TY - JOUR A1 - Alcalde-Vasquez, R. A1 - Moreno-Pedraza, A. A1 - Rosas-Román, I. A1 - Guillén-Alonso, H. A1 - Riedel, Jens A1 - Partida-Martínez, L. P. A1 - Winkler, R. T1 - MoBiMS: A Modular Miniature Mass Analyzer for the Real-Time 2 Monitoring of Gases and Volatile Compounds in Biological Systems JF - Microchemical Journal N2 - Living organisms constantly interact with their environment, including through the exchange of gases. Respiration and the release and uptake of volatile organic compounds (VOCs) create dynamic microenvironments in biological systems. Studying the kinetics of volatiles in biological systems requires expensive instruments, and data analysis is challenging. Therefore, we aimed to design a minimal analytical device for measuring the composition of gaseous mixtures in realtime. We built the ‘Modular Biological Mass Spectrometer’ (MoBiMS) from 3D-printed parts and custom sensors to fit a wide array of experimental set-ups. We tested the chemical detection range and temporal resolution of the MoBiMS employing pure compounds and complex biological samples. Compounds with a higher than 0.4 mmHg vapor pressure and a molecular weight up to 154 g/mol were reliably sensed within seconds. The generated electron impact (EI) spectra were directly comparable with standard databases like the NIST EI library. Under a direct analysis approach, the MoBiMS identified the characteristic odor of banana (Musa sp.), that is, isoamyl acetate; tracked the dynamics of CO2 release while the Alka-Seltzer® reaction occurred showed the kinetics of the transient production and consumption of carbon dioxide during photosynthesis. MoBiMS also discriminated between volatile compounds ions coming from tobacco (Nicotiana benthamiana) leaves and the surrounding air through untargeted analysis. The small footprint of the MoBiMS and its relatively low energy consumption facilitate in situ analyses, as an additional gas supply is not necessary with EI ionization. The MoBiMS is easy to assemble, and its construction and operation are very cost-efficient compared to commercial devices. The analytical performance of the MoBiMS is suitable for real-time studies of biological systems, environmental monitoring, and medical diagnostics. KW - Mass spectrometry KW - Instrumentation KW - Miniaturization KW - Volatiles PY - 2022 DO - https://doi.org/10.1016/j.microc.2021.107090 SN - 0026-265X VL - 175 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-56528 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - van Wasen, S. A1 - You, Yi A1 - Beck, S. A1 - Riedel, Jens A1 - Volmer, D. A. T1 - Laser Ablation Secondary Electrospray Ionization for In Situ Mass Spectrometric Interrogation of Acoustically-Levitated Droplets JF - Analytical Chemistry N2 - The composition of acoustically levitated droplets was probed by a novel combination of mid-IR laser evaporation and subsequent postionization via secondary electrospray ionization. The combination of microliter samples and subnanoliter sampling provided time-resolved interrogation of droplets and allowed for a kinetic investigation of the laser-induced release of the analyte, which was found to strongly depend on the analytes. The observed substancespecific delayed release of the analytes permitted baseline-separated discrimination of the analytes, ideal for the study of complex samples. The additionally applied postionization scheme was found to enable efficient detection of small volatile compounds as well as peptides. The detection of small molecules and peptides occurred under very different sampling geometries, pointing to two distinct underlying ionization mechanisms. Overall, our results suggest that the experimental setup presented in this study can serve as a widely applicable platform to study chemical reactions in acoustically levitated droplets as model reactors. KW - Acoustic levitation KW - Mass spectrometry KW - Electrospray KW - Laser ablation PY - 2022 DO - https://doi.org/10.1021/acs.analchem.2c03800 SN - 0003-2700 VL - 2022 SP - 1 EP - 5 PB - ACS Publications CY - Washington AN - OPUS4-56531 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Epping, Ruben A1 - Bliesener, Lilly A1 - Weiss, Tilman A1 - Koch, Matthias T1 - Marker Substances in the Aroma of Truffles JF - Molecules N2 - The aim of this study was to identify specific truffle marker substances within the truffle aroma. The aroma profile of different truffle species was analyzed using static headspace sampling with gas chromatography mass spectrometry analysis (SHS/GC-MS). Possible marker substances were identified, taking the additional literature into account. The selected marker substances were tested in an experiment with 19 truffle dogs. The hypothesis “If trained truffle dogs recognize the substances as supposed truffles in the context of an experiment, they can be regarded as specific” was made. As it would be nearly impossible to investigate every other possible emitter of the same compounds to determine their specificity, this hypothesis was a reasonable approximation. We were interested in the question of what it is the dogs actually search for on a chemical level and whether we can link their ability to find truffles to one or more specific marker substances. The results of the dog experiment are not as unambiguous as could have been expected based on the SHS/GC-MS measurements. Presumably, the truffle aroma is mainly characterized and perceived by dogs by dimethyl sulfide and dimethyl disulfide. However, as dogs are living beings and not analytical instruments, it seems unavoidable that one must live with some degree of uncertainty regarding these results. KW - Truffle KW - Volatile organic compounds; KW - Gas chromatography KW - Mass spectrometry KW - Canine olfactometry PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-556116 DO - https://doi.org/10.3390/molecules27165169 SN - 1420-3049 VL - 27 IS - 16 SP - 1 EP - 19 PB - MDPI CY - Basel AN - OPUS4-55611 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Cruz-Alonso, M. A1 - Fernandez, B. A1 - Álvarez, L. A1 - González-Iglesias, H. A1 - Traub, Heike A1 - Jakubowski, Norbert A1 - Pereiro, R. T1 - Bioimaging of metallothioneins in ocular tissue sections by laser ablation-ICP-MS using bioconjugated gold nanoclusters as specific tags JF - Microchimica Acta N2 - An immunohistochemical method is described to visualize the distribution of metallothioneins 1/2 (MT 1/2) and metallothionein 3 (MT 3) in human ocular tissue. It is making use of (a) antibodies conjugated to gold nanoclusters (AuNCs) acting as labels, and (b) laser ablation (LA) coupled to inductively coupled plasma – mass spectrometry (ICP-MS).Water-soluble fluorescent AuNCs (with an average size of 2.7 nm) were synthesized and then conjugated to antibody by carbodiimide coupling. The surface of the modified AuNCs was then blocked with hydroxylamine to avoid nonspecific interactions with biological tissue. Immunoassays for MT 1/2 and MT 3 in ocular tissue sections (5 μm thick) from two post mortem human donors were performed. Imaging studies were then performed by fluorescence using confocal microscopy, and LA-ICP-MS was performed in the retina to measure the signal for gold. Signal amplification by the >500 gold atoms in each nanocluster allowed the antigens (MT 1/2 and MT 3) to be imaged by LA-ICP-MS using a laser spot size as small as 4 μm. The image patterns found in retina are in good agreement with those obtained by conventional fluorescence immunohistochemistry which was used as an established reference method. KW - Metal nanoclusters KW - Fluorescence KW - Protein imaging KW - Thin tissue sections KW - Immunohistochemistry KW - Bioconjugation KW - Carbodiimide crosslinking KW - Laser ablation KW - Mass spectrometry PY - 2018 DO - https://doi.org/10.1007/s00604-017-2597-1 VL - 185 IS - 1 SP - 1 EP - 9 PB - Springer AN - OPUS4-44022 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ribar, A. A1 - Huber, S. E. A1 - Smiałek, M. A. A1 - Tanzer, K. A1 - Neustetter, M. A1 - Schürmann, Robin A1 - Bald, Ilko A1 - Denifl, S. T1 - Hydroperoxyl radical and formic acid formation from common DNA stabilizers upon low energy electron attachment JF - Physical chemistry chemical physics N2 - 2-Amino-2-(hydroxymethyl)-1,3-propanediol (TRIS) and ethylene-diaminetetraacetic acid (EDTA) are key components of biological buffers and are frequently used as DNA stabilizers in irradiation studies. Such surface or liquid phase studies are done with the aim to understand the fundamental mechanisms of DNA radiation damage and to improve cancer radiotherapy. When ionizing radiation is used, abundant secondary electrons are formed during the irradiation process, which are able to attach to the molecular compounds present on the surface. In the present study we experimentally investigate low energy electron attachment to TRIS and methyliminodiacetic acid (MIDA), an analogue of EDTA, supported by quantum chemical calculations. The most prominent dissociation channel for TRIS is through hydroperoxyl radical formation, whereas the dissociation of MIDA results in the formation of formic and acetic acid. These compounds are well-known to cause DNA modifications, like strand breaks. The present results indicate that buffer compounds may not have an exclusive protecting effect on DNA as suggested previously. KW - DEA KW - DNA KW - Low energy electrons KW - Mass spectrometry PY - 2018 UR - http://pubs.rsc.org/en/content/articlepdf/2018/cp/c7cp07697e DO - https://doi.org/10.1039/c7cp07697e VL - 20 IS - 8 SP - 5578 EP - 5585 PB - Royal Society of Chemistry AN - OPUS4-44703 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Kuhring, M. A1 - Doellinger, J. A1 - Nitsche, A. A1 - Muth, Thilo A1 - Renard, B. Y. T1 - TaxIt: An Iterative Computational Pipeline for Untargeted Strain-Level Identification Using MS/MS Spectra from Pathogenic Single-Organism Samples JF - Journal of Proteome Research N2 - Untargeted accurate strain-level classification of a priori unidentified organisms using tandem mass spectrometry is a challenging task. Reference databases often lack taxonomic depth, limiting peptide assignments to the species level. However, the extension with detailed strain information increases runtime and decreases statistical power. In addition, larger databases contain a higher number of similar proteomes. We present TaxIt, an iterative workflow to address the increasing search space required for MS/MS-based strain-level classification of samples with unknown taxonomic origin. TaxIt first applies reference sequence data for initial identification of species candidates, followed by automated acquisition of relevant strain sequences for low level classification. Furthermore, proteome similarities resulting in ambiguous taxonomic assignments are addressed with an abundance weighting strategy to increase the confidence in candidate taxa. For benchmarking the performance of our method, we apply our iterative workflow on several samples of bacterial and viral origin. In comparison to noniterative approaches using unique peptides or advanced abundance correction, TaxIt identifies microbial strains correctly in all examples presented (with one tie), thereby demonstrating the potential for untargeted and deeper taxonomic classification. TaxIt makes extensive use of public, unrestricted, and continuously growing sequence resources such as the NCBI databases and is available under open-source BSD license at https://gitlab.com/rki_bioinformatics/TaxIt. KW - Bioinformatics KW - Mass spectrometry KW - Microbial proteomics KW - Strain identification KW - MS/MS PY - 2020 DO - https://doi.org/10.1021/acs.jproteome.9b00714 VL - 19 IS - 6 SP - 2501 EP - 2510 PB - ACS AN - OPUS4-50942 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Jaeger, Carsten T1 - Improving Small Molecule Annotation In Nontargeted Soft Ionization GC/LC High Resolution Mass Spectrometry N2 - High-resolution mass spectrometry, either combined with gas or liquid chromatography (GC/LC-HR-MS), is currently the most powerful analytical option for broad nontargeted small molecule analysis. To transform HR-MS raw data from metabolomics or environmental studies into chemically meaningful data, several computational steps are needed, including peak detection, deconvolution of these peaks into compounds and putative identification of compounds using databases. Each of these steps can cause problems and still requires methodological advancements. Computational compound annotation as one of the steps, however, has proven particularly challenging, mainly due to the chemical diversity of organic analytes. In addition, different annotation approaches are needed for the different platforms used in HR-MS screening. We recently introduced InterpretMSSpectrum as an annotation workflow for GC-HR-MS using atmospheric pressure chemical ionization (APCI). InterpretMSSpectrum locates molecular ion, fragment and adduct peaks, calculates their most likely sum formula combination and graphically summarizes results as an annotated mass spectrum. As a complementary approach for LC-HR-MS, we presented findMAIN, which scores MS1 spectra based on explained intensity, mass accuracy and isotope charge agreement of adducts and related electrospray ionization (ESI) products to determine the neutral mass of unknown compounds. Both approaches were validated against large spectral libraries containing more than 600 compounds, for which correct annotation was achieved in over 80% of the cases. Based on the experiences from this validation, we here compare soft-ionization GC/LC-HR-MS regarding “annotatability” of unknown compounds from a computational perspective. As a main advantage for GC-HR-MS, the relatively uniform ionization behavior of commonly used trimethylsilyl (TMS) derivatives observed under APCI allowed differentiation of molecular ion peaks from in-source fragments based on relatively compact set of rules. By contrast, neutral mass inference in ESI required a more complex evaluation scheme, due to the higher diversity of ionization products observable in ESI. We demonstrate such differences by practical examples of both software packages applied to metabolomics studies and discuss the challenges connected to transferring the approaches to environmental screening. T2 - ACS Fall Meeting 2019 CY - San Diego, CA, USA DA - 25.09.2019 KW - Compound annotation KW - Mass spectrometry KW - Electrospray ionization KW - Nontarget analysis KW - Accurate mass PY - 2019 AN - OPUS4-49200 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Lörchner, Dominique A1 - Dittmann, Daniel A1 - Braun, Ulrike A1 - Kroh, L. W. A1 - Köppen, Robert T1 - Investigation of two triazine-based heterocyclic brominated flame retardants by coupled thermogravimetry-Fourier transform infrared spectroscopy JF - Journal of Analytical and Applied Pyrolysis N2 - In this study, the thermal decomposition of 1,3,5-tris-(2,3-dibromopropyl)-1,3,5-triazine-2,4,6-trione (TDBPTAZTO) and 2,4,6-tris-(2,4,6-tribromo-phenoxy)-1,3,5-triazine (TTBP-TAZ) were investigated for the first time by thermogravimetric analysis. Both compounds were thermal degraded between 225 and 350 °C (TDBP-TAZTO) as well as 300 and 400 °C (TTBP-TAZ). As a result, mass loss (%) of 98.5% for TTBP-TAZ and 95.4% for TDBP-TAZTO at 600 °C under N2 were observed. The major pyrolytic degradation products of TTBP-TAZ were formed in a single step and identified by FTIR analysis as 2,4,6-tribromophenol and further bromine-substituted aromatic compounds. In comparison, TDBP-TAZTO was pyrolytic degraded in two steps, whereby on the first step the release of hydrogen Bromide and 1,3,5-triallyl-1,3,5-triazine-2,4,6-trione could be detected. In the second minor step, isocyanic acid could be additionally identified. Subsequently, the obtained products of the TGA-FTIR measurements were used for a targeted search for mass fragments in mass spectrometry measurements. For TTBP-TAZ, only the degradation product 1,3,5-tribromobenzene could be detected by MS/MS analyzes. No comparable thermal degradation products, except hydrogen bromide, were observed in the MS/MS spectra of TDBP-TAZTO. Therefore, the search of further mass fragments was not possible compared to the findings of the TGA-FTIR measurements. KW - Pyrolysis KW - Thermal decomposition KW - TGA-FTIR KW - Mass spectrometry PY - 2019 DO - https://doi.org/10.1016/j.jaap.2019.104635 VL - 141 SP - 104635-1 EP - 104635-5 PB - Elsevier B.V. AN - OPUS4-48506 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Jaeger, Carsten A1 - Lisec, Jan T1 - Rapid characterization of ionization products for chemistry-aware compound annotation in nontargeted small molecule mass spectrometry N2 - Nontargeted high-resolution mass spectrometry (HRMS) is widely used for small molecule screening in biotic or abiotic samples. However, HRMS approaches like metabolomics or environmental nontarget screening currently still lack confidence in chemical annotation, i.e. computational structure assignment to all measured mass signals. As a crucial step within the annotation pipeline, molecular weight inference (MWI) deduces a compound’s intact mass from diagnostic mass differences between MS1 peaks, allowing precise database queries in subsequent steps. As the common practice of considering all possible ionization products such as adducts, multimers, multiple charges etc. in MWI suffers from high false positive rates, we aimed at selecting candidate ionization products in a chemically sensitive way. Generally, electrospray ionization produces different types of adducts depending on chromatographic system and sample matrix, necessitating application-specific optimization for optimum MWI performance. To avoid, however, the tedious and potentially biased manual data curation connected to optimization, we established an R-based workflow for automating this task. The workflow consists of two parts. Part 1 creates an MS1 spectral library by performing peak detection, spectral deconvolution and target peak assignment based on density estimation. Part 2 analyzes ion relationships within the library and returns a list of detected ionization products ranked by their frequency. We applied the workflow to a commercial 634-compound library that was acquired for two chromatographic methods (reverse phase, RP; hydrophilic liquid interaction chromatographic, HILIC) and the two ESI modes (positive, negative). As expected, different frequency distributions of ionization products were found for the two chromatographies. Interestingly, however, some of the differences were expected in terms of solvent chemistries (e.g. [M+NH4]+ in ammonium formate-buffered HILIC) while others indicated more complex ion competition (e.g. abundant [M+K]+, [M+2K-H]+ in HILIC). This demonstrated the relevance of this empirical approach. We further show that MWI accuracy clearly benefitted from derived optimized adduct lists – by adding filters or weighting terms – and present FDR calculations supporting this observation. We conclude that chemistry-aware compound annotation based on the combination of high-throughput library acquisition and statistical analysis holds significant potential for further improvements in nontargeted small molecule HRMS. T2 - ANAKON 2019 CY - Münster, Germany DA - 25.03.2019 KW - Compound annotation KW - Mass spectrometry KW - Electrospray ionization KW - Nontarget analysis KW - Accurate mass PY - 2019 AN - OPUS4-48332 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen T1 - The triple-isotope calibration approach: a universal and standard-free calibration approach for obtaining absolute isotope ratios of multi-isotopic elements JF - Analytical and Bioanalytical Chemistry N2 - The theory of a new calibration approach for obtaining absolute isotope ratios of multi-isotopic elements without the use of any standard has been developed. The calibration approach basically uses the difference in the instrumental isotope fractionation of two different types of mass spectrometers, leading to two different fractionation lines in a three-isotope diagram. When measuring the same sample with both mass spectrometers, the different fractionation lines have one point in common: this is the ‘true’ logarithmized isotope ratio pair of the sample. Thus, the intersection of both fractionation lines provides us with the absolute isotope ratios of the sample. This theory has been tested in practice by measuring Cd and of Pb isotope ratios in the certified reference materials BAM-I012 and NIST SRM981 by thermal ionization mass spectrometry and by inductively coupled plasma mass spectrometry while varying the ionization conditions for both mass spectrometers. With this experiment, the theory could be verified, and absolute isotope ratios were obtained, which were metrologically compatible with the certified isotope ratios. The so-obtained absolute isotope ratios are biased by − 0.5% in average, which should be improved with further developments of the method. This calibration approach is universal, as it can be applied to all elements with three or more isotopes and it is not limited to the type of mass spectrometers applied; it can be applied as well to secondary ion mass spectrometry or others. Additionally, this approach provides information on the fractionation process itself via the triple-isotope fractionation exponent θ. KW - Triple isotope fractionation KW - Absolute isotope ratio KW - Mass spectrometry KW - Calibration KW - Uncertainty PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-516364 DO - https://doi.org/10.1007/s00216-020-03050-4 VL - 413 IS - 3 SP - 821 EP - 826 PB - Springer Verlag AN - OPUS4-51636 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -