TY - JOUR A1 - Taylor, Tristen L. A1 - Tukhmetova, Dariya A1 - Duong, Thi Phuong Thanh A1 - Böwe, Anna-Maria A1 - Meermann, Björn A1 - Gundlach-Graham, Alexander T1 - Comparative study of the vibrating capillary nebulizer (VCN) and commercially available interfaces for on-line coupling of capillary electrophoresis with ICP-MS N2 - Capillary electrophoresis (CE) is a powerful and sensitive tool for speciation analysis when combined with inductively coupled plasma mass spectrometry (ICP-MS); however, the performance of this technique can be limited by the nature of pneumatic nebulizers. This study compares two commercially available pneumatic nebulizers to a newly introduced vibrating capillary nebulizer (VCN) for on-line coupling of CE with ICP-MS. The VCN is a low-cost, non-pneumatic nebulizer that is based on the design of capillary vibrating sharp-edge spray ionization. As a piezoelectrically driven nebulization source, the VCN creates an aerosol independent of gas flows and does not produce a low-pressure region at the nebulizer orifice. To compare the systems, we performed replicate analyses of sulfate in river water with each nebulizer and the same CE and ICP-MS instruments and determined the figures of merit of each setup. With the CE-VCN-ICP-MS setup, we achieved around 2–4 times lower sensitivity compared to the commercial setups. However, the VCN-based setup provided lower noise levels and better linear correlation from the analysis of calibration standards, which resulted in indistinguishable LOD and LOQ values from the in-house-built VCN-based and commercial setups for CE-ICP-MS analysis. The VCN is found to have the highest baseline stability with a standard deviation of 3500 cts s−1, corresponding to an RSD of 2.7%. High reproducibility is found with the VCN with a peak area RSD of 4.1% between 3 replicate measurements. KW - Speciation analysis KW - Analytical chemistry KW - Surface water PY - 2024 DO - https://doi.org/10.1007/s00216-024-05162-7 SN - 1618-2650 SP - 1 EP - 9 PB - Springer CY - Berlin AN - OPUS4-59472 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Chazeau, E. A1 - Fabre, C. A1 - Privat, M. A1 - Godard, A. A1 - Racoeur, C. A1 - Bodio, E. A1 - Busser, B. A1 - Wegner, Karl David A1 - Sancey, L. A1 - Paul, C. A1 - Goze, C. T1 - Comparison of the In Vitro and In Vivo Behavior of a Series of NIR-II-Emitting Aza-BODIPYs Containing Different Water-Solubilizing Groups and Their Trastuzumab Antibody Conjugates N2 - The development of new fluorescent organic probes effective in the NIR-II region is currently a fast-growing field and represents a challenge in the domain of medical imaging. In this study, we have designed and synthesized an innovative series of aza-boron dipyrromethenes emitting in the NIR-II region. We have investigated the effect of different water-solubilizing groups not only on the photophysical properties of the compounds but also on their in vitro and in vivo performance after bioconjugation to the antibody trastuzumab. Remarkably, we discovered that the most lipophilic compound unexpectedly displayed the most favorable in vivo properties after bioconjugation. This underlines the profound influence that the fluorophore functionalization approach can have on the efficiency of the resulting imaging agent. KW - NIR-II KW - In vivo imaging KW - Fluorescence KW - Spectroscopy KW - Antibody conjugates PY - 2024 DO - https://doi.org/10.1021/acs.jmedchem.3c02139 SN - 1520-4804 VL - 67 IS - 5 SP - 3679 EP - 3691 PB - ACS Publications CY - Washington, DC AN - OPUS4-59607 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kornev, Roman A1 - Gornushkin, Igor B. A1 - Shabarova, Lubov A1 - Kadomtseva, Alena A1 - Mochalov, Georgy A1 - Rekunov, Nikita A1 - Romanov, Sergey A1 - Medov, Vitaly A1 - Belousova, Darya A1 - Maleev, Nikita T1 - Plasma-Chemical Disposal of Silicon and Germanium Tetrachlorides Waste by Hydrogen Reduction N2 - The processes of hydrogen reduction of silicon and germanium chlorides under the conditions of high-frequency (40.68 MHz) counteracted arc discharge stabilized between two rod electrodes are investigated. The main gas-phase and solid products of plasma-chemical transformations are determined. Thermodynamic analysis of SiCl4 + H2 and GeCl4 + H2 systems for optimal process parameters was carried out. Using the example of hydrogen reduction of SiCl4 by the method of numerical modeling, gas-dynamic and thermal processes for this type of discharge are investigated. The impurity composition of gas-phase and solid reaction products is investigated. The possibility of single-stage production of high-purity Si and Ge mainly in the form of compact ingots, as well as high-purity chlorosilanes and trichlorogermane, is shown. KW - High-frequency arc discharge KW - Hydrogen reduction KW - Silicon chlorides KW - Germanium chlorides PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-594454 DO - https://doi.org/10.3390/sci6010001 VL - 6 IS - 1 SP - 1 EP - 12 PB - MDPI AG AN - OPUS4-59445 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Blaeß, Carsten A1 - Müller, Ralf A1 - Boccaccini, A. R. T1 - Sintering and crystallization kinetics of bioactive glass 13-93 N2 - This study investigates the sintering and crystallization behavior and kinetic of the bioactive glass (BG) 13–93 with nominal composition (in mol%): 54.6 SiO2 - 1.7 P2O3 - 22.1 CaO - 6.0 Na2O - 7.9 K2O - 7.7 MgO. Sintering and crystallization were investigated non-isothermally for various particle size fractions smaller than 315 μm as well as for bulk samples. Densification was not hindered by the presence of crystalline phases across all particle size fractions. Afterwards, wollastonite was found as the dominant crystal phase at higher temperature which resorb primary surface precipitation-like quartz crystallites. The growth direction shifts into volume when the sample surface is nearly covered. The crystal growth rate of wollastonite was calculated from the crystalline surface layer thickness measured during heating. The findings of this study are relevant for the high temperature processing of BG 13–93. KW - Bioactive glass KW - Sintering KW - Crystallization PY - 2024 DO - https://doi.org/10.1016/j.jnoncrysol.2023.122790 SN - 0022-3093 VL - 627 SP - 1 EP - 7 PB - Elsevier AN - OPUS4-59337 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bayerlein, Bernd A1 - Schilling, Markus A1 - Birkholz, Henk A1 - Jung, Matthias A1 - Waitelonis, Jörg A1 - Mädler, Lutz A1 - Sack, Harald T1 - PMD Core Ontology: Achieving semantic interoperability in materials science N2 - Knowledge representation in the Materials Science and Engineering (MSE) domain is a vast and multi-faceted challenge: Overlap, ambiguity, and inconsistency in terminology are common. Invariant (consistent) and variant (context-specific) knowledge are difficult to align cross-domain. Generic top-level semantic terminology often is too abstract, while MSE domain terminology often is too specific. In this paper, an approach how to maintain a comprehensive MSE-centric terminology composing a mid-level ontology–the Platform MaterialDigital Core Ontology (PMDco)–via MSE community-based curation procedures is presented. The illustrated findings show how the PMDco bridges semantic gaps between high-level, MSE-specific, and other science domain semantics. Additionally, it demonstrates how the PMDco lowers development and integration thresholds. Moreover, the research highlights how to fuel it with real-world data sources ranging from manually conducted experiments and simulations with continuously automated industrial applications. KW - Ontology KW - Materials science and engineering KW - Knowledge representation KW - Reproducibility KW - Semantic interoperability KW - Semantic data integration PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-592948 DO - https://doi.org/10.1016/j.matdes.2023.112603 SN - 0264-1275 VL - 237 SP - 1 EP - 12 PB - Elsevier CY - Amsterdam AN - OPUS4-59294 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Isleyen, Alper A1 - Can, Suleyman Z. A1 - Cankur, Oktay A1 - Engin, Betul Ari A1 - Vogl, Jochen A1 - Koenig, Maren A1 - Horvat, Milena A1 - Jacimovic, Radojko A1 - Zuliani, Tea A1 - Fajon, Vesna A1 - Jotanovic, Aida A1 - Gaževic, Luka A1 - Milosevic, Milena A1 - Ochsenkuehn–Petropoulou, Maria A1 - Tsopelas, Fotis A1 - Lymberopoulou, Theopisti A1 - Tsakanika, Lamprini-Areti A1 - Serifi, Olga A1 - Ochsenkuehn, Klaus M. A1 - Bulska, Ewa A1 - Tomiak, Anna A1 - Kurek, Eliza A1 - Cakılbahçe, Zehra A1 - Aktas, Gokhan A1 - Altuntas, Hatice A1 - Basaran, Elif A1 - Kısacık, Barıs A1 - Gumus, Zeynep T1 - Certification of the total element mass fractions in UME EnvCRM 03 soil sample via a joint research project N2 - Soil certified reference material (CRM), UME EnvCRM 03 was produced by a collaborative approach among national metrology institutes, designated institutes and university research laboratories within the scope of the EMPIR project: Matrix Reference Materials for Environmental Analysis. This paper presents the sampling and processing methodology, homogeneity, stability, characterization campaign, the assignment of property values and their associated uncertainties in compliance with ISO 17034:2016. The material processing methodology involves blending a natural soil sample with a contaminated soil sample obtained by spiking elemental solutions for 8 elements (Cd, Co, Cu, Hg, Ni, Pb, Sb and Zn) to reach the level of warning risk monitoring values specified for metals and metalloids of soils in Europe. Comparative homogeneity and stability test data were obtained by two different institutes, ensuring the reliability and back up of the data. The certified values and associated expanded uncertainties for the total mass fractions of thirteen elements (As, Cd, Co, Cr, Cu, Fe, Hg, Mn, Ni, Pb, Sb, V and Zn) are established. The developed CRM can be used for the development and validation of measurement procedures for the determination of the total mass fractions of elements in soil and also for quality control/assurance purposes. The developed CRM is the first example of a soil material originating from Türkiye. KW - Total element content KW - Soil KW - CRM KW - Certification KW - Environmental pollution monitoring PY - 2024 DO - https://doi.org/10.1007/s00769-024-01597-8 SN - 1432-0517 VL - 29 SP - 293 EP - 301 PB - Springer Science and Business Media LLC AN - OPUS4-59954 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Gehrenkemper, Lennart A1 - Rühl, Isabel A1 - Westphalen, Tanja A1 - Simon, Fabian A1 - von der Au, Marcus A1 - Cossmer, Antje A1 - Meermann, Björn T1 - Investigating the uptake and fate of per- and polyfluoroalkylated substances (PFAS) in bean plants (Phaseolus vulgaris): comparison between target MS and sum parameter analysis via HR-CS-GFMAS N2 - AbstractIn this study, we present a screening method based on molecular absorption spectrometry to study PFAS uptake and fate in plants. To evaluate the suitability of this method we analyzed plant extracts with molecular absorption spectrometry (MAS) as well as liquid chromatography–tandem mass spectrometry (LC–MS/MS) for mass balance studies (w(F)). French bean plants (Phaseolus vulgaris) were grown on soil spiked using eight PFAS substances that vary in chain length and functional group composition. Specifically, these include three short-chained (C4–C5), five long-chained (C7–C10) carboxylic acids, one sulfonic acid and one sulfonic amide moieties. To investigate substance-specific PFAS uptake systematically, PFAS were spiked as single substance spike. Additionally, we studied one mixture of the investigated substances in equal proportions regarding w(F) and four PFAS mixtures of unknown composition. After 6 weeks, the plants were separated into four compartments. We analyzed the four compartments as well as the soil for extractable organically bound fluorine (EOF) by high resolution-continuum source-graphite furnace-molecular absorption spectrometry (HR-CS-GFMAS) as well as for sum of ten target-PFAS by LC–MS/MS. All three short-chained PFAS perfluorobutanoic acid (PFBA), perfluorobutanoic sulfonic acid (PFBS) and perfluoropentanoic acid (PFPeA) were determined in high concentrations mainly in the fruits of the investigated plants while long-chained PFAS perfluorononanoic acid (PFNA) and perfluorodecanoic acid (PFDA) were mainly determined in roots. PFBS was determined in remarkably high concentrations in leaves compartment by both quantification methods. Overall, comprehensive results of single substance spikes were in good agreement for both methods except for a few cases. Hence, two phenomena were identified: for mixed PFAS spikes of unknown composition huge differences between EOF and sum of target PFAS were observed with systematically higher EOF values. Overall, both methods indicate comparable results with MS being more reliable for known PFAS contamination and MAS being more valuable to identify PFAS exposure of unknown composition. Graphical Abstract KW - Pollution PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-589394 DO - https://doi.org/10.1186/s12302-023-00811-7 VL - 35 IS - 104 SP - 1 EP - 13 PB - Springer Science and Business Media LLC AN - OPUS4-58939 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bilsel, M. A1 - Gökçen, T. A1 - Binici, B. A1 - Isleyen, A. A1 - Piechotta, Christian A1 - Kar-wai, A. A1 - Krylov, A. A1 - Miheeva, A. A1 - Beliakov, M. A1 - Palagina, M. A1 - chenko, Irina Tka A1 - Perkola, N. A1 - Lewin, M. A1 - Hua, T. T1 - High polarity analyte(s) in aqueous media: determination of L-PFOA and L-PFOS in ground water N2 - The CCQM-K156 comparison was coordinated by TUBITAK UME on behalf of the Organic Analysis Working Group (OAWG) of the Comité Consultatif pour la Quantité de Matière (CCQM) for National Measurement Institutes (NMIs) and Designated Institutes (DIs) which provide measurement services in organic analysis under the 'Comité International des Poids et Mesures' Mutual Recognition Arrangement (CIPM MRA). Perfluoro alkyl substances (PFAS) such as PFOS and PFOA have been used in numerous industrial applications and products. Because of their high stability and resistance to biodegradation, atmospheric photooxidation, direct photolysis and hydrolysis, they are extremely persistent in the environment. The European Union (EU) Water Framework Directive lists PFOS as a priority hazardous substance that poses a significant risk to the aquatic environment. The use of PFOS-containing Aqueous Film-Forming Foams (AFFFs) has been banned since June 2011 in the EU. As relatively water-soluble, effectively non-degradable compounds, PFOS and PFOA migrate to ground water. They are not removed in the conventional drinking water treatment, and therefore cause health risks in polluted areas. The EU Drinking Water Directive and the European Commission has proposed a limit value of 100 ng/L for the sum of 20 PFAS, including PFOS and PFOA. This study provides the means for assessing measurement capabilities for determination of high polarity measurands in a procedure that requires extraction, clean-up, analytical separation and detection. Successful participation in CCQM-K156 demonstrates measurement capabilities in determining mass fraction of organic compounds, with a molecular mass of 200 g/mol to 700 g/mol, having high polarity pKow -2, in a mass fraction range from 0.5 ng/kg to 500 ng/kg in aqueous media. Nine NMIs and DIs participated in the CCQM-K156 key comparison. Seven institutes reported their results. SPE was applied in the sample pre-treatment and LC-MS was applied for detection. All participating laboratories applied isotope dilution mass spectrometry (IDMS) techniques for quantification. Participants established the metrological traceability of their results using certified reference materials (CRMs) from NMIs with stated traceability; where commercially available high purity materials were used the purity was determined in-house. The CCQM-K156 results for L-PFOA and L-PFOS range from 2.75 ng/kg to 5.50 ng/kg with a % RSD of 19.5 % for L-PFOA and from 2.04 ng/kg to 4.45 ng/kg with a % RSD of 21.3 % for L-PFOS. The KCRV was assigned using a Hierarchical Bayesian Random Effects Model (HB REM) estimator from the values reported by six of the participants. One participant result of L-PFOS and one result of L-PFOA were excluded from the KCRV for technical reasons. The KCRV was 4.9 ng/kg ± 0.4 ng/kg for L-PFOA and 3.8 ng/kg ± 0.4 ng/kg for L-PFOS. The six institutes that were included in the assignment of consensus KCRV all agreed within their standard uncertainties. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). KW - PFAS KW - Surface water KW - ILC KW - CCQM PY - 2022 DO - https://doi.org/10.1088/0026-1394/59/1A/08016 VL - 59 IS - 1A SP - 1 EP - 3 PB - IOP Publishing AN - OPUS4-58941 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Reynolds, Merilie A. A1 - Kamber, Balz S. A1 - McKenna, Cora A. A1 - Oelze, Marcus A1 - Gleeson, Sarah A. T1 - Repeat, fast, and high-resolution mapping of fine-scale trace element distribution in pyrite and marcasite by LA-Q-ICP-MS with the Aerosol Rapid Introduction System (ARIS) N2 - AbstractThe minor and trace element composition of minerals provides critical insights into a variety of geological processes. Multi-element mapping by laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS) is an important technique applied for this purpose and although the method is rapidly advancing, there remains a fundamental compromise between spatial resolution, detection limit, and experiment duration when using sequential mass analyzers. To address the limitation of limited analyte selection for high spatial resolution maps imposed by the sequential nature of typical quadrupole (Q)-ICP-MS, we tested the Aerosol Rapid Introduction System (ARIS) for repeat mapping of the same area. The ARIS is a high-speed transfer tubing system that reduces aerosol washout times, permitting resolution of individual pulses at 40–60 Hz. Here, the ARIS was tested not for pulse resolution but with novel operating conditions optimized to perform fast, high spatial resolution mapping of minor and trace element distribution in pyrite and marcasite. For this purpose, ablation was conducted with a 5 µm beam aperture, a repetition rate of 50 Hz, and a continuous stage scan speed of 40 µm s−1. For each LA-Q-ICP-MS map, data were acquired for six elements with an acquisition time of 20 ms per element. This deliberately reduced the individual pulse resolution of the ARIS but instead exploited the spatial resolution and sensitivity gains afforded by the high-laser repetition rate combined with efficient aerosol transfer. The new method successfully mapped trace elements at single to double-digit parts per million levels, and the maps reveal fine-scale zoning of trace elements with an effective x and y resolution of 5 µm, while white light interferometry showed that for each experiment, only ca. 1 µm of the sample was removed. Repeated mapping of the same area showed excellent correspondence not only between element concentrations in successive experiments but also in the shape, dimension, and location of regions of interest defined by concentration criteria. The very good repeatability of the elemental maps indicates that for studies requiring more analytes, successive mapping of additional elements is possible. By contrast with conventional very small spot (i.e., 5 µm) analysis, fast repetition rate and stage scan speed mapping avoids down-hole fractionation effects and minimizes accidental analysis of buried invisible inclusions. Compared to conventional LA-ICP-MS mapping, the method reduces the experiment time by 4–8 times. KW - Geochemistry and Petrology KW - Geophysics PY - 2023 DO - https://doi.org/10.2138/am-2022-8168 SN - 0003-004X VL - 108 IS - 1 SP - 18 EP - 30 PB - Mineralogical Society of America AN - OPUS4-59038 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rieger, Philip A1 - Magnall, Joseph M. A1 - Gleeson, Sarah A. A1 - Oelze, Marcus T1 - Pyrite chemistry records a multistage ore forming system at the Proterozoic George Fisher massive sulfide Zn-Pb-Ag deposit, Mount Isa, Australia N2 - Trace element (TE) analysis of pyrite via LA-ICP-MS can be used to reconstruct the conditions of pyrite formation in complex mineral systems. The Carpentaria province in northern Australia is host to some of the world’s highest value Zn-Pb (+Ag, Cu) deposits. The genesis of many of these deposits is controversial, with competing models of single-vs. multi-stage ore formation. In this study, LA-ICP-MS data of paragenetically constrained pyrite from the George Fisher Zn-Pb-Ag deposit has been analysed to investigate the chemistry of different stages of ore formation. Pyrite from correlative unmineralized host rocks has also been analysed to investigate evidence of distal hydrothermal anomalism. All LA-ICP-MS data have been statistically evaluated (principal component analysis) and interpreted together with whole rock lithogeochemical data of the same samples. Pre-ore diagenetic pyrite is compositionally similar to other Proterozoic diagenetic pyrite, with some evidence of minor hydrothermal anomalism that with further analysis could help define distal alteration. Pyrite from the different ore stages are compositionally distinct, consistent with a multi-stage system. Ore stage 1 pyrite exceeds background contents of Co, Cu, Zn, As, Ag, Sb, Tl, and Pb and has elevated Co/Ni ratios, whereas only Ni and Co are above background abundances in ore stage 2 and 3 pyrite, of which only ore stage 3 pyrite has high Co/Ni ratios. Ore stage 1 pyrite has a similar composition to hydrothermal pyrite in the undeformed northern Carpentaria CD-type deposits and was likely syn-diagenesis. Ore stage 2 was syn-deformation, and resulted in replacement and recrystallization of pre-existing pyrite, and the expulsion of incompatible TEs. Ore stage 3 formed via a later Cu-Zn-Pb mineralizing event that resulted in a new geochemically distinct generation of Co-rich pyrite. Overall, this study demonstrates the value of paragenetically-constrained pyrite TE data for refining genetic models in complex sediment hosted mineral systems. KW - General Earth and Planetary Sciences PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-590407 DO - https://doi.org/10.3389/feart.2023.892759 VL - 11 SP - 1 EP - 19 PB - Frontiers Media SA AN - OPUS4-59040 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Inostroza, Manuel A1 - Fernandez, Bárbara A1 - Aguilera, Felipe A1 - Layana, Susana A1 - Walter, Thomas R. A1 - Zimmer, Martin A1 - Rodríguez-Díaz, Augusto A1 - Oelze, Marcus T1 - Physical and chemical characteristics of active sulfur flows observed at Lastarria volcano (northern Chile) in January 2019 N2 - Molten sulfur is found in various subaerial volcanoes. However, limited records of the pools and flows of molten sulfur have been reported: therefore, questions remain regarding the physicochemical processes behind this phenomenon. A suite of new sulfur flows, some of which active, was identified at the Lastarria volcano (northern Chile) and studied using satellite imagery, in situ probing, and temperature and video recording. This finding provides a unique opportunity to better understand the emplacement mechanisms and mineral and chemical compositions of molten sulfur, in addition to gaining insight into its origin. Molten sulfur presented temperatures of 124–158°C, with the most prolonged sulfur flow reaching 12 m from the source. Photogrammetric tools permitted the identification of levees and channel structures, with an estimated average flow speed of 0.069 m/s. Field measurements yielded a total volume of 1.45 ± 0.29 m3 of sulfur (equivalent to ∼2.07 tons) mobilized during the January 2019 event for at least 408 min. Solidified sulfur was composed of native sulfur with minor galena and arsenic- and iodine-bearing minerals. Trace element analysis indicated substantial enrichment of Bi, Sb, Sn, Cd, as well as a very high concentration of As (&gt;40.000 ppm). The January 2019 molten sulfur manifestations in Lastarria appear to be more enriched in As compared to the worldwide known volcanoes with molten sulfur records, such as the Shiretoko-Iozan and Poás volcanoes. Furthermore, their rheological properties suggest that the “time of activity” in events such as this could be underestimated as flows in Lastarria have moved significantly slower than previously thought. The origin of molten sulfur is ascribed to the favorable S-rich chemistry of fumarolic gases and changes in host rock permeability (fracture opening). Molten sulfur in Lastarria correlates with a peak in activity characterized by high emissions of SO2 and other acid species, such as HF and HCl, in addition to ground deformation. Consequently, molten sulfur was framed within a period of volcanic unrest in Lastarria, triggered by changes in the magmatic-hydrothermal system. The appearance of molten sulfur is related to physicochemical perturbations inside the volcanic system and is perhaps a precursor of eruptive activity, as observed in the Poás and Turrialba volcanoes. KW - General Earth and Planetary Sciences PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-590331 DO - https://doi.org/10.3389/feart.2023.1197363 VL - 11 SP - 1 EP - 18 PB - Frontiers Media SA AN - OPUS4-59033 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ruyle, B. A1 - Pickard, H. A1 - Schultes, L. A1 - Fredriksson, F. A1 - Heffernan, A. A1 - Knappe, D. A1 - Lord, H. A1 - Meng, P. A1 - Mills, M. A1 - Ndungu, K. A1 - Roesch, Philipp A1 - Rundberget, J.T. A1 - Tettenhorst, D. A1 - Van Buren, J. A1 - Vogel, Christian A1 - Westerman, D. A1 - Yeung, L. A1 - Sunderland, E. T1 - Interlaboratory Comparison of Extractable Organofluorine Measurements in Groundwater and Eel (Anguilla rostrata): Recommendations for Methods Standardization N2 - Research on per- and polyfluoroalkyl substances (PFAS) frequently incorporates organofluorine measurements, particularly because they could support a class-based approach to regulation. However, standardized methods for organofluorine analysis in a broad suite of matrices are currently unavailable, including a method for extractable organofluorine (EOF) measured using combustion ion chromatography (CIC). Here, we report the results of an international interlaboratory comparison. Seven laboratories representing academia, government, and the private sector measured paired EOF and PFAS concentrations in groundwater and eel (Anguilla rostrata) from a site contaminated by aqueous film-forming foam. Among all laboratories, targeted PFAS could not explain all EOF in groundwater but accounted for most EOF in eel. EOF results from all laboratories for at least one replicate extract fell within one standard deviation of the interlaboratory mean for groundwater and five out of seven laboratories for eel. PFAS spike mixture recoveries for EOF measurements in groundwater and eel were close to the criterion (±30%) for standardized targeted PFAS methods. Instrumental operation of the CIC such as replicate sample injections was a major source of measurement uncertainty. Blank contamination and incomplete inorganic fluorine removal may introduce additional uncertainties. To elucidate the presence of unknown organofluorine using paired EOF and PFAS measurements, we recommend that analysts carefully consider confounding methodological uncertainties such as differences in precision between measurements, data processing steps such as blank subtraction and replicate analyses, and the relative recoveries of PFAS and other fluorine compounds. KW - Extractable organic fluorine KW - Per- and Polyfluoroalkyl substances (PFAS) KW - Combustion ion chromatography (CIC) PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-587726 DO - https://doi.org/10.1021/acs.est.3c04560 SN - 0013-936X VL - 57 IS - 48 SP - 20159 EP - 20168 PB - American Chemical Society (ACS) AN - OPUS4-58772 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Yang, Chunliang A1 - Yan, Fan A1 - Meng, Xiangmeng A1 - Putra, Stephen Nugraha A1 - Bachmann, Marcel A1 - Rethmeier, Michael T1 - Phase-field simulation of the dendrite growth in aluminum alloy AA5754 during alternating current electromagnetic stirring laser beam welding N2 - Electromagnetic stirring is known to promote material flow, reduce porosity, uniform elements distribution, and refine grain in laser beam welding (LBW), which enhances the applicability of LBW in various industries. In this study, a phase-field model of dendrite growth in AA5754 Al alloy electromagnetic stirring laser beam welding was established. The model considered the thermal electromagnetic Lorentz force resulting from the interaction between the electric field generated by the Seebeck effect and the magnetic field, as well as the temperature gradient and solidification rate of the solidification interface obtained from the computational fluid dynamics electromagnetic stirring LBW model. The variation rules of dendrite growth with different magnetic parameters and effects are analyzed. Comprehensively, the magnetic field promotes the solidification rate, thus promoting interfacial instability and a large magnetic flux density leads to a faster interface instability. The solidification rate as well as the temperature gradient affect the growth rate, and the accelerated growth caused by the so lidification rate with a high frequency and a large magnetic flux density effectively inhibits the slow growth caused by the temperature gradient. The thermal electromagnetic Lorentz force is the main factor for the branch increment at low frequencies, while both thermal electromagnetic Lorentz force and temperature gradient in crease the number of branches at high frequencies. The calculated average branch numbers considering various factors in the stable stage under different magnetic parameters were consistent with the results of the scanning electron microscope tests. KW - Laser beam welding KW - Electromagnetic KW - Aluminum alloys KW - Phase field method KW - Dendrite growth PY - 2024 DO - https://doi.org/10.1016/j.ijheatmasstransfer.2023.124754 SN - 0017-9310 VL - 218 SP - 1 EP - 16 PB - Elsevier Ltd. AN - OPUS4-58489 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - van Wasen, S. A1 - You, Yi A1 - Beck, S. A1 - Riedel, Jens A1 - Volmer, D. A. T1 - Miniaturized Protein Digestion Using Acoustic Levitation with Online High Resolution Mass Spectrometry N2 - The combination of acoustically levitated droplets, mid-IR laser evaporation, and subsequent post-ionization by secondary electrospray ionization was applied for monitoring the enzymatic digestion of various proteins. Acoustically levitated droplets are an ideal, wall-free model reactor, readily allowing compartmentalized microfluidic trypsin digestions. Time-resolved interrogation of the droplets yielded real-time information on the progress of the reaction and thus provided insights into reaction kinetics. After 30 min of digestion in the acoustic levitator, the obtained protein sequence coverages were identical to the reference overnight digestions. Importantly, our results clearly demonstrate that the applied experimental setup can be used for the real-time investigation of chemical reactions. Furthermore, the described methodology only uses a fraction of the typically applied amounts of solvent, analyte, and trypsin. Thus, the results exemplify the use of acoustic levitation as a green analytical chemistry alternative to the currently used batch reactions. KW - Acoustic levitation KW - Protein analysis KW - Mass spectrometry PY - 2023 DO - https://doi.org/10.1021/acs.analchem.2c05334 VL - 95 SP - 4190 EP - 4195 PB - ACS Publications AN - OPUS4-57053 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Alcalde-Vasquez, R. A1 - Moreno-Pedraza, A. A1 - Rosas-Román, I. A1 - Guillén-Alonso, H. A1 - Riedel, Jens A1 - Partida-Martínez, L. P. A1 - Winkler, R. T1 - MoBiMS: A Modular Miniature Mass Analyzer for the Real-Time 2 Monitoring of Gases and Volatile Compounds in Biological Systems N2 - Living organisms constantly interact with their environment, including through the exchange of gases. Respiration and the release and uptake of volatile organic compounds (VOCs) create dynamic microenvironments in biological systems. Studying the kinetics of volatiles in biological systems requires expensive instruments, and data analysis is challenging. Therefore, we aimed to design a minimal analytical device for measuring the composition of gaseous mixtures in realtime. We built the ‘Modular Biological Mass Spectrometer’ (MoBiMS) from 3D-printed parts and custom sensors to fit a wide array of experimental set-ups. We tested the chemical detection range and temporal resolution of the MoBiMS employing pure compounds and complex biological samples. Compounds with a higher than 0.4 mmHg vapor pressure and a molecular weight up to 154 g/mol were reliably sensed within seconds. The generated electron impact (EI) spectra were directly comparable with standard databases like the NIST EI library. Under a direct analysis approach, the MoBiMS identified the characteristic odor of banana (Musa sp.), that is, isoamyl acetate; tracked the dynamics of CO2 release while the Alka-Seltzer® reaction occurred showed the kinetics of the transient production and consumption of carbon dioxide during photosynthesis. MoBiMS also discriminated between volatile compounds ions coming from tobacco (Nicotiana benthamiana) leaves and the surrounding air through untargeted analysis. The small footprint of the MoBiMS and its relatively low energy consumption facilitate in situ analyses, as an additional gas supply is not necessary with EI ionization. The MoBiMS is easy to assemble, and its construction and operation are very cost-efficient compared to commercial devices. The analytical performance of the MoBiMS is suitable for real-time studies of biological systems, environmental monitoring, and medical diagnostics. KW - Mass spectrometry KW - Instrumentation KW - Miniaturization KW - Volatiles PY - 2022 DO - https://doi.org/10.1016/j.microc.2021.107090 SN - 0026-265X VL - 175 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-56528 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Riedel, Soraya A1 - Jaut, Valerie A1 - Schneider, Rudolf T1 - Ergometrine sensing in rye flour by a magnetic bead-based immunoassay followed by flow injection analysis with amperometric detection N2 - A certain group of mycotoxins, the ergot alkaloids, has caused countless deaths throughout human history. They are found in rye and other cereals and ingesting contaminated foods can cause serious health problems. To identify contaminated food exceeding the legal limits for ergot alkaloids, a portable and cost-effective test system is of great interest to the food industry. Rapid analysis can be achieved by screening for a marker compound, for which we chose ergometrine. We developed a magnetic bead-based immunoassay for ergometrine with amperometric detection in a flow injection system using a handheld potentiostat and a smartphone. With this assay a limit of detection of 3 nM (1 μg/L) was achieved. In spiked rye flour, ergometrine levels from 25 to 250 μg/kg could be quantified. All results could be verified by optical detection. The developed assay offers great promise to meet the demand for on-site ergometrine detection in the food industry. KW - Ergot alkaloids KW - Amperometry KW - Magnetic beads KW - Immunoassay KW - Food analysis KW - Fow injection analysis KW - Mycotoxins PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-565906 DO - https://doi.org/10.1016/j.talanta.2022.124172 SN - 0039-9140 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. VL - 254 IS - 124172 SP - 1 EP - 8 PB - Elsevier B.V. AN - OPUS4-56590 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Simon, Fabian A1 - Gehrenkemper, Lennart A1 - Becher, S. A1 - Dierckes, G. A1 - Langhammer, Nicole A1 - Cossmer, Antje A1 - von der Au, Marcus A1 - Göckener, B. A1 - Fliedner, A. A1 - Rüdel, H. A1 - Koschorreck, J. A1 - Meermann, Björn T1 - Quantification and characterization of PFASs in suspended particulate matter (SPM) of German rivers using EOF, dTOPA, (non-)target HRMS N2 - In this study, we compare analytical methods for PFAS determination–target analysis, non-target screening (NTS), direct total oxidizable precursor assay (dTOPA) and extractable organically bound fluorine (EOF). Therefore, suspended particulate matter (SPM) samples from German rivers at different locations in time series from2005 to 2020 were analyzed to investigate temporal and spatially resolved trends. In this study 3 PFAS mass balances approaches were utilized: (i) PFAA target vs. PFAS dTOPA, (ii) PFAS target vs. EOF and (iii) PFAS target vs. PFAS dTOPA vs. organofluorines NTS vs. EOF. Mass balance approach (i) revealed high proportions of precursor substances in SPM samples. For the time resolved analysis an increase from 94% (2005) to 97% in 2019 was observable. Also for the spatial resolved analysis precursor proportions were high with >84% at all sampling sites. Mass balance approach (ii) showed that the unidentified EOF (uEOF) fraction increased over time from82% (2005) to 99% (2019). Furthermore, along the river courses the uEOF increased. In the combined mass balance approach (iii) using 4 different analytical approaches EOF fractions were further unraveled. The EOF pattern was fully explainable at the sampling sites at Saar and Elbe rivers. For the time resolved analysis, an increased proportion of the EOF was now explainable. However, still 27% of the EOF for the time resolved analysis and 25% of the EOF for the spatial resolved analysis remained unknown. Therefore, in a complementary approach, both the EOF and dTOPA reveal unknown gaps in the PFAS mass balance and are valuable contributions to PFAS risk assessment. Further research is needed to identify organofluorines summarized in the EOF parameter. KW - PFAS KW - HR-CS-GFMAS KW - Fluorine KW - SPM KW - LC-MS/MS PY - 2023 DO - https://doi.org/10.1016/j.scitotenv.2023.163753 SN - 1879-1026 SN - 0048-9697 VL - 885 SP - 1 EP - 12 PB - Elsevier Science CY - Amsterdam AN - OPUS4-57429 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Flierl, Lukas A1 - Rienitz, Olaf A1 - Vogl, Jochen A1 - Pramann, Axel T1 - An advancement of the gravimetric isotope mixture method rendering the knowledge of the spike purity superfluous N2 - The gravimetric isotope mixture method is the primary method to determine absolute isotope ratios. This method, however, depends on the existence of suitable spike materials and knowledge of their purities. Determining the purity of the spikes can be tedious and labour-intensive. In this publication, an advancement of the gravimetric isotope mixture method, rendering the determination of the purity of the spike materials unnecessary, is presented. The advancement combines mass spectrometry and ion chromatography leading to an approach being independent of the purity of the spike materials. In the manuscript the mathematical background and the basic idea of the novel approach are described using a two-isotope system like copper or lithium. KW - Isotope amount ratios KW - Metrology KW - Mass spectrometry KW - Ion chromatography PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-608726 DO - https://doi.org/10.1007/s00216-024-05465-9 SP - 5325 EP - 5333 PB - Springer Science and Business Media LLC AN - OPUS4-60872 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Scholz, Philipp A1 - Vogel, Christian A1 - Schuck, G. A1 - Simon, Franz-Georg T1 - Speciation of copper and zinc compounds relevant for the hazard property (HP) 14 classification of municipal solid waste incineration bottom and fly ashes N2 - The analysis of the presence and content of substances that are toxic to aquatic life in waste is essential for classification of waste with regard to hazard property (HP) 14 ‘ecotoxic’. For the determination of HP14 classified copper (Cu) and zinc (Zn) compounds in various municipal solid waste incineration bottom ashes (IBA) and one fly ash (FA) from Germany we applied X-ray absorption near-edge structure (XANES) spectroscopy in combination with linear combination fitting. The analysis showed that approx. 50–70% of Cu in the IBA are Cu(I) compounds and elemental Cu(0), but these compounds were not equally distributed in the different IBA. In contrast, the majority (approx. 50–70%) of Zn in all IBA is elemental zinc, which originates from brass or other alloys and galvanized metals with a large content of zinc in the waste. The FA contain higher mass fraction on Zn and other toxic elements, but similar Cu and Zn species. Additional performed selective extraction at a pH of 4 with an organic acid of some IBA showed that the ecotoxic Zn fraction is mainly elemental zinc and zinc oxide. In contrast, for the ecotoxic Cu fraction within the IBA no specific compound could be identified. Furthermore, the XANES analysis showed that the HP14 properties of especially Cu in IBA is overestimated with current best-practice guidelines for sample processing for the current substance-related approach with the 0.1% cut-off rule for each substance. However, it should be considered whether it would not be better from an environmental point of view to take the ecotoxicologically leachable copper and zinc as a reference value. KW - Thermal waste treatment KW - XANES spectroscopy KW - Ecotoxicity KW - Incineration bottom ash PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-609584 DO - https://doi.org/10.1016/j.wasman.2024.09.001 VL - 189 SP - 421 EP - 426 PB - Elsevier Ltd. AN - OPUS4-60958 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Wegner, Karl David A1 - Hildebrandt, N. T1 - Near infrared quantum dots for biosensing and bioimaging N2 - Quantum dots (QDs) possess unique optoelectronic properties, which make them very attractive to be used as optical probes in biosensing and bioimaging applications. The strong absorbance and light scattering of biological compounds like tissue and blood in the visible range pose a problem. However, if optical probes emitting in the near-infrared (NIR) range are used, scattering, absorption, and autofluorescence of biological components are strongly reduced. This allows for an increased light penetration depth and higher spatial and temporal resolution for the investigation of biological processes. The synthesis and application of NIR emitting QDs is a fast-growing research field and the benefits of using QDs were demonstrated for a variety of applications, such as photoelectrochemical biosensor, in vivo vascular imaging, and fluorescence-guided surgery. This article reviews the state-of-the-art developments in the preparation of NIR/IR QDs and highlights the latest research about their utilization in biosensing and bioimaging applications. KW - Quantum dots KW - Near-infrared KW - Biosensing KW - Bioimaging KW - NIR-II / SWIR KW - Advanced nanomaterials PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-609708 SN - 0165-9936 DO - https://doi.org/10.1016/j.trac.2024.117922 VL - 180 SP - 1 EP - 19 PB - Elsevier B.V. AN - OPUS4-60970 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fischer, Janina A1 - Kaufmann, Jan Ole A1 - Weller, Michael G. T1 - Simple Determination of Affinity Constants of Antibodies by Competitive Immunoassays N2 - The affinity constant, also known as the equilibrium constant, binding constant, equilibrium association constant, or the reciprocal value, the equilibrium dissociation constant (Kd), can be considered as one of the most important characteristics for any antibody–antigen pair. Many methods based on different technologies have been proposed and used to determine this value. However, since a very large number of publications and commercial datasheets do not include this information, significant obstacles in performing such measurements seem to exist. In other cases where such data are reported, the results have often proved to be unreliable. This situation may indicate that most of the technologies available today require a high level of expertise and effort that does not seem to be available in many laboratories. In this paper, we present a simple approach based on standard immunoassay technology that is easy and quick to perform. It relies on the effect that the molar IC50 approaches the Kd value in the case of infinitely small concentrations of the reagents. A two-dimensional dilution of the reagents leads to an asymptotic convergence to Kd. The approach has some similarity to the well-known checkerboard titration used for the optimization of immunoassays. A well-known antibody against the FLAG peptide, clone M2, was used as a model system and the results were compared with other methods. This approach could be used in any case where a competitive assay is available or can be developed. The determination of an affinity constant should belong to the crucial parameters in any quality control of antibody-related products and assays and should be mandatory in papers using immunochemical protocols. N2 - Die Affinitätskonstante, auch bekannt als Gleichgewichtskonstante, Bindungskonstante, Gleichgewichtsassoziationskonstante oder der reziproke Wert, die Gleichgewichtsdissoziationskonstante (Kd), kann als eine der wichtigsten Eigenschaften für jedes Antikörper-Antigen-Paar angesehen werden. Zur Bestimmung dieses Wertes wurden zahlreiche Methoden auf der Grundlage verschiedener Technologien vorgeschlagen und verwendet. Da jedoch eine sehr große Anzahl von Veröffentlichungen und kommerziellen Datenblätter diese Information nicht enthalten, scheint es erhebliche Hindernisse bei der Durchführung solcher Messungen zu geben. In anderen Fällen, in denen solche Daten angegeben werden, haben sich die Ergebnisse häufig als unzuverlässig erwiesen. Diese Situation könnte darauf hindeuten, dass die meisten der heute verfügbaren Technologien ein hohes Maß an Fachwissen und Aufwand erfordern, das in vielen Labors nicht vorhanden zu sein scheint. In diesem Beitrag stellen wir einen einfachen Ansatz vor, der auf der Standard-Immunoassay-Technologie basiert und einfach und schnell durchführbar ist. Er beruht auf dem Effekt, dass sich der molare IC50 bei unendlich kleinen Reagenz-Konzentrationen dem Kd-Wert annähert. Eine zweidimensionale Verdünnung der Reagenzien führt zu einer asymptotischen Konvergenz zu Kd. Der Ansatz hat eine gewisse Ähnlichkeit mit der bekannten Checkerboard-Titration, die zur Optimierung von Immunoassays verwendet wird. Ein bekannter Antikörper gegen das FLAG-Peptid, Klon M2, wurde als Modellsystem verwendet, und die Ergebnisse wurden mit anderen Methoden verglichen. Dieser Ansatz kann überall dort angewendet werden, wo ein kompetitiver Assay verfügbar ist oder entwickelt werden kann. Die Bestimmung einer Affinitätskonstante sollte zu den entscheidenden Parametern bei der Qualitätskontrolle von Antikörperprodukten und -assays gehören und in Arbeiten, die immunchemische Protokolle verwenden, obligatorisch sein. KW - ELISA KW - Competitive immunoassay KW - IC50 KW - Test midpoint KW - Point of inflection KW - Equilibrium constant KW - Dissociation constant KW - Binding strength KW - Antibody antigen complex KW - Bio-interaction KW - SPR KW - Surface-plasmon resonance KW - FLAG peptide KW - Clone M2 KW - Antibody quality PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-605607 DO - https://doi.org/10.3390/mps7030049 SN - 2409-9279 VL - 7 IS - 3 SP - 1 EP - 11 PB - MDPI CY - Basel AN - OPUS4-60560 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hill, Sarah A1 - Infante, Heidi Goenaga A1 - Entwisle, John A1 - Strekopytov, Stanislav A1 - Ward-Deitrich, Christian A1 - Cowen, Simon A1 - Rienitz, Olaf A1 - Roethke, Anita A1 - Goerlitz, Volker A1 - Schulz, Ursula A1 - Pape, Carola A1 - Vogl, Jochen A1 - Koenig, Maren A1 - Jacimovic, Radojko A1 - Fisicaro, Paola A1 - Ren, Tongxiang A1 - Wang, Song A1 - Song, Panshu A1 - Li, Haifeng A1 - Linsky, Maré A1 - Sobina, Egor A1 - Lozano, Hernán Ezequiel A1 - Puelles, Mabel A1 - Yamani, Randa A1 - Haraldsson, Conny T1 - CCQM-K160: platinum group elements in automotive catalyst N2 - The platinum group elements (PGEs) play an important role in reducing emissions from automotive vehicles through their use in catalytic convertors but also for catalysis in the pharmaceutical industry. The immense economic value of platinum (Pt), palladium (Pd) and rhodium (Rh) highlights the importance of highly accurate measurements. Therefore, there is a need for National Metrology Institutes (NMIs) and Designated Institutes (DIs) to demonstrate measurement capability in this space. A pilot comparison (CCQM-P63) for precious metals in automotive catalyst took place in 2006, but with a limited number of institutes participating. Furthermore, this study was performed over 17 years ago. Therefore, there was a need to maintain existing capability and demonstrate new capability in a key comparison, in order to claim calibration and measurement capability claims (CMCs). With the core capability matrix, this study falls into the "Difficult to dissolve metals/metal oxides" which will support CMC categories 8 (Metal and metal alloys), 9 (Advanced materials) and 14 (Other materials). Eleven NMIs and DIs participated in the Key Comparison CCQM-K160 Platinum Group Elements in Automotive Catalyst. Participants were requested to evaluate the mass fractions of Pt, Pd and Rh in mg/kg in an unused autocatalyst material (cordierite ceramic base). The Key Comparison Reference Values (KCRVs) and Degrees of Equivalence (DoEs) were calculated utilising the NIST Decision Tree for the measurands. The participants utilised a number of sample preparation and analytical methods including hot plate digestion, microwave digestion and sodium fusion, followed by either atomic absorption spectroscopy (AAS), inductively coupled plasma optical emission spectroscopy (ICP-OES) or inductively coupled plasma mass spectrometry (ICP-MS) detection. Several calibration techniques were used, namely external calibration, standard addition, isotope dilution mass spectrometry (IDMS) and an exact matching procedure. Additionally, one participant employed instrumental neutron activation analysis (INAA) with k0 standardisation which is a direct solid analysis method. The majority of participants claimed traceability to NIST primary calibrants or their own CRMs. Furthermore, several matrix CRMs were included or spiked samples for quality control. All institutes were required to determine the dry mass fraction using the stipulated protocol. The NIST decision tree was implemented for the calculation of the KCRVs and DoEs. The participant results overall showed good agreement with the KCRV, despite the variety of dissolution procedures and measurement techniques for this highly complex matrix and challenging measurands. Successful participation in CCQM-K160 demonstrated measurement capabilities for the determination of mass fraction of Pt, Pd and Rh in the mg/kg range and will support broad scope CMC claims for a wide range of challenging matrices. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCQM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). KW - Metrology in Chemistry KW - Traceability KW - Uncertainty PY - 2024 DO - https://doi.org/10.1088/0026-1394/61/1A/08011 VL - 61 IS - 1A SP - 1 EP - 39 PB - IOP Publishing AN - OPUS4-60524 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Simon, Fabian A1 - von der Au, Marcus A1 - Gehrenkemper, Lennart A1 - Meermann, Björn T1 - An optimized method for PFAS analysis using HR–CS–GFMAS via GaF detection N2 - The analysis of per- and polyfluoroalkyl substances (PFAS) via sum parameters like extractable organic fluorine (EOF) in combination with high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR–CS–GFMAS) is highly promising regarding fluorine sensitivity and selectivity. However, the HR–CS–GFMAS method includes several drying and heating steps which can lead to losses of volatile PFAS before the molecular formation step using e.g., GaF formation. Hence, the method leads to a strong discrimination of PFAS within the EOF depending on their physical/chemical properties and is therefore associated with reduced accuracy. To reduce this discrepancy and to indicate realistic PFAS pollution values, an optimization of the HR–CS–GFMAS method for PFAS analysis is needed. Hence, we determined fluorine response factors of several PFAS with different physical/chemical properties upon application of systematic optimization steps. We could therefore improve the method's sensitivity for PFAS analysis using a modifier drying pre-treatment step followed by a sequential injection of sample solutions. The highest improvement in sensitivity of volatile PFAS was shown upon addition of a Mg modifier during drying pre-treatment. Thereby, during optimization the relative standard deviation of fluorine response factors could be reduced from 55 % (initial method) to 27 % (optimized method) leading to a more accurate determination of organofluorine sum parameters. The method provides an instrumental LOD and LOQ of β(F) 1.71 μg/L and 5.13 μg/L, respectively. Further validation aimed to investigate several matrix effects with respect to water matrices. Here, substance-specific behavior was observed. For example, perfluorooctanoic acid (PFOA) which was used as calibrator, showed signal suppressions upon high chloride concentrations (>50 mg/L). Hence, a thorough separation of Cl from analytes during sample preparation is needed for accurate sum parameter analysis. KW - PFAS KW - GF-AAS KW - Method optimization PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-610522 DO - https://doi.org/10.1016/j.talanta.2024.126811 VL - 281 SP - 1 EP - 11 PB - Elsevier B.V. AN - OPUS4-61052 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Chronakis, Michail Ioannis A1 - Meermann, Björn A1 - von der Au, Marcus T1 - The evolution of data treatment tools in single-particle and single-cell ICP-MS analytics N2 - Single-particle inductively coupled plasma-mass spectrometry (sp-ICP-MS) is one of the most powerful tools in the thriving field of nanomaterial analysis. Along the same lines, single-cell ICP-MS (sc-ICP-MS) has become an invaluable tool in the study of the variances of cell populations down to a per-cell basis. Their importance and application fields have been listed numerous times, across various reports and reviews. However, not enough attention has been paid to the immense and ongoing development of the tools that are currently available to the analytical community for the acquisition, and more importantly, the treatment of single-particle and single-cell-related data. Due to the ever-increasing demands of modern research, the efficient and dependable treatment of the data has become more important than ever. In addition, the field of single-particle and single-cell analysis suffers due to a large number of approaches for the generated data—with varying levels of specificity and applicability. As a result, finding the appropriate tool or approach, or even comparing results, can be challenging. This article will attempt to bridge these gaps, by covering the evolution and current state of the tools at the disposal of sp-ICP-MS users. Graphical Abstract KW - Trend Article KW - spICP-MS KW - Data Tools KW - scICP-MS PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-610539 DO - https://doi.org/10.1007/s00216-024-05513-4 SP - 1 EP - 7 PB - Springer Science and Business Media LLC AN - OPUS4-61053 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Mangarova, Dilyana Branimirova A1 - Reimann, Carolin A1 - Kaufmann, Jan Ole A1 - Möckel, Jana A1 - Kader, Avan A1 - Adams, Lisa Christine A1 - Ludwig, Antje A1 - Onthank, David A1 - Robinson, Simon A1 - Karst, Uwe A1 - Helmer, Rebecca A1 - Botnar, Rene A1 - Hamm, Bernd A1 - Makowski, Marcus Richard A1 - Brangsch, Julia T1 - Elastin-specific MR probe for visualization and evaluation of an interleukin-1β targeted therapy for atherosclerosis N2 - Atherosclerosis is a chronic inflammatory condition of the arteries and represents the primary cause of various cardiovascular diseases. Despite ongoing progress, finding effective anti-inflammatory therapeutic strategies for atherosclerosis remains a challenge. Here, we assessed the potential of molecular magnetic resonance imaging (MRI) to visualize the effects of 01BSUR, an anti-interleukin-1β monoclonal antibody, for treating atherosclerosis in a murine model. Male apolipoprotein E-deficient mice were divided into a therapy group (01BSUR, 2 × 0.3 mg/kg subcutaneously, n = 10) and control group (no treatment, n = 10) and received a high-fat diet for eight weeks. The plaque burden was assessed using an elastin-targeted gadolinium-based contrast probe (0.2 mmol/kg intravenously) on a 3 T MRI scanner. T1-weighted imaging showed a significantly lower contrast-to-noise (CNR) ratio in the 01BSUR group (pre: 3.93042664; post: 8.4007067) compared to the control group (pre: 3.70679168; post: 13.2982156) following administration of the elastin-specific MRI probe (p < 0.05). Histological examinations demonstrated a significant reduction in plaque size (p < 0.05) and a significant decrease in plaque elastin content (p < 0.05) in the treatment group compared to control animals. This study demonstrated that 01BSUR hinders the progression of atherosclerosis in a mouse model. Using an elastin-targeted MRI probe, we could quantify these therapeutic effects in MRI. N2 - Atherosklerose ist eine chronische entzündliche Erkrankung der Arterien und die Hauptursache für verschiedene Herz-Kreislauf-Erkrankungen. Trotz kontinuierlicher Fortschritte ist es nach wie vor schwierig, wirksame entzündungshemmende Therapiestrategien für Atherosklerose zu finden. In dieser Studie haben wir das Potenzial der molekularen Magnetresonanztomographie (MRT) zur Visualisierung der Wirkung von 01BSUR, einem monoklonalen Anti-Interleukin-1β-Antikörper, zur Behandlung von Atherosklerose in einem Mausmodell untersucht. Männliche Apolipoprotein-E-defiziente Mäuse wurden in eine Therapiegruppe (01BSUR, 2 × 0,3 mg/kg subkutan, n = 10) und eine Kontrollgruppe (keine Behandlung, n = 10) aufgeteilt und erhielten acht Wochen lang eine fettreiche Diät. Die Plaquebelastung wurde mit einer auf Elastin ausgerichteten Kontrastmittelsonde auf Gadoliniumbasis (0,2 mmol/kg intravenös) auf einem 3-T-MRT-Scanner untersucht. Die T1-gewichtete Bildgebung zeigte ein signifikant niedrigeres Kontrast-Rausch-Verhältnis (CNR) in der 01BSUR-Gruppe (vorher: 3,93042664; nachher: 8,4007067) im Vergleich zur Kontrollgruppe (vorher: 3,70679168; nachher: 13,2982156) nach Verabreichung der Elastin-spezifischen MRT-Sonde (p < 0,05). Histologische Untersuchungen zeigten eine signifikante Verringerung der Plaque-Größe (p < 0,05) und eine signifikante Abnahme des Plaque-Elastin-Gehalts (p < 0,05) in der Behandlungsgruppe im Vergleich zu den Kontrolltieren. Diese Studie zeigte, dass 01BSUR das Fortschreiten der Atherosklerose in einem Mausmodell hemmt. Mit einer auf Elastin ausgerichteten MRT-Sonde konnten wir diese therapeutischen Effekte in der MRT quantifizieren. KW - Atherosclerosis KW - Molecular imaging KW - Elastin KW - Plaque KW - Interleukin-1beta PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-610858 DO - https://doi.org/10.1038/s41598-024-71716-5 VL - 14 IS - 1 SP - 1 EP - 11 PB - Springer Science and Business Media LLC AN - OPUS4-61085 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ecke, Alexander A1 - Westphalen, Tanja A1 - Retzmann, Anika A1 - Schneider, Rudolf T1 - Factors affecting the hydrolysis of the antibiotic amoxicillin in the aquatic environment N2 - The environmental fate of the frequently used broad-spectrum β-lactam antibiotic amoxicillin (AMX) is of high concern regarding the potential evolution of antimicrobial resistance (AMR). Moreover, it is known that AMX is prone to hydrolysis, yielding a variety of hydrolysis products (HPs) with yet unknown effects. Studies to identify those HPs and investigate their formation mechanisms have been reported but a long-term study on their stability in real water samples was missing. In this regard, we investigated the hydrolysis of AMX at two concentration levels in four distinct water types under three different storage conditions over two months. Concentrations of AMX and four relevant HPs were monitored by an LC-MS/MS method revealing pronounced differences in the hydrolysis rate of AMX in tap water and mineral water on the one hand (fast) and surface water on the other(slow). In this context, the occurrence, relative intensities, and stability of certain HPs are more dependent on the water type than on the storage condition. As clarified by ICP-MS, the main difference between the water types was the content of the metals copper and zinc which are supposed to catalyze AMX hydrolysis demonstrating an effective method to degrade AMX at ambient conditions. KW - β-lactam KW - Stability KW - Degradation KW - Hydrolysis products KW - LC-MS/MS KW - ICP-MS PY - 2023 DO - https://doi.org/10.1016/j.chemosphere.2022.136921 SN - 0045-6535 VL - 311 SP - 136921 PB - Elsevier Ltd. AN - OPUS4-56124 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Sieber, M. J. A1 - Wilke, M. A1 - Appelt, O. A1 - Oelze, Marcus A1 - Koch-Müller, M. T1 - Melting relations of Ca–Mg carbonates and trace element signature of carbonate melts up to 9 GPa – A proxy for melting of carbonated mantle lithologies N2 - The most profound consequences of the presence of Ca–Mg carbonates (CaCO3–MgCO3) in the Earth’s upper mantle may be to lower the melting temperatures of the mantle and control the melt composition. Low-degree partial melting of a carbonate-bearing mantle produces CO2-rich, silica-poor melts compositionally imposed by the melting relations of carbonates. Thus, understanding the melting relations in the CaCO3–MgCO3 system facilitates the interpretation of natural carbonate-bearing silicate systems. We report the melting relations of the CaCO3–MgCO3 system and the partition coefficient of trace elements between carbonates and carbonate melt from experiments at high pressure (6 and 9 GPa) and temperature (1300–1800 ◦C) using a rocking multi-anvil press. In the absence of water, Ca–Mg carbonates are stable along geothermal gradients typical of subducting slabs. Ca–Mg carbonates (∼ Mg0.1–0.9Ca0.9–0.1CO3) partially melt beneath mid-ocean ridges and in plume settings. Ca–Mg carbonates melt incongruently, forming periclase crystals and carbonate melt between 4 and 9 GPa. Furthermore, we show that the rare earth element (REE) signature of Group-I kimberlites, namely strong REE fractionation and depletion of heavy REE relative to the primitive mantle, is resembled by carbonate melt in equilibrium with Ca-bearing magnesite and periclase at 6 and 9 GPa. This suggests that the dolomite–magnesite join of the CaCO3–MgCO3 system might be useful to approximate the REE signature of carbonate-rich melts parental to kimberlites. KW - High pressure experiments KW - Laser Ablation PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-561197 DO - https://doi.org/10.5194/ejm-34-411-2022 SN - 0935-1221 VL - 34 IS - 5 SP - 411 EP - 424 PB - Copernicus Publications CY - Göttingen AN - OPUS4-56119 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Steinhäuser, Lorin A1 - Piechotta, Christian A1 - Westphalen, Tanja A1 - Kaminski, Katja T1 - Evaluation, comparison and combination of molecularly imprinted polymer solid phase extraction and classical solid phase extraction for the preconcentration of endocrine disrupting chemicals from representative whole water samples N2 - Estrogens are endocrine disrupting chemicals and of high concerns due to demonstrated harmful effects on the environment and low effect levels. For monitoring and risk assessment, several estrogens were included in the "watch list" of the EU Water Framework Directive which sets very low environmental quality standard (EQS) levels for Estrone (E1) and 17β-Estradiol (E2) of 0.4 ng L−1 and for 17α-Ethinylestradiol (EE2) of 0.035 ng L−1 requiring sensitive detection methods, as well as extensive sample preparation. A sensitive, derivatization-free, isotope dilution calibration HPLC-MS/MS method for a panel of 5 selected estrogens (including the 3 estrogens of the EU WFD watchlist), and a procedure for the reproducible preparation of a representative whole water matrix including mineral water, humic acids and solid particulate matter are presented. These are used in a diligent comparison of classical solid phase extraction (SPE) on hydrophilic-lipophilic balanced (HLB) phase to SPE on an estrogen-specific molecularly imprinted polymer phase (MISPE) for ultra-trace levels of the analytes (1–10 ng L−1). Additionally, a two-step procedure combining HLB SPE disks followed by MISPE is evaluated. The tow-step procedure provides superior enrichment, matrix removal and sample throughput while maintaining comparable recovery rates to simple cartridge SPE. Estimated method quantification limits (MQLs) range from 0.109–0.184 ng L−1 and thus meet EQS-levels for E1 and E2, but not EE2. The representative whole water matrix provides a reproducible comparison of sample preparation methods and lays the foundation for a certified reference material for estrogen analysis. The presented method will serve as the basis for an extended validation study to assess its use for estrogen monitoring in the environment. KW - Estrogens KW - Whole water samples KW - Molecular imprinted polymers KW - EU-WFD PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-563547 DO - https://doi.org/10.1016/j.talo.2022.100163 SN - 2666-8319 VL - 6 SP - 1 EP - 5 PB - Elsevier B.V. CY - Amsterdam, Niederlande AN - OPUS4-56354 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tatzel, M. A1 - Frings, P. J. A1 - Oelze, Marcus A1 - Herwartz, D. A1 - Lünsdorf, K. A1 - Wiedenbeck, M. T1 - Chert oxygen isotope ratios are driven by Earth's thermal evolution N2 - The 18O/16O ratio of cherts (δ18Ochert) increases nearly monotonically by ~15‰ from the Archean to present. Two end-member explanations have emerged: cooling seawater temperature (TSW) and increasing seawater δ18O (δ18Osw). Yet despite decades of work, there is no consensus, leading some to view the δ18Ochert record as pervasively altered. Here, we demonstrate that cherts are a robust archive of diagenetic temperatures, despite metamorphism and exposure to meteoric fluids, and show that the timing and temperature of quartz precipitation and thus δ18Ochert are determined by the kinetics of silica diagenesis. A diagenetic model shows that δ18Ochert is influenced by heat flow through the sediment column. Heat flow has decreased over time as planetary heat is dissipated, and reasonable Archean-modern heat flow changes account for ~5‰ of the increase in δ18Ochert, obviating the need for extreme TSW or δ18Osw reconstructions. The seawater oxygen isotope budget is also influenced by solid Earth cooling, with a recent reconstruction placing Archean δ18OSW 5 to 10‰ lower than today. Together, this provides an internally consistent view of the δ18Ochert record as driven by solid Earth cooling over billion-year timescales that is compatible with Precambrian glaciations and biological. KW - Climate KW - Oxygen isotope ratios KW - Silica diagenesis KW - Early Earth KW - Heat flow PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-569359 DO - https://doi.org/10.1073/pnas.2213076119 SN - 0027-8424 VL - 119 IS - 51 SP - 1 EP - 7 PB - National Academy of Sciences CY - Washington, DC AN - OPUS4-56935 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Chan, George C.-Y. A1 - Engelhard, Carsten A1 - Wiley, Joshua S. A1 - Shoulds, Ayanna U. A1 - Cooks, R. Graham A1 - Hieftje, Gary M. A1 - Shelley, Jacob T. T1 - Characterization of a Low-Temperature Plasma (LTP) Ambient Ionization Source Using Temporally Resolved Monochromatic Imaging Spectrometry N2 - The low-temperature plasma (LTP) probe is a common plasma-based source used for ambient desorption–ionization mass spectrometry (MS). While the LTP probe has been characterized in detail with MS, relatively few studies have used optical spectroscopy. In this paper, two-dimensional (2D) imaging at selected wavelengths is used to visualize important species in the LTP plasma jet. First, 2D steady-state images of the LTP plume for N2+ (391.2 nm), He I (706.5 nm), and N2 (337.1 nm) emissions were recorded under selected plasma conditions. Second, time-resolved 2D emission maps of radiative species in the LTP plasma jet were recorded through the use of a 200 ns detection gate and varying gate delays with respect to the LTP trigger pulse. Emission from He I, N2+, and N2 in the plasma jet region was found to show a transient behavior (often referred to as plasma bullets) lasting only a few microseconds. The N2+ and He I maps were highly correlated in spatial and temporal structure. Further, emission from N2 showed two maxima in time, one before and one after the maximum emission for N2+ and He I, due to an initial electronic excitation wave and ion–electron recombination, respectively. Third, the interaction of the LTP probe with a sample substrate and an electrically grounded metallic needle was studied. Emission from a fluorophore on the sample substrate showed an initial photon-induced excitation from plasma-generated photons followed by electronic excitation by other plasma species. The presence of a grounded needle near the plasma jet significantly extended the plasma jet lifetime and also generated a long-lived corona discharge on the needle. The effect of LTP operating parameters on emission spectra was correlated with mass-spectral results including reagent-ion signals. Lastly, five movies provide a side-by-side comparison of the temporal behavior of emitting species and insights into the interactions of the emission clouds with a sample surface as well as an external needle. Temporally and spatially resolved imaging provided insights into important processes in the LTP plasma jet, which will help improve analyte ion sampling in LTP–MS. KW - Optical spectroscopy KW - Low-temperature plasma KW - Ambient ionization KW - Instrumentation PY - 2023 DO - https://doi.org/10.1177/00037028231184501 VL - 77 IS - 8 SP - 940 EP - 956 PB - SAGE Publications AN - OPUS4-61269 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vo, P. H.N. A1 - Vogel, Christian A1 - Nguyen, H. T.M. A1 - Hamilton, B. R. A1 - Thai, P. K. A1 - Roesch, Philipp A1 - Simon, Franz-Georg A1 - Mueller, J. F. T1 - µ-X-ray fluorescence (XRF) and fluorine K-edge µ-X-ray absorption near-edge structure (XANES) spectroscopy for detection of PFAS distribution in the impacted concrete N2 - An improved understanding of the distribution of per- and polyfluoroalkyl substances (PFAS) in PFAS-impacted concrete is important for risk management and decontamination of PFAS. This study incorporates µ-X-ray fluorescence (µ-XRF) and fluorine K-edge µ-X-ray absorption near-edge structure (µ-XANES) spectroscopy to gain non-destructive insights into PFAS distribution in the impacted concrete. The μ-XRF and μ-XANES spectroscopy provided additional details on the detection of PFAS, which were not detected by the desorption electrospray ionization (DESI) imaging method conducted previously. The shorter chain PFAS were found on the top part of the concrete core (0.5 cm), and longer chain PFAS were mostly at the bottom part of the concrete core (5 cm). The inorganic fluorine fraction was also detected, and it likely hampered the detection of organic fluorine such as PFAS in the concrete. Thus, this non-destructive technique is an complementary approach to detect PFAS in contaminated concrete. KW - Beton KW - Per- and Polyfluoroalkyl substances (PFAS) KW - XANES spectroscopy PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-616711 DO - https://doi.org/10.1016/j.hazl.2024.100134 SN - 2666-9110 VL - 5 SP - 1 EP - 5 PB - Elsevier B.V. AN - OPUS4-61671 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Zweigle, J. A1 - Schmidt, A. A1 - Bugsel, B. A1 - Vogel, Christian A1 - Simon, Fabian A1 - Zwiener, C. T1 - Perfluoroalkyl acid precursor or weakly fluorinated organic compound? A proof of concept for oxidative fractionation of PFAS and organofluorines N2 - Organofluorine mass balance approaches are increasingly applied to investigate the occurrence of per- and polyfluoroalkyl substances (PFAS) and other organofluorines in environmental samples more comprehensively. Usually, complex samples prevent the identification and quantification of every fluorine-containing molecule. Consequently, large unidentified fractions between fluorine sum parameters such as extractable organic fluorine (EOF) and the sum of quantified analytes are frequently reported. We propose using oxidative conversion to separate (unidentified) weakly fluorinated compounds (e.g., pesticides, pharmaceuticals) from PFAA-precursors (perfluoroalkyl chain lengths ≥ C6). We show with three organofluorine model substances (flufenamic acid, diflufenican, pantoprazole) that CF3-groups or aromatic fluorine can be quantitatively converted to inorganic fluoride and trifluoroacetic acid (TFA) by applying PhotoTOP oxidation (UV/TiO2). The principle of fluorine separation in mixtures is demonstrated by the oxidation of the three weakly fluorinated compounds together with the PFAA-precursor 6:2/6:2 fluorotelomer mercaptoalkyl phosphate diester (FTMAP). After oxidation, the products F− and TFA were separated from PFCAs ( C4) by SPE, and the fractions were analyzed individually. Closed mass balances both with and without the addition of organic matrix were achieved. Eventually, the fluorine balance was verified by total fluorine measurements with combustion ion chromatography (CIC). The proposed methods should be considered a proof of concept to potentially explain unidentified fractions of the EOF, especially if compounds with low fluorine content such as pesticides, pharmaceuticals, and their transformation products contribute largely to the EOF. Future studies are needed to show the applicability to the complexity of environmental samples. Graphical Abstract KW - Combustion Ion Chromatography KW - Per- and Polyfluoroalkyl substances (PFAS) PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-613429 DO - https://doi.org/10.1007/s00216-024-05590-5 SP - 1 EP - 10 PB - Springer Science and Business Media LLC AN - OPUS4-61342 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Lancaster, Shaun T. A1 - Sahlin, Eskil A1 - Oelze, Marcus A1 - Ostermann, Markus A1 - Vogl, Jochen A1 - Laperche, Valérie A1 - Touze, Solène A1 - Ghestem, Jean-Philippe A1 - Dalencourt, Claire A1 - Gendre, Régine A1 - Stammeier, Jessica A1 - Klein, Ole A1 - Pröfrock, Daniel A1 - Košarac, Gala A1 - Jotanovic, Aida A1 - Bergamaschi, Luigi A1 - Di Luzio, Marco A1 - D’Agostino, Giancarlo A1 - Jaćimović, Radojko A1 - Eberhard, Melissa A1 - Feiner, Laura A1 - Trimmel, Simone A1 - Rachetti, Alessandra A1 - Sara-Aho, Timo A1 - Roethke, Anita A1 - Michaliszyn, Lena A1 - Pramann, Axel A1 - Rienitz, Olaf A1 - Irrgeher, Johanna T1 - Evaluation of X-ray fluorescence for analysing critical elements in three electronic waste matrices: A comprehensive comparison of analytical techniques N2 - As the drive towards recycling electronic waste increases, demand for rapid and reliable analytical methodology to analyse the metal content of the waste is increasing, e.g. to assess the value of the waste and to decide the correct recycling routes. Here, we comprehensively assess the suitability of different x-ray fluorescence spectroscopy (XRF)-based techniques as rapid analytical tools for the determination of critical raw materials, such as Al, Ti, Mn, Fe, Co, Ni, Cu, Zn, Nb, Pd and Au, in three electronic waste matrices: printed circuit boards (PCB), light emitting diodes (LED), and lithium (Li)-ion batteries. As validated reference methods and materials to establish metrological traceability are lacking, several laboratories measured test samples of each matrix using XRF as well as other independent complementary techniques (instrumental neutron activation analysis (INAA), inductively coupled plasma mass spectrometry (ICP-MS) and ICP optical emission spectrometry (OES)) as an inter-laboratory comparison (ILC). Results highlighted key aspects of sample preparation, limits of detection, and spectral interferences that affect the reliability of XRF, while additionally highlighting that XRF can provide more reliable data for certain elements compared to digestion-based approaches followed by ICP-MS analysis (e.g. group 4 and 5 metals). A clear distinction was observed in data processing methodologies for wavelength dispersive XRF, highlighting that considering the metals present as elements (rather than oxides) induces overestimations of the mass fractions when compared to other techniques. Eventually, the effect of sample particle size was studied and indicated that smaller particle size (<200 μm) is essential for reliable determinations. KW - XRF KW - WEEE KW - PCB KW - Battery KW - LED KW - Recycling PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-614060 DO - https://doi.org/10.1016/j.wasman.2024.10.015 VL - 190 SP - 496 EP - 505 PB - Elsevier B.V. AN - OPUS4-61406 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Jurković, J. A1 - Kazlagić, Anera A1 - Sulejmanović, J. A1 - Smječanin, N. A1 - Karalija, E. A1 - Prkić, A. A1 - Nuhanović, M. A1 - Kolar, M. A1 - Albuquerque, A. T1 - Assessment of heavy metals bioaccumulation in Silver Birch (Betula pendula Roth) from an AMD active, abandoned gold mine waste N2 - Acid mine drainage (AMD) is generally outlined as one of the largest environmental concerns, characterized by very low pH value of mine waste, heavy metals and high sulphate content. This extremely hostile environment reduces plant ability to develop and grow. Present study focuses on a silver birch (Betula pendula Roth), a pioneer species that grows on an extremely hostile gold mine waste, to investigate the bioaccumulation of rare metals (thallium (Tl) and indium (In)), as well as nine other more common heavy metals (bismuth (Bi), cadmium (Cd), cobalt (Co), copper (Cu), lead (Pb), manganese (Mn), nickel (Ni), silver (Ag) and zinc (Zn)), and to asses phytoextraction and phytostabilization potential of silver birch. Additionally, parameters determining AMD process and overall contamination (pH, electrical conductivity (EC), sulphates (SO₄²⁻), arsenic (As), iron (Fe), oxidation–reduction potential (ORP), turbidity, dissolved oxygen (DO), total dissolved solids (TDS), acidity, hardness, X-ray diffraction (XRD) and radioactivity) were determined in mine waste and drainage water samples. To assess the heavy metals bioaccumulation and mine waste status, statistical geochemical indices were determined: bioaccumulation factor (BCF), pollution load index (PLI), geochemical abundance index (GAI) and exposure index (EI). The results show that silver birch bioaccumulates the essential elements Cu, Ni, Mn and Zn, and the nonessential elements Tl (average BCF = 24.99), In (average BC = 23.01) and Pb (average BCF = 0.84). Investigated mine waste was enriched by Bi, Ag and Cd according to positive values of GAI index. Present research provides a novel insight into bioaccumulation of nonessential heavy metals in silver birches who grow on the extremely hostile mine waste, and they exhibit significant phytoremediation potential. KW - Heavy metals KW - Acid mine drainage KW - Bioaccumulation KW - Silver birch KW - Thallium PY - 2023 DO - https://doi.org/10.1007/s10653-023-01774-7 SN - 1573-2983 SN - 0142-7245 VL - 45 SP - 9855 EP - 9873 PB - Springer Science and Business Media CY - Dordrecht AN - OPUS4-61835 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kader, A. A1 - Snellings, J. A1 - Adams, L. C. A1 - Gottheil, P. A1 - Mangarova, D. B. A1 - Heyl, J. L. A1 - Kaufmann, Jan Ole A1 - Moeckel, J. A1 - Brangsch, J. A1 - Auer, T. A. A1 - Collettini, F. A1 - Sauer, F. A1 - Hamm, B. A1 - Käs, J. A1 - Sack, I. A1 - Makowski, M. R. A1 - Braun, J. T1 - Sensitivity of magnetic resonance elastography to extracellular matrix and cell motility in human prostate cancer cell line-derived xenograft models N2 - Prostate cancer (PCa) is a significant health problem in the male population of the Western world. Magnetic resonance elastography (MRE), an emerging medical imaging technique sensitive to mechanical properties of biological tissues, detects PCa based on abnormally high stiffness and viscosity values. Yet, the origin of these changes in tissue properties and how they correlate with histopathological markers and tumor aggressiveness are largely unknown, hindering the use of tumor biomechanical properties for establishing a noninvasive PCa staging system. To infer the contributions of extracellular matrix (ECM) components and cell motility, we investigated fresh tissue specimens from two PCa xenograft mouse models, PC3 and LNCaP, using magnetic resonance elastography (MRE), diffusion-weighted imaging (DWI), quantitative histology, and nuclear shape analysis. Increased tumor stiffness and impaired water diffusion were observed to be associated with collagen and elastin accumulation and decreased cell motility. Overall, LNCaP, while more representative of clinical PCa than PC3, accumulated fewer ECM components, induced less restriction of water diffusion, and exhibited increased cell motility, resulting in overall softer and less viscous properties. Taken together, our results suggest that prostate tumor stiffness increases with ECM accumulation and cell adhesion - characteristics that influence critical biological processes of cancer development. MRE paired with DWI provides a powerful set of imaging markers that can potentially predict prostate tumor development from benign masses to aggressive malignancies in patients. Statement of significance: Xenograft models of human prostate tumor cell lines, allowing correlation of microstructure-sensitive biophysical imaging parameters with quantitative histological methods, can be investigated to identify hallmarks of cancer. KW - Magnetic resonance elastography KW - Quantitative histology KW - Prostate cancer KW - Stiffness KW - Fluidity KW - Apparent diffusion coefficient KW - Collagen KW - Elastin KW - Nuclear shape PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-618361 DO - https://doi.org/10.1016/j.bioadv.2024.213884 SN - 2772-9508 VL - 161 SP - 1 EP - 12 PB - Elsevier CY - Amsterdam AN - OPUS4-61836 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bosc-Bierne, Gaby A1 - Ewald, Shireen A1 - Kreuzer, Oliver J. A1 - Weller, Michael G. T1 - Efficient Quality Control of Peptide Pools by UHPLC and Simultaneous UV and HRMS Detection N2 - Peptide pools consist of short amino acid sequences and have proven to be versatile tools in various research areas in immunology and clinical applications. They are commercially available in many different compositions and variants. However, unlike other reagents that consist of only one or a few compounds, peptide pools are highly complex products which makes their quality control a major challenge. Quantitative peptide analysis usually requires sophisticated methods, in most cases isotope-labeled standards and reference materials. Usually, this would be prohibitively laborious and expensive. Therefore, an approach is needed to provide a practical and feasible method for quality control of peptide pools. With insufficient quality control, the use of such products could lead to incorrect experimental results, worsening the well-known reproducibility crisis in the biomedical sciences. Here we propose the use of ultra-high performance liquid chromatography (UHPLC) with two detectors, a standard UV detector at 214 nm for quantitative analysis and a high-resolution mass spectrometer (HRMS) for identity confirmation. To be cost-efficient and fast, quantification and identification are performed in one chromatographic run. An optimized protocol is shown, and different peak integration methods are compared and discussed. This work was performed using a peptide pool known as CEF advanced, which consists of 32 peptides derived from cytomegalovirus (CMV), Epstein–Barr virus (EBV) and influenza virus, ranging from 8 to 12 amino acids in length. N2 - Peptidpools bestehen aus kurzen Aminosäuresequenzen und haben sich als vielseitige Werkzeuge in verschiedenen Forschungsbereichen der Immunologie und bei klinischen Anwendungen erwiesen. Sie sind in vielen verschiedenen Zusammensetzungen und Varianten im Handel erhältlich. Im Gegensatz zu anderen Reagenzien, die nur aus einer oder wenigen Verbindungen bestehen, sind Peptidpools jedoch hochkomplexe Produkte, was ihre Qualitätskontrolle zu einer großen Herausforderung macht. Die quantitative Peptidanalyse erfordert in der Regel ausgefeilte Methoden, in den meisten Fällen isotopenmarkierte Standards und Referenzmaterialien. Dies ist in der Regel sehr aufwändig und teuer. Daher wird ein Ansatz benötigt, der eine praktische und praktikable Methode zur Qualitätskontrolle von Peptidpools bietet. Bei unzureichender Qualitätskontrolle könnte die Verwendung solcher Produkte zu falschen Versuchsergebnissen führen, was das bekannte Problem der Reproduzierbarkeit in den biomedizinischen Wissenschaften noch verschärfen würde. Hier schlagen wir die Verwendung der Ultrahochleistungs-Flüssigkeitschromatographie (UHPLC) mit zwei Detektoren vor, einem Standard-UV-Detektor bei 214 nm für die quantitative Analyse und einem hochauflösenden Massenspektrometer (HRMS) für die Identitätsbestätigung. Um kosteneffizient und schnell zu sein, werden Quantifizierung und Identifizierung in einem einzigen chromatographischen Lauf durchgeführt. Es wird ein optimiertes Protokoll gezeigt, und es werden verschiedene Peak-Integrationsmethoden verglichen und diskutiert. Für diese Arbeit wurde ein Peptidpool verwendet, der als CEF advanced bekannt ist und aus 32 Peptiden besteht, die vom Cytomegalovirus (CMV), Epstein-Barr-Virus (EBV) und Influenzavirus stammen und zwischen 8 und 12 Aminosäuren lang sind. KW - Synthetic peptides KW - Quality control KW - Impurites KW - Byproducts KW - Degradation KW - Mass spectrometry KW - Orbitrap PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-602299 DO - https://doi.org/10.3390/separations11050156 SN - 2297-8739 VL - 11 IS - 5 SP - 1 EP - 18 PB - MDPI CY - Basel AN - OPUS4-60229 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Pavasarytė, Lina A1 - Azevedo do Nascimento, Allana A1 - Cysne Barbosa, Ana Paula A1 - Trappe, Volker A1 - Melo, Daniel T1 - Effects of particle size and particle concentration of poly (ethylene-co-methacrylic acid) on properties of epoxy resin N2 - Self-healing polymers have been developed to improve durability and reduce costs associated with maintenance during service. The addition of thermoplastics to thermosets to produce mendable polymers appears as a promising selfhealing technique. In this study, poly (ethylene-co-methacrylic acid) (EMAA) was added to epoxy resin and the effects of EMAA addition on epoxy properties were evaluated. Specimens with two different contents of thermoplastic and particles sizes were manufactured. A two-level full factorial experimental design was used to evaluate the effect of particle size and particle content on properties of epoxy modified with addition of EMAA. Tensile tests and dynamic mechanical analysis (DMA) were used and the evaluated responses were tensile strength, modulus of elasticity, and glass transition temperature (Tg). X-ray computed tomography (XCT) was used to investigate particle size and concentration after manufacturing. It was found that the particle concentration has greater effects on stress–strain behavior of epoxy while Tg was not significantly affected by neither of the analyzed entrance variables. KW - Fracture KW - Self-healing KW - Epoxy KW - Thermoplastic PY - 2024 DO - https://doi.org/10.1002/app.55677 SN - 0021-8995 SP - 1 EP - 14 PB - Wiley online library AN - OPUS4-60205 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Groschke, Matthias A1 - Becker, Roland T1 - Comparison of carrier gases for the separation and quantification of mineral oil hydrocarbon (MOH) fractions using online coupled high performance liquid chromatography-gas chromatography-flame ionisation detection N2 - On-line coupled high performance liquid chromatography-gas chromatography-flame ionisation detection (HPLC-GC-FID) was used to compare the effect of hydrogen, helium and nitrogen as carrier gases on the chromatographic characteristics for the quantification of mineral oil hydrocarbon (MOH) traces in food related matrices. After optimisation of chromatographic parameters nitrogen carrier gas exhibited characteristics equivalent to hydrogen and helium regarding requirements set by current guidelines and standardisation such as linear range, quantification limit and carry over. Though nitrogen expectedly led to greater peak widths, all required separations of standard compounds were sufficient and humps of saturated mineral oil hydrocarbons (MOSH) and aromatic mineral oil hydrocarbons (MOAH) were appropriate to enable quantitation similar to situations where hydrogen or helium had been used. Slightly increased peak widths of individual hump components did not affect shapes and widths of the MOSH and MOAH humps were not significantly affected by the use of nitrogen as carrier gas. Notably, nitrogen carrier gas led to less solvent peak tailing and smaller baseline offset. Overall, nitrogen may be regarded as viable alternative to hydrogen or helium and may even extend the range of quantifiable compounds to highly volatile hydrocarbon eluting directly after the solvent peak. KW - Mineral oil hydrocarobons KW - Food KW - Liquid chromatography KW - Gas chromatography KW - MOSH/MOAH PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-601450 DO - https://doi.org/10.1016/j.chroma.2024.464946 SN - 0021-9673 VL - 1726 SP - 1 EP - 7 PB - Elsevier CY - New York, NY AN - OPUS4-60145 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schmidt, Stephan A1 - Hoffmann, Holger A1 - Garbe, Leif-Alexander A1 - Harrer, Andrea A1 - Steiner, Markus A1 - Himly, Martin A1 - Schneider, Rudolf T1 - Re-assessment of monoclonal antibodies against diclofenac for their application in the analysis of environmental waters N2 - The non-steroidal anti-inflammatory drug (NSAID) diclofenac (DCF) is an important environmental contaminant occurring in surface waters all over the world, because, after excretion, it is not adequately removed from wastewater in sewage treatment plants. To be able to monitor this pollutant, highly efficient analytical methods are needed, including immunoassays. In a medical research project, monoclonal antibodies against diclofenac and its metabolites had been produced. Based on this monoclonal anti-DCF antibody, a new indirect competitive enzyme-linked immunosorbent assay (ELISA) was developed and applied for environmental samples. The introduction of a spacer between diclofenac and the carrier protein in the coating conjugate led to higher sensitivity. With a test midpoint of 3 mg L−1 and a measurement range of 1–30 mg L−1, the system is not sensitive enough for direct analysis of surface water. However, this assay is quite robust against matrix influences and can be used for wastewater. Without adjustment of the calibration, organic solvents up to 5%, natural organic matter (NOM) up to 10 mg L−1, humic acids up to 2.5 mg L−1, and salt concentrations up to 6 g L−1 NaCl and 75 mg L−1 CaCl2 are tolerated. The antibody is also stable in a pH range from 3 to 12. Cross-reactivity (CR) of 1% or less was determined for the metabolites 40-hydroxydiclofenac (40-OH-DCF), 5-hydroxydiclofenac (5-OH-DCF), DCF lactam, and other NSAIDs. Relevant cross-reactivity occurred only with an amide derivative of DCF, 6-aminohexanoic acid (DCF-Ahx), aceclofenac (ACF) and DCF methyl ester (DCF-Me) with 150%, 61% and 44%, respectively. These substances, however, have not been found in samples. Only DCF-acyl glucuronide with a cross-reactivity of 57% is of some relevance. For the first time, photodegradation products were tested for cross-reactivity. With the ELISA based on this antibody, water samples were analysed. In sewage treatment plant effluents, concentrations in the range of 1.9–5.2 mg L−1 were determined directly, with recoveries compared to HPLC-MS/MS averaging 136%. Concentrations in lakes ranged from 3 to 4.4 ng L−1 and were, after pre-concentration, determined with an average recovery of 100% KW - Antikörper KW - Immunoassay KW - Wasser PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-602280 DO - https://doi.org/10.1039/d3ay01333b VL - 16 IS - 21 SP - 3349 EP - 3363 PB - Royal Society of Chemistry (RSC) CY - London AN - OPUS4-60228 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tatzel, Michael A1 - Oelze, Marcus A1 - Frick, Daniel A. A1 - Di Rocco, Tommaso A1 - Liesegang, Moritz A1 - Stuff, Maria A1 - Wiedenbeck, Michael T1 - Silicon and oxygen isotope fractionation in a silicified carbonate rock N2 - Silicon isotope fractionation during silicification is poorly understood and impedes our ability to decipher paleoenvironmental conditions from Si isotopes in ancient cherts. To investigate isotope fractionation during silica-for-carbonate replacement we analyzed the microscale Si and O isotope composition in different silica phases in a silicified zebra dolostone as well as their bulk δ18O and Δ’17O compositions. The subsequent replacement of carbonate layers is mimicked by decreasing δ18O and δ30Si. The textural relationship and magnitude of Si and O isotope fractionation is best explained by near-quantitative silica precipitation in an open system with finite Si. A Rayleigh model for silicification suggests positive Ɛ30/28Si during silicification, conforming with predictions for isotope distribution at chemical equilibrium from ab-initio models. Application of the modelled Ɛ30Si-T relationship yields silicification temperatures of approx. 50°C. To reconcile the δ18Ochert composition with these temperatures, the δ18O of the fluid must have been between -2.5 and -4 ‰, compositions for which the quartz phases fall close to the oxygen equilibrium fractionation line in three-isotope space. Diagenetic silica replacement appears to occur in O and Si isotopic equilibrium allowing reconstructions of temperatures of silicification from Si isotopes and derive the δ18O composition of the fluid – a highly desired value needed for accurate reconstructions of the temperature- and δ18O histories of the oceans. KW - Silicon isotopes PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-603430 DO - https://doi.org/10.1016/j.chemgeo.2024.122120 SN - 0009-2541 VL - 658 SP - 1 EP - 13 PB - Elsevier B.V. AN - OPUS4-60343 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tukhmetova, Dariya A1 - Langhammer, Nicole A1 - Vogl, Jochen A1 - Meermann, Björn T1 - Online Isotope Analysis of Sulfur in Proteins via Capillary Electrophoresis Coupled With Multicollector ICP‐MS (CE/MC‐ICP‐MS): A Proof of Concept Study N2 - Isotope ratio analysis of sulfur in biological samples using inductively coupled plasma-mass spectrometry (ICP-MS) has gained significant interest for applications in quantitative proteomics. Advancements like coupling separation techniques with multicollector ICP-MS (MC-ICP-MS) enhance the throughput of species-specific sulfur isotope ratio measurements, fostering new avenues for studying sulfur metabolism in complex biological matrices. This proof-of-concept study investigates the feasibility of online CE/MC-ICP-MS for directly analyzing sulfur isotope ratios in proteins (albumin). Leveraging our previous work on the applicability of CE/ICP-MS for quantifying sulfur-containing biological molecules, we explore its potential for sulfur isotope analysis. Our results demonstrate that direct analysis of sulfur isotopes in albumin protein using online capillary electrophoresis MC-ICP-MS (CE/MC-ICP-MS) eliminates the need for laborious pretreatment steps, while yielding isotope ratios comparable to the reference values. Although initial precision can be improved through further system optimization and protein injection techniques, this approach paves the way for future analysis of mixtures of various biological compounds in, for example, clinical diagnosis studies. KW - Isotope analysis KW - Sulfur KW - CE/MC-ICP-MS PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-612439 DO - https://doi.org/10.1002/elps.202400128 SN - 0173-0835 SP - 1 EP - 6 PB - Wiley-Blackwell CY - Weinheim AN - OPUS4-61243 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Duddi, M. A1 - Kocherla, Amarteja A1 - Subramaniam, K. V. L. T1 - Evaluation of boundary and material influences on the dynamic electromechanical impedance response of the embedded PZT sensor in concrete N2 - The article explores the material and structural size effects on the electrical impedance response of the embedded PZT sensors which are used as an sensors/actuator in the smart sensing concrete. An embedded sensor for use in concrete monitoring is developed using Lead Zirconate Titanate (PZT). The electrical impedance (EI) response of the sensor placed inside concrete cubes of varying sizes are recorded for an applied electrical input of varying frequency. The dynamic response of the embedded PZT sensor to the applied electrical input is determined by the mechanical impedance offered by the surrounding medium. In specimens of finite size, the mechanical impedance to the motion of the PZT is influenced by the combination of dynamic material resistance and structural stiffness (size of the structure). The current study examined the effects of material resistance and size of structure on the EI response of the PZT sensor embedded in concrete medium. The influence of the size of the medium from small-sized specimens is seen in the form of secondary peaks, identified as structural peaks, on the resonance spectrum of the embedded PZT sensor. The structural peaks on the EI response are very dominant in the concrete cube of small sizes. Magnitude and intensity of these structural peaks attenuate with an increase in the size of the concrete cube. Also, the EI response from the smaller-sized specimens registered a significant leftward shift in the spectrum. The observations from the numerical simulations also showed the mechanical impedance of the surrounding concrete medium to the motion of the PZT also increases and converges at 150 mm. At sizes below 150 mm, the structural effects that are influenced by the specimen boundary must be separated to access the material properties of the concrete medium from the measured EI response of an embedded PZT sensor. At sizes, larger than 150 mm, the dynamic motion of the PZT patch and the measured EI response of the PZT patch depend only on the dynamic material properties of the concrete medium around the sensor. The zone of influence and its importance in standardization of PZT sensor for SHM of concrete structures is discussed. KW - PZT KW - Concrete sensor KW - Embedded KW - Electrical impedance KW - Mechanical impedance KW - Zone of influence PY - 2023 DO - https://doi.org/10.1016/j.sna.2023.114575 SN - 0924-4247 SN - 1873-3069 VL - 361 SP - 1 EP - 12 PB - Elsevier CY - Amsterdam AN - OPUS4-61791 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dhull, N. A1 - Kaur, Gurpreet A1 - Jindal, K. A1 - Verma, M. A1 - Tomar, M. T1 - Microfluidics integrated NiO based electrolyte-gated FETs for the detection of cortisol N2 - A microfluidics integrated EGFET has been devised using NiO for the detection of cortisol. A low detection limit of 0.5 fg mL−1 has been achieved and human salivary cortisol has been successfully assessed. KW - Microfluidic system KW - Field effect transistor (FET) KW - Electrolyte gating PY - 2022 DO - https://doi.org/10.1039/d2tb01652d SN - 2050-7518 VL - 10 IS - 44 SP - 9226 EP - 9234 PB - Royal Society of Chemistry (RSC) CY - London AN - OPUS4-61772 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Li, Yue A1 - Colnaghi, Timoteo A1 - Gong, Yilun A1 - Zhang, Huaide A1 - Yu, Yuan A1 - Wei, Ye A1 - Gan, Bin A1 - Song, Min A1 - Marek, Andreas A1 - Rampp, Markus A1 - Zhang, Siyuan A1 - Pei, Zongrui A1 - Wuttig, Matthias A1 - Ghosh, Sheuly A1 - Körmann, Fritz A1 - Neugebauer, Jörg A1 - Wang, Zhangwei A1 - Gault, Baptiste T1 - Machine learning‐enabled tomographic imaging of chemical short‐range atomic ordering N2 - In solids, chemical short‐range order (CSRO) refers to the self‐organization of atoms of certain species occupying specific crystal sites. CSRO is increasingly being envisaged as a lever to tailor the mechanical and functional properties of materials. Yet quantitative relationships between properties and the morphology, number density, and atomic configurations of CSRO domains remain elusive. Herein, it is showcased how machine learning‐enhanced atom probe tomography (APT) can mine the near‐atomically resolved APT data and jointly exploit the technique's high elemental sensitivity to provide a 3D quantitative analysis of CSRO in a CoCrNi medium‐entropy alloy. Multiple CSRO configurations are revealed, with their formation supported by state‐of‐the‐art Monte‐Carlo simulations. Quantitative analysis of these CSROs allows establishing relationships between processing parameters and physical properties. The unambiguous characterization of CSRO will help refine strategies for designing advanced materials by manipulating atomic‐scale architectures. KW - Chemical short-range order (CSRO) KW - Atom probe tomography (APT) KW - Machine learning PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-623777 DO - https://doi.org/10.1002/adma.202407564 SN - 1521-4095 VL - 36 IS - 44 SP - 1 EP - 9 PB - Wiley-VCH CY - Weinheim AN - OPUS4-62377 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Arıkan, Muzaffer A1 - Demir, Tuğçe Kahraman A1 - Yıldız, Zeynep A1 - Nalbantoğlu, Özkan Ufuk A1 - Korkmaz, Nur Damla A1 - Yılmaz, Nesrin H. A1 - Şen, Aysu A1 - Özcan, Mutlu A1 - Muth, Thilo A1 - Hanoğlu, Lütfü A1 - Yıldırım, Süleyman T1 - Metaproteogenomic analysis of saliva samples from Parkinson's disease patients with cognitive impairment N2 - Cognitive impairment (CI) is very common in patients with Parkinson’s Disease (PD) and progressively develops on a spectrum from mild cognitive impairment (PD-MCI) to full dementia (PDD). Identification of PD patients at risk of developing cognitive decline, therefore, is unmet need in the clinic to manage the disease. Previous studies reported that oral microbiota of PD patients was altered even at early stages and poor oral hygiene is associated with dementia. However, data from single modalities are often unable to explain complex chronic diseases in the brain and cannot reliably predict the risk of disease progression. Here, we performed integrative metaproteogenomic characterization of salivary microbiota and tested the hypothesis that biological molecules of saliva and saliva microbiota dynamically shift in association with the progression of cognitive decline and harbor discriminatory key signatures across the spectrum of CI in PD. We recruited a cohort of 115 participants in a multi-center study and employed multi-omics factor analysis (MOFA) to integrate amplicon sequencing and metaproteomic analysis to identify signature taxa and proteins in saliva. Our baseline analyses revealed contrasting interplay between the genus Neisseria and Lactobacillus and Ligilactobacillus genera across the spectrum of CI. The group specific signature profiles enabled us to identify bacterial genera and protein groups associated with CI stages in PD. Our study describes compositional dynamics of saliva across the spectrum of CI in PD and paves the way for developing non-invasive biomarker strategies to predict the risk of CI progression in PD. KW - Cognitive impairment (CI) KW - Parkinson’s disease (PD) KW - Metaproteogenomic analysis KW - Saliva sample PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-624228 DO - https://doi.org/10.1038/s41522-023-00452-x SN - 2055-5008 VL - 9 IS - 1 SP - 1 EP - 10 PB - Nature Publ. Group CY - London AN - OPUS4-62422 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bosc-Bierne, Gaby A1 - Weller, Michael G. T1 - Investigation of impurities in peptide pools T1 - Untersuchung von Verunreinigungen in Peptidpools N2 - Peptide pools are important research tools in different biomedical fields. They consist of a complex mixture of defined peptides, which places high demands on the production and quality control of these products. Previously it has been shown that the combination of UHPLC with high-resolution mass-spectrometry (HRMS) is a fast and powerful method to confirm the relative concentration and the structural identity of all peptides expected to be in the pool. In this work, the additional information contained in the UV chromatograms and mass spectra is used to search for impurities due to synthesis by-products and degradation during storage and transportation and to identify possible analytical artifacts. It was shown that most impurities are only present in trace amounts and can be considered uncritical for most applications. The most frequent and perhaps unexpected impurities were homo- and heterodimers caused by the free cysteines contained in these peptide pools. Furthermore, pyroglutamate and aspartimide formation, deamidation, methionine oxidation, and amino acid deletions could be found. This list is not intended to be comprehensive, but rather a brief guide to quickly identify impurities and, in the long term, to suggest possible changes in the composition of the peptide pools to avoid such impurities by design or by special precautions. N2 - Peptidpools sind wichtige Forschungsinstrumente in verschiedenen biomedizinischen Bereichen. Sie bestehen aus einer komplexen Mischung definierter Peptide, was hohe Anforderungen an die Herstellung und Qualitätskontrolle dieser Produkte stellt. In der Vergangenheit wurde gezeigt, dass die Kombination von UHPLC mit hochauflösender Massenspektrometrie (HRMS) eine schnelle und leistungsfähige Methode zur Bestätigung der relativen Konzentration und der strukturellen Identität aller Peptide ist, die in dem Pool erwartet werden. In dieser Arbeit werden die zusätzlichen Informationen, die in den UV-Chromatogrammen und Massenspektren enthalten sind, genutzt, um nach Verunreinigungen zu suchen, die auf Nebenprodukte der Synthese und den Abbau während der Lagerung und des Transports zurückzuführen sind, und um mögliche analytische Artefakte zu identifizieren. Es hat sich gezeigt, dass die meisten Verunreinigungen nur in Spuren vorhanden sind und für die meisten Anwendungen als unkritisch angesehen werden können. Die häufigsten und vielleicht unerwartetsten Verunreinigungen waren Homo- und Heterodimere, die durch die in diesen Peptidpools enthaltenen freien Cysteine verursacht werden. Außerdem wurden Pyroglutamat- und Aspartimidbildung, Deamidierung, Methioninoxidation und Aminosäuredeletionen festgestellt. Diese Liste erhebt keinen Anspruch auf Vollständigkeit, sondern ist eher ein kurzer Leitfaden, um Verunreinigungen schnell zu identifizieren und langfristig mögliche Änderungen in der Zusammensetzung der Peptidpools vorzuschlagen, um solche Verunreinigungen durch Design oder besondere Vorsichtsmaßnahmen zu vermeiden. KW - Synthetic peptides KW - Degradation KW - Synthetic artifacts KW - Peptide losses KW - Stability of peptides KW - Disulfide dimers KW - Cysteine dimers KW - Isomers KW - Cysteine alkylation PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-626987 DO - https://doi.org/10.3390/separations12020036 SN - 2297-8739 VL - 12 IS - 2 SP - 1 EP - 18 PB - MDPI AG CY - Basel AN - OPUS4-62698 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Völker, Tobias A1 - Mensing, Friedrich A1 - Kruschwitz, Sabine T1 - Estimation of cement content in concrete by spatially resolved laser induced breakdown spectroscopy N2 - The cement content in concrete significantly influences critical properties such as durability, permeability, strength, and workability. Traditional methods for estimating the cement content face limitations. These include the need for comprehensive chemical and solubility knowledge, extensive sample preparation, and their time-consuming and destructive nature. This study investigates the application of laser-induced breakdown spectroscopy (LIBS) as an alternative method. It involves probing concrete samples with high spatial resolution and analyzing the resultant spectra. The methodology is first tested on mesoscale concrete models to assess limitations and inherent errors. Subsequently, the methodology is applied to actual concrete samples with varying cement content and aggregate size distributions. The results demonstrate a promising accuracy, with an average relative error of approximately 8%. This paper offers a comprehensive evaluation of the method's advantages, limitations, and factors influencing its practical applicability in field conditions. KW - LIBS KW - Spectroscopy KW - Cement content KW - Elemental mapping PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-620112 DO - https://doi.org/10.1016/j.cemconres.2024.107714 SN - 1873-3948 VL - 189 SP - 1 EP - 10 PB - Elsevier Ltd. AN - OPUS4-62011 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Saeidfirozeh, H. A1 - Kubelík, P. A1 - Laitl, V. A1 - Krivkova, A. A1 - Vrabel, J. A1 - Rammelkamp, K. A1 - Schroder, S. A1 - Gornushkin, Igor B. A1 - Kepes, E. A1 - Zabka, J. A1 - Ferus, M. A1 - Porízka, P. A1 - Kaiser, J. T1 - Laser-induced breakdown spectroscopy in space applications: Review and prospects N2 - This review describes the principles and summarizes the challenges of analytical methods based on optical emission spectroscopy (OES) in space applications, with a particular focus on Laser-Induced Breakdown Spectroscopy (LIBS). Over the past decade, LIBS has emerged as a powerful analytical technique for space exploration and In-Situ Resource Utilization (ISRU) of celestial bodies. Its implementation has been suggested for various segments of the Space Resources Value Chain, including prospecting, mining, and beneficiation. Current missions to Mars, including the ChemCam instrument on the Curiosity rover, the SuperCam on the Perseverance rover, and the MarSCoDe on the Zhurong rover, are considered flagship applications of LIBS. Despite neither the Pragyan rover nor the Vikram lander waking from the lunar night, the success of the Chandrayaan-3 mission marks another milestone in the development of LIBS instruments, with further missions, including commercial ones, anticipated. This paper reviews the deployment of LIBS payloads on Mars rovers, upcoming missions prospecting the Moon and asteroids, and LIBS analysis of meteorites. Additionally, it highlights the importance of data processing specific to space applications, emphasizing recent trends in transfer learning. Furthermore, LIBS combined with other spectroscopic techniques (e.g., Raman Spectroscopy, Mass Spectrometry, and Fourier-Transform Infrared Spectroscopy) represents an intriguing platform with comprehensive analytical capabilities. The review concludes by emphasizing the significance of LIBS-based contributions in advancing our understanding of celestial bodies and paving the way for future space exploration endeavors KW - Laser-induced breakdown spectroscopy KW - Mars KW - Moon KW - Asteroids KW - Meteorites KW - Machine learning KW - Transfer learning KW - Raman spectroscopy KW - Mass spectrometry KW - Fourier-transform infrared spectroscopy PY - 2024 DO - https://doi.org/10.1016/j.trac.2024.117991 VL - 181 SP - 1 EP - 22 PB - Elsevier B.V. AN - OPUS4-62142 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Westwood, Steven A1 - Lippa, Katrice A1 - Shimuzu, Yoshitaka A1 - Lalerle, Beatrice A1 - Saito, Takeshi A1 - Duewer, David A1 - Dai, Xinhua A1 - Davies, Stephen A1 - Ricci, Marina A1 - Baldan, Annarita A1 - Lang, Brian A1 - Sarge, Stefan A1 - Wang, Haifeng A1 - Pratt, Ken A1 - Josephs, Ralf A1 - Mariassy, Mikael A1 - Pfeifer, Dietmar A1 - Warren, John A1 - Bremser, Wolfram A1 - Ellison, Stephen A1 - Toman, Blaza A1 - Nelson, Michael A1 - Huang, Ting A1 - Fajgelj, Ales A1 - Gören, Ahmet A1 - Mackay, Lindsey A1 - Wielgosz, Robert T1 - Methods for the SI-traceable value assignment of the purity of organic compounds (IUPAC Technical Report) N2 - The “purity” of an organic compound typically refers, in practice, to an assignment of the mass fraction content of the primary organic component present in the material. The “purity” value of an organic primary calibrator material is the ultimate source of metrological traceability of any quantitative measurement of the content of that compound in a given matrix. The primary calibrator may consist of a Certified Reference Material (CRM) whose purity has been assigned by the CRM producer or a laboratory may choose to value-assign a material to the extent necessary for their intended application by using appropriately valid methods. This report provides an overview of the approach, performance and applicability of the principal methods used to determine organic purity including mass balance, quantitative NMR, thermal methods and direct-assay techniques. A statistical section reviews best practice for combination of data, value assignment as the upper limit values corresponding to 100 % purity are approached and how to report and propagate the standard uncertainty associated with the assigned values. KW - Calibration hierarchy KW - Certified reference materials KW - Mass balance KW - Measurement uncertainty KW - Organic purity KW - Primary calibrators KW - Primary reference materials KW - qNMR KW - Related-structure impurities KW - Thermal analysis KW - Titrimetry KW - Traceability KW - Uncertainty propagation KW - Volatile organic compounds KW - Water content PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-627932 DO - https://doi.org/10.1515/pac-2020-0804 SN - 0033-4545 SN - 1365-3075 VL - 95 IS - 1 SP - 1 EP - 77 PB - de Gruyter CY - Berlin AN - OPUS4-62793 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Sparber‐Sauer, Monika A1 - Ferrari, Andrea A1 - Kosztyla, Daniel A1 - Ladenstein, Ruth A1 - Cecchetto, Giovanni A1 - Kazanowska, Bernarda A1 - Scarzello, Giovanni A1 - Ljungman, Gustaf A1 - Milano, Giuseppe Maria A1 - Niggli, Felix A1 - Alaggio, Rita A1 - Vokuhl, Christian A1 - Casanova, Michela A1 - Klingebiel, Thomas A1 - Zin, Angelica A1 - Koscielniak, Ewa A1 - Bisogno, Gianni T1 - Long‐term results from the multicentric European randomized phase 3 trial CWS/RMS‐96 for localized high‐risk soft tissue sarcoma in children, adolescents, and young adults N2 - Background: CWS/RMS‐96 was an international multicenter trial with randomization between two therapy arms of the standard four‐drug therapy (vincristine, ifosfamide, adriamycin, dactinomycin [VAIA]) versus an intensified six‐drug regimen (carboplatin, epirubicin, vincristine, dactinomycin, ifosfamide, and etoposide [CEVAIE]) for high‐risk rhabdomyosarcoma (RMS), extraskeletal Ewing sarcoma (EES), and undifferentiated sarcoma (UDS) in children, adolescents, and young adults aiming to improve their survival. Intensified chemotherapy with CEVAIE did not improve outcome. Methods: Patients younger than 21 years with a previously untreated localized HR‐RMS, EES, and UDS were enrolled from Cooperative Weichteilsarkom Studiengruppe (CWS) centers in Germany, Austria, Poland, Switzerland, and from Italian Soft Tissue Sarcoma Committee (STSC) centers. Randomization (1:1) to receive either 9 × 21 days cycles of VAIA or CEVAIE was performed separately in CWS and STSC. Hyperfractionated accelerated radiotherapy (32–44.8 Gy) was added at week 9–12 according to histology and response to chemotherapy. A secondary microscopically complete nonmutilating resection was performed if possible. Primary endpoints were response to chemotherapy, event‐free (EFS) and overall survival (OS). Results: Five hundred fifty‐seven patients (HR‐RMS:n = 416, EES and UDS:n = 141) underwent randomization: VAIA (n = 273) or CEVAIE (n = 284). Radiotherapy was given to 70% of patients in both groups. A secondary resection was performed in 47% and 48% patients, respectively. The 5‐year EFS and OS for the VAIA and CEVAIE treatment arms were 59.8% and 60.8% (p = .89), and 74.2% and 68.3% (p = .16), respectively. No differences in response, toxicity, or second malignancies emerged in the two groups. Conclusion: The use of an intensified regimen failed to show a significant improvement in tumor response and outcome of patients with localized HR‐RMS, EES, and UDS. KW - CEVAIE KW - CWS-96 KW - High-risk soft tissue sarcoma KW - Randomization KW - Rhabdomyosarcoma KW - RMS-96 KW - VAIA PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-629373 DO - https://doi.org/10.1002/pbc.29691 SN - 1545-5017 VL - 69 IS - 9 SP - 1 EP - 11 PB - Wiley CY - New York, NY AN - OPUS4-62937 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Capozzi, Staci L. A1 - Xia, Chunjie A1 - Shuwal, Matthew A1 - Zaharias Miller, Gillian A1 - Gearhart, Jeff A1 - Bloom, Erica A1 - Gehrenkemper, Lennart A1 - Venier, Marta T1 - From watersheds to dinner plates: Evaluating PFAS exposure through fish consumption in Southeast Michigan N2 - Muscle tissue and organ samples of six different fish species were collected from ten locations in Southeast Michigan’s Huron and Rouge watersheds. Per- and polyfluoroalkyl substances (PFAS) were analyzed in 36 samples comprising filets, liver, gut, and eggs using targeted analysis and the direct total oxidizable precursor (dTOP) assay on a subset of six samples. The median concentrations of the ΣPFAS in filets from the Huron and Rouge watersheds were 13 and 6.3 ng/g wet weight (w.w.), respectively. Perfluorooctane sulfonate (PFOS) was the most detected and abundant compound in fish organs, with the liver having the largest overall burden of PFAS. The highest percent increase in targeted PFAS after the dTOP assay was observed in the Catfish filet (552%) while the smallest increase was in the Catfish liver (32%) accounting for 1.3 and 8.1 nMole F/g dry weight (d.w.), respectively. The positive matrix factorization (PMF) analysis revealed three distinct PFAS sources, of which the one attributed to PFOS explained 73% of the data. Results from this work have important implications for fish consumption in Michigan waterways. Among the filet samples analyzed, the calculated daily consumption limit of total PFOS was exceeded in approximately 82% and 91% of samples for adults and children over the age of seven years old, respectively. KW - Direct total oxidizable precursor (dTOP) assay KW - Huron river watershed KW - Positive matrix factorization (PMF) KW - Rouge river watershed KW - Consumption limit PY - 2023 DO - https://doi.org/10.1016/j.chemosphere.2023.140454 SN - 1879-1298 VL - 345 SP - 1 EP - 9 PB - Elsevier CY - Amsterdam AN - OPUS4-62423 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Zöllner, Moritz T. A1 - Dariz, Petra A1 - Riedel, Jens A1 - Schmid, Thomas T1 - Dolomite and Mg Calcite as Mineral Thermometers in Mortar Binders. A High Resolution Raman Spectroscopic Study N2 - This paper suggests the use of high‐resolution Raman scattering bands of MgCa carbonates as posteriori thermometer minerals in archaeometric studies. Therefore, the thermal behavior of two dolomite samples and the hydration and carbonation reaction in air of the decomposition products were investigated by Raman microspectroscopy. The increase in the calcination temperature resulted in the formation of – Raman silent MgO and – inert Mg calcite at 700°C–750°C. In contrast, the decarbonation, hydration, and recarbonation of sample material exposed to 750°C–900°C in a muffle furnace led to the appearance of Mg‐free calcite. High spectral resolution Raman spectroscopy enabled a spectral distinction between these two groups due to differences in the band parameters (peak position, bandwidth) of the vibrational (v1, v4, L) modes of calcite. In combination with Raman microspectroscopic mapping, this spectral information represents a new approach for the estimation of burning temperatures of medieval high‐fired gypsum mortars via natural dolomite impurities. Thus, the results of this work highlight the importance and potential of Raman microspectroscopy as a thermometric tool for elucidating the thermal history of anthropogenic fired materials, with potential applications for archaeometry and art technology, as well as for quality controls in the frame of the production of mineral mortar binders and ceramics or bricks, respectively. KW - Mineral thermometry KW - Raman spectroscopy KW - Dolomite KW - High-fired gypsum mortar PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-630270 DO - https://doi.org/10.1002/jrs.6810 SN - 1097-4555 SP - 1 EP - 13 PB - Wiley AN - OPUS4-63027 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bühling, Benjamin A1 - Maack, Stefan A1 - Strangfeld, Christoph T1 - Fluidic Ultrasound Generation for Non‐Destructive Testing N2 - AbstractAir‐coupled ultrasonic testing (ACU) is a pioneering technique in non‐destructive testing (NDT). While contact testing and fluid immersion testing are standard methods in many applications, the adoption of ACU is progressing slowly, especially in the low ultrasonic frequency range. A main reason for this development is the difficulty of generating high amplitude ultrasonic bursts with equipment that is robust enough to be applied outside a laboratory environment. This paper presents the fluidic ultrasonic transducer as a solution to this challenge. This novel aeroacoustic source uses the flow instability of a sonic jet in a bistable fluidic switch to generate ultrasonic bursts up to 60 kHz with a mean peak pressure of 320 Pa. The robust design allows operation in adverse environments, independent of the operating fluid. Non‐contact through‐transmission experiments are conducted on four materials and compared with the results of conventional transducers. For the first time, it is shown that the novel fluidic ultrasonic transducer provides a suitable acoustic signal for NDT tasks and has potential of furthering the implementation of ACU in industrial applications.This article is protected by copyright. All rights reserved KW - Aeroacoustics KW - Air-coupled ultrasound KW - Fluidics KW - Harsh environment KW - Laser Doppler vibrometer KW - Non-destructive testing PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-594273 DO - https://doi.org/10.1002/adma.202311724 SN - 0935-9648 SP - 1 EP - 14 PB - Wiley AN - OPUS4-59427 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Capitain, C. A1 - Schüßler, M. A1 - Bugsel, B. A1 - Zweigle, J. A1 - Vogel, Christian A1 - Leube, Peter A1 - Zwiener, C. T1 - Implementation of matrix-matched semiquantification of PFAS in AFFF-contaminated soil N2 - This study presents a novel semiquantification approach for nontarget screening (NTS), combining matrix-matched calibration and ionization class-specific average calibration curves (ACCs) to address the lack of analytical reference standards for most per- and polyfluoroalkyl substances (PFAS). Ionization class-specific ACCs for carboxylic and sulfonic acids, sulfonamides, and cationic PFAS result in high accuracy, with median absolute accuracy quotients below 2.27×. The approach was applied to soil impacted by aqueous film-forming foam (AFFF) contamination. A total of 96 tentatively identified PFAS were semiquantified in addition to 28 quantified compounds based on available standards. Semiquantified concentrations exceeded those of target analytes, demonstrating the critical role of this method in capturing broader PFAS contamination. In this case, validation against extractable organofluorine (EOF) showed a 102% closed mass balance. The innovative approach not only enables comprehensive PFAS contamination assessment in complex matrices but also expands the scope of the NTS for environmental monitoring, remediation, and risk assessment of AFFF-contaminated sites. KW - Per- and polyfluoroalkyl substances (PFAS) KW - Contaminated soil KW - Non-targeted analysis KW - Combustion ion chromatography PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-629445 DO - https://doi.org/10.1021/acs.est.4c14255 SN - 1520-5851 VL - 59 IS - 14 SP - 7338 EP - 7347 PB - American Chemical Society (ACS) CY - Columbus, Ohio AN - OPUS4-62944 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Mugani, R. A1 - El Khalloufi, F. A1 - Kasada, M. A1 - Redouane, E. M. A1 - Haida, M. A1 - Aba, R. P. A1 - Essadki, Y. A1 - Zerrifi, S. E. A. A1 - Herter, Sven-Oliver A1 - Hejjaj, A. A1 - Aziz, F. A1 - Ouazzani, N. A1 - Azevedo, J. A1 - Campos, A. A1 - Putschew, A. A1 - Grossart, H.-P. A1 - Mandi, L. A1 - Vasconcelos, V. A1 - Oudra, B. T1 - Monitoring of toxic cyanobacterial blooms in Lalla Takerkoust reservoir by satellite imagery and microcystin transfer to surrounding farms N2 - Cyanobacterial harmful algal blooms (CyanoHABs) threaten public health and freshwater ecosystems worldwide. In this study, our main goal was to explore the dynamics of cyanobacterial blooms and how microcystins (MCs) move from the Lalla Takerkoust reservoir to the nearby farms. We used Landsat imagery, molecular analysis, collecting and analyzing physicochemical data, and assessing toxins using HPLC. Our investigation identified two cyanobacterial species responsible for the blooms: Microcystis sp. and Synechococcus sp. Our Microcystis strain produced three MC variants (MC-RR, MC-YR, and MC-LR), with MC-RR exhibiting the highest concentrations in dissolved and intracellular toxins. In contrast, our Synechococcus strain did not produce any detectable toxins. To validate our Normalized Difference Vegetation Index (NDVI) results, we utilized limnological data, including algal cell counts, and quantified MCs in freeze-dried Microcystis bloom samples collected from the reservoir. Our study revealed patterns and trends in cyanobacterial proliferation in the reservoir over 30 years and presented a historical map of the area of cyanobacterial infestation using the NDVI method. The study found that MC-LR accumulates near the water surface due to the buoyancy of Microcystis. The maximum concentration of MC-LR in the reservoir water was 160 μg/L. In contrast, 4 km downstream of the reservoir, the concentration decreased by a factor of 5.39 to 29.63 μg/L, indicating a decrease in MC-LR concentration with increasing distance from the bloom source. Similarly, the MC-YR concentration decreased by a factor of 2.98 for the same distance. Interestingly, the MC distribution varied with depth, with MC-LR dominating at the water surface and MC-YR at the reservoir outlet at a water depth of 10 m. Our findings highlight the impact of nutrient concentrations, environmental factors, and transfer processes on bloom dynamics and MC distribution. We emphasize the need for effective management strategies to minimize toxin transfer and ensure public health and safety. KW - Cyanotoxin KW - HPLC-MS/MS KW - Surface water PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-623672 DO - https://doi.org/10.1016/j.hal.2024.102631 SN - 1568-9883 VL - 135 SP - 1 EP - 14 PB - Elsevier CY - Amsterdam AN - OPUS4-62367 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Isleyen, Alper A1 - Özcan, Kemal A1 - Tunc, Murat A1 - Boztepe, Aylin A1 - Coşkun, Fatma Gonca A1 - Moshammer, Kai A1 - Shehab, Moaaz A1 - Stratulat, Camelia A1 - Bratu, Adriana A1 - Hafner-Vuk, Katarina A1 - Vogl, Jochen A1 - Strzelec, Michał A1 - Calvo, Mariana Villegas A1 - Frey, Anne Mette A1 - Strauss, Helena T1 - Development of three biofuel CRMs for the quality parameters in biodiesel and wood pellet via a joint research project N2 - Biomass is a key element in biofuels which can be defined as a fuel produced through contemporary biological processes, and its increased use can support the EU’s aims of reducing greenhouse gas emissions. Information on the nature and the quality of the biomass or biofuel is important in order to support the optimization of their combustion with respect to realizing higher efficiencies and lower emissions during energy production. Three reference materials were produced by a collaborative approach among national metrology institutes and designated institutes within the scope of the EMPIR project: BIOFMET. The project was aimed to establish advanced traceable measurement standards for the determination of the calorific value, impurities, and other parameters such as density, kinematic viscosity, moisture, and ash. This paper presents the sampling and processing methodology, homogeneity, stability, characterization campaign, the assignment of property values, and their associated uncertainties in compliance with ISO 17034 for biofuel reference materials: biodiesel, wood pellet powder, and wood pellet. Parameters of interest in biodiesel reference material-UME BIOFMET CRM 01 are gross calorific value (GCV), density, viscosity, and mass fractions of Ca, K, Mg, Na, P, and S elements. Parameters to be certified in wood pellet powder reference material-UME BIOFMET CRM 02 are GCV, moisture, ash, and mass fractions of Al, Cr, K, Mg, Mn, Ni, S, and Zn elements. Parameters to be certified in the wood pellet reference material-UME BIOFMET CRM 03 are GCV and moisture. The homogeneity and stability of the materials were assessed in accordance with ISO 33405. The materials were characterized by interlaboratory comparison studies among competent metrology institute and designated institute laboratories. Assigned values and uncertainties of the certified values were calculated in accordance with ISO 33405, and uncertainties include characterization, homogeneity, and stability components. The developed CRMs are intended to be used for the development and validation of measurement procedures for the determination and quality control/assurance purposes of the quality parameters for biofuels. It should be emphasized that the UME BIOFMET CRM 01-Biodiesel CRM is the first biodiesel reference material certified for calorific value. Among the developed wood CRMs, the pellet form, UME BIOFMET CRM 03, was found to be more stable than the powder one, UME BIOFMET CRM 02, for the moisture parameter. Sixfold lower relative uncertainty value for short-term stability at 45 °C and twofold lower relative uncertainty value for long-term stability at 22 °C were obtained for the moisture parameter of the CRM in pellet form compared to the CRM in powder form. KW - Biofuel KW - Biodiesel KW - Wood pellet KW - Calorific value KW - Moisture KW - Element content PY - 2024 DO - https://doi.org/10.1007/s00216-024-05694-y SP - 1 EP - 13 PB - Springer Science and Business Media LLC AN - OPUS4-62371 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fiedler, Saskia A1 - Ton-That, C. A1 - Phillips, M. R. T1 - Defect-free ZnO nanorods with high angular distribution for enhanced excitonic emission N2 - AbstractLow-temperature hydrothermal growth has emerged as a popular method for the fabrication of ZnO nanorods (NRs), increasing the functionality and utility of ZnO-based devices. In this work, we study the influence of growth time, temperature and seed layer on the dimensions and angular distribution of ZnO NRs. High-quality NRs with a crisscrossed 60° angular distribution have been grown with a 20–60 nm diameter and 600 nm length. We show that, within the ideal range of growth parameters, the growth time and temperature have no controllable influence on NR diameter and length, while the deposition method and size of the pre-growth deposited ZnO seeds affects diameter and NR angular alignment. We demonstrate advantages of using crisscross-aligned NRs over planar ZnO for the enhancement of ZnO excitonic emission by optical coupling with gold nanoparticles. These results can be readily adapted for applications that involve surface coating-mediated enhancement of both light emission and injection. KW - ZnO nanorod KW - Hydrothermal growth KW - Seed layer KW - Angular distribution KW - Cathodoluminescence KW - Luminescence enhancement PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-618176 DO - https://doi.org/10.1557/s43578-023-00941-x SN - 0884-2914 SN - 2044-5326 VL - 38 IS - 8 SP - 2145 EP - 2155 PB - Springer CY - Berlin AN - OPUS4-61817 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schardt, Annika A1 - Schmitt, Johannes A1 - Engelhard, Carsten T1 - Single particle inductively coupled plasma mass spectrometry with nanosecond time resolution N2 - In this proof-of-principle study, we present our contribution to single particle inductively coupled plasma mass spectrometry (spICP-MS) developments with a novel in-house built data acquisition system with nanosecond time resolution (nanoDAQ) and a matching data processing approach. The new system can continuously sample the secondary electron multiplier (SEM) detector signal and enables the detection of gold nanoparticles (AuNP) as small as 7.5 nm with the commercial single quadrupole ICP-MS instrument used in this study. Recording of the SEM signal by the nanoDAQ is performed with a dwell time of approximately 4 ns. A tailored method was developed to process this type of transient data, which is based on determining the temporal distance between detector events that is denoted as event gap (EG). We found that the inverse logarithm of EG is proportional to the particle size and that the number of detector events corresponding to a particle signal distribution can be used to calibrate and determine the particle number concentration (PNC) of a nanoparticle dispersion. Due to the high data acquisition frequency, a statistically significant number of data points can be obtained in 60 s or less and the main time limitation for analyses is merely the sample uptake time and rinsing step between analyte solutions. At this stage, the data processing method provides average information on complete data sets only and will be adapted to enable particle-by-particle analysis with future hardware/software revision. KW - ICP-MS KW - Nanoparticles KW - Nanosecond time resolution KW - Single particle detection PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-612706 DO - https://doi.org/10.1039/d3ja00373f SN - 1364-5544 SN - 0267-9477 VL - 39 IS - 2 SP - 389 EP - 400 PB - Royal Society of Chemistry (RSC) CY - Cambridge AN - OPUS4-61270 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Willms, Thomas A1 - Kryk, Holger A1 - Hampel, Uwe T1 - An improved method for the analysis of the reaction products of the partial isobutane oxidation by gas chromatography using a green solvent N2 - In the present work, an improved GC/MS method for the analysis of the reaction products of the non-catalytic partial oxidation of isobutane at a starting temperature of 25°C has been developed and compared to a previously published method. Two green GC-solvents (dimethyl carbonate (DMC), diethyl carbonate (DEC)) and several internal standards (hexane, benzene, toluene, DEC) have been tested. The separation of the products at low retention times (methanol, isobutane, isobutene) could be significantly improved. For the liquid main products t-butyl hydroperoxide (TBHP), di-t-butyl peroxide (DTBP), t-butanol (TBA), propanone and most trace products (isopropanol, isobutanal, methyl formate, isopropyl formate, t-butyl formate and some acetates – 24 compounds in total), a coefficient of determination CoD > 0.999 has been achieved. For all weighing and GC/MS measurements uncertainties have been calculated. The peaks of the peak pair t-butyl formate (CoD >0.999) and isobutanol (CoD >0.99), have been simulated and used for calibrations. DEC gave a decomposition of TBHP on the column. With DMC, a problem of peak purity of TBHP appeared but could be resolved. The use of the green solvent DMC, compared to also studied traditional solvents like heptane and toluene, proved to be very beneficial, since it also permitted to determine polar compounds, e. g. carbon dioxide and formic acid. KW - GC/MS KW - Green solvent KW - Method development KW - Peroxide KW - Measurement uncertainties KW - Oxidation products KW - Simulation PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-632026 DO - https://doi.org/10.1016/j.chroma.2025.465892 SN - 0021-9673 VL - 1753 SP - 1 EP - 16 PB - Elsevier CY - Amsterdam AN - OPUS4-63202 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Chazeau, E. A1 - Pipier, A. A1 - Wegner, Karl David A1 - Ghiringhelli, F. A1 - Sancey, L. A1 - Paul, C. A1 - Goze, C. T1 - NIR-II aza-BODIPY Platform for the Development of a Fluorescent Antibody Drug Conjugate N2 - Real-time imaging of antibody-drug conjugates (ADCs) offers valuable insights for assessing tumor targeting specificity, monitoring therapeutic efficacy, and detecting off-target accumulation that may cause adverse effects. To enable precise tracking, we developed a versatile fluorescent platform based on an NIR-II emitting aza-BODIPY dye, which can be site-specifically grafted onto an IgG1 antibody to generate well-defined fluorescent ADCs. As a proof of concept, we synthesized an HER2-targeting trastuzumab immunoconjugate bearing a NIR-II aza-BODIPY fluorophore. The cytotoxic monomethyl auristatin E (MMAE) payload was introduced in the final step, resulting in a trackable and homogeneous ADC suitable for both in vitro and in vivo investigations. The resulting Trastu-azaNIRII-MMAE selectively accumulated in HER2-positive subcutaneous tumors, significantly reducing the tumor growth. Using NIR-II optical imaging, a single injection of the NIR-II-ADC allowed for the detection of the conjugate over a period of more than one month, highlighting its potential for long-term tracking and therapeutic applications. KW - NIR-II KW - Fluorescence KW - Quality assurance KW - Antibody drug conjugate KW - In vivo imaging PY - 2025 DO - https://doi.org/10.1021/acs.jmedchem.4c02777 VL - 68 IS - 7 SP - 7232 EP - 7242 PB - ACS Publications AN - OPUS4-63025 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Müller, Axel A1 - Altmann, Korinna A1 - Kocher, B. A1 - Braun, U. T1 - Determination of tire wear markers in soil samples and their distribution in a roadside soil N2 - Tire wear (TW) constitutes a significant source of microplastic in terrestrial ecosystems. It is known that particles emitted by roads can have an effect up to 100 m into adjacent areas. Here, we apply for the first-time thermal extraction desorption gas chromatography-mass spectrometry (TED-GC/MS) to determine TW in soil samples by detection of thermal decomposition products of styrene-butadiene rubber (SBR), without additional enrichment. Additionally, zinc contents were determined as an elemental marker for TW. Mixed soil samples were taken along three transects along a German motorway in 0.3, 2.0, and 5.0 m distance from the road. Sampling depths were 0–2, 2–5, 5–10, and 10–20 cm. Four fine fractions, 1 000–500, 500–100, 100–50, and <50 μm, were analyzed. TW contents based on SBR ranged from 155 to 15 898 mg kg−1. TW contents based on zinc were between 413 and 44 812 mg kg−1. Comparison of individual values of SBR and zinc reveals SBR as a more specific marker. Results confirm that most TW ends up in the topsoil within a 2 m distance. The sampling strategy resulted in representative data for a larger area. Standard deviations of quadruple TED-GC/MS determination of SBR were <10% for all grain size fractions. TED-GC/MS is a suitable analytical tool for determining TW in soil samples without the use of toxic chemicals, enrichment, or special sample preparation. KW - Microplastic KW - TED-GC/MS KW - Tire wear PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-543344 DO - https://doi.org/10.1016/j.chemosphere.2022.133653 VL - 294 SP - 1 EP - 8 PB - Elsevier Ltd. AN - OPUS4-54334 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogel, Christian A1 - Roesch, Philipp A1 - Wittwer, Philipp A1 - Sommerfeld, Thomas A1 - Riedel, Maren A1 - Leube, Peter A1 - Kalbe, Ute A1 - Schoknecht, Ute A1 - Simon, Franz-Georg T1 - Per- and Polyfluoroalkyl Substances (PFAS) in Ski Waxes and Snow from Cross-Country Skiing in Germany - Comparative study of Sum Parameter and Target Analysis N2 - Per- and polyfluoroalkyl substances (PFAS) are often environmentally exposed via discharge through human consumer products, such as ski waxes. In our study we analyzed various ski waxes from the 1980s and 2020s, to determine both the sum parameter values total fluorine (TF), extractable organically bound fluorine (EOF), hydrolysable organically bound fluorine (HOF) as well as targeted PFAS analysis. This showed that modern high-performance waxes contain up to 6 % TF, but also PFAS-free labelled ski waxes contain traces of PFAS with EOF/HOF values in the low mg kg-1 range. With the ban of all fluorine-based waxes with the start of the 2023/2024 winter season this will probably change soon. Moreover, we applied our analysis methods to snow samples from a frequently used cross country ski trail (Kammloipe) in the Ore Mountain region in Germany, assessing the potential PFAS entry/discharge through ski waxes. Melted snow samples from different spots were analyzed by the adsorbable organically bound fluorine (AOF) sum parameter and PFAS target analysis and confirmed the abrasion of the ski waxes into the snow. Moreover, on a PFAS hotspot also soil samples were analyzed, which indicate that PFAS from the ski waxes adsorb after snow melting into the soil. KW - Combustion Ion Chromatography KW - Per- and Polyfluoroalkyl substances (PFAS) KW - Consumer Products PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-612760 DO - https://doi.org/10.1016/j.hazadv.2024.100484 VL - 16 SP - 1 EP - 5 PB - Elsevier B.V. AN - OPUS4-61276 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogel, Christian A1 - Scholz, Philipp A1 - Kalbe, Ute A1 - Caliebe, W. A1 - Tayal, A. A1 - Vasala, S. J. A1 - Simon, Franz-Georg T1 - Speciation of antimony and vanadium in municipal solid waste incineration ashes analyzed by XANES spectroscopy N2 - The use of ashes from municipal solid waste incineration as secondary building materials is an important pillar for the circular economy in Germany. However, leaching of potential toxic elements from these materials must be at environmentally acceptable levels. Normally, a three-month ageing period immobilizes most hazardous heavy metals, but antimony (Sb) and vanadium (V) showed previously unusual leaching. In order to clarify the mechanisms, we analyzed the Sb and V species in various bottom and fly ashes from municipal waste incineration by XANES spectroscopy. Antimony oxidizes from Sb(+ III) species used as flame retardants in plastics to Sb(+ V) compounds during waste incineration. However, owing to the similarity of different Sb(+ V) compound in the Sb K- and L-edge XANES spectra, it was not possible to accurately identify an exact Sb(+ V) species. Moreover, V is mainly present as oxidation state + V compound in the analyzed ashes. However, the coarse and magnetic fraction of the bottom ashes contain larger amounts of V(+ III) and V(+ IV) compounds which might enter the waste incineration from vanadium carbide containing steel tools. Thus, Sb and V could be critical potential toxic elements in secondary building materials and long-term monitoring of the release should be taken into account in the future. KW - Müllverbrennung KW - XANES spectroscopy PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-599841 DO - https://doi.org/10.1007/s10163-024-01954-2 SN - 1438-4957 VL - 26 SP - 2152 EP - 72158 PB - Springer Science and Business Media LLC AN - OPUS4-59984 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dittmar, Stefan A1 - Weyrauch, Steffen A1 - Reemtsma, Thorsten A1 - Eisentraut, Paul A1 - Altmann, Korinna A1 - Ruhl, Aki S. A1 - Jekel, Martin T1 - Settling Velocities of Tire and Road Wear Particles: Analyzing Finely Graded Density Fractions of Samples from a Road Simulator and a Highway Tunnel N2 - The terminal settling velocity is considered the most critical parameter determining the transport of tire and road wear particles (TRWP) in aquatic environments. Nonetheless, no respective empirical data has been reported so far. In this study, particle samples from a road simulator and a highway tunnel were investigated with a validated imaging method. Different density and size fractions of both samples were measured separately, acquiring sizes and settling velocities of more than 30,000 individual particles. In addition, tire marker polymers were analyzed for each fraction via thermal extraction desorption-gas chromatography/mass spectrometry. Finally, the acquired particle data was combined according to the fractions’ estimated tire contents in order to deduce detailed probability distributions of particle size and settling velocity for the actual TRWP from both samples. Weighted by TRWP-incorporated tire mass, median diameters of 54 and 44 μm as well as median settling velocities of 0.65 and 0.22 mm/s were found for TRWP from the road simulator and highway tunnel, respectively. This study thus provides the first ever empirical data on TRWP settling velocities in water, which can be highly valuable input for modeling the environmental transport of TRWP and for dimensioning TRWP retention systems. KW - Microplastics KW - TED-GC/MS KW - Sinking velocities KW - Transport PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-635449 DO - https://doi.org/10.1021/acs.est.5c04165 SN - 0013-936X VL - 59 IS - 26 SP - 13434 EP - 13446 PB - American Chemical Society (ACS) AN - OPUS4-63544 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Visileanu, Emilia A1 - Altmann, Korinna A1 - Stepa, Raluca A1 - Haiducu, Maria A1 - Miclea, Paul Tiberiu A1 - Vladu, Alina A1 - Dondea, Felicia A1 - Grosu, Marian Catalin A1 - Scarlat, Razvan T1 - Comparative Analysis of Airborne Particle Concentrations in Textile Industry Environments Throughout the Workday N2 - This paper addresses the growing concern surrounding microplastic pollution, particularly within the textile industry, and the associated potential health risks linked to the inhalation and ingestion of microplastic particles. Microplastics, defined as plastic particles smaller than five millimeters, are increasingly found not only in aquatic environments, but also in soils, air, and food. Although research on the health impacts of microplastics is still emerging, early studies indicate that these particles could contribute to health issues, including oxidative stress, inflammation, and cardiovascular diseases. Notably, individuals with higher concentrations of plastics in arterial plaques are more susceptible to heart attacks and strokes. In the textile industry, synthetic fibers such as polyester, nylon, and acrylic release microplastics into the air during production. The paper discusses a study conducted in a textile company that processes polyester yarns, where airborne microplastic concentrations were measured at various locations throughout the day. Particle sizes ranging from 0.3 nm to 10 nm were analyzed, revealing the presence of polyester polymers in the particulate matter. These findings underscore the widespread nature of microplastic pollution, particularly in industrial settings, and raise concerns about the health risks associated with prolonged exposure to airborne microplastics. While further research is necessary to fully understand the extent of these health impacts, preliminary data suggest a troubling link between microplastic inhalation and cardiovascular conditions. KW - Microplastics KW - Textiles KW - Airborne particles PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-635269 DO - https://doi.org/10.3390/microplastics4020034 SN - 2673-8929 VL - 4 IS - 2 SP - 1 EP - 17 PB - MDPI AG AN - OPUS4-63526 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kalbe, Ute A1 - Piechotta, Christian A1 - Bandow, Nicole T1 - Comparing PFAS analysis in batch leaching and column leaching tests N2 - Laboratory leaching tests are tools to assess the mobility of environmental contaminants released from granular materials. Comparative leaching tests were performed using four PFAS-contaminated soils whose concentration patterns of 10 selected perfluoroalkyl and polyfluoroalkyl substances (PFAS) differed due to the two types of contamination sources. This study aimed to evaluate the equivalence of two usual laboratory-scale leaching test procedures, batch and column percolation tests, at liquid-to-solid ratios (L/S) of 2 l/kg, which is the current practice within the German assessment framework, and 10 l/kg (relevant for some EU regulations such as the landfill directive). The differences between the replicates of leaching tests investigating PFAS were smaller for column percolation tests than for batch tests, probably mainly due to the greater sample size and the better representativeness of the sample portion analyzed. It was observed that batch tests overestimate the release of shorter-chain PFAS, whereby the effect was greater with carboxylic than with sulfonic acids. Currently, the limits of detection of analyses given by the DIN standard with regard to PFCA and PFSA in soils are partly not sufficient to detect very low contents, whereas the detection of selected PFCA and PFSA in eluates is more sensitive, in accordance with the available standards. This results in limitations when calculating mass balances. KW - PFAS KW - Leaching KW - Column percolation test KW - Batch test KW - Soil PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-618187 DO - https://doi.org/10.1007/s11356-024-35510-0 SN - 1614-7499 VL - 31 IS - 57 SP - 65233 EP - 65251 PB - Springer Science and Business Media LLC AN - OPUS4-61818 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fricke, F. A1 - Brandalero, M. A1 - Liehr, Sascha A1 - Kern, Simon A1 - Meyer, Klas A1 - Kowarik, Stefan A1 - Hierzegger, R. A1 - Westerdick, S. A1 - Maiwald, Michael A1 - Hübner, M. T1 - Artificial Intelligence for Mass Spectrometry and Nuclear Magnetic Resonance Spectroscopy Using a Novel Data Augmentation Method N2 - Mass Spectrometry (MS) and Nuclear Magnetic Resonance Spectroscopy (NMR) are valuable analytical and quality control methods for most industrial chemical processes as they provide information on the concentrations of individual compounds and by-products. These processes are traditionally carried out manually and by a specialist, which takes a substantial amount of time and prevents their utilization for real-time closed-loop process control. This paper presents recent advances from two projects that use Artificial Neural Networks (ANNs) to address the challenges of automation and performance-efficient realizations of MS and NMR. In the first part, a complete toolchain has been realized to develop simulated spectra and train ANNs to identify compounds in MS. In the second part, a limited number of experimental NMR spectra have been augmented by simulated spectra, to train an ANN with better prediction performance and speed than state-of-the-art analysis. These results suggest that, in the context of the digital transformation of the process industry, we are now on the threshold of a strongly simplified use of MS and MRS and the accompanying data evaluation by machine-supported procedures, and can utilize both methods much wider for reaction and process monitoring or quality control. KW - Industry 4.0 KW - Cyber-Physical Systems KW - Artificial Neural Networks KW - Mass Spectrometry KW - Nuclear Magnetic Resonance Spectroscopy KW - Modular Production PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-539412 UR - https://ieeexplore.ieee.org/stamp/stamp.jsp?tp=&arnumber=9638378 DO - https://doi.org/10.1109/TETC.2021.3131371 SN - 2168-6750 VL - 10 IS - 1 SP - 87 EP - 98 PB - IEEE AN - OPUS4-53941 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bornemann-Pfeiffer, Martin A1 - Kern, Simon A1 - Maiwald, Michael A1 - Meyer, Klas T1 - Calibration-Free Chemical Process and Quality Control Units as Enablers for Modular Production N2 - Modular chemical production is a tangible translation of the digital transformation of the process industry for specialty chemicals. In particular, it enables the speeding-up of process development and thus a quicker time to market by flexibly connecting and orchestrating standardised physical modules and bringing them to life (i.e., parameterising them) with digitally accumulated process knowledge. We focus on the specific challenges of chemical process and quality control, which in its current form is not well suited for modular production and provide possible approaches and examples of the change towards direct analytical methods, analytical model transfer or machine-supported processes. KW - Modular Production KW - Chemical Process Control KW - Process Analytical Technology KW - Digital Transformation KW - Industry 4.0 PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-517264 DO - https://doi.org/10.1002/cite.202000150 SN - 1522-2640 VL - 93 IS - 1-2 SP - 62 EP - 70 PB - Wiley-VCH CY - Weinheim AN - OPUS4-51726 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Gottu Mukkula, A. R. A1 - Kern, Simon A1 - Salge, M. A1 - Holtkamp, M. A1 - Guhl, Svetlana A1 - Fleischer, C. A1 - Meyer, Klas A1 - Remelhe, M. A1 - Maiwald, Michael A1 - Engell, S. T1 - An Application of Modifier Adaptation with Quadratic Approximation on a Pilot Scale Plant in Industrial Environment N2 - The goal of this work is to identify the optimal operating input for a lithiation reaction that is performed in a highly innovative pilot scale continuous flow chemical plant in an industrial environment, taking into account the process and safety constraints. The main challenge is to identify the optimum operation in the absence of information about the reaction mechanism and the reaction kinetics. We employ an iterative real-time optimization scheme called modifier adaptation with quadratic approximation (MAWQA) to identify the plant optimum in the presence of plant-model mismatch and measurement noise. A novel NMR PAT-sensor is used to measure the concentration of the reactants and of the product at the reactor outlet. The experiment results demonstrate the capabilities of the iterative optimization using the MAWQA algorithm in driving a complex real plant to an economically optimal operating point in the presence of plant-model mismatch and of process and measurement uncertainties. KW - Process Analytical Technology KW - Online NMR Spectroscopy KW - Process Industry KW - Iterative real-time optimization KW - Modifier adaptation KW - Plant-model mismatch KW - Reactor control KW - CONSENS PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-524531 DO - https://doi.org/10.1016/j.ifacol.2020.12.685 SN - 1522-2640 VL - 53 IS - 2 SP - 11773 EP - 11779 PB - Elsevier CY - Amsterdam AN - OPUS4-52453 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bruno, Francesco A1 - Fiorucci, Letizia A1 - Vignoli, Alessia A1 - Meyer, Klas A1 - Maiwald, Michael A1 - Ravera, Enrico T1 - pyIHM: Indirect Hard Modeling, in Python N2 - NMR is a powerful analytical technique that combines an exquisite qualitative power, related to the unicity of the spectra of each molecule in a mixture, with an intrinsic quantitativeness, related to the fact that the integral of each peak only depends on the number of nuclei (i.e., the amount of substance times the number of equivalent nuclei in the signal), regardless of the molecule. Signal integration is the most common approach in quantitative NMR but has several drawbacks (vide infra). An alternative is to use hard modeling of the peaks. In this paper, we present pyIHM, a Python package for the quantification of the components of NMR spectra through indirect hard modeling, and we discuss some numerical details of the implementation that make this approach robust and reliable. KW - Algorithms KW - Chemical Structure KW - Deconvolution KW - Mixtures KW - NMR spectroscopy PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-626793 DO - https://doi.org/10.1021/acs.analchem.4c06484 SN - 1520-6882 VL - 97 IS - 8 SP - 4598 EP - 4605 PB - ACS Publications CY - Washington D.C. AN - OPUS4-62679 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - You, Zengchao A1 - Akkus, Asli A1 - Weisheit, W. A1 - Giray, Thorsten A1 - Penk, Sibylle A1 - Buttler, Sabine A1 - Recknagel, Sebastian A1 - Abad Andrade, Carlos Enrique T1 - Multielement analysis in soils using nitrogen microwave inductively coupled atmospheric pressure plasma mass spectrometry N2 - In this study, we employed nitrogen microwave inductively coupled atmospheric-pressure plasma (MICAP) combined with quadrupole mass spectrometry (MS) and a liquid sample introduction system to analyze heavy metals in soils. The vanadium, cobalt, nickel, zinc, copper, chromium, arsenic, lead, and cadmium contents in seven reference and three environmental soil samples determined using MICAP-MS were within the uncertainty of the reference values, indicating that MICAP-MS is promising for soil analysis similar to the conventional inductively coupled plasma mass spectrometry (ICP-MS) technique. In addition, the limits of detection (LODs) and sensitivity of both techniques using N2 and Ar plasma were of the same order of magnitude. Furthermore, the performance of MICAP-MS under different N2 purity was investigated, and we found that the plasma formation and ionization efficiency were not influenced by the impurities in the gas. A prominent advantage of MICAP-MS is the low operating cost associated with gas consumption. In this work, MICAP-MS used nitrogen, which is cheaper than argon, and consumed 25% less gas than ICP-MS. Using low-purity N2 can further reduce the gas cost, making MICAP-MS more cost effective than ICP-MS. These results suggest that MICAP-MS is a promising alternative to ICP-MS for the analysis of heavy metals in the soil. KW - Soil KW - Microwave inductively coupled atmospheric pressure mass spectrometry (MICAP-MS) KW - Nitrogen plasma KW - Multi-element analysis PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-561048 DO - https://doi.org/10.1039/d2ja00244b SN - 0267-9477 VL - 37 IS - 12 SP - 2556 EP - 2562 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-56104 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Neamtu, M. A1 - Nadejde, C. A1 - Brinza, L. A1 - Dragos, O. A1 - Gherghel, D. A1 - Paul, Andrea T1 - Iron phthalocyanine-sensitized magnetic catalysts for BPA photodegradation N2 - The catalytic behavior of iron phthalocyanine (FePc)-sensitized magnetic nanocatalysts was evaluated for their application in the oxidative treatment of Bisphenol A (BPA) under mild environmental conditions. Two types of FePc (Fe(II)Pc and Fe(III)Pc), which are highly photosensitive compounds, were immobilized on the surface of functionalized magnetite. The nanomaterials were characterized by high resolution transmission electron microscopy (HR-TEM), X-ray difraction (XRD), Fourier transform infrared spectroscopy (FTIR) and thermogravimetric analyses (TGA). The generation of singlet Oxygen by nanomaterials was also investigated. In the presence of UVA light exposure (365nm) and 15mM H2O2, the M@Fe(III)Pc photocatalyst gave the best results; for a catalyst concentration of 2.0gL −1, around 60% BPA was removed after 120min of reaction. These experimental conditions were further tested under natural solar light exposure, for which also M@Fe(III)Pc exhibited enhanced oxidative catalytic activity, being able to remove 83% of BPA in solution. The water samples were less cytotoxic after treatment, this being confrmed by the MCF-7 cell viability assay. KW - Photosensitization KW - magnetic nanocatalysts KW - BPA removal PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-506368 DO - https://doi.org/10.1038/s41598-020-61980-6 VL - 10 IS - 1 SP - 5376 AN - OPUS4-50636 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Eifert, T. A1 - Eisen, K. A1 - Maiwald, Michael A1 - Herwig, C. T1 - Current and future requirements to industrial analytical infrastructure—part 2: smart sensors N2 - Complex processes meet and need Industry 4.0 capabilities. Shorter product cycles, flexible production needs, and direct assessment of product quality attributes and raw material attributes call for an increased need of new process analytical technologies (PAT) concepts. While individual PAT tools may be available since decades, we need holistic concepts to fulfill above industrial needs. In this series of two contributions, we want to present a combined view on the future of PAT (process analytical technology), which is projected in smart labs (Part 1) and smart sensors (Part 2). Part 2 of this feature article series describes the future functionality as well as the ingredients of a smart sensor aiming to eventually fuel full PAT functionality. The smart sensor consists of (i) chemical and process information in the physical twin by smart field devices, by measuring multiple components, and is fully connected in the IIoT 4.0 environment. In addition, (ii) it includes process intelligence in the digital twin, as to being able to generate knowledge from multi-sensor and multi-dimensional data. The cyber-physical system (CPS) combines both elements mentioned above and allows the smart sensor to be self-calibrating and self-optimizing. It maintains its operation autonomously. Furthermore, it allows—as central PAT enabler—a flexible but also target-oriented predictive control strategy and efficient process development and can compensate variations of the process and raw material attributes. Future cyber-physical production systems—like smart sensors—consist of the fusion of two main pillars, the physical and the digital twins. We discuss the individual elements of both pillars, such as connectivity, and chemical analytics on the one hand as well as hybrid models and knowledge workflows on the other. Finally, we discuss its integration needs in a CPS in order to allow is versatile deployment in efficient process development and advanced optimum predictive process control. KW - Smart sensors KW - Industry 4.0 KW - Digital twins KW - Process intelligence KW - Process analytical technology KW - Physical twin KW - Cyber-physical system PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-503980 DO - https://doi.org/10.1007/s00216-020-02421-1 SN - 1618-2642 VL - 412 IS - 9 SP - 2037 EP - 2045 PB - Springer CY - Berlin Heidelberg AN - OPUS4-50398 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Eisen, K A1 - Eifert, T A1 - Herwig, C A1 - Maiwald, Michael T1 - Current and future requirements to industrial analytical infrastructure—part 1: process analytical laboratories N2 - The competitiveness of the chemical and pharmaceutical industry is based on ensuring the required product quality while making optimum use of plants, raw materials, and energy. In this context, effective process control using reliable chemical process analytics secures global competitiveness. The setup of those control strategies often originate in process development but need to be transferable along the whole product life cycle. In this series of two contributions, we want to present a combined view on the future of PAT (process analytical technology), which is projected in smart labs (part 1) and smart sensors (part 2). In laboratories and pilot plants, offline chemical analytical methods are frequently used, where inline methods are also used in production. Here, a transferability from process development to the process in operation would be desirable. This can be obtained by establishing PAT methods for production already during process development or scale-up. However, the current PAT (Bakeev 2005, Org Process Res 19:3–62; Simon et al. 2015, Org Process Res Dev 19:3–62) must become more flexible and smarter. This can be achieved by introducing digitalization-based knowledge management, so that knowledge from product development enables and accelerates the integration of PAT. Conversely, knowledge from the production process will also contribute to product and process development. This contribution describes the future role of the laboratory and develops requirements therefrom. In part 2, we examine the future functionality as well as the ingredients of a smart sensor aiming to eventually fuel full PAT functionality—also within process development or scale-up facilities (Eifert et al. 2020, Anal Bioanal Chem). KW - Smart test laboratories KW - Laboratory 4.0 KW - Sustainable Production KW - Industry 4.0 PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-504020 DO - https://doi.org/10.1007/s00216-020-02420-2 SN - 1618-2642 VL - 412 IS - 9 SP - 2027 EP - 2035 PB - Springer CY - Berlin Heidelberg AN - OPUS4-50402 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Thiele, Isabel A1 - Santolin, Lara A1 - Meyer, Klas A1 - Machatschek, Rainhard A1 - Bölz, Uwe A1 - Tarazona, Natalia A. A1 - Riedel, Sebastian L. T1 - Microbially synthesized poly(hydroxybutyrate-co-hydroxyhexanoate) with low to moderate hydroxyhexanoate content: Properties and applications N2 - Plastic pollution is the biggest environmental concern of our time. Breakdown products like micro- and nanoplastics inevitably enter the food chain and pose unprecedented health risks. In this scenario, bio-based and biodegradable plastic alternatives have been given a momentum aiming to bridge a transition towards a more sustainable future. Polyhydroxyalkanoates (PHAs) are one of the few thermoplastic polymers synthesized 100 % via biotechnological routes which fully biodegrade in common natural environments. Poly(hydroxybutyrate-cohydroxyhexanoate) [P(HB-co-HHx)] is a PHA copolymer with great potential for the commodity polymers industry, as its mechanical properties can be tailored through fine-tuning of its molar HHx content. We have recently developed a strategy that enables for reliable tailoring of the monomer content of P(HB-co-HHx). Nevertheless, there is often a lack of comprehensive investigation of the material properties of PHAs to evaluate whether they actually mimic the functionalities of conventional plastics. We present a detailed study of P(HB-co-HHx) copolymers with low to moderate hydroxyhexanoate content to understand how the HHx monomer content influences the thermal and mechanical properties and to link those to their abiotic degradation. By increasing the HHx fractions in the range of 2 – 14 mol%, we impart an extension of the processing window and application range as the melting temperature (Tm) and glass temperature (Tg) of the copolymers decrease from Tm 165 ◦C to 126 ◦C, Tg 4 ◦C to − 5.9 ◦C, accompanied by reduced crystallinity from 54 % to 20 %. Elongation at break was increased from 5.7 % up to 703 % at 14 mol% HHx content, confirming that the range examined was sufficiently large to obtain ductile and brittle copolymers, while tensile strength was maintained throughout the studied range. Finally, accelerated abiotic degradation was shown to be slowed down with an increasing HHx fraction decreasing from 70 % to 55 % in 12 h. KW - Molecular Biology KW - General Medicine KW - Biochemistry KW - Structural Biology PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-595636 DO - https://doi.org/10.1016/j.ijbiomac.2024.130188 VL - 263 SP - 1 EP - 9 PB - Elsevier B.V. AN - OPUS4-59563 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Villajos Collado, José Antonio A1 - Balderas‐Xicohténcatl, Rafael A1 - Al Shakhs, Ali N. A1 - Berenguer‐Murcia, Ángel A1 - Buckley, Craig E. A1 - Cazorla‐Amorós, Diego A1 - Charalambopoulou, Georgia A1 - Couturas, Fabrice A1 - Cuevas, Fermin A1 - Fairen‐Jimenez, David A1 - Heinselman, Karen N. A1 - Humphries, Terry D. A1 - Kaskel, Stefan A1 - Kim, Hyunlim A1 - Marco‐Lozar, Juan P. A1 - Oh, Hyunchul A1 - Parilla, Philip A. A1 - Paskevicius, Mark A1 - Senkovska, Irena A1 - Shulda, Sarah A1 - Silvestre‐Albero, Joaquin A1 - Steriotis, Theodore A1 - Tampaxis, Christos A1 - Hirscher, Michael A1 - Maiwald, Michael T1 - Establishing ZIF‐8 as a reference material for hydrogen cryoadsorption: An interlaboratory study N2 - AbstractHydrogen storage by cryoadsorption on porous materials has the advantages of low material cost, safety, fast kinetics, and high cyclic stability. The further development of this technology requires reliable data on the H2 uptake of the adsorbents, however, even for activated carbons the values between different laboratories show sometimes large discrepancies. So far no reference material for hydrogen cryoadsorption is available. The metal‐organic framework ZIF‐8 is an ideal material possessing high thermal, chemical, and mechanical stability that reduces degradation during handling and activation. Here, we distributed ZIF‐8 pellets synthesized by extrusion to 9 laboratories equipped with 15 different experimental setups including gravimetric and volumetric analyzers. The gravimetric H2 uptake of the pellets was measured at 77 K and up to 100 bar showing a high reproducibility between the different laboratories, with a small relative standard deviation of 3–4 % between pressures of 10–100 bar. The effect of operating variables like the amount of sample or analysis temperature was evaluated, remarking the calibration of devices and other correction procedures as the most significant deviation sources. Overall, the reproducible hydrogen cryoadsorption measurements indicate the robustness of the ZIF‐8 pellets, which we want to propose as a reference material. KW - Physical and theoretical chemistry KW - Atomic and molecular physics, and optics PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-594788 DO - https://doi.org/10.1002/cphc.202300794 SN - 1439-7641 SP - 1 EP - 7 PB - Wiley CY - Weinheim AN - OPUS4-59478 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kowalewska, Z. A1 - Abad Andrade, Carlos Enrique A1 - Okruss, M. A1 - Recknagel, Sebastian T1 - Feasibility of high-resolution continuum source molecular absorption spectrometry for vanadium determination N2 - This work aimed to evaluate high-resolution continuum source molecular absorption spectrometry (HR-CS MAS), traditionally used to determine non-metals, for the determination of a new element, metal, vanadium. VO was selected as a target molecule because it is relatively stable and was expected to be spontaneously generated in a flame or a graphite furnace (GF). The high-resolution overview spectra of the molecule were obtained in a wide range of 480–630 nm, and absorption due to the X4Σ−–C4Σ− electronic transition was registered. A unique instrumental setup, comprising a prototype Modular Simultaneous Echelle Spectrograph (MOSES) and a commercial HR-CS MAS apparatus, was applied in the research. Finally, the spectral area centered at 550.6230 nm was selected for analysis. A method was developed to determine V in solutions of catalysts of heavy petroleum oil hydroprocessing using a commercial HR-CS spectrometer in a flame version. Although sensitivity was relatively poor (characteristic concentration 380 mg L−1), an extremely low noise enabled reaching a satisfactory detection limit (20 mg L−1 in solution, i.e. 0.1% m:m in the catalyst). For the first time vanadium was determined using ordinary air-acetylene flame. The obtained results were consistent with the results of atomic absorption spectrometry with N2O-C2H2 flame. Unfortunately, only a small population of VO molecules could have been generated in GF measurements. Furthermore, the observed VO molecules appeared only at unfavorably high temperatures. The work shows the potential of HR-CS MAS as a scientific tool for investigating the mechanism of processes occurring in the GF. This work can inspire other research of new analytes for HR-CS MAS. KW - Vanadium KW - HR-CS-MAS KW - VO KW - High-resolution continuum source molecular absorption spectrometry KW - AAS PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-569899 DO - https://doi.org/10.1039/D2JA00281G SN - 0267-9477 VL - 38 IS - 2 SP - 472 EP - 483 PB - Royal Society of Chemistry AN - OPUS4-56989 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Zhuang, Y. A1 - Spahr, S. A1 - Lutze, H.V. A1 - Reith, C.J. A1 - Hagemann, N. A1 - Paul, Andrea A1 - Haderlein, S.B. T1 - Persulfate activation by biochar and iron: Effect of chloride on formation of reactive species and transformation of N,N-diethyl-m-toluamide (DEET) N2 - Fenton-like processes using persulfate for oxidative water treatment and contaminant removal can be enhanced by the addition of redox-active biochar, which accelerates the reduction of Fe(III) to Fe(II) and increases the yield of reactive species that react with organic contaminants. However, available data on the formation of non-radical or radical species in the biochar/Fe(III)/persulfate system are inconsistent, which limits the evaluation of treatment efficiency and applicability in different water matrices. Based on competition kinetics calculations, we employed different scavengers and probe compounds to systematically evaluate the effect of chloride in presence of organic matter on the formation of major reactive species in the biochar/Fe(III)/persulfate system for the transformation of the model compound N,N‑diethyl-m-toluamide (DEET) at pH 2.5. We show that the transformation of methyl phenyl sulfoxide (PMSO) to methyl phenyl sulfone (PMSO2) cannot serve as a reliable indicator for Fe(IV), as previously suggested, because sulfate radicals also induce PMSO2 formation. Although the formation of Fe(IV) cannot be completely excluded, sulfate radicals were identified as the major reactive species in the biochar/Fe(III)/persulfate system in pure water. In the presence of dissolved organic matter, low chloride concentrations (0.1 mM) shifted the major reactive species likely to hydroxyl radicals. Higher chloride concentrations (1 mM), as present in a mining-impacted acidic surface water, resulted in the formation of another reactive species, possibly Cl2•−, and efficient DEET degradation. To tailor the application of this oxidation process, the water matrix must be considered as a decisive factor for reactive species formation and contaminant removal. KW - Water treatment KW - Organic contaminants KW - Fenton-like systems KW - Radicals KW - ESR PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-608677 DO - https://doi.org/10.1016/j.watres.2024.122267 SN - 0043-1354 VL - 265 SP - 1 EP - 8 PB - Elsevier Ltd. AN - OPUS4-60867 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - You, Zengchao A1 - Winckelmann, Alexander A1 - Vogl, Jochen A1 - Recknagel, Sebastian A1 - Abad Andrade, Carlos Enrique T1 - Determination of calcium, iron, and selenium in human serum by isotope dilution analysis using nitrogen microwave inductively coupled atmospheric pressure plasma mass spectrometry (MICAP-MS) N2 - In this study, we demonstrate the applicability of nitrogen microwave inductively coupled atmospheric pressure mass spectrometry (MICAP-MS) for Ca, Fe, and Se quantification in human serum using isotope dilution (ID) analysis. The matrix tolerance of MICAP-MS in Na matrix was investigated, uncovering that high Na levels can suppress the signal intensity. This suppression is likely due to the plasma loading and the space charge effect. Moreover, 40Ca and 44Ca isotopic fractionation was noted at elevated Na concentration. Nine certified serum samples were analyzed using both external calibration and ID analysis. Overestimation of Cr, Zn, As, and Se was found in the results of external calibration, which might be resulted from C-induced polyatomic interference and signal enhancement, respectively. Further investigations performed with methanol showed a similar enhancement effect for Zn, As, and Se, potentially supporting this assumption. The mass concentrations determined with ID analysis show metrological compatibility with the reference values, indicating that MICAP-MS combined with ID analysis can be a promising method for precise Ca, Fe, and Se determination. Moreover, this combination reduces the influences of matrix effects, broadening the applicability of MICAP-MS for samples with complex matrix. KW - Selenium KW - Nitrogen microwave inductively coupled atmospheric pressure mass spectrometry KW - Isotope dilution KW - Human serum KW - Calcium KW - Iron PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-598664 DO - https://doi.org/10.1007/s00216-024-05274-0 SN - 1618-2642 SP - 3117 EP - 3125 PB - Springer CY - Berlin AN - OPUS4-59866 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tchipilov, Teodor A1 - Meyer, Klas A1 - Weller, Michael G. T1 - Quantitative 1H nuclear magnetic resonance (qNMR) of aromatic amino acids for protein quantification N2 - Hydrolysis of protein samples into amino acids facilitates the use of NMR spectroscopy for protein and peptide quantification. Different conditions have been tested for quantifying aromatic amino acids and proteins. The pH-dependent signal shifts in the aromatic region of amino acid samples were examined. A pH of 12 was found to minimize signal overlap of the four aromatic amino acids. Several aromatic compounds, such as terephthalic acid, sulfoisophthalic acid, and benzene tricarboxylic acid, were applied as internal standards. The quantification of amino acids from an amino acid standard was performed. Using the first two suggested internal standards, recovery was ~97% for histidine, phenylalanine, and tyrosine at a concentration of approximately 1 mM in solution. Acidic hydrolysis of a certified reference material (CRM) of bovine serum albumin (BSA) and subsequent quantification of Phe and Tyr yielded recoveries of 98% ± 2% and 88% ± 4%, respectively, at a protein concentration of 16 g/L or 250 µM. KW - Amino acid analysis KW - Aromatic amino acid analysis KW - AAA KW - AAAA KW - Protein hydrolysis KW - Hydrochloric acid KW - Metrology KW - Traceability KW - NIST KW - Reference materials KW - Internal standard KW - Calibration KW - Compound-independent calibration KW - Histidine KW - Tyrosine KW - Tryptophan KW - Phenylalanine KW - Terephthalic acid KW - Benzene-1,3,5-tricarboxylic acid KW - Bovine serum albumin (BSA) KW - Quantitative protein analysis KW - Phenylketonuria PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-570943 DO - https://doi.org/10.3390/mps6010011 VL - 6 IS - 1 SP - 1 EP - 13 PB - MDPI CY - Basel, Schweiz AN - OPUS4-57094 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Recknagel, Sebastian A1 - Vogel, Kristin T1 - Berechnung und Nutzung von Unsicherheiten zertifizierter Werte von Referenzmaterialien N2 - Referenzmaterialien sind entscheidend für die Qualitätssicherung von Laboratorien. Diese nutzen Referenzmaterialien unter anderem zur Validierung von Messverfahren und zur Kalibrierung von Geräten. Dabei ist zu beachten, dass zertifizierte Merkmalswerte stets eine gewisse Unsicherheit aufweisen. Die Ermittlung dieser Unsicherheitsbeiträge ist Gegenstand des Artikels. Des Weiteren wird ein Vorgehen beschrieben, wie die Unsicherheit von Messverfahren mit Hilfe eines Referenzmaterials bestimmt werden kann. KW - Referenzmaterial KW - Messunsicherheit KW - Qualitätskontrolle PY - 2024 DO - https://doi.org/10.1515/teme-2024-0025 SN - 0171-8096 SP - 1 EP - 7 PB - De Gruyter AN - OPUS4-60564 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Mangoufis-Giasin, I. A1 - Piqué, O. A1 - Khanipour, Peyman A1 - Mayrhofer, K. J. J. A1 - Calle-Vallejo, F. A1 - Katsounaros, I. T1 - Different promoting roles of ruthenium for the oxidation of primary and secondary alcohols on PtRu electrocatalysts N2 - This study shows remarkably different features between the oxidation of secondary and primary C₃-C₅ alcohols. The oxidation of primary alcohols is controlled by the oxidative removal of blocking adsorbates, such as CO, formed after the dissociative adsorption of alcohol molecules. Conversely, secondary alcohols do not undergo dissociative adsorption and therefore their oxidation is purely controlled by the energetics of the elementary reaction steps. In this respect, a different role of ruthenium is revealed for the electrooxidation of primary and secondary alcohols on bimetallic platinum-ruthenium catalysts. Ruthenium enhances the oxidation of primary alcohols via the established bifunctional mechanism, in which the adsorption of (hydr)oxide species that are necessary to remove the blocking adsorbates is favored. In contrast, the oxidation of secondary alcohols is enhanced by the Ru-assisted stabilization of an O-bound intermediate that is involved in the potential-limiting step. This alternative pathway enables the oxidation of secondary alcohols close to the equilibrium potential. KW - Electrocatalysis KW - Electrochemical energy conversion KW - Direct alcohol fuel cells KW - Alcohol electrooxidation KW - Secondary alcohols KW - 2-propanol oxidation KW - Electrochemical real-time mass spectrometry KW - Density functional theory PY - 2021 DO - https://doi.org/10.1016/j.jcat.2021.05.028 SN - 0021-9517 VL - 400 SP - 166 EP - 172 PB - Elsevier CY - San Diego, Calif. AN - OPUS4-56999 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fricke, F. A1 - Mahmood, S. A1 - Hoffmann, J. A1 - Brandalero, M. A1 - Liehr, Sascha A1 - Kern, Simon A1 - Meyer, Klas A1 - Kowarik, S. A1 - Westerdick, S. A1 - Maiwald, Michael A1 - Hübner, M. T1 - Artificial Intelligence for Mass Spectrometry and Nuclear Magnetic Resonance Spectroscopy N2 - Mass Spectrometry (MS) and Nuclear Magnetic Resonance Spectroscopy (NMR) are critical components of every industrial chemical process as they provide information on the concentrations of individual compounds and by-products. These processes are carried out manually and by a specialist, which takes a substantial amount of time and prevents their utilization for real-time closed-loop process control. This paper presents recent advances from two projects that use Artificial Neural Networks (ANNs) to address the challenges of automation and performance-efficient realizations of MS and NMR. In the first part, a complete toolchain has been developed to develop simulated spectra and train ANNs to identify compounds in MS. In the second part, a limited number of experimental NMR spectra have been augmented by simulated spectra to train an ANN with better prediction performance and speed than state-of-theart analysis. These results suggest that, in the context of the digital transformation of the process industry, we are now on the threshold of a possible strongly simplified use of MS and MRS and the accompanying data evaluation by machine-supported procedures, and can utilize both methods much wider for reaction and process monitoring or quality control. T2 - 2021 Design, Automation & Test in Europe Conference & Exhibition (DATE) CY - Grenoble, France DA - 01.02.2021 KW - Industry 4.0, KW - Cyber-physical systems KW - Artificial neural networks KW - Mass spectrometry KW - Nuclear magnetic resonance spectroscopy PY - 2021 DO - https://doi.org/10.23919/DATE51398.2021.9473958 SP - 615 EP - 620 PB - IEEE AN - OPUS4-55360 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Anderhalten, L. A1 - Silva, R. V. A1 - Morr, A. A1 - Wang, S. A1 - Smorodchenko, A. A1 - Saatz, Jessica A1 - Traub, Heike A1 - Mueller, S. A1 - Boehm-Sturm, P. A1 - Rodriguez-Sillke, Y. A1 - Kunkel, D. A1 - Hahndorf, J. A1 - Paul, F. A1 - Taupitz, M. A1 - Sack, I. A1 - Infante-Duarte, C. T1 - Different Impact of Gadopentetate and Gadobutrol on Inflammation-Promoted Retention and Toxicity of Gadolinium Within the Mouse Brain N2 - Objectives: Using a murine model of multiple sclerosis, we previously showed that repeated administration of gadopentetate dimeglumine led to retention of gadolinium (Gd) within cerebellar structures and that this process was enhanced with inflammation. This study aimed to compare the kinetics and retention profiles of Gd in inflamed and healthy brains after application of the macrocyclic Gd-based contrast agent (GBCA) gadobutrol or the linear GBCA gadopentetate. Moreover, potential Gd-induced neurotoxicity was investigated in living hippocampal slices ex vivo. Materials and Methods: Mice at peak of experimental autoimmune encephalomyelitis (EAE; n = 29) and healthy control mice (HC; n = 24) were exposed to a cumulative dose of 20 mmol/kg bodyweight of either gadopentetate dimeglumine or gadobutrol (8 injections of 2.5 mmol/kg over 10 days). Magnetic resonance imaging (7 T) was performed at baseline as well as at day 1, 10, and 40 post final injection (pfi) of GBCAs. Mice were sacrificed after magnetic resonance imaging and brain and blood Gd content was assessed by laser ablation-inductively coupled plasma (ICP)-mass spectrometry (MS) and ICP-MS, respectively. In addition, using chronic organotypic hippocampal slice cultures, Gd-induced neurotoxicity was addressed in living brain tissue ex vivo, both under control or inflammatory (tumor necrosis factor α [TNF-α] at 50 ng/μL) conditions. Results: Neuroinflammation promoted a significant decrease in T1 relaxation times after multiple injections of both GBCAs as shown by quantitative T1 mapping of EAE brains compared with HC. This corresponded to higher Gd retention within the EAE brains at 1, 10, and 40 days pfi as determined by laser ablation-ICP-MS. In inflamed cerebellum, in particular in the deep cerebellar nuclei (CN), elevated Gd retention was observed until day 40 after last gadopentetate application (CN: EAE vs HC, 55.06 ± 0.16 μM vs 30.44 ± 4.43 μM). In contrast, gadobutrol application led to a rather diffuse Gd content in the inflamed brains, which strongly diminished until day 40 (CN: EAE vs HC, 0.38 ± 0.08 μM vs 0.17 ± 0.03 μM). The analysis of cytotoxic effects of both GBCAs using living brain tissue revealed an elevated cell death rate after incubation with gadopentetate but not gadobutrol at 50 mM. The cytotoxic effect due to gadopentetate increased in the presence of the inflammatory mediator TNF-α (with vs without TNF-α, 3.15% ± 1.18% vs 2.17% ± 1.14%; P = 0.0345). Conclusions: In the EAE model, neuroinflammation promoted increased Gd retention in the brain for both GBCAs. Whereas in the inflamed brains, efficient clearance of macrocyclic gadobutrol during the investigated time period was observed, the Gd retention after application of linear gadopentetate persisted over the entire observational period. Gadopentetate but not gadubutrol appeared to be neurotoxic in an ex vivo paradigm of neuronal inflammation. KW - Imaging KW - ICP-MS KW - Gadolinium KW - Contrast agent KW - Laser ablation KW - Brain KW - Multiple sclerosis PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-546910 DO - https://doi.org/10.1097/RLI.0000000000000884 SN - 0020-9996/22/0000–0000 VL - 57 IS - 10 SP - 677 EP - 688 PB - Wolters Kluwer N.V. CY - Alphen aan den Rijn, The Netherlands AN - OPUS4-54691 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kader, A. A1 - Kaufmann, Jan Ole A1 - Mangarova, D. B. A1 - Moeckel, J. A1 - Brangsch, J. A1 - Adams, L. C. A1 - Zhao, J. A1 - Reimann, C. A1 - Saatz, Jessica A1 - Traub, Heike A1 - Buchholz, R. A1 - Karst, U. A1 - Hamm, B. A1 - Makowski, M. R. T1 - Iron Oxide Nanoparticles for Visualization of Prostate Cancer in MRI N2 - Prostate cancer (PCa) is one of the most common cancers in men. For detection and diagnosis of PCa, non-invasive methods, including magnetic resonance imaging (MRI), can reduce the risk potential of surgical intervention. To explore the molecular characteristics of the tumor, we investigated the applicability of ferumoxytol in PCa in a xenograft mouse model in two different tumor volumes, 500 mm3 and 1000 mm3. Macrophages play a key role in tumor progression, and they are able to internalize iron-oxide particles, such as ferumoxytol. When evaluating T2*-weighted sequences on MRI, a significant decrease of signal intensity between pre- and post-contrast images for each tumor volume (n = 14; p < 0.001) was measured. We, furthermore, observed a higher signal loss for a tumor volume of 500 mm3 than for 1000 mm3. These findings were confirmed by histological examinations and laser ablation inductively coupled plasma-mass spectrometry. The 500 mm3 tumors had 1.5% iron content (n = 14; sigma = 1.1), while the 1000 mm3 tumors contained only 0.4% iron (n = 14; sigma = 0.2). In vivo MRI data demonstrated a correlation with the ex vivo data (R2 = 0.75). The results of elemental analysis by inductively coupled plasma-mass spectrometry correlated strongly with the MRI data (R2 = 0.83) (n = 4). Due to its long retention time in the blood, biodegradability, and low toxicity to patients, ferumoxytol has great potential as a contrast agent for visualization PCa. KW - Imaging KW - Nanoparticle KW - Cancer KW - Iron oxide KW - ICP-MS KW - Magnetic resonance imaging PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-550075 DO - https://doi.org/10.3390/cancers14122909 VL - 14 IS - 12 SP - 1 EP - 13 PB - MDPI CY - Basel, Switzerland AN - OPUS4-55007 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Golusda, L. A1 - Kühl, A. A. A1 - Lehmann, M. A1 - Dahlke, K. A1 - Mueller, S. A1 - Boehm-Sturm, P. A1 - Saatz, Jessica A1 - Traub, Heike A1 - Schnorr, J. A1 - Freise, C. A1 - Taupitz, M. A1 - Biskup, K. A1 - Blanchard, V. A1 - Klein, O. A1 - Sack, I. A1 - Siegmund, B. A1 - Paclik, D. T1 - Visualization of inflammation in experimental colitis by magnetic resonance imaging using very small superparamagnetic iron oxide particles N2 - Inflammatory bowel diseases (IBD) comprise mainly ulcerative colitis (UC) and Crohn´s disease (CD). Both forms present with a chronic inflammation of the (gastro) intestinal tract, which induces excessive changes in the composition of the associated extracellular matrix (ECM). In UC, the inflammation is limited to the colon, whereas it can occur throughout the entire gastrointestinal tract in CD. Tools for early diagnosis of IBD are still very limited and highly invasive and measures for standardized evaluation of structural changes are scarce. To investigate an efficient non-invasive way of diagnosing intestinal inflammation and early changes of the ECM, very small superparamagnetic iron oxide nanoparticles (VSOPs) in magnetic resonance imaging (MRI) were applied in two mouse models of experimental colitis: the dextran sulfate sodium (DSS)-induced colitis and the transfer model of colitis. For further validation of ECM changes and inflammation, tissue sections were analyzed by immunohistochemistry. For in depth ex-vivo investigation of VSOPs localization within the tissue, Europium-doped VSOPs served to visualize the contrast agent by imaging mass cytometry (IMC). VSOPs accumulation in the inflamed colon wall of DSS-induced colitis mice was visualized in T2* weighted MRI scans. Components of the ECM, especially the hyaluronic acid content, were found to influence VSOPs binding. Using IMC, colocalization of VSOPs with macrophages and endothelial cells in colon tissue was shown. In contrast to the DSS model, colonic inflammation could not be visualized with VSOP-enhanced MRI in transfer colitis. VSOPs present a potential contrast agent for contrast-enhanced MRI to detect intestinal inflammation in mice at an early stage and in a less invasive manner depending on hyaluronic acid content. KW - Inflammation KW - Imaging KW - Immunohistochemistry KW - MRI KW - Nanoparticle KW - Extracellular matrix KW - Laser ablation KW - ICP-MS PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-555395 DO - https://doi.org/10.3389/fphys.2022.862212 SN - 1664-042X VL - 13 IS - July 2022 SP - 1 EP - 15 PB - Frontiers Research Foundation CY - Lausanne AN - OPUS4-55539 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Völker, Tobias A1 - Gornushkin, Igor B. T1 - Mass and mole fractions in calibration-free LIBS N2 - This technical note highlights the fact that CF-LIBS algorithms work in mole fractions, while results of spectrochemical analysis are usually reported in mass fractions or mass percent. Ignoring this difference and not converting mole fractions to mass fractions can lead to errors in reported concentrations determined by the CF-LIBS method and inadequate comparison of these concentrations with certified concentrations. Here, the key points of the CF-LIBS algorithm are reproduced and the formulae for converting a mole fraction to a mass fraction and vice versa are given. Several numerical examples are also given, which show that the greater the difference between the molar mass of an individual element in a sample and the average molar mass, the greater the discrepancy between the mole and mass fractions. KW - Spectroscopy KW - Analytical Chemistry KW - LIBS KW - Calibration PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-598611 DO - https://doi.org/10.1039/d4ja00028e SN - 0267-9477 VL - 39 IS - 4 SP - 1030 EP - 1032 PB - Royal Society of Chemistry (RSC) AN - OPUS4-59861 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kirilina, E. A1 - Helbling, S. A1 - Morawski, M. A1 - Pine, K. A1 - Reimann, K. A1 - Jankuhn, S. A1 - Dinse, J. A1 - Deistung, A. A1 - Reichenbach, J. R. A1 - Trampel, R. A1 - Geyer, S. A1 - Müller, Larissa A1 - Jakubowski, Norbert A1 - Arendt, T. A1 - Bazin, P.-L. A1 - Weiskopf, N. T1 - Superficial white matter imaging: Contrast mechanisms and whole-brain in vivo mapping N2 - Superficial white matter (SWM) contains the most cortico-cortical white matter connections in the human brain encompassing the short U-shaped association fibers. Despite its importance for brain connectivity, very little is known about SWM in humans, mainly due to the lack of noninvasive imaging methods. Here, we lay the groundwork for systematic in vivo SWM mapping using ultrahigh resolution 7 T magnetic resonance imaging. Using biophysical modeling informed by quantitative ion beam microscopy on postmortem brain tissue, we demonstrate that MR contrast in SWM is driven by iron and can be linked to the microscopic iron distribution. Higher SWM iron concentrations were observed in U-fiber–rich frontal, temporal, and parietal areas, potentially reflecting high fiber density or late myelination in these areas. Our SWM mapping approach provides the foundation for systematic studies of interindividual differences, plasticity, and pathologies of this crucial structure for cortico-cortical connectivity in humans. KW - Magnetic resonance imaging KW - Laser ablation KW - ICP-MS KW - Brain KW - Imaging PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-514430 DO - https://doi.org/10.1126/sciadv.aaz9281 SN - 2375-2548 VL - 6 IS - 41 SP - eaaz9281 PB - American Association for the Advancement of Science (Science/AAAS) CY - Washington, DC, USA AN - OPUS4-51443 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Garcia-Fernandez, J. A1 - Turiel, D. A1 - Bettmer, J. A1 - Jakubowski, Norbert A1 - Panne, Ulrich A1 - Rivas Garcia, L. A1 - Llopis, J. A1 - Sanchez-Gonzalez, C. A1 - Montes-Bayon, M. T1 - In vitro and in situ experiments to evaluate the biodistribution and cellular toxicity of ultrasmall iron oxide nanoparticles potentially used as oral iron supplements N2 - Well-absorbed iron-based nanoparticulated materials are a promise for the oral management of iron deficient anemia. In this work, a battery of in vitro and in situ experiments are combined for the evaluation of the uptake, distribution and toxicity of new synthesized ultrasmall (4 nm core) Fe2O3 nanoparticles coated with tartaric/adipic acid with potential to be used as oral Fe supplements. First, the in vitro simulated gastric acid solubility studies by TEM and HPLC-ICP-MS reveal a partial reduction of the core size of about 40% after 90 min at pH3. Such scenario confirms the arrival of the nanoparticulate material in the small intestine. In the next step, the in vivo absorption through the small intestine by intestinal perfusion experiments is conducted using the sought nanoparticles in Wistar rats. The quantification of Fe in the NPs Suspension before and after perfusion shows Fe absorption levels above 79%, never reported for other Fe treatments. Such high absorption levels do not seem to compromise cell viability, evaluated in enterocytes-like models (Caco-2 and HT-29) using cytotoxicity, ROS production, genotoxicity and lipid peroxidation tests. Moreover, regional differences in terms of Fe concentration are obtained among different parts of the small intestine as duodenum>jejunum>ileum. Complementary transmission electron microscopy (TEM) images show the presence of the intact particles around the intestinal microvilli without significant tissue damage. These studies show the high potential of these NP preparations for their use as oral management of anemia. KW - Iron nanoparticles KW - Anemia KW - ICP-MS KW - In vitro KW - In situ PY - 2020 DO - https://doi.org/10.1080/17435390.2019.1710613 VL - 14 IS - 3 SP - 388 EP - 403 PB - Taylor & Francis Online CY - London AN - OPUS4-50314 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dorgerloh, Ute A1 - Becker, Roland A1 - Freeling, Finnian A1 - Scheurer, Marco A1 - Gökçe, Emine A1 - Strack, Katrin A1 - Sommerfeld, Thomas A1 - Sauer, Andreas A1 - Schneider, Jonas A1 - Härtel, Christoph A1 - Valkov, Vassil A1 - Moser, Lukas A1 - Geiß, Sabine A1 - Junginger, Sabine A1 - Schlittenbauer, Linda A1 - Suess, Elke A1 - Ruth, Katinka T1 - Standardising the quantification of trifluoroacetic acid in water: Interlaboratory validation trial using liquid chromatography-mass spectrometric detection (LC–MS/MS) N2 - The concentration of trifluoroacetic acid in surface and groundwaters and drinking water is subject of increasing concern. Since there is no standardised procedure available, German standardisation body DIN has initiated the first draft standard using liquid chromatography with tandem mass spectrometry (LC–MS/MS) after direct injection for the concentration range between ≥ 0.1 and 3 µg/L. Herein, the interlaboratory comparison as final validation step involving 12 expert laboratories from Germany and Switzerland is described. Two groundwaters, two surface waters, a drinking water, a rainwater, and two fortified groundwater samples displayed relative repeatability standard deviations between 3 and 7% and relative reproducibility standard deviation between 6 and 20% and were included in the draft standard DIN 38407-53. KW - TFA Trifluoroacetic acid KW - LC/MS-MS KW - Validation trial PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-632376 DO - https://doi.org/10.1007/s00769-025-01640-2 SN - 0949-1775 SP - 1 EP - 4 PB - Springer CY - Berlin AN - OPUS4-63237 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rosenkranz, D. A1 - Kriegel, F. L. A1 - Mavrakis, E. A1 - Pergantis, S. A. A1 - Reichardt, P. A1 - Tentschert, J. A1 - Jakubowski, N. A1 - Laux, P. A1 - Panne, Ulrich A1 - Luch, A. T1 - Versatile dual-inlet sample introduction system for multi-mode single particle inductively coupled plasma mass spectrometry N2 - Metal-containing nanoparticles (NP) can be characterized with inductively coupled plasma mass spectrometers (ICP-MS) in terms of their size and number concentration by using the single-particle mode of the instrument (spICP-MS). The accuracy of measurement depends on the setup, operational conditions of the instrument and specific parameters that are set by the user. The transport efficiency of the ICP-MS is crucial for the quantification of the NP and usually requires a reference material with homogenous size distribution and a known particle number concentration. Currently, NP reference materials are available for only a few metals and in limited sizes. If particles are characterized without a reference standard, the results of both size and particle number may be biased. Therefore, a dual-inlet Setup for characterizing nanoparticles with spICP-MS was developed to overcome this problem. This setup is based on a conventional introduction system consisting of a pneumatic nebulizer (PN) for nanoparticle solutions and a microdroplet Generator (μDG) for ionic calibration solutions. A new and flexible interface was developed to facilitate the coupling of μDG, PN and the ICP-MS system. The interface consists of available laboratory components and allows for the calibration, nanoparticle (NP) characterization and cleaning of the arrangement, while the ICP-MS instrument is still running. Three independent analysis modes are available for determining particle size and number concentration. Each mode is based on a different calibration principle. While mode I (counting) and mode III (μDG) are known from the literature, mode II (sensitivity), is used to determine the transport efficiency by inorganic ionic Standard solutions only. It is independent of NP reference materials. The μDG based inlet system described here guarantees superior analyte sensitivities and, therefore, lower detection limits (LOD). The size dependent LODs achieved are less than 15 nm for all NP (Au, Ag, CeO2) investigated. KW - Spectrometer KW - Reference KW - Calibration PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-536121 DO - https://doi.org/10.3791/61653 SN - 1940-087X IS - 163 SP - 1 EP - 19 PB - MyJoVE Corporation CY - Cambridge, MA, USA AN - OPUS4-53612 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hachenberger, Y. U. A1 - Rosenkranz, D. A1 - Kriegel, Fabian L. A1 - Krause, B. A1 - Matschaß, René A1 - Reichardt, P. A1 - Tentschert, J. A1 - Laux, P. A1 - Jakubowski, Norbert A1 - Panne, Ulrich A1 - Luch, A. T1 - Tackling Complex Analytical Tasks: An ISO/TS-Based Validation Approach for Hydrodynamic Chromatography Single Particle Inductively Coupled Plasma Mass Spectrometry N2 - Nano-carrier systems such as liposomes have promising biomedical applications. Nevertheless, characterization of these complex samples is a challenging analytical task. In this study a coupled hydrodynamic chromatography-single particle-inductively coupled plasma mass spectrometry (HDC-spICP-MS) approach was validated based on the technical specification (TS) 19590:2017 of the international organization for standardization (ISO). The TS has been adapted to the hyphenated setup. The quality criteria (QC), e.g., linearity of the calibration, transport efficiency, were investigated. Furthermore, a cross calibration of the particle size was performed with values from dynamic light scattering (DLS) and transmission electron microscopy (TEM). Due to an additional Y-piece, an online-calibration routine was implemented. This approach allows the calibration of the ICP-MS during the dead time of the chromatography run, to reduce the required time and enhance the robustness of the results. The optimized method was tested with different gold nanoparticle (Au-NP) mixtures to investigate the characterization properties of HDC separations for samples with increasing complexity. Additionally, the technique was successfully applied to simultaneously determine both the hydrodynamic radius and the Au-NP content in liposomes. With the established hyphenated setup, it was possible to distinguish between different subpopulations with various NP loads and different hydrodynamic diameters inside the liposome carriers. KW - Single particle ICP-MS KW - Nanoparticle characterization KW - Nano-carrier KW - Iposomes KW - Hydrodynamic chromatography (HDC) KW - Validation PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-506609 DO - https://doi.org/10.3390/ma13061447 VL - 13 IS - 6 SP - 1 EP - 14 CY - Basel, Switzerland AN - OPUS4-50660 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rosenkranz, Daniel A1 - Kriegel, F. L. A1 - Mavrakis, E. A1 - Pergantis, S. A. A1 - Reichardt, Ph. A1 - Tentschert, J. A1 - Jakubowski, Norbert A1 - Laux, P. A1 - Panne, Ulrich T1 - Improved validation for single particle ICP-MS analysis using a pneumatic nebulizer / microdroplet generator sample introduction system for multi-mode nanoparticle determination N2 - This study reports on the development of a single-particle (sp) inductively coupled plasma mass spectrometry (ICP-MS) technique suitable for the multi-mode determination of nanoparticle (NP) metal mass fraction and number concentration. The described technique, which is based on a dual inlet System consisting of a pneumatic nebulizer (PN) and a microdroplet generator (MDG), allows for the sequential introduction of ionic metal calibrant solutions and nanoparticle suspensions via all combinations of the two inlets; thus allowing for a combination of three independent modes of analysis. A novel interface, assembled using standard analytical components (a demountable quartz ICP-MS torch, flexible nonconducting silicon tubing and various connectors), was used to interface the dual inlet system to an ICP-MS. The interface provided improved functionality, compared to a previous design. It is now possible to conveniently exchange and introduce standard solutions and samples via all inlet combinations, analyze them, and also wash the sample inlet systems while the whole setup is still connected to an operating ICP-MS. This setup provided seamless and robust operation in a total of three analysis modes, i.e. three ways to independently determine the metal mass fraction and NP number concentration. All three analyses modes could be carried out within a single analytical run lasting approximately 20 min. The unique feature of the described approach is that each analysis mode is based on a different. KW - Nanomaterials KW - Nanoparticles KW - Single particle ICP-MS KW - Microdroplet generator PY - 2020 DO - https://doi.org/10.1016/j.aca.2019.11.043 VL - 1099 SP - 16 EP - 25 PB - Elsevier B.V. AN - OPUS4-50361 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Nourinejhad Zarghani, S. A1 - Monavari, Mehran A1 - Nourinejhad Zarghani, A. A1 - Nouri, S. A1 - Ehlers, J. A1 - Hamacher, J. A1 - Bandte, M. A1 - Büttner, C. T1 - Quantifying plant viruses: Evolution from bioassay to infectivity dilution curves along the model of tobamoviruses N2 - This review describes the development of the bioassay as a means of quantifying plant viruses, with particular attention to tobamovirus. It delves into various models used to establish a correlation between virus particle concentration and the number of induced local lesions (the infectivity dilution curve), including the Poisson, Furumoto and Mickey, Kleczkowski, Growth curve, and modified Poisson models. The parameters of each model are described, and their application or performance in the context of the tobacco mosaic virus is explored. This overview highlights the enduring value of the infectivity dilution curve in tobamovirus quantification, providing valuable insights for researchers or practitioners of bioassays and theoreticians of modeling. KW - Local lesion KW - qRT-PCR KW - Nicotiana KW - ToBRFV KW - TMV KW - CGMMV KW - Disinfection KW - Modeling PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-630142 DO - https://doi.org/10.3390/v16030440 SN - 1999-4915 VL - 16 IS - 3 SP - 1 EP - 14 PB - MDPI CY - Basel AN - OPUS4-63014 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Quackatz, Lukas A1 - Gornushkin, Igor B. A1 - Griesche, Axel A1 - Kannengießer, Thomas A1 - Treutler, Kai A1 - Wesling, Volker T1 - In situ chemical analysis of duplex stainless steel weld by laser induced breakdown spectroscopy N2 - The high corrosion resistance and good mechanical properties of duplex stainless steel (DSS) are due to its special chemical composition, which is a balanced phase ratio of ferrite (α) and austenite (γ). Many industrial applications require the integration of DSS components. For this, Gas tungsten arc welding (GTAW) is an excellent choice, as it allows an automated operation with high reproducibility. However, when the weld pool solidifies, critical ratios of α- and γ- phases can occur, which lead to solidification cracking, increased susceptibility to corrosion, and a decrease in ductility and critical strength. Previous studies have shown that these defects can be caused by the accumulation of manganese and chromium in the heat affected zone (HAZ), requiring ongoing monitoring of this accumulation. A suitable method for such monitoring is laser-induced breakdown spectroscopy (LIBS), which can be used in two operating modes: calibration using standard reference samples and calibration-free. Unlike conventional quantitative LIBS measurements, which require reference samples to generate a calibration curve, calibration-free LIBS (CF-LIBS) allows chemical compositions to be determined solely from the emission spectrum of the plasma. Numerous publications show that CF-LIBS is a fast and efficient analytical method for the quantitative analysis of metal samples. In this work, CF-LIBS is applied to spectra obtained during GTAW DSS welding and the result is compared with those obtained by PLS analysis. A good correlation was found between both types of analysis, demonstrating the suitability of the CF-LIBS method for this application. The CF-LIBS method has a significant advantage over conventional LIBS due to the rapid in situ measurement of concentrations of major alloying elements without calibration procedure. This, combined with fast feedback and appropriate adjustment of welding parameters, helps prevent welding defects. KW - Duplex stainless steels KW - In situ measurement KW - LIBS KW - GMAW PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-597940 DO - https://doi.org/10.1016/j.sab.2024.106899 SN - 0584-8547 VL - 214 SP - 1 EP - 7 PB - Elsevier B.V. AN - OPUS4-59794 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rabin, Ira A1 - Gordon, Nehemia A1 - Calvillo, Nelson T1 - Coloured Ruling Lines in a Fourteenth-Century Ashkenazi Torah Scroll N2 - Rabbinic halakhah required that Torah scrolls be written with relief ruling, horizontal and vertical lines scored into the writing surface. Coloured ruling with plummet or ink would make a Torah scroll invalid for liturgical use. Ruling lines on some leather fragments of early Oriental Torah scrolls give the impression of being produced with plummet or ink but may be the result of dust in the furrows of relief ruling. Erfurt 9, a fourteenth-century Ashkenazi Torah scroll, has coloured ruling, which is partial secondary re-ruling, on 13 of the 34 surviving sheets, perhaps added when the original relief ruling was difficult to see. The coloured ruling in Erfurt 9 is brown, giving it the visual appearance of iron-gall ink. However, XRF tests showed that the only meaningful metallic component in the coloured ruling was lead (Pb). Lead plummet ruling tends to have a greyish appearance. Black specks under near infrared light are consistent with minium, a lead-based red ink that can turn brown when it degrades. Partial secondary re-ruling added with minium to the original relief ruling comes as a surprise as it would have rendered Erfurt 9 unfit for liturgical use. KW - Hebrew Manuscripts KW - Ruling Lines KW - Coloured Ink PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-628531 DO - https://doi.org/10.25592/uhhfdm.17169 SN - 2410-0951 VL - 10 SP - 9 EP - 32 AN - OPUS4-62853 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fink, Friedrich A1 - Stawski, Tomasz M. A1 - Emmerling, Franziska A1 - Falkenhagen, Jana T1 - A novel machine-learning approach to unlock technical lignin classification by NIR spectroscopy - bench to handheld N2 - In this research, the utilization of near-infrared (NIR) spectroscopy in conjunction with advanced machine learning methods is investigated for categorizing technical lignins obtained from different biomass sources and industrial procedures. Technical lignins, such as kraft, organosolv and lignosulfonates, have different chemical compositions, which continue to make uniform characterization and application in sustainable sectors extremely difficult. Fast, universally accessible analytics combined with data analysis is still an open question. For the first time three distinct NIR spectrometers—a high-performance benchtop system, a mid-priced compact device, and an economical handheld unit—were utilized to record NIR spectra of 31 unique lignin samples. The spectra underwent pre-processing through standard normal variate (SNV) transformation and Savitzky-Golay derivatives to amplify spectral features and decrease noise. Principal component analysis (PCA) was employed to reduce data complexity and extract crucial characteristics for classification purposes. Subsequently, four machine learning algorithms—Support Vector Machines (SVM), Gaussian Naive Bayes (GNB), Gaussian Process Classification (GPC), and Decision Tree Classification (DTC)—were implemented for the classification of the lignin samples. The DTC model exhibited the highest accuracy among them across different spectrometers. Although the benchtop spectrometer produced the most precise outcomes, the compact NeoSpectra system also displayed potential as a cost-efficient option. Nonetheless, the restricted spectral coverage of the handheld NIRONE spectrometer resulted in reduced classification accuracy. Our discoveries highlight the capability of NIR spectroscopy, combined with robust data analysis techniques, for the swift and non-destructive classification of technical lignins, facilitating their improved utilization in sustainable fields. KW - Technical lignins KW - NIR spectroscopy KW - Classification KW - Machine-learning KW - PCA PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-634518 DO - https://doi.org/10.1016/j.chemolab.2025.105467 SN - 0169-7439 VL - 264 SP - 1 EP - 10 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-63451 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - El-Athman, Rukeia A1 - Rädler, Jörg A1 - Löhmann, Oliver A1 - Ariza de Schellenberger, Angela A1 - Muth, Thilo T1 - The BAM Data Store N2 - As a partner in several NFDI consortia, the Bundesanstalt für Materialforschung und -prüfung (BAM, German federal institute for materials science and testing) contributes to research data standardization efforts in various domains of materials science and engineering (MSE). To implement a central research data management (RDM) infrastructure that meets the requirements of MSE groups at BAM, we initiated the Data Store pilot project in 2021. The resulting infrastructure should enable researchers to digitally document research processes and store related data in a standardized and interoperable manner. As a software solution, we chose openBIS, an open-source framework that is increasingly being used for RDM in MSE communities. The pilot project was conducted for one year with five research groups across different organizational units and MSE disciplines. The main results are presented for the use case “nanoPlattform”. The group registered experimental steps and linked associated instruments and chemicals in the Data Store to ensure full traceability of data related to the synthesis of ~400 nanomaterials. The system also supported researchers in implementing RDM practices in their workflows, e.g., by automating data import and documentation and by integrating infrastructure for data analysis. Based on the promising results of the pilot phase, we will roll out the Data Store as the central RDM infrastructure of BAM starting in 2023. We further aim to develop openBIS plugins, metadata standards, and RDM workflows to contribute to the openBIS community and to foster RDM in MSE. T2 - 1st Conference on Research Data Infrastructure DA - 12.09.2023 KW - Research Data Infrastructure KW - Electronic Lab Notebook (ELN) KW - openBIS KW - Research Data Management PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-596032 DO - https://doi.org/10.52825/CoRDI.v1i.229 VL - 1 SP - 1 EP - 5 AN - OPUS4-59603 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Drescher, D. A1 - Büchner, T. A1 - Schrade, P. A1 - Traub, Heike A1 - Werner, S. A1 - Guttmann, P. A1 - Bachmann, S. A1 - Kneipp, J. T1 - Influence of Nuclear Localization Sequences on the Intracellular Fate of Gold Nanoparticles N2 - Directing nanoparticles to the nucleus by attachment of nuclear localization sequences (NLS) is an aim in many applications. Gold nanoparticles modified with two different NLS were studied while crossing barriers of intact cells, including uptake, endosomal escape, and nuclear translocation. By imaging of the nanoparticles and by characterization of their molecular interactions with surface-enhanced Raman scattering (SERS), it is shown that nuclear translocation strongly depends on the particular incubation conditions. After an 1 h of incubation followed by a 24 h chase time, 14 nm gold particles carrying an adenoviral NLS are localized in endosomes, in the cytoplasm, and in the nucleus of fibroblast cells. In contrast, the cells display no nanoparticles in the cytoplasm or nucleus when continuously incubated with the nanoparticles for 24 h. The ultrastructural and spectroscopic data indicate different processing of NLS-functionalized particles in endosomes compared to unmodified particles. NLS functionalized nanoparticles form larger intraendosomal aggregates than unmodified gold nanoparticles. SERS spectra of cells with NLS-functionalized gold nanoparticles contain bands assigned to DNA and were clearly different from those with unmodified gold nanoparticles. The different processing in the presence of an NLS is influenced by a continuous exposure of the cells to nanoparticles and an ongoing nanoparticle uptake. This is supported by mass-spectrometry-based quantification that indicates enhanced uptake of NLS-functionalized nanoparticles compared to unmodified particles under the same conditions. The results contribute to the optimization of nanoparticle analysis in cells in a variety of applications, e.g., in theranostics, biotechnology, and bioanalytics. KW - Nanoparticle KW - Laser ablation KW - SERS KW - ICP-MS PY - 2021 DO - https://doi.org/10.1021/acsnano.1c04925 SN - 1936-086X VL - 15 IS - 9 SP - 14838 EP - 14849 PB - American Chemical Society AN - OPUS4-54047 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Theiner, S. A1 - Löhr, Konrad A1 - Köllensperger, G. A1 - Müller, L. A1 - Jakubowski, Norbert T1 - Single-cell analysis by use of ICP-MS N2 - This tutorial review article is highlighting the fundamentals, instrumentation, and most recent trends of single-cell analysis by use of inductively coupled plasma-mass spectrometry (ICP-MS). It is shown that metals and hetero-elements being intrinsically present in cells, taken up by cells (for instance engineered metallic nanoparticles) or binding to a cell can be detected qualitatively by existing ICP-MS Technologies on a single cell level. Adding a quantitative dimension to single-cell analysis by (laser ablation-) ICP-MS requires dedicated calibration and validation strategies, which are currently being established and are being critically discussed. In a tutorial part, the ICP-MS instruments, the measurement conditions, and the sample introduction and preparation techniques are introduced. The application section focuses on the state-of-the-art of single-cell analysis in suspension, using laser ablation or (imaging) mass cytometry. Finally, future trends are critically assessed. KW - Cell KW - ICP KW - ICP-MS KW - Laser ablation PY - 2020 DO - https://doi.org/10.1039/d0ja00194e SN - 0267-9477 VL - 35 IS - 9 SP - 1784 EP - 1813 PB - The Royal Society of Chemistry CY - Cambridge, UK AN - OPUS4-51448 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -