TY - CONF A1 - Gehrenkemper, Lennart T1 - Detection of PFAS pollution in environmental samples - A Fast & sensitive PFAS sum parameter method using HR CS GFMAS N2 - The substance class of per- and polyfluorinated alkyl substances (PFAS) comprises more than 5300 organic compounds. PFAS are completely fluorinated on at least one carbon atom. They are associated with negative impacts on human and animal health, are extremely persistent in the environment, and bioaccumulate along food chains. Therefore, PFAS are classified as emerging pollutants. At the same time, their physicochemical properties make them attractive for use in diverse technical applications. They are both hydrophobic and lipophobic and show high thermal as well as chemical resistance due to the strong C-F bond. First regulations of some PFAS in combination with the technically excellent properties generated an innovation pressure and led to an enormous increase in the number of fluorinated substitution compounds. Due to the increasing complexity of this substance class, target analysis is not able to cover such a variety and multitude of analytes. Therefore, a suitable PFAS sum parameter method is necessary for an accurate detection of PFAS pollution in the environment, the identification of PFAS hotspots and an evaluation of appropriate remediation measures. Here we provide insights into the current state of PFAS sum parameter development and present our latest results on method development for the quantitative analysis of PFAS as extractable organically bound fluorine (EOF) in environmental samples using high-resolution molecular absorption spectrometry (HR-CS-GFMAS). For this purpose, we optimized the extraction of PFAS from different solid matrices with simultaneous separation of inorganic fluoride. For quantification resulting extracts were measured using a fluorine specific HR-CS-GFMAS method. By adding gallium salt solutions as modifiers in HR-CS-GFMAS, fluorine can be indirectly quantified very selectively by the in situ formation of GaF with low limits of quantification (instrumental LOQ c(F) < 3 µg/L). Here we will show results from real soil samples from sites with and without known contamination. T2 - 6. Doktorandenseminar des Deutschen Arbeitskreises für Analytische Spektroskopie (DAAS) CY - Online meeting DA - 20.09.2021 KW - High resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) KW - Soil KW - Per- and polyfluorinated alkyl substances (PFASs) KW - Extractable organically bound fluorine (EOF) KW - Solid-liquid extraction PY - 2021 AN - OPUS4-53333 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Meermann, Björn T1 - Elemental Analysis Methods in Material- & Environmental Analysis N2 - Materials in contact with the environment release e.g., metal-ions, elemental species and/or (nano-)particles. Once these species and/or particles are released, they are ingested by organisms and cells and thus, might have a negative impact on the environment. Thus, identification as well as quantification of potentially harmful substances is of utmost importance and highly needed to assess ecotoxicological impact of (emerging) pollutants. The oral presentation provides an overview on the power of elemental analytical techniques, in particular ICP-MS as well as HR-CS-GFMAS in environmental research. Current research topics from Division 1.1 @ BAM - Inorganic trace analysis will be highlighted: i) Elemental Speciation & Isotope analysis - new tools: Among elemental species separation and quantification, one of the main challenges in environmental elemental speciation analysis is the distinction between anthropogenic and natural elemental species. The on-line combination of elemental speciation and isotope analysis combines “the best from both worlds” - species specific isotopic information becomes available. As an application example the analysis of current anti-fouling agents via CE/MCICP-MS will be highlighted. ii) HR-CS-GFMAS for PFC analysis: Per- and polyfluorinated compounds (PFC) are emerging contaminants in particular in soil and surface water samples. Due to the large number of compounds (>4700), target analytical methods are not sufficient and sum parameter methods for organically bound fluorine are highly needed. High resolution-continuum source-graphite furnace molecular absorption spectroscopy (HR-CS-GFMAS) based methods for organically bound fluorine analysis will be presented. Application examples (soil and surface water) will be highlighted. iii) Single cell-ICP-ToF-MS - ecotox. assessment: Single cell and single organism analysis for e.g. ecotoxicological/medicinal assessment are hot topics in the research field of ICP-MS. In particular ICP-ToF-MS is a powerful, emerging techniques in terms of single cell/particle analysis. Automated single cell/diatom-ICP-ToF-MS as a potential tool in ecotoxicological testing will be presented. T2 - Eingeladener Fachvortrag Kolloquium AG Bleiner - EMPA CY - Online meeting DA - 03.11.2021 KW - HR-CS-GFMAS KW - SC-ICP-ToF-MS KW - Speciation analysis PY - 2021 AN - OPUS4-53707 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Meermann, Björn T1 - HR-CS-GFMAS as a promising new tool for the analysis of organo-fluorine in environmental samples N2 - New HR-CS-GFMAS based methods for organically bound fluorine detection in environmental samples. T2 - CRC Colloquium - CRC 1349 Fluorine-Specific Interactions CY - Online meeting DA - 21.06.2021 KW - HR-CS-GFMAS KW - Organically bound fluorine KW - Fluorine sum parameter KW - Environmental analysis PY - 2021 AN - OPUS4-52844 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Riedel, Soraya T1 - Immunomagnetic ochratoxin a assay with electrochemical 3,3′,5,5′ tetramethylbenzidine detection N2 - The demand for miniaturized analytical devices monitoring important parameters in the food and medical industry has increased strongly in the past decades. With fast progress, smart technologies are finding their way into our everyday life. For the future, it is, therefore, a major goal to also link analytical methods with smart technologies to create user-friendly on-site devices. In food industry the monitoring of harmful substances such as dioxins, heavy metals or mycotoxins plays a key role, since the European Commission prescribes legal limits for various food products and beverages[1]. Therefore, companies often have their own laboratories and trained personnel. For one of the most abundant and toxic mycotoxins, Ochratoxin A (OTA) we want to present an electrochemical detection system in which the read-out can be performed with a smartphone connected via Bluetooth to a miniaturized potentiostat. The recognition of OTA is performed with specific antibodies in a competitive assay format. Anti-OTA-antibodies were captured on magnetic beads on which the competitive binding between OTA and an OTA horseradish peroxidase (HRP) tracer was performed. To quantify OTA, the enzymatic reaction of the tracer with 3,3′,5,5′-tetramethylbenzidine (TMB) and H2O2 is employed. Oxidized TMB, which is enzymatically produced by the reduction of H2O2, is quantified by amperometry with screen-printed electrodes in a custom-made flow system. The results of the electrochemical detection method are in good correlation with the photometric detection of TMB. To demonstrate the applicability, we tested our system with OTA-spiked beer and performed the measurement via smartphone. T2 - 23rd JCF-Frühjahrssymposium 2021 CY - Online meeting DA - 29.03.2021 KW - Ochratoxin A KW - Cyclic voltammetry KW - Amperometry KW - Electrochemistry PY - 2021 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. AN - OPUS4-52390 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Riedel, Soraya T1 - Enzyme activity determination of human monoamine oxidase B (Mao B) by amperometric hydrogen peroxide detection N2 - Parkinson’s disease (PD) is one of the most common neurodegenerative disorders worldwide. About 0.3 % of the global population and approximately 2 % of people older than 80 years are affected by PD. Monoamine oxidase B (Mao B) is an enzyme, which is a drug target in Parkinson’s disease (PD), since it is involved in dopamine metabolism. Several Mao B inhibitors are well established as medication for PD patients. However, the medical treatment is only little personalized since the monitoring of the patients Mao B activity is complex and requires sophisticated laboratory equipment. Here a sensorial Mao B activity determination system has been developed which has potential in the personalization of the medical PD treatment. The enzyme activity is quantified by amperometric detection of enzymatically produced H2O2. Therefore, the enzyme is enriched from the solution via cellulose particles which are functionalized with antibodies against human Mao B. The successful capturing of the enzyme can be verified by SDS-PAGE. For activity determination the enzyme is brought in contact with a suitable substrate - here benzylamine. Selectivity of the amperometric hydrogen peroxide detection in the presence of co-reactants has been verified. Within the time span of 30 min, a linear dependency of enzymatically produced H2O2 with the substrate incubation time can be observed. This allows the evaluation of the Mao B activity. The results have been correlated to an optical detection method. Furthermore, the method has been tested for different amounts of enzyme used in the experiments and found to be sensitive enough for Mao B analysis in blood samples. T2 - European Biosensor Symposium 2021 CY - Online meeting DA - 09.03.2021 KW - Monoamine oxidase B KW - Amperometry KW - Prussian blue KW - Parkinsons's disease KW - Screen-printed electrodes PY - 2021 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. AN - OPUS4-52270 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von Törne, Wipert T1 - Einblicke in das Transformationsverhalten von Warfarin N2 - Warfarin ist seit den 1950er Jahren einer der bekanntesten Gerinnungshemmer. Beim Menschen wird es bis heute zur Vorbeugung von Thrombose und der Behandlung von Vorhofflimmern und Arrhythmie eingesetzt. Doch schon einige Jahre vor der FDA Zulassung (U. S. Food and Drug Administration) als Medikament wurde es zur Bekämpfung von Nagetieren eingesetzt. Bis heute wurde es nur teilweise durch weitaus wirksamere antikoagulante Rodentizide (ARs) der zweiten Generation, sogenannte Superwarfarine, ersetzt. In der Fachliteratur finden sich zahlreiche Beispiele für sekundäre und tertiäre Vergiftung mit ARs bei Wildtieren und anderen Nichtzieltieren. Bisher ist relativ wenig über die Bioakkumulation, Persistenz und Toxizität von ARs und deren Transformationsprodukten (TPs) in der terrestrischen und aquatischen Umwelt, sowie der Nahrungskette bekannt. Um Einblicke in mögliche Transformations- und Abbauprozesse von Warfarin unter umweltrelevanten Bedingungen zu erhalten und um diese Einflüsse zu simulieren, wurde hierzu eine elektrochemische Reaktorzelle in Kopplung mit einem Massenspektrometer verwendet. So können mögliche Phase-I-Metabolite und sogenannte „oxidation products“ vorausgesagt werden. Hinsichtlich des Verhaltens während der Wasseraufbereitung in Kläranlagen wurden technisch relevante Prozesse wie Chlorung, UV-Bestrahlung und Ozonung in Modellen simuliert, um hier potenzielle Abbauwege und -prozesse aufzuklären. Gebildete Transformationsprodukte wurden daher mittels unterschiedlicher chromatografischer Verfahren getrennt und massenspektrometrisch identifiziert, charakterisiert und verglichen. T2 - 49. Jahrestagung der GUS CY - Online meeting DA - 24.03.2021 KW - Warfarin KW - Transformationsprodukte PY - 2021 AN - OPUS4-52354 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Vogl, Jochen T1 - New applications and reference materials in isotope analysis N2 - The presentations gives a short introduction into isotope analysis illustrated by examples from archaeology, food provenancing and metrology. In the main part current projects on the provenance determination of cement, the intercalibration of Mg isotope reference materials and the development of a new calibration approach for isotope measurements are presented. T2 - Seminar der Abteilung 1 CY - Online meeting DA - 19.01.2021 KW - Isotope reference material KW - Absolute isotope ratio KW - Metrology KW - Traceability KW - Triple isotope fractionation KW - Archaeometry KW - Cement PY - 2021 AN - OPUS4-52460 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Simon, Fabian T1 - A fast and simple extraction method for the determination of PFASs in soil samples - HR-CS-GFMAS a new screening tool N2 - Here, we describe a fast and simple extraction method for the determination of per- and polyfluorinated alkyl substances (PFASs) utilizing extractable organic fluorine (EOF) sum parameter analysis and high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) in soil samples. After extraction and separation of inorganic and organic fluorinated compounds, organically bound fluorinated compounds were indirectly determined via the molecular absorption of gallium mono-fluoride at 211.248 nm. The implementation of the decisions of the “Stockholm Convention on persistent organic pollutants” as well as the “Protocol on Persistent Organic Pollutants of the Convention on Long-Range Transboundary Air Pollution” of the UNECE include the reduction respectively the elimination of PFASs in the environment. Currently, regulations aim to target single compounds (mostly C8-PFAS). But the PFAS group includes over 4700 substances, which are potentially persistent and toxic, thus target analytical approaches are not suitable for a holistic approach investigating the PFAS pollutant situation. Furthermore, forbidden PFASs are substituted by short-chain PFASs, thus the number of unknow substances is steadily increasing. For this reason, sum parameter approaches are more suitable to investigate and assess the pollution situation as well as derive exposure limits. Our developed analytical method was successfully applied to determine PFASs in soil samples from a polluted site. In view of steadily increasing numbers of PFAS substances, our method will make an important contribution in assessing the pollution situation as well as support policy makers in deriving exposure limits for PFAS in the future. T2 - ENSOr - International Workshop on Emerging policy challenges on new soil contaminants CY - Online Meeting DA - 06.05.2021 KW - PFAS KW - HR-CS-GFMAS KW - Soil KW - Sediment PY - 2021 AN - OPUS4-52608 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Meermann, Björn T1 - Elemental analytical methods in material- & environmental analysis N2 - Materials in contact with the environment release e.g., metal-ions, elemental species and/or (nano-)particles. Once these species and/or particles are released, they are ingested by organisms and cells and thus, might have a negative impact on the environment. Thus, identification as well as quantification of potentially harmful substances is of utmost importance and highly needed to assess ecotoxicological impact of (emerging) pollutants. The oral presentation provides an overview on the power of elemental analytical techniques, in particular: ICP-MS as well as HR-CS-GFMAS in environmental research. Current research topics will be highlighted: i) Elemental Speciation analysis of anti-fouling agents via CE/MC-ICP-MS; ii) Per- and polyfluorinated compounds in surface waters via high resolution-continuum source-graphite furnace molecular absorption spectroscopy (HR-CS-GFMAS); iii) Single cell and single organism analysis via sc-ICP-ToF-MS and ETV/ICP-MS for ecotoxicological assessment. T2 - Analytical Colloquium at MERCK CY - Online meeting DA - 19.04.2021 KW - HR-CS-GFMAS KW - ICP-ToF-MS KW - ETV/ICP-MS PY - 2021 AN - OPUS4-52566 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Vogl, Jochen T1 - The triple isotope calibration approach BT - A new calibration approach for obtaining absolute isotope ratios of multi-isotopic elements N2 - The calibration of isotope ratio measurements is an ongoing challenge since instrumental isotope fractionation (IIF) has been detected in mass spectrometry (MS). There is a variety of approaches which either bypass IIF such as delta measurements or refer to reference materials (RMs) and thus shifting the problem of calibration to somebody else: the RM producer. For certifying isotope RMs with absolute isotope ratios only a few approaches are available, namely the isotope mixture approach, the double spike approach, the mass bias regression model and total evaporation in TIMS. All of them require either enriched isotopes, isotope RMs of another element or an RM for correcting residual error. As the enriched isotopes required for the isotope mixture and the double spike approach need to be fully characterized beforehand, all mentioned calibration approaches require a standard. Here, a new and standard-free calibration approach for obtaining absolute isotope ratios of multi-isotopic elements has been developed. The underlying principle is that each MS suffers from IIF and thus yields a specific isotope fractionation line in a three-isotope diagram. When applying a second MS featuring a different ionization mechanism, we obtain a second isotope fractionation line with a different slope in the same three-isotope diagram. In both cases the absolute isotope ratios range somewhere on the isotope fractionation line. Consequentially, the intersect of both lines yield the absolute isotope ratios of the measured sample. This theory has been tested by measuring Cd and Pb isotope ratios of suitable isotope RMs with a TIMS and an ICP-MS, both equipped with multi-collector array. During the measurements the ionization conditions were changed such that different extent of the isotope fractionation has been achieved. With the resulting data set the theory described above could be verified. The obtained absolute isotope ratios were metrologically compatible with the certified isotope ratios. The remaining average bias of -5 ‰ can be reduced with further improvements. The calibration approach is universal and can be applied to any multi-isotopic element and it is not limited by the type of the mass spectrometer. T2 - Virtual Goldschmidt 2021 CY - Online meeting DA - 04.07.2021 KW - Absolute isotope ratio KW - Traceability KW - Metrology KW - Calibration KW - Uncertainty KW - Triple isotope fractionation PY - 2021 AN - OPUS4-53023 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -