TY - JOUR A1 - Richter, Maria A1 - Güttler, Arne A1 - Pauli, Jutta A1 - Vogel, K. A1 - Homann, Christian A1 - Würth, Christian A1 - Resch-Genger, Ute T1 - Extending certified spectral fluorescence standards for the calibration and performance validation of fluorescence instruments to the NIR—closing the gap from 750 to 940 nm with two novel NIR dyes N2 - Fluorescence techniques such as fluorescence spectroscopy, microfluorometry, and fluorescence microscopy, providing spectral, intensity, polarization, and lifetime information, are amongst the most broadly utilized analytical methods in the life and materials sciences. However, the measured fluorescence data contain sample- and instrument-specific contributions, which hamper their comparability across instruments and laboratories. Comparable, instrument-independent fluorescence data require the determination of the fluorescence instrument’s wavelength-dependent spectral responsivity, also termed emission correction curve, for the same instrument settings as those used for the fluorescence measurements as a prerequisite for the subsequent correction of the measured instrument-specific data. Such a spectral correction is essential for the performance comparison of different fluorescent labels and reporters, quantitative fluorescence measurements, the determination of the fluorescence quantum yield, and the spectroscopic measure for the fluorescence efficiency of a fluorophore. Simple-to-use tools for obtaining emission correction curves are chromophore-based reference materials (RMs), referred to as fluorescence standards, with precisely known, preferably certified instrument-independent fluorescence spectra. However, for the increasingly used near-infrared (NIR) wavelength region >700 nm, at present, no spectral fluorescence standards are available. To close this gap, we developed two novel spectral fluorescence standards, BAM F007 and BAM-F009, with broad emission bands from about 580 to 940 nm in ethanolic solution. These liquid fluorescence standards currently under certification, which will be released in 2025, will expand the wavelength range of the already available certified Calibration Kit BAM F001b-F005b from about 300–730 to 940 nm. In this research article, we will detail the criteria utilized for dye and matrix selection and the homogeneity and stability tests accompanying dye certification as well as the calculation of the wavelength-dependent uncertainty budgets of the emission spectra BAM F007 and BAM-F009, determined with the traceably calibrated BAM reference spectrofluorometer. These fluorescence standards can provide the basis for comparable fluorescence measurements in the ultraviolet, visible, and NIR for the fluorescence community. KW - Quality assurance KW - Reference material KW - Fluorescence KW - Dye KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - NIR KW - Instrument performance validation PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-626317 DO - https://doi.org/10.1007/s00216-024-05723-w SN - 1618-2650 SP - 1 EP - 15 PB - Springer AN - OPUS4-62631 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - GEN A1 - Resch-Genger, Ute A1 - Richter, Maria A1 - Güttler, Arne A1 - Pauli, Jutta A1 - Vogel, K. A1 - Homann, Christian A1 - Würth, Christian T1 - Extending Certified Spectral Fluorescence Standards for the Calibration and Performance Validation of Fluorescence Instruments to the NIR – Closing the Gap from 750 nm to 940 nm with Two Novel NIR Dyes N2 - Fluorescence methods provide spectral, intensity, polarization, and lifetime information, which contain sample- and instrument-specific contributions. Fluorescence data, comparable across instruments and laboratories, require validated calibration procedures and certified fluorescence standards. KW - Quality assurance KW - Reference material KW - Fluorescence KW - Dye KW - Traceability KW - Metrology KW - Calibration KW - Reference data KW - Reference product KW - Digital certificate KW - NIR KW - Instrument performance validation PY - 2025 SP - 1 EP - 4 PB - Springer Nature CY - London AN - OPUS4-62739 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Hennig, Andreas A1 - Borcherding, H. A1 - Jäger, Christian A1 - Hatami, Soheil A1 - Würth, Christian A1 - Hoffmann, Angelika A1 - Hoffmann, Katrin A1 - Thiele, T. A1 - Schedler, U. A1 - Resch-Genger, Ute T1 - Scope and limitations of surface functional quantification methods: exploratory study with poly(acrylic acid)-grafted micro- and nanoparticles N2 - The amount of grafted poly(acrylic acid) on poly(methyl methacrylate) micro- and nanoparticles was quantified by conductometry, 13C solid-state NMR, fluorophore labeling, a supramolecular assay based on high-affinity binding of cucurbit[7]uril, and two colorimetric assays based on toluidine blue and nickel complexation by pyrocatechol violet. The methods were thoroughly validated and compared with respect to reproducibility, sensitivity, and ease of use. The results demonstrate that only a small but constant fraction of the surface functional groups is accessible to covalent surface derivatization independently of the total number of surface functional groups, and different contributing factors are discussed that determine the number of probe molecules which can be bound to the polymer surface. The fluorophore labeling approach was modified to exclude artifacts due to fluorescence quenching, but absolute quantum yield measurements still indicate a major uncertainty in routine fluorescence-based surface group quantifications, which is directly relevant for biochemical assays and medical diagnostics. Comparison with results from protein labeling with streptavidin suggests a porous network of poly(acrylic acid) chains on the particle surface, which allows diffusion of small molecules (cutoff between 1.6 and 6.5 nm) into the network. KW - Polymers KW - Surface groups KW - Quantification KW - Fluorescence PY - 2012 DO - https://doi.org/10.1021/ja302649g SN - 0002-7863 SN - 1520-5126 VL - 134 IS - 19 SP - 8268 EP - 8276 PB - American Chemical Society CY - Washington, DC AN - OPUS4-26002 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Zhang, X.-X. A1 - Würth, Christian A1 - Zhao, L. A1 - Resch-Genger, Ute A1 - Ernsting, N.P. T1 - Femtosecond broadband fluorescence upconversion spectroscopy: improved setup and photometric correction N2 - A setup for fluorescence upconversion spectroscopy (FLUPS) is described which has 80 fs temporal response (fwhm) for emission in the spectral range 425–750 nm. Broadband phase matching is achieved with tilted gate pulses at 1340 nm. Background from harmonics of the gate pulse is removed and sensitivity increased compared to previous designs. Photometric calibration of the upconversion process is performed with a set of fluorescent dyes. For Coumarin 153 in methanol the peak position, bandwidth, and asymmetry depending on delay time are reported. KW - Calibration KW - Fluorescence spectroscopy KW - Fluorescence KW - Upconversion KW - Standard KW - Spectral responsivity KW - Time-resolved fluorescence KW - Quality assurance KW - Coumarin 153 KW - Solvation dynamics PY - 2011 DO - https://doi.org/10.1063/1.3597674 SN - 0034-6748 SN - 1089-7623 VL - 82 SP - 063108-1 EP - 063108-8 PB - American Institute of Physics CY - Melville, NY AN - OPUS4-23886 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Koshkina, Olga A1 - Bantz, C. A1 - Würth, Christian A1 - Lang, Thomas A1 - Resch-Genger, Ute A1 - Maskos, Michael T1 - Fluorophore-labeled siloxane-based nanoparticles for biomedical applications N2 - We present the synthesis and characterization of multifunctional fluorophore-labeled poly(organosiloxane) nanoparticles with core-shell architecture, where the fluorescent dye is incorporated into the core. Grafting of heterobifunctional poly(ethylene oxide) (PEO) onto the particle surface leads to water-soluble biocompatible nanoparticles. Two different strategies have been used for the synthesis: The encapsulation of dye-labeled monomers during the polycondensation with additional PEO coating and subsequent dye labeling by covalent attachment of the fluorescent dye rhodamine B to the (chloromethylphenyl)siloxane groups in the core after polymerization and grafting of PEO onto the surface. Comparison of the fluorescence quantum yields of the nanoparticles before and after PEO coating show a decrease in quantum yield after PEO coating. KW - Biocompatibility KW - Core-shell KW - Fluorescence KW - Fluorescence quantum yield KW - Nanoparticles KW - PEO PY - 2011 DO - https://doi.org/10.1002/masy.201100041 SN - 1022-1360 SN - 0258-0322 SN - 1521-3900 VL - 309/310 IS - 1 SP - 141 EP - 146 PB - Wiley-VCH Verl. CY - Weinheim AN - OPUS4-25143 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Amouroux, B. A1 - Würth, Christian A1 - Roux, C. A1 - Eftekhari, A. A1 - Sliwa, M. A1 - Bouchet, A. A1 - Micheau, J.-C. A1 - Resch-Genger, Ute A1 - Coudret, C. T1 - Time-Resolved Rate Equation Analysis Disclose Kinetics Controlling Luminescence of Nanometer Tm-Upconverting Nanoparticles N2 - Upconversion luminescence of lanthanide-based upconversion nanoparticles (UCNPs) is a nonlinear step-wise process in which the consecutive absorption of multiple, low-energy photons results in the subsequent emission of a high-energy photon. The primary upconversion mechanism is energy transfer upconversion (ETU) from a sensitizer (Yb3+) to an activator (Tm3+). It requires the absorption of several excitation lowenergy photons by Yb3+, followed by the sequential energy transfer to Tm3+ions. Excited states relax to their ground states either radiatively by emitting a high-energy photon or non-radiatively by multiphonon relaxation through the crystalline host matrix. The time-resolved rise and decay luminescence curves of a set of five ultrasmall have been recorded under varying power near-infrared μs pulses. Six wavelengths have been used to monitor the evolution of the main Yb and Tm excited states. We use an average rate equations model to decipher the relationships between the compositional constraints and size of these ultrasmall UCNPs and the luminescence kinetic parameters. Several rate constants of ETU and other depopulation processes involving the multiple states of the Tm3+ energy scaffold have been retrieved from the simultaneous fit of the recorded curves. Their values have been interpreted by considering bulk and surface quenching, radiative and multi-phonon relaxations, and ion-to-ion hopping. Energy transfer between Yb3+ and Tm3+ is mainly occurring within neighbor atoms. The importance of mismatches on multiphonon relaxations, ETUs, and back-transfers has also been highlighted. For these numerical modeling, it appears that changing the composition and synthesis conditions with the aim to improve a single-specific parameter could remain a major challenge as this modification would automatically impact other properties with immediate consequences on UCNP dynamics. KW - Nano KW - Particle KW - Synthesis KW - Shell KW - Fluorescence KW - Lifetime KW - Decay kinetics KW - Method KW - Modelling KW - Quality assurance KW - Energy transfer KW - Upconversation PY - 2024 DO - https://doi.org/10.1021/acs.jpcc.4c04969 VL - 128 IS - 44 SP - 18836 EP - 18848 PB - ACS Publications AN - OPUS4-61645 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Würth, Christian A1 - Lochmann, Cornelia A1 - Spieles, Monika A1 - Pauli, Jutta A1 - Hoffmann, Katrin A1 - Schüttrigkeit, T. A1 - Franzl, T. A1 - Resch-Genger, Ute T1 - Evaluation of a commercial integrating sphere setup for the determination of absolute photoluminescence quantum yields of dilute dye solutions N2 - The commercial availability of stand-alone setups for the determination of absolute photoluminescence quantum yields (φf) in conjunction with the increasing use of integrating sphere accessories for spectrofluorometers is expected to have a considerable influence not only on the characterization of chromophore systems for use in optical and opto-electronic devices, but also on the determination of this key parameter for (bio)analytically relevant dyes and functional luminophores. Despite the huge potential of systems measuring absolute φf values and the renewed interest in dependable data, evaluated protocols for even the most elementary case, the determination of the fluorescence quantum yield of transparent dilute solutions of small organic dyes with integrating sphere methods, are still missing. This encouraged us to evaluate the performance and sources of uncertainty of a simple commercial integrating sphere setup with dilute solutions of two of the best characterized fluorescence quantum yield standards, quinine sulfate dihydrate and rhodamine 101, strongly differing in spectral overlap between absorption and emission. Special attention is dedicated to illustrate common pitfalls of this approach, thereby deriving simple procedures to minimize measurement uncertainties and improve the comparability of data for the broad community of users of fluorescence techniques. KW - Lifetime KW - Fluorescence KW - Luminescence KW - Quantum yield KW - Quantum efficiency KW - Integrating sphere KW - Reabsorption KW - Rhodamine 101 KW - Quinine sulfate dihydrate KW - Method KW - Photoluminescence KW - Standard KW - Emission KW - Spectral correction KW - Excitation KW - Anisotropy PY - 2010 DO - https://doi.org/10.1366/000370210791666390 SN - 0003-7028 SN - 1943-3530 VL - 64 IS - 7 SP - 733 EP - 741 PB - Society for Applied Spectroscopy CY - Frederick, Md. AN - OPUS4-22089 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Würth, Christian A1 - González, M.G. A1 - Niessner, R. A1 - Panne, Ulrich A1 - Haisch, C. A1 - Resch-Genger, Ute T1 - Determination of the absolute fluorescence quantum yield of rhodamine 6G with optical and photoacoustic methods - providing the basis for fluorescence quantum yield standards N2 - To establish the methodical basis for the development and certification of fluorescence quantum yield standards, we determined the fluorescence quantum yield Φf of rhodamine 6G (R6G) with two absolute methods with complementary measurement principles, here optical spectroscopy using an integrating sphere setup and pulsed laser photoacoustic spectroscopy (PAS). For the assessment of aggregation- and reabsorption-induced distortions of measured fluorescence quantum yields and procedures for the reliable consideration of such effects, this systematic comparison was performed in ethanol and in water employing different concentrations of R6G. In addition, the relative and absolute fluorescence quantum yields of these solutions were obtained with a calibrated spectrofluorometer and a commercialized integrating sphere setup. Based upon this systematic comparison, experimental advantages and systematic sources of variation were identified for both methods. KW - Fluorescence KW - Photoluminescence KW - Quantum yield KW - Photoacoustic spectroscopy KW - Absolute quantum yield KW - Integrating sphere KW - Rhodamine 6G KW - Aggregation KW - Reabsorption PY - 2012 DO - https://doi.org/10.1016/j.talanta.2011.12.051 SN - 0039-9140 VL - 90 SP - 30 EP - 37 PB - Elsevier CY - Amsterdam AN - OPUS4-25710 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Resch-Genger, Ute A1 - Behnke, Thomas A1 - Brehm, Robert A1 - Grabolle, Markus A1 - Hennig, Andreas A1 - Hoffmann, Angelika A1 - Hoffmann, Katrin A1 - Linck, Lena A1 - Lochmann, Cornelia A1 - Pauli, Jutta A1 - Spieles, Monika A1 - Würth, Christian T1 - Funktionelle Chromophor-Systeme, innovative Validierungskonzepte und rückführbare Standards für die fluoreszenzbasierte multiparametrische Bioanalytik N2 - Unter dem Motto „Innovation und Qualitätssicherung in der (Bio)Analytik“ werden in der Arbeitsgruppe Fluoreszenzspektroskopie der BAM, Bundesanstalt für Materialforschung und -prüfung, funktionelle Chromophor-Systeme, einfache Signalverstärkungs- und Multiplexingstrategien sowie innovative Validierungs- und rückführbare Standardisierungskonzepte für verschiedene fluorometrische Messgrößen und Methoden entwickelt. Im Mittelpunkt stehen dabei molekulare Fluorophore, Nanokristalle mit größenabhängigen optischen Eigenschaften (sogenannte Quantenpunkte, QDs) und fluoreszierende Partikel variabler Größe sowie Sonden und Sensormoleküle für neutrale und ionische Analyte und für die Charakterisierung von funktionellen Gruppen. Dabei erfolgen auch methodische Entwicklungen für die Fluoreszenzspektroskopie, die Fluoreszenzmikroskopie, die Milcrofluorometrie, die Sensorik und die Mikroarraytechnologie. Ziele sind u. a. das Design und die Untersuchung von multiplexfähigen selektiven und sensitiven Sonden für die Biomarkeranalytik, die Entwicklung von Methoden zur Charakterisierung der signalrelevanten Eigenschaften dieser Chromophor-Systeme und zur Charakterisierung von funktionellen Gruppen an Oberflächen und ihre Validierung sowie die Entwicklung und Bereitstellung von formatadaptierbaren, flexibel ersetzbaren Standards für die fluoreszenzbasierte Multiparameteranalytik. T2 - 5. Senftenberger Innovationsforum Multiparameteranalytik CY - Senftenberg, Deutschland DA - 10.03.2011 KW - Multiparametric KW - Multiplexing KW - Fluorescence KW - Nanoparticles KW - NIR dyes KW - Surface analysis KW - Quantum yield KW - Quantum dot KW - Lifetime PY - 2011 SP - 86 EP - 108 CY - Senftenberg AN - OPUS4-23635 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Würth, Christian A1 - Pauli, Jutta A1 - Lochmann, Cornelia A1 - Spieles, Monika A1 - Resch-Genger, Ute T1 - Integrating sphere setup for the traceable measurement of absolute photoluminescence quantum yields in the near infrared N2 - There is an increasing interest in chromophores absorbing and emitting in the near-infrared (NIR) spectral region, e.g., for applications as fluorescent reporters for optical imaging techniques and hence, in reliable methods for the characterization of their signal-relevant properties like the fluorescence quantum yield (Φf) and brightness. The lack of well established Φf standards for the NIR region in conjunction with the need for accurate Φf measurements in transparent and scattering media encouraged us to built up an integrating sphere setup for spectrally resolved measurements of absolute fluorescence traceable to radiometric scales. Here, we present the design of this setup and its characterization and validation including an uncertainty budget for the determination of absolute Φf in the visible and NIR. To provide the basis for better measurements of Φf in the spectral window from ca. 600 to 1000 nm used, e.g., for optical imaging, the absolute Φf of a set of NIR chromophores covering this spectral region are measured and compared to relative values obtained using rhodamine 101 as Φf standard. Additionally, the absolute Φf values of some red dyes that are among the most commonly used labels in the life sciences are presented as well as the absolute quantum yield of an optical probe for tumor imaging. KW - Fluorescence KW - Photoluminescence KW - Quantum yield KW - Dye KW - Uncertainty KW - Absolute quantum yield KW - Method comparison KW - Integrating sphere KW - NIR KW - Standards KW - Quantum yield standards PY - 2012 DO - https://doi.org/10.1021/ac2021954 SN - 0003-2700 SN - 1520-6882 VL - 84 IS - 3 SP - 1345 EP - 1352 PB - American Chemical Society CY - Washington, DC AN - OPUS4-25504 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Fiedler, Saskia A1 - Frenzel, Florian A1 - Würth, Christian A1 - Tavernaro, Isabella A1 - Grüne, M. A1 - Schweizer, S. A1 - Engel, A. A1 - Resch-Genger, Ute T1 - Interlaboratory Comparison on Absolute Photoluminescence Quantum Yield Measurements of Solid Light Converting Phosphors with Three Commercial Integrating Sphere Setups N2 - Scattering luminescent materials dispersed in liquid and solid matrices and luminescent powders are increasingly relevant for fundamental research and industry. Examples are luminescent nano- and microparticles and phosphors of different compositions in various matrices or incorporated into ceramics with applications in energy conversion, solid-state lighting, medical diagnostics, and security barcoding. The key parameter to characterize the performance of these materials is the photoluminescence/fluorescence quantum yield (Φf), i.e., the number of emitted photons per number of absorbed photons. To identify and quantify the sources of uncertainty of absolute measurements of Φf of scattering samples, the first interlaboratory comparison (ILC) of three laboratories from academia and industry was performed by following identical measurement protocols. Thereby, two types ofcommercial stand-alone integrating sphere setups with different illumination and detection geometries were utilized for measuring the Φf of transparent and scattering dye solutions and solid phosphors, namely, YAG:Ce optoceramics of varying surface roughness, used as converter materials for blue light emitting diodes. Special emphasis was dedicated to the influence of the measurement geometry, the optical properties of the blank utilized to determine the number of photons of the incident excitation light absorbed by the sample, and the sample-specific surface roughness. While the Φf values of the liquid samples matched between instruments, Φf measurements of the optoceramics with different blanks revealed substantial differences. The ILC results underline the importance of the measurement geometry, sample position, and blank for reliable Φf data of scattering the YAG:Ce optoceramics, with the blank’s optical properties accounting for uncertainties exceeding 20%. KW - Nano KW - Fluorescence KW - Reference material KW - Luminescence KW - Quantitative spectroscopy KW - Particle KW - Quantum yield KW - Quality assurance KW - Phosphor KW - Converter material KW - Lifetime KW - Interlaboratory comparison KW - Method KW - Uncertainty PY - 2024 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-600945 DO - https://doi.org/10.1021/acs.analchem.4c00372 SN - 0003-2700 SP - 6730 EP - 6737 PB - ACS Publications AN - OPUS4-60094 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Resch-Genger, Ute A1 - Carl, F. A1 - Grauel, Bettina A1 - Pons, Monica A1 - Würth, Christian A1 - Haase, M. T1 - LiYF4:Yb/LiYF4 and LiYF4:Yb,Er/LiYF4 core/shell nanocrystals with luminescence decay times similar to YLF laser crystals and the upconversion quantum yield of the Yb,Er doped nanocrystals N2 - We developed a procedure to prepare luminescent LiYF4:Yb/LiYF4 and LiYF4:Yb,Er/LiYF4 core/shell nanocrystals with a size of approximately 40 nm revealing luminescence decay times of the dopant ions that approach those of high-quality laser crystals of LiYF4:Yb (Yb:YLF) and LiYF4:Yb,Er (Yb,Er:YLF) with identical doping concentrations. As the luminescence decay times of Yb3+ and Er3+ are known to be very sensitive to the presence of quenchers, the long decay times of the core/shell nanocrystals indicate a very low number of defects in the core particles and at the core/shell interfaces. This improvement in the performance was achieved by introducing two important modifications in the commonly used oleic acid based synthesis. First, the shell was prepared via anewly developed method characterized by a very low nucleation rate for particles of pure LiYF4 shell material. Second, anhydrous acetates were used as precursors and additional drying steps were applied to reduce the incorporation of OH− in the crystal lattice, known to quench the emission of Yb3+ ions. Excitation power density (P)-dependent absolute measurements of the upconversion luminescence quantum yield (Φ,UC) of LiYF4:Yb,Er/LiYF4 core/shell particles reveal a maximum value of 1.25% at P of 180 W·cm−2. Although lower than the values reported for NaYF4:18%Yb,2%Er core/shell nanocrystals with comparable sizes, these Φ, UC values are the highest reported so far for LiYF4:18%Yb,2%Er/LiYF4 nanocrystals without additional dopants. Further improvements May nevertheless be possible by optimizing the dopant concentrations in the LiYF4 nanocrystals. KW - Nano KW - Crystal KW - Quantum yield KW - LiYF4 KW - Synthesis KW - Lifetime KW - Fluorescence KW - NIR KW - Photoluminescence KW - Lanthanide KW - Upconversion nanoparticle KW - Nanomaterial PY - 2020 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-515395 DO - https://doi.org/10.1007/s12274-020-3116-y SN - 1998-0124 VL - 14 IS - 3 SP - 797 EP - 806 PB - Springer AN - OPUS4-51539 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Würth, Christian A1 - Grabolle, Markus A1 - Pauli, Jutta A1 - Spieles, Monika A1 - Resch-Genger, Ute T1 - Comparison of methods and achievable uncertainties for the relative and absolute measurement of photoluminescence quantum yields N2 - The photoluminescence quantum yield (Φf) that presents a direct measure for the efficiency of the conversion of absorbed photons into emitted photons is one of the spectroscopic key parameters of functional fluorophores. It determines the suitability of such materials for applications in, for example, (bio)analysis, biosensing, and fluorescence imaging as well as as active components in optical devices. The reborn interest in accurate Φf measurements in conjunction with the controversial reliability of reported Φf values of many common organic dyes encouraged us to compare two relative and one absolute fluorometric method for the determination of the fluorescence quantum yields of quinine sulfate dihydrate, coumarin 153, fluorescein, rhodamine 6G, and rhodamine 101. The relative methods include the use of a chain of Φf transfer standards consisting of several 'standard dye' versus 'reference dye' pairs linked to a golden Φf standard that covers the ultraviolet and visible spectral region, and the use of different excitation wavelengths for standard and sample, respectively. Based upon these measurements and the calibration of the instruments employed, complete uncertainty budgets for the resulting Φf values are derived for each method, thereby providing evaluated standard operation procedures for Φf measurements and, simultaneously, a set of assessed Φf standards. KW - Fluorescence KW - Photoluminescence KW - Quantum yield KW - Dye KW - Uncertainty KW - Absolute quantum yield KW - Method comparison KW - Integrating sphere PY - 2011 DO - https://doi.org/10.1021/ac2000303 SN - 0003-2700 SN - 1520-6882 VL - 83 IS - 9 SP - 3431 EP - 3439 PB - American Chemical Society CY - Washington, DC AN - OPUS4-23632 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Güttler, Arne A1 - Richter, Maria A1 - Würth, Christian A1 - Resch-Genger, Ute T1 - New Reference Materials for the Quantification and Standardization of Fluorescence-based Measurements N2 - Luminescence techniques are amongst the most commonly used analytical methods in the life and material sciences due to their high sensitivity and non-destructive and multiparametric character. Photoluminescence signals are, however, affected by wavelength-, polarization-, and time-dependent instrument specific effect and the compound-specific photoluminescence quantum yield. The former hamper the comparability of fluorescence measurements, while the relative determination of the latter requires suitable quantum yield standards with well-known photoluminescence quantum yields (QY). For the simple correction of instrument specific effects in the wavelength region of 300 nm to 950 nm, the set of the five certified spectral fluorescence standards BAM-F001 – BAM-F005, has been extended to the NIR range by including two new fluorescence standards currently under certification. For the reliable and accurate determination of QY which is the key performance parameter for the comparison of different luminophores, we certified a set of 12 quantum yield standards, which absorb and emit in the wavelength range from 300 nm to 1000 nm. T2 - Methods and Applications in Fluorescence CY - Gothenburg, Sweden DA - 11.09.2022 KW - Luminescence KW - Photoluminescence KW - Fluorescence KW - Quantum yield KW - Certified reference material KW - Standard PY - 2022 AN - OPUS4-55914 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Würth, Christian A1 - Geißler, Daniel A1 - Behnke, Thomas A1 - Kaiser, Martin A1 - Resch-Genger, Ute T1 - Critical review of the determination of photoluminescence quantum yields of luminescent reporters N2 - A crucial variable for methodical performance evaluation and comparison of luminescent reporters is the photoluminescence quantum yield (Φ pl). This quantity, defined as the number of emitted photons per number of absorbed photons, is the direct measure of the efficiency of the conversion of absorbed photons into emitted light for small organic dyes, fluorescent proteins, metal–ligand complexes, metal clusters, polymeric nanoparticles, and semiconductor and up-conversion nanocrystals. Φ pl determines the sensitivity for the detection of a specific analyte from the chromophore perspective, together with its molar-absorption coefficient at the excitation wavelength. In this review we discuss different optical and photothermal methods for measuring Φ pl of transparent and scattering systems for the most common classes of luminescent reporters, and critically evaluate their potential and limitations. In addition, reporter-specific effects and sources of uncertainty are addressed. The ultimate objective is to provide users of fluorescence techniques with validated tools for the determination of Φ pl, including a series of Φ pl standards for the ultraviolet, visible, and near-infrared regions, and to enable better judgment of the reliability of literature data. KW - Fluorescence KW - Photoluminescence KW - Quantum yield KW - Organic dye KW - Nanoparticle KW - Quantum dot KW - Up-conversion nanocrystal KW - Optical probe KW - Standard KW - Quality assurance KW - Integrating sphere spectroscopy KW - Photoacoustic spectroscopy KW - Thermal lensing KW - Nanocavity PY - 2015 DO - https://doi.org/10.1007/s00216-014-8130-z SN - 1618-2642 SN - 1618-2650 VL - 407 IS - 1 SP - 59 EP - 78 PB - Springer CY - Berlin AN - OPUS4-32406 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Frenzel, F. A1 - Fiedler, S. A1 - Bardan, A. A1 - Güttler, Arne A1 - Würth, Christian A1 - Resch-Genger, Ute T1 - Influence of Measurement Geometry and Blank on Absolute Measurements of Photoluminescence Quantum Yields of Scattering Luminescent Films N2 - For a series of 500 μm-thick polyurethane films containing different concentrations of luminescent and scattering YAG:Ce microparticles, we systematically explored and quantified pitfalls of absolute measurements of photoluminescence quantum yields (Φf) for often employed integrating sphere (IS) geometries, where the sample is placed either on a sample holder at the bottom of the IS surface or mounted in the IS center. Thereby, the influence of detection and illumination geometry and sample position was examined using blanks with various scattering properties for measuring the number of photons absorbed by the sample. Our results reveal that (i) setup configurations where the scattering sample is mounted in the IS center and (ii) transparent blanks can introduce systematic errors in absolute Φf measurements. For strongly scattering, luminescent samples, this can result in either an under- or overestimation of the absorbed photon flux and hence an under- or overestimation of Φf. The size of these uncertainties depends on the scattering properties of the sample and instrument parameters, such as sample position, IS size, wavelength-dependent reflectivity of the IS surface coating, and port configuration. For accurate and reliable absolute Φf measurements, we recommend (i) a blank with scattering properties closely matching those of the sample to realize similar distributions of the diffusely scattered excitation photons within the IS, and (ii) a sufficiently high sample absorption at the excitation wavelength. For IS setups with center-mounted samples, measurement geometries should be utilized that prevent the loss of excitation photons by reflections from the sample out of the IS. KW - Quality assurance KW - Fluorescence KW - Nano KW - Particle KW - Quantum yield KW - Characterization KW - Silica KW - Scattering KW - Uncertainty KW - Film KW - Pphosphor KW - YAG:Ce KW - LED KW - Converter material KW - Solid material KW - Polymer KW - Composite material KW - Advanced material PY - 2025 UR - https://nbn-resolving.org/urn:nbn:de:kobv:b43-638304 DO - https://doi.org/10.1021/acs.analchem.4c06726 SN - 1520-6882 SP - 1 EP - 8 PB - ACS Publications AN - OPUS4-63830 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Resch-Genger, Ute A1 - Pilch, A. A1 - Würth, Christian A1 - Kaiser, Martin A1 - Wawrzynczyk, D. A1 - Kurnatowska, M. A1 - Arabasz, S. A1 - Prorok, K. A1 - Samoc, M. A1 - Strek, W. A1 - Bednarkiewicz, A. T1 - Shaping luminescent properties of Yb3+ and Ho3+ co-doped upconverting core-shell ß-NaYF4 nanoparticles by dopant distribution and spacing N2 - At the core of luminescence color and lifetime Tuning of rare earth doped upconverting nanoparticles (UCNPs), is the understanding of the Impact of the particle architecture for commonly used sensitizer (S) and activator (A) Ions. In this respect, a series of core@Shell NaYF4 UCNPs doped with Yb3+ and Ho3+ ions are presented here, where the same dopant concentrations are distributed in different particle architectures following the scheme: YbHo core and YbHo@..., ...@YbHo, Yb@Ho, Ho@Yb, YbHo@Yb, and Yb@YbHo core-Shell NPs. As refealed by quantitative steady-state and time-resolved luminescence studies, the relative spatial Distribution of the A and S ions in the UCNPs and their protection from surface quenching has a critical Impact on ther luminescence characteristics. Although the increased amount of Yb3+ Ions boosts UCNP Performance by amplifying the Absorption, the Yb3+ ions can also efficiently dissipate the energy stored in the material through energy Migration to the surface, thereby reducing the Overall energy Transfer Efficiency to the activator ions. The results provide yet another proof that UC Phosphor chemistry combined with materials Engineering through intentional core@shell structures may help to fine-tune the luminescence Features of UCNPs for their specific future applications in biosensing, bioimaging, photovoltaics, and Display technologies. KW - Fluorescence KW - Upconversion KW - NIR KW - Nonlinear KW - Nanoparticle KW - Surface chemistry KW - Quantum yield KW - Liftetime KW - Nanocrystal KW - Lanthanide KW - Ho(III) KW - Yb(III) KW - Mechanism KW - Absolute flourescence KW - Excitation power density dependence PY - 2017 DO - https://doi.org/10.1002/smll.201701635 SN - 1613-6810 VL - 13 IS - 47 SP - 1701635, 1 EP - 13 PB - WILEY-VCH Verlag GmbH & co. KGaA CY - Weinheim AN - OPUS4-43629 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Behnke, Thomas A1 - Würth, Christian A1 - Hoffmann, Katrin A1 - Hübner, Martin A1 - Panne, Ulrich A1 - Resch-Genger, Ute T1 - Encapsulation of hydrophobic dyes in polystyrene micro- and nanoparticles via swelling procedures N2 - Aiming at the derivation of a generalized procedure for the straightforward preparation of particles fluorescing in the visible and near-infrared (NIR) spectral region, different swelling procedures for the loading of the hydrophobic polarity-probe Nile Red into nano- and micrometer sized polystyrene particles were studied and compared with respect to the optical properties of the resulting particles. The effect of the amount of incorporated dye on the spectroscopic properties of the particles was investigated for differently sized beads with different surface chemistries, i.e., non-functionalized, aminomodified and PEG-grafted surfaces. Moreover, photostability and leaking studies were performed. The main criterion for the optimization of the dye loading procedures was a high and thermally and photochemically stable fluorescence output of the particles for the future application of these systems as fluorescent labels. KW - Fluorescence KW - Nile red KW - Polystyrene KW - Nanoparticles KW - Microparticles KW - Encapsulation KW - Swelling PY - 2011 DO - https://doi.org/10.1007/s10895-010-0632-2 SN - 1053-0509 SN - 1573-4994 VL - 21 IS - 3 SP - 937 EP - 944 PB - Plenum Publ. Corp. CY - New York, NY AN - OPUS4-22692 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Napp, J. A1 - Behnke, Thomas A1 - Fischer, L. A1 - Würth, Christian A1 - Wottawa, M. A1 - Katschinski, D.M. A1 - Alves, F. A1 - Resch-Genger, Ute A1 - Schäferling, M. T1 - Targeted luminescent near-infrared polymer-nanoprobes for in vivo imaging of tumor hypoxia N2 - Polystyrene nanoparticles (PS-NPs) were doped with an oxygen-sensitive near-infrared (NIR)-emissive palladium meso-tetraphenylporphyrin and an inert reference dye which are both excitable at 635 nm. The nanosensors were characterized with special emphasis on fundamental parameters such as absolute photoluminescence quantum yield and fluorescence lifetime. The PS-NPs were employed for ratiometric dual-wavelength and lifetime-based photoluminescent oxygen sensing. They were efficiently taken up by cultured murine alveolar macrophages, yielding a characteristic and reversible change in ratiometric response with decreasing oxygen concentration. This correlated with the cellular hypoxic status verified by analysis of hypoxia inducible factor-1α (HIF-1α) accumulation. In addition, the surface of PS-NPs was functionalized with polyethylene glycol (PEG) and the monoclonal antibody herceptin, and their binding to HER2/neu-overexpressing tumor cells was confirmed in vitro. First experiments with tumor-bearing mouse revealed a distinctive ratiometric response within the tumor upon hypoxic condition induced by animal sacrifice. These results demonstrate the potential of these referenced NIR nanosensors for in vitro and in vivo imaging that present a new generation of optical probes for oncology. KW - Fluorescence KW - NIR KW - Nanoparticles KW - Microparticles KW - Imaging KW - Cancer KW - Oxygene PY - 2011 DO - https://doi.org/10.1021/ac201870b SN - 0003-2700 SN - 1520-6882 VL - 83 IS - 23 SP - 9039 EP - 9046 PB - American Chemical Society CY - Washington, DC AN - OPUS4-24987 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Resch-Genger, Ute A1 - Hoffmann, Katrin A1 - Würth, Christian A1 - Behnke, Thomas A1 - Hoffmann, Angelika A1 - Pfeifer, Dietmar A1 - Engel, A. T1 - The toolbox of fluorescence standards: Flexible calibration tools for the standardization of fluorescence-based measurements N2 - To improve the reliability of fluorescence data in the life and material sciences and to enable accreditation of fluorescence techniques, standardization concepts are required that guarantee and improve the comparability of fluorescence measurements. At the core of such concepts are simple and evaluated fluorescence standards for the consideration of instrument-specific spectral and intensity distortions of measured signals and for instrument performance validation (IPV). Similarly in need are fluorescence intensity standards for the quantification from measured intensities and for signal referencing, thereby accounting for excitation light-induced intensity fluctuations. These standards should be preferably certified, especially for use in regulated areas like medical diagnostics. This encouraged us to develop liquid and solid standards for different fluorescence parameters and techniques for use under routine measurement conditions in different formates. Special emphasis was dedicated to the determination and control of the spectral responsivity of detection systems, wavelength accuracy, homogeneity of illumination, and intensity referencing for e.g. spectrofluorometers, fluorescence sensors and confocal laser scanning fluorescence microscopes. Here, we will present design concepts and examples for mono- and multifunctional fluorescence standards that provide traceability to radiometric units and present a first step towards a toolbox of standards. KW - Fluorescence KW - Fluorescence standard KW - Calibration tool KW - Spectral fluorescence standard KW - Intensity standard KW - Instrument performance validation KW - Quality assurance KW - Traceability KW - Glass KW - Liquid standard PY - 2010 DO - https://doi.org/10.1117/12.853133 SN - 0277-786X SN - 0038-7355 SN - 0361-0748 VL - 7666 IS - 76661J SP - 1 EP - 12 PB - Soc. CY - Redondo Beach, Calif. [u.a.] AN - OPUS4-21595 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Pauli, Jutta A1 - Hoffmann, Katrin A1 - Würth, Christian A1 - Behnke, Thomas A1 - Resch-Genger, Ute T1 - Standardization of fluorescence measurements in the UV/vis/NIR/IR N2 - Photoluminescence techniques are amongst the most widely used Tools in the life sciences, with new and exciting applications in medical diagnostics and molecular Imaging continuously emerging. Advantages include their comparative ease of use, unique sensitivity, non-invasive character, and potential for Multiplexing, remote sensing, and miniaturization. General drawbacks are, however, signals, that contain unwanted wavelength- and polarization contributions from Instrument-dependent effects, which are also time-dependent due to aging of Instrument-components, and difficulties to measure absolute flourescence entensities. Moreover, scattering Systems require Special measurement geometries and the interest in new optical Reporters with Emission > 1000 nm strategies for reliable measurements in the second diagnostic for the comparison of material Performance and the rational designg of new flourophores with improved properties. Here, we present strategies to versatile method-adaptable liquid and solid flourescence Standards for different flourescence paramters including traceable Instrument calibration procedures and the design of integrating spere setups for the absolute measurements of emission spectra and Quantum yields in the wavelength Region of 350 to 1600 nm. Examples are multi-Emitter glasses, spectral flourescence Standards, and quantum yield Standards for the UV/vis/NIR. T2 - Conference on Molecular-Guided Surgery - Molecules, Devices, and Applications III CY - San Francisco, CA, USA DA - 28.01.2017 KW - Fluorescence KW - Reference material KW - Standard KW - Calibration KW - Nanoparticle KW - Absolute flourometry KW - Integrating sphere spectroscopy KW - NIR KW - IR KW - Quantum yield standard KW - Emission standards PY - 2017 SN - 978-1-5106-0539-8 DO - https://doi.org/10.1117/12.2255728 SN - 0277-786X VL - 10049 SP - 1 PB - Proceedings of SPIE AN - OPUS4-41783 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Behnke, Thomas A1 - Mathejczyk, J.E. A1 - Brehm, Robert A1 - Würth, Christian A1 - Gomes, F.R. A1 - Dullin, C. A1 - Napp, J. A1 - Alves, F. A1 - Resch-Genger, Ute T1 - Target-specific nanoparticles containing a broad band emissive NIR dye for the sensitive detection and characterization of tumor development N2 - Current optical probes including engineered nanoparticles (NPs) are constructed from near infrared (NIR)-emissive organic dyes with narrow absorption and emission bands and small Stokes shifts prone to aggregation-induced self-quenching. Here, we present the new asymmetric cyanine Itrybe with broad, almost environment-insensitive absorption and emission bands in the diagnostic window, offering a unique flexibility of the choice of excitation and detection wavelengths compared to common NIR dyes. This strongly emissive dye was spectroscopically studied in different solvents and encapsulated into differently sized (15, 25, 100 nm) amino-modified polystyrene NPs (PSNPs) via a one-step staining procedure. As proof-of-concept for its potential for pre-/clinical imaging applications, Itrybe-loaded NPs were surface-functionalized with polyethylene glycol (PEG) and the tumor-targeting antibody Herceptin and their binding specificity to the tumor-specific biomarker HER2 was systematically assessed. Itrybe-loaded NPs display strong fluorescence signals in vitro and in vivo and Herceptin-conjugated NPs bind specifically to HER2 as demonstrated in immunoassays as well as on tumor cells and sections from mouse tumor xenografts in vitro. This demonstrates that our design strategy exploiting broad band-absorbing and -emitting dyes yields versatile and bright NIR probes with a high potential for e.g. the sensitive detection and characterization of tumor development and progression. KW - Nanoparticle KW - Fluorescence KW - In vitro test KW - In vivo test KW - Surface modification KW - Cytotoxicity PY - 2013 DO - https://doi.org/10.1016/j.biomaterials.2012.09.028 SN - 0142-9612 VL - 34 IS - 1 SP - 160 EP - 170 PB - Elsevier CY - Oxford AN - OPUS4-26877 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Pauli, Jutta A1 - Würth, Christian A1 - Güttler, Arne A1 - Resch-Genger, Ute T1 - Reliable Determination of the Signal-Relevant Spectroscopic Key Characteristics of Luminescent Reporters and Optical Probes for Imaging in the vis/NIR/SWIR N2 - Introduction. Comparing different emitter classes and rationally designing the next generation of molecular and nanoscale probes for bioimaging applications require accurate and quantitative methods for the measurement of the key parameter photoluminescence quantum yield f.1 f equals the number of emitted per number of absorbed photons. This is particularly relevant for increasingly used fluorescence imaging in the short wave-infrared region (SWIR) ≥ 900 nm providing deeper penetration depths, a better image resolution, and an improved signal-to-noise or tumor-to-background ratio.2, 3 However, spectroscopic measurements in the SWIR are more challenging and require specific calibrations and standards. T2 - EMIM 2021 CY - Göttingen, Germany DA - 24.08.2021 KW - Fluorescence KW - Optical probe KW - Dye KW - Photophysics KW - Quantum yield KW - Mechanism KW - NIR KW - SWIR KW - Imaging KW - Reference material KW - Reliability KW - Nano KW - Particle KW - Method KW - Quality assurance PY - 2021 AN - OPUS4-53233 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Güttler, Arne A1 - Richter, Maria A1 - Würth, Christian A1 - Resch-Genger, Ute T1 - New Reference Materials for Quantification and Standardization of Fluorescence-based Measurements N2 - Luminescence techniques are amongst the most commonly used analytical methods in the life and material sciences due to their high sensitivity and non-destructive and multiparametric character. Photoluminescence signals are, however, affected by wavelength-, polarization-, and time-dependent instrument specific effects and the compound-specific photoluminescence quantum yield. The former hamper the comparability of fluorescence measurements, while the relative determination of the latter requires suitable quantum yield standards with well-known photoluminescence quantum yields. For the simple correction of instrument specific effects in the wavelength region of 300 nm to 950 nm, the set of the five certified spectral fluorescence standards (CRM) BAM-F001 – BAM-F005, has been extended to the NIR range by including two new fluorescence standards currently under certification. T2 - 17th conference on Methods and Applications in Fluorescence CY - Gothenburg, Sweden DA - 11.09.2022 KW - Luminescence KW - Photoluminescence KW - Fluorescence KW - Quantum yield KW - Certified reference material KW - Standard PY - 2022 AN - OPUS4-56441 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Huang, K. A1 - Liu, H. A1 - Kraft, Marco A1 - Shikha, S. A1 - Zheng, X. A1 - Agren, H. A1 - Würth, Christian A1 - Resch-Genger, Ute A1 - Zhang, Y. T1 - A protected excitation-energy reservoir for efficient upconversion luminescence N2 - Lanthanide-doped upconversion nanoparticles (UCNPs) are of great interest for biomedical applications. Currently, the applicability of UCNP bionanotechnology is hampered by the generally low luminescence intensity of UCNPs and inefficient energy Transfer from UCNPs to surface-bound chromophores used e.g. for photodynamic therapy or analyte sensing. In this work, we address the low-Efficiency issue by developing versatile core-Shell nanostructures, where high-concentration sensitizers and activators are confined in the core and Shell Region of representative hexagonal NaYF2:Yb,Er UCNPs. After Doping concentration optimization, the sensitizer-rich core is able to harvest/accumulate more excitation energy and generate almost one order of Magnitude higher luminescence intesity than conventional homogeneously doped nanostructures. At the same time, the activator Ions located in the Shell enable a ~6 times more efficient resonant energy Transfer from UCNPs to surface-bound acceptor dye molecules due to the short distance between donor-acceptor pairs. Our work provides new insights into the rational design of UCNPs and will greatly encrease the General applicability of upconversion nanotechnologies. KW - Fluorescence KW - Lanthanide KW - Upconversion KW - Brightness KW - Quantification KW - Nanoparticle KW - Absolute fluorometry KW - NIR KW - IR KW - Quantum yield KW - Integrating sphere spectroscopy KW - Method KW - Energy transfer KW - Shell KW - Particle architecture PY - 2017 DO - https://doi.org/10.1039/c7nr06900f SN - 2040-3372 SN - 2040-3364 VL - 10 IS - 1 SP - 250 EP - 259 PB - The Royal Society of Chemistry AN - OPUS4-43893 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Pauli, J. A1 - Güttler, Arne A1 - Schneider, T. A1 - Würth, Christian A1 - Resch-Genger, Ute T1 - Fluorescence Quantum Yield Standards for the UV/Visible/NIR: Development, Traceable Characterization, and Certification N2 - The rational design of next generation molecular and nanoscale reporters and the comparison of different emitter classes require the determination of the fluorometric key performance parameter fluorescence quantum yield (Φf), i.e., the number of emitted photons per number of absorbed photons. Main prerequisites for reliable Φf measurements, which are for transparent luminophore solutions commonly done relative to a reference, i.e., a fluorescence quantum yield standard of known Φf, are reliable and validated instrument calibration procedures to consider wavelength-, polarization-, and time-dependent instrument specific signal contributions, and sufficiently well characterized fluorescence quantum yield standards. As the standard’s Φf value directly contributes to the calculation of the sample’s Φf, its accuracy presents one of the main sources of uncertainty of relative Φf measurements. To close this gap, we developed a first set of 12 fluorescence quantum yield standards, which absorb and emit in the wavelength region of 330−1000 nm and absolutely determined their Φf values with two independently calibrated integrating sphere setups. Criteria for standard selection and the configuration of these novel fluorescence reference materials are given, and the certification procedure is presented including homogeneity and stability studies and the calculation of complete uncertainty budgets for the certified Φf values. The ultimate goal is to provide the community of fluorescence users with available reference materials as a basis for an improved comparability and reliability of quantum yield data since the measurement of this spectroscopic key property is an essential part of the characterization of any new emitter. KW - Optical spectroscopy KW - Traceability KW - Reference product KW - Dye KW - Fluorescence KW - Quantum yield KW - Reference material KW - Reference data KW - Certification KW - Quality assurance PY - 2023 DO - https://doi.org/10.1021/acs.analchem.2c05530 VL - 95 SP - 5671 EP - 5677 PB - American Chemical Society AN - OPUS4-58151 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Würth, Christian A1 - Resch-Genger, Ute T1 - Determination of photoluminescence quantum yields of scattering media with an integrating sphere: direct and indirect illumination N2 - The ever-increasing use of fluorescent nanomaterials and micrometer-sized beads in the life and material sciences requires reliable procedures for the measurement of the key performance parameter fluorescence quantum yield (Ff) of scattering particle dispersions and reference systems to evaluate the performance of such measurements. This encouraged us to systematically study, both theoretically and experimentally, the optical determination of photoluminescent quantum yield as a function of the scattering and absorption properties of the sample and the illumination geometry with an integrating sphere method. The latter included measurements with a direct and an indirect illumination. As a representative and easy-to-prepare reference system, we used ethanolic dispersions of 250 nm sized silica particles and the dye rhodamine 101 and systematically varied the concentration of the dye and particles within the typical ranges of spectroscopic and (bio)analytical applications of fluorescent nanomaterials. Based on our measurements, we recommend indirect sample illumination geometry for the accurate measurement of Ff of samples with low or unknown absorption and high scattering coefficients such as dispersions of luminescent particles or fluorescent reporters in biological matrices. This finding is of utmost relevance for all (bio)analytical applications of fluorescent nanomaterials ranging from particle labels and probes over assay platforms to safety barcodes. KW - Absolute quantum yield KW - Fluorescence KW - Integrating sphere KW - Particles KW - Photoluminescence KW - Rhodamine 101 KW - Scattering media PY - 2015 DO - https://doi.org/10.1366/14-07679 SN - 0003-7028 SN - 1943-3530 VL - 69 IS - 6 SP - 749 EP - 759 PB - Society for Applied Spectroscopy CY - Frederick, Md. AN - OPUS4-33819 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Behnke, Thomas A1 - Würth, Christian A1 - Laux, Eva-Maria A1 - Hoffmann, Katrin A1 - Resch-Genger, Ute T1 - Simple strategies towards bright polymer particles via one-step staining procedures N2 - In order to develop simple and versatile procedures for the preparation of red emissive particles, various one-step swelling procedures for the loading of fluorophores into nanometer- and micrometer-sized polystyrene particles were systematically assessed. Parameters studied for model dyes from common dye classes include the composition of the swelling medium, dye charge and polarity, dye concentration, and particle surface chemistry. The dye loading procedures were compared based upon the efficiency of dye incorporation, fluorescence intensity, and colloidal stability of the resulting particles as well as the absence of dye leaking as determined by absorption and fluorescence spectroscopy, flow cytometry, and measurements of zeta potentials. In addition, for the first time, the influence of the amount of incorporated dye on the absolute fluorescence quantum yield and brightness of the fluorescent particles was investigated for selected chromophores in differently sized particles using a custom-made calibrated integrating sphere setup. Our results demonstrate the general suitability of these one-step loading procedures for efficient particle staining with neutral, zwitterionic, and charged fluorophores like oxazines, coumarines, squaraines, xanthenes, and cyanines emitting in the visible and near infrared. Dye polarity was identified as a suitable tool to estimate the loading efficiency of fluorophores into these polymer particles. KW - Fluorescence KW - Polystyrene KW - Particles KW - Encapsulation KW - Quantum yield KW - Zeta potential KW - Method KW - Label KW - Particle KW - Polymer KW - Absolute fluorescence quantum yield KW - Fluorophore KW - Dye content KW - Surface groups KW - Size KW - Brightness PY - 2012 DO - https://doi.org/10.1016/j.dyepig.2012.01.021 SN - 0143-7208 SN - 1873-3743 VL - 94 IS - 2 SP - 247 EP - 257 PB - Elsevier Ltd. CY - Kidlington AN - OPUS4-25481 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -