TY - JOUR A1 - Wildgruber, M. A1 - Lueg, C. A1 - Borgmeyer, S. A1 - Karimov, I. A1 - Braun, Ulrike A1 - Kiechle, M. A1 - Meier, R. A1 - Koehler, M. A1 - Ettl, J. A1 - Berger, H. T1 - Polyurethane versus silicone catheters for central venous port devices implanted at the forearm N2 - Purpose: We aimed to analyse short and long-term complications of polyurethane (PU) versus silicone catheters used in totally implantable venous-access ports (TIVAPs) implanted at the forearm. Methods: Retrospective analysis of 698 consecutively implanted TIVAPs was performed. Primary end-points were defined as rates of major complications associated with either type of central venous port catheter. Technical success rate, device service interval as well as minor complications not requiring port explantation were defined as secondary end-points. Results: A total of 698 port devices were implanted in 681 patients, 396 equipped with a PU catheter, 302 with a silicone catheter. The technical success rate was 99.9% with no major periprocedural complications. During follow-up a total of 211 complications in 146 patients were observed (1.0/1000 catheter days), 183 occurred associated with PU catheters (1.8/100 catheter days), 28 (0.3/1000 catheter days) with silicone catheters (log rank test p < 0.0001). Catheter-related bloodstream infections as well as thrombotic complications occurred significantly more frequently with PU catheters, while silicone catheters exhibited a trend towards a higher rate of mechanical failure such as disconnection or catheter rupture. Major complications requiring explantation of the device occurred more frequently with PU-based catheters (10.6%) compared to silicone catheter carrying ports (4.6%, log rank test p < 0.001). KW - Chemotherapy KW - Central venous access port KW - Complication KW - Thrombosis KW - Infection PY - 2016 U6 - https://doi.org/10.1016/j.ejca.2016.02.011 SN - 0959-8049 VL - 59 SP - 113 EP - 124 PB - Elsevier Ltd. CY - Oxford, UK AN - OPUS4-36288 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Paul, Andrea A1 - Breitfeld, S. A1 - Sauer, Andreas A1 - Becker, Roland A1 - Kalbe, Ute A1 - Bremser, Wolfram A1 - Braun, Ulrike T1 - A Raman spectroscopic method for the monitoring of microplastics in soil N2 - Due to the increasing presence of microplastic particles (MP) in the Environment and the unknown risks arising from them, there is an urgent need for analytical methods that allow for an efficient identification and quantification of microplastics (MP), i.e. particles < 5 mm, in environmental samples. So far, mostly timeconsuming (micro) infrared or micro-Raman spectroscopic methods are applied. Here an faster alternative Approach is presented based on a Raman processspectrometer with fiber-optical probes in combination with multivariate data analysis. T2 - 12. Kolloquium Prozessanalytik CY - Berlin, Germany DA - 28.11.2016 KW - Raman spectroscopy KW - Microplastics KW - Soil KW - Chemometrics KW - PLS-DA PY - 2016 SP - P10, 59 EP - 60 AN - OPUS4-38392 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Koshkina, Olga A1 - Westmeier, D. A1 - Lang, Thomas A1 - Bantz, C. A1 - Hahlbrock, A. A1 - Würth, Christian A1 - Resch-Genger, Ute A1 - Braun, Ulrike A1 - Thiermann, Raphael A1 - Weise, C. A1 - Eravci, M. A1 - Mohr, B. A1 - Schlaad, H. A1 - Stauber, R. H. A1 - Docter, D. A1 - Bertin, Annabelle A1 - Maskos, M. T1 - Tuning the surface of nanoparticles: Impact of poly(2-ethyl-2-oxazoline) on protein adsorption in serum and cellular uptake N2 - Due to the adsorption of biomolecules, the control of the biodistribution of nanoparticles is still one of the major challenges of nanomedicine. Poly(2-ethyl-2-oxazoline) (PEtOx) for surface modification of nanoparticles is applied and both protein adsorption and cellular uptake of PEtOxylated nanoparticles versus nanoparticles coated with poly(ethylene glycol) (PEG) and non-coated positively and negatively charged nanoparticles are compared. Therefore, fluorescent poly(organosiloxane) nanoparticles of 15 nm radius are synthesized, which are used as a scaffold for surface modification in a grafting onto approach. With multi-angle dynamic light scattering, asymmetrical flow field-flow fractionation, gel electrophoresis, and liquid chromatography-mass spectrometry, it is demonstrated that protein adsorption on PEtOxylated nanoparticles is extremely low, similar as on PEGylated nanoparticles. Moreover, quantitative microscopy reveals that PEtOxylation significantly reduces the non-specific cellular uptake, particularly by macrophage-like cells. Collectively, studies demonstrate that PEtOx is a very effective alternative to PEG for stealth modification of the surface of nanoparticles. KW - Poloxazolines KW - Protein corona KW - Cellular uptake PY - 2016 U6 - https://doi.org/10.1002/mabi.201600074 SN - 1616-5187 SN - 1616-5195 VL - 16 IS - 9 SP - 1287 EP - 1300 AN - OPUS4-37369 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Jaunich, Matthias A1 - Böhning, Martin A1 - Braun, Ulrike A1 - Teteris, G. A1 - Stark, W. T1 - Investigation of the curing state of ethylene/vinyl acetate copolymer (EVA) for photovoltaic applications by gel content determination, rheology, DSC and FTIR N2 - The cure behaviour of a specific ethylene vinyl acetate material as used for encapsulation of photovoltaic modules was analysed by rheometer, differential scanning calorimetry and Fourier transform infrared spectroscopy to test for a suitable replacement for the laborious determination of gel content. The results show that all applied methods are capable of describing the effects of the cross-linking process. Some provide results very similar to those yielded by analysis of the insoluble content, but the question remains as to whether indirect methods should be preferred over the direct measurement of physical properties, e.g. as performed by the curemeter. A material stored for one year was also tested to demonstrate the effect of extended storage on cure behaviour and how this is detected by different methods. This complements the other methods, which were clearly able to detect the different cure behaviour of the aged EVA, whereas determination of the gel content could not. KW - Photovoltaic cell encapsulation KW - EVA KW - Incoming goods control KW - Insoluble content KW - Rheometer PY - 2016 U6 - https://doi.org/10.1016/j.polymertesting.2016.03.013 SN - 0142-9418 VL - 52 IS - 1 SP - 133 EP - 140 PB - Elsevier Ltd. CY - Oxford, UK AN - OPUS4-35918 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Deußing, G. A1 - Dümichen, Erik A1 - Braun, Ulrike T1 - Polymeranalytik Intelligent kombiniert N2 - Durch die intelligente Anbindung an die Stir Bar Sorptive Extraction (SBSE) lassen sich Zersetzungsprodukte der thermogravimetrischen Analyse (TGA) von Polymeren auf effiziente Weise anreichern, auftrennen und mittels GC/MS identifizieren. KW - Thermische Degradation KW - Mikroplastik PY - 2016 UR - http://www.gerstel.cn/de/GA51_Polymeranalytik.htm IS - 51 SP - 9 EP - 11 PB - Gerstel AN - OPUS4-42146 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - JOUR A1 - Braun, Ulrike A1 - Lorenz, Edelgard A1 - Weimann, Christiane A1 - Sturm, Heinz A1 - Karimov, I. A1 - Ettl, J. A1 - Meier, R. A1 - Wohlgemuth, W. A. A1 - Berger, H. A1 - Wildgruber, M. T1 - Mechanic and surface properties of central-venous port catheters after removal: A comparison of polyurethane and Silicon rubber materials N2 - Central venous port devices made of two different polymeric materials, thermoplastic polyurethane (TPU)and silicone rubber (SiR), were compared due their material properties. Both naïve catheters as well as catheters after removal from patients were investigated. In lab experiments the influence of various chemo-therapeutic solutions on material properties was investigated, where as the samples after removal were compared according to the implanted time inpatient. The macroscopic,mechanical performance was assessed with dynamic, specially adapted tests for elasticity. The degradation status of the materials was determined with common tools of polymer characterisation, such as infrared spectroscopy, molecular weight measurements and various methods of thermal analysis. The surface morphology was an alysed using scanning electron microscopy. A correlation between material properties and clinical performance was proposed. The surface morphology and chemical composition of the polyurethane catheter materials can potentially result in increased susceptibility of the catheter to bloodstream infections and thrombotic complications. The higher mechanic failure,especially with increasing implantation time of the silicone catheters is related to the lower mechanical performance compared to the polyurethane material as well as loss of barium sulphate filler particles near the surface of the catheter. This results in preformed microscopic notches, which act as predetermined sites of fracture. KW - Thermoplastic polyurethane (TPU) KW - Silicone rubber (SiR) KW - Catheters KW - Central venous access port KW - Complication KW - Structure propertyrelationship KW - Mechanical testing PY - 2016 U6 - https://doi.org/10.1016/j.jmbbm.2016.08.002 SN - 1751-6161 SN - 1878-0180 VL - 64 SP - 281 EP - 291 PB - Elsevier Ltd. AN - OPUS4-37178 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike A1 - Dümichen, Erik A1 - Barthel, Anne-Kathrin A1 - Bannick, Claus Gerhard T1 - Charakterisierung von Mikroplastik Partikeln aus Umweltproben N2 - Durch den stetig steigenden Einsatz von Plastikprodukten sind unerwünschte Kunststofffunde in verschiedenen Umweltmatrices mittlerweile allgegenwärtig. Besonderes Interesse erfahren die Kunststoffpartikel von einer Größe kleiner 5 mm, sogenannte Mikroplastik (MP) Partikel. Es werden aktuell Risiken für Mensch und Umwelt diskutiert, die von diesen Partikeln ausgehen. T2 - Wasser 2016 - Jahrestagung der Wasserchemischen Gesellschaft CY - Bamberg, Deutschland DA - 02.05.2016 KW - Mikroplastik KW - Polymere KW - Polyethylen (PE) KW - Polystyrol (PS) KW - Poly(ethylenterephthalat) (PET) PY - 2016 SN - 978-3-936028-94-2 SP - 74 EP - 76 PB - Wasserchemische Gesellschaft CY - Mülheim an der Ruhr, Germany AN - OPUS4-36782 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike A1 - Dümichen, Erik A1 - Eisentraut, Paul A1 - Barthel, Anne-Kathrin A1 - Bannick, Claus Gerhard T1 - Mikroplastik: Ein Material oder ein Problem? N2 - Durch den stetig steigenden Einsatz von Plastikprodukten sind unerwünschte Kunststofffunde in verschiedenen Umweltmatrices mittlerweile allgegenwärtig. Bereits seit Jahren werden die Funde in aquatischen Bereich kommentiert, mittlerweile erfahren auch terrestrische Kompartimente erhöhte Aufmerksamkeit. Besonderes Interesse erfahren Kunststoffpartikel von einer Größe kleiner 5 mm, sogenannte Mikroplastik (MP) Partikel. T2 - Kunststofftechnisches Kolloquium CY - Chemnitz, Germany DA - 10.05.2016 KW - Mikroplastik KW - Polymere PY - 2016 AN - OPUS4-36757 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike A1 - Dümichen, Erik A1 - Becker, Roland A1 - Barthel, Anne-Kathrin A1 - Bannick, Claus Gerhard T1 - General aspects of polymers and and polymers in environment N2 - The presentation addresses first general aspects of polymers and polymers in environment (definitions, types, phases, morphology, density) and than summarise the possibilities of analysis. T2 - 1. meeting of ISO/TC 61/SC 5/AHG "microplastics" CY - Berlin, Germany DA - 12.04.2016 KW - Microplastics KW - Polymers PY - 2016 AN - OPUS4-36758 LA - eng AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER - TY - CONF A1 - Braun, Ulrike A1 - Dümichen, Erik A1 - Barthel, Anne-Kathrin A1 - Bannick, Claus Gerhard T1 - Charakterisierung von Mikroplastik Partikeln aus Umweltproben N2 - Durch den stetig steigenden Einsatz von Plastikprodukten sind unerwünschte Kunststofffunde in verschiedenen Umweltmatrices mittlerweile allgegenwärtig. Besonderes Interesse erfahren die Kunststoffpartikel von einer Größe kleiner 5 mm, sogenannte Mikroplastik (MP) Partikel und deren ausgehende Risiken auf Mensch und Umwelt T2 - Wasser 2016 - Jahrestagung der Wasserchemischen Gesellschaft CY - Bamberg, Germany DA - 02.05.2016 KW - Mikroplastik KW - Polymere PY - 2016 AN - OPUS4-36785 LA - deu AD - Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin, Germany ER -