TY - JOUR A1 - Moraleja, I. A1 - Mena, M. L. A1 - Lázaro, A. A1 - Neumann, B. A1 - Tejedor, A. A1 - Jakubowski, Norbert A1 - Gómez-Gómez, M. M. A1 - Esteban-Fernández, D. T1 - An approach for quantification of platinum distribution in tissues by LA-ICP-MS imaging using isotope dilution analysis N2 - Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) has been revealed as a convenient technique for trace elemental imaging in tissue sections, providing elemental 2D distribution at a quantitative level. For quantification purposes, in the last years several approaches have been proposed in the literature such as the use of CRMs or matrix matched standards. The use of Isotope Dilution (ID) for quantification by LA-ICP-MS has been also described, being mainly useful for bulk analysis but not feasible for spatial measurements so far. In this work, a quantification method based on ID analysis was developed by printing isotope-enriched inks onto kidney slices from rats treated with antitumoral Pt-based drugs using a commercial ink-jet device, in order to perform an elemental quantification in different areas from bio-images. For the ID experiments ¹⁹⁴Pt enriched platinum was used. The methodology was validated by deposition of natural Pt standard droplets with a known amount of Pt onto the surface of a control tissue, where could be quantified even 50 pg of Pt, with recoveries higher than 90%. The amount of Pt present in the whole kidney slices was quantified for cisplatin, carboplatin and oxaliplatin-treated rats. The results obtained were in accordance with those previously reported. The amount of Pt distributed between the medullar and cortical areas was also quantified, observing different behavior for the three drugs. KW - Isotopic dilution KW - LA-ICP-MS KW - Quantification KW - Imaging KW - Kidney KW - Pt-based drugs PY - 2018 DO - https://doi.org/10.1016/j.talanta.2017.09.031 SN - 0039-9140 VL - 178 SP - 166 EP - 171 PB - Elsevier AN - OPUS4-43627 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Nxumalo, T. A1 - Akhdhar, A. A1 - Mueller, V. A1 - Simon, Fabian A1 - von der Au, Marcus A1 - Cossmer, Antje A1 - Pfeifer, Jens A1 - Krupp, E. A1 - Meermann, Björn A1 - Kindness, A. A1 - Feldmann, J. T1 - EOF and target PFAS analysis in surface waters affected by sewage treatment effluents in Berlin, Germany N2 - Per- and polyfuoroalkyl substances (PFAS) are emerging organic pollutants and can occur in surface and groundwater. To identify the degree of pollution in surface water with PFAS, often targeted HPLC–ESI–MS/MS has been employed in which commonly 30–40 compounds are analyzed. However, other PFAS and organofuorines remain undetected. We sampled surface water of the river Spree and the Teltow Canal in Berlin, Germany, which are afected by the efuent discharge of wastewater treatment plants. Here, we employed high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) for measuring extractable organofuorines (EOF) and compared in a mass balance approach the total fuorine to the identifed and quantifed PFAS from the targeted analysis. The analysis highlights that the EOF are in the range expected for an urban river system (Winchell et al. in Sci Total Environ 774, 2021). However, downstream of an efuent discharge, the EOF increased by one order of magnitude, e.g., 40.3 to 574 ng F L−1, along the Teltow Canal. From our target analytes, mostly short-chained perfuorinated carboxylic acids and sulfonates occur in the water, which however makes up less than 10% of the EOF. The increase in EOF in the Teltow Canal correlates well with the increase of perfuorohexanoic acid (PFHxA), indicating that PFHxA is characteristic for the discharged EOF but not responsible for the increase. Hence, it points to PFHxA precursor discharge. The study highlights that EOF screening using HR-CS-GFMAS is necessary to identify the full scale of pollution with regard to PFAS and other organofuorines such as pharmaceutical compounds from the efuent of WWTPs. KW - PFAS KW - WWTP KW - GF-MAS PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-580341 DO - https://doi.org/10.1007/s00216-022-04500-x VL - 415 SP - 1195 EP - 1204 PB - Springer AN - OPUS4-58034 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Phukphatthanachai, P. A1 - Panne, Ulrich A1 - Traub, Heike A1 - Pfeifer, Jens A1 - Vogl, Jochen T1 - Quantification of sulphur in copper and copper alloys by GDMS and LA-ICP-MS, demonstrating metrological traceability to the international system of units N2 - The quantification of the sulphur mass fraction in pure copper and copper alloys by GDMS and LA-ICP-MS revealed a lack of traceability mainly due to a lack of suitable certified reference materials for calibrating the instruments. Within this study GDMS and LA-ICP-MS were applied as routine analytical tools to quantify sulphur in copper samples by applying reference materials as calibrators, which were characterized for their sulphur mass fraction by IDMS beforehand. Different external calibration strategies were applied including a matrix cross type calibration. Both techniques with all calibration strategies were validated by using certified reference materials (others than those used for calibration) and good agreement with the reference values was achieved except for the matrix cross type calibration, for which the agreement was slightly worse. All measurement results were accompanied by an uncertainty statement. For GDMS, the relative expanded (k = 2) measurement uncertainty ranged from 3% to 7%, while for LA-ICP-MS it ranged from 11% to 33% when applying matrix-matched calibration in the sulphur mass fraction range between 25 mg kg-1 and 1300 mg kg-1. For cross-type calibration the relative expanded (k = 2) measurement uncertainty need to be increased to at least 12% for GDMS and to at least 54% for LA-ICP-MS to yield metrological compatibility with the reference values. The so obtained measurement results are traceable to the international system of units (SI) via IDMS reference values, which is clearly illustrated by the unbroken chain of calibrations in the metrological traceability scheme. KW - Sulfur KW - Copper KW - GDMS KW - LA-ICP-MS KW - Uncertainty KW - Traceability KW - SI PY - 2021 DO - https://doi.org/10.1039/d1ja00137j SN - 0267-9477 VL - 36 IS - 11 SP - 2404 EP - 2414 PB - Royal Society of Chemistry AN - OPUS4-53412 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Phukphatthanachai, Pranee A1 - Panne, Ulrich A1 - Jakubowski, Norbert A1 - Vogl, Jochen T1 - SI-traceable quantification of sulphur in copper metal and its alloys by ICP-IDMS N2 - Previously applied methods for the quantification of sulphur in copper and other pure metals revealed a lack of SI-traceability and additionally showed inconsistent results, when different methods were compared. Therefore, a reference procedure is required which allows SI-traceable values accompanied by a Sound uncertainty budget. In this study a procedure was developed for the quantification of total sulphur in copper at low concentration levels using inductively coupled plasma-isotope dilution mass spectrometry (ICP-IDMS). The major part of the copper matrix was separated by adding ammonia which forms a complex with the copper while releasing the sulphur followed by chromatographic separation using a weak cation resin. After that the sulphur fraction was further purified by chromatographic means using first an anion exchange method and second a chelating resin. The developed procedure shows high performance, especially concerning high efficiency in matrix removal (>99.999%) while keeping the recovery of sulphur above 80%. Procedure blanks are in the order of 3–53 ng resulting in LOD and LOQ values of 0.2 mg g1 and 0.54 mg g1, respectively. The procedure is sufficient to facilitate value assignment of the total sulphur mass fraction in reference materials. Additionally, relative measurement uncertainties were calculated to be below 1% and the measurement results were traceable to the SI. The procedure reported in this study is a new reference procedure for sulphur measurement in copper, being fit for two major purposes, certification of reference materials and assignment of reference values for inter-laboratory comparison. KW - Traceability KW - Measuremment uncertainty KW - Isotope dilution mass spectrometry KW - Reference measurements PY - 2018 DO - https://doi.org/10.1039/c7ja00338b SN - 0267-9477 SN - 1364-5544 VL - 33 IS - 1 SP - 90 EP - 101 PB - Royal Society of Chemistry CY - Cambridge, UK AN - OPUS4-43614 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Phukphatthanachai, Pranee A1 - Vogl, Jochen A1 - Traub, Heike A1 - Jakubowski, Norbert A1 - Panne, Ulrich T1 - A new approach of using polyethylene frits for the quantification of sulphur in copper metals by isotope dilution LA-ICP-MS and comparison with conventional IDMS techniques N2 - Polyethylene (PE) frits were used to quantify sulphur in copper and its alloys by isotope dilution combined with LA-ICP-MS as an alternative approach to conventional sample preparation: the copper samples were spiked, the spiked samples were dissolved, the resulting solutions were absorbed in the PE frits and finally the PE frits were analysed by LA-ICP-MS. A prerequisite for such a support material is a low sulphur blank and thus PE was selected for this purpose. The absorption efficiency of the PE frits was studied for varying sulphur amounts ranging from 2 mg S to 80 mg S showing that more than 99.5% of the loaded sulphur was absorbed by the frit. The so prepared PE frits were measured by LA-ICP-MS and yielded a good linearity (R2 ¼ 0.999) for the sulphur ion intensities corresponding to sulphur amounts up to 40 mg S; the associated sensitivity is approximately 3.4 x 10⁴ cps μg⁻¹ for ³²S. For the validation of the developed procedure the reference materials BAM-M376a, BAM-228 and BAM-227 were applied such that 2 μg S, 5 μg S and 11 μg S were absorbed in the PE frits, respectively. These samples were pre-quantified for the adsorbed sulphur amount by external calibration LA-ICP-MS yielding sulphur amounts of 0.9 μg, 5.1 μg and 8.5 μg (quantified for ³²S only), respectively. Relative Standard deviations of the isotope ratios were below 5% in average (n ¼ 3 lines) in all cases (except for the pure spike solution). These samples were then analysed by LA-ICP-IDMS and the measurement results were validated by comparing them with the results obtained by conventional ICP-IDMS. The obtained relative expanded measurement uncertainties ranged between 10% and 26%. Pearson's coefficient was used to express the correlation between both techniques; the obtained value was 0.999 demonstrating a strong correlation. Contrary to most published LA-ICP-IDMS procedures, the developed procedure enables SI-traceability for the measurement results. The metrological traceability to the SI for the sulphur mass fractions in copper was established by an unbroken chain of comparisons, each accompanied by an uncertainty budget. Thus, the measurement results are considered reliable, acceptable and comparable within the stated measurement uncertainty. The metrological traceability chain from the kg down to mass fraction in the samples obtained by LA-ICP-IDMS is presented as well. KW - Laser ablation KW - IDMS KW - Traceability KW - Uncertainty KW - Purity PY - 2018 DO - https://doi.org/10.1039/c8ja00116b SN - 0267-9477 SN - 1364-5544 VL - 33 IS - 9 SP - 1506 EP - 1517 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-45899 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Pisonero, J. A1 - Bouzas-Ramos, D. A1 - Traub, Heike A1 - Cappella, Brunero A1 - Alvarez-Llamas, C. A1 - Richter, Silke A1 - Mayo, J. C. A1 - Costa-Fernandez, J. M. A1 - Bordel, N. A1 - Jakubowski, Norbert T1 - Critical evaluation of fast and highly resolved elemental distribution in single cells using LA-ICP-SFMS N2 - The analytical potential of a nanosecond laser ablation inductively coupled plasma mass spectrometer (ns-LA-ICP-SFMS) system, equipped with an ultra-fast wash-out ablation chamber, is critically investigated for fast and highly spatially resolved (∼μm) qualitative elemental distribution within single cells. Initially, a low surface roughness (< 10 nm) thin In–SnO2 layer (total coating thickness ∼200 nm) deposited on glass is employed to investigate the size, morphology and overlapping of laser-induced craters obtained at different laser repetition rates, making use of Atomic Force Microscopy (AFM). Conical craters with a surface diameter of about 2 µm and depths of about 100 nm were measured after a single laser shot. Furthermore, the influence of the sampling distance (i.e. distance between the sample surface and the inner sniffer of the ablation chamber) on the LA-ICP-MS ion signal wash-out time is evaluated. A significant decrease of the transient 120Sn+ ion signal is noticed after slight variations (±200 μm) around the optimum sampling position. Ultra-fast wash-outs (< 10 ms) are achieved reducing the aerosol mixing from consecutive laser shots even when operating the laser at high repetition rates (25 – 100 Hz). Fast and highly spatially resolved images of elemental distribution within mouse embryonic fibroblast cells (NIH/3T3 fibroblast cells) and human cervical carcinoma cells (HeLa cells), incubated with gold nanoparticles (Au NPs) and Cd-based quantum dots (QDs), respectively, are determined at the optimized operating conditions. Elemental distribution of Au and Cd in single cells is achieved using a high scanning speed (50 µm/s) and high repetition rate (100 Hz). The results obtained for the distribution of fluorescent Cd-based QDs within the HeLa cells are in good agreement with those obtained by confocal microscopy. The size, morphology and overlapping of laser-induced craters in the fixed cells are also investigated using AFM, observing conical craters with a surface diameter of about 2.5 µm and depths of about 800 nm after a single laser shot. KW - Laser ablation KW - ICP-MS KW - Nanoparticle KW - Atomic Force Microscopy KW - Cell PY - 2018 DO - https://doi.org/10.1039/c8ja00096d SP - 1 EP - 9 PB - Royal Society of Chemistry CY - Cambridge AN - OPUS4-45564 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Pramann, A. A1 - Vogl, Jochen A1 - Rienitz, O. T1 - The Uncertainty Paradox: Molar Mass of Enriched Versus Natural Silicon Used in the XRCD Method N2 - The X-ray crystal density method uses silicon spheres highly enriched in 28Si as a primary method for the dissemination of the SI base unit kilogram yielding smallest possible uncertainties associated with the mass m within a few parts in 10-8. This study compares different available and newly developed analytical methods and their results for the determination of the molar mass M of silicon highly enriched in 28Si (Me) and of silicon (Mx) with an almost natural isotopic distribution. While for Me relative uncertainties urel(Me) in the lower 10-9 range are obtained routinely, it was not possible to fall below a value of urel(Mx) < 4 x 10-6 in the case of natural silicon, which is approximately three orders of magnitude larger. The application of the state-of the-art isotope ratio mass spectrometry accompanied with sophisticated thoroughly investigated methods suggests an intrinsic cause for the large uncertainty associated with the molar mass of natural silicon compared to the enriched material. KW - silicon KW - Molar mass KW - Isotope ratios KW - SI KW - Kilogram KW - Mole KW - XRCD method PY - 2020 DO - https://doi.org/10.1007/s12647-020-00408-y VL - 35 SP - 499 EP - 510 PB - Springer Verlag AN - OPUS4-51637 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Prohaska, T. A1 - Irrgeher, J. A1 - Benefield, J. A1 - Böhlke, J. K. A1 - Chesson, L. A. A1 - Coplen, T. B. A1 - Ding, T. A1 - Dunn, P. J. H. A1 - Gröning, M. A1 - Holden, N. E. A1 - Meijer, H. A. J. A1 - Moossen, H. A1 - Possolo, A. A1 - Takahashi, Y. A1 - Vogl, Jochen A1 - Walczyk, T. A1 - Wang, J. A1 - Wieser, M. E. A1 - Yoneda, S. A1 - Zhu, X.-K. A1 - Meija, J. T1 - Standard atomic weights of the elements 2021 (IUPAC Technical Report) N2 - Following the reviews of atomic-weight determinations and other cognate data in 2015, 2017, 2019 and 2021, the IUPAC (International Union of Pure and Applied Chemistry) Commission on Isotopic Abundances and Atomic Weights (CIAAW) reports changes of standard atomic weights. The symbol Ar(E) was selected for standard atomic weight of an element to distinguish it from the atomic weight of an element E in a specific substance P, designated Ar(E, P). The CIAAW has changed the values of the standard atomic weights of five elements based on recent determinations of terrestrial isotopic abundances: Ar (argon): from 39.948 ± 0.001 to [39.792, 39.963] Hf (hafnium): from 178.49 ± 0.02 to 178.486 ± 0.006 Ir (iridium): from 192.217 ± 0.003 to 192.217 ± 0.002 Pb (lead): from 207.2 ± 0.1 to [206.14, 207.94] Yb (ytterbium): from 173.054 ± 0.005 to 173.045 ± 0.010 The standard atomic weight of argon and lead have changed to an interval to reflect that the natural variation in isotopic composition exceeds the measurement uncertainty of Ar(Ar) and Ar(Pb) in a specific substance. The standard atomic weights and/or the uncertainties of fourteen elements have been changed based on the Atomic Mass Evaluations 2016 and 2020 accomplished under the auspices of the International Union of Pure and Applied Physics (IUPAP). Ar of Ho, Tb, Tm and Y were changed in 2017 and again updated in 2021: Al (aluminium), 2017: from 26.981 5385 ± 0.000 0007 to 26.981 5384 ± 0.000 0003 Au (gold), 2017: from 196.966 569 ± 0.000 005 to 196.966 570 ± 0.000 004 Co (cobalt), 2017: from 58.933 194 ± 0.000 004 to 58.933 194 ± 0.000 003 F (fluorine), 2021: from 18.998 403 163 ± 0.000 000 006 to 18.998 403 162 ± 0.000 000 005 (Ho (holmium), 2017: from 164.930 33 ± 0.000 02 to 164.930 328 ± 0.000 007) Ho (holmium), 2021: from 164.930 328 ± 0.000 007 to 164.930 329 ± 0.000 005 Mn (manganese), 2017: from 54.938 044 ± 0.000 003 to 54.938 043 ± 0.000 002 Nb (niobium), 2017: from 92.906 37 ± 0.000 02 to 92.906 37 ± 0.000 01 Pa (protactinium), 2017: from 231.035 88 ± 0.000 02 to 231.035 88 ± 0.000 01 Pr (praseodymium), 2017: from 140.907 66 ± 0.000 02 to 140.907 66 ± 0.000 01 Rh (rhodium), 2017: from 102.905 50 ± 0.000 02 to 102.905 49 ± 0.000 02 Sc (scandium), 2021: from 44.955 908 ± 0.000 005 to 44.955 907 ± 0.000 004 (Tb (terbium), 2017: from 158.925 35 ± 0.000 02 to 158.925 354 ± 0.000 008) Tb (terbium), 2021: from 158.925 354 ± 0.000 008 to 158.925 354 ± 0.000 007 (Tm (thulium), 2017: from 168.934 22 ± 0.000 02 to 168.934 218 ± 0.000 006) Tm (thulium), 2021: from 168.934 218 ± 0.000 006 to 168.934 219 ± 0.000 005 (Y (yttrium), 2017: from 88.905 84 ± 0.000 02 to 88.905 84 ± 0.000 01) Y (yttrium), 2021: from 88.905 84 ± 0.000 01 to 88.905 838 ± 0.000 002 KW - Argon KW - Ciaaw.org KW - Hafnium KW - Iridium KW - Lead KW - LSVEC KW - Ytterbium PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-548443 DO - https://doi.org/10.1515/pac-2019-0603 SN - 0033-4545 VL - 94 IS - 5 SP - 573 EP - 600 PB - De Gruyter Verlag CY - Berlin AN - OPUS4-54844 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Resch-Genger, Ute A1 - Meermann, Björn A1 - Koch, Matthias A1 - Weller, Michael G. T1 - Editorial: Analytical methods and applications in materials and life sciences N2 - Current trends in materials and life sciences are flanked by the need to push detection limits to single molecules or single cells, enable the characterization of increasingly complex matrices or sophisticated nanostructures, speed up the time of analysis, reduce instrument complexity and costs, and improve the reliability of data. This requires suitable analytical tools such as spectroscopic, separation and imaging techniques, mass spectrometry, and hyphenated techniques as well as sensors and their adaptation to application-specific challenges in the environmental, food, consumer product, health sector, nanotechnology, and bioanalysis. Increasing concerns about health threatening known or emerging pollutants in drinking water, consumer products, and food and about the safety of nanomaterials led to a new awareness of the importance of analytical sciences. Another important driver in this direction is the increasing demand by legislation, particularly in view of the 17 sustainable development goals by the United Nations addressing clean energy, industry, and innovation, sustainable cities, clean water, and responsible consumption and production. In this respect, also the development of analytical methods that enable the characterization of material flows in production processes and support recycling concepts of precious raw materials becomes more and more relevant. In the future, this will provide the basis for greener production in the chemical industry utilizing recycled or sustainable starting materials. This makes analytical chemistry an essential player in terms of the circular economy helping to increase the sustainability of production processes. In the life sciences sector, products based on proteins, such as therapeutic and diagnostic antibodies, increase in importance. These increasingly biotechnologically produced functional biomolecules pose a high level of complexity of matrix and structural features that can be met only by highly advanced methods for separation, characterization, and detection. In addition, metrological traceability and target definition are still significant challenges for the future, particularly in the life sciences. However, innovative reference materials as required for the health and food sector and the characterization of advanced materials can only be developed when suitable analytical protocols are available. The so-called reproducibility crisis in sciences underlines the importance of improved measures of quality control for all kinds of measurements and material characterization. This calls for thorough method validation concepts, suitable reference materials, and regular interlaboratory comparisons of measurements as well as better training of scientists in analytical sciences. The important contribution of analytical sciences to these developments is highlighted by a broad collection of research papers, trend articles, and critical reviews from these different application fields. Special emphasis is dedicated to often-overlooked quality assurance and reference materials. T2 - 150 years BAM: Science with impact CY - Berlin, Germany DA - 01.01.2021 KW - Analysis KW - Life sciences KW - Analytical sciences KW - Quality assurance KW - Reference material KW - Fluorescence KW - Nanoparticle KW - Sensor KW - Material sciences KW - Pollutant KW - Environment KW - Method KW - Limit of detection KW - 150th anniversary KW - ABC KW - BAM KW - Collection KW - Editorial KW - Special issue KW - Bundesanstalt für Materialforschung und -prüfung KW - Jahrestag PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-550720 DO - https://doi.org/10.1007/s00216-022-04082-8 SN - 1618-2642 SN - 1618-2650 VL - 414 IS - Topical collection: Analytical methods and applications in the materials and life sciences SP - 4267 EP - 4268 PB - Springer CY - Berlin AN - OPUS4-55072 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Retzmann, Anika A1 - Faßbender, Sebastian A1 - Rosner, M. A1 - von der Au, Marcus A1 - Vogl, Jochen T1 - Performance of second generation ICP-TOFMS for (multi-)isotope ratio analysis: a case study on B, Sr and Pb and their isotope fractionation behavior during the measurements N2 - The performance of second generation ICP-TOFMS, equipped with a micro-channel plate (MCP) enabling multi-isotope detection, in terms of isotope ratio precision and instrumental isotopic fractionation (IIF) for (multi-)isotope ratio analysis was thoroughly assessed for B, Sr and Pb. Experimental isotope ratio precision of 0.14 % for 11B/10B intensity ratio, 0.15 % for 87Sr/86Sr intensity ratio and 0.07% for 208Pb/206Pb intensity ratio were obtained at high signal levels ($500 mg L−1) which is comparable to first generation ICP-TOFMS. The long-term stability of isotope ratios, measured over several hours and expressed as repeatability, is between 0.05 % and 1.8 % for B, Sr and Pb. The observed IIF per mass unit is negative for B (i.e., −11 % for 11B/10B) which is in accordance with measurements using sector field (MC) ICP-MS. But the observed IIF per mass unit is positive for Sr (i.e., 2 % for 87Sr/86Sr) and Pb (i.e., 4.5 % for 208Pb/206Pb) which is not in accordance with measurements using sector field (MC) ICP-MS. Furthermore, different IIFs per mass unit were observed for different isotope pairs of the same isotopic system (i.e., Sr, Pb) and adjacent isotopic systems (i.e., Pb vs. Tl). This and the observations from three-isotope plots for Sr and Pb show that ion formation, ion extraction, ion transmission, ion separation and ion detection in second generation ICP-TOFMS is subject to IIF that does not follow the known mass dependent fractionation laws and is possibly caused by mass independent fractionation and/or multiple (contradictory) fractionation processes with varying contributions. The non-mass dependent IIF behavior observed for second generation ICP TOFMS has profound consequences for the IIF correction of isotope raw data, including application of multi-isotope dilution mass spectrometry (IDMS) using ICP-TOFMS. Hence, only IIF correction models that correct also for mass independent fractionation are applicable to calculate reliable isotope ratios using second generation ICP-TOFMS. In the present study, reliable d11B values, and absolute B, Sr and Pb isotope ratios could be determined using the SSB approach in single-element solutions as well as in a mixture of B, Sr and Pb, where the isotopes were measured simultaneously. KW - ICP-TOFMS KW - Isotope delta value KW - Isotope amount ratio KW - Conventional isotope ratio KW - Instrumental isotope fractionation PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-582239 DO - https://doi.org/10.1039/d3ja00084b SN - 0267-9477 VL - 38 IS - 10 SP - 2144 EP - 2158 PB - Royal Society of Chemistry AN - OPUS4-58223 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Reynolds, Merilie A. A1 - Kamber, Balz S. A1 - McKenna, Cora A. A1 - Oelze, Marcus A1 - Gleeson, Sarah A. T1 - Repeat, fast, and high-resolution mapping of fine-scale trace element distribution in pyrite and marcasite by LA-Q-ICP-MS with the Aerosol Rapid Introduction System (ARIS) N2 - AbstractThe minor and trace element composition of minerals provides critical insights into a variety of geological processes. Multi-element mapping by laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS) is an important technique applied for this purpose and although the method is rapidly advancing, there remains a fundamental compromise between spatial resolution, detection limit, and experiment duration when using sequential mass analyzers. To address the limitation of limited analyte selection for high spatial resolution maps imposed by the sequential nature of typical quadrupole (Q)-ICP-MS, we tested the Aerosol Rapid Introduction System (ARIS) for repeat mapping of the same area. The ARIS is a high-speed transfer tubing system that reduces aerosol washout times, permitting resolution of individual pulses at 40–60 Hz. Here, the ARIS was tested not for pulse resolution but with novel operating conditions optimized to perform fast, high spatial resolution mapping of minor and trace element distribution in pyrite and marcasite. For this purpose, ablation was conducted with a 5 µm beam aperture, a repetition rate of 50 Hz, and a continuous stage scan speed of 40 µm s−1. For each LA-Q-ICP-MS map, data were acquired for six elements with an acquisition time of 20 ms per element. This deliberately reduced the individual pulse resolution of the ARIS but instead exploited the spatial resolution and sensitivity gains afforded by the high-laser repetition rate combined with efficient aerosol transfer. The new method successfully mapped trace elements at single to double-digit parts per million levels, and the maps reveal fine-scale zoning of trace elements with an effective x and y resolution of 5 µm, while white light interferometry showed that for each experiment, only ca. 1 µm of the sample was removed. Repeated mapping of the same area showed excellent correspondence not only between element concentrations in successive experiments but also in the shape, dimension, and location of regions of interest defined by concentration criteria. The very good repeatability of the elemental maps indicates that for studies requiring more analytes, successive mapping of additional elements is possible. By contrast with conventional very small spot (i.e., 5 µm) analysis, fast repetition rate and stage scan speed mapping avoids down-hole fractionation effects and minimizes accidental analysis of buried invisible inclusions. Compared to conventional LA-ICP-MS mapping, the method reduces the experiment time by 4–8 times. KW - Geochemistry and Petrology KW - Geophysics PY - 2023 DO - https://doi.org/10.2138/am-2022-8168 SN - 0003-004X VL - 108 IS - 1 SP - 18 EP - 30 PB - Mineralogical Society of America AN - OPUS4-59038 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rieger, Philip A1 - Magnall, Joseph M. A1 - Gleeson, Sarah A. A1 - Oelze, Marcus T1 - Pyrite chemistry records a multistage ore forming system at the Proterozoic George Fisher massive sulfide Zn-Pb-Ag deposit, Mount Isa, Australia N2 - Trace element (TE) analysis of pyrite via LA-ICP-MS can be used to reconstruct the conditions of pyrite formation in complex mineral systems. The Carpentaria province in northern Australia is host to some of the world’s highest value Zn-Pb (+Ag, Cu) deposits. The genesis of many of these deposits is controversial, with competing models of single-vs. multi-stage ore formation. In this study, LA-ICP-MS data of paragenetically constrained pyrite from the George Fisher Zn-Pb-Ag deposit has been analysed to investigate the chemistry of different stages of ore formation. Pyrite from correlative unmineralized host rocks has also been analysed to investigate evidence of distal hydrothermal anomalism. All LA-ICP-MS data have been statistically evaluated (principal component analysis) and interpreted together with whole rock lithogeochemical data of the same samples. Pre-ore diagenetic pyrite is compositionally similar to other Proterozoic diagenetic pyrite, with some evidence of minor hydrothermal anomalism that with further analysis could help define distal alteration. Pyrite from the different ore stages are compositionally distinct, consistent with a multi-stage system. Ore stage 1 pyrite exceeds background contents of Co, Cu, Zn, As, Ag, Sb, Tl, and Pb and has elevated Co/Ni ratios, whereas only Ni and Co are above background abundances in ore stage 2 and 3 pyrite, of which only ore stage 3 pyrite has high Co/Ni ratios. Ore stage 1 pyrite has a similar composition to hydrothermal pyrite in the undeformed northern Carpentaria CD-type deposits and was likely syn-diagenesis. Ore stage 2 was syn-deformation, and resulted in replacement and recrystallization of pre-existing pyrite, and the expulsion of incompatible TEs. Ore stage 3 formed via a later Cu-Zn-Pb mineralizing event that resulted in a new geochemically distinct generation of Co-rich pyrite. Overall, this study demonstrates the value of paragenetically-constrained pyrite TE data for refining genetic models in complex sediment hosted mineral systems. KW - General Earth and Planetary Sciences PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-590407 DO - https://doi.org/10.3389/feart.2023.892759 VL - 11 SP - 1 EP - 19 PB - Frontiers Media SA AN - OPUS4-59040 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rienitz, O. A1 - Jährling, R. A1 - Noordmann, J. A1 - Pape, C. A1 - Röhker, K. A1 - Vogl, Jochen A1 - Manzano, J. V. L. A1 - Kozlowski, W. A1 - Caciano de Siena, R. A1 - Marques Rodrigues, J. A1 - Galli, A. H. A1 - Yim, Y.-H. A1 - Lee, K.-S. A1 - Lee, J. H. A1 - Min, H.-S. A1 - Chingbo, C. A1 - Naijie, S. A1 - Qian, W. A1 - Ren, T. A1 - Jun, W. A1 - Tangpaisarnkul, N. A1 - Suzuki, T. A1 - Nonose, N. A1 - Mester, Z. A1 - Yang, L. A1 - Pagliano, E. A1 - Greenberg, P. A1 - Mariassy, M. A1 - Näykki, T. A1 - Cankur, O. A1 - Coskun, F. G. A1 - Ari, B. A1 - Can, S. Z. T1 - CCQM-K122 "Anionic impurities and lead in salt solutions" N2 - The determination of the mass fractions of bromide, sulfate, and lead as well as the isotopic composition of the lead (expressed as the molar mass and the amount fractions of all four stable lead isotopes) in an aqueous solution of sodium chloride with a mass fraction of 0.15 g/g was the subject of this comparison. Even though the mass fractions ranged from 3 μg/g (bromide) to 50 ng/g (lead), almost all results reported agreed with the according KCRVs. KW - Absolute isotope ratio KW - Lead isotope ratios KW - Metrology KW - Traceability PY - 2020 DO - https://doi.org/10.1088/0026-1394/57/1A/08012 VL - 57 IS - 1A SP - 8012 PB - IOP Science AN - OPUS4-51156 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rienitz, O. A1 - Pramann, A. A1 - Vogl, Jochen A1 - Lee, K.-S. A1 - Yim, Y.-H. A1 - Malinovskiy, D. A1 - Hill, S. A1 - Dunn, P. A1 - Goenaga-Infante, H. A1 - Ren, T. A1 - Wang, J. A1 - Vocke jr., R. D. A1 - Rabb, S: A. A1 - Narukawa, T. A1 - Yang, L. A1 - Mester, Z. A1 - Meija, J. A1 - Aref'ev, D. G. A1 - Marchin, V. A1 - Sharin, A. G. A1 - Bulanov, A. D. A1 - Potapov, A. M. A1 - Otopkova, P. A. A1 - Kessel, R. T1 - The comparability of the determination of the molar mass of silicon highly enriched in 28Si: results of the CCQM-P160 interlaboratory comparison and additional external measurements N2 - An international comparison study on the accurate determination of the molar mass M(Si) of silicon artificially enriched in 28Si (x(28Si) > 0.9999 mol mol−1) has been completed. The measurements were part of the high level CCQM-P160 pilot study assessing the ability of National Metrology Institutes (NMIs) and Designated Institutes (DIs) to make such measurements at the lowest possible levels of measurement uncertainty and to identify possible difficulties when measuring this kind of sample. This study supports the molar mass measurements critical to disseminating the silicon route to realizing the new definitions for the kilogram and the mole. Measurements were also made by one external research institute and an external company. The different institutes were free to choose their experimental (mass spectrometric) set-ups and equipment, thereby enabling also the comparison of different techniques. The investigated material was a chemically pure, polycrystalline silicon material. The subsequent modified single crystalline secondary product of this material was intended for the production of silicon which was used for two additional spheres in the context of the redetermination of the Avogadro constant NA, required for the revision of the International System of Units (SI) via fundamental constants which came into force from May 2019. The CCQM pilot study was organized by Physikalisch-Technische Bundesanstalt (PTB). Aqueous silicon solutions were shipped to all participating institutions. The data analysis as well as the uncertainty modelling and calculation of the results was predefined. The participants were provided with an uncertainty budget as a GUM Workbench® file as well as a free software license for the duration of the comparison. The agreement of the values of the molar mass (M(Si) = 27.976 942 577 g mol−1) was excellent with ten out of 11 results reported within the range of relative uncertainty of 1 × 10−8 required for the revision of the SI. KW - Absolute isotope ratio KW - Molar mass KW - Avogadro constant KW - Revision of the SI PY - 2020 DO - https://doi.org/10.1088/1681-7575/abbdbf VL - 57 IS - 6 SP - 065028 PB - IOP Science CY - Cambridge AN - OPUS4-51500 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rosenkranz, D. A1 - Kriegel, F. L. A1 - Mavrakis, E. A1 - Pergantis, S. A. A1 - Reichardt, P. A1 - Tentschert, J. A1 - Jakubowski, N. A1 - Laux, P. A1 - Panne, Ulrich A1 - Luch, A. T1 - Versatile dual-inlet sample introduction system for multi-mode single particle inductively coupled plasma mass spectrometry N2 - Metal-containing nanoparticles (NP) can be characterized with inductively coupled plasma mass spectrometers (ICP-MS) in terms of their size and number concentration by using the single-particle mode of the instrument (spICP-MS). The accuracy of measurement depends on the setup, operational conditions of the instrument and specific parameters that are set by the user. The transport efficiency of the ICP-MS is crucial for the quantification of the NP and usually requires a reference material with homogenous size distribution and a known particle number concentration. Currently, NP reference materials are available for only a few metals and in limited sizes. If particles are characterized without a reference standard, the results of both size and particle number may be biased. Therefore, a dual-inlet Setup for characterizing nanoparticles with spICP-MS was developed to overcome this problem. This setup is based on a conventional introduction system consisting of a pneumatic nebulizer (PN) for nanoparticle solutions and a microdroplet Generator (μDG) for ionic calibration solutions. A new and flexible interface was developed to facilitate the coupling of μDG, PN and the ICP-MS system. The interface consists of available laboratory components and allows for the calibration, nanoparticle (NP) characterization and cleaning of the arrangement, while the ICP-MS instrument is still running. Three independent analysis modes are available for determining particle size and number concentration. Each mode is based on a different calibration principle. While mode I (counting) and mode III (μDG) are known from the literature, mode II (sensitivity), is used to determine the transport efficiency by inorganic ionic Standard solutions only. It is independent of NP reference materials. The μDG based inlet system described here guarantees superior analyte sensitivities and, therefore, lower detection limits (LOD). The size dependent LODs achieved are less than 15 nm for all NP (Au, Ag, CeO2) investigated. KW - Spectrometer KW - Reference KW - Calibration PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-536121 DO - https://doi.org/10.3791/61653 SN - 1940-087X IS - 163 SP - 1 EP - 19 PB - MyJoVE Corporation CY - Cambridge, MA, USA AN - OPUS4-53612 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Rosenkranz, Daniel A1 - Kriegel, F. L. A1 - Mavrakis, E. A1 - Pergantis, S. A. A1 - Reichardt, Ph. A1 - Tentschert, J. A1 - Jakubowski, Norbert A1 - Laux, P. A1 - Panne, Ulrich T1 - Improved validation for single particle ICP-MS analysis using a pneumatic nebulizer / microdroplet generator sample introduction system for multi-mode nanoparticle determination N2 - This study reports on the development of a single-particle (sp) inductively coupled plasma mass spectrometry (ICP-MS) technique suitable for the multi-mode determination of nanoparticle (NP) metal mass fraction and number concentration. The described technique, which is based on a dual inlet System consisting of a pneumatic nebulizer (PN) and a microdroplet generator (MDG), allows for the sequential introduction of ionic metal calibrant solutions and nanoparticle suspensions via all combinations of the two inlets; thus allowing for a combination of three independent modes of analysis. A novel interface, assembled using standard analytical components (a demountable quartz ICP-MS torch, flexible nonconducting silicon tubing and various connectors), was used to interface the dual inlet system to an ICP-MS. The interface provided improved functionality, compared to a previous design. It is now possible to conveniently exchange and introduce standard solutions and samples via all inlet combinations, analyze them, and also wash the sample inlet systems while the whole setup is still connected to an operating ICP-MS. This setup provided seamless and robust operation in a total of three analysis modes, i.e. three ways to independently determine the metal mass fraction and NP number concentration. All three analyses modes could be carried out within a single analytical run lasting approximately 20 min. The unique feature of the described approach is that each analysis mode is based on a different. KW - Nanomaterials KW - Nanoparticles KW - Single particle ICP-MS KW - Microdroplet generator PY - 2020 DO - https://doi.org/10.1016/j.aca.2019.11.043 VL - 1099 SP - 16 EP - 25 PB - Elsevier B.V. AN - OPUS4-50361 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Sargent, M A1 - Goenaga-Infante, H A1 - Inagaki, K A1 - Ma, L A1 - Meija, J A1 - Pramann, A A1 - Rienitz, O A1 - Sturgeon, R A1 - Vogl, Jochen A1 - Wang, J A1 - Yang, L T1 - The role of ICP-MS in inorganic chemical metrology N2 - ICP-MS has played a key role in inorganic chemical metrology for 25 years, from the 1993 CIPM feasibility study which led to establishment of the CCQM. Since that time, the Inorganic Analysis Working Group of the CCQM has organised 56 international comparisons involving measurements by ICP-MS and, in a recent comparison, 16 different national institutes submitted their results using the technique. Metrological applications of ICP-MS currently address an enormous range of measurements using a wide variety of instrumentation, calibration strategies and methodologies. This review provides an overview of the ICP-MS field with an emphasis on developments which are of particular relevance to chemical metrology. Examples from CCQM comparisons and the services available from the participants are used to illustrate how the capability and scope of ICP-MS methods have expanded far beyond the expectations of 1993. This is due in part to the research and development Programmes of the national institutes which participate in the CCQM. They have played a key role in advancing new instrumentation and applications for elemental analysis, isotope dilution mass spectrometry, determination of isotopic ratio or composition, and speciation of organometallic compounds. These developments are continuing today, as demonstrated by work in new fields such as heteroatom quantitation of proteins, characterisation and counting of nanoparticles using spICP-MS, and LA-ICP-MS analysis of solid materials. KW - CCQM KW - Metrology KW - Interlaboratory comparison KW - ICP-MS KW - Mass spectrometry KW - Hyphenated KW - Isotope ratio PY - 2019 DO - https://doi.org/10.1088/1681-7575/ab0eac VL - 56 IS - 3 SP - 034005 PB - IOP Publishing AN - OPUS4-47929 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schmidt, C. A1 - Hensen, C. A1 - Wallmann, K. A1 - Liebetrau, V. A1 - Tatzel, Michael A1 - Schurr, S. L. A1 - Kutterolf, S. A1 - Haffert, L. A1 - Geilert, Sonja A1 - Hübscher, C. A1 - Lebas, E. A1 - Heuser, A. A1 - Schmidt, M. A1 - Strauss, H. A1 - Vogl, Jochen A1 - Hansteen, T. T1 - Origin of High Mg and SO4 Fluids in Sediments of the Terceira Rift, Azores‐Indications for Caminite Dissolution in a Waning Hydrothermal System N2 - During R/V Meteor cruise 141/1, pore fluids of near surface sediments were investigated to find indications for hydrothermal activity in the Terceira Rift (TR), a hyperslow spreading center in the Central North Atlantic Ocean. To date, submarine hydrothermal fluid venting in the TR has only been reported for the D. João de Castro seamount, which presently seems to be inactive. Pore fluids sampled close to a volcanic cone at 2,800‐m water depth show an anomalous composition with Mg, SO4, and total alkalinity concentrations significantly higher than seawater and a nearby reference core. The most straightforward way of interpreting these deviations is the dissolution of the hydrothermally formed mineral caminite (MgSO4 0.25 Mg (OH)2 0.2H2O). This interpretation is corroborated by a thorough investigation of fluid isotope systems (δ26Mg, δ30Si, δ34S, δ44/42Ca, and 87Sr/86Sr). Caminite is known from mineral assemblages with anhydrite and forms in hydrothermal recharge zones only under specific conditions such as high fluid temperatures and in altered oceanic crust, which are conditions generally met at the TR. We hypothesize that caminite was formed during hydrothermal activity and is now dissolving during the waning state of the hydrothermal system, so that caminite mineralization is shifted out of its stability zone. Ongoing fluid circulation through the basement is transporting the geochemical signal via slow advection toward the seafloor. KW - Isotope ratio KW - Delta value KW - Pore fluids KW - Magnesium KW - Hydrothermal fluid PY - 2019 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-503430 DO - https://doi.org/10.1029/2019GC008525 VL - 20 IS - 12 SP - 6078 EP - 6094 PB - Wiley AN - OPUS4-50343 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Sieber, M. J. A1 - Wilke, M. A1 - Appelt, O. A1 - Oelze, Marcus A1 - Koch-Müller, M. T1 - Melting relations of Ca–Mg carbonates and trace element signature of carbonate melts up to 9 GPa – A proxy for melting of carbonated mantle lithologies N2 - The most profound consequences of the presence of Ca–Mg carbonates (CaCO3–MgCO3) in the Earth’s upper mantle may be to lower the melting temperatures of the mantle and control the melt composition. Low-degree partial melting of a carbonate-bearing mantle produces CO2-rich, silica-poor melts compositionally imposed by the melting relations of carbonates. Thus, understanding the melting relations in the CaCO3–MgCO3 system facilitates the interpretation of natural carbonate-bearing silicate systems. We report the melting relations of the CaCO3–MgCO3 system and the partition coefficient of trace elements between carbonates and carbonate melt from experiments at high pressure (6 and 9 GPa) and temperature (1300–1800 ◦C) using a rocking multi-anvil press. In the absence of water, Ca–Mg carbonates are stable along geothermal gradients typical of subducting slabs. Ca–Mg carbonates (∼ Mg0.1–0.9Ca0.9–0.1CO3) partially melt beneath mid-ocean ridges and in plume settings. Ca–Mg carbonates melt incongruently, forming periclase crystals and carbonate melt between 4 and 9 GPa. Furthermore, we show that the rare earth element (REE) signature of Group-I kimberlites, namely strong REE fractionation and depletion of heavy REE relative to the primitive mantle, is resembled by carbonate melt in equilibrium with Ca-bearing magnesite and periclase at 6 and 9 GPa. This suggests that the dolomite–magnesite join of the CaCO3–MgCO3 system might be useful to approximate the REE signature of carbonate-rich melts parental to kimberlites. KW - High pressure experiments KW - Laser Ablation PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-561197 DO - https://doi.org/10.5194/ejm-34-411-2022 SN - 0935-1221 VL - 34 IS - 5 SP - 411 EP - 424 PB - Copernicus Publications CY - Göttingen AN - OPUS4-56119 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Simon, Fabian A1 - Gehrenkemper, Lennart A1 - Becher, S. A1 - Dierckes, G. A1 - Langhammer, Nicole A1 - Cossmer, Antje A1 - von der Au, Marcus A1 - Göckener, B. A1 - Fliedner, A. A1 - Rüdel, H. A1 - Koschorreck, J. A1 - Meermann, Björn T1 - Quantification and characterization of PFASs in suspended particulate matter (SPM) of German rivers using EOF, dTOPA, (non-)target HRMS N2 - In this study, we compare analytical methods for PFAS determination–target analysis, non-target screening (NTS), direct total oxidizable precursor assay (dTOPA) and extractable organically bound fluorine (EOF). Therefore, suspended particulate matter (SPM) samples from German rivers at different locations in time series from2005 to 2020 were analyzed to investigate temporal and spatially resolved trends. In this study 3 PFAS mass balances approaches were utilized: (i) PFAA target vs. PFAS dTOPA, (ii) PFAS target vs. EOF and (iii) PFAS target vs. PFAS dTOPA vs. organofluorines NTS vs. EOF. Mass balance approach (i) revealed high proportions of precursor substances in SPM samples. For the time resolved analysis an increase from 94% (2005) to 97% in 2019 was observable. Also for the spatial resolved analysis precursor proportions were high with >84% at all sampling sites. Mass balance approach (ii) showed that the unidentified EOF (uEOF) fraction increased over time from82% (2005) to 99% (2019). Furthermore, along the river courses the uEOF increased. In the combined mass balance approach (iii) using 4 different analytical approaches EOF fractions were further unraveled. The EOF pattern was fully explainable at the sampling sites at Saar and Elbe rivers. For the time resolved analysis, an increased proportion of the EOF was now explainable. However, still 27% of the EOF for the time resolved analysis and 25% of the EOF for the spatial resolved analysis remained unknown. Therefore, in a complementary approach, both the EOF and dTOPA reveal unknown gaps in the PFAS mass balance and are valuable contributions to PFAS risk assessment. Further research is needed to identify organofluorines summarized in the EOF parameter. KW - PFAS KW - HR-CS-GFMAS KW - Fluorine KW - SPM KW - LC-MS/MS PY - 2023 DO - https://doi.org/10.1016/j.scitotenv.2023.163753 SN - 1879-1026 SN - 0048-9697 VL - 885 SP - 1 EP - 12 PB - Elsevier Science CY - Amsterdam AN - OPUS4-57429 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Simon, Fabian A1 - Gehrenkemper, Lennart A1 - Dierckes, G. A1 - Göckener, B. A1 - Koschorreck, J. A1 - Meermann, Björn T1 - Analyse von PFAS mittels (Non-)Target HRMS, dTOPA und EOF in Schwebstoff-Proben aus Zeitreihen der Umweltprobenbank N2 - Per- und polyfluorierte Alkylverbindungen (PFAS) sind eine Substanzklasse bestehend aus mehreren tausenden Einzelverbindungen. Unter PFAS zählen alle Verbindungen, die mindestens eine perfluorierte Methyl- (-CF3) oder Methylen-Gruppe (-CF2-) besitzen. Somit zählt auch ein großer Teil der fluorierten Medikamente, Pflanzenschutzmittel oder Biozide in die PFAS-Substanzklasse. Aufgrund der Persistenz, Bioakkumulation in Nahrungsketten, Toxizität und der ubiquitären Verbreitung von PFAS zählen sie zu den „emerging pollutants“. Die PFAS-Analytik ist wegen ihrer vielfältigen physikalisch-chemischen Eigenschaften und der sehr hohen Anzahl an PFAS eine große Herausforderung. Da es für viele Anwendungen noch ungewiss ist, welche PFAS verwendet werden, wie sie in die Umwelt gelangen und welche Transformations- bzw. Abbauprodukte entstehen, stoßen analytische Ansätze, die nur auf Einzelsubstanzen abzielen, hier an ihre Grenzen. PFAS-Summenparameter-Methoden, die darauf abzielen ein möglichst vollständiges Abbild der PFAS-Belastungssituation widerzuspiegeln, werden daher immer wichtiger. KW - PFAS KW - HR-CS-GFMAS KW - Fluorine KW - SPM KW - LC-MS/MS KW - Schwebstoffe PY - 2023 VL - 5 SP - 1 EP - 6 AN - OPUS4-57544 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Simon, Fabian A1 - Gehrenkemper, Lennart A1 - von der Au, Marcus A1 - Wittwer, Phillip A1 - Roesch, Philipp A1 - Pfeifer, Jens A1 - Cossmer, Antje A1 - Meermann, Björn T1 - A fast and simple PFAS extraction method utilizing HR–CS–GFMAS for soil samples N2 - Here, we describe an optimized fast and simple extraction method for the determination of per- and polyfluorinated alkyl substances (PFASs) in soils utilizing high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR–CS–GFMAS). To omit the bias of the solid phase extraction (SPE) step commonly used during the analysis of extractable organically bound fluorine (EOF) we optimized a fast and simple SPE-free extraction method. The developed extraction method consists of a liquid-solid extraction using acidified methanol without any additional SPE. Four extraction steps were representative to determine a high proportion of the EOF (>80% of eight extractions). Comparison of the optimized method with and without an additional SPE clean-up step revealed a drastic underestimation of EOF concentrations using SPE. Differences of up to 94% were observed which were not explainable by coextracted inorganic fluoride. Therefore, not only a more accurate but also a more economic as well as ecologic method (bypassing of unnecessary SPE) was developed. The procedural limit of quantification (LOQ) of the developed method was 10.30 μg/kg which was sufficient for quantifying EOF concentrations in all tested samples. For future PFAS monitoring and potential regulative decisions the herein presented optimized extraction method can offer a valuable contribution. KW - Per- and polyfluorinated alkly substances (PFASs) KW - High resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) KW - Soils KW - Solid phase extraction (SPE) KW - Extractable organically bound fluorine (EOF) KW - Solid-liquid extraction PY - 2022 DO - https://doi.org/10.1016/j.chemosphere.2022.133922 VL - 295 SP - 1 EP - 9 PB - Elsevier Ltd. AN - OPUS4-54359 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Solovyev, N. A1 - El-Khatib, Ahmed A1 - Costas-Rodrigues, M. A1 - Schwab, K. A1 - Griffin, E. A1 - Raab, A. A1 - Platt, B. A1 - Theuring, F. A1 - Vogl, Jochen A1 - Vanhaecke, F. T1 - Cu, Fe and Zn isotope ratios in murine Alzheimer's disease models suggest specific signatures of amyloidogenesis and tauopathy N2 - Alzheimer’s disease (AD) is characterized by accumulation of tau and amyloid-beta in the brain, and recent evidence suggests a correlation between associated protein aggregates and trace elements, such as copper, iron and zinc. In AD, distorted brain redox homeostasis and complexation by amyloid-beta and hyperphosphorylated tau May alter the isotopic composition of essential mineral elements. Therefore, high-precision isotopic analysis may reveal changes in the homeostasis of these elements. We used inductively coupled plasma-mass spectrometry (ICP-MS)-based techniques to determine the total Cu, Fe and Zn contents in the brain, as well as their isotopic compositions in both mouse brain and serum. Results for male transgenic tau (Line 66, L66) and amyloid/presenilin (5xFAD) mice were compared to those for the corresponding age- and gendermatched wild-type control mice (WT). Our data show that L66 brains showed significantly higher Fe levels than the corresponding WT. Significantly less Cu, but more Zn was found in 5xFAD brains. We observed significantly lighter isotopic compositions of Fe (enrichment in the lighter isotopes) in the brain, and in serum of L66 mice compared to WT. For 5xFAD mice, Zn exhibited a trend towards a lighter isotopic composition in brain and a heavier isotopic composition in serum compared to WT. Neither mouse model yielded differences in the isotopic composition of Cu. Our findings indicate significant pathology-specific alterations of Fe and Zn brain homeostasis in mouse models of AD. The associated changes in isotopic composition May serve as a marker for proteinopathies Underlying AD and other types of dementia. KW - Alzheimer’s disease KW - Tau KW - Amyloid-beta KW - Copper KW - Iron KW - Zinc KW - Multi-collector inductively coupled plasma-mass spectrometry (ICP-MS) KW - Brain KW - Serum KW - Isotopic analysis KW - Total element determination PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-520214 DO - https://doi.org/10.1016/j.jbc.2021.100292 VL - 296 SP - 100292 PB - Elsevier Inc. CY - Amsterdam AN - OPUS4-52021 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Spallanzani, Roberta A1 - Koga, K. A1 - Cichy, S. A1 - Wiedenbeck, M. A1 - Schmidt, B. A1 - Oelze, Marcus A1 - Wilke, M. T1 - Lithium and boron difusivity and isotopic fractionation in hydrated rhyolitic melts N2 - Lithium and boron are trace components of magmas, released during exsolution of a gas phase during volcanic activity. In this study, we determine the difusivity and isotopic fractionation of Li and B in hydrous silicate melts. Two glasses were synthesized with the same rhyolitic composition (4.2 wt% water), having diferent Li and B contents; these were studied in difusion-couple experiments that were performed using an internally heated pressure vessel, operated at 300 MPa in the temperature range 700–1250 °C for durations from 0 s to 24 h. From this we determined activation energies for Li and B difusion of 57±4 kJ/mol and 152±15 kJ/mol with pre-exponential factors of 1.53 × 10–7 m2/s and 3.80× 10–8 m2 /s, respectively. Lithium isotopic fractionation during difusion gave β values between 0.15 and 0.20, whereas B showed no clear isotopic fractionation. Our Li difusivities and isotopic fractionation results difer somewhat from earlier published values, but overall confrm that Li difusivity increases with water content. Our results on B difusion show that similarly to Li, B mobility increases in the presence of water. By applying the Eyring relation, we confrm that B difusivity is limited by viscous fow in silicate melts. Our results on Li and B difusion present a new tool for understanding degassing-related processes, ofering a potential geospeedometer to measure volcanic ascent rates. KW - Hydrated silicate melts KW - Stable isotopes KW - Diffusion KW - Isotopic fractionation PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-554496 DO - https://doi.org/10.1007/s00410-022-01937-2 VL - 177 IS - 8 SP - 1 EP - 17 PB - Springer AN - OPUS4-55449 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tatzel, M. A1 - Frings, P. J. A1 - Oelze, Marcus A1 - Herwartz, D. A1 - Lünsdorf, K. A1 - Wiedenbeck, M. T1 - Chert oxygen isotope ratios are driven by Earth's thermal evolution N2 - The 18O/16O ratio of cherts (δ18Ochert) increases nearly monotonically by ~15‰ from the Archean to present. Two end-member explanations have emerged: cooling seawater temperature (TSW) and increasing seawater δ18O (δ18Osw). Yet despite decades of work, there is no consensus, leading some to view the δ18Ochert record as pervasively altered. Here, we demonstrate that cherts are a robust archive of diagenetic temperatures, despite metamorphism and exposure to meteoric fluids, and show that the timing and temperature of quartz precipitation and thus δ18Ochert are determined by the kinetics of silica diagenesis. A diagenetic model shows that δ18Ochert is influenced by heat flow through the sediment column. Heat flow has decreased over time as planetary heat is dissipated, and reasonable Archean-modern heat flow changes account for ~5‰ of the increase in δ18Ochert, obviating the need for extreme TSW or δ18Osw reconstructions. The seawater oxygen isotope budget is also influenced by solid Earth cooling, with a recent reconstruction placing Archean δ18OSW 5 to 10‰ lower than today. Together, this provides an internally consistent view of the δ18Ochert record as driven by solid Earth cooling over billion-year timescales that is compatible with Precambrian glaciations and biological. KW - Climate KW - Oxygen isotope ratios KW - Silica diagenesis KW - Early Earth KW - Heat flow PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-569359 DO - https://doi.org/10.1073/pnas.2213076119 SN - 0027-8424 VL - 119 IS - 51 SP - 1 EP - 7 PB - National Academy of Sciences CY - Washington, DC AN - OPUS4-56935 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tatzel, Michael A1 - Oelze, Marcus A1 - Frick, Daniel A. A1 - Di Rocco, Tommaso A1 - Liesegang, Moritz A1 - Stuff, Maria A1 - Wiedenbeck, Michael T1 - Silicon and oxygen isotope fractionation in a silicified carbonate rock N2 - Silicon isotope fractionation during silicification is poorly understood and impedes our ability to decipher paleoenvironmental conditions from Si isotopes in ancient cherts. To investigate isotope fractionation during silica-for-carbonate replacement we analyzed the microscale Si and O isotope composition in different silica phases in a silicified zebra dolostone as well as their bulk δ18O and Δ’17O compositions. The subsequent replacement of carbonate layers is mimicked by decreasing δ18O and δ30Si. The textural relationship and magnitude of Si and O isotope fractionation is best explained by near-quantitative silica precipitation in an open system with finite Si. A Rayleigh model for silicification suggests positive Ɛ30/28Si during silicification, conforming with predictions for isotope distribution at chemical equilibrium from ab-initio models. Application of the modelled Ɛ30Si-T relationship yields silicification temperatures of approx. 50°C. To reconcile the δ18Ochert composition with these temperatures, the δ18O of the fluid must have been between -2.5 and -4 ‰, compositions for which the quartz phases fall close to the oxygen equilibrium fractionation line in three-isotope space. Diagenetic silica replacement appears to occur in O and Si isotopic equilibrium allowing reconstructions of temperatures of silicification from Si isotopes and derive the δ18O composition of the fluid – a highly desired value needed for accurate reconstructions of the temperature- and δ18O histories of the oceans. KW - Silicon isotopes PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-603430 DO - https://doi.org/10.1016/j.chemgeo.2024.122120 SN - 0009-2541 VL - 658 SP - 1 EP - 13 PB - Elsevier B.V. AN - OPUS4-60343 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tatzel, Michael A1 - Vogl, Jochen A1 - Rosner, M. A1 - Henehan, M. J. A1 - Tütken, T. T1 - Triple Isotope Fractionation Exponents of Elements Measured by MC-ICP-MS - An Example of Mg N2 - In most chemical reactions, stable isotopes are fractionated in a mass-dependent manner, yielding correlated isotope ratios in elements with three or more stable isotopes. The proportionality between isotope ratios is set by the triple isotope fractionation exponent θ that can be determined precisely for, e.g., sulfur and oxygen by IRMS, but not for metal(loid) elements due to the lower precision of MC-ICP-MS analysis and smaller isotopic variations. Here, using Mg as a test case, we compute a complete metrologically robust uncertainty budget for apparent θ values and, with reference to this, present a new measurement Approach that reduces uncertainty on θ values by 30%. This approach, namely, direct educt-product bracketing (sample−sample bracketing), allows apparent θ values of metal(loid) isotopes to be determined precisely enough to distinguish slopes in three-isotope space. For the example of Mg, we assess appropriate quality Control standards for interference-to-signal ratios and Report apparent θ values of carbonate−seawater pairs. We determined apparent θ values for marine biogenic carbonates, where the foraminifera Globorotalia menardii yields 0.514 ± 0.005 (2 SD), the coral Porites, 0.515 ± 0.006 (2 SD), and two specimens of the giant clam Tridacna gigas, 0.508 ± 0.007 (2 SD) and 0.509 ± 0.006 (2 SD), documenting differences in the uptake pathway of Mg among marine calcifiers. The capability to measure apparent θ values more precisely adds a new dimension to metal(loid) δ values, with the potential to allow us to resolve different modes of fractionation in industrial and natural processes. KW - Isotope fractionation KW - Delta value KW - Biogenic carbonates KW - Calcification KW - Magnesium isotope ratios KW - Measurement uncertainty KW - Sample-sample bracketing PY - 2019 DO - https://doi.org/10.1021/acs.analchem.9b02699 VL - 91 IS - 22 SP - 14314 EP - 14322 PB - ACS Publications AN - OPUS4-49818 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Taylor, Tristen L. A1 - Tukhmetova, Dariya A1 - Duong, Thi Phuong Thanh A1 - Böwe, Anna-Maria A1 - Meermann, Björn A1 - Gundlach-Graham, Alexander T1 - Comparative study of the vibrating capillary nebulizer (VCN) and commercially available interfaces for on-line coupling of capillary electrophoresis with ICP-MS N2 - Capillary electrophoresis (CE) is a powerful and sensitive tool for speciation analysis when combined with inductively coupled plasma mass spectrometry (ICP-MS); however, the performance of this technique can be limited by the nature of pneumatic nebulizers. This study compares two commercially available pneumatic nebulizers to a newly introduced vibrating capillary nebulizer (VCN) for on-line coupling of CE with ICP-MS. The VCN is a low-cost, non-pneumatic nebulizer that is based on the design of capillary vibrating sharp-edge spray ionization. As a piezoelectrically driven nebulization source, the VCN creates an aerosol independent of gas flows and does not produce a low-pressure region at the nebulizer orifice. To compare the systems, we performed replicate analyses of sulfate in river water with each nebulizer and the same CE and ICP-MS instruments and determined the figures of merit of each setup. With the CE-VCN-ICP-MS setup, we achieved around 2–4 times lower sensitivity compared to the commercial setups. However, the VCN-based setup provided lower noise levels and better linear correlation from the analysis of calibration standards, which resulted in indistinguishable LOD and LOQ values from the in-house-built VCN-based and commercial setups for CE-ICP-MS analysis. The VCN is found to have the highest baseline stability with a standard deviation of 3500 cts s−1, corresponding to an RSD of 2.7%. High reproducibility is found with the VCN with a peak area RSD of 4.1% between 3 replicate measurements. KW - Speciation analysis KW - Analytical chemistry KW - Surface water PY - 2024 DO - https://doi.org/10.1007/s00216-024-05162-7 SN - 1618-2650 SP - 1 EP - 9 PB - Springer CY - Berlin AN - OPUS4-59472 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Trog, S. A1 - El-Khatib, Ahmed A1 - Beck, S. A1 - Makowski, M. A1 - Jakubowski, Norbert A1 - Linscheid, M. T1 - Complementarity of molecular and elemental mass spectrometric imaging of Gadovist™ in mouse tissues N2 - Drug biodistribution analyses can be considered a key issue in pharmaceutical discovery and development. Here, mass spectrometric imaging can be employed as a powerful tool to investigate distributions of drug compounds in biologically and medically relevant tissue sections. Both matrix-assisted laser desorption ionization–mass spectrometric imaging as molecular method and laser ablation inductively coupled plasma–mass spectrometric imaging as elemental detection method were applied to determine drug distributions in tissue thin sections. Several mouse organs including the heart, kidney, liver, and brain were analyzed with regard to distribution of Gadovist™, a gadolinium-based contrast agent already approved for clinical investigation. This work demonstrated the successful detection and localization of Gadovist™ in several organs. Furthermore, the results gave evidence that gadolinium-based contrast agents in general can be well analyzed by mass spectrometric imaging methods. In conclusion, the combined application of molecular and elemental mass spectrometry could complement each other and thus confirm analytical results or provide additional information. KW - Laser ablation inductively coupled plasma–mass spectrometry imaging (LA-ICP-MSI) KW - Gadolinium-based contrast agents (GBCAs) KW - Matrix-assisted laser desorption ionization–mass spectrometry imaging (MALDI-MSI) PY - 2019 DO - https://doi.org/10.1007/s00216-018-1477-9 SN - 1618-2642 SN - 1618-2650 VL - 411 IS - 3 SP - 629 EP - 637 PB - Springer AN - OPUS4-47371 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tukhmetova, Dariya A1 - Lisec, Jan A1 - Vogl, Jochen A1 - Meermann, Björn T1 - Development of an Online Isotope Dilution CE/ICP–MS Method for the Quantification of Sulfur in Biological Compounds N2 - We report an analytical methodology for the quantification of sulfur in biological molecules via a speciesunspecific postcolumn isotope dilution (online ID) approach using capillary electrophoresis (CE) coupled online with inductively coupled plasma−mass spectrometry (online ID CE/ICP−MS). The method was optimized using a mixture of standard compounds including sulfate, methionine, cysteine, cystine, and albumin, yielding compound recoveries between 98 and 105%. The quantity of sulfur is further converted to the quantity of the compounds owing to the prior knowledge of the sulfur content in the molecules. The limit of detection and limit of quantification of sulfur in the compounds were 1.3−2.6 and 4.1−8.4 mg L−1, respectively, with a correlation coefficient of 0.99 within the concentration range of sulfur of 5−100 mg L−1. The capability of the method was extended to quantify albumin in its native matrix (i.e., in serum) using experimentally prepared serum spiked with a pure albumin standard for validation. The relative expanded uncertainty of the method for the quantification of albumin was 6.7% (k = 2). Finally, we tested the applicability of the method on real samples by the analysis of albumin in bovine and human sera. For automated data assessment, a software application (IsoCor) which was developed by us in a previous work was developed further for handling of online ID data. The method has several improvements compared to previously published setups: (i) reduced adsorption of proteins onto the capillary wall owing to a special capillary-coating procedure, (ii) baseline separation of the compounds in less than 30 min via CE, (iii) quantification of several sulfur species within one run by means of the online setup, (iv) SI traceability of the quantification results through online ID, and (v) facilitated data processing of the transient signals using the IsoCor application. Our method can be used as an accurate approach for quantification of proteins and other biological molecules via sulfur analysis in complex matrices for various fields, such as environmental, biological, and pharmaceutical studies as well as clinical diagnosis. Sulfur is an essential element in living organisms, where it plays important roles in various biological processes, such as protein synthesis, enzyme activity, and antioxidant defense. However, the biological effects of different sulfur species can vary widely, and imbalances in sulfur speciation have been observed in a range of diseases, including cancer, Alzheimer’s disease, and diabetes.1−3 The accurate quantification of sulfur and its species in biological samples requires sensitive and selective analytical techniques. In recent years, separation techniques coupled online with inductively coupled plasma−mass spectrometry (ICP−MS) have emerged as powerful online analytical tools complementary to molecular spectrometric methods for speciation analysis of biological compounds. External calibration4−9 and isotope dilution (ID)10−15 are common calibration approaches applied for online quantification of sulfur species in complex samples. The ID analysis is advantageous over. KW - Analytical Chemistry KW - CE/MC-ICP-MS KW - species-specific isotope information PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-594736 DO - https://doi.org/10.1021/acs.analchem.3c03553 SN - 0003-2700 SP - 1 EP - 8 PB - American Chemical Society (ACS) AN - OPUS4-59473 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tukhmetova, Dariya A1 - Lisec, Jan A1 - Vogl, Jochen A1 - Meermann, Björn T1 - Data processing made easy: standalone tool for automated calculation of isotope ratio from transient signals – IsoCor N2 - Despite numerous advantages offered by hyphenation of chromatography and electrokinetic separation methods with multicollector (MC) ICP-MS for isotope analysis, the main limitation of such systems is the decrease in precision and increase in uncertainty due to generation of short transient signals. To minimize this limitation, most authors compare several isotope ratio calculation methods and establish a multi-step data processing routine based on the precision and accuracy of the methods. However, to the best of our knowledge, there is no universal data processing tool available that incorporates all important steps of the treatment of the transient signals. Thus, we introduce a data processing application (App) IsoCor that facilitates automatic calculation of isotope ratios from transient signals and eases selection of the most suitable method. The IsoCor App performs baseline subtraction, peak detection, mass bias correction, isotope ratio calculation and delta calculation. The feasibility and reliability of the App was proven by reproducing the results from isotope analysis of three elements (neodymium, mercury and sulfur) measured on-line via hyphenated systems. The IsoCor App provides trackability of the results to ensure quality control of the analysis. KW - Isotope ratio KW - Transient signal KW - MC-ICP-MS KW - Data processing KW - App PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-559068 DO - https://doi.org/10.1039/D2JA00208F VL - 37 IS - 11 SP - 2401 EP - 2409 PB - Royal Society of Chemistry AN - OPUS4-55906 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Tvrdonova, M. A1 - Vlcnovska, M. A1 - Pompeiano Vanickova, L. A1 - Kanicky, V. A1 - Adam, V. A1 - Ascher, Lena A1 - Jakubowski, Norbert A1 - Vaculovicova, M. A1 - Vaculovic, T. T1 - Gold nanoparticles as labels for immunochemical analysis using laser ablation inductively coupled plasma mass spectrometry N2 - In this paper, we describe the labelling of antibodies by gold nanoparticles (AuNPs) with diameters of 10 and 60 nm with detection by laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS). Additionally, the AuNPs labelling strategy is compared with commercially available labelling reagents based on MeCAT (metal coded affinity tagging). Proof of principle experiments based on dot blot experiments were performed. The two labelling methods investigated were compared by sensitivity and limit of detection (LOD). The absolute LODs achieved were in the range of tens of picograms for AuNP labelling compared to a few hundred picograms by the MeCAT labelling. KW - Nanoparticle KW - LA-ICP-MS KW - Labeling PY - 2019 DO - https://doi.org/10.1007/s00216-018-1300-7 SN - 1618-2642 SN - 1618-2650 VL - 411 IS - 3 SP - 559 EP - 564 PB - Springer-Verlag GmbH CY - Berlin, Heidelberg AN - OPUS4-47093 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogel, Christian A1 - Roesch, Philipp A1 - Wittwer, Philipp A1 - Piechotta, Christian A1 - Lisec, Jan A1 - Sommerfeld, Thomas A1 - Kluge, Stephanie A1 - Herzel, Hannes A1 - Huthwelker, T. A1 - Borca, C. A1 - Simon, Franz-Georg T1 - Levels of per- and polyfluoroalkyl substances (PFAS) in various wastewater-derived fertilizers - Analytical investigations from different perspectives N2 - Solid wastewater-based fertilizers were screened for per- and polyfluoroalkyl substances (PFAS) by the extractable organic fluorine (EOF) sum parameter method. The EOF values for ten sewage sludges from Germany and Switzerland range from 154 to 7209 mg kg−1. For thermal treated sewage sludge and struvite the EOF were lower with values up to 121 mg kg−1. Moreover, the application of PFAS targeted and suspect screening analysis of selected sewage sludge samples showed that only a small part of the EOF sum parameter values can be explained by the usually screened legacy PFAS. The hitherto unknown part of EOF sum parameter contains also fluorinated pesticides, pharmaceutical and aromatic compounds. Because these partly fluorinated compounds can degrade to (ultra-)short PFAS in wastewater treatment plants they should be considered as significant sources of organic fluorine in the environment. The combined results of sum parameter analysis and suspect screening reveal the need to update current regulations, such as the German fertilizer ordinance, to focus not solely on a few selected PFAS such as perfluorooctane sulfonic acid (PFOS) and perfluorooctanoic acid (PFOA) but consider an additional sum parameter approach as a more holistic alternative. Moreover, diffusion gradient in thin-films (DGT) passive samplers were utilized as an alternative simplified extraction method for PFAS in solid wastewater-based fertilizers and subsequently quantified via combustion ion chromatography. However, the DGT method was less sensitive and only comparable to the EOF values of the fertilizers in samples with >150 mg kg−1, because of different diffusion properties for various PFAS, but also kinetic exchange limitations. KW - Combustion ion chromatography KW - Per- and polyfluoroalkyl substances (PFAS) KW - Sewage sludge KW - Fertilizer PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-583429 DO - https://doi.org/10.1039/d3va00178d SN - 2754-7000 VL - 2 IS - 10 SP - 1436 EP - 1445 PB - Royal Society of Chemistry (RSC) CY - London AN - OPUS4-58342 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen T1 - The triple-isotope calibration approach: a universal and standard-free calibration approach for obtaining absolute isotope ratios of multi-isotopic elements N2 - The theory of a new calibration approach for obtaining absolute isotope ratios of multi-isotopic elements without the use of any standard has been developed. The calibration approach basically uses the difference in the instrumental isotope fractionation of two different types of mass spectrometers, leading to two different fractionation lines in a three-isotope diagram. When measuring the same sample with both mass spectrometers, the different fractionation lines have one point in common: this is the ‘true’ logarithmized isotope ratio pair of the sample. Thus, the intersection of both fractionation lines provides us with the absolute isotope ratios of the sample. This theory has been tested in practice by measuring Cd and of Pb isotope ratios in the certified reference materials BAM-I012 and NIST SRM981 by thermal ionization mass spectrometry and by inductively coupled plasma mass spectrometry while varying the ionization conditions for both mass spectrometers. With this experiment, the theory could be verified, and absolute isotope ratios were obtained, which were metrologically compatible with the certified isotope ratios. The so-obtained absolute isotope ratios are biased by − 0.5% in average, which should be improved with further developments of the method. This calibration approach is universal, as it can be applied to all elements with three or more isotopes and it is not limited to the type of mass spectrometers applied; it can be applied as well to secondary ion mass spectrometry or others. Additionally, this approach provides information on the fractionation process itself via the triple-isotope fractionation exponent θ. KW - Triple isotope fractionation KW - Absolute isotope ratio KW - Mass spectrometry KW - Calibration KW - Uncertainty PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-516364 DO - https://doi.org/10.1007/s00216-020-03050-4 VL - 413 IS - 3 SP - 821 EP - 826 PB - Springer Verlag AN - OPUS4-51636 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Kipphardt, Heinrich A1 - Richter, Silke A1 - Bremser, Wolfram A1 - Arvizu Torres, M. A1 - Lara Manzano, J. A1 - Buzoianu, M. A1 - Hill, S. A1 - Petrov, P. A1 - Goenaga-Infante, H. A1 - Sargent, M. A1 - Fisicaro, P. A1 - Labarraque, G. A1 - Zhou, T. A1 - Turk, G. C. A1 - Winchester, M. A1 - Miura, T. A1 - Methven, B. A1 - Sturgeon, R. A1 - Jährling, R. A1 - Rienitz, O. A1 - Mariassy, M. A1 - Hankova, Z. A1 - Sobina, E. A1 - Krylov, A. I. A1 - Kustikov, Y. A. A1 - Smirnov, V. V. T1 - Establishing comparability and compatibility in the purity assessment of high purity zinc as demonstrated by the CCQM-P149 intercomparison N2 - For the first time, an international comparison was conducted on the determination of the purity of a high purity element. Participants were free to choose any analytical approach appropriate for their institute’s applications and services. The material tested was a high purity zinc, which had earlier been assessed for homogeneity and previously used in CCQM-K72 for the determination of six defined metallic impurities. Either a direct metal assay of the Zn mass fraction was undertaken by EDTA titrimetry, or an indirect approach was used wherein all impurities, or at least the major ones, were determined and their sum subtracted from ideal purity of 100 %, or 1 kg/kg. Impurity assessment techniques included glow discharge mass spectrometry, inductively coupled plasma mass spectrometry and carrier gas hot extraction/combustion analysis. Up to 91 elemental impurities covering metals, non-metals and semi-metals/metalloids were quantified. Due to the lack of internal experience or experimental capabilities, some participants contracted external laboratories for specific analytical tasks, mainly for the analysis of non-metals. The reported purity, expressed as zinc mass fraction in the high purity zinc material, showed excellent agreement for all participants, with a relative standard deviation of 0.011 %. The calculated reference value, w(Zn) = 0.999 873 kg/kg, was assigned an asymmetric combined uncertainty of + 0.000025 kg/kg and – 0.000028 kg/kg. Comparability amongst participating metrology institutes is thus demonstrated for the purity determination of high purity metals which have no particular difficulties with their decomposition / dissolution process when solution-based analytical methods are used, or which do not have specific difficulties when direct analysis approaches are used. Nevertheless, further development is required in terms of uncertainty assessment, quantification of non-metals and the determination of purity of less pure elements and/or for those elements suffering difficulties with the decomposition process. KW - Purity assessment KW - Direct metal assay KW - Impurity assessment KW - Non-metal analysis KW - High-purity elements KW - SI-traceability PY - 2018 DO - https://doi.org/10.1088/1681-7575/aaa677 SN - 1681-7575 SN - 0026-1394 VL - 55 IS - 2 SP - 211 EP - 221 PB - Institute of Physics Publishing CY - Bristol AN - OPUS4-44257 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Paz, B. A1 - Völling, E. T1 - On the ore provenance of the Trojan silver artefacts N2 - Lead isotopes are a well-established tool to trace the geographic origin of samples and artefacts in archaeology and geochemistry. In archaeology, lead isotopes are often applied to gain information on the provenance of the used ores especially in lead and silver artefacts. The assignment of a specific and unambiguous provenance in most cases is not possible or at least hindered due to several limitations such as ore deposits overlapping in their lead isotopic composition, a large spread within one ore deposit or a missing overlap with known mining sites. Such difficult cases can only be solved by using information from sources being independent of the isotope data. This information can be of chemical nature such as concentrations of key elements or they can be of archaeological nature such as cultural or trade route information.Within this study, we combined lead isotope data of ores and artefacts with silver mass fractions in the ore deposits, Au/Ag-ratios in ores and artefacts and finally archaeological Information on the cultural context in the Mediterranean and Anatolian Region. This approach enabled us to significantly reduce the potential number of mining regions. Finally, the potential sources could be narrowed down to the three remaining locations the Central Taurus, Arap Dağ and the Eastern Troad. Beneath these three locations, the Central Taurus shows the highest probability for the geographic origin of the galena which has been used to create the Trojan silver artefacts. KW - Trojan silver artefacts KW - Priam's treasure KW - Lead isotope composition KW - Central Taurus KW - Ore provenance KW - Elemental composition PY - 2019 DO - https://doi.org/10.1007/s12520-018-0756-x SN - 1866-9557 SN - 1866-9565 VL - 11 IS - 7 SP - 3267 EP - 3277 PB - Springer Verlag AN - OPUS4-48272 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Rienitz, O. A1 - Pramann, A. A1 - Flierl, L. T1 - Scale Conversion and Uncertainty Calculations in Isotope Delta Measurements N2 - Isotope ratio applications are on the increase and a major part of which are delta measurements, because they are easier to perform than the determination of absolute isotope ratios while offering lower measurement uncertainties. Delta measurements use artefact-based scales and therefore scale conversions are required due to the lack of the scale defining standards. Such scale conversions often form the basis for comparing data being generated in numerous projects andtherefore need to be as accurate as possible. In practice, users are tempted to apply linear approximations, which are not sufficiently exact, because delta values are defined by nonlinear relationships. The bias of such approximations often is beyond typical measurement uncertainties and its extent can hardly be predicted. Therefore, exact calculations are advised. Here, the exact equations and the bias of the approximations are presented, and calculations are illustrated by real-world examples. Measurement uncertainty is indispensable in this context and therefore, its calculation is described as well for determining delta values but also for scale conversions. Approaches for obtaining a single delta measurement and for repeated measurements are presented. For the latter case, a new approach for calculating the measurement uncertainty is presented, which considers covariances between the isotope ratios. KW - Delta isotope standard KW - Delta scale KW - In-house calibration solution KW - Isotope ratios KW - Isotope reference material KW - Measurement uncertainty KW - Scale conversion PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-557254 DO - https://doi.org/10.1111/ggr.12450 SN - 1639-4488 VL - 46 IS - 4 SP - 773 EP - 787 PB - Wiley AN - OPUS4-55725 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Rosner, M. A1 - Curbera, J. A1 - Peltz, U. A1 - Peplinski, Burkhard T1 - Lead isotope analysis in magic artefacts from the Berlin museums N2 - A set of 59 ancient magical artefacts, mainly made of lead, was selected from the collections of the Staatliche Museen zu Berlin in order to unravel their origins. All the selected artefacts have been studied for their Pb isotope compositions, which covered the whole range of the Mediterranean ore deposits. However, the majority (≈86%) were made of lead matching the small compositional range of the Laurion ore deposits. Only eight out of the 59 artefacts were made of recycled lead or lead from other ore deposits. Additionally, all but two were approximately dated based on their inscriptions. The lead isotopic composition together with information obtained from the inscriptions, the resulting dating, the context of the find and the known history of each item allowed us to gain more detailed information about the origins of these magical artefacts. The Attic provenance of 36 curse tablets was confirmed, whereas for 11 curse tablets previously classified as non-Attic, the provenance was either confirmed and specified (six artefacts) or changed to Attic (five artefacts). Surprisingly, the majority (six out of eight) of the analysed curse tablets from the Egyptian collection showed a lead isotopic composition closely matching that of Laurion. A Laurion-like lead isotopic composition was also observed for three of the four analysed oracular tablets from Dodona. Together with the dating information, this points to Laurion as the major and dominant lead source in the Aegean, at least during the fourth–third century B.C. The few curse tablets from earlier than the fourth–third century B.C. point to the use of multiple and thus isotopically more variable lead sources compared with the Roman times. KW - Lead isotopes KW - Pb isotopes KW - Greek curse tablets KW - Antikensammlung Berlin KW - Ägyptisches Museum Berlin PY - 2018 DO - https://doi.org/10.1007/s12520-016-0445-6 SN - 1866-9557 SN - 1866-9565 VL - 10 IS - 5 SP - 1111 EP - 1127 PB - Springer Verlag CY - Berlin AN - OPUS4-45628 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Rosner, M. A1 - Kasemann, S. A. A1 - Kraft, R. A1 - Meixner, A. A1 - Noordmann, J. A1 - Rabb, S. A1 - Rienitz, O. A1 - Schuessler, J. A. A1 - Tatzel, Michael A1 - Vocke, R. D. T1 - Intercalibration of Mg isotope delta scales and realisation of SI traceability for Mg isotope amount ratios and isotope delta values N2 - The continuous improvement of analytical procedures using multi-collector technologies in ICP-mass spectrometry has led to an increased demand for isotope standards with improved homogeneity and reduced measurement uncertainty. For magnesium, this has led to a variety of available standards with different quality levels ranging from artefact standards to isotope reference materials certified for absolute isotope ratios. This required an intercalibration of all standards and reference materials, which we present in this interlaboratory comparison study. The materials Cambridge1, DSM3, ERMAE143, ERM-AE144, ERM-AE145, IRMM-009 and NIST SRM 980 were cross-calibrated with expanded measurement uncertainties (95% confidence level) of less than 0.030‰ for the δ25/24Mg values and less than 0.037‰ for the δ26/24Mg values. Thus, comparability of all magnesium isotope delta (δ) measurements based on these standards and reference materials is established. Further, ERM-AE143 anchors all magnesium δ-scales to absolute isotope ratios and therefore establishes SI traceability, here traceability to the SI base unit mole. This applies especially to the DSM3 scale, which is proposed to be maintained. With ERM-AE144 and ERM-AE145, which are product and educt of a sublimation-condensation process, for the first time a set of isotope reference materials is available with a published value for the apparent triple isotope fractionation exponent θapp, the fractionation relationship ln α(25/24Mg)/ln α(26/24Mg). KW - Delta scale KW - Traceability KW - Scale anchor KW - Absolute isotope ratio KW - Comparability KW - Triple isotope fractionation PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-511557 DO - https://doi.org/10.1111/ggr.12327 SN - 1751-908X VL - 44 IS - 3 SP - 439 EP - 457 PB - Wiley-Blackwell CY - Oxford AN - OPUS4-51155 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Vogl, Jochen A1 - Yim, Y.-H. A1 - Lee, K.-S. A1 - Goenaga-Infante, H A1 - Malinovskiy, D. A1 - Hill, S. A1 - Ren, T. A1 - Wang, J. A1 - Vocke, R. D. A1 - Murphy, K. E. A1 - Nonose, N. A1 - Rienitz, O. A1 - Noordmann, J. T1 - Certification of ERM-EB400, the first matrix reference material for lead isotope amount ratios, and ERM-AE142, a lead solution providing a lead isotopic composition at the edge of natural variation N2 - Lead isotope amount ratios are commonly used in diverse fields such as archaeometry, geochemistry and forensic science. Currently, five reference materials with certified lead isotope amount ratios are available, namely NIST SRM 981, 982 and 983, GBW-04442 and NMIJ 3681-a. Only NIST SRM 981 and NMIJ 3681-a have approximately natural isotopic compositions, and NIST SRM 981 is predominantly used for correcting mass discrimination/mass fractionation in the applied mass spectrometric procedures. Consequently, there is no other certified reference material available to be used for validation and/or quality control of the analytical procedures applied to lead isotope amount ratio measurements. To fill this gap, two new reference materials have been produced and certified for their lead isotope amount ratios. For both certified reference materials, complete uncertainty budgets have been calculated and SI traceability has been established. This provides the users with independent means for validating and verifying their analytical procedures and for conducting quality control measures. ERM-EB400 is a bronze material with a nominal lead mass fraction of 45 mg kg-1 and certified lead isotope amount ratios of n(206Pb)/n(204Pb) = 18.072(17) mol mol-1, n(207Pb)/n(204Pb) = 15.578(18) mol mol-1 and n(208Pb)/n(204Pb) = 38.075(46) mol mol-1 with the associated expanded uncertainties (k = 2) given in brackets. ERM-AE142 is a high-purity solution of lead in 2% nitric acid with a nominal mass fraction of 100 mg kg-1 and certified Pb isotope amount ratios of n(206Pb)/n(204Pb) = 21.114(17) mol mol-1, n(207Pb)/n(204Pb) = 15.944(17) mol mol-1 and n(208Pb)/n(204Pb) = 39.850(44) mol mol-1 with the associated expanded uncertainties (k = 2) given in brackets. Both materials are specifically designed to fall within the natural lead isotopic variation and to assist users with the validation and verification of their analytical procedures. Note that while one of these reference materials requires the chemical separation of Pb from its matrix (ERM-EB400), the other does not (ERM-AE142). As additional information, δ208/206PbNIST SRM981 values are provided for both materials. For ERM-AE142, a delta value of δ208/206PbNIST SRM981 = -28.21(30) ‰ was obtained, and for ERM-EB400, a delta value of δ208/206PbNIST SRM981 = -129.47(38) ‰ was obtained, with the associated expanded uncertainties (k = 2) given in brackets. KW - Lead isotope variations KW - Radiogenic isotopes KW - Isotope reference material KW - Metrology in chemistry KW - Measurement uncertainty PY - 2019 DO - https://doi.org/10.1111/ggr.12253 SN - 1751-908X SN - 1639-4488 VL - 43 IS - 1 SP - 23 EP - 37 PB - John Wiley & Sons AN - OPUS4-47383 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Von der Au, M. A1 - Borovinskaya, O. A1 - Büchel, C. A1 - Meermann, Björn T1 - Der Fingerabdruck der Kieselalge N2 - Elementanalytische Verfahren werden heute zu mehr eingesetzt als bloß zur Bestimmung von Metallgesamtgehalten in diversen Probenmatrizes. Sie stellen heute ein wichtiges Werkzeug zur Beantwortung lebenswissenschaftlicher Fragen aus Umwelt, Medizin und Biologie dar. In der vorliegenden Arbeit kommt die single-cell-ICP-Flugzeitmassenspektrometrie (single-cell-ICP-TOF-MS) zur Multielementanalytik in einzelnen Kieselalgen (Diatomeen) zum Einsatz und wird zukünftig ein wichtiges Werkzeug z.B. bei der Beantwortung ökotoxikologischer Fragestellung sein. KW - Diatomeenanalytik KW - Single cell-ICP-ToF-MS KW - Multielementanalytik in kurzen transienten Signalen PY - 2019 UR - https://q-more.chemie.de/q-more-artikel/299/der-fingerabdruck-der-kieselalge.html SP - 1 EP - 3 PB - LUMITOS AN - OPUS4-49749 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Von der Au, M. A1 - Meermann, Björn T1 - Screening von Organofluorverbindungen im Oberflächenwasser N2 - Organofluorverbindungen kommen in vielen Bereichen zum Einsatz und gelangen in die aquatische Umwelt. Die Belastungssituation der Gewässer ist bisher wenig untersucht – hier kommt die HR-CS GF MAS als leistungsstarkes Screeningtool zum Einsatz. KW - HR-CS-GFMAS KW - Extrahierbar organisch gebundenes Fluor (EOF) KW - Oberflächengewässer PY - 2019 UR - https://analytik.news/Fachartikel/Volltext/bam5.pdf SP - 1 EP - 3 PB - Dr. Beyer Internet-Beratung AN - OPUS4-49750 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - von der Au, Marcus A1 - Borovinskaya, O. A1 - Flamigni, L. A1 - Kuhlmeier, K. A1 - Büchel, C. A1 - Meermann, Björn T1 - Single cell-inductively coupled plasma-time of flight-mass spectrometry approach for ecotoxicological testing N2 - Diatoms play a key role in assessing the eco-toxicology of metals and are already part of several national and international guidelines and regulations. Data on metal uptake and its correlation with a natural metal composition of the diatoms are mostly lacking on a cellular basis - mainly due to the lack of a suitable method on both the preparation and detection side. Therefore, within this work a fully automated approach based on the on-line coupling of a high performance liquid chromatograph (HPLC) and an inductively coupled plasma-time of flight-mass spectrometer (ICP-ToF-MS) was applied to analyze single cells of the alga Cyclotella meneghiniana multi-elementally in order to provide a deeper insight into the metal composition and its response to environmental stress. Multi-elemental analysis in single diatoms also enables assessment of combined toxicity of a set of metals. A set of four fingerprint elements, characteristically for diatoms (Mg, P, Si, Fe), were identified and hence the investigation of environmental stress onto the cells was enabled by performing incubation experiments with environmentally relevant toxic elements. It could be shown at moderate environmental stress caused by increasing the metal concentration in the medium (zinc) that the fingerprint element concentrations remained stable and thus the suitability of the selected elements for algae tracing was demonstrated. With regard to further ecotoxicological assessments, a multivariate approach was successfully applied allowing for cell classification upon different incubation concentration levels. This multivariate approach also facilitated an effective identification of three different diatom species (Cyclotella meneghiniana, Thalassiosira weissflogii and Thalassiosira pseudonana). KW - ICP-TOF-MS KW - Ecotoxicological Testing KW - Diatoms PY - 2020 DO - https://doi.org/10.1016/j.algal.2020.101964 VL - 49 SP - 101964 PB - Elsevier B.V. AN - OPUS4-50875 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - von der Au, Marcus A1 - Faßbender, Sebastian A1 - Chronakis, Michail A1 - Vogl, Jochen A1 - Meermann, Björn T1 - Size determination of nanoparticles by ICP-ToF-MS using isotope dilution in microdroplets N2 - Within this work, the combination of a microdroplet generator and an ICP-ToF-MS for nanoparticle analysis is presented. For the size determination of platinum nanoparticles an on-line isotope dilution analysis approach was developed. The 194Pt/195Pt isotopic ratio was used for the characterization of the particles, while the 182W/183W isotopic ratio was monitored simultaneously for mass bias correction. The on-line ID-MDG-sp-ICP-ToF-MS approach was deployed for the size determination of three platinum nanoparticle samples (50 nm, 63 nm, 70 nm); for validation, complementary size characterization techniques (sp-ICP-ToF-MS and TEM) were used. The robustness of this technique was evidenced, by using sodium chloride concentrations up to 100 mg L−1 as a matrix component. Our new on-line ID MDG-sp-ICP-ToF-MS approach is a promising tool for the fast and reliable determination of nanoparticles' size in severe matrix concentrations, e.g., environmental samples. KW - ICP-ToF-MS KW - Nanoparticles KW - Isotope Dilution PY - 2022 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-552727 DO - https://doi.org/10.1039/D2JA00072E SN - 0267-9477 VL - 37 IS - 6 SP - 1203 EP - 1207 PB - Royal Society of Chemistry AN - OPUS4-55272 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - von der Au, Marcus A1 - Karbach, H. A1 - Bell, A. M. A1 - Bauer, O. B. A1 - Karst, U. A1 - Meermann, Björn T1 - Determination of metal uptake in single organisms, Corophiumvolutator, via complementary electrothermal vaporization/inductively coupled plasma mass spectrometry and laserablation/inductively coupled plasma mass spectrometry N2 - Rationale: (Eco-)toxicological effects are mostly derived empirically and are notcorrelated with metal uptake. Furthermore, if the metal content is determined,mostly bulk analysis of the whole organism population is conducted; thus, biologicalvariability is completely disregarded, and this may lead to misleading results. Toovercome this issue, we compared two different solid sampling techniques for theanalysis of single organisms.Methods: In this study, complementary electrothermal vaporization/inductivelycoupled plasma mass spectrometry (ETV/ICP-MS) , laser ablation/inductivelycoupled plasma mass spectrometry (LA/ICP-MS)-based methods for the analysisof individual organisms were developed and the results obtained were comparedwith the concentrations obtained after digestion and measured using ICP-MS.For this purpose, a common (eco-)toxicological test organism, the mud shrimpCorophium volutator, was selected. As proof-of-concept application, these organismswere incubated with environmentally relevant metals from galvanic anodes, whichare often used for protection against metal corrosion in, for example, offshorewind farms.Results: The bulk analysis revealed that large quantities of the incubated elementswere detectable. Using the ETV/ICP-MS method, we could identify a highbiovariability within the population of organisms tested. Using the LA/ICP-MSmethod, it could be determined that the large quantities of the elements detectedwere due to adsorption of the metals and not due to uptake, which correlates wellwith the absence of (eco-)toxicological effects.Conclusions: The results obtained imply the efficiency of complementary methods toexplain the absence or presence of (eco-)toxicological effects. In particular, methodsthat allow for single-organism analysis or provide even a spatial resolution supportthe interpretation of ecotoxicological findings. KW - ICP-MS KW - Metal toxicity KW - Metal uptake KW - Corophium volutator KW - Seawater PY - 2021 DO - https://doi.org/10.1002/rcm.8953 VL - 35 IS - 2 SP - e8953 AN - OPUS4-52077 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - von der Au, Marcus A1 - Schwinn, M. A1 - Kuhlmeier, K. A1 - Büchel, C. A1 - Meermann, Björn T1 - Development of an automated on-line purification HPLC single cell-ICP-MS approach for fast diatom analysis N2 - The most challenging part in performing a single cell ICP-MS (sc-ICP-MS) approach is the sample preparation, in particular the reduction of the ionic background. This step is, in many cases, time-consuming and required for each sample separately. Furthermore, sc-ICP-MS measurements are mostly carried out "manually", given the fact that present systems are not allowing for an automated change of samples. Thus, within this work, we developed an approach based on a HPLC system coupled on-line with sc-ICP-MS via a set of switching valves as well as an in-line filter for automated cell washing. This set-up enables the ionic background removal as well as analysis of single cells completely automated without any manual sample pretreatment. Our approach was applied for the analysis of the single celled diatom species Cyclotella meneghiniana, a marine diatom species, on the basis of Mg24 and facilitates testing in 11 min per sample, requiring only around 10,000 cells in a volume of 10 µL and approx. 10 mL of a 5% MeOH/95% deionized water (v/v) mixture. Even at extremely saline culturing media concentrations (up to 1000 mg L-1 magnesium) our on-line approach worked sufficiently allowing for distinction of ionic and particulate fractions. Furthermore, a set of diatom samples was analyzed completely automated without the need for changing samples manually. So, utilizing this approach enables analyzing a high quantity of samples in a short time and therefore in future the investigation of ecotoxicological effects is simplified for example in terms of metal accumulation by taking biovariability into account. KW - Single cell-ICP-MS KW - Diatoms KW - Ecotoxicology testing KW - Automated system PY - 2019 DO - https://doi.org/10.1016/j.aca.2019.05.045 SN - 1873-4324 VL - 1077 SP - 87 EP - 94 PB - Elsevier AN - OPUS4-48567 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Völzke, Jule L. A1 - Smatty, Sarah A1 - Döring, Sarah A1 - Ewald, Shireen A1 - Oelze, Marcus A1 - Fratzke, Franziska A1 - Flemig, Sabine A1 - Konthur, Zoltán A1 - Weller, Michael G. T1 - Efficient Purification of Polyhistidine-Tagged Recombinant Proteins Using Functionalized Corundum Particles N2 - Immobilized metal affinity chromatography (IMAC) is a widely used technique for purifying polyhistidine-tagged recombinant proteins. However, it often has practical limitations that require complex optimizations and additional steps for purification. In this study, we introduce functionalized corundum particles as a novel, efficient, and economical method for purifying recombinant proteins in a column-free format. The corundum surface is modified with amino silane APTES, followed by EDTA dianhydride, and then loaded with nickel ions. We used the Kaiser test to monitor the modification process and ICP-MS to quantify the metal-binding capacity. To evaluate the system, we used His-tagged protein A/G (PAG) mixed with bovine serum albumin (BSA). The corundum particles exhibited a binding capacity of approximately 3 mg of protein per gram of corundum or 2.4 mg per 1 mL of corundum suspension. We also examined cytoplasm obtained from different E. coli strains as an example of a complex matrix. Varying the imidazole concentration in the loading and washing buffers showed that higher concentrations during loading improved purity. Even with sample volumes as large as one liter, we successfully isolated recombinant proteins down to a concentration of 1 µg/mL. We found higher purity levels with corundum when comparing the corundum material to standard Ni–NTA agarose beads. We successfully purified His6-MBP-mSA2, a fusion protein comprising monomeric streptavidin and maltose-binding protein, from E. coli cytoplasm, demonstrating the method's applicability. We also purified SARS-CoV-2-S-RBD-His8 expressed in human Expi293F cells, confirming its suitability for mammalian cell culture supernatants. The material cost of the nickel-loaded corundum material (without regeneration) is estimated to be less than 30 cents per gram of functionalized support or 10 cents per milligram of isolated protein. Another advantage of this system is the exceptional physical and chemical stability of corundum particles. Overall, we have demonstrated that this novel material offers an efficient, robust, and cost-effective purification platform for His-tagged proteins, even in challenging, complex matrices and large sample volumes with low product concentrations. This method has potential applications in both small laboratories and large-scale industrial settings. N2 - Die immobilisierte Metallaffinitätschromatographie (IMAC) ist eine weit verbreitete Technik zur Reinigung von rekombinanten Proteinen mit Polyhistidin-Markierung. Sie hat jedoch oft praktische Einschränkungen, die komplexe Optimierungen und zusätzliche Schritte für die Aufreinigung erfordern. In dieser Studie stellen wir funktionalisierte Korundpartikel als neuartige, effiziente und wirtschaftliche Methode zur Reinigung rekombinanter Proteine in einem säulenfreien Format vor. Die Korundoberfläche wird mit dem Aminosilan APTES und anschließend mit EDTA-Dianhydrid modifiziert und dann mit Nickelionen beladen. Wir haben den Kaiser-Test zur Überwachung des Modifizierungsprozesses und ICP-MS zur Quantifizierung der Metallbindungskapazität verwendet. Zur Charakterisierung des Systems verwendeten wir His-markiertes Protein A/G (PAG) in Kombination mit Rinderserumalbumin (BSA). Die Korundpartikel wiesen eine Bindungskapazität von etwa 3 mg Protein pro Gramm Korund oder 2,4 mg pro 1 ml Korundsuspension auf. Als Beispiel für eine komplexe Matrix untersuchten wir auch Zytoplasma, das aus verschiedenen E. coli-Stämmen gewonnen wurde. Die Variation der Imidazolkonzentration in den Lade- und Waschpuffern zeigte, dass höhere Konzentrationen beim Laden die Reinheit verbesserten. Selbst bei einem Probenvolumen von bis zu einem Liter konnten wir rekombinante Proteine von 1 µg/mL isolieren. Beim Vergleich des Korundmaterials mit Standard-Ni-NTA-Agarose-Beads stellten wir einen höheren Reinheitsgrad mit Korund fest. Wir reinigten erfolgreich His6-MBP-mSA2 aus E. coli-Zytoplasma, ein Fusionsprotein, das aus monomerem Streptavidin und Maltose-bindendem Protein besteht. Wir reinigten auch SARS-CoV-2-S-RBD-His8, das in humanen Expi293F-Zellen exprimiert wurde, und bestätigten damit die Eignung des Materials für Zellkulturüberstände von Säugetieren. Die Materialkosten für nickelbeladenen Korund (ohne Regenerierung) werden auf weniger als 30 Cent pro Gramm funktionalisierten Trägers oder 10 Cent pro Milligramm isolierten Proteins geschätzt. Ein weiterer Vorteil dieses Systems ist die außergewöhnliche physikalische und chemische Stabilität der Korundpartikel. Insgesamt haben wir gezeigt, dass dieses neuartige Material eine effiziente, robuste und kostengünstige Reinigungsplattform für His-markierte Proteine bietet, selbst bei schwierigen, komplexen Matrices und großen Probenmengen mit niedrigen Produktkonzentrationen. Diese Methode könnte sowohl in kleinen Labors als auch in der Großindustrie eingesetzt werden. KW - Aluminum oxide KW - Sapphire KW - Ethylenediaminetetraacetic acid KW - Nickel chelate KW - EDTAD KW - HexaHis-Tag KW - His6 KW - 6xHis KW - His8 KW - Bioseparation KW - IMAC purification KW - Immunocapture KW - Affinity chromatography KW - Carrier KW - Nickel KW - Recombinant protein KW - Escherichia coli KW - Bacterial lysates KW - Cytoplasm KW - Polishing KW - Downstream processing PY - 2023 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-575672 DO - https://doi.org/10.3390/biotech12020031 VL - 12 IS - 2 SP - 1 EP - 18 PB - MDPI CY - Basel, Schweiz AN - OPUS4-57567 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Weber, A. A1 - von Randow, M. A1 - Voigt, A.-L. A1 - von der Au, Marcus A1 - Fischer, E. A1 - Meermann, Björn A1 - Wagner, M. T1 - Ingestion and toxicity of microplastics in the freshwater gastropodLymnaea stagnalis: No microplastic-induced effects alone or incombination with copper N2 - The interaction of microplastics with freshwater biota and their interaction with other stressors is still not very well understood. Therefore, we investigated the ingestion, excretion and toxicity of microplastics in the freshwater gastropod Lymnaea stagnalis. MP ingestion was analyzed as tissues levels in L. stagnalis after 6–96 h of exposure to 5–90 μm spherical polystyrene (PS) microplastics. To understand the excretion, tissue levels were determined after 24 h of exposure followed by a 12 h–7 d depuration period. To assess the toxicity, snails were exposed for 28 d to irregular PS microplastics (<63 μm, 6.4–100,000 particles mL−1), both alone and in combination with copper as additional stressor. To compare the toxicity of natural and synthetic particles, we also included diatomite particles. Microplastics ingestion and excretion significantly depended on the particle size and the exposure/depuration duration. An exposure to irregular PS had no effect on survival, reproduction, energy reserves and oxidative stress. However, we observed slight effects on immune cell phagocytosis. Exposure to microplastics did not exacerbate the reproductive toxicity of copper. In addition, there was no pronounced difference between the effects of microplastics and diatomite. The tolerance towards microplastics may originate from an adaptation of L. stagnalis to particle-rich environments or a general stress resilience. In conclusion, despite high uptake rates, PS fragments do not appear to be a relevant stressor for stress tolerant freshwater gastropods considering current environmental levels of microplastics. KW - Lymnaea stagnalis KW - Microplastic-induced effects KW - Mixture toxicity PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-513551 DO - https://doi.org/10.1016/j.chemosphere.2020.128040 VL - 263 SP - 128040 PB - Elsevier Ltd. AN - OPUS4-51355 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Weber, K. A1 - Weber, M. A1 - Menneken, M. A1 - Kral, A. G. A1 - Mertz-Kraus, R. A1 - Geisler, T. A1 - Vogl, Jochen A1 - Tütken, T. T1 - Diagenetic stability of non-traditional stable isotope systems (Ca, Sr, Mg, Zn) in teeth – An in-vitro alteration experiment of biogenic apatite in isotopically enriched tracer solution N2 - Stable isotope ratios and trace element concentrations of fossil bones and teeth are important geochemical proxies for the reconstruction of diet and past environment in archaeology and palaeontology. However, since diagenesis can significantly alter primary diet-related isotope signatures and elemental compositions, it is important to understand and quantify alteration processes. Here, we present the results of in-vitro Alteration experiments of dental tissues from a modern African elephant molar reacted in aqueous solutions at 30 °C and 90 °C for 4 to 63 days. Dental cubes with ≈ 3 mm edge length, comprising both enamel and dentin, were placed into 2 mL of acidic aqueous solution enriched in different isotopes (25Mg, 44Ca, 67Zn, 86Sr, initial pH 1). Element and isotope distribution profiles across the reacted cubes were measured with LA-(MC-)ICP-MS and EMPA, while potential effects on the bioapatite crystal structure were characterised by Raman spectroscopy. In all experiments isotope ratios measured by LA-(MC-)ICP-MS revealed an alteration of the enamel in the outer ≈ 200–300 μm. In contrast, dentin was fully altered (≈ 1.4 mm) after one week at 90 °C while the alteration did not exceed a depth of 150–200 μm during the 30 °C experiments. Then, the tracer solution started also to penetrate through the enamel-dentin junction into the innermost enamel, however, leaving the central part of the enamel unaltered, even after three months. The Raman spectra suggest an initial demineralisation in the acidic environment while organic matter (i.e. collagen) is still preserved. In the 90 °C experiment, Raman spectra of the v1 PO4) band of the dentin shift over time towards synthetic hydroxylapatite patterns and the Ca (and Sr) concentrations in the respective solutions decrease. This indicates precipitation of newly formed apatite. Isotope and element concentration profiles across the dental tissues reveal different exchange mechanisms for different isotope systems. Magnesium is leached from enamel and dentin, while Zn is incorporated into the apatite crystal structure. However, the distribution of both elements is not affected in the innermost enamel where their concentrations do not change over the whole duration of the experiments. We found no correlation of reaction depth in the cubes and experimental duration, which might be caused by natural variability of the dental material already at the beginning of the experiment. Our alteration experiments in a closed system at high temperatures ≤90 °C and low initial pH demonstrate that at least the central part of mm-thick mammalian enamel apatite seems to be resistant against alteration preserving its pristine bioapatite mineral structure as well as its in-vivo elemental and isotopic composition. The experiments assess diagenetic alteration in a novel multi-proxy approach using in-situ analyses in high spatial resolution. It is demonstrated that the isotopes of Ca, Sr, Zn and Mg in the dentin are prone for diagenetic alteration, while enamel is more resistant against alteration and could be used for dietary and physiological reconstructions in fossil teeth. KW - Bioapatite KW - Isotopes KW - Raman spectroscopy KW - Diagenesis KW - LA-(MC-)ICP-MS KW - EPMA PY - 2021 DO - https://doi.org/10.1016/j.chemgeo.2021.120196 VL - 572 SP - 120196 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-52447 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Winckelmann, Alexander A1 - Morcillo Garcia-Morato, Dalia A1 - Richter, Silke A1 - Recknagel, Sebastian A1 - Riedel, Jens A1 - Vogl, Jochen A1 - Panne, Ulrich A1 - Abad Andrade, Carlos Enrique T1 - Determination of lithium in human serum by isotope dilution atomic absorption spectrometry N2 - The therapeutic dose of lithium (Li) compounds, which are widely used for the treatment of psychiatric and hematologic disorders, is close to its toxic level; therefore, drug monitoring protocols are mandatory. Herein, we propose a fast, simple, and low-cost analytical procedure for the traceable determination of Li concentration in human serum, based on the monitoring of the Li isotope dilution through the partially resolved isotope shift in its electronic transition around 670.80 nm using a commercially available high-resolution continuum source graphite furnace atomic absorption spectrometer. With this technique, serum samples only require acidic digestion before analysis. The procedure requires three measurements—an enriched 6Li spike, a mixture of a certified standard solution and spike, and a mixture of the sample and spike with a nominal 7Li/6Li ratio of 0.82. Lanthanum has been used as an internal spectral standard for wavelength correction. The spectra are described as the linear superposition of the contributions of the respective isotopes, each consisting of a spin-orbit doublet, which can be expressed as Gaussian components with constant spectral position and width and different relative intensity, reflecting the isotope ratio in the sample. Both the spectral constants and the correlation between isotope ratio and relative band intensity have been experimentally obtained using commercially available materials enriched with Li isotopes. The Li characteristic mass (mc) obtained corresponds to 0.6 pg. The procedure has been validated using five human serum certified reference materials. The results are metrologically comparable and compatible to the certified values. The measurement uncertainties are comparable to those obtained by the more complex and expensive technique, isotope dilution mass spectrometry. KW - Lithium KW - Human serum KW - Isotope dilution KW - Atomic absorption spectrometry KW - High-resolution continuum source graphite furnace atomic absorption spectrometry PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-532446 DO - https://doi.org/10.1007/s00216-021-03636-6 VL - 414 IS - 1 SP - 251 EP - 256 PB - Springer CY - Berlin AN - OPUS4-53244 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -