TY - JOUR A1 - Ecke, Alexander A1 - Westphalen, Tanja A1 - Hornung, J. A1 - Voetz, M. A1 - Schneider, Rudolf T1 - A rapid magnetic bead-based immunoassay for sensitive determination of diclofenac JF - Analytical and Bioanalytical Chemistry N2 - Increasing contamination of environmental waters with pharmaceuticals represents an emerging threat for the drinking water quality and safety. In this regard, fast and reliable analytical methods are required to allow quick countermeasures in case of contamination. Here, we report the development of a magnetic bead-based immunoassay (MBBA) for the fast and cost-effective determination of the analgesic diclofenac (DCF) in water samples, based on diclofenac-coupled magnetic beads and a robust monoclonal anti-DCF antibody. A novel synthetic strategy for preparation of the beads resulted in an assay that enabled for the determination of diclofenac with a significantly lower limit of detection (400 ng/L) than the respective enzyme-linked immunosorbent assay (ELISA). With shorter incubation times and only one manual washing step required, the assay demands for remarkably shorter time to result (< 45 min) and less equipment than ELISA. Evaluation of assay precision and accuracy with a series of spiked water samples yielded results with low to moderate intra- and inter-assay variations and in good agreement with LC–MS/MS reference analysis. The assay principle can be transferred to other, e.g., microfluidic, formats, as well as applied to other analytes and may replace ELISA as the standard immunochemical method. KW - Immunoassay KW - Magnetic beads KW - Diclofenac KW - Water analysis KW - LC-MS/MS PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-542346 DO - https://doi.org/10.1007/s00216-021-03778-7 SN - 1618-2650 VL - 414 SP - 1563 EP - 1573 PB - Springer CY - Heidelberg AN - OPUS4-54234 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Riedel, Soraya A1 - Jaut, Valerie A1 - Schneider, Rudolf T1 - Ergometrine sensing in rye flour by a magnetic bead-based immunoassay followed by flow injection analysis with amperometric detection JF - Talanta N2 - A certain group of mycotoxins, the ergot alkaloids, has caused countless deaths throughout human history. They are found in rye and other cereals and ingesting contaminated foods can cause serious health problems. To identify contaminated food exceeding the legal limits for ergot alkaloids, a portable and cost-effective test system is of great interest to the food industry. Rapid analysis can be achieved by screening for a marker compound, for which we chose ergometrine. We developed a magnetic bead-based immunoassay for ergometrine with amperometric detection in a flow injection system using a handheld potentiostat and a smartphone. With this assay a limit of detection of 3 nM (1 μg/L) was achieved. In spiked rye flour, ergometrine levels from 25 to 250 μg/kg could be quantified. All results could be verified by optical detection. The developed assay offers great promise to meet the demand for on-site ergometrine detection in the food industry. KW - Ergot alkaloids KW - Amperometry KW - Magnetic beads KW - Immunoassay KW - Food analysis KW - Fow injection analysis KW - Mycotoxins PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-565906 DO - https://doi.org/10.1016/j.talanta.2022.124172 SN - 0039-9140 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. VL - 254 IS - 124172 SP - 1 EP - 8 PB - Elsevier B.V. AN - OPUS4-56590 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ecke, Alexander A1 - Bohm, Konrad A1 - Schneider, Rudolf T1 - Magnetpartikelbasierte Immunoassays als vielseitiges Werkzeug für die Umweltanalytik JF - Mitteilungen der Fachgruppe Umweltchemie und Ökotoxikologie N2 - Der Bedarf nach schnellen und kostengünstigen Analysemethoden in der (Trink-)Wasseranalytik steigt mit dem zunehmenden Eintrag pharmazeutisch aktiver Substanzen in die Umwelt. Insbesondere die Kontamination mit Antibiotika erscheint neben ökotoxikologischen Effekten vor allem in Hinblick auf die Evolution resistenter Keime im Wasser besorgniserregend. Wir berichten hier über die Entwicklung eines magnetpartikelbasierten Immunoassays (MBBA) zur Detektion des Breitbandantibiotikums Amoxicillin (AMX) in Wasserproben, der im Vergleich zum entsprechenden Enzyme-linked Immunosorbent Assay (ELISA) eine wesentlich kürzere Analysendauer sowie einen breiteren Messbereich mit niedrigerer Nachweisgrenze aufweist. Das angewendete Assayprinzip ist dabei leicht auf weitere Analyten übertragbar und ermöglicht die Implementierung in einen Immunosensor. KW - Magnetpartikel KW - Amoxicillin KW - Immunoassay KW - Wasser KW - Antikörper PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-527366 UR - https://www.gdch.de/netzwerk-strukturen/fachstrukturen/umweltchemie-und-oekotoxikologie/publikationen/mitteilungen/archiv/2021.html SN - 1618-3258 VL - 27 IS - 2 SP - 48 EP - 51 PB - Gesellschaft Deutscher Chemiker (GDCh) CY - Frankfurt/Main AN - OPUS4-52736 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ramos, I. I. A1 - Carl, Peter A1 - Schneider, Rudolf A1 - Segundo, M. A. T1 - Automated lab-on-valve sequential injection ELISA for determination of carbamazepine JF - Analytica Chimica Acta N2 - The development of an automated miniaturized analytical system that allows for the rapid monitoring of carbamazepine (CBZ) levels in serum and wastewater is proposed. Molecular recognition of CBZ was achieved through its selective interaction with microbeads carrying anti-CBZ antibodies. The proposed method combines the advantages of the micro-bead injection spectroscopy and of the flow-based platform lab-on-valve for implementation of automatic immunosorbent renewal, rendering a new recognition surface for each sample. The sequential (or simultaneous) perfusion of CBZ and the horseradish peroxidase-labelled CBZ through the microbeads is followed by real-time on-column Monitoring of substrate (3,30,5,50-tetramethylbenzidine) oxidation by colorimetry. The evaluation of the initial oxidation rate and also the absorbance value at a fixed time point provided a linear response versus the logarithm of the CBZ concentration. Under the selected assay conditions, a single analysis was completed after only 11 min, with a quantification range between 1.0 and 50 µg L⁻¹. Detection of CBZ levels in undiluted wastewater samples was feasible after a simple filtration step while good recoveries were attained for spiked certified human serum, analyzed without sample clean-up. KW - Automation KW - Bead injection spectroscopy KW - Human serum KW - Microparticles KW - Therapeutic Drug Monitoring KW - Wastewater PY - 2019 DO - https://doi.org/10.1016/j.aca.2019.05.017 SN - 0003-2670 VL - 1076 SP - 91 EP - 99 PB - Elsevier CY - Amsterdam AN - OPUS4-48317 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Chapartegui-Arias, Ander A1 - Villajos Collado, José Antonio A1 - Myxa, Anett A1 - Beyer, Sebastian A1 - Falkenhagen, Jana A1 - Schneider, Rudolf A1 - Emmerling, Franziska T1 - Covalently Fluorophore-Functionalized ZIF‑8 Colloidal Particles as a Sensing Platform for Endocrine-Disrupting Chemicals Such as Phthalates Plasticizers JF - ACS Omega N2 - We present the optical sensing of phthalate Esters (PAEs), a group of endocrine-disrupting chemicals. The sensing takes place as changes in the fluorescence emission intensity of aminopyrene covalently bound to the organic ligands of the metal−organic framework compound ZIF-8. In the presence of PAEs, a quenching of the fluorescence emission is observed. We evaluated strategies to engineer colloidal size distribution of the sensing particles to optimize the sensory response to PAEs. A thorough characterization of the modified ZIF-8 nanoparticles included powder X-ray diffractometry, transmission electron microscopy, high-performance liquid chromatography, and photophysical characterization. The presented capability of the fluorophore-functionalized ZIF-8 to sense PAEs complements established methods such as chromatography-based procedures, which cannot be used on-site and paves the way for future developments such as hand-held quick sensing devices. KW - Sensing KW - MOF PY - 2019 DO - https://doi.org/10.1021/acsomega.9b01051 VL - 4 IS - 17 SP - 17090 EP - 17097 PB - ACS AN - OPUS4-49562 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ecke, Alexander A1 - Westphalen, Tanja A1 - Retzmann, Anika A1 - Schneider, Rudolf T1 - Factors affecting the hydrolysis of the antibiotic amoxicillin in the aquatic environment JF - Chemosphere N2 - The environmental fate of the frequently used broad-spectrum β-lactam antibiotic amoxicillin (AMX) is of high concern regarding the potential evolution of antimicrobial resistance (AMR). Moreover, it is known that AMX is prone to hydrolysis, yielding a variety of hydrolysis products (HPs) with yet unknown effects. Studies to identify those HPs and investigate their formation mechanisms have been reported but a long-term study on their stability in real water samples was missing. In this regard, we investigated the hydrolysis of AMX at two concentration levels in four distinct water types under three different storage conditions over two months. Concentrations of AMX and four relevant HPs were monitored by an LC-MS/MS method revealing pronounced differences in the hydrolysis rate of AMX in tap water and mineral water on the one hand (fast) and surface water on the other(slow). In this context, the occurrence, relative intensities, and stability of certain HPs are more dependent on the water type than on the storage condition. As clarified by ICP-MS, the main difference between the water types was the content of the metals copper and zinc which are supposed to catalyze AMX hydrolysis demonstrating an effective method to degrade AMX at ambient conditions. KW - β-lactam KW - Stability KW - Degradation KW - Hydrolysis products KW - LC-MS/MS KW - ICP-MS PY - 2023 DO - https://doi.org/10.1016/j.chemosphere.2022.136921 SN - 0045-6535 VL - 311 SP - 136921 PB - Elsevier Ltd. AN - OPUS4-56124 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fernández, L. A1 - Lin, Z. A1 - Schneider, Rudolf A1 - Esteves, V. I. A1 - Cunha, Â. A1 - Tomé, J. P. C. T1 - Antimicrobial Photodynamic Activity of Cationic Nanoparticles Decorated with Glycosylated Photosensitizers for Water Disinfection JF - ChemPhotoChem N2 - The antimicrobial photodynamic approach has been demonstrated as an efficient and sustainable process for the eradication of microbial pathogens. In this work, silica-coated Magnetite nanoparticles (NPs) were used as carriers of glycosylated porphyrins and phthalocyanines. Their subsequent cationization resulted in the production of stable antimicrobial photosensitizing materials, effective against E. coli. Suspensions of the photocatalysts in water present bimodal size distributions formed by big clusters and small NPs with hydrodynamic diameters between 8 and 38 nm. The presence of small NPs in the suspensions is related to an effective photodynamic inactivation (PDI) of E. coli cells. Glycosylation of the PS showed a positive effect on the PDI performance, which could be related to a higher accumulation of the photocatalyst over the bacterial cell membrane. In addition, these biocidal agents proved to be photostable and their photoactive performance decreased only between 23% and 28% upon 5 PDI cycles, mostly because of the loss of material between cycles, which makes them promising materials for water disinfection purposes. KW - Photodynamische Inaktivierung KW - E. coli KW - Photokatalyse KW - Bakterien KW - Nanopartikel KW - Photosensibilisator KW - Porphyrin KW - Phthalocyanin PY - 2018 DO - https://doi.org/10.1002/cptc.201700169 SN - 2367-0932 VL - 2 IS - 7 SP - 596 EP - 605 PB - Wiley VHC CY - Weinheim AN - OPUS4-45684 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Riedel, Soraya A1 - Hülagü, Deniz A1 - Bennet, Francesca A1 - Carl, Peter A1 - Flemig, Sabine A1 - Schmid, Thomas A1 - Schenk, J. A. A1 - Hodoroaba, Vasile-Dan A1 - Schneider, Rudolf T1 - Electrochemical Immunomagnetic Ochratoxin A Sensing: Steps Forward in the Application of 3,3’,5,5’- Tetramethylbenzidine in Amperometric Assays JF - ChemElectroChem N2 - Electrochemical methods offer great promise in meeting the demand for user-friendly on-site devices for Monitoring important parameters. The food industry often runs own lab procedures, for example, for mycotoxin analysis, but it is a major goal to simplify analysis, linking analytical methods with smart technologies. Enzyme-linked immunosorbent assays, with photometric detection of 3,3’,5,5’-tetramethylbenzidine (TMB),form a good basis for sensitive detection. To provide a straightforward approach for the miniaturization of the detectionstep, we have studied the pitfalls of the electrochemical TMB detection. By cyclic voltammetry it was found that the TMB electrochemistry is strongly dependent on the pH and the electrode material. A stable electrode response to TMB could be achieved at pH 1 on gold electrodes. We created a smartphonebased, electrochemical, immunomagnetic assay for the detection of ochratoxin A in real samples, providing a solid basis forsensing of further analytes. KW - Ochratoxin A KW - Amperometry KW - Cyclic voltammetry KW - Electrochemistry KW - Immunoassay PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-530421 DO - https://doi.org/10.1002/celc.202100446 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. VL - 8 IS - 13 SP - 2597 EP - 2606 AN - OPUS4-53042 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schmidt, Stephan A1 - Hoffmann, Holger A1 - Garbe, L.-A. A1 - Schneider, Rudolf T1 - Liquid chromatography-tandem mass spectrometry detection of diclofenac and related compounds in water samples JF - Journal of Chromatography A N2 - A frequently studied environmental contaminant is the active substance diclofenac, which is removed insufficiently in sewage treatment plants. Since its inclusion in the watch list of the EU Water Framework Directive, the concentrations in surface waters will be determined throughout Europe. For this, still, more precise analytical methods are needed. As a reference, HPLC-MS is frequently employed. One of the major metabolites is 4’-hydroxydiclofenac (4’-OH-DCF). Also, diclofenac lactam is important for assessing degradation and transformation. Aceclofenac (ACF), the glycolic acid ester of diclofenac is used as a drug, too, and could potentially be cleaved to yield diclofenac again. In various sewage treatment plant influent samples, diclofenac, 4’-OH-DCF, DCF lactam and ACF could be determined with detection limits of 3 µg/L, 0.2 µg/L, 0.17 µg/L and 10 ng/L, respectively. KW - Diclofenac KW - Festphasenextraktion KW - HPLC-MS KW - Probenvorbereitung KW - Kläranlage KW - Abwasser KW - Zyklisierung KW - Diclofenac KW - Solid-phase extraction KW - Sewage treatment plant KW - Cyclization KW - Sample treatment KW - HPLC–MS PY - 2018 DO - https://doi.org/10.1016/j.chroma.2018.01.037 SN - 0021-9673 SN - 1873-3778 VL - 1538 SP - 112 EP - 116 PB - Elsevier B.V. CY - Amsterdam, NL AN - OPUS4-44183 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schmidt, Stephan A1 - Hoffmann, Holger A1 - Garbe, Leif-Alexander A1 - Harrer, Andrea A1 - Steiner, Markus A1 - Himly, Martin A1 - Schneider, Rudolf T1 - Re-assessment of monoclonal antibodies against diclofenac for their application in the analysis of environmental waters JF - Analytical Methods N2 - The non-steroidal anti-inflammatory drug (NSAID) diclofenac (DCF) is an important environmental contaminant occurring in surface waters all over the world, because, after excretion, it is not adequately removed from wastewater in sewage treatment plants. To be able to monitor this pollutant, highly efficient analytical methods are needed, including immunoassays. In a medical research project, monoclonal antibodies against diclofenac and its metabolites had been produced. Based on this monoclonal anti-DCF antibody, a new indirect competitive enzyme-linked immunosorbent assay (ELISA) was developed and applied for environmental samples. The introduction of a spacer between diclofenac and the carrier protein in the coating conjugate led to higher sensitivity. With a test midpoint of 3 mg L−1 and a measurement range of 1–30 mg L−1, the system is not sensitive enough for direct analysis of surface water. However, this assay is quite robust against matrix influences and can be used for wastewater. Without adjustment of the calibration, organic solvents up to 5%, natural organic matter (NOM) up to 10 mg L−1, humic acids up to 2.5 mg L−1, and salt concentrations up to 6 g L−1 NaCl and 75 mg L−1 CaCl2 are tolerated. The antibody is also stable in a pH range from 3 to 12. Cross-reactivity (CR) of 1% or less was determined for the metabolites 40-hydroxydiclofenac (40-OH-DCF), 5-hydroxydiclofenac (5-OH-DCF), DCF lactam, and other NSAIDs. Relevant cross-reactivity occurred only with an amide derivative of DCF, 6-aminohexanoic acid (DCF-Ahx), aceclofenac (ACF) and DCF methyl ester (DCF-Me) with 150%, 61% and 44%, respectively. These substances, however, have not been found in samples. Only DCF-acyl glucuronide with a cross-reactivity of 57% is of some relevance. For the first time, photodegradation products were tested for cross-reactivity. With the ELISA based on this antibody, water samples were analysed. In sewage treatment plant effluents, concentrations in the range of 1.9–5.2 mg L−1 were determined directly, with recoveries compared to HPLC-MS/MS averaging 136%. Concentrations in lakes ranged from 3 to 4.4 ng L−1 and were, after pre-concentration, determined with an average recovery of 100% KW - Antikörper KW - Immunoassay KW - Wasser PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-602280 DO - https://doi.org/10.1039/d3ay01333b VL - 16 IS - 21 SP - 3349 EP - 3363 PB - Royal Society of Chemistry (RSC) CY - London AN - OPUS4-60228 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -