TY - JOUR A1 - Hachenberger, Y. U. A1 - Rosenkranz, D. A1 - Kriegel, Fabian L. A1 - Krause, B. A1 - Matschaß, René A1 - Reichardt, P. A1 - Tentschert, J. A1 - Laux, P. A1 - Jakubowski, Norbert A1 - Panne, Ulrich A1 - Luch, A. T1 - Tackling Complex Analytical Tasks: An ISO/TS-Based Validation Approach for Hydrodynamic Chromatography Single Particle Inductively Coupled Plasma Mass Spectrometry N2 - Nano-carrier systems such as liposomes have promising biomedical applications. Nevertheless, characterization of these complex samples is a challenging analytical task. In this study a coupled hydrodynamic chromatography-single particle-inductively coupled plasma mass spectrometry (HDC-spICP-MS) approach was validated based on the technical specification (TS) 19590:2017 of the international organization for standardization (ISO). The TS has been adapted to the hyphenated setup. The quality criteria (QC), e.g., linearity of the calibration, transport efficiency, were investigated. Furthermore, a cross calibration of the particle size was performed with values from dynamic light scattering (DLS) and transmission electron microscopy (TEM). Due to an additional Y-piece, an online-calibration routine was implemented. This approach allows the calibration of the ICP-MS during the dead time of the chromatography run, to reduce the required time and enhance the robustness of the results. The optimized method was tested with different gold nanoparticle (Au-NP) mixtures to investigate the characterization properties of HDC separations for samples with increasing complexity. Additionally, the technique was successfully applied to simultaneously determine both the hydrodynamic radius and the Au-NP content in liposomes. With the established hyphenated setup, it was possible to distinguish between different subpopulations with various NP loads and different hydrodynamic diameters inside the liposome carriers. KW - Single particle ICP-MS KW - Nanoparticle characterization KW - Nano-carrier KW - Iposomes KW - Hydrodynamic chromatography (HDC) KW - Validation PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-506609 DO - https://doi.org/10.3390/ma13061447 VL - 13 IS - 6 SP - 1 EP - 14 CY - Basel, Switzerland AN - OPUS4-50660 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Phukphatthanachai, P. A1 - Panne, Ulrich A1 - Traub, Heike A1 - Pfeifer, Jens A1 - Vogl, Jochen T1 - Quantification of sulphur in copper and copper alloys by GDMS and LA-ICP-MS, demonstrating metrological traceability to the international system of units N2 - The quantification of the sulphur mass fraction in pure copper and copper alloys by GDMS and LA-ICP-MS revealed a lack of traceability mainly due to a lack of suitable certified reference materials for calibrating the instruments. Within this study GDMS and LA-ICP-MS were applied as routine analytical tools to quantify sulphur in copper samples by applying reference materials as calibrators, which were characterized for their sulphur mass fraction by IDMS beforehand. Different external calibration strategies were applied including a matrix cross type calibration. Both techniques with all calibration strategies were validated by using certified reference materials (others than those used for calibration) and good agreement with the reference values was achieved except for the matrix cross type calibration, for which the agreement was slightly worse. All measurement results were accompanied by an uncertainty statement. For GDMS, the relative expanded (k = 2) measurement uncertainty ranged from 3% to 7%, while for LA-ICP-MS it ranged from 11% to 33% when applying matrix-matched calibration in the sulphur mass fraction range between 25 mg kg-1 and 1300 mg kg-1. For cross-type calibration the relative expanded (k = 2) measurement uncertainty need to be increased to at least 12% for GDMS and to at least 54% for LA-ICP-MS to yield metrological compatibility with the reference values. The so obtained measurement results are traceable to the international system of units (SI) via IDMS reference values, which is clearly illustrated by the unbroken chain of calibrations in the metrological traceability scheme. KW - Sulfur KW - Copper KW - GDMS KW - LA-ICP-MS KW - Uncertainty KW - Traceability KW - SI PY - 2021 DO - https://doi.org/10.1039/d1ja00137j SN - 0267-9477 VL - 36 IS - 11 SP - 2404 EP - 2414 PB - Royal Society of Chemistry AN - OPUS4-53412 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Loehr, Konrad A1 - Jakubowski, Norbert A1 - Wanka, Antje Jutta A1 - Traub, Heike A1 - Panne, Ulrich T1 - Quantification of metals in single cells by LA-ICP-MS comparison of single spot analysis and imaging N2 - LA-ICP-MS is increasingly used for single cell analysis in two different detection modes using either the imaging mode with subcellular resolution or alternatively single spot analysis of cells with a larger laser spot size. This study compares the analytical figures of merit of both detection modes (signal to noise, precision, accuracy, throughput), as well as ease of operation and data evaluation. Adherent 3T3 fibroblast cells were stained with two metal dyes (mDOTA-Ho, Ir-DNA-intercalator) and several dozen cells were measured using both modes. We found a ten times higher throughput for single spot analysis, which has as well a straightforward data analysis, shortening the total analysis time further. The signal to noise ratio for single spot analysis was found to be slightly better compared to the signal to noise of pixels in imaging. The mean metal intensity per single cell differed by only 10% between both modes and obtained distributions were found to show no statistically significant differences. Using matrix matched calibration based on standards spotted onto nitrocellulose membrane, we achieved detection limits (10 σ) of 12 fg for Ir and 30 fg for Ho and quantified 57 ± 35 fg Ir and 1,192 ± 707 fg Ho per single cell. Compared to a conventional ICP-MS measurement of a digest of ~60,000 cells, 54 % of Ir content and 358 % Ho content was found using quantitative LA-ICP-MS. The difference might be a consequence of the two metal dyes binding to different structures of the cell and therefore might behave differently in sample preparation for conventional and LA-ICP-MS. KW - Single cell analysis KW - LA-ICP-MS PY - 2018 DO - https://doi.org/10.1039/c8ja00191j SN - 0267-9477 SN - 1364-5544 VL - 33 IS - 9 SP - 1579 EP - 1587 PB - Royal Society of Chemistry AN - OPUS4-45903 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schwaar, Timm A1 - Lettow, Maike A1 - Remmler, Dario A1 - Börner, H. G. A1 - Weller, Michael G. T1 - Efficient Screening of Combinatorial Peptide Libraries by Spatially Ordered Beads Immobilized on Conventional Glass Slides N2 - Screening of one-bead-one-compound (OBOC) libraries is a proven procedure for the identification of protein-binding ligands. The demand for binders with high affinity and specificity towards various targets has surged in the biomedical and pharmaceutical field in recent years. The traditional peptide screening involves tedious steps such as affinity selection, bead picking, sequencing, and characterization. Herein, we present a high-throughput “all-on-one chip” system to avoid slow and technically complex bead picking steps. On a traditional glass slide provided with an electrically conductive tape, beads of a combinatorial peptide library are aligned and immobilized by application of a precision sieve. Subsequently, the chip is incubated with a fluorophore-labeled target protein. In a fluorescence scan followed by matrix-assisted laser desorption/ionization (MALDI)-time of flight (TOF) mass spectrometry, high-affinity binders are directly and unambiguously sequenced with high accuracy without picking of the positive beads. The use of an optimized ladder sequencing approach improved the accuracy of the de-novo sequencing step to nearly 100%. The new technique was validated by employing a FLAG-based model system, identifying new peptide binders for the monoclonal M2 anti-FLAG antibody, and was finally utilized to search for IgG-binding peptides. In the present format, more than 30,000 beads can be screened on one slide. KW - Peptide library KW - HTS KW - Target KW - MALDI KW - Mass spectrometry KW - Biochip KW - Lab-on-a-Chip KW - Array KW - Screening KW - Ladder sequencing KW - Binder KW - Pharmaceutical PY - 2019 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-478973 UR - https://www.mdpi.com/2571-5135/8/2/11 DO - https://doi.org/10.3390/ht8020011 VL - 8 IS - 2 SP - 1 EP - 15 PB - MDPI CY - Basel AN - OPUS4-47897 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fernández, L. A1 - Lin, Z. A1 - Schneider, Rudolf A1 - Esteves, V. I. A1 - Cunha, Â. A1 - Tomé, J. P. C. T1 - Antimicrobial Photodynamic Activity of Cationic Nanoparticles Decorated with Glycosylated Photosensitizers for Water Disinfection N2 - The antimicrobial photodynamic approach has been demonstrated as an efficient and sustainable process for the eradication of microbial pathogens. In this work, silica-coated Magnetite nanoparticles (NPs) were used as carriers of glycosylated porphyrins and phthalocyanines. Their subsequent cationization resulted in the production of stable antimicrobial photosensitizing materials, effective against E. coli. Suspensions of the photocatalysts in water present bimodal size distributions formed by big clusters and small NPs with hydrodynamic diameters between 8 and 38 nm. The presence of small NPs in the suspensions is related to an effective photodynamic inactivation (PDI) of E. coli cells. Glycosylation of the PS showed a positive effect on the PDI performance, which could be related to a higher accumulation of the photocatalyst over the bacterial cell membrane. In addition, these biocidal agents proved to be photostable and their photoactive performance decreased only between 23% and 28% upon 5 PDI cycles, mostly because of the loss of material between cycles, which makes them promising materials for water disinfection purposes. KW - Photodynamische Inaktivierung KW - E. coli KW - Photokatalyse KW - Bakterien KW - Nanopartikel KW - Photosensibilisator KW - Porphyrin KW - Phthalocyanin PY - 2018 DO - https://doi.org/10.1002/cptc.201700169 SN - 2367-0932 VL - 2 IS - 7 SP - 596 EP - 605 PB - Wiley VHC CY - Weinheim AN - OPUS4-45684 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Riedel, Soraya A1 - Hülagü, Deniz A1 - Bennet, Francesca A1 - Carl, Peter A1 - Flemig, Sabine A1 - Schmid, Thomas A1 - Schenk, J. A. A1 - Hodoroaba, Vasile-Dan A1 - Schneider, Rudolf T1 - Electrochemical Immunomagnetic Ochratoxin A Sensing: Steps Forward in the Application of 3,3’,5,5’- Tetramethylbenzidine in Amperometric Assays N2 - Electrochemical methods offer great promise in meeting the demand for user-friendly on-site devices for Monitoring important parameters. The food industry often runs own lab procedures, for example, for mycotoxin analysis, but it is a major goal to simplify analysis, linking analytical methods with smart technologies. Enzyme-linked immunosorbent assays, with photometric detection of 3,3’,5,5’-tetramethylbenzidine (TMB),form a good basis for sensitive detection. To provide a straightforward approach for the miniaturization of the detectionstep, we have studied the pitfalls of the electrochemical TMB detection. By cyclic voltammetry it was found that the TMB electrochemistry is strongly dependent on the pH and the electrode material. A stable electrode response to TMB could be achieved at pH 1 on gold electrodes. We created a smartphonebased, electrochemical, immunomagnetic assay for the detection of ochratoxin A in real samples, providing a solid basis forsensing of further analytes. KW - Ochratoxin A KW - Amperometry KW - Cyclic voltammetry KW - Electrochemistry KW - Immunoassay PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-530421 DO - https://doi.org/10.1002/celc.202100446 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. VL - 8 IS - 13 SP - 2597 EP - 2606 AN - OPUS4-53042 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schmidt, Stephan A1 - Hoffmann, Holger A1 - Garbe, L.-A. A1 - Schneider, Rudolf T1 - Liquid chromatography-tandem mass spectrometry detection of diclofenac and related compounds in water samples N2 - A frequently studied environmental contaminant is the active substance diclofenac, which is removed insufficiently in sewage treatment plants. Since its inclusion in the watch list of the EU Water Framework Directive, the concentrations in surface waters will be determined throughout Europe. For this, still, more precise analytical methods are needed. As a reference, HPLC-MS is frequently employed. One of the major metabolites is 4’-hydroxydiclofenac (4’-OH-DCF). Also, diclofenac lactam is important for assessing degradation and transformation. Aceclofenac (ACF), the glycolic acid ester of diclofenac is used as a drug, too, and could potentially be cleaved to yield diclofenac again. In various sewage treatment plant influent samples, diclofenac, 4’-OH-DCF, DCF lactam and ACF could be determined with detection limits of 3 µg/L, 0.2 µg/L, 0.17 µg/L and 10 ng/L, respectively. KW - Diclofenac KW - Festphasenextraktion KW - HPLC-MS KW - Probenvorbereitung KW - Kläranlage KW - Abwasser KW - Zyklisierung KW - Diclofenac KW - Solid-phase extraction KW - Sewage treatment plant KW - Cyclization KW - Sample treatment KW - HPLC–MS PY - 2018 DO - https://doi.org/10.1016/j.chroma.2018.01.037 SN - 0021-9673 SN - 1873-3778 VL - 1538 SP - 112 EP - 116 PB - Elsevier B.V. CY - Amsterdam, NL AN - OPUS4-44183 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schmidt, Stephan A1 - Hoffmann, Holger A1 - Garbe, Leif-Alexander A1 - Harrer, Andrea A1 - Steiner, Markus A1 - Himly, Martin A1 - Schneider, Rudolf T1 - Re-assessment of monoclonal antibodies against diclofenac for their application in the analysis of environmental waters N2 - The non-steroidal anti-inflammatory drug (NSAID) diclofenac (DCF) is an important environmental contaminant occurring in surface waters all over the world, because, after excretion, it is not adequately removed from wastewater in sewage treatment plants. To be able to monitor this pollutant, highly efficient analytical methods are needed, including immunoassays. In a medical research project, monoclonal antibodies against diclofenac and its metabolites had been produced. Based on this monoclonal anti-DCF antibody, a new indirect competitive enzyme-linked immunosorbent assay (ELISA) was developed and applied for environmental samples. The introduction of a spacer between diclofenac and the carrier protein in the coating conjugate led to higher sensitivity. With a test midpoint of 3 mg L−1 and a measurement range of 1–30 mg L−1, the system is not sensitive enough for direct analysis of surface water. However, this assay is quite robust against matrix influences and can be used for wastewater. Without adjustment of the calibration, organic solvents up to 5%, natural organic matter (NOM) up to 10 mg L−1, humic acids up to 2.5 mg L−1, and salt concentrations up to 6 g L−1 NaCl and 75 mg L−1 CaCl2 are tolerated. The antibody is also stable in a pH range from 3 to 12. Cross-reactivity (CR) of 1% or less was determined for the metabolites 40-hydroxydiclofenac (40-OH-DCF), 5-hydroxydiclofenac (5-OH-DCF), DCF lactam, and other NSAIDs. Relevant cross-reactivity occurred only with an amide derivative of DCF, 6-aminohexanoic acid (DCF-Ahx), aceclofenac (ACF) and DCF methyl ester (DCF-Me) with 150%, 61% and 44%, respectively. These substances, however, have not been found in samples. Only DCF-acyl glucuronide with a cross-reactivity of 57% is of some relevance. For the first time, photodegradation products were tested for cross-reactivity. With the ELISA based on this antibody, water samples were analysed. In sewage treatment plant effluents, concentrations in the range of 1.9–5.2 mg L−1 were determined directly, with recoveries compared to HPLC-MS/MS averaging 136%. Concentrations in lakes ranged from 3 to 4.4 ng L−1 and were, after pre-concentration, determined with an average recovery of 100% KW - Antikörper KW - Immunoassay KW - Wasser PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-602280 DO - https://doi.org/10.1039/d3ay01333b VL - 16 IS - 21 SP - 3349 EP - 3363 PB - Royal Society of Chemistry (RSC) CY - London AN - OPUS4-60228 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Becker, Roland T1 - Non-invasive cancer detection using volatile biomarkers: Is urine superior to breath? N2 - In recent years numerous reports have highlighted the options of chemical breath analysis with regard to noninvasive cancer detection. Certain volatile organic compounds (VOC) supposedly present in higher amounts or in characteristic patterns have been suggested as potential biomarkers. However, so far no clinical application based on a specific set of compounds appears to exist. Numerous reports on the capability of sniffer dogs and sensor arrays or electronic noses to distinguish breath of cancer patients and healthy controls supports the concept of genuine cancer-related volatile profiles. However, the actual compounds responsible for the scent are completely unknown and there is no correlation with the potential biomarkers suggested on basis of chemical trace analysis. It is outlined that specific features connected with the VOC analysis in breath – namely small concentrations of volatiles, interfering background concentrations, considerable sampling effort and sample instability, impracticability regarding routine application - stand in the way of substantial progress. The underlying chemicalanalytical challenge can only be met considering the severe susceptibility of VOC determination to these adverse conditions. Therefore, the attention is drawn to the needs for appropriate quality assurance/quality control as the most important feature for the reliable quantification of volatiles present in trace concentration. Consequently, the advantages of urine as an alternative matrix for volatile biomarker search in the context of diagnosing lung and other cancers are outlined with specific focus on quality assurance and practicability in clinical chemistry. The headspace over urine samples as the VOC source allows adapting gas chromatographical procedures well-established in water analysis. Foremost, the selection of urine over breath as non-invasive matrix should provide considerably more resilience to adverse effects during sampling and analysis. The most important advantage of urine over breath is seen in the option to partition, dispense, mix, spike, store, and thus to dispatch taylor-made urine samples on demand for quality control measures. Although it is still open at this point if cancer diagnosis supported by non-invasively sampled VOC profiles will ultimately reach clinical application the advantages of urine over breath should significantly facilitate urgently required steps beyond the current proof-of-concept stage and towards standardisation. KW - VOC KW - Breath KW - Urine KW - Lung cancer KW - Volatile organic compounds PY - 2020 DO - https://doi.org/10.1016/j.mehy.2020.110060 VL - 143 SP - 110060 PB - Elsevier Ltd. AN - OPUS4-51066 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Junge, Florian A1 - Wittwer, Philipp A1 - Sommerfeld, Thomas A1 - Gehrenkemper, Lennart A1 - Zoister, Christian A1 - Nickl, Philip A1 - Koch, Matthias A1 - Meermann, Björn A1 - Haag, Rainer T1 - Adsorber Charge Dominates over Hydrophobic or Fluorophilic Functionalization in Influencing Adsorption of PFCA onto Polystyrene Resins N2 - A systematic series of industrial-relevant polystyrene-based anion exchange resins that are functionalized with hydro- or fluorocarbon chains are compared regarding their adsorption behavior toward perfluorocarboxylic acids (PFCA) in respect to their charge, chain length, and type of chain. The results clearly show the dominance of electrostatic interactions in the adsorption process as uncharged adsorber materials showed no adsorption at all. In contrast, the charged adsorber materials showed in general a PFCA removal of 80% to 30% over the experiment depending on effluent fraction. Unexpectedly, for perfluorobutanoic acid (PFBA) the highest removal rate is found with consistently >90%. Despite observing significant benefits in the adsorption of PFCA for fluoroalkylated adsorbers in comparison to their non-fluorinated counterparts, this effect of fluoroalkylation is comparatively small and can not be clearly attributed to fluorophilic interactions between the fluoroalkyl chains. These findings help clarifying that the introduction of fluorocarbon moieties in adsorber materials is not necessary in order to remove fluorocarbon molecules from the environment. KW - PFAS KW - Remediation KW - Adsorption KW - Fluorophilic interactions PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-601883 DO - https://doi.org/10.1002/admi.202400199 SN - 2196-7350 SP - 1 EP - 10 PB - John Wiley & Sons CY - Hoboken, New Jersey, USA AN - OPUS4-60188 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -