TY - JOUR A1 - Faßbender, Sebastian A1 - Rodiouchkina, K. A1 - Vanhaecke, F. A1 - Meermann, Björn T1 - Method development for on-line species-specific sulfur isotopic analysis by means of capillary electrophoresis/multicollector ICP-mass spectrometry N2 - In this work, a method for species-specific isotopic analysis of sulfur via capillary electrophoresis hyphenated on-line with multicollector ICP-MS (CE/MC-ICP-MS) was developed. Correction for the mass bias caused by instrumental mass discrimination was realized via external correction with multiple-injection sample-standard bracketing. By comparing the isotope ratio measurement results obtained using the newly developed on-line CE/MC-ICP-MS method with those obtained via traditional MC-ICP-MS measurement after analyte/matrix separation by anion exchange chromatography for isotopic reference materials and an in-house bracketing standard, the most suitable data evaluation method could be identified. The repeatability for the sulfate-δ34S value (calculated from 18 measurements of a standard conducted over seven measurement sessions) was 0.57‰ (2SD) and thereby only twice that obtained with off-line measurements (0.30‰, n = 68). As a proof of concept for analysis of samples with a real matrix, the determination of the sulfur isotopic composition of naturally present sulfate was performed for different river systems. The CE/MC-ICP-MS results thus obtained agreed with the corresponding off-line MC-ICP-MS results within the 2SD ranges, and the repeatability of consecutive δ34S measurements (n = 3) was between 0.3‰ and 1.3‰ (2SD). Finally, the isotopic analysis of two different S-species in a river water sample spiked with 2-pyridinesulfonic acid (PSA) was also accomplished. KW - River water sulfate KW - Environmental speciation KW - Sulfur isotopes KW - On-line CE/MC-ICP-MS KW - Multiple-injection sample-standard bracketing PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-509549 DO - https://doi.org/10.1007/s00216-020-02781-8 SN - 1618-2642 VL - 412 IS - 23 SP - 5637 EP - 5646 PB - Springer CY - Berlin, Heidelberg AN - OPUS4-50954 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Meermann, Björn T1 - Metal based pollutant assessment via diatoms - new possibilities via automated single cell-ICP-ToF-MS N2 - Diatoms are located at the bottom of the food chain. Toxicological relevant metals can possibly accumulate within the food web and cause harmful effects. Diatoms are a test system in ecotoxicology. Toxicological effects weaken the growth of algae which is by default investigated by means of fluorescence detection. On basis of the expose concentration as well as obtained fluorescence data potential threshold exceedance in e.g. surface waters is assessed. However, this approach does not allow for the determination of “real” accumulated metal concentration in diatoms. Common approaches are based on bulk analysis via e.g. ICP-MS, ICP-OES or AAS. But, biological variability is completely disregarded. To tackle this problem, alternative approaches are highly needed. Within the last years, sp-ICP-MS for nanoparticle as well as single cell analysis turned out as a powerful technique to analyze metal contents as well as size distributions on broad size range. But, common ICP-MS systems do not allow for multi-element detection within single particle/cell events. Thus, simultaneous MS detection devices are needed - just recently, ICP-ToF-MS experienced a revival. Within our previous work, we developed an automated sample introduction system based on a HPLC system on-line with single particle-ICP-MS, which allowed for ionic background separation and single algae analysis. However, for unambiguous tracing several fingerprint elements and multielement analysis in single algae (diatoms) is needed. Thus, we coupled our previous setup on-line to ICP-ToF-MS. Test diatom species were exposed to test substances (Zn) as well as nanoparticles (FeNPs). The developed setup allowed for a fast, automated and multielement analysis in single diatoms. Furthermore, we combined our approach with multivariate data assessment - multielement detection of characteristic fingerprint elements allowed for an unambiguous diatom tracing. Clustering of diatoms according to metal exposure concentration levels was enabled. Our approach is a new potential tool in ecotoxicological testing. T2 - 53. DGMS & 27th ICP-MS user meeting CY - Münster, Germany DA - 01.03.2020 KW - Single cell-ICP-ToF-MS KW - Diatoms KW - New ecotoxicological testing tools PY - 2020 AN - OPUS4-50644 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Alfeld, M. A1 - Eckhardt, H.-S. A1 - Kraft, J. A1 - Maiwald, Michael A1 - Meermann, Björn A1 - Merz, K. A1 - Prikler, S. A1 - Richert, J. A1 - Steiner, G. A1 - von Tümpling, W. T1 - Trendbericht Analytische Chemie N2 - Zusätzlich zu Methodenentwicklung, Miniaturisierung und Kopplungsverfahren zeigen sich die Hyperspektroskopie zusammen mit Imaging‐Verfahren, der Einzelmolekülnachweis und der Einsatz von 3‐D‐Druckern als neue Schwerpunkte. Hinzu kommen künstliche Intelligenz bei Sensoren, Bildgebungsverfahren und Prozesssteuerung sowie die Vernetzung von Analyse‐ und Laborgeräten. Trends und Forschungsthemen aus der analytischen Chemie, zusammengestellt von elf Autoren, koordiniert von Günter Gauglitz. KW - Multielementanalytik KW - Multimodale Analytik KW - Kristallolgraphie KW - Prozessanalytik KW - Prozessindustrie KW - Industrielle Analytik KW - Chemometrik KW - Chemometrie PY - 2020 DO - https://doi.org/10.1002/nadc.20204095786 SN - 1868-0054 VL - 68 IS - 4 SP - 52 EP - 60 PB - Wiley CY - Weinheim AN - OPUS4-50609 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Maiwald, Michael T1 - The internet of things in the lab and in process - The digital transformation challenges for the laboratory 4.0 T1 - Das Internet of Things in Labor und Prozess - Herausforderungen des digitalen Wandels für das Labor 4.0 N2 - It is a fact that much of the time spent on analytical laboratory instrumentation these days goes into system maintenance. Digital transformation could give us more time again for creativity and our actual laboratory work – if we shape it the right way. N2 - Fakt ist: Einen Großteil der Zeit, der an analytischen Laborgeräten verbracht wird, nimmt heute die Systempflege in Anspruch. Der digitale Wandel kann uns endlich wieder mehr Zeit für Kreativität und die eigentliche Laborarbeit geben – wenn wir ihn richtig gestalten. KW - Lab of the Future KW - Digitalisation KW - Automation KW - Data Analysis KW - Instrument Communication KW - Labor der Zukunft KW - Digitale Transformation KW - Automatisierung KW - Gerätekommunikation PY - 2020 IS - 4 SP - 1 EP - 3 PB - Lumitos AG CY - Darmstadt AN - OPUS4-50618 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von der Au, Marcus T1 - SPE HR-CS GF MAS – A New Sum Parameter Method for Extractable Organically Bound Fluorine (EOF) in Surface Water N2 - Introducing fluorine into organic molecules leads to new chemical/physical properties. Up to now, the OECD identified and categorized 4730 per- and polyfluoroalkyl substances-related CAS numbers. Especially in the field of technical applications (e.g. surface coatings, fire extinguishing foams) as well as pharmaceuticals, fluorinated substances gain in importance. Thus, an increasing release of fluorinated compounds into the environment is expected. The high environmental persistence of perfluorinated compounds leads to the risk of bioaccumulation. Partially fluorinated substances (polyfluorinated compounds) undergo degradation; thus, further possible fluorine species occur, which may exhibit different toxic/chemical properties. However, current target methods (e.g., HPLC/MS-MS) are not applicable for a comprehensive screening as well as assessment of pollution. Thus, the poster presents a new sum parameter method for quantitative determination of extractable organically bound fluorine (EOF) in surface water samples. The method is based on solid-phase extraction (SPE) for fluorinated compounds as well as quantitative separation of interfering inorganic fluoride in combination with high-resolution-continuum source graphite furnace molecular absorption spectrometry (HR-CS GF MAS) for quantitative analysis. By means of this technique, the detection of high resolution molecular absorption spectra (MAS) is enabled upon the addition of a modifier and the formation of a diatomic molecule (e.g. GaF) After successful optimization of the SPE procedure (maximum concentration of extractable organic fluorine), enrichment factors of about 1000 were achieved, allowing for highly sensitive fluorine detection. Next to a species-unspecific response, limits of detection in the low nanogram per liter range were achieved and real surface water samples were analyzed. EOF values in the range of about 50-300 ng/L were detected. The developed method allows for a fast and sensitive as well as selective screening of organically bound fluorine (EOF) in surface water samples, helping to elucidate pollution hotspots as well as discharge routes. T2 - 2020 Winter Conference on Plasma Spectrochemistry CY - Tucson, AZ, USA DA - 12.01.2020 KW - SPE HR-CS GF MAS KW - Sum parameter KW - Fluorine Detection PY - 2020 AN - OPUS4-50352 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Hampel, U. A1 - Schütze, A. A1 - Rädle, M. A1 - Rück, T. A1 - Krawczyk-Becker, M. A1 - Musch, T. A1 - Maiwald, Michael A1 - Fröhlich, H. J. A1 - Zeck, S. T1 - Positionspapier Sensorik für die Digitalisierung chemischer Produktionsanlagen N2 - Die chemische Industrie steht derzeit, wie viele andere Industriebereiche, vor den Herausforderungen einer Digitalisierung der Produktion. Sie ist der Schlüssel für die Flexibilisierung von Prozessen und Anlagen, für die Verkürzung von Produkteinführungszeiten sowie für den Zuschnitt der Produktion auf wechselnde Nachfrage und kürzere Produktlebenszyklen. Die Messtechnik und Sensorik spielt neben der intelligenten Datenverarbeitung eine Schlüsselrolle für die Digitalisierung. Flexiblere Anlagen benötigen Sensorik zur Überwachung des Anlagenzustandes, zur Früherkennung nicht bestimmungsgemäßer Betriebszustände sowie für eine bedarfsgerechte Wartung. Da die Entwicklung neuer und verbesserter Messtechnik und Sensorik grundlegend aus verschiedenen Richtungen gedacht werden muss, haben sich Akteure aus verschiedenen Branchen zusammengetan und dieses Positionspapier erstellt. Es basiert auf einer grundlegenden Analyse des Ist-Stands sowie des Bedarfs der Industrie, die unter anderem auf einem eigens dafür durchgeführten Workshop mit Sensorentwicklern, Anlagenherstellern sowie Anlagenbetreibern am 18. Juni 2019 bei der DECHEMA in Frankfurt a. M. diskutiert wurden. Diese Aktivitäten wurden maßgeblich von der Initiative Wanted Technologies der ProcessNet sowie dem AMA Verband für Sensorik und Messtechnik e.V. initiiert. KW - Prozessindustrie KW - Smarte Sensoren KW - Prozessanalytik KW - DECHEMA KW - Positionspapier PY - 2020 UR - https://dechema.de/Sensorik SP - 1 EP - 20 PB - DECHEMA CY - Frankfurt am Main AN - OPUS4-50403 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - von der Au, Marcus A1 - Borovinskaya, O. A1 - Flamigni, L. A1 - Kuhlmeier, K. A1 - Büchel, C. A1 - Meermann, Björn T1 - Single cell-inductively coupled plasma-time of flight-mass spectrometry approach for ecotoxicological testing N2 - Diatoms play a key role in assessing the eco-toxicology of metals and are already part of several national and international guidelines and regulations. Data on metal uptake and its correlation with a natural metal composition of the diatoms are mostly lacking on a cellular basis - mainly due to the lack of a suitable method on both the preparation and detection side. Therefore, within this work a fully automated approach based on the on-line coupling of a high performance liquid chromatograph (HPLC) and an inductively coupled plasma-time of flight-mass spectrometer (ICP-ToF-MS) was applied to analyze single cells of the alga Cyclotella meneghiniana multi-elementally in order to provide a deeper insight into the metal composition and its response to environmental stress. Multi-elemental analysis in single diatoms also enables assessment of combined toxicity of a set of metals. A set of four fingerprint elements, characteristically for diatoms (Mg, P, Si, Fe), were identified and hence the investigation of environmental stress onto the cells was enabled by performing incubation experiments with environmentally relevant toxic elements. It could be shown at moderate environmental stress caused by increasing the metal concentration in the medium (zinc) that the fingerprint element concentrations remained stable and thus the suitability of the selected elements for algae tracing was demonstrated. With regard to further ecotoxicological assessments, a multivariate approach was successfully applied allowing for cell classification upon different incubation concentration levels. This multivariate approach also facilitated an effective identification of three different diatom species (Cyclotella meneghiniana, Thalassiosira weissflogii and Thalassiosira pseudonana). KW - ICP-TOF-MS KW - Ecotoxicological Testing KW - Diatoms PY - 2020 DO - https://doi.org/10.1016/j.algal.2020.101964 VL - 49 SP - 101964 PB - Elsevier B.V. AN - OPUS4-50875 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von der Au, Marcus T1 - HR-CS-GF-MAS as a new screening method for emerging pollutants - per- and polyfluorinated substances in the environment N2 - The introduction of fluorine in organic molecules leads to new chemical/physical properties. Especially in the field of pharmaceuticals, fluorinated organic molecules are becoming more and more popular and at present amount up to 25% of market share, with an upward trend. The main benefits of fluorinated pharmaceuticals are: (i) enhanced fat solubility; (ii) enhanced interaction of catalytic-center of enzymes with fluorine-drugs; (iii) a delayed metabolism within the human body. Highly fluorinated organic substances are also used in technical applications (e.g. coatings, fire-extinguishing agents). Due to the broad variety of fluorinated substances and increasing production volumes numerous and up to date unknown fluorine-species are most likely to be present in the (aquatic) environment. Analytical methods to assess the degree of contamination of surface waters with organically bound fluorine are highly needed and up to now only combustion ion chromatography based method is available, which is relatively laborious. Since a few years’ high resolution-continuum source-graphite furnace atomic absorption spectrometers (HR-CS-GFAAS) are commercially available from Analytik Jena. By means of this technique, the detection of high resolution molecular absorption spectra (MAS) is enabled. Thus, fluoride is detectable upon the addition of a modifier and the formation of a diatomic molecule (e.g. GaF). Just recently, we applied this technique for total fluorine (mainly dissolved fluoride) analysis in river water samples. In the present work a HR-CS-GFMAS method for extractable organically bound fluorine (EOF) analysis in surface water samples was developed by us. The method is based on SPE extraction of organically bound fluorine even in the presence of high fluoride concentrations followed by HR-CS-GFMAS analysis upon elution. Due to high enrichment factors, LODs in the low ng/L range were achieved. We successfully applied our SPE HR-CS-GFMAS method to Rhine water samples and EOF in the range of about 50-300 ng/L was detectable. T2 - DGMS 2020 CY - Münster, Germany DA - 01.03.2020 KW - SPE HR-CS GF MAS PY - 2020 AN - OPUS4-50503 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von der Au, Marcus T1 - HR-CS-GF-MAS as a new screening method for emerging pollutants - per- and polyfluorinated substances in the environment N2 - The introduction of fluorine in organic molecules leads to new chemical/physical properties. Especially in the field of pharmaceuticals, fluorinated organic molecules are becoming more and more popular and at present amount up to 25% of market share, with an upward trend. The main benefits of fluorinated pharmaceuticals are: (i) enhanced fat solubility; (ii) enhanced interaction of catalytic-center of enzymes with fluorine-drugs; (iii) a delayed metabolism within the human body. Highly fluorinated organic substances are also used in technical applications (e.g. coatings, fire-extinguishing agents). Due to the broad variety of fluorinated substances and increasing production volumes numerous and up to date unknown fluorine-species are most likely to be present in the (aquatic) environment. Analytical methods to assess the degree of contamination of surface waters with organically bound fluorine are highly needed and up to now only combustion ion chromatography based method is available, which is relatively laborious. Since a few years’ high resolution-continuum source-graphite furnace atomic absorption spectrometers (HR-CS-GFAAS) are commercially available from Analytik Jena. By means of this technique, the detection of high resolution molecular absorption spectra (MAS) is enabled. Thus, fluoride is detectable upon the addition of a modifier and the formation of a diatomic molecule (e.g. GaF). Just recently, we applied this technique for total fluorine (mainly dissolved fluoride) analysis in river water samples. In the present work a HR-CS-GFMAS method for extractable organically bound fluorine (EOF) analysis in surface water samples was developed by us. The method is based on SPE extraction of organically bound fluorine even in the presence of high fluoride concentrations followed by HR-CS-GFMAS analysis upon elution. Due to high enrichment factors, LODs in the low ng/L range were achieved. We successfully applied our SPE HR-CS-GFMAS method to Rhine water samples and EOF in the range of about 50-300 ng/L was detectable. T2 - DGMS 2020 CY - Münster, Germany DA - 01.03.2020 KW - SPE HR-CS GF MAS PY - 2020 AN - OPUS4-50504 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Gojani, Ardian T1 - Combined Raman and LIBS using Spatial Heterodyne Spectrometer with High Repetition Rate Laser N2 - Spatial heterodyne spectroscopy is used for Raman and laser-induced breakdown spectroscopy of six rocks with various mineral content, using high repetition rate diode-pumped solid state lasers. While LIBS data are obtained for all samples, Raman signal was determined only for half of those. This work shows that it is possible to combine LIBS and Raman spectrometry into a single instrument consisting of a DPSS laser for excitation and SHS for spectral recording. For better results and gain from complementary spectrochemical information that the system provides, it is necessary to optimize light collection. T2 - 2020 Winter Conference on Plasma Spectrochemistry CY - Tucson, Arizona, USA DA - 13.01.2020 KW - Laser-Induced Background Spectroscopy KW - Spatial heterodyne spectroscopy KW - Raman spectroscopy PY - 2020 AN - OPUS4-50265 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Zimmermann, T. A1 - von der Au, Marcus A1 - Reese, A. A1 - Klein, O. A1 - Hildebrandt, L. A1 - Pröfrock, D. T1 - Substituting HF by HBF4 – an optimized digestion method for multi-elemental sediment analysis via ICP-MS/MS N2 - Determination of elemental mass fractions in sediments plays a major role in evaluating the environmental status of aquatic ecosystems. Herewith, the optimization of a new total digestion protocol and the subsequent analysis of 48 elements in different sediment reference materials (NIST SRM 2702, GBW 07313, GBW 07311 and JMC-2) based on ICP-MS/MS detection is presented. The developed method applies microwave acid digestion and utilizes HBF4 as fluoride source for silicate decomposition. Similar to established protocols based on HF, HBF4 ensures the dissolution of the silicate matrix, as well as other refractory oxides. As HBF4 is not acutely toxic; no special precautions have to be made and digests can be directly measured via ICP-MS without specific sample inlet systems, evaporation steps or the addition of e.g. H3BO3, in order to mask excess HF. Different acid mixtures with and without HBF4 were evaluated in terms of digestion efficiency based on the trace metal recovery. The optimized protocol (5 mL HNO3, 2 mL HCL, 1 mL HBF4) allows a complete dissolution of the analyzed reference materials, as well as quantitative recoveries for a wide variety of certified analytes. Low recoveries for e.g. Sr, Ba and rare earth elements due to fluoride precipitation of HF-based digestions protocols, can be avoided by the usage of HBF4 instead. Based on the usage of high purity HBF4 all relevant trace, as well as matrix elements can be analyzed with sufficiently low LOQs (0.002 μg L−1 for U up to 6.7 μg L−1 for Al). In total, 34 elements were within a recovery range of 80%–120% for all three analyzed reference materials GBW 07313, GBW 07311 and JMC-2. 14 elements were outside a recovery range of 80%–120% for at least one of the analyzed reference materials. KW - Reference Materials KW - Sediment KW - HF free Digestion PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-510480 DO - https://doi.org/10.1039/D0AY01049A SP - 1 EP - 10 PB - Royal Society of Chemistry AN - OPUS4-51048 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Maiwald, Michael T1 - On the way to the cyber-physical lab T1 - Auf dem Weg zum cyber-physical Lab N2 - Laboratories tend to be central hubs for chemical, biotechnological, pharmaceutical or foodstuff production. They play a key role in research and development, chemical analysis, quality assurance, maintenance and process control. For process development and optimization, process analytical technology (PAT) has proven to be a powerful tool to improve our understanding of processes, increase productivity, reduce waste and costs and shorten processing times. N2 - Im Allgemeinen sind Laboratorien zentrale Drehscheiben für die chemische, biotechnologische, pharmazeutische oder Lebensmittelproduktion. Sie spielen eine Schlüsselrolle in Forschung und Entwicklung, chemischer Analytik, Qualitätssicherung, Instandhaltung und Prozesskontrolle. Für die Prozessentwicklung und -Optimierung hat sich die Prozessanalysentechnik (PAT) als leistungsfähiges Werkzeug erwiesen, um das Prozessverständnis zu verbessern, die Produktivität zu steigern, Abfall und Kosten zu reduzieren und die Prozesszeit zu verkürzen. KW - Lab of the Future KW - Digitalisation KW - Cyber-physical Lab KW - Smart Laboratoy PY - 2020 SP - 1 EP - 3 PB - Lumitos AG CY - Darmstadt AN - OPUS4-51036 LA - mul AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Climent Terol, Estela T1 - Design and synthesis of lateral flow tests containing antibody-gated materials N2 - Methods for the rapid and sensitive detection of target analytes are gaining importance in medical diagnostics and environmental monitoring, in the security, occupational health and safety as well as food sectors. Among all the methods employed for rapid tests, lateral flow assays (LFAs) are the most commonly used. However, some drawbacks are that most of these tests either indicate the analyte only indirectly, and in other cases the sensitivity is not high enough. Keeping in mind these limitations, the use of stimuli-responsive materials for small-molecule sensing relying on chemical signal amplification and utilizing specific interactions between biomolecules such as antibodies and the corresponding analytes are particularly attractive. Such materials can also be incorporated on paper strips for lateral-flow assays in a straightforward manner. These sensing materials comprise mesoporous silica nanoparticles loaded with indicator molecules and containing certain hapten derivative molecules covalently grafted at the surface, which bind to the respective antibody and inhibit the release of a dye as reporter (Scheme 1). In presence of the designated analyte, a displacement of the antibody from the material is observed, allowing the release of dye. Because a large number of indicator molecules can be released when a single analyte molecule binds to an antibody cap, a strong signal amplification is observed. Considering the modularity, high sensitivity and selectivity of these antibody-gated indicator delivery systems, the presentation discusses general aspects of system design as well as analytical performance and highlights the integration into a lateral-flow assay, showing as an example the determination of the explosives TATP, TNT and PETN with a fluorescence readout, in single-substance and multiplexing modes. T2 - International Webinar on Biosensors and Bioelectronics CY - Online meeting DA - 20.07.2020 KW - Multiplexing detection KW - Rapid testing methods KW - Gated hybrid materials KW - Lateral flow tests KW - Explosives detection PY - 2020 UR - https://europeanmeetings.net/conferences/biosensor-bioelectronics AN - OPUS4-51038 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Silva, C. P. A1 - Carvalho, T. A1 - Schneider, Rudolf A1 - Esteves, V. I. A1 - Lima, D. L. D. T1 - ELISA as an effective tool to determine spatial and seasonal occurrence of emerging contaminants in the aquatic environment N2 - During the last two decades, studies related to the occurrence and fate of emerging contaminants in the aquatic environment have received great attention from the international scientific community. The monitoring of the presence of these compounds is particularly important since they are known to induce adverse effects in aquatic environments, even at extremely low concentrations. This work aimed to apply a simple and effective methodology, such as enzyme-linked immunosorbent assay (ELISA), in the monitoring of 17a-ethinylestradiol (EE2) and 17b-estradiol (E2) (a synthetic and a natural hormone, respectively), carbamazepine (CBZ, an antiepileptic), cetirizine (CET, an antihistamine) and caffeine (CAF, a stimulant) in water matrices with differing salinity and organic matter contents. ELISA was proven to be a valid and practical tool, especially for screening purposes in contrast to traditional chromatographic techniques which are prohibitively expensive for an application on a broader base. The main originality of this work was to establish seasonal and spatial effects on the occurrence of the referred contaminants by using the effectiveness of ELISA to screen those compounds in samples with different characteristics. This work reports both the seasonal and spatial quantification of the referred contaminants in the aquatic environment of the central region of Portugal, with concentrations ranging as follows: 5–87 ng L-1, for E2, 2–17 ng L-1, for EE2, 10–1290 ng L-1, for CBZ, 10–190 ng L-1, for CET, and 62–6400 ng L-1, for CAF. KW - Estrogen KW - Estradiol KW - Ethinylestradiol KW - Carbamazepine KW - Portugal KW - Wasser PY - 2020 DO - https://doi.org/10.1039/D0AY00426J SN - 1759-9679 VL - 12 IS - 19 SP - 2517 EP - 2526 PB - Royal Society of Chemistry RSC CY - London AN - OPUS4-51340 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Raysyan, Anna A1 - Eremin, S. A. A1 - Beloglazova, N. V. A1 - De Saeger, S. A1 - Gravel, I. V. T1 - Immunochemical approaches for detection of aflatoxin B1 in herbal medicines N2 - Introduction Aflatoxin B1 (AFB1) is a toxic low‐molecular‐weight secondary metabolite of Aspergillus flavus and A. parasiticus. AFB1 was classified as a Group I carcinogen by the World Health Organisation for Research on Cancer in 1993. AFB1 is an unavoidable natural contaminant of some herbal medicine, able to cause serious health issues for humans consuming the related medicine. Objective Therefore, this study aimed to develop an efficient fluorescence polarisation immunoassay (FPIA) and a rapid, low‐cost, and easy‐to‐use membrane‐based flow‐through immunoassay (MBA) for determination of AFB1 in herbal medicine Origanum vulgare L., Rubus idaeus L., Urtica dioica L. and Sorbus aucuparia L. Results A cut‐off level of the developed MBA was 0.8 ppb. Validation of the developed test was performed with blank and spiked samples. Using three naturally contaminated or three artificially spiked samples. The FPIA showed a linear working range of 8.6 to 64 ppb, and a half maximal inhibitory concentration (IC50) of 24 ppb. Conclusion The results were in good correlation with the enzymelinked immunosorbent assay (ELISA) results (the IC50 0.1 ppb). Both the sample preparation and analysis are simple, cost‐effective and easy to perform on‐site in non‐laboratory environments. Liquid chromatography with tandem mass spectrometry (LC‐MS/MS) was used as a confirmatory technique. KW - Fluorescence Polarisation Immunoassay KW - membrane-based flow-through immunoassay KW - Mycotoxins KW - Mykotoxine KW - Schnelltest PY - 2020 DO - https://doi.org/10.1002/pca.2931 SN - 1099-1565 VL - 31 IS - 5 SP - 662 EP - 669 PB - Wiley Online Libary CY - Hoboken, NJ, USA AN - OPUS4-51341 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Raysyan, Anna A1 - Galvidis, I. A. A1 - Schneider, Rudolf A1 - Eremin, S. A. A1 - Burkin, M. A. T1 - Development of a latex particles-based lateral flow immunoassay for group determination of macrolide antibiotics in breast milk N2 - A lateral flow immunoassay (LFIA) using latex particles labeled with antibody to BSAclarithromycin (CLA) was developed for the rapid simultaneous group determination of six macrolide antibiotics. Optimization of antigen spotting on the membrane and latex probe loading allowed improving visual detectability (vLOD) 100 times, which was 1, 1, 10, 10, 50, and 1000 ng/mL for CLA, roxithromycin, erythromycin, dirithromycin, azithromycin, and oleandomycin in buffer, respectively. The calculated limits of instrumental detection (cLOD) were respectively 0.12, 0.15, 1.4, 2.1, 2.4, and 3.3 ng/mL. To avoid a strong influence of breast milk of a very diverse and variable composition, a sample pretreatment is proposed. The six macrolides mentioned can be visually detected in breast milk after 20 minutes pretreatment at concentrations of 10-1000 ng / ml or instrumentally with cLOD of 4.0, 2.5, 30, 42, 42 and 180 ng/mL. The recovery rate from the spiked samples carried out using a strip scanner device ranged from 71% to 110%, and precision expressed as relative standard deviation was between 3-14%. The described rapid on-site diagnostic assay format can be useful for monitoring the content of antibiotics in breast milk during macrolide treatment to ensure safe breastfeeding of infants. KW - Clarithromycin KW - Roxithromycin KW - Erythromycin KW - Dirithromycin KW - Azithromycin KW - Oleandomycin KW - Macrolide antibiotics KW - Makrolide KW - Antibiotika PY - 2020 DO - https://doi.org/10.1016/j.jpba.2020.113450 SN - 0731-7085 VL - 189 IS - 113450 SP - 1 EP - 9 PB - Elsevier CY - Amsterdam, NL AN - OPUS4-51343 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ecke, Alexander T1 - Magnetpartikelbasierte Immunoassays für die Online-Trinkwasseranalyik N2 - Wir berichten hier über die Entwicklung plattenbasierter Magnetpartikel-Immunoassays für die Bestimmung des Breitbandantibiotikums Amoxicillin und des nicht-steroidalen Antirheumatikums Diclofenac im Trinkwasser. Für die beiden Analyten wurden unterschiedliche Ansätze entwickelt, mit partikelgebundenen Antikörpern für AMX und Analyt-gekoppelten Partikeln für DCF, die die simultane Detektion beider Kontaminanten im nanomolaren Konzentrationsbereich erlauben. T2 - Interdisziplinäres Doktorandenseminar des AK Prozessanalytik CY - Online meeting DA - 21.09.2020 KW - Magnetpartikel KW - Immunoassay KW - Trinkwasseranalytik KW - Diclofenac KW - Amoxicillin PY - 2020 AN - OPUS4-51345 LA - mul AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ecke, Alexander A1 - Schneider, Rudolf T1 - Immunanalytische Bestimmung von β-Lactam-Antibiotika und deren Hydrolyseprodukten im Trinkwasser N2 - Die speziellen Herausforderungen der immunanalytischen Bestimmung von β-Lactam-Antibiotika in (Trink-)Wasserproben werden anhand eines neu entwickelten ELISA (enzyme-linked immunosorbent assay) für das Breitbandantibiotikum Amoxicillin diskutiert. Es konnte gezeigt werden, dass die Hydrolyse des Analyten einen positiven Einfluss auf die Erkennung durch den Antikörper hat. Das entsprechende Hydrolyseprodukt, zu dem der Antikörper folglich eine höhere Affinität aufweist, wurde identifiziert und ein entsprechendes Verhalten wurde auch für andere β-Lactam-Antibiotika aus der Gruppe der Penicilline beobachtet. Eine Strategie für die Analyse von Wasserproben hinsichtlich Kontamination mit den hier betrachteten β-Lactamen und deren Hydrolyseprodukten wurde ausgearbeitet. T2 - Wasser 2020 - Jahrestagung der Wasserchemischen Gesellschaft CY - Meeting was canceled DA - 18.05.2020 KW - ELISA KW - Immunoassay KW - Antibiotika KW - Hydrolyseprodukte KW - Amoxicillin PY - 2020 UR - https://www.wasserchemische-gesellschaft.de/images/pdf_veranstaltungen/2020/Wasserchemische_Gesellschaft_Tagungsband_Wasser_2020_s.pdf SP - 466 EP - 470 AN - OPUS4-51061 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Weber, A. A1 - von Randow, M. A1 - Voigt, A.-L. A1 - von der Au, Marcus A1 - Fischer, E. A1 - Meermann, Björn A1 - Wagner, M. T1 - Ingestion and toxicity of microplastics in the freshwater gastropodLymnaea stagnalis: No microplastic-induced effects alone or incombination with copper N2 - The interaction of microplastics with freshwater biota and their interaction with other stressors is still not very well understood. Therefore, we investigated the ingestion, excretion and toxicity of microplastics in the freshwater gastropod Lymnaea stagnalis. MP ingestion was analyzed as tissues levels in L. stagnalis after 6–96 h of exposure to 5–90 μm spherical polystyrene (PS) microplastics. To understand the excretion, tissue levels were determined after 24 h of exposure followed by a 12 h–7 d depuration period. To assess the toxicity, snails were exposed for 28 d to irregular PS microplastics (<63 μm, 6.4–100,000 particles mL−1), both alone and in combination with copper as additional stressor. To compare the toxicity of natural and synthetic particles, we also included diatomite particles. Microplastics ingestion and excretion significantly depended on the particle size and the exposure/depuration duration. An exposure to irregular PS had no effect on survival, reproduction, energy reserves and oxidative stress. However, we observed slight effects on immune cell phagocytosis. Exposure to microplastics did not exacerbate the reproductive toxicity of copper. In addition, there was no pronounced difference between the effects of microplastics and diatomite. The tolerance towards microplastics may originate from an adaptation of L. stagnalis to particle-rich environments or a general stress resilience. In conclusion, despite high uptake rates, PS fragments do not appear to be a relevant stressor for stress tolerant freshwater gastropods considering current environmental levels of microplastics. KW - Lymnaea stagnalis KW - Microplastic-induced effects KW - Mixture toxicity PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-513551 DO - https://doi.org/10.1016/j.chemosphere.2020.128040 VL - 263 SP - 128040 PB - Elsevier Ltd. AN - OPUS4-51355 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Jenning, M. A1 - Marklein, B. A1 - Ytterberg, J. A1 - Zubarev, R. A. A1 - Joshua, V. A1 - van Schaardenburg, D. A1 - van de Stadt, L. A1 - Catrina, A. I. A1 - Nonhoff, U. A1 - Häupl, T. A1 - Konthur, Zoltán A1 - Burmester, G. R. A1 - Skriner, K. T1 - Bacterial citrullinated epitopes generated by Porphyromonas gingivalis infection—a missing link for ACPA production N2 - Objectives: Porphyromonas gingivalis (P.g.) is discussed to be involved in triggering self-reactive immune responses. The aim of this study was to investigate the autocitrullinated prokaryotic peptidylarginine deiminase (PPAD) from P.g. CH2007 (RACH2007-PPAD) from a rheumatoid arthritis (RA) patient and a synthetic citrullinated PPAD peptide (CPP) containing the main autocitrullination site as potential targets for antibody reactivity in RA and to analyse the possibility of citrullinating native human proteins by PPAD in the context of RA. Methods: Recombinant RACH2007-PPAD was cloned and expressed in Escherichia coli. Purified RACH2007-PPAD and its enzymatic activity was analysed using two-dimensional electrophoresis, mass spectrometry, immunoblot and ELISA. Autoantibody response to different modified proteins and peptides was recorded and bioinformatically evaluated. Results: RACH2007-PPAD was capable to citrullinate major RA autoantigens, such as fibrinogen, vimentin, hnRNP-A2/B1, histone H1 and multiple peptides, which identify a common RG/RGG consensus motif. 33% of RA patients (n=30) revealed increased reactivity for α-cit-RACH2007-PPAD before RA onset. 77% of RA patients (n=99) presented α-cit-specific signals to CPP amino acids 57-71 which were positively correlated to α-CCP2 antibody levels. Interestingly, 48% of the α-CPP-positives were rheumatoidfactor IgM/anti-citrullinated peptide/protein antibodies (ACPA)-negative. Anti-CPP and α-RACH2007-PPAD antibody levels increase with age. Protein macroarrays that were citrullinated by RACH2007-PPAD and screened with RA patient sera (n=6) and controls (n=4) uncovered 16 RACH2007-PPAD citrullinated RA autoantigens and 9 autoantigens associated with lung diseases. We showed that the α-CPP response could be an important determinant in parenchymal changes in the lung at the time of RA diagnosis (n=106; p=0.018). Conclusions: RACH2007-PPAD induced internal citrullination of major RA autoantigens. Anti-RACH2007-PPAD correlates with ACPA levels and interstitial lung disease autoantigen reactivity, supporting an infection-based concept for induction of ACPAs via enzymatic mimicry. KW - Zitrullin KW - Rheumatoide Arthritis KW - Proteinmodifikation KW - Autoantikörper PY - 2020 DO - https://doi.org/10.1136/annrheumdis-2019-216919 VL - 79 IS - 9 SP - 1194 EP - 1202 PB - BMJ Publishing Group Ltd AN - OPUS4-51372 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Riedel, Soraya T1 - An ELISA-inspired electrochemical detection system for the quantification of ochratoxin A N2 - Toxic secondary metabolites produced by fungi (aka mycotoxins) are often identified as contaminants in cereals or other food and pose a severe health risk, since they can be carcinogenic, neurotoxic, mutagenic or harmful in other respect. The highly nephrotoxic ochratoxin A (OTA) is one of the most important and widespread mycotoxins found in grains such as rye, wheat or oats. Thus, the fast and cost-effective detection of OTA is of great importance, to ensure that the toxin concentration is below the legal limits. Currently, the mycotoxin detection in mill operations is typically performed by means of the antibody-based ELISA method, which requires extensive laboratory equipment and specialized personnel. It was the aim of this study to develop an ELISA-inspired sensor method enabling the on-site quantification of OTA. Based on a classical competitive ELISA, an electrochemical detection system was developed. Anti-OTA-antibodies were captured on magnetic beads on which the competitive affinity reaction between OTA and an OTA- horseradish peroxidase (HRP) tracer was performed. Here the OTA concentration is inversely proportional to the amount of the bound OTA-HRP tracer, which was electrochemically quantified. For this purpose, oxidized tetramethylbenzidine (TMB), which is enzymatically produced by the reduction of H2O2 by the OTA-HRP tracer, is quantified by amperometry with screen-printed electrodes in a flow system. With this method a detection limit of 250 ng L-1 could be achieved which meets the current requirements for the detection of OTA in food and feed. T2 - Interdisziplinäres Doktorandenseminar des AK Prozessanalytik CY - Online meeting DA - 21.09.2020 KW - Ochratoxin A KW - Cyclic voltammetry KW - Amperometry KW - ELISA KW - Magnetic beads PY - 2020 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. AN - OPUS4-51375 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Gojani, Ardian A1 - Gornushkin, Igor B. T1 - Räumlich aufgelöste Heterodyne-Spektroskopie N2 - Die räumlich aufgelöste Heterodyn-Spektroskopie (SHS) kombiniert dispersive und interferometrische Techniken zur Gewinnung spektroskopischer Informationen und kann in einem anpassbaren Instrument realisiert werden. SHS wird für die chemische Analyse von verschiedenen Materialien mit Hilfe der Laser-induzierten Plasmaspektroskopie (LIBS) und der Raman-Spektroskopie eingesetzt. KW - Raman spectroscopy KW - Spatial heterodyne spectroscopy KW - Laser-induced breakdown spectroscopy PY - 2020 UR - https://analyticalscience.wiley.com/do/10.1002/was.00080118/full VL - 5 SP - 2 EP - 4 PB - Wiley CY - Weinheim AN - OPUS4-51397 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bröcker, S. A1 - Klaer, R.-H. A1 - Kockmann, N. A1 - Maiwald, Michael T1 - Process Industry United in the Digital World - The current landscape of research for the digitalization of the process industry N2 - The Tutzing Symposium "100 % digital: survival strategies for the process industry" (see 4.1) in April 2018 was characterized by a great momentum which has been taken up and continued until today. The aim was to implement the ideas from the Tutzing Symposium in a coordinated and targeted manner. For this purpose, development needs as well as the numerous currently planned or already started research and development activities in the context of digitalisation were first compiled and analysed. This resulted in the current research landscape for digitalization in the process industry. It now enables to identify open topics and to translate them into research funding programs as well as to define new projects in the dialogue between users, suppliers and research, which are to be meaningfully interlinked and consolidated with existing projects. Due to the strong interest in digitalisation, activities are constantly being added, so that this paper can only provide a snapshot of the situation in the period 2019-2020. KW - Process Industry KW - Digitalisation KW - Tutzing Symposion KW - research landscape KW - Process Control PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-512265 UR - https://processnet.org/en/TAKdigital.html SP - 1 EP - 15 AN - OPUS4-51226 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bhuckory, S. A1 - Wegner, Karl David A1 - Qiu, X. A1 - Wu, Y.T. A1 - Jennings, T. L. A1 - Incamps, A. A1 - Hildebrandt, N. T1 - Triplexed CEA-NSE-PSA Immunoassay Using Time-Gated Terbium-to-Quantum Dot FRET N2 - Time-gated Förster resonance energy transfer (TG-FRET) between Tb complexes and luminescent semiconductor quantum dots (QDs) provides highly advantageous photophysical properties for multiplexed biosensing. Multiplexed Tb-to-QD FRET immunoassays possess a large potential for in vitro diagnostics, but their performance is often insufficient for their application under clinical conditions. Here, we developed a homogeneous TG-FRET immunoassay for the quantification of carcinoembryonic antigen (CEA), neuron-specific enolase (NSE), and prostatespecific antigen (PSA) from a single serum sample by multiplexed Tb-to-QD FRET. Tb–IgG antibody donor conjugates were combined with compact QD-F(ab’)2 antibody acceptor conjugates with three different QDs emitting at 605, 650, and 705 nm. Upon antibody–antigen–antibody Sandwich complex formation, the QD acceptors were sensitized via FRET from Tb, and the FRET ratios of QD and Tb TG luminescence intensities increased specifically with increasing antigen concentrations. Although limits of detection (LoDs: 3.6 ng/mL CEA, 3.5 ng/mL NSE, and 0.3 ng/mL PSA) for the triplexed assay were slightly higher compared to the single-antigen assays, they were still in a clinically relevant concentration range and could be quantified in 50 μL serum samples on a B·R·A·H·M·S KRYPTOR Compact PLUS clinical immunoassay plate reader. The simultaneous quantification of CEA, NSE, and PSA at different concentrations from the same serum sample demonstrated actual multiplexing Tb-to-QD FRET immunoassays and the potential of this technology for translation into clinical diagnostics. KW - Lanthanides KW - Nanoparticles KW - Biosensing KW - Multiplexing KW - FRET KW - Fluorescence KW - PSA KW - NSE KW - CEA PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-512290 DO - https://doi.org/10.3390/molecules25163679 VL - 25 IS - 16 SP - 3679 PB - MDPI AN - OPUS4-51229 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hildebrandt, L. A1 - von der Au, Marcus A1 - Zimmermann, T. A1 - Reese, A. A1 - Ludwig, J. A1 - Pröfrock, D. T1 - A metrologically traceable protocol for the quantification of trace metals in different types of microplastic N2 - The presence of microplastic (MP) particles in aquatic environments raised concern About possible enrichment of organic and inorganic pollutants due to their specific surface and chemical properties. In particular the role of metals within this context is still poorly understood. Therefore, the aim of this work was to develop a fully validated acid digestion protocol for metal analysis in different polymers, which is a prerequisite to study such interactions. The proposed digestion protocol was validated using six different certified reference materials in the microplastic size range consisting of polyethylene, polypropylene, acrylonitrile butadiene styrene and polyvinyl chloride. As ICP-MS/MS enabled time-efficient, sensitive and robust analysis of 56 metals in one measurement, the method was suitable to provide mass fractions for a multitude of other elements beside the certified ones (As, Cd, Cr, Hg, Pb, Sb, Sn and Zn). Three different microwaves, different acid mixtures as well as different temperatures in combination with different hold times were tested for optimization purposes. With the exception of Cr in acrylonitrile butadiene styrene, recovery rates obtained using the optimized protocol for all six certified reference materials fell within a range from 95.9% ± 2.7% to 112% ± 7%. Subsequent optimization further enhanced both precision and recoveries ranging from 103% ± 5% to 107 ± 4% (U; k = 2 (n = 3)) for all certified metals (incl. Cr) in acrylonitrile butadiene styrene. The results clearly show the analytical challenges that come along with metal analysis in chemically resistant plastics. Addressing specific analysis Tools for different sorption scenarios and processes as well as the underlying kinetics was beyond this study’s scope. However, the future application of the two recommended thoroughly validated total acid digestion protocols as a first step in the direction of harmonization of metal analysis in/on MP will enhance the significance and comparability of the generated data. It will contribute to a better understanding of the role of MP as vector for trace metals in the environment. KW - Microplatic KW - Digestion KW - Analytical Chemistry PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-510727 DO - https://doi.org/10.1371/journal.pone.0236120 VL - 15 IS - 7 SP - e0236120 AN - OPUS4-51072 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Riedel, Soraya T1 - Immunomagnetic ochratoxin A assay with smartphone-based amperometric detection N2 - Toxic secondary metabolites produced by fungi (aka mycotoxins) are often identified as contaminants in food and pose a severe health risk. Ochratoxin A (OTA) is one of the most abundant mycotoxins found in grains such as wheat, rye or oats. Thus, the fast and cost-effective detection of OTA is of great importance, to ensure that the toxin concentration is below the legal limits. Currently, mycotoxin detection in mill operations is typically performed by means of the antibody-based ELISA (Enzyme-linked Immunosorbent Assay) method, which requires extensive laboratory equipment and specialized personnel. The aim of this project is to establish a more user-friendly detection method. Thus, we developed a magnetic bead-based assay enabling the quantification of OTA with a miniaturized potentiostat connected to a smartphone. T2 - SALSA Make and Measure 2020 CY - Online meeting DA - 15.10.2020 KW - ochratoxin A KW - smartphone KW - amperometry KW - cyclic voltammetry KW - magnetic beads KW - screen-printed electrodes PY - 2020 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. AN - OPUS4-51445 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Steinhäuser, Lorin A1 - Piechotta, Christian T1 - A project to deliver reliable measurements of estrogens for better monitoring surveys and risk assessments N2 - EDC WFD project to deliver reliable measurements of estrogens for better monitoring surveys and risk assessments. Collaboration between National Metrology Institutes and advanced research institutes from 6 European countries A Balance of expertise: development and certification of RM, proficiency tests / interlaboratory comparison design, method development and validation, standardisation A 3 years project: September 2019- August 2022 Strong engagement with stakeholders (Advisory Group) T2 - SETAC 2020 CY - Online meeting DA - 03.05.2020 KW - EDC KW - Endocrine disrupting chemicals PY - 2020 AN - OPUS4-51739 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von Törne, Wipert T1 - Structural Investigations of Transformation Products of the Anticoagulant Drug and Rodenticide Warfarin N2 - Herein, we present liquid and gas chromatographic methods coupled to (high-resolution) mass spectrometry for the analysis of warfarin’s transformation products. Methodologies such as UV-irradiation, ozonation, and chlorination were utilised to simulate technical water treatment. Resulting compounds were elucidated and examined by numerous analytical methods. Moreover, the oxidative phase I metabolism was mimicked by an electrochemical flow cell to synthesise and confirm major metabolites. The further aim is the toxicological assessment of all substances formed, as well as, the quantification of warfarin and its TPs under environmentally relevant conditions employing the introduced methods. T2 - SETAC 2020 CY - Online meeting DA - 03.05.2020 KW - Warfarin PY - 2020 AN - OPUS4-51738 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Raysyan, Anna A1 - Moerer, R. A1 - Coesfeld, Bianca A1 - Eremin, S. A1 - Schneider, Rudolf T1 - Fluorescence polarization immunoassay for the determination of diclofenac in wastewater N2 - Pharmacologically active compounds are often detected in wastewater and surface waters. The nonsteroidal anti-inflammatory drug diclofenac (DCF) was included in the European watch list of substances that requires its environmental monitoring in the member states. DCF may harmfully influence the ecosystem already at concentrations ≤ 1 μg L−1. The fast and easy quantification of DCF is becoming a subject of global importance. Fluorescence polarization immunoassay (FPIA) is a homogeneous mix-and-read method which does not require the immobilization of reagents. FPIA can be performed in one phase within 20–30 min, making it possible to analyse wastewater without any complicated pre-treatment. In this study, new tracermolecules with different structures, linking fluorophores to derivatives of the analyte, were synthesized, three homologous tracers based on DCF, two including a C6 spacer, and one heterologous tracer derived from 5-hydroxy-DCF. The tracer molecules were thoroughly assessed for performance. Regarding sensitivity of the FPIA, the lowest limit of detection reached was 2.0 μg L−1 with a working range up to 870 μg L−1. The method was validated for real wastewater samples against LC-MS/MS as reference method with good agreement of both methods. KW - Abwasser KW - Umweltschadstoffe KW - Antikörper KW - Antibody PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-518398 DO - https://doi.org/10.1007/s00216-020-03058-w SN - 1618-2642 VL - 413 IS - 4 SP - 999 EP - 107 PB - Springer CY - Heidelberg AN - OPUS4-51839 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Almeida, Â. A1 - Calisto, V. A1 - Esteves, V. A1 - Schneider, Rudolf A1 - Figueira, E. A1 - Soares, A. A1 - Freitas, R. T1 - Can ocean warming alter sub-lethal effects of antiepileptic and antihistaminic pharmaceuticals in marine bivalves? N2 - The negative effects induced in marine organisms by Climate Change related abiotic factors consequences, namely ocean warming, are well-known. However, few works studied the combined impacts of ocean warming and contaminants, as pharmaceutical drugs. Carbamazepine (CBZ) and cetirizine (CTZ) occur in the marine environment, showing negative effects in marine organisms. This study aimed to evaluate the impacts of Ocean warming on the effects of CBZ and CTZ, when acting individually and combined (drug vs drug), in the edible clam Ruditapes philippinarum. For that, drugs concentration, bioconcentration factors and biochemical parameters, related with clam’s metabolic capacity and oxidative stress, were evaluated after 28 days exposure to environmentally relevant scenarios of these stressors. The results showed limited impacts of the drugs (single and combined) at control and warming condition. Indeed, it appeared that warming improved the oxidative status of contaminated clams (higher reduced to oxidized glutathione ratio, lower lipid peroxidation and Protein carbonylation levels), especially when both drugs were combined. This may result from clam’s defence mechanisms activation and reduced metabolic capacity that, respectively, increased elimination and limited production of reactive oxygen species. At low stress levels, defence mechanisms were not activated which resulted into oxidative stress. The present findings highlighted that under higher stress levels clams may be able to activate defence strategies that were sufficient to avoid cellular damages and loss of redox homeostasis. Nevertheless, low concentrations were tested in the present study and the observed responses may greatly Change under increased pollution levels or temperatures. Further research on this topic is needed since marine heat waves are increasing in frequency and intensity and pollution levels of some pharmaceuticals are also increasing in coastal systems. KW - Klimaerwärmung KW - Meer KW - Antiepileptika KW - Antihistaminika KW - Muscheln KW - Immunoassay KW - ELISA KW - Carbamazepine KW - Cetirizine PY - 2021 DO - https://doi.org/10.1016/j.aquatox.2020.105673 SN - 0166-445X VL - 230 SP - 105673 PB - Elsevier B.V. AN - OPUS4-51840 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Schneider, Rudolf T1 - Immunoanalytical platforms for on-site environmental health and food safety testing N2 - An overview of methods developed in the division. T2 - FoodSmartphone.eu - Online Symposium CY - Online meeting DA - 25.11.2020 KW - Umweltanalytik KW - Lebensmittelanalytik KW - Immunoassay KW - Biosensor PY - 2020 AN - OPUS4-51841 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Pignatelli, Giuseppe T1 - Optical Sensor for Monitoring Quality of 3D Metal Printing N2 - Additive Manufacturing (AM) becomes widespread in many technological fields including the precise machining of steel. To assure quality of final products, thorough monitoring of online process is required. We test several monitoring techniques during the AM printing to quickly detect and possibly correct flaws while building a workpiece. Here we show how optical emission spectroscopy can be used to recognize defects that are artificially introduced on a steel printing substrate. T2 - Scix 2020 CY - Online meeting DA - 12.10.2020 KW - Additive manufacturing KW - Spectroscopy KW - Analytical chemistry PY - 2020 AN - OPUS4-51858 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Bell, A. M. A1 - von der Au, Marcus A1 - Regnery, J. A1 - Schmid, M. A1 - Meermann, Björn A1 - Reifferscheid, G. A1 - Ternes, T. A1 - Buchinger, S. T1 - Does galvanic cathodic protection by aluminum anodes impact marine organisms? N2 - Background: Cathodic protection by sacrifcial anodes composed of aluminum-zinc-indium alloys is often applied to protect ofshore support structures of wind turbines from corrosion. Given the considerable growth of renewable energies and thus ofshore wind farms in Germany over the last decade, increasing levels of aluminum, Indium and zinc are released to the marine environment. Although these metals are ecotoxicologically well-studied, data regarding their impact on marine organisms, especially sediment-dwelling species, as well as possible ecotoxicological efects of galvanic anodes are scarce. To investigate possible ecotoxicological efects to the marine environment, the diatom Phaedactylum tricornutum, the bacterium Aliivibrio fscheri and the amphipod Corophium volutator were exposed to dissolved galvanic anodes and solutions of aluminum and zinc, respectively, in standardized laboratory tests using natural seawater. In addition to acute toxicological efects, the uptake of these elements by C. volutator was investigated. Results: The investigated anode material caused no acute toxicity to the tested bacteria and only weak but signifcant efects on algal growth. In case of the amphipods, the single elements Al and Zn showed signifcant efects only at the highest tested concentrations. Moreover, an accumulation of Al and In was observed in the crustacea species. Conclusions: Overall, the fndings of this study indicated no direct environmental impact on the tested marine organisms by the use of galvanic anodes for cathodic protection. However, the accumulation of metals in, e.g., crustaceans might enhance their trophic transfer within the marine food web. KW - Galvanic anodes KW - Metal toxicity KW - Metal uptake KW - Corophium volutator KW - Seawater PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-520769 DO - https://doi.org/10.1186/s12302-020-00441-3 VL - 32 IS - 1 SP - Article number 157 AN - OPUS4-52076 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - von der Au, Marcus A1 - Karbach, H. A1 - Bell, A. M. A1 - Bauer, O. B. A1 - Karst, U. A1 - Meermann, Björn T1 - Determination of metal uptake in single organisms, Corophiumvolutator, via complementary electrothermal vaporization/inductively coupled plasma mass spectrometry and laserablation/inductively coupled plasma mass spectrometry N2 - Rationale: (Eco-)toxicological effects are mostly derived empirically and are notcorrelated with metal uptake. Furthermore, if the metal content is determined,mostly bulk analysis of the whole organism population is conducted; thus, biologicalvariability is completely disregarded, and this may lead to misleading results. Toovercome this issue, we compared two different solid sampling techniques for theanalysis of single organisms.Methods: In this study, complementary electrothermal vaporization/inductivelycoupled plasma mass spectrometry (ETV/ICP-MS) , laser ablation/inductivelycoupled plasma mass spectrometry (LA/ICP-MS)-based methods for the analysisof individual organisms were developed and the results obtained were comparedwith the concentrations obtained after digestion and measured using ICP-MS.For this purpose, a common (eco-)toxicological test organism, the mud shrimpCorophium volutator, was selected. As proof-of-concept application, these organismswere incubated with environmentally relevant metals from galvanic anodes, whichare often used for protection against metal corrosion in, for example, offshorewind farms.Results: The bulk analysis revealed that large quantities of the incubated elementswere detectable. Using the ETV/ICP-MS method, we could identify a highbiovariability within the population of organisms tested. Using the LA/ICP-MSmethod, it could be determined that the large quantities of the elements detectedwere due to adsorption of the metals and not due to uptake, which correlates wellwith the absence of (eco-)toxicological effects.Conclusions: The results obtained imply the efficiency of complementary methods toexplain the absence or presence of (eco-)toxicological effects. In particular, methodsthat allow for single-organism analysis or provide even a spatial resolution supportthe interpretation of ecotoxicological findings. KW - ICP-MS KW - Metal toxicity KW - Metal uptake KW - Corophium volutator KW - Seawater PY - 2021 DO - https://doi.org/10.1002/rcm.8953 VL - 35 IS - 2 SP - e8953 AN - OPUS4-52077 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Gehrenkemper, Lennart T1 - Detection of PFAS pollution in environmental samples - A Fast & sensitive PFAS sum parameter method using HR CS GFMAS N2 - The substance class of per- and polyfluorinated alkyl substances (PFAS) comprises more than 5300 organic compounds. PFAS are completely fluorinated on at least one carbon atom. They are associated with negative impacts on human and animal health, are extremely persistent in the environment, and bioaccumulate along food chains. Therefore, PFAS are classified as emerging pollutants. At the same time, their physicochemical properties make them attractive for use in diverse technical applications. They are both hydrophobic and lipophobic and show high thermal as well as chemical resistance due to the strong C-F bond. First regulations of some PFAS in combination with the technically excellent properties generated an innovation pressure and led to an enormous increase in the number of fluorinated substitution compounds. Due to the increasing complexity of this substance class, target analysis is not able to cover such a variety and multitude of analytes. Therefore, a suitable PFAS sum parameter method is necessary for an accurate detection of PFAS pollution in the environment, the identification of PFAS hotspots and an evaluation of appropriate remediation measures. Here we provide insights into the current state of PFAS sum parameter development and present our latest results on method development for the quantitative analysis of PFAS as extractable organically bound fluorine (EOF) in environmental samples using high-resolution molecular absorption spectrometry (HR-CS-GFMAS). For this purpose, we optimized the extraction of PFAS from different solid matrices with simultaneous separation of inorganic fluoride. For quantification resulting extracts were measured using a fluorine specific HR-CS-GFMAS method. By adding gallium salt solutions as modifiers in HR-CS-GFMAS, fluorine can be indirectly quantified very selectively by the in situ formation of GaF with low limits of quantification (instrumental LOQ c(F) < 3 µg/L). Here we will show results from real soil samples from sites with and without known contamination. T2 - 6. Doktorandenseminar des Deutschen Arbeitskreises für Analytische Spektroskopie (DAAS) CY - Online meeting DA - 20.09.2021 KW - High resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) KW - Soil KW - Per- and polyfluorinated alkyl substances (PFASs) KW - Extractable organically bound fluorine (EOF) KW - Solid-liquid extraction PY - 2021 AN - OPUS4-53333 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Vogl, Jochen T1 - Metrology for the recycling of Technology Critical Elements to support Europe´s circular economy agenda N2 - Technology critical elements (TCE) are key materials for high-tech products such as smartphones, notebooks and monitors. Their demand is expected to increase exponentially due to the shift towards greener economy through the deployment of renewable energy and electro mobility solutions (European Green Deal). This and the lack of TCE production within the EU, puts the market under pressure and leads to increasing prices. The need to secure TCE supply has become even more pressing under the current health crisis and it is a major objective of the Covid-19 Recovery Plan aimed at reinforcing Europe’s resilience and autonomy. A sustainable solution is effective recycling and finally a circular economy. However, the analysis of waste streams is difficult and requires suitable SI traceable analytical solutions to allow for comparability of measurement results throughout the recycling process. The required reference materials (RMs) certified for TCE in the corresponding waste or secondary raw materials and specific documentary standards for TCE to comply with ISO/IEC 17025 requirements, however, are lacking. Furthermore, waste from the so-called “urban mine” is extremely heterogeneous making the estimation of its TCE content difficult. Currently, there is a lack of knowledge at the European level about the TCE stocks and flows in the urban mine. Given the high volume of waste generated and received, fast reliable analytical methods as well as sampling and sample preparation strategies are needed to determine the economic value of the waste and of the final product and to develop and improve recycling procedures. Within this project we focus on TCEs (Co, Ga, Ge, In, Ta, Nd, Pr, Dy, Gd, La, Au, Pt, Pd, Rh), which have been selected from the list of critical elements for Europe established by the EU in 2020. The overall objective of the project is to provide the reliable and SI traceable determination of TCE in urban waste material at µg/g levels to increase the efficiency and accuracy of TCE recycling. This will be realized by developing validated SI-traceable reference methods, developing traceable and validated RMs for the TCEs, validating the use of the routine methods and RMs for real world applications and facilitating the take up of the technology and measurement infrastructure developed in the project by the measurement supply chain, standards developing organizations and end users. This presentation will give an overview of the consortium, the project structure and the planned research. First activities will be presented as well. T2 - International Conference on Raw Materials and Circular Economy CY - Athens, Greece DA - 05.09.2021 KW - TCE KW - Reference material KW - Metrology PY - 2021 AN - OPUS4-53282 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ecke, Alexander T1 - An Electrochemical Immunosensor for Pharmaceutical Contaminants in Water N2 - Contamination of drinking water with pharmaceuticals is an emerging problem and a potential health threat. Immunochemical methods based on the binding of the analyte to specific antibodies enable fast & cost-effective on-site analyses. Magnetic bead-based immunoassays allow for implementation into an immunosensor for online testing. Particles are prepared by either coupling the analyte molecule (diclofenac) or the respective antibody (amoxicillin) to the surface. For miniaturization, detection is performed electrochemically (chronoamperometry) on a microfluidic chip. The developed immunosensor will enable detection of pharmaceuticals directly in water supply pipes. T2 - SALSA Make & Measure CY - Online meeting DA - 16.09.2021 KW - Immunosensor KW - Magnetic Beads KW - Amoxicillin KW - Diclofenac KW - Microfluidics KW - Chronoamperometry PY - 2021 AN - OPUS4-53307 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von Törne, Jannik Wipert A1 - Piechotta, Christian A1 - Karnstedt, L. T1 - Einblicke in das Transformationsverhalten von Warfarin N2 - Warfarin ist seit den 1950er Jahren einer der bekanntesten Gerinnungshemmer. Beim Menschen wird es bis heute zur Vorbeugung von Thrombose und der Behandlung von Vorhofflimmern und Arrhythmie eingesetzt. Doch schon einige Jahre vor der FDA Zulassung (U. S. Food and Drug Administration) als Medikament wurde es zur Bekämpfung von Nagetieren eingesetzt. Bis heute wurde es nur teilweise durch weitaus wirksamere antikoagulante Rodentizide (ARs) der zweiten Generation, sogenannte Superwarfarine, ersetzt. In der Fachliteratur finden sich zahlreiche Beispiele für sekundäre und tertiäre Vergiftung mit ARs bei Wildtieren und anderen Nichtzieltieren. Bisher ist relativ wenig über die Bioakkumulation, Persistenz und Toxizität von ARs und deren Transformationsprodukten (TPs) in der terrestrischen und aquatischen Umwelt, sowie der Nahrungskette bekannt. Um Einblicke in mögliche Transformations- und Abbauprozesse von Warfarin unter umweltrelevanten Bedingungen zu erhalten und um diese Einflüsse zu simulieren, wurde hierzu eine elektrochemische Reaktorzelle in Kopplung mit einem Massenspektrometer verwendet. So können mögliche Phase-I-Metabolite und sogenannte „oxidation products“ vorausgesagt werden. Hinsichtlich des Verhaltens während der Wasseraufbereitung in Kläranlagen wurden technisch relevante Prozesse wie Chlorung, UV-Bestrahlung und Ozonung in Modellen simuliert, um hier potenzielle Abbauwege und -prozesse aufzuklären. Gebildete Transformationsprodukte wurden daher mittels unterschiedlicher chromatografischer Verfahren getrennt und massenspektrometrisch identifiziert, charakterisiert und verglichen. T2 - 49. Jahrestagung der GUS 2020 CY - Stutensee, Blankenloch DA - 24.03.2021 KW - Warfarin KW - Transformationsprodukte KW - Umweltverhalten PY - 2021 SN - 978-3-9818507-6-5 VL - 49 SP - 265 EP - 272 AN - OPUS4-53963 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Dorgerloh, Ute A1 - Becker, Roland A1 - Riedel, Juliane A1 - Hofmann, Andrea T1 - Comparison of gas- and liquid chromatography-mass spectrometry for trace analysis of anilines in groundwater N2 - Three chromatographic procedures were investigated regarding their potential for the quantification of aniline and 19 of its methylated and chlorinated derivatives in groundwater. These methods were based on liquid-liquid-extraction in combination with gas chromatography and single quadrupole mass spectrometry (GC/MS) according to German standard DIN 38407-16:1999 and its extension using tandem mass spectrometry (GC/MS-MS), both following liquid-liquid extraction, and as third alternative the direct injection of the water sample into a liquid chromatograph coupled to tandem mass spectrometry (LC/MS-MS). Results were compared using fortified water and real-world contaminated groundwater used in an interlaboratory comparison. It could be shown that GC/MS and GC/MS-MS yielded results deviating less than 10% from each other while all three procedure displayed quantification results deviating less than 15% from the intercomparison reference values in case of each analyte in the concentration range between 1 and 45 µg L-1. Though GC/MS-MS displays a ten-fold higher sensitivity than single quadrupole GC/MS, the precision of both methods in the concentration range was similar. LC/MS-MS has the advantage of no further sample preparation due to the direct injection and leads for methylanilines and meta-, para- substituted chloroanilines to results sufficiently equivalent to the standardised GC/MS method. However, LC/MS-MS is not suitable for ortho-chloroaniline derivates due to significantly lower ion yields than meta- and para-substituted chloroanilines. KW - Interlaboratory comparison KW - Aniline KW - Chloroanilines KW - Methylanilines KW - Groundwater KW - GC/MS KW - GC/MS-MS KW - LC/MS-MS PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-535793 DO - https://doi.org/10.1080/03067319.2021.1987423 VL - 103 IS - 19 SP - 8465 EP - 8477 PB - Taylor & Francis CY - London AN - OPUS4-53579 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Maiwald, Michael T1 - OPC-UA für Prozessindustrie und Laborwelt - Offener Standard für die störungsfreie Kommuniktaion N2 - Im privaten Umfeld machen wir uns keine Gedanken mehr über die Einbindung von elektronischen Geräten in Netzwerke, sei es bei der Verbindung des Mobiltelephons mit dem W-Lan-Netzwerk des Hotels oder von Freunden oder des drahtlosen Kopfhörers mit dem Notebook: Wir erwarten, dass es auf Anhieb funktioniert. Dahinter stecken offene Standards für die Konnektivität. Dafür braucht man immer zwei Dinge: Einen physischen Kommunikationskanal, auf dem die Datenpakete zwischen den Komponenten und Geräten hin- und herwandern und eine einheitliche Kommunikationssprache, so dass sich die Komponenten und Geräte auch untereinander verstehen, weil sie dieselbe Sprache sprechen und automatisch wissen, was sie miteinander zu besprechen haben. Wenn man eine sichere Konnektivität bevorzugt, werden weitere Informationen benötigt, wie etwa ein Passwort. Auch für die störungsfreie Kommunikation aller Automatisierungskomponenten der Prozessindustrie untereinander oder etwa für die Kommunikation von Geräten in einem analytischen Labor wird ein einheitliches Protokoll (Kommunikationssprache) und ein einheitlicher Feldbus (Kommunikationskanal) benötigt. Mittlerweile gilt der Standard OPC Unified Architecture (OPC-UA) für das Erstere als gesetzt und kann als ein kleiner Triumph von Industrie 4.0 betrachtet werden. OPC-UA kann auf verschiedenen Kommunikationskanälen laufen – am liebsten solchen, die eine angemessene Bandbreite oder Datenrate haben. Was den dazu notwendigen physischen Kommunikationskanal betrifft, findet man aus verschiedenen Gründen einen sehr spezialisierten und historisch über Jahrzehnte gewachsenen Lebensraum vor, der Operational Technology (OT) genannt wird. Zumeist findet man Zwei- und Vierdraht-Techniken mit Strom- und Spannungs-Signalübertragung, zum Beispiel mit 4–20 mA. Kommunikationsstandards wie eingangs erwähnt, wie Ethernet oder W-Lan sind absolut die Seltenheit. Dieses wird für die sogenannte Informationstechnik (Information Technology, IT) eingesetzt und ist aus Sicherheitsgründen heute komplett von der OT getrennt. Auch wenn es der OT nicht gefällt, wachsen die beiden Welten IT und OT aber stetig weiter zusammen, da die IT bereits viele funktionierende und preislich attraktive Lösungen bereithält. Die vermeintliche Sicherheit, Verfügbarkeit und Echtzeitfähigkeit der klassischen OT, aber auch ihre Komplexität müssen im Zuge der Digitalen Transformation nun sorgfältig abgewogen werden. KW - Prozessindustrie KW - Laborkommunikation KW - Konnektivität KW - OPC-UA KW - Digital Transformation KW - Labor-IT PY - 2021 UR - https://analyticalscience.wiley.com/do/10.1002/was.00170277/full/blaetterkatalog_git1021.pdf SN - 0016-3538 VL - 65 IS - 10 SP - 1 EP - 4 PB - Wiley CY - Weinheim AN - OPUS4-53566 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Scala-Benuzzi, M. L. A1 - Soler-Illia, G. J. A. A. A1 - Rabia, J. A1 - Battaglini, F. A1 - Schneider, Rudolf A1 - Pereira, S. V. A1 - Messina, G. A. T1 - Immunosensor based on porous gold and reduced graphene platform for the determination of EE2 by electrochemical impedance spectroscopy N2 - In this work, we report an electrochemical immunosensor to detect ethinylestradiol in water samples, using electrochemical impedance spectroscopy (EIS) as a detection technique. For the development of this immunosensor, the direct modification of the working electrode of a screen-printed carbon electrode was carried out. First, to reduce the resistance of the electrode, electroreduced graphene was incorporated on the surface. Second, a porous gold structure was electrodeposited on reduced graphene by electrodeposition and the dynamic hydrogen bubble template assisted method. Thus, a marked increase in surface area was obtained for anti-EE2 antibodies immobilization. Subsequently, the specific anti-EE2 antibodies were covalently immobilized using α-lipoic acid for attaching them to the gold surface. The electrode modified with the antibodies was incubated for 30 min in the samples containing EE2, producing the specific Antigen antibody binding. As the charge transfer resistance of a redox probe in the electrode surface is governed by the surface blocking effects, the charge transfer resistance was related to the amount of EE2 captured to realize a quantitative determination. For this, the EIS measurements were performed in a 4 mM [Fe(CN)6]4−/3− solution in 0.1 M KCl. The obtained Nyquist diagrams were adjusted using the Randles circuit as an equivalent circuit to obtain the corresponding resistances. The developed methodology showed good selectivity, precision, and sensitivity; although the LOD obtained was higher than those presented in other published articles, it turned out to be an alternative that allows the determination of ethinylestradiol using a simple disposable electrode. KW - Ethinylestradiol KW - Biosensor KW - Elektrochemisch KW - Impedanz PY - 2021 DO - https://doi.org/10.1016/j.jelechem.2021.115604 SN - 1572-6657 VL - 897 SP - 115604 EP - 115611 PB - Elsevier B.V. CY - Amsterdam AN - OPUS4-54048 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Raysyan, Anna A1 - Schneider, Rudolf T1 - Development of a Lateral Flow Immunoassay (LFIA) to Screen for the Release of the Endocrine Disruptor Bisphenol A from Polymer Materials and Products N2 - One of the most important chemicals used in the production of polymer plastics and coatings is bisphenol A. However, despite the large number of studies on the toxicity and hormonal activity of BPA, there are still open questions and thus considerable media attention regarding BPA toxicity. Hence, it is necessary to develop a sensitive, simple, cost-efficient, specific, portable, and rapid method for monitoring bisphenol A and for high sample throughput and on-site screening analysis. Lateral flow immunoassays have potential as rapid tests for on-site screening. To meet sensitivity criteria, they must be carefully optimized. A latex microparticle-based LFIA for detection of BPA was developed. The sensitivity of the assay was improved by non-contact printing of spot grids as the control and test lines with careful parameter optimization. Results of the test could be visually evaluated within 10 min with a visual cut-off of 10 µg/L (vLOD). Alternatively, photographs were taken, and image analysis performed to set up a calibration, which allowed for a calculated limit of detection (cLOD) of 0.14 µg/L. The method was validated for thermal paper samples against ELISA and LC–MS/MS as reference methods, showing good agreement with both methods KW - Endokrine Disruptoren KW - Polymere KW - Schnelltest KW - Endocrine disruptor KW - Bisphenol A KW - LFIA KW - Immunoassay PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-540492 DO - https://doi.org/10.3390/bios11070231 VL - 11 IS - 7 SP - 1 EP - 11 PB - MDPI CY - Basel AN - OPUS4-54049 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Schneider, Rudolf T1 - Portable, antikörperbasierte Analysenverfahren für die Schadstofferfassung im Wasserkreislauf N2 - Die Wasseranalytik wird bestimmt von zwei Hauptbedarfen, einerseits der Erfassung der Gewässergüte und Grundwassergüte als Basis für die Bereitstellung einwandfreien Trinkwassers und zum anderen des Monitorings der Wasserqualität in den Prozessen der Wasseraufbereitung und der Abwasserreinigung. Für beide Anwendungsbereiche ist eine laborbasierte Analytik eher hinderlich, werden die Analysenergebnisse doch eigentlich vor Ort benötigt, um etwa Beprobungskampagnen rasch anpassen zu können. Des Weiteren werden die Daten auch schnell benötigt, um zeitnah in die technischen Prozesse einzugreifen. Organische Kontaminanten werden meist über chromatographische Verfahren, häufig gekoppelt mit einer massenspektrometrischen Detektion, in instrumentellen Labormethoden erfasst, und die hochauflösende Massenspektrometrie und die Non-target-Analytik haben viel zum Wissen über das Vorkommen, insbesondere von sogenannten „Emerging Contaminants“, beigetragen, sowie zum Verständnis der beteiligten Eintragspfade und Abbauwege. Moderne, portable Analysenmethoden und insbesondere Sensoren werden zukünftig aber ihren Platz in der Schadstofferfassung im Wasserkreislauf einnehmen. Dies gilt insbesondere für antikörperbasierte Analysenverfahren, da erst diese oft die nötige Selektivität und Sensitivität in die Detektionsmethoden einbringen. KW - Biosensoren KW - Schnelltests KW - Immunoassay KW - Abwasser PY - 2021 DO - https://doi.org/10.1002/vomw.202100005 VL - 119 IS - 2 SP - 35 EP - 53 PB - Wiley-VCH GmbH CY - Weinheim AN - OPUS4-54050 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Schneider, Rudolf T1 - Portable, antikörperbasierte Analysenverfahren für die Schadstofferfassung im Wasserkreislauf N2 - Die Wasseranalytik wird bestimmt von zwei Hauptbedarfen, einerseits der Erfassung der Gewässergüte und Grundwassergüte als Basis für die Bereitstellung einwandfrei-en Trinkwassers und zum anderen des Monitorings der Wasserqualität in den Prozes-sen der Wasseraufbereitung und der Abwasserreinigung. Für beide Anwendungsbe-reiche ist eine laborbasierte Analytik eher hinderlich, werden die Analysenergebnisse doch eigentlich vor Ort benötigt, um etwa Beprobungskampagnen rasch anpassen zu können. Desweiteren werden die Daten auch schnell benötigt, um zeitnah in die technischen Prozesse einzugreifen. Organische Kontaminanten werden meist über chromatographische Verfahren, häufig gekoppelt mit einer massenspektrometrischen Detektion, in instrumentelle Labormethoden erfasst und die hochauflösende Massen-spektrometrie und die Non-target-Analytik haben viel zum Wissen über das Vorkom-men, insbesondere von sog. „emerging contaminants“, beigetragen, sowie zum Ver-ständnis der beteiligten Eintragspfade und Abbauwege. Moderne, portable Analy-senmethoden und insbesondere Sensoren werden zukünftig aber ihren Platz in der Schadstofferfassung im Wasserkreislauf einnehmen. Dies gilt insbesondere für anti-körperbasierte Analysenverfahren, da erst diese oft die nötige Selektivität und Sensitivität in die Detektionsmethoden einbringen. T2 - Wasser 2021 - Jahrestagung der Wasserchemischen Gesellschaft in der GDCh CY - Online meeting DA - 10.05.2021 KW - Immunoassay KW - Biosensoren KW - Schnelltest KW - Elektrochemisch PY - 2021 AN - OPUS4-54051 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Schneider, Rudolf T1 - Environmental monitoring by high-throughput immunoanalytical methods and portable devices N2 - A vast number of emerging pollutants has been detected in the environment over the last decades. Analytical methods suitable for trace analysis are needed that are desirably also fast, inexpensive and, if possible, robust and portable. Immunoanalytical, i.e. antibody-based, methods which are available in a broad range of formats, can be profitably used here to analyse for the distribution and the trends of concentration levels of contaminants in the environment. Some of these formats are single-analyte but high-throughput methods. In order to use them wisely, indicator substances, sometimes called anthropogenic markers, should be selected and used in screening approaches. Other methods are suitable to be performed on portable instrumentation in the field (on-site) or in facilities such as wastewater treatment plants for on-line monitoring. Furthermore, there are array technologies that allow for parallel (multiplex) analysis of several analytes of interest. The microtiter-plate based ELISA (Enzyme-linked Immunosorbent Assay) is the method of choice for the analysis of a large number of samples [1]. ELISAs are available to monitor for anthropogenic markers such as the antiepileptic carbamazepine, the analgesic diclofenac, the antihistaminic cetirizine, the steroid hormone estrone, the antimicrobial sulfamethoxazole, the stimulants caffeine and cocaine, the priority pollutant bisphenol A, and the bile acid isolithocholic acid. For on-site screening and monitoring, simpler formats, like mix-and-read assays, e.g. the Fluorescence Polarization Immunoassay (FPIA) [2] or Lateral-flow Immunoassays (LFIA) [3] are more suitable tools, the latter based on dipsticks or cassettes, that is why they are also called pregnancy test-like assays. The suitability of multi-analyte formats such as immunomicroarrays depends on the choice of a signal-producing system that provides small uncertainties and good reproducibility of the measurements. Bead-based (“suspension”) arrays read out in flow cytometers are a powerful platform for multiplex assays [4]. Electrochemical formats run on portable devices provide additional advantages as no light source is required. They are most promising for stand-alone analysers and biosensors T2 - European Biosensor Symposium 2021 CY - Online meeting DA - 09.03.2021 KW - Biosensoren KW - Schnelltest KW - Elektrochemisch KW - Immunoassay PY - 2021 AN - OPUS4-54052 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Riedel, Soraya T1 - How to trace a cereal killer? N2 - How to trace the "cereal" killer? Fungal secondary metabolites, aka mycotoxins, pose a severe health risk for humans and animals. Since their occurrence in food, such as cereal products, is inevitable, the detection is of considerable importance. I want to demonstrate how the trace analysis of mycotoxins in foodstuff can be performed outside the laboratory. For this purpose, sensitive, antibody-based methods are combined with electrochemical smartphone-based detection. T2 - Bionnale 2021 CY - Online meeting DA - 12.05.2021 KW - Mycotoxins KW - Electrochemistry KW - Ochratoxin A KW - Food safety PY - 2021 N1 - Geburtsname von Riedel, Soraya: Höfs, S. - Birth name of Riedel, Soraya: Höfs, S. AN - OPUS4-52658 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ecke, Alexander A1 - Bohm, Konrad A1 - Schneider, Rudolf T1 - Magnetpartikelbasierte Immunoassays als vielseitiges Werkzeug für die Umweltanalytik N2 - Der Bedarf nach schnellen und kostengünstigen Analysemethoden in der (Trink-)Wasseranalytik steigt mit dem zunehmenden Eintrag pharmazeutisch aktiver Substanzen in die Umwelt. Insbesondere die Kontamination mit Antibiotika erscheint neben ökotoxikologischen Effekten vor allem in Hinblick auf die Evolution resistenter Keime im Wasser besorgniserregend. Wir berichten hier über die Entwicklung eines magnetpartikelbasierten Immunoassays (MBBA) zur Detektion des Breitbandantibiotikums Amoxicillin (AMX) in Wasserproben, der im Vergleich zum entsprechenden Enzyme-linked Immunosorbent Assay (ELISA) eine wesentlich kürzere Analysendauer sowie einen breiteren Messbereich mit niedrigerer Nachweisgrenze aufweist. Das angewendete Assayprinzip ist dabei leicht auf weitere Analyten übertragbar und ermöglicht die Implementierung in einen Immunosensor. KW - Magnetpartikel KW - Amoxicillin KW - Immunoassay KW - Wasser KW - Antikörper PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-527366 UR - https://www.gdch.de/netzwerk-strukturen/fachstrukturen/umweltchemie-und-oekotoxikologie/publikationen/mitteilungen/archiv/2021.html SN - 1618-3258 VL - 27 IS - 2 SP - 48 EP - 51 PB - Gesellschaft Deutscher Chemiker (GDCh) CY - Frankfurt/Main AN - OPUS4-52736 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Ecke, Alexander A1 - Schneider, Rudolf T1 - Pitfalls in the Immunochemical Determination of β-Lactam Antibiotics in Water N2 - Contamination of waters with pharmaceuticals is an alarming problem as it may support the evolution of antimicrobial resistance. Therefore, fast and cost-effective analytical methods for potential on-site analysis are desired in order to control the water quality and assure the safety of its use as a source of drinking water. Antibody-based methods, such as the enzyme-linked immunosorbent assay (ELISA), can be helpful in this regard but can also have certain pitfalls in store, depending on the analyte. As shown here for the class of β-lactam antibiotics, hydrolysis of the β-lactam ring is a key factor in the immunochemical analysis as it influences antibody recognition. With the antibody used in this study, the limit of detection (LOD) in the immunoassay could be significantly reduced by hydrolysis for the five tested penicillins, with the lowest LOD for carbenicillin (0.2 nmol/L) and the greatest impact on penicillins G and V (reduction by 85%). In addition to enhanced quantification, our strategy also provides access to information about the degree of hydrolysis in water samples as shown for the most abundant penicillin amoxicillin. KW - ELISA KW - Immunoassay KW - β-Lactam KW - Amoxicillin KW - Hydrolysis PY - 2021 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-522832 DO - https://doi.org/10.3390/antibiotics10030298 SN - 2079-6382 VL - 10 IS - 3 SP - 298 PB - MDPI CY - Basel, CH AN - OPUS4-52283 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Ecke, Alexander T1 - Magnetic Bead-Based Immunoassays for Online Sensing Applications N2 - The demand for quick and reliable online sensors is rising not only in the field of environmental analysis with new and diversifying sources of pollution. A major aspect of this is the contamination of (drinking) water with pharmaceuticals representing a potential health threat and requiring timely countermeasures in case of contamination. In this regard, antibody-based methods can be helpful as they enable fast and cost-effective on-site and potentially online analyses. The transition from the conventional plate-based formats like ELISA (enzyme-linked immunosorbent assay) to an immunosensor fit for online sensing can be achieved by using magnetic microparticles as a mobile sensing platform. Beads are prepared by either coupling the antibody or the analyte to the surface. Both these strategies are presented using the example of two relevant contaminants in water: diclofenac (DCF) & amoxicillin (AMX), and can be transferred to other analytes. The developed magnetic bead-based assays (MBBAs) exhibit lower LODs, enhanced sensitivity as well as shorter time of analysis than the respective ELISAs and can be integrated into an immunosensor with electrochemical detection on a microfluidic chip. T2 - European Biosensor Symposium 2021 CY - Online meeting DA - 09.03.2021 KW - Immunoassay KW - Magnetic Beads KW - Diclofenac KW - Amoxicillin KW - Online Sensing PY - 2021 AN - OPUS4-52285 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - von Törne, Wipert T1 - Einblicke in das Transformationsverhalten von Warfarin N2 - Warfarin ist seit den 1950er Jahren einer der bekanntesten Gerinnungshemmer. Beim Menschen wird es bis heute zur Vorbeugung von Thrombose und der Behandlung von Vorhofflimmern und Arrhythmie eingesetzt. Doch schon einige Jahre vor der FDA Zulassung (U. S. Food and Drug Administration) als Medikament wurde es zur Bekämpfung von Nagetieren eingesetzt. Bis heute wurde es nur teilweise durch weitaus wirksamere antikoagulante Rodentizide (ARs) der zweiten Generation, sogenannte Superwarfarine, ersetzt. In der Fachliteratur finden sich zahlreiche Beispiele für sekundäre und tertiäre Vergiftung mit ARs bei Wildtieren und anderen Nichtzieltieren. Bisher ist relativ wenig über die Bioakkumulation, Persistenz und Toxizität von ARs und deren Transformationsprodukten (TPs) in der terrestrischen und aquatischen Umwelt, sowie der Nahrungskette bekannt. Um Einblicke in mögliche Transformations- und Abbauprozesse von Warfarin unter umweltrelevanten Bedingungen zu erhalten und um diese Einflüsse zu simulieren, wurde hierzu eine elektrochemische Reaktorzelle in Kopplung mit einem Massenspektrometer verwendet. So können mögliche Phase-I-Metabolite und sogenannte „oxidation products“ vorausgesagt werden. Hinsichtlich des Verhaltens während der Wasseraufbereitung in Kläranlagen wurden technisch relevante Prozesse wie Chlorung, UV-Bestrahlung und Ozonung in Modellen simuliert, um hier potenzielle Abbauwege und -prozesse aufzuklären. Gebildete Transformationsprodukte wurden daher mittels unterschiedlicher chromatografischer Verfahren getrennt und massenspektrometrisch identifiziert, charakterisiert und verglichen. T2 - 49. Jahrestagung der GUS CY - Online meeting DA - 24.03.2021 KW - Warfarin KW - Transformationsprodukte PY - 2021 AN - OPUS4-52354 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -