TY - JOUR A1 - Vogl, Jochen T1 - The triple-isotope calibration approach: a universal and standard-free calibration approach for obtaining absolute isotope ratios of multi-isotopic elements N2 - The theory of a new calibration approach for obtaining absolute isotope ratios of multi-isotopic elements without the use of any standard has been developed. The calibration approach basically uses the difference in the instrumental isotope fractionation of two different types of mass spectrometers, leading to two different fractionation lines in a three-isotope diagram. When measuring the same sample with both mass spectrometers, the different fractionation lines have one point in common: this is the ‘true’ logarithmized isotope ratio pair of the sample. Thus, the intersection of both fractionation lines provides us with the absolute isotope ratios of the sample. This theory has been tested in practice by measuring Cd and of Pb isotope ratios in the certified reference materials BAM-I012 and NIST SRM981 by thermal ionization mass spectrometry and by inductively coupled plasma mass spectrometry while varying the ionization conditions for both mass spectrometers. With this experiment, the theory could be verified, and absolute isotope ratios were obtained, which were metrologically compatible with the certified isotope ratios. The so-obtained absolute isotope ratios are biased by − 0.5% in average, which should be improved with further developments of the method. This calibration approach is universal, as it can be applied to all elements with three or more isotopes and it is not limited to the type of mass spectrometers applied; it can be applied as well to secondary ion mass spectrometry or others. Additionally, this approach provides information on the fractionation process itself via the triple-isotope fractionation exponent θ. KW - Triple isotope fractionation KW - Absolute isotope ratio KW - Mass spectrometry KW - Calibration KW - Uncertainty PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-516364 DO - https://doi.org/10.1007/s00216-020-03050-4 VL - 413 IS - 3 SP - 821 EP - 826 PB - Springer Verlag AN - OPUS4-51636 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Vogl, Jochen T1 - The triple isotope calibration approach BT - A new calibration approach for obtaining absolute isotope ratios of multi-isotopic elements N2 - The calibration of isotope ratio measurements is an ongoing challenge since instrumental isotope fractionation (IIF) has been detected in mass spectrometry (MS). There is a variety of approaches which either bypass IIF such as delta measurements or refer to reference materials (RMs) and thus shifting the problem of calibration to somebody else: the RM producer. For certifying isotope RMs with absolute isotope ratios only a few approaches are available, namely the isotope mixture approach, the double spike approach, the mass bias regression model and total evaporation in TIMS. All of them require either enriched isotopes, isotope RMs of another element or an RM for correcting residual error. As the enriched isotopes required for the isotope mixture and the double spike approach need to be fully characterized beforehand, all mentioned calibration approaches require a standard. Here, a new and standard-free calibration approach for obtaining absolute isotope ratios of multi-isotopic elements has been developed. The underlying principle is that each MS suffers from IIF and thus yields a specific isotope fractionation line in a three-isotope diagram. When applying a second MS featuring a different ionization mechanism, we obtain a second isotope fractionation line with a different slope in the same three-isotope diagram. In both cases the absolute isotope ratios range somewhere on the isotope fractionation line. Consequentially, the intersect of both lines yield the absolute isotope ratios of the measured sample. This theory has been tested by measuring Cd and Pb isotope ratios of suitable isotope RMs with a TIMS and an ICP-MS, both equipped with multi-collector array. During the measurements the ionization conditions were changed such that different extent of the isotope fractionation has been achieved. With the resulting data set the theory described above could be verified. The obtained absolute isotope ratios were metrologically compatible with the certified isotope ratios. The remaining average bias of -5 ‰ can be reduced with further improvements. The calibration approach is universal and can be applied to any multi-isotopic element and it is not limited by the type of the mass spectrometer. T2 - Virtual Goldschmidt 2021 CY - Online meeting DA - 04.07.2021 KW - Absolute isotope ratio KW - Traceability KW - Metrology KW - Calibration KW - Uncertainty KW - Triple isotope fractionation PY - 2021 AN - OPUS4-53023 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - INPR A1 - Tchipilov, Teodor A1 - Meyer, Klas A1 - Weller, Michael G. T1 - Quantitative 1H nuclear magnetic resonance (qNMR) of aromatic amino acids for protein quantification N2 - qNMR is a valuable technique for metrological studies due to the uniformity of its signal response for all chemical species of an isotope of interest, which enables compound-independent calibration. However, protein quantification remained challenging as large molecules produce wide, low-intensity signals that reduce the already low sensitivity. Combining qNMR with the hydrolysis of protein samples into amino acids circumvents many of these issues and facilitates the use of NMR spectroscopy for absolute protein and peptide quantification.In this work, different conditions have been tested for quantifying aromatic amino acids and proteins. First, we examined the pH-based signal shifts in the aromatic region. The preferable pH depends on the selection of the amino acids for quantification and which internal standard substance should be used to avoid peak overlap. Several aromatic compounds, such as terephthalic acid, sulfoisophthalic acid, and benzene tricarboxylic acid, have been applied as internal standards. The quantification of amino acids from an amino acid standard, as well as from a certified reference material (bovine serum albumin), was performed. Using the first two suggested internal standards, recovery was ~ 97 % for histidine, phenylalanine, and tyrosine at a concentration of approximately 1 mM in solution. Acidic hydrolysis of a certified reference material (CRM) of bovine serum albumin (BSA) and subsequent quantification of Phe and Tyr yielded recoveries of 98 ± 2 and 88 ± 4 %, respectively, at a protein concentration of 16 g/L or 250 µM. KW - Amino acid analysis KW - AAA KW - Protein hydrolysis KW - Metrology KW - Traceability KW - Reference materials KW - Internal standards KW - Calibration PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-564520 DO - https://doi.org/10.20944/preprints202211.0569.v1 SP - 1 EP - 13 PB - MDPI CY - Basel AN - OPUS4-56452 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tchipilov, Teodor A1 - Meyer, Klas A1 - Weller, Michael G. T1 - Quantitative 1H nuclear magnetic resonance (qNMR) of aromatic amino acids for protein quantification N2 - Hydrolysis of protein samples into amino acids facilitates the use of NMR spectroscopy for protein and peptide quantification. Different conditions have been tested for quantifying aromatic amino acids and proteins. The pH-dependent signal shifts in the aromatic region of amino acid samples were examined. A pH of 12 was found to minimize signal overlap of the four aromatic amino acids. Several aromatic compounds, such as terephthalic acid, sulfoisophthalic acid, and benzene tricarboxylic acid, were applied as internal standards. The quantification of amino acids from an amino acid standard was performed. Using the first two suggested internal standards, recovery was ~97% for histidine, phenylalanine, and tyrosine at a concentration of approximately 1 mM in solution. Acidic hydrolysis of a certified reference material (CRM) of bovine serum albumin (BSA) and subsequent quantification of Phe and Tyr yielded recoveries of 98% ± 2% and 88% ± 4%, respectively, at a protein concentration of 16 g/L or 250 µM. KW - Amino acid analysis KW - Aromatic amino acid analysis KW - AAA KW - AAAA KW - Protein hydrolysis KW - Hydrochloric acid KW - Metrology KW - Traceability KW - NIST KW - Reference materials KW - Internal standard KW - Calibration KW - Compound-independent calibration KW - Histidine KW - Tyrosine KW - Tryptophan KW - Phenylalanine KW - Terephthalic acid KW - Benzene-1,3,5-tricarboxylic acid KW - Bovine serum albumin (BSA) KW - Quantitative protein analysis KW - Phenylketonuria PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-570943 DO - https://doi.org/10.3390/mps6010011 VL - 6 IS - 1 SP - 1 EP - 13 PB - MDPI CY - Basel, Schweiz AN - OPUS4-57094 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Völker, Tobias A1 - Gornushkin, Igor B. T1 - Mass and mole fractions in calibration-free LIBS N2 - This technical note highlights the fact that CF-LIBS algorithms work in mole fractions, while results of spectrochemical analysis are usually reported in mass fractions or mass percent. Ignoring this difference and not converting mole fractions to mass fractions can lead to errors in reported concentrations determined by the CF-LIBS method and inadequate comparison of these concentrations with certified concentrations. Here, the key points of the CF-LIBS algorithm are reproduced and the formulae for converting a mole fraction to a mass fraction and vice versa are given. Several numerical examples are also given, which show that the greater the difference between the molar mass of an individual element in a sample and the average molar mass, the greater the discrepancy between the mole and mass fractions. KW - Spectroscopy KW - Analytical Chemistry KW - LIBS KW - Calibration PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-598611 DO - https://doi.org/10.1039/d4ja00028e SN - 0267-9477 VL - 39 IS - 4 SP - 1030 EP - 1032 PB - Royal Society of Chemistry (RSC) AN - OPUS4-59861 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rosenkranz, D. A1 - Kriegel, F. L. A1 - Mavrakis, E. A1 - Pergantis, S. A. A1 - Reichardt, P. A1 - Tentschert, J. A1 - Jakubowski, N. A1 - Laux, P. A1 - Panne, Ulrich A1 - Luch, A. T1 - Versatile dual-inlet sample introduction system for multi-mode single particle inductively coupled plasma mass spectrometry N2 - Metal-containing nanoparticles (NP) can be characterized with inductively coupled plasma mass spectrometers (ICP-MS) in terms of their size and number concentration by using the single-particle mode of the instrument (spICP-MS). The accuracy of measurement depends on the setup, operational conditions of the instrument and specific parameters that are set by the user. The transport efficiency of the ICP-MS is crucial for the quantification of the NP and usually requires a reference material with homogenous size distribution and a known particle number concentration. Currently, NP reference materials are available for only a few metals and in limited sizes. If particles are characterized without a reference standard, the results of both size and particle number may be biased. Therefore, a dual-inlet Setup for characterizing nanoparticles with spICP-MS was developed to overcome this problem. This setup is based on a conventional introduction system consisting of a pneumatic nebulizer (PN) for nanoparticle solutions and a microdroplet Generator (μDG) for ionic calibration solutions. A new and flexible interface was developed to facilitate the coupling of μDG, PN and the ICP-MS system. The interface consists of available laboratory components and allows for the calibration, nanoparticle (NP) characterization and cleaning of the arrangement, while the ICP-MS instrument is still running. Three independent analysis modes are available for determining particle size and number concentration. Each mode is based on a different calibration principle. While mode I (counting) and mode III (μDG) are known from the literature, mode II (sensitivity), is used to determine the transport efficiency by inorganic ionic Standard solutions only. It is independent of NP reference materials. The μDG based inlet system described here guarantees superior analyte sensitivities and, therefore, lower detection limits (LOD). The size dependent LODs achieved are less than 15 nm for all NP (Au, Ag, CeO2) investigated. KW - Spectrometer KW - Reference KW - Calibration PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-536121 DO - https://doi.org/10.3791/61653 SN - 1940-087X IS - 163 SP - 1 EP - 19 PB - MyJoVE Corporation CY - Cambridge, MA, USA AN - OPUS4-53612 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hoffmann, V. A1 - Uhlemann, M. A1 - Richter, Silke A1 - Pfeifer, jens T1 - Calibration capacity of hot-pressed hydrogen standards for glow discharge optical emission and mass spectrometry N2 - Mixed copper and titanium hydride powder was hot-pressed and characterized by Carrier Gas Hot Extraction, XRay Diffraction, Thermal Gravimetric Analysis coupled with Mass Spectrometry, and Scanning Electron Microscopy. The hot-pressed and five conventional samples were applied for calibration of hydrogen in Glow Discharge Optical Emission and Mass Spectrometry. Up to the introduction of 15 ng/s hydrogen the Emission yield model is useful in Glow Discharge Optical Emission Spectrometry. A correlation between saturation and even reversal of the emission yield of the spectral lines H121, H486 and H656 and low sputtering rates was found. Hydrogen effects exist for the spectral lines of Cu(II) 219 and Ti(I) 399. In Glow Discharge Mass Spectrometry, a linear dependency of the 1H ion current on the sputtered mass per time exists over the total range of hydrogen content investigated. Hydrogen effects also exist for the sensitivity of 48Ti and 63Cu. The sputtering rate of two-phase materials depends linearly on the sputtered mass per time of one phase, which allows the sputtering rate of two-phase materials with known composition to be predicted. KW - Hot-pressing KW - GD-OES KW - GD-MS KW - Calibration KW - Hydrogen KW - Titanium hydride KW - Sputtering KW - Two-phase system PY - 2021 DO - https://doi.org/10.1016/j.sab.2020.106039 VL - 176 SP - 106039 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-52074 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute A1 - Fallisch, A. A1 - Petrov, E. P. A1 - Salhany, R. A1 - Forthman, C. A1 - Guttenberg, Z. A1 - Nitschke, R. T1 - Spectral fluorescence standards for the calibration and performance validation of fluorescence microscopes N2 - The standardization and calibration of fluorescence microscopy have become increasingly vital due to the wide-spread use of in life and materials sciences. As the demand for reliable and user-friendly methods to assess micro-scope performance grows, universal calibration tools accessible to both researchers and vendors are needed. To support the standardization of characterization methods in microscopy, it is crucial to provide calibration tools together with standardized operating procedures for their effective implementation. The public-funded project "FluMiKal"* develops calibration tools in the shape of typical microscopic slides to assess key parameters such as spatial resolution, point spread function, spectral sensitivity, linearity and sensitivity of the detection system. The focus is on creating calibration tools that are user-friendly, robust, and versatile in their application. This work addresses the critical parameter of wavelength-dependent spectral sensitivity, which affects the meas-ured signals from the instrument side, yielding instrument-specific data and instrument aging-induced changes over time. For this purpose, μ-slides from ibidi with six channels are used, allowing them to be filled with different solutions containing molecular or nanoscale fluorophores with well-characterized absorption and fluorescence properties. The certified spectral fluorescence standards BAM-F003, F004, F005, and F007 assessed provided as ethanolic solutions by the Federal Institute for Materials Research and Testing (BAM), cover a broad spectral range from the blue to the near-infrared [1], [2]. Dye-based slide prototypes have been used to determine the spectral sensitivity of confocal microscopes from different vendors with various detector types by acquiring the spectral data of the BAM dyes under standardized measurement conditions, demonstrating the applicability of this concept. Proof-of-concept experiments could demonstrate the proper sealing of the slides. Further experiments will explore long-term stability and their potential as standards for relative intensity calibrations. * FluMiKal is funded by the Federal Ministry for Economic Affairs and Climate Action, Germany (WIPANO FKZ 03TN0047B) [1] doi: 10.1007/4243_2008_028. [2] doi: 10.1007/s00216-024-05723-w. T2 - European Light Microscopy Initiative - ELMI 2025 CY - Heidelberg, Germany DA - 03.06.2025 KW - Fluorescence KW - Advanced material KW - Validation KW - Calibration KW - Method comparison KW - Reference material KW - Dye KW - Fluorescence standard KW - Microscopy KW - CLSM KW - Imaging PY - 2025 AN - OPUS4-64206 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Auxillos, J. A1 - Crouigneau, R. A1 - Li, Y.-F. A1 - Dai, Y. A1 - Stigliani, A. A1 - Tavernaro, Isabella A1 - Resch-Genger, Ute A1 - Sandelin, A. A1 - Marie, R. A1 - Pedersen, S. F. T1 - Spatially resolved analysis of microenvironmental gradient impact on cancer cell phenotypes N2 - Despite the physiological and pathophysiological significance of microenvironmental gradients, e.g., for diseases such as cancer, tools for generating such gradients and analyzing their impact are lacking. Here, we present an integrated microfluidic-based workflow that mimics extracellular pH gradients characteristic of solid tumors while enabling high-resolution live imaging of, e.g., cell motility and chemotaxis, and preserving the capacity to capture the spatial transcriptome. Our microfluidic device generates a pH gradient that can be rapidly controlled to mimic spatiotemporal microenvironmental changes over cancer cells embedded in a 3D matrix. The device can be reopened allowing immunofluorescence analysis of selected phenotypes, as well as the transfer of cells and matrix to a Visium slide for spatially resolved analysis of transcriptional changes across the pH gradient. This workflow is easily adaptable to other gradients and multiple cell types and can therefore prove invaluable for integrated analysis of roles of microenvironmental gradients in biology. KW - Bioimaging KW - Fluorescence KW - Cell KW - Cancer KW - Method KW - Microfluids KW - Model KW - Calibration KW - Sensor KW - Ph KW - Probe KW - Workflow PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-604631 DO - https://doi.org/10.1126/sciadv.adn3448 VL - 19 IS - 18 SP - 1 EP - 17 AN - OPUS4-60463 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Richter, Maria A1 - Güttler, Arne A1 - Pauli, Jutta A1 - Vogel, K. A1 - Homann, Christian A1 - Würth, Christian A1 - Resch-Genger, Ute T1 - Extending certified spectral fluorescence standards for the calibration and performance validation of fluorescence instruments to the NIR—closing the gap from 750 to 940 nm with two novel NIR dyes N2 - Fluorescence techniques such as fluorescence spectroscopy, microfluorometry, and fluorescence microscopy, providing spectral, intensity, polarization, and lifetime information, are amongst the most broadly utilized analytical methods in the life and materials sciences. However, the measured fluorescence data contain sample- and instrument-specific contributions, which hamper their comparability across instruments and laboratories. Comparable, instrument-independent fluorescence data require the determination of the fluorescence instrument’s wavelength-dependent spectral responsivity, also termed emission correction curve, for the same instrument settings as those used for the fluorescence measurements as a prerequisite for the subsequent correction of the measured instrument-specific data. Such a spectral correction is essential for the performance comparison of different fluorescent labels and reporters, quantitative fluorescence measurements, the determination of the fluorescence quantum yield, and the spectroscopic measure for the fluorescence efficiency of a fluorophore. Simple-to-use tools for obtaining emission correction curves are chromophore-based reference materials (RMs), referred to as fluorescence standards, with precisely known, preferably certified instrument-independent fluorescence spectra. However, for the increasingly used near-infrared (NIR) wavelength region >700 nm, at present, no spectral fluorescence standards are available. To close this gap, we developed two novel spectral fluorescence standards, BAM F007 and BAM-F009, with broad emission bands from about 580 to 940 nm in ethanolic solution. These liquid fluorescence standards currently under certification, which will be released in 2025, will expand the wavelength range of the already available certified Calibration Kit BAM F001b-F005b from about 300–730 to 940 nm. In this research article, we will detail the criteria utilized for dye and matrix selection and the homogeneity and stability tests accompanying dye certification as well as the calculation of the wavelength-dependent uncertainty budgets of the emission spectra BAM F007 and BAM-F009, determined with the traceably calibrated BAM reference spectrofluorometer. These fluorescence standards can provide the basis for comparable fluorescence measurements in the ultraviolet, visible, and NIR for the fluorescence community. KW - Quality assurance KW - Reference material KW - Fluorescence KW - Dye KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - NIR KW - Instrument performance validation PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-626317 DO - https://doi.org/10.1007/s00216-024-05723-w SN - 1618-2650 SP - 1 EP - 15 PB - Springer AN - OPUS4-62631 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tavernaro, Isabella A1 - Sander, P. A1 - Andresen, Elina A1 - Schedler, U. A1 - Resch-Genger, Ute T1 - Potentiometric and Optical Titration for Cost- Efficient Quantification of Surface Functional Groups on Silica Nanoparticles N2 - Surface chemistry of engineered nanomaterials (NMs) plays a critical role not only in determining their interactions with the environment but also in their stability, safety, and functionality across diverse applications ranging from catalysis to biomedicine. Accurate quantification of surface functional groups (FGs) is therefore essential for quality control, risk assessment, and performance optimization.[1] However, many existing analytical techniques are either cost-intensive, require specialized instrumentation, or lack scalability for routine use. In this study, we present a comparative evaluation of potentiometric and optical titration as two simple, cost-efficient, and automatable methods for quantifying surface functional groups on a variety of surface-modified silica nanoparticles (SiO₂ NPs). These NPs were chosen as they are among the most frequently utilized engineered NMs in the life and material sciences. Potentiometric titration, based on pH monitoring during acid-base neutralization, offers a direct and label-free approach to determine the total amount of FGs. Optical titration provides a complementary method with potential for high-throughput screening. To examine the accuracy and robustness of our stepwise-optimized workflows and the achievable relative standard deviations (RSDs), measurements were performed by multiple operators in two laboratories. Method validation was conducted through cross-comparison with traceable, chemo-selective quantitative nuclear magnetic resonance spectroscopy (qNMR) and thermogravimetric analysis (TGA). A comparison with optical assays highlights the importance of measuring both quantities for comprehensive characterization of surface-modified NMs.[2] A combined NM surface analysis using optical assays and pH titration will simplify quality control of NM production processes and stability studies, and can yield large datasets for NM grouping in sustainable and safe(r)-by-design studies. T2 - eMRS Fall Meeting 2025 CY - Warsaw, Poland DA - 15.09.2025 KW - Fluorescence KW - Advanced material KW - Synthesis KW - Characterization KW - Nano KW - Particle KW - Silica KW - Surface analysis KW - Validation KW - qNMR KW - Fluram assay KW - Functional group KW - Quantification KW - Potentiometry KW - Amino groups KW - Fluorescamine KW - Calibration KW - Method comparison PY - 2025 AN - OPUS4-64205 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Güttler, Arne T1 - Certified Reference Materials for the Quantification and Standardization of Fluorescence-based Measurements N2 - The size and shape of photoluminescence signals is affected by wavelength-, polarization-, and time-dependent instrumentspecific contributions and the compound- and environment-specific photoluminescence quantum yield. The former hamper the comparability of fluorescence measurements performed on different measuring devices. The commonly relatively done determination of the performance parameter requires suitable quantum yield standards with well-known. The performance of such measurements is, e.g., described in the written standard IEC 62607 currently revised. T2 - Colloquium für Optische Spektrometrie 2025 CY - Jena, Germany DA - 24.09.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Advanced material KW - Calibration KW - Characterization KW - Fluorescence quantum yield KW - Phosphor KW - Absolute KW - Integrating sphere spectroscopy KW - Dye KW - Standardization KW - Reference material KW - Interlaboratory comparison KW - Uncertainty PY - 2025 AN - OPUS4-64213 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - Measurements of Photoluminescence Quantum Yields of Scattering LED Converter Materials N2 - How to Get it Right with the Absolute Measurement of Photoluminescence Quantum Yields of Scattering LED Converter Materials Saskia Fiedler+,a, Florian Frenzel+,a, Christian Würth a, Isabella Tavernaro a, Michelle Grüne c, Stefan Schweizer c,d, Axel Engel e, and Ute Resch-Genger a* a Division Biophotonics, Federal Institute for Materials Research and Testing (BAM), Richard-Willstaetter-Strasse 11, D-12489 Berlin, Germany; email: ute.resch@bam.de b Present address: Photonic Materials, NWO-Institute AMOLF, Science Park 104, 1098 XG Amsterdam, The Netherlands c Faculty of Electrical Engineering, South Westphalia University of Applied Sciences, Lübecker Ring 2, 59494, Soest, Germany d Fraunhofer Application Center for Inorganic Phosphors, Branch Lab of Fraunhofer Institute for Microstructure of Materials and Systems IMWS, Lübecker Ring 2, 59494, Soest, Germany e Schott AG Technical Services, Hattenbergstrasse 10, D-55122 Mainz, Germany Optical measurements of scattering materials such as luminescent nano- and microparticles and phosphors dispersed in liquid and solid matrices play an important role in energy conversion, solid-state lighting, medical diagnostics, and security barcoding. A key performance parameter is the photoluminescence quantum yield QY, i.e., the number of emitted per number of absorbed photons. QY of transparent luminophore solutions can be obtained relative to a fluorescence QY standard of known QY, meanwhile available as certified reference materials.[1] The determination of QY of scattering liquid and solid samples like nanoparticle dispersions, phosphors, and optoceramics requires, however, absolute measurements with an integrating sphere setup. Despite the need for reliable absolute QY measurements, no interlaboratory comparison (ILCs) on measurement uncertainties has been performed and scattering standards with known QY are not available. We present the results of an ILC of 3 labs from academia and industry on measurements of transparent and scattering dye solutions and solid phosphors and converter materials like YAG:Ce optoceramics with commercial stand-alone integrating sphere setups of different illumination and detection geometries. Special emphasis was dedicated to the influence of measurement geometry, optical properties of the blank for determining the number of incident photons absorbed by the sample, and sample-specific surface roughness. Matching QY values could be obtained for transparent dye solutions and scattering dispersions with a blank with scattering properties closely matching those of the sample, QY measurements of optoceramic samples with different blanks revealed substantial differences of more than 20 %. Based on our data, we recommend non-absorbing blank materials with a high reflectivity (>95 %) such as a 2 mm-thick PTFE target placed on the sample holder as blanks. T2 - eMRS CY - Strasbourg, France DA - 26.05.2025 KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Quantum yield KW - NIR KW - Characterization KW - Electron microscopy KW - Film KW - Integrating sphere spectroscopy KW - Calibration KW - Lifetime KW - Advanced materials KW - LED converter KW - YAG:Ce KW - ILC KW - Measurement uncertainty KW - Absolute quantum yield PY - 2025 AN - OPUS4-63327 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Resch-Genger, Ute A1 - Richter, Maria A1 - Güttler, Arne A1 - Pauli, Jutta A1 - Vogel, K. A1 - Homann, Christian A1 - Würth, Christian T1 - Extending Certified Spectral Fluorescence Standards for the Calibration and Performance Validation of Fluorescence Instruments to the NIR – Closing the Gap from 750 nm to 940 nm with Two Novel NIR Dyes N2 - Fluorescence methods provide spectral, intensity, polarization, and lifetime information, which contain sample- and instrument-specific contributions. Fluorescence data, comparable across instruments and laboratories, require validated calibration procedures and certified fluorescence standards. KW - Quality assurance KW - Reference material KW - Fluorescence KW - Dye KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - NIR KW - Instrument performance validation PY - 2025 SP - 1 EP - 4 PB - Springer Nature CY - London AN - OPUS4-62739 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Abad Andrade, Carlos Enrique T1 - The use of reference materials to improve the calibration strategy in glow discharge optical emission spectroscopy with machine learning N2 - Glow discharge optical emission spectroscopy (GD-OES) is a technique for the analysis of solids such as metals, semiconductors, and ceramics. A low-pressure glow discharge plasma is applied in this system, which ‘sputters’ and promotes the sample atoms to a higher energy state. When the atoms return to their ground state, they emit light with characteristic wavelengths, which a spectrometer can detect. Thus, GD-OES combines the advantages of ICP-OES with solid sampling techniques, which enables it to determine the bulk elemental composition and depth profiles. However, direct solid sampling methods such as glow-discharge spectroscopy require reference materials for calibration due to the strong matrix effect. Reference materials are essential when the accuracy and reliability of measurement results need to be guaranteed to generate confidence in the analysis. These materials are frequently used to determine measurement uncertainty, validate methods, suitability testing, and quality assurance. In addition, they guarantee that measurement results can be compared to recognized reference values. Unfortunately, the availability of certified reference materials suited to calibrate all elements in different matrix materials is limited. Therefore various calibration strategies and the preparation of traceable matrix-matched calibration standards will be discussed. Machine learning is an essential component of the growing field of data science. Through statistical methods, algorithms are trained to make classifications or predictions, uncovering key insights within data mining projects. Therefore, it was tried in our work to combine GD-OES with machine learning strategies to establish a new and robust calibration model, which can be used to identify the elemental composition and concentration of metals from a single spectrum. For this purpose, copper reference materials from different manufacturers, which contain various impurity elements, were investigated using GD-OES. The obtained spectra information are evaluated with different algorithms (e.g., gradient boosting and artificial neural networks), and the results are compared and discussed in detail. T2 - Winter Conference on Plasma Chemistry 2022 CY - Tucson, AZ, USA DA - 17.01.2022 KW - GDOES KW - Machine learning KW - Reference materials KW - Calibration KW - Cooper PY - 2022 AN - OPUS4-56497 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tavernaro, Isabella A1 - Rajotte, Isabelle A1 - Thibeault, Marie-Pier A1 - Sander, Philipp C. A1 - Kodra, Oltion A1 - Lopinski, Gregory A1 - Radnik, Jörg A1 - Johnston, Linda J. A1 - Brinkmann, Andreas A1 - Resch-Genger, Ute T1 - Quantifying surface groups on aminated silica nanoparticles of different size, surface chemistry, and porosity with solution NMR, XPS, optical assays, and potentiometric titration N2 - We assessed the quantification of surface amino functional groups (FGs) for a large set of commercial and custom-made aminated silica nanoparticles (SiO2 NPs) with sizes of 20–100 nm, prepared with different sol–gel routes, different amounts of surface amino FGs, and different porosity with four methods providing different, yet connected measurands in a bilateral study of two laboratories, BAM and NRC, with the overall aim to develop standardizable measurements for surface FG quantification. Special emphasis was dedicated to traceable quantitative magnetic resonance spectroscopy (qNMR) performed with dissolved SiO2 NPs. For the cost efficient and automatable screening of the amount of surface amino FGs done in a first step of this study, the optical fluorescamine assay and a potentiometric titration method were utilized by one partner, i.e., BAM, yielding the amount of primary amino FGs accessible for the reaction with a dye precursor and the total amount of (de)protonatable FGs. These measurements, which give estimates of the minimum and maximum number of surface amino FGs, laid the basis for quantifying the amount of amino silane molecules with chemo-selective qNMR with stepwise fine-tuned workflows, involving centrifugation, drying, weighting, dissolution, measurement, and data evaluation steps jointly performed by BAM and NRC. Data comparability and relative standard deviations (RSDs) obtained by both labs were used as quality measures for method optimization and as prerequisites to identify method-inherent limitations to be later considered for standardized measurement protocols. Additionally, the nitrogen (N) to silicon (Si) ratio in the near-surface region of the SiO2 NPs was determined by both labs using X-ray photoelectron spectroscopy (XPS), a well established surface sensitive analytical method increasingly utilized for microparticles and nano-objects which is currently also in the focus of international standardization activities. Overall, our results underline the importance of multi-method characterization studies for quantifying FGs on NMs involving at least two expert laboratories for effectively identifying sources of uncertainty, validating analytical methods, and deriving NM structure–property relationships. KW - Advanced Materials KW - Amino Groups KW - Calibration KW - Characterization KW - Functional groups KW - Method Comparison KW - Nano Particle KW - Validation KW - XPS KW - Optical Assay KW - Quantification KW - Surface Analysis KW - Reference Materials KW - Synthesis KW - Fluorescence PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-649992 DO - https://doi.org/10.1039/d5na00794a VL - 7 IS - 21 SP - 6888 EP - 6900 PB - Royal Society of Chemistry AN - OPUS4-64999 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tavernaro, Isabella A1 - Abram, Sarah-Luise A1 - Homann, Christian A1 - Scholtz, Lena A1 - Radnik, Jörg A1 - Resch-Genger, Ute T1 - Validated and standardized measurements and quantification of surface functionalities on nanoparticles N2 - Surface-functionalized organic and inorganic engineered nanomaterials (NMs) are widely applied in the life and materials sciences. NM performance depends on key factors such as particle size and shape, crystal phase, morphology, chemical composition, and surface chemistry, i.e., surface coatings, functional groups (FGs), and ligands.1 The latter controls their processability and interaction with the environment and largely their possible toxicity. Thus, methods for FG quantification are important tools for quality control of NM production processes and can foster the sustainable development of functional and safe(r) NMs. This underlines the importance of validated and standardized analytical methods for surface analysis and reference materials.2 This encouraged us to explore simple and versatile tools for quantifying common bioanalytically relevant FGs such as optical assays, electrochemical titration methods, quantitative nuclear magnetic resonance spectroscopy (qNMR), and X-Ray photoelectron spectroscopy (XPS) and to perform a first interlaboratory comparison (ILC) on surface FG quantification.3,4 In a follow-up ILC, BAM and NRC explored qNMR sample preparation, measurement, and data evaluation protocols for commercial and custom-made aminated SiO2 NPs with sizes of 20-100 nm, different amounts of surface amino FGs, and different porosity.5,6 First, the number of amino FGs accessible for a dye reporter was determined with a cost-efficient, automated optical fluorescamine assay. Then, qNMR workflows and protocols were stepwise fine-tuned. The qNMR ILC was complemented by joint XPS measurements. BAM also examined the applicability of fast and automatable potentiometric titrations to screen the total amount of (de)protonable FGs on aminated SiO2 NPs. Our results underline the need to evaluate protocols for FG quantification in ILCs and the advantages of multi-method characterization strategies for efficient method cross validation. T2 - Surface and Micro/Nano Analysis Working Group CY - Paris, France DA - 08.04.2025 KW - Quality assurance KW - Reference material KW - Fluorescence KW - Nano KW - Particle KW - Surface analysis KW - XPS KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - QNMR KW - Potentiometry PY - 2025 AN - OPUS4-62969 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute T1 - How to get it right with photoluminescence measurements in the UV/VIS/NIR N2 - Optical measurements of transparent solutions of organic dyes and semiconductor quantum dots and scattering materials such as luminescent nanocomposites and microparticles and phosphors dispersed in liquid and solid matrices play an important role in energy conversion, solid-state lighting, medical diagnostics, and security barcoding. A key performance parameter is the photoluminescence quantum yield QY, i.e., the number of emitted per number of absorbed photons. QY of transparent luminophore solutions can be obtained relative to a fluorescence QY standard of known QY. Such relative QY measurements require a calibrated spectrofluorometer. Here, we address typical sources of uncertainty for luminescence measurements and relative and absolute QY measurements from the instrument and material side and provide recommendations to circumvent or at least minimize them for transparent and scattering samples in the UV/vis/NIR/SWIR. T2 - Workshop "Photoluminescence Spectroscopy and its application in Materials Research" CY - Vienna, Austria DA - 19.02.2026 KW - Quality assurance KW - Reference material KW - Fluorescence KW - Dye KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - NIR KW - Instrument performance validation PY - 2026 AN - OPUS4-65575 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -