances of cell populations down to a per-cell basis. Their importance and application fields have been listed numerous times, across various reports and reviews. However, not enough attention has been paid to the immense and ongoing development of the tools that are currently available to the analytical community for the acquisition, and more importantly, the treatment of single-particle and single-cell-related data. Due to the ever-increasing demands of modern research, the efficient and dependable treatment of the data has become more important than ever. In addition, the field of single-particle and single-cell analysis suffers due to a large number of approaches for the generated data—with varying levels of specificity and applicability. As a result, finding the appropriate tool or approach, or even comparing results, can be challenging. This article will attempt to bridge these gaps, by covering the evolution and current state of the tools at the disposal of sp-ICP-MS users. Graphical Abstract KW - Trend Article KW - spICP-MS KW - Data Tools KW - scICP-MS PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-610539 DO - https://doi.org/10.1007/s00216-024-05513-4 SP - 1 EP - 7 PB - Springer Science and Business Media LLC AN - OPUS4-61053 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -