TY - CONF A1 - Bresch, Harald T1 - Nanomaterial characterisation - The long way to standardisation N2 - In 1981 the OECD published the Test Guideline on Particle size and size distribution. This TG is still a valid document for the measurement of particles all over the world. When nanomaterials gained importance, ISO set up a technical commitee for Nanotechnologies in 2005 and the OECD followed this step in 2006 with the Working Party on Manufactured Nanomaterials. In the following years ISO and OECD published several documents about nanomaterials and the systematisation developed. In 2017 it was finally clear that nanomaterials need to be adressed in another way than chemicals and in 2020 ECHA revised the REACH-Annexes accordingly and included nanomaterials. Unfortunately there is a little problem with this: Only a few applicable test guidelines exit for the measurement of the nanomaterials. Several test guidelines date from 1981 and do not address nanomaterials. The logical next step for the OECD would be to publish a series of test guidelines which are indeed currently prepared and will be shown in this talk. Finally there is an additional need for the future of NM standardisation: Digitalisation. T2 - Bilateral workshop with Uni Bermingham CY - Online meeting DA - 10.03.2021 KW - Nano KW - Standardisation KW - Test guideline KW - OECD KW - Nanomaterial PY - 2021 AN - OPUS4-53822 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Schühle, Florian A1 - Richter, Matthias T1 - Determination and classification of seven-day uptake rates for indoor air VOCs into tube type diffusive samplers with Tenax® TA N2 - Axial passive sampling of VOCs with Tenax® TA and thermal desorption GC-MS analysis is an accepted alternative to active sampling in occupational hygiene. In theory, the uptake into the passive sampler is only dependent on the diffusion coefficient of the analyte in air and the geometry of the sampler (ideal adsorption). For characterization, the uptake rate (UR) is used defined as the ratio of the mass adsorbed and the product of ambient concentration and exposure time. Various reported effective uptake rates (UR,eff) differ to an increasing degree from ideal values (UR,id) with increasing exposure doses (denominator of the given definition of UR), (Tolnai, 2001). In national and international standards, uptake rates are essentially sorted by the applicable exposure time, while detailed information about the corresponding concentration range is lacking. Moreover, especially for long exposure periods as applied in indoor air monitoring, the number of itemized substances is limited. Therefore it is the aim of this contribution to review and expand uptake rate data by comparison of literature and own laboratory values of assured quality. Passive samplers were exposed to nine single compound atmospheres of known concentrations for seven days. Concentrations were checked twice a day via active sampling.The determined uptake rates are considered accurate in terms of RSD and comparability to literature values and can be recommended for exposure times of seven days at 50 – 100 μg m-3 (approximately 100 – 300 ppm min). Seven-day uptake rates in ISO16017-2 and ASTM D6196 are not generally suited for this purpose, as has been exemplarily shown for benzene. Thus, it is crucial for optimization of the method to produce more reliable uptake rate data, including specific information about applicable exposure times and concentrations, which will be promoted in the course of this ongoing study. T2 - Healthy Buildings America 2021 CY - Online meeting DA - 18.01.2022 KW - Diffusive sampling KW - Volatile organic compounds (VOC) KW - Effective uptake rate PY - 2022 UR - https://www.isiaq.org/docs/HB2021America262.pdf AN - OPUS4-54372 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Even, Morgane A1 - Juritsch, Elevtheria A1 - Richter, Matthias T1 - Development of an efficient procedure for the analysis of the emissions of very volatile organic compounds (VVOCs) in emission measurements from construction products and in the indoor air N2 - The ISO 16000-6 standard describes a method for the determination of volatile organic compounds (VOCs) in indoor and test chamber air by sorbent-based active sampling, thermal desorption and gas chromatography coupled with mass spectrometry (GC/MS). It also gives directions to adapt this methodology to very volatile organic compounds (VVOCs). Indeed, toxicologically based guideline values are being implemented for these compounds and it becomes necessary to measure them. But a comprehensive and robust measurement method is lacking. Investigations on the use of gaseous standards, the suitability of chromatography columns, the suitable sorbent combinations and the water removal are required. The talk will highlight the points that need to be explored towards the standardisation of a suitable procedure and provide appropriate preliminary results. T2 - Emissions and Odours from Materials 2021, Certech CY - Online meeting DA - 07.10.2021 KW - VVOCs KW - Analytical method KW - ISO 16000-6 KW - Thermal desorption KW - Gas chromatography PY - 2021 AN - OPUS4-53485 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Abram, Sarah-Luise A1 - Mrkwitschka, Paul A1 - Prinz, Carsten A1 - Rühle, Bastian A1 - Kuchenbecker, Petra A1 - Hodoroaba, Vasile-Dan A1 - Resch-Genger, Ute T1 - Monodisperse iron oxide nanoparticles as reference material candidate for particle size measurements N2 - In order to utilize and rationally design materials at the nanoscale the reliable characterization of their physico-chemical properties is highly important, especially with respect to the assessment of their environmental or biological impact. Furthermore, the European Commission’s REACH Regulations require the registration of nanomaterials traded in quantities of at least 1 ton. Powders or dispersions where 50% (number distribution) of the constituent particles have sizes ≤ 100 nm in at least one dimension are defined as nanomaterials. This creates a need for industrial manufacturers and research or analytical service facilities to reliably characterize potential nanomaterials. Currently, BAM is developing reference nanoparticles, which shall expand the scarce list of worldwide available nano reference materials certified for particle size distribution and will also target other key parameters like shape, structure, porosity or functional properties. In this respect, materials like iron oxide or titanium dioxide are considered as candidates to complement the already available silica, Au, Ag, and polystyrene reference nanoparticles. The thermal decomposition of iron oleate precursors in high boiling organic solvents can provide large quantities of iron oxide nanoparticles that can be varied in size and shape.[1, 2] The presence of oleic acid or other hydrophobic ligands as capping agents ensures stable dispersion in nonpolar solvents. Such monodisperse, spherical particles were synthesized at BAM and pre-characterized by electron microscopy (TEM, SEM including the transmission mode STEM-in-SEM) and dynamic light scattering comparing cumulants analysis and frequency power spectrum. 1. REACH regulations and nanosafety concerns create a strong need for nano reference materials with diverse properties. 2. Iron oxide nanoparticles are under development as new candidate reference material at BAM. 3. Narrow particle size distribution confirmed by light scattering and electron microscopy. T2 - Nanosafety 2020 CY - Online meeting DA - 05.10.2020 KW - Iron oxide nanoparticles KW - Reference material KW - Particle size KW - Electron microscopy KW - Nanoplattform PY - 2020 AN - OPUS4-52774 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Seneschal-Merz, Karine A1 - Bücker, Michael A1 - Wachtendorf, Volker A1 - Heidrich, Christa A1 - Sander, Christoph T1 - New UV-protective glazing for the conservation of cultural assets N2 - For centuries, churches, secular buildings and museums have been furnished with valuable works of art. Many works of art are UV- and light-sensitive. It is well known that especially UV radiation causes damage and discoloration in paintings, textiles, plastics, wood and other materials. In particular, the wavelengths between 280 and 410 nm cause color changes, embrittlement or destruction of e.g. plastics over time. Therefore, strongly UV-absorbing glasses are advantageous for architecture and are necessary for the protection of cultural assets. As restorers in all disciplines become more and more aware, the demands placed on buildings and their furnishings in terms of climate are becoming increasingly detailed and precise. The aim in each case is to preserve the valuable artwork. For some years now, the industry has been offering the protection of cultural objects by installing special UV-protective glasses. Currently, UV protection for church buildings is realized by laminated safety glass equipped with appropriate UV-protective plastic films. Technically, this always means a second pane of glass in front of the windows, which is installed outside the building. This creates climatic gaps that are difficult to control and deterioration due to ageing effects can be expected. At the same time, this protective glazing is not invisible and has a considerable aesthetic influence on the interior and exterior appearance of the building. Meanwhile, the preservation of historical monuments accepts such aesthetic cuts on buildings in order to protect the artwork in the interior from UV light. To this day, however, the long-term durability of UV protection provided by inserted plastic films is still controversial. To date the only available alternative on the market is a mouth-blown UV protection glass which uses a so-called overlay to provide UV protection. This shows that UV protection can also be achieved by glasses without plastic films thus realizing an exclusive inorganic protection which normally is more stable than a polymeric one. So far there are not enough studies to prove long-term durability. The aim of this project is to provide existing glazing or new glazing to be created with a highly transparent layer that ensures this UV-protective filter function below 400 nm. The glass coating is to be applied to the glass over a large area and fired into the surface like a classic ceramic enamel paint with the same technics. In addition, it should be long-term durable in its function. It means, that the glass has to be fused at temperatures below 630 °C during the firing process, its chemical durability has to be high, its coefficient of thermal expansion has to be as close as the one of the substrate (usually float glass) and the glass has to absorb the UV-radiation within a thin thickness (thinner as 100 µm). In this project, the long-term durability of commercial UV-protective glasses is examined. New low melting glasses containing UV-absorbing ions are being developed. Their UV-absorption as a thin layer is analyzed as well as their chemical durability and their thermal properties. We are grateful to BMWI for the financial support in the frame of the Central Innovation Programme for SMEs (ZIM). T2 - HVG-DGG: 94. Glastechnische Tagung CY - Online meeting DA - 10.05.2021 KW - Glass KW - Low melting KW - Chemical durability KW - Weathering tests KW - Aging test KW - UV absorption KW - UV protection KW - Architecture KW - Optical properties PY - 2021 AN - OPUS4-52648 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Even, Morgane A1 - Juritsch, Elevtheria A1 - Richter, Matthias T1 - Development of a procedure for the analysis of the emissions of VVOCs – Results of a research project N2 - Since 1997, the Committee for Health-related Evaluation of Building Products (AgBB) has been developing the basis for building regulations for protection against indoor health risks that may arise when building products are used. In 2009, the AgBB decided to include relevant VVOCs in the assessment procedure. The ISO 16000-6 (2021) states that the use of the thermal desorption gas chromatography technique is appropriate for VVOCs, if adaptations are considered. However, a suitable method for the trustworthy quantification of VVOC emissions from building products and in the indoor air is still missing. This webinar will present the results of a research project on the development of a procedure for VVOC analysis: • identification of the gaps towards standardization • investigations on gas standards and the suitability of chromatography columns • investigations on sorbent combinations and water management • validation of the method and screening of VVOC emissions from building products. T2 - Webinar organized by Umweltbundesamt and BAM CY - Online meeting DA - 13.09.2022 KW - Analytical method KW - EN 16516 KW - ISO 16000-6 KW - VVOCs PY - 2022 AN - OPUS4-55693 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Richter, Matthias A1 - Horn, Wolfgang A1 - Mölders, N. A1 - Sauerwald, T. A1 - Schultealbert, C. A1 - Mull, B. T1 - Two new promising approaches for quality assurance measures for materials emissions testing N2 - Two new approaches towards an emission reference material for use in quality assurance measures for materials emissions testing were developed and intensively tested. The overall goal was to obtain solid materials with homogenous and reproducible (S)VOC release. Since the application in inter-laboratory comparisons is aimed at, it should furthermore be long-term stable to ensure safe shipment to the customer without sustaining compound losses. In the first approach, thermoplastic polyurethane (TPU) was impregnated with the VOC texanol under high-pressure with liquid CO2 as solvent. In the second, styrene (VOC) and the SVOC 2,6-diisopropylnaphthalene (DIPN) were spiked into vacuum grease (VG) and a mixture of paraf-fin/squalane (P/S). For the prediction of the emission rates a finite element model (FEM) was developed for the VG and P/S type materials. All requirements for reference materials were fulfilled, whereas the TPU samples need to be aged for about 10 days until repeatable and reproducible emission rates were obtained. T2 - Indoor Air 2018 CY - Philadelphia, PA, USA DA - 22.07.2018 KW - Emission reference material KW - Volatile organic compounds KW - CO2 assisted impregnation KW - FEM model PY - 2018 AN - OPUS4-45611 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Bresch, Harald A1 - Schmidt, Alexandra A1 - Bachmann, V. A1 - Kämpf, K. A1 - Kuhlbusch, T. A1 - Schwirn, K. A1 - Völker, D. T1 - OECD Test Guideline on particle size and size distribution of manufactured nanomaterials N2 - The properties of nanomaterials are influenced not only by their chemical composition but also by physical properties (such as size, geometry and crystal structure). For the reliable determination and assessment of behaviour and effects of nanomaterials as well as for the determination of the exposure of humans and environment a comprehensive physical-chemical characterization of nanomaterials is essential. This is an important prerequisite to identify them as nanomaterials and to interpret and compare test results and - in future – to forecast interaction and effects of nanomaterials. In 2006, the OECD launched a sponsorship program for the testing of nanomaterials in which 11 nanomaterials were thoroughly investigated using a variety of methods. The aim of the project was, among other things, to find out where problems occur and where there are gaps in the measurement and test procedures and where are changes required. An important outcome of the sponsorship program was the finding that the OECD Test Guidelines should in several cases be extended to the specific needs in testing of nanomaterials. The existing standardized test methods of the OECD for physical-chemical characterization have not been developed for nanomaterials in particular. A high demand for an extension of the test guidelines was identified. Germany complied with the OECD's request in 2017 and has agreed to extend the “Test Guideline on Particle Size Distribution / Fiber Length and Diameter Distributions Test Guideline” for Manufactured Nanomaterials (MN). UBA commissioned BAM and BAuA with the preparation of the Test Guideline. The aim of the project is the development of a harmonized test protocol for a valid and reproducible determination of particle size and size distribution which is one of the most relevant physical-chemical properties for MNs. Different measuring methods provide different results for the size distribution of the particles. This is caused by the different measuring principles of the methods. Each method measures a specific parameter that ultimately determines particle size. First, the measured quantity differs for each method (Scattered light intensity, 2D image / projection, electric mobility, etc.). Second, the calculated diameters of the MN may differ (Feret Diameter, Area Projection, Mobility Diameter, Aerodynamic Diameter, Hydrodynamic Diameter). Third, a measuring method provides a size distribution which is measured either mass-based, surface-based or number-based. A conversion between the results requires additional parameters and thus possibly increases the measurement error. In addition to the technical differences, the individual parameters are strongly influenced by the structure and material of the nanoparticles. For example, a surface functionalization can lead to very different results in the size distribution. The suitability of measurement methods differs with the material of the MN. As a result, two very different results can be measured for the particle size distribution using two different methods, which are nevertheless both correct. Several large projects in recent years therefore concluded that nanomaterials should be characterized by at least two complementary method. Imaging techniques are regarded as one of these methods for the characterization, the complementary methods are supposed to be statistical methods. The different results for the size distribution of nanomaterials become problematic for the registration of new MN. A comparable and reproducible size distribution is a prerequisite for a standardized registration. In the future, the particle size distribution in the EU will also decide on the classification of a substance as a nanomaterial or as a non-nanomaterial. Especially in borderline cases, a standardized and comparable measurement methodology is therefore essential. T2 - BAM-PTB Workshop 2018 CY - Berlin, Germany DA - 14.05.2018 KW - Nano KW - OECD KW - Test Guideline KW - Size distribution KW - Particle PY - 2018 AN - OPUS4-45106 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Horn, Wolfgang A1 - Jann, Oliver A1 - Hahn, Oliver T1 - The BEMMA-Scheme helpful for low VOC inside display cases? N2 - Museums worldwide are equipped with different display cases. Exhibit display cases should protect cultural objects from dust as well as from mechanical and physical damage. To ensure a stable climate inside the display cases, a low air exchange rate is maintained. Typically, air exchange rates are often smaller than 0.1 d-1, which can result in rising concentrations of potential harmful VOC inside of the display cases due to emissions from materials. Especially organic acids, e. g. formic or acetic acid which can emit from e.g. sealing materials or wood-based materials, can produce damage of cultural objects. In 2012 BAM introduced a procedure which is called: BEMMA-Scheme (Bewertung von Emissionen aus Materialien für Museums¬ausstattungen) which stands for: “Assessment of Emissions from Materials for Museum Equipment”. Regarding the testing procedure only construction materials were evaluated, not the display cases their self. Micro chambers are used for VOC emission tests of display case construction materials, e.g. plastics, sealing materials, coatings, textiles and others. Each sampling procedure is carried out in duplicate. Emissions like formic acid, acetic acid, formaldehyde, piperidine-derivates and oximes are excluded and the sum of emissions of VVOCs, VOCs and SVOCs is limited. For a positive assessment all listed criteria must be fulfilled; otherwise the display construction material fails the BEMMA scheme. T2 - Indoor Air Quality in Heritage and Historic Environments CY - Online meeting DA - 12.10.2020 KW - Museum KW - VOC-emission KW - BEMMA KW - Display case PY - 2020 AN - OPUS4-51470 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Bresch, Harald T1 - Standard Operating Procedures in the digital context N2 - The intention of the presentation is to inspire a discussion on the needs and challenges for the digitalisation of SOPs. SOPs are available for the disciplines of physical-chemical characterisation, for toxicology, for environmental applications and for exposure applications. Furthermore the SOPs need to be integrated in the regulatory framework and need to be detailed enough for the digital processing. This is highlighted in this presentation. T2 - Nanosafety 2020 CY - Online meeting DA - 07.10.2020 KW - Nano KW - SOP KW - NFDI KW - Digitalisation KW - Operating Procedure PY - 2020 AN - OPUS4-51491 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Seneschal-Merz, Karine A1 - Bücker, Michael A1 - Sander, C. T1 - Development of low melting glasses as durable transparent enamel colors for the manufacture of decorated glass panels N2 - Today glass is broadly used in modern architecture. For applications indoor it is possible to produce decor glass by using enamel colors and glass painting techniques without any problems. However, this is more limited for applications outdoor. Humidity and environmental pollution attack the surface of the coating and damage it strongly. There are only few colors on the market which are resistant towards acids and bases until now. Additionally, most of those colors are opaque. In order to extend the color palette, chemically resistant colored glasses are being developed which are transparent, relatively low melting and intensively toned even in thin coat thickness. To achieve such ambitious aim, many parameters have to match which act in complex manner. Metal oxides were used to color the glasses. A lead-free glass composition was developed to avoid an exposition of heavy metals to the environment. The glasses were characterized, in particular in terms of their thermal properties, their crystallization and corrosion behavior as well as their chemical and environmental durability. Different practical applications will be shown. T2 - 15th International Conference on the Physics of Non-Crystalline Solids & 14th European Society of Glass Conference CY - Saint-Malon, France DA - 08.07.2018 KW - Glass KW - Chemical durability KW - Corrosion KW - Enamel colors KW - Low melting KW - Environmental friendly KW - Architecture KW - Thermal properties PY - 2018 AN - OPUS4-45755 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Bartelmeß, Jürgen A1 - Gawlitza, Kornelia A1 - Kraus, Werner A1 - Chlvi-Iborra, Katherine A1 - Tiebe, Carlo A1 - Noske, Reinhard A1 - Bartholmai, Matthias A1 - Rurack, Knut T1 - Developments Towards a BODIPY-based fluorometric sensing device for multiple hazardous gases N2 - Fluorometric sensing is a versatile approach for trace analysis outside of the laboratory, requiring suitable sensor materials and their integration into sensing devices. The versatility of fluorophores as probes, especially in terms of the possibility to tailor their optical as well as their recognition properties by synthetic modifications in a wide range, renders them as superior active component for the preparation of optical sensing devices. Recent works at BAM in this field include, for example, the detection of nerve gas agents, illustrating impressively the aforementioned benefits of fluorophores in optical sensing applications. In the interdisciplinary project presented here, we target hazardous gases such as ammonia, benzene, and hydrogen sulfide, next to others, which pose a major threat to human health and environmental safety and for which the availability of a sensitive and reliable detection method is highly desirable. The dyes presented follow a “turn-on” fluorescence schematic, which allows for the selective and sensitive detection of the respective gaseous analyte. The immobilization of the probe in polymeric matrices is then the next step toward the fabrication of a prototype device for molecular sensing. Further steps in the project include the assembly of instruments for test-atmosphere generation, the referencing of the sensor system, development and implementation of an optical setup, and the testing of the prototype device under laboratory conditions and in the field. In this presentation, we give an overview over the recent developments on this topic in our groups. Highlights are hydrogen sulfide sensitive, BODIPY based transition metal complexes, which allow for a sensitive as well as selective detection of the toxic gas. In addition, we present a novel class of highly substituted BODIPY derivatives – pocket-BODIPYs – which are of a synthetically high versatility and can readily be modified to create pockets in the periphery of the molecule of defined geometries. This is illustrated on the successful encapsulation of benzene by a pocket-BODIPY derivative, confirmed by X-ray crystallographic analysis as well as by further spectroscopic and analytical methods. T2 - International Conference on Porphyrins and Phthalocyanins (ICPP-10) CY - Munich, Germany DA - 01.07.2018 KW - BODIPY dye KW - Fluorometric sensing KW - Gas sensing PY - 2018 AN - OPUS4-45646 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Kuhlbusch, T. A1 - Schmidt, Alexandra A1 - Bresch, Harald A1 - Kämpf, K. A1 - Bachmann, V. A1 - Völker, D. A1 - Schwirn, K. T1 - Test guideline on particle size and size distribution of manufactured nanomaterials N2 - The particle size distribution is considered the most relevant information for nanoscale property identification and material characterization. The current OECD test guideline on particle size and size distribution (TG 110) is not applicable to ‘nano-sized’ objects. In this project we thus develop a new OECD test guideline for the measurement of the size and size distribution of particles and fibers with at least one dimension in the range of 1 - 1000 nm. A fiber is defined as an object having an aspect ratio of length/diameter l/d >3. The width and length of each fiber should be measured concurrently. In order to measure the particle size distributions, many techniques are available. 9 methods for particles and 2 methods for fibres have been tested in a prevalidation study and appropriate methods will be compared in an interlaboratory round robin test starting in February 2019. T2 - Expertenaustausch Nanocluster und Malta-Initiative CY - Vienna, Austria DA - 30.10.2018 KW - Nanopartikel KW - Nano KW - Nanomaterial KW - Nano particle KW - Test guideline PY - 2018 AN - OPUS4-46671 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Schmidt, Alexandra A1 - Bresch, Harald T1 - OECD test guideline on particle size and size distribution: Comparability of measurement results N2 - OECD-Prüfrichtlinien werden weltweit für regulatorische Zwecke angewandt, da sie eine harmonisierte und über internationale Ringversuche validierte Vorschrift zur Analyse vorgeben. Die bestehende Prüfrichtlinie TG 110 zur Größencharakterisierung von Partikeln und Fasern ist für viele Methoden und Materialklassen nicht geeignet, weswegen es Bedarf an einer Neuentwicklung gibt. Im Fokus der neuen TG steht die reproduzierbare und vergleichbare Anwendung von Methoden zur Bestimmung von Partikelgröße sowie Größenverteilung annährend sphärischer und faserförmiger Nanomaterialien im Bereich 1-1000 nm. Berücksichtigt werden die häufig angewandten Methoden zur Größencharakterisierung mit ihren unterschiedlichen physikalisch-technischen Limitierungen und die Bestimmung unterschiedlicher Durchmesser (hydrodynamischer, Feret…). Weiterhin sind materialspezifische Einflüsse durch die Partikel sowie das Messmedium zu berücksichtigen. Jede der Methoden soll unter Berücksichtigung von ISO-Standards und den materialspezifischen Limitierungen in der neuen Prüfrichtlinie beschrieben und über einen Ringversuch validiert werden. T2 - BAM BfR Workshop CY - Berlin, Germany DA - 21.03.2019 KW - OECD KW - Nanoparticle KW - Particle Size KW - Nano KW - Characterization PY - 2019 AN - OPUS4-47633 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Seeger, Stefan T1 - Overview on the EMPIR Project "AEROMET" N2 - Overview on the ongoing EU EMPIR Project "AEROMET" Aerosol Metrology for Atmospheric Science and Air Quality, Overview on preliminary resuilts in Workpackages 3 "Calibration of MPSS and CPCs" and 4 "Quantifying airborne particle compositions in the field" T2 - Meeting des ISO TC24/SC4, Joint session of Working Groups 9 "Single particle light interaction methods" and 12 "Electrical mobility and number concentration analysis for aerosol particles" CY - Graz, Austria DA - 11.04.2019 KW - Ambient air pollution KW - EU EMPIR KW - Chemical analysis KW - Total reflection X-ray spectroscopy KW - Airborne Particles PY - 2019 AN - OPUS4-47820 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Heuser, Lina A1 - de Oliveira Guilherme Buzanich, Ana A1 - Radtke, Martin A1 - Agudo Jácome, Leonardo A1 - Feldmann, Ines A1 - Deubener, J. T1 - Silver diffusion in low-melting alkali zinc borate model glasses studied by means of SNMS, TEM and XAS N2 - In many late-breaking research fields as in photovoltaics, microelectronics, nuclear waste glasses or at least mirror glasses silver diffusion in glasses is relevant to the issues of high-level functionality and recycling. The present study is focused on silver diffusion in innovative, low-melting alkali zinc borate glasses (X2O-ZnO-B2O3, X = Li, Na, K, Rb) potentially usable for silver metallization-pastes in solar cells. The glasses were coated with a thin metallic silver layer and heat treatments in air and nitrogen close to Tg at 470 °C for 2 h were performed. After heat treatment under air and nitrogen atmospheres the coating thickness, measured by a white light interferometer, was about 1.8 µm thick. Silver depth profiles determined by means of secondary neutral mass spectrometry (SNMS) indicate the fastest silver diffusion to a depth of 3.5 µm for Li2O-ZnO-B2O3 (LZB) glass. Nevertheless, the influence of the different alkali ions on the silver diffusion is small. The oxygen availability determines the silver diffusion into the glasses. The oxygen promotes the oxidation of the silver layer enabling Ag+ to diffuse into the glass and to precipitate as Ag0. Both species were detected by x-ray absorption spectroscopy (XAS). The precipitated metallic silver particles in Na2O-ZnO-B2O3 (NZB) glass have a mean size of 5.9 nm ± 1.2 nm diameter, which was determined using transmission electron microscopy (TEM). Phase separation in zinc-rich and zinc-poor phases with a mean diameter of 75 nm ± 20 nm occurred in NZB glass after heat treatment. Ion diffusion of the glasses into the silver layer was suggested by EDX-line scans. T2 - 94. Glastechnische Tagung CY - Online meeting DA - 10.05.2021 KW - Silver diffusion KW - Alkali zinc borate glass KW - Metallic silver precipitates KW - Phase separation PY - 2021 AN - OPUS4-52861 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Wilke, Olaf T1 - Emission Testing for Indoor Products N2 - The chemical emissions from products are tested by means of emission test chambers under defined conditions (climate, loading, air change rate). The standard method for the determination of volatile organic compounds (VOC) is the sampling onto Tenax-tubes followed by thermal desorption (TDS) and gas chromatography-mass spectrometry (GC-MS) analysis. The EU-LCI list includes some very volatile organic compounds (VVOC) and some VOC for which there are limitations when using the standard method. For VVOC additional sampling is required using stronger absorbers like Carbotrap or multi-bed adsorption tubes. The analysis of VVOC also requires a different GC oven program and a different column for the separation. For the determination of formaldehyde and other low boiling aldehydes (e.g. acetaldehyde, acetone, propanal, propenal) DNPH-cartridges are used which are extracted with acetonitrile followed by liquid chromatography (HPLC-UV) analysis. The derivatisation of propenal and other unsaturated aldehydes (e.g. 2-butenal) with DNPH might lead to lower findings due to incomplete derivatization and forming of by-products. For a better quantification of acetic acid the use of ion chromatography (IC) is recommended because the analysis of acetic acid with the standard method (TDS) leads to lower findings due to break through during sampling. The use of ion chromatography for the analysis of organic acids requires a third sampling technique. The acids are adsorbed onto silica-gel and extracted with water. T2 - ISESISIAQ 2019 CY - Kaunas, Lituania DA - 18.08.2019 KW - Indoor Air Quality KW - Emission Testing KW - Indoor Products KW - VOC Emission KW - Construction Products PY - 2019 AN - OPUS4-49406 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - de Krom, I. A1 - Heikens, D. A1 - Richter, Matthias A1 - Horn, Wolfgang A1 - Baldan, A. T1 - CRMs for (semi-)VOCs in sorbent tubes N2 - EN 16516 sets the test method and requirements for the determination of emissions of Volatile Organic Compounds (VOCs) from building materials into indoor air. To address the quality control requirements for the class of semi-VOCs (SVOCs), VSL developed gaseous reference materials. A novel home-made dynamic gas mixture preparation system, operating according to ISO 6145-4 (continuous injection method), has recently been developed and validated. Thanks to the stable temperature control up to 100 oC, the system can prevent condensation of the SVOCs in air at indoor air concentration levels. The in-situ obtained SVOC gas standards can be sampled in sorbent tubes to obtain SVOC transfer standards. A study was performed to determine the optimal sorbent material and storage conditions. This study will be presented together with the results of the 2018 Round Robin test for emission test chamber measurements organised by BAM. Using the novel system, VSL prepared transfer standards with known amounts of VOCs and SVOCs for participants to evaluate their analytical performance. T2 - Emissions and Odours from Materials CY - Brussels, Belgium DA - 07.10.2019 KW - CRM KW - VOC transfer standards KW - Round robin test KW - Material emissions testing PY - 2019 AN - OPUS4-49423 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Pollakowski-Herrmann, Beatrix A1 - Seeger, Stefan A1 - Gross, A. A1 - Kayser, Y. A1 - Osan, J. A1 - Stabile, L. A1 - Beckhoff, B. T1 - Quantitative elemental analysis of ambient aerosol particles using portable TXRF N2 - A reliable analysis of aerosol particle is curial for enforcing EU air quality regulations to protect human health, and for research on climate change effects [1]. Although metrics such as PM10 and PM2.5 are currently in use, the level of uncertainty of aerosol metrics is too high and the traceability is insufficient. Within the AEROMET project [2] procedures are developed aiming at reducing the uncertainties of particle mass, size, and number concentration measurements including the characterization of regulated components in airborne particles. Here, we present an approach how to improve the uncertainties of the particle mass by mobile total reflection x-ray fluorescence (TXRF) analysis. The combination of TXRF and aerosols sampling techniques supported by reference-free synchrotron radiation-based XRF enables a quantitative real-time analysis of particle mass. During in-field campaigns, the procedure was tested, monitoring the size dependent mass concentrations of specific elements in ambient aerosols under dynamic conditions. This approach allows a direct time and size-resolved analysis without laborious digestion steps and a reduced risk of contamination. Aerosol particles were sampled in a 13-stage DLPI impactor on acrylic discs. TXRF analysis was performed on-site with the transportable spectrometer S2 PICOFOX (Bruker Nano GmbH). The TXRF quantification was based on internal standardization. At moderate air pollution levels (PM10 20 µg/m³) sampling times of less than 2 hours were enough to detect elements in different particle size bins. The on-site approach and the high sensitivity of TXRF enables the observation of rather quick changes in the quantity and distribution of elements in an ambient aerosol on the day of sampling. The analysis of the morning and afternoon sampling shifts reveals the occurrence of the elements Fe, Ca and Si in different size bins as well as their temporal change in respective mass concentrations over the day while the distributions of several other elements remain unchanged. T2 - 11th International Conference on “Instrumental Methods of Analysis” CY - Ioannina, Greece DA - 22.09.2019 KW - Air quality KW - Novel sample preparation techniques KW - Analytical chemistry KW - Metals PY - 2019 AN - OPUS4-49249 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Seeger, Stefan T1 - Instruments for Fine and Ultrafine Particle Emission Characterization N2 - An overview on instruments and methods for the measurement and chemical characterization of particulate emissions from local sources is presented T2 - BAM - JBMIA (Japan Business Machine and Information System Industries Association) Meeting, JBMIA Emissions Working Group Meeting CY - Tokyo, Japan DA - 04.05.2019 KW - Aerosol spectroscopy KW - Particulate emission KW - Chemical analysis KW - Total reflection X-ray spectroscopy KW - Calibration PY - 2019 AN - OPUS4-47816 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -