TY - CONF A1 - Schuehle, Florian T1 - Quality assured uptake rates for passive sampling of indoor air VOCs N2 - Generation of experimental uptake rates (UR) for > 70 VOCs (7d /100 μg m-3) Application in German Environmental Survey – VOC exposure assessment Assurance of transparency concerning UR generation and uncertainty creates basis for decision making and comparability of measurement results. Exemplarily the uncertainty u(UR) of toluene UR was modelled and determined statistically with own laboratory data (relative standard deviation of actively sampled test gas atmosphere – RSDa and n = 5 exposed passive samplers – RSDp) in line with EN 838 as well as from variation of literature UR – RSD UR,lit. T2 - Airmon 2022, 10th International Symposium on modern principles of air monitoring and biomonitoring CY - Bristol, UK DA - 06.11.2022 KW - IAQ KW - VOC KW - Diffusive sampling KW - Uptake rate PY - 2022 AN - OPUS4-56583 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Horn, Wolfgang T1 - VOC-Sampling inside display cases with focus on BEMMA evaluation N2 - Museum display cases aim to protect art and cultural goods not only from theft and vandalism, but also from dirt and dust. An often-heard argument within a call for tender of new display cases is that there should be no emission inside of the display case. This is nearly impossible for new ones especially when they must also fulfill the requirement of an air change rate of less than 0.1 per day. For example, constructions or decorations in exhibition spaces, storage containers in depot rooms, and exhibition display cases release volatile substances that might lead to irreversible changes or, under certain circumstances, damage to the stored or displayed objects. In Germany beneath the well-established Oddy-test also the BEMMA procedure is used to evaluate the construction materials. For larger amounts of display cases control measurements of VOCs at the end of construction or buildup is quite common. However, what is necessary to measure the air inside of display cases. What are the circumstances when the display cases were closed before the test period? Is it possible to organize a clean environment for reliable measurements. How comparable are the resulting VOCs with results from BEMMA test and control measurement at the end. The BEMMA procedure should include a further VOC-test at 60 °C. What are the benefits of this enlargement. Mainly materials that have a high potential to store VOCs inside of the material in combination with a less permeable surface should be evaluated stricter than others. T2 - IAQ 2024 The International Conference on Indoor Air Quality in Heritage and Historic Environments CY - New York, USA DA - 18.11.2024 KW - Emission KW - Air quality KW - Display cases KW - VOC PY - 2024 AN - OPUS4-61967 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Klein, Luise T1 - Miniaturization Options in VOC Emissions Assessments of Consumer Products: Comparison of Test Chambers and Dynamic Headspace Extraction Systems N2 - Different emission testing devices were compared to measure reliable initial indoor air concen-trations of volatile organic compounds (VOC) emitting from consumer products. Compared to conventional medium- or large-scale test chambers, smaller devices including micro chambers and dynamic headspace extraction systems offer options for increased sample throughput, automation as well as reduced costs, which could facilitate their application in routine analysis of consumer product-related VOC emissions. T2 - Indoor Air 2024 CY - Honolulu, Hawaii, USA DA - 07.07.2024 KW - VOC KW - Emissions KW - Consumer product KW - Indoor air PY - 2024 AN - OPUS4-62098 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Klein, Luise T1 - Bestimmung der Aufnahmeraten von Carbonylen und VVOCs aus Raumluft in Passivprobenahmesystemen N2 - Für die Bewertung der Innenraumluftqualität könnte die Bestimmung der Konzentration von sehr flüchtigen organischen Verbindungen (VVOCs) von großer Wichtigkeit sein, aber für viele Komponenten fehlen die Daten. Auch und im Besonderen bei den GerES-Studien des UBAs erfolgt die Probenahme von flüchtigen organischen Verbindungen (VOCs) üblicherweise mit Passivsammlern über deren substanzspezifische Aufnahmeraten Rückschlüsse auf die Konzentration der einzelnen Verbindungen gezogen werden können. Reale Aufnahmeraten für die Berechnung der Innenraumluftkonzentration fehlen für viele der VVOCs. Ziel dieses UFOPLAN-Projektes ist es von 69 ausgewählten VVOCs und flüchtigen Carbonylen die Aufnahmeraten auf Passivsammler experimentell zu bestimmen. Dafür werden Erkenntnisse von zwei Vorgängerprojekten zur Entwicklung von Analysemethoden für VVOCs mittels Thermodesorptions-Gaschromatographie-Massenspektrometrie (TD-GC-MS) und zur Bestimmung der Aufnahmeraten von VOCs angewendet und angepasst. T2 - Innenraumluft 2024 CY - Dessau, Germany DA - 06.05.2024 KW - VVOC KW - Passivsammler KW - Innenraumluft PY - 2024 AN - OPUS4-62097 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Tang, Chi-Long T1 - Sub-4 nm particles from FFF-3D printing measured with the TSI 1 nm CPC and the Airmodus A11 nCNC N2 - Concerns have been raised as Fused Filament Fabrication (FFF) desktop 3D printer emits harmful ultrafine particles (dP < 100 nm) during operation in indoor spaces. However, the vast majority of previous emission studies have neglected the possible occurrence of sub-4 nm particles by using conventional condensation particle counter (CPC) for detection. Thus, the total particle emission could be systematically underestimated. This study has compared two diethylene glycol (DEG) based instruments to evaluate their suitability for measuring organic FFF particles in the sub-4 nm size range either as particle counter or as a particle size spectrometer. T2 - European Aerosol Conference 2024 CY - Tampere, Finland DA - 25.08.2024 KW - Air pollution KW - Emission testing KW - FFF-3D printing KW - Sub-4nm particles KW - Ultrafine particles PY - 2024 AN - OPUS4-60930 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Bonnerot, Olivier T1 - Are X-rays safe for manuscripts’ materials? N2 - In the last decade, applications of X-rays to the study of manuscripts significantly spread in both diversity and extent. They range from writing material analysis, mostly with X-ray fluorescence (XRF), permitting non-invasive characterization of inks and pigments used, to the investigation of the origin of writing supports. In addition, XRF mapping has proved to be an invaluable tool for recovering erased text. Finally, computed-tomography (CT) has shown potential in virtually unrolling rolls, making text readable without using-damaging mechanical methods. Despite their growing use, little attention has been paid to the side effects of such analytical tools. We observed irreversible parchment colour changes during some experiments on dead-sea scrolls with synchrotron radiation sources. Furthermore, partial photo-reduction of iron under high intensity beam during X-ray absorption near edge structure spectroscopy (XANES) measurements of iron-gall ink on paper has been reported several times [5,6]. Such phenomena have mostly been overlooked so far, although there is an increasing awareness of the necessity to study them. We conducted experiments at the Deutsches Elektronen-Synchrotron (DESY) facilities to investigate X-ray induced structural alteration of paper and parchment to see whether the presence of absorption centres (ink and pigments) has an impact. In addition to better understanding degradation processes, we are aiming to define an appropriate methodology of analysis of manuscripts with a tolerable risk of damage. The first results concerning X-ray induced damage of cellulose materials have already been presented at the Synchrotron Radiation and Neutrons in Art and Archaeology (SR2A). We are focusing here on the results on parchment materials. T2 - Cultural and Natural Heritage Workshop CY - Grenoble, France DA - 22.01.2020 KW - X-rays KW - Manuscripts KW - Parchment KW - Synchrotron PY - 2020 AN - OPUS4-50305 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Röllig, Mathias T1 - Thermografie zur Vergleichbarkeit von Emissionsmessung im 3D-Druck N2 - Die Bedeutung der Extrudertemperatur bei kleinen Desktop 3D Druckern der Fused Filament Fabrication (FFF) Technologie auf die Emission von ultrafeinen Partikeln im Druckbetrieb ist nicht zu unterschätzen. Allerdings zeigen kommerzielle 3D Drucker häufig eine systematische Temperaturabweichung zum Sollwert, d.h. die eingestellte Temperatur weicht von der tatsächlichen Temperatur ab. Da die Partikelemission meist temperaturabhängig ist, ist eine genaue Messung der tatsächlichen Extrudertemperatur für den Vergleich von Emissionsdaten aus unterschiedlichen Druckermodellen relevant. T2 - DGZfP Jahrestagung 2025 CY - Berin, Germany DA - 26.05.2025 KW - Thermografie KW - Fused Filament Fabrication (FFF) KW - Ultrafeine Partikel KW - Extrudertemperatur KW - Temperaturmessung PY - 2025 AN - OPUS4-63328 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Strzelczyk, Rebecca T1 - Zeolites loaded with VOCs as reference for material emissions testing N2 - Nowadays, people spend most of their time indoors. Thus, a good indoor air quality is important. Emissions of volatile organic compounds (VOCs) from furniture and building materials can cause health complaints1. Quantitative VOC-emission testing is carried out under standardized conditions in emission test chambers. In the presented project an emission reference material (ERM) is developed that emits a defined mixture of VOCs which is required for quality assurance and -control (QA/QC) measures. Porous materials (e.g zeolites, activated carbons, MOFs or aerogels) are used as reservoir materials and impregnated with VOC. The porous materials are selected, among others, by their pore size, pore size distribution, polarity and availability. Due to their regular pore structure zeolites are tested at first. For a prediction of the emission profile, the ERM is supposed to exhibit a constant emission rate over time. The aim is a stability of ≤ 10 % change in the emission rate over a minimum of 14 days. Method For impregnation, the material is placed into an autoclave inside a rotatable basket. The VOC is added and the autoclave is closed. Afterwards, CO2 is inserted. The closed system is then heated to the supercritical point of CO2 (31 °C, 73.75 bar). In this state, the CO2 acts as solvent for the VOC. By rotating the basket, the distribution of the VOC is ensured. After a few minutes, the pressure is decreased slowly and the CO2 is released. For the determination of the emission profile, the impregnated sample is placed into an emission test chamber. These chambers can be operated either with dry or humid air (50 ± 5 % rel. humidity). Every second to third day, air samples are taken and analyzed by gas chromatography. For an ideal impregnation, several different pressures and temperatures as well as impregnation times are tested. Results Two zeolite materials tested in dry air conditions reach emission profiles with a decrease of less than 10 % over 14 days (heptane and toluene, respectively). Further it was discovered that smaller pellets of the same zeolite show better results than bigger particles. When the pore size of a zeolite is too small, e.g. 0.3 nm, the VOC cannot be absorbed sufficiently. The main disadvantage of zeolites is their hygroscopicity because it has a large impact on the release of VOC when they are used in emission test chambers under standardized test conditions (23 °C, 50 % rel. humidity). Activated carbons have emission profiles with a larger change over 14 days. However, the high hydrophobicity allows measurements in humid air conditions which was not possible with the before mentioned hygroscopic zeolites. It is possible to impregnate powdered materials as well, and thus powdered non-hygroscopic (n.h.) zeolites were impregnated. Their emission profiles are comparable to those of the activated carbons. The use of methylated hygroscopic zeolites with a decrease in hygroscopicity did not yield successful emission measurements. The change over 14 days is calculated only for the stable phase (~250–300 h). The desired stability of ≤ 10 % change of the emission rate over 14 days could already be reached under dry testing conditions. Further investigations under humid conditions show that zeolites with high Si/Al-ratios are non-hygroscopic and comparable to activated carbons (20–30 % change). The next step is to reduce the change in the emission rate of these materials to the aimed ≤ 10 % over 14 days. T2 - Deutsche Zeolithtagung CY - Jena, Germany DA - 28.02.2024 KW - VOC KW - Emission KW - Quality assurance KW - Reference material KW - Zeolite PY - 2024 AN - OPUS4-59843 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Horn, Wolfgang A1 - Richter, Matthias A1 - Wilke, Olaf T1 - Volatile organic compounds from building products - Results from seven proficiency tests with emission test chambers conducted between 2008 and 2021 N2 - Emission testing of volatile organic compounds (VOC) from materials and products is commonly based on emission test chamber measurements. To ensure the comparability of results from different testing laboratories their measurement performance must be verified. For this purpose, Bundesanstalt für Materialforschung und -prüfung (BAM) organizes an international proficiency test every two years using well-characterised test materials (one sealant, one furniture board and four times a lacquer) with defined VOC emissions. The materials fulfilled the requirements of homogeneity, reproducibility, and stability. Altogether, 41 VOCs were included of which 37 gave test chamber air concentrations between 10 and 98 µg/m³. This is the typical concentration range to be expected and to be quantified when performing chamber tests. Four compounds had higher concentrations between 250 and 1105 µg/m³. The relative standard deviations (RSD) of BAM proficiency tests since 2008 are compared and the improvement of the comparability of the emission chamber testing is shown by the decrease of the mean RSD down to 23% in 2021. In contrast, the first large European interlaboratory comparison in 1999 showed a mean RSD of 51%. T2 - Proficiency Testing in Analytical Chemistry, Microbiology and Laboratory Medicine CY - Windsor, United Kingdom DA - 25.09.2023 KW - Proficiency Test KW - VOC-Emission KW - Chamber-test PY - 2023 AN - OPUS4-59694 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Horn, Wolfgang A1 - Müller, Birgit T1 - Low-odour and low-emission indoor products - Nose vers. odour threshold values N2 - To achieve good air quality in buildings, it is necessary to remove, avoid or reduce cleaning loads through ventilation. To reduce the cleaning loads and their effect on the odour perception of the occupants, a sensory test in indoor spaces and on building products is necessary in addition to VOC emission measurements of the contaminants. ISO 16000-28 provides the basis for odour assessment of products using a human olfactory panel. Through many years of experience and testing of these methods, it can now be assumed that reproducible odour ratings are available for many products used indoors. An odour measurement is mandatory for the Blue Angel for carpets and is being tested for environmentally friendly resilient floor coverings. Based on the results of an earlier project and new results of the upcoming study, the ISO should be revised and completed. The project investigates the odour and VOC emissions of building products based on natural raw materials. The data generated is used to address several issues, which are briefly explained below. Floor coverings based on linoleum and rubber are included in the Blue Angel eco-label with an odour assessment. The aim of the project is to clarify whether other assessment methods in addition to intensity can provide better information. Positive hedonics have an influence on the intensity assessment. Typically, building products show high intensity with low hedonic values. Results for wood-based products often show high intensity and positive hedonics. Based on these examples, an appropriate odour assessment of wood-based products for the Blue Angel, supported by research results, should be possible. The concept of the odour guide value which is based on odour threshold values was renewed in 2023. This approach will now be compared with the standardised odour measurement (16000-28). This assessment is based on GC/MS measurements and uses odour thresholds analytically when panel-based odour assessment is not possible. For this procedure well defined odour thresholds are necessary T2 - AIRMON 2025 The 11th International Symposium on Modern Principles of Air Monitoring and Biomonitoring. CY - Loen, Norway DA - 15.06.2025 KW - Emission KW - Indoor air quality KW - Materials emissions test KW - VOC KW - Geruch PY - 2025 UR - https://stami.no/en/wp-content/uploads/sites/2/2025/06/Programhefte-Airmon-2025-Juni-10-YTH.pdf AN - OPUS4-63634 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -